HIKVISION DS-2TD1217B-6/PA, DS-2TD2617B-3/PA Installation & Configuration Guide

HIKVISION DS-2TD1217B-6/PA, DS-2TD2617B-3/PA Installation & Configuration Guide

Fever Screening Thermal Camera

Installation & Configuration Guide

Temperature measurement range: 30-45

Working temperature: 10-35

Accuracy: ±0.5

Resolution:160 × 120

Bullet:DS-2TD2617B Turret:DS-2TD1217B

1 Installation-Tripod quick temporary deployment

Tripod Adapter For

Tripod Adapter for

A. Tripod connector fixed on the adapter

Turret

Bullet

B. Fix the adapter on the camera

C. Install the camera on the tripod

Installation Parameter Recommended

 

Model

 

Distance

 

(between human&camera)

 

 

 

 

 

 

 

 

 

 

DS-2TD1217B-3/PA

 

0.8-1.5m

 

 

 

 

 

DS-2TD2617B-3/PA

 

 

 

 

 

 

 

 

 

 

 

 

DS-2TD1217B-6/PA

 

1.5-3m

 

 

 

 

 

DS-2TD2617B-6/PA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Model

 

Height

 

Elevation angle

 

 

 

 

 

 

 

DS-2TD1217B-3/PA

 

 

 

 

 

 

 

 

 

 

 

DS-2TD2617B-3/PA

 

 

 

≤20°

 

 

 

1.5m

 

 

DS-2TD1217B-6/PA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DS-2TD2617B-6/PA

 

 

 

 

D. Installation complete

 

 

 

 

 

 

 

 

 

 

2 Configuration

1.Select Body Thermography as VCA Resource Type. 4.Configure Camera 01 (Optical) Settings.

2. Go to Local Configuration interface, enable and save the following settings:

3. Go to Body Thermography Basic Settings. Configure and save the following settings:

Emissivity: For human skin, this value is normally set as 0.98.

Distance: The actual distance between the camera and measured object.

Check Enable Face Detection and Display Temperature. Suggest uncheck Upload Captured Face Image and Display Face Temperature Position if its not a must.

Set Alarm Temp. and Pre-alarm Temp.

Press Draw Area to adjust face detection region.

Click Max. Pupil Distance and Min. Pupil Distance to draw width filter frame, Max and Min distance refer to camera detection distance.

5. Configure Camera 02 (Thermal) Settings.

3 Advice

Uncheck Enable Blackbody Correction If no blackbody is used. Check Enable of Body Temperature Compensation and keep parameters as default.

The environment would easily influence the performance of thermal camera. Therefore, it is required to be used in a stable indoor environment without wind, and the ambient temperature should be consistent. Here’s some advice:

1.Set up a one-way screening zone, and ensure that camera can clearly see the face of person whose body temperature is being measured.

2.Avoid backgrounds that are too crowded or bright.

3.30 minutes after turn on the camera, then go body temperature measurement, because the camera itself needs to be steady first.

4.If there is a large gap between indoor and outdoor temperature, it is highly suggested to wait more than 5 minutes then measure body temperature of whom just got in.

5.Follow the guidance of required temperature measurement distance.