L6081A
Aquastat® Controller
INSTALLATION INSTRUCTIONS
APPLICATION
The L6081A is a triple function controller; it is used as a
high limit, low limit, and circulator control. This controller
is designed to replace the controls listed in Table 1.
Table 1. Controls L6081A Replaces.
Honeywell General Controls
L401A
L4010B
L4062A
L4062B
L4081A,B
L6062A
L6081A,C
a
All require a 124904 Well Adapter to fit L6081A case to well (order separately, see form 68-0040, Wells and Fittings
for Temperature Controllers).
b
Well for L6081A must be added when replacing these models.
L52F
L52G
L52H
L52J
L52K
L52L
a
480
DA-36-153
INSTALLATION
When Installing This Product…
1. Read these instructions carefully. Failure to follow
them could damage the product or cause a
hazardous condition.
2. Check the ratings given in the instructions and on
the product to make sure the product is suitable for
your application.
3. Installer must be a trained, experienced service
technician.
4. After installation is complete, check out product
operation as provided in these instructions.
WARNING
Explosion Hazard.
Can cause serious injury, death or property
damage.
Use this product only in systems with pressure
relief valves.
When replacing some of these controls, only 1 or 2 of the
L6081A switching functions may be needed. Refer to Fig.
2 to determine which terminals are used in a given
application.
Mercoid
a
443AT
447G 4432X
4420 4433
4420X 4433X
4422 4436
4422X 4437
Penn
b
4432
a
White Rodgers
11A61
11C61
b
1131
1135
11C30
1115
11C31
WARNING
Electrical Shock Hazard.
Can cause serious injury or death.
Disconnect power supply before installation to
prevent electrical shock or equipment damage.
NOTE: If new immersion well is needed, refer to form
68-0040 Wells and Fittings for Temperature
Controllers for ordering information.
Mounting
Follow the mounting system manufacturer instructions if
available; otherwise, proceed as follows:
1. Turn off power and drain the system if necessary.
2. Remove old control.
3. loosen the captive cover mounting screw on the
L6081A and pull off the cover.
b
a
® U.S. Registered Trademark
Copyright © 2001 Honeywell • All Rights Reserved
95- 6701- 5
L6081A AQUASTAT® CONTROLLER
IMPORTANT:
Best thermal response is obtained with a well
that snugly fits the sensing bulb.
If well does not snugly fit bulb, use the
heat-conductive grease. Fold the plastic bag of
grease lengthwise and twist it gently. Then snip
off end of bag and work the open end all the
way into the well. Slowly pull out bag while
squeezing it firmly to distribute grease evenly in
the well. Insert bulb all the way to the bottom of
the well. Bend the tubing, if necessary, to hold
the bulb against the bottom of the well and to
hold the outer end of the bulb in firmly to the
side of the well (Fig. 1). Wipe excess grease off
outer end of the well.
CONTROLLER
CASE
IMMERSION
WELL CLAMP
IMMERSION
WELL SPUD
ADAPTER
(C)
2
CAPILLARY
TUBE
SETSCREW
IMMERSION
WELL CLAMP SCREW
1
BEND THE CAPILLARY TUBE TO HOLD THE SENSING BULB IN GOOD THERMAL CONTACT WITH THE IMMERSION WELL AT POINTS (A) AND (B).
2
ASSURE THAT CAPILLARY TUBE FITS FREELY IN THE ADAPTER SO THE TENSION OF THE CAPILLARY TUBE AT POINT (C)
HOLDS THE SENSING BULB IN GOOD THERMAL CONTACT WITH THE IMMERSION WELL AT POINT (D).
BOILER
1
SHORT TUBE
FITS IN CENTRAL
RECESS OF ADAPTER
Fig. 1. Use of adapter and proper position of sensing bulb in immersion well.
6. If the immersion well from the old control does not
fit into the clamp on the back of L6081A case, use
the 124904 Well Adapter (order separately, see
form 68-0040, Wells and Fittings for Temperature
Controllers).
a. Fit sensing element capillary into the adapter
groove (Fig. 1).
b. Fasten the adapter to the case. Make certain
that the clamp is properly positioned on the
adapter and the adapter flange fits snugly into
the opening in the case.
c. Apply heat-conductive grease as instructed
above.
d. Insert the bulb and adapter into the well. The
sensing element bulb must bottom in the well
for best temperature response. (A slight bend
in the tubing will hold the bulb against the well.)
See Fig. 1.
e. Tighten the setscrew. Make certain that the
adapter groove is on the opposite side of the
setscrew (Fig. 1).
4. Loosen the insertion well clamp screw (Fig. 1).
5. If the immersion well from the old control fits into
the clamp on the back of the L6081 case, proceed
as follows. If it does not, go to step 6 below.
a. Apply heat-conductive grease as instructed
above.
b. Insert the sensing element into the well.
c. Fasten the case to the well with the clamp
screw. Make certain that the clamp is properly
positioned on the well and the well flange fits
snugly into the opening in the case. The
sensing element bulb must bottom in the well
for best temperature response. (A slight bend
in the tubing holds the bulb against the well.)
See Fig. 1.
(A)
BACK OF
CONTROLLER
CASE
OLD IMMERSION
WELL ASSEMBLY
IMMERSION
WELL CLAMP
SCREW
HEAT-CONDUCTIVE COMPOUND
ADAPTER
IMMERSION
WELL CLAMP
SENSING
BULB
(B)
M8830
(D)
WIRING
WARNING
Electrical Shock Hazard.
Can cause serious injury, death or property
damage.
Disconnect power supply to prevent electrical
shock or equipment damage.
All wiring must comply with local electrical codes. Make
certain that the electrical requirement of the controlled
equipment does not exceed the rating shown inside the
L6081A cover. Follow the hookup instructions furnished
by the system manufacturer.
IMPORTANT:
L6081A terminals are approved for copper wire
only.
Use Fig. 2 to determine which switches to use when
replacing the existing control. For example, if the existing
control has high limit and circulator control switch
terminals, use terminals R and B on the left side of the
switch for the high limit and the R and W terminals on the
right side of the switch to control the circulator.
95-6701—5 2