9
CAUTION
■ The tool can get hot while operating.
■ You could burn your hands.
■ Do not disassemble the tool while
it is hot. Let the tool cool down.
8. Care and maintenance
8.1 Care of the tool
The outer casing of the tool is manufactured from impactresistant plastic. The grip comprises a synthetic rubber
section.
The ventilation slots must be unobstructed and kept
clean at all times. Do not permit foreign objects to enter
the interior of the tool. Use a slightly damp cloth to clean
the outside of the tool at regular intervals. Do not use a
spray or steam-cleaning system for cleaning.
8.2 Maintenance
Check all external parts of the tool for damage at regular intervals and check that all controls operate properly. Do not operate the tool when parts are damaged or
when the controls do not operate properly. If necessary ,
have the tool repaired at a Hilti service centre.
Servicing the tool
The tool should be serviced if:
1. Cartridges misfire
2. Fastener driving power is inconsistent
3. If you notice that:
●contact pressure increases,
●trigger force increases,
●power regulation is difficult to adjust (stiff),
●the cartridge strip is difficult to remove.
8.3 Disassemble the tool
1. Ensure that no cartridge strip or fastener is loaded in
the tool. Remove the cartridge strip by pulling it
upwards out of the tool and remove the fastener from
the fastener guide.
2. Press the release button at the side.
3. Unscrew the fastener guide / magazine.
4. Remove the buffer by bending it away from the fastener guide / magazine.
5. Remove the piston.
8.4 Check the buffer and piston for wear
Replace the buffer if:
●the metal ring is loose or broken,
●the buffer no longer holds on the fastener guide,
● excessive, uneven wear beneath the metal ring is
noticed.
Replace the piston if:
●it is broken,
●the tip is heavily worn (i.e. a 90°segment is chipped
off),
●piston rings are broken or missing,
●it is bent (check by rolling on an even surface).
NOTE
Do not use worn pistons. Do not modify or grind pistons.
8.5 Check the fastener guide for wear
The nosepiece of the X-460-FIE and X-460-FIE-L fastener guides should be replaced if the tubular section is
damaged (e.g. bent, widened or cracked). For instructions on replacing the nosepiece, please refer to sections 6.3 and 8.5.
1. Ensure that no cartridge strip or fastener is loaded
in the tool. Remove the cartridge strip by pulling it
upwards out of the tool and remove the fastener
from the fastener guide.
2. Press the release button on the side of the fastener
guide.
3. Unscrew the fastener guide.
4. Check the buffer and piston for wear (see care and
maintenance instructions).
5. Pull the moveable ring downwards and unscrew and
remove the clamping nut.
6. Replace the fastener guide nosepiece.
7. Pull the moveable ring downwards and screw on
the clamping nut.
8. Push the piston into the tool as far as it will go.
9. Press the buffer onto the fastener guide until it
engages.
10. Push the fastener guide firmly onto the exhaust gas
piston return unit.
11. Screw the fastener guide onto the tool until it engages.
8.6 Clean the threaded section of the fastener guide/
magazine
1. Clean the thread with the flat brush.
2. Spray the thread lightly with Hilti spray .
8.7 Disassemble the piston return unit
1. Press the release button at the gripping part.
2. Unscrew the piston return unit.
8.8 Clean the piston return unit
1. Clean the spring with the flat brush.
2. Clean the front end with the flat brush.
3. Use the small round brush to clean the two holes at
the end face.
4. Spray the piston return unit lightly with Hilti spray .
8.9 Clean inside the housing
1. Use the large round brush to clean inside the hous-
ing.
2. Spray the inside of the housing lightly with Hilti spray .