Sparco™ Aquamix™
(Old Style AM Series Models)
Replacement Element
INSTALLATION INSTRUCTIONS
GENERAL
Hard water conditions may result in scale deposits causing
binding of internal parts in extreme cases. Cleaning the
internal parts will usually restore the valve to proper operating
conditions. In some cases it may be necessary to replace the
lower assembly.
SCREW
LOWER NUT
CH
MIX
Fig. 1. Exploded View
LABEL
HANDWHEEL
UPPER NUT
HANDWHEEL
STOP
TOP ASSEMBLY
LOWER ASSEMBLY
DIFFUSER
SPRING
VALVE BODY
M25028
INSTALLATION
To clean and/or replace the lower assembly,
shut off water and:
1. Pierce center circle of nameplate with a screwdriver and
remove screw and hand wheel.
2. Loosen upper nut (do not remove) to allow engaging an
adjustable wrench on lower nut. Unscrew lower nut
(counterclockwise). This removes top assembly.
3. Brass top assembly will pop up. Remove lower
assembly, diffuser and spring.
4. Carefully remove any scaling (calcium deposits) or
foreign particles from valve seat and other internal parts.
Use vinegar to remove calcium. Soak parts until calcium
becomes soft and can be scrubbed and washed off. Do
not use solvents or scratch metallic/Teflon
surfaces.
5. Replace cleaned spring, diffuser and lower assembly
following instructions below or use new replacement kit
assembly. For correct kit number, refer to table on
page 2.
a. Some older models have a brass washer between
the spring and body. Remove this washer, as it is no
longer needed with the latest replacement kits.
b. Insert spring with larger diameter on the bottom into
the body.
c. Fit valve top assembly into lower assembly and
insert into valve.
d. Tighten lower nut. Line up U slot with raised dot or
arrow on body and hold while tightening upper nut.
e. Turn plastic star wheel clockwise until it stops, then
back as shown in table.
f. Place hand wheel on position 1 over raised arrow
and insert screw. turn hand wheel to desired
temperature setting.
®
coated
62-3105—01
SPARCO™ AQUAMIX™ (OLD STYLE AM SERIES MODELS) REPLACEMENT ELEMENT
Calibration of Sparco™ Aquamix™ Valves when testing apparatus is not available. (Field Calibration)
With handwheel removed, turn stem clockwise as far
Lower Assembly
Model
70° - 100 ° F (21° - 38° C) (“B”) AM 100-009 RP 1/4 turn
90° - 120° F (32° - 49° C) (“C”) AM 100-001 RP 3/8 turn
110° - 145° F (43° - 63° C) AM 100-002 RP 5/8 turn
110° - 145° F (43° - 63° C) (“H”) AM 100-003 RP 1/4 turn
80° - 180° F (27° - 82° C) (“R”) AM 100-010 RP
110° - 145° F (43° - 63° C) (“J”) AM 100-011 RP
Field calibration is less accurate than calibration performed on the testing apparatus.
It is important that the element replacement is done
exactly as indicated above to prevent malfunction,
temperature range inaccuracies or possible scalding.
Please consult the factory if you need further
assistance.
Replacement Kit
as possible. Then turn counterclockwise by the
amount shown below and install handwheel so that #1
is in-line with red arrow on body.
Aquamix™ is trademark of Wattts Cazzaniga, S.p.A.
Sparco™ is trademark of Honeywell International Inc.
®
Tefl on
is a registered trademark of E.I. Du Pont De Nemours and Company.
Automation and Control Solutions
Honeywell International Inc. Honeywell Limited-Honeywell Limitée
1985 Douglas Drive North 35 Dynamic Drive
Golden Valley, MN 55422 Toronto, Ontario M1V 4Z9
customer.honeywell.com
® U.S. Registered Trademark
© 2007 Honeywell International Inc.
62-3105—01 J.I. 09-07