Honeywell DPTM52, DPTM112, DPTM552, DPTM1102, DPTM102 User Manual

...
DPTM52-5002
2-WIRE DIFFERENTIAL PRESSURE TRANSMITTERS
CURRENT OUTPUT
SPECIFICATION DATA & MOUNTING INSTRUCTIONS
FEATURES
Monitoring gaseous, non-aggressive media
Piezo-resistive pressure transducer
Up to 40 kPa (60 kPa) overload capacity
Rugged design; protection class IP54
Easy installation and wiring connection
Measurement range adjustable by jumper
Response time adjustable by jumper
Re-zeroing possible by pushbutton
NOTE: These sensors are not suitable for use in
installations under periodic inspection by the U.S. Food and Drug Administration.
The differential pressure transmitters of the DPTM series are used for measuring differential pressure, positive pressure, and vacuum. The transmitters are suitable for:
air-conditioning,
building automation,
environmental protection,
valve and flap control,
filter and blower monitoring,
fluid and level monitoring, and
control of air flows.
Models
order no.
DPTM52 -50...0...+50 Pa1) n.a. 20 kPa 40 kPa
DPTM112 -100...0...+100 Pa1 n.a. 20 kPa 40 kPa
DPTM552 -500...0...+500 Pa1) n.a. 20 kPa 40 kPa
DPTM1102 -1 kPa...0...+ 1 kPa2) n.a. 40 kPa 70 kPa
DPTM102 0...100 Pa1) 0...250 Pa 20 kPa 40 kPa
DPTM252 0...250 Pa1) 0...500 Pa 20 kPa 40 kPa
DPTM502 0...500 Pa1) 0...1 kPa 20 kPa 40 kPa
DPTM1002 0...1 kPa2) 0...2.5 kPa 40 kPa 70 kPa
DPTM5002 0...5 kPa3) 0...10 kPa 60 kPa 120 kPa
1)
Temperature error at 0…50 °C ≤ ± 5% of FS
2)
Temperature error at 0…50 °C ≤ ± 2.5% of FS
3)
Temperature error at 0…50 °C ≤ ± 1% of FS
pressure range
1
(pre-set)
2
overload
capacity
bursting
pressure
SPECIFICATION
Supply voltage 16...32 Vdc Output signal 4...20 mA, two-wire Pressure medium Air + non-aggressive gases Working temperature 0...50 °C Linearity and hysteresis error ≤ ± 1.0% of FS Temperature error see "Models" Storage temperature -10...+70 °C Humidity 0...95% rh, non-condensing Repetition accuracy ≤ ± 0.2% of FS Response time 1 s (switchable to 100 ms) Process connection 6 mm hose pipe Electrical connection Screw terminal block for wire
up to 1.5 mm² Housing material ABS and POM Cable entry M20x1.5 (polyamide) Protection class IP54 as per EN60529 EMV EN60770, EN61326 Weight approx. 130 g
FUNCTION
DPTM52-5002 Two-Wire Differential Pressure Transmitters are equipped with an integrated piezo-resistive pressure transducer. The pressure to be measured is applied to and thus deflects a thin membrane made of mono-silicon. The membrane's semiconductor resistors (arranged to simul­taneously compensate for the temperature response) detect this deflection and generate an electrical output signal. The output signal is converted into a 4…20 mA analog signal
® U.S. Registered Trademark EN0B-0467GE51 R1007
Copyright © 2007 Honeywell Inc. • All rights reserved
DPTM52-5002 TWO-WIRE DIFFERENTIAL PRESSURE TRANSMITTERS
j
j
A
V
AD (
which changes (within the specified error limits) in proportion to the applied pressure.
NOTE: The devices are factory pre-set to pressure range 1
This can be changed (except for +/- models) to pressure range 2 by removing the corresponding jumper (see Fig. 3).
NOTE: The devices are factory pre-set to a response time
of 1 second. This can be changed to 100 ms by removing the corresponding jumper (see Fig. 3).
ACCESSORIES
DPSK: Included in delivery. Duct Kit, incl. 2 m of silicone
hose and two joining pipes
DPSL: Ordered separately. L-shaped mounting brackets
with screws.
DIMENSIONS
4.5
8.5
IP 54
WIRING
Range:
high
low
Response:
fastmAslow Mode: Output:
V
Jumper yes Jumper no
Differential Pressure Transmitter
2341
REZEROING: Remove tubes and push button approx. 5 sec.
1000
PRESSURE RANGE jumper inserted = pressure range 1 jumper removed = pressure range 2
RESPONSE TIME
umper inserted = 1 sec
umper removed 100 msec=
NALOG OUTPUT (4...20 mA)
jumper without function
LOAD
P2 (LOW PRESSURE)
P1 (HIGH PRESSURE)
STORAGE POSITION FOR JUMPERS
Fig. 3. Wiring details
NOT USED
1
16...32Vdc
2
4...20 mA
3
DO NOT USE!
4
DO NOT USE!
Fig. 1. Dimensions (in mm)
MOUNTING
800
OHMS)
600
400
E LO
200
0
RESISTI
32 34302826242220181614
SUPPLY VOLTAGE (Vdc)
Fig. 4. Permissible load vs. supply voltage
Fig. 2. Mounting
Manufactured for and on behalf of the Environmental and Combustion Controls Division of Honeywell Technologies Sàrl, Ecublens, Route du Bois 37, Switzerland by its Authorized Representative:
Fema Controls Honeywell GmbH
P.O. Box 1254 71099 Schönaich Germany
phone: (49) 7031-637-02 fax: (49) 7031-637-850 http://www.honeywell.de/fema
Subject to change without notice. EN0B-0467GE51 R1007