Dear Customer,
Thank you for choosing a HANNA product.
Please read this instruction manual carefully
before using the instrument.
For additional technical information, do not
hesitate to e-mail us at tech@hannainst.com
These instruments are in compliance with the
directives.
WARRANTY
All Hanna Instruments meters are warranted
for two years against defects in workman-
ship and materials when used for their intended
purpose and maintained according to the instructions. Probes are warranted for 6 months.
Damages due to accidents, misuse, tampering
or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge. If service is required,
contact the dealer from whom you purchased
the instrument. If under warranty, report the model
number, date of purchase, serial number and
the nature of the failure. If the repair is not covered by the warranty, you will be notified of the
charge for repair or replacement. If the instrument is to be returned to Hanna Instruments,
obtain a Return Goods Authorization from the
Customer Service Department first and then send
it with shipment cost prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
CE Declaration of Conformity ................... 15
PRELIMINARY EXAMINATION
Remove the instrument from the packing
material and examine it to make sure that
no damage has occurred during shipping. If
there is any damage, notify your Dealer.
Note: Save all packing material until you
are sure that the instrument functions correctly. All defective items
must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
HI 8564 and HI 93640 portable thermo-hygrom-
eters provide reading with high precision in a
very short time for both RH (Relative Humidity) and temperature.
HI93640 is a compact, portable and versatile
instrument to monitor relative humidity anywhere. The built-in thin-film capacitance sensor assures accurate measurements from 5
to 95% RH with a resolution of 0.1%.
Note: The RH probe sensor must never
come into contact with water or
other liquids.
HI8564 is supplied complete with HI70601/2
RH probe, 9V battery and instructions.
HI93640 is supplied complete with probe sin-
Display Symbols:
°Cto indicate temperature readings in °C
Hto indicate RH readings
Vto indicate that battery needs to be replaced
45
Page 4
FUNCTIONAL DESCRIPTION OF HI93640
SPECIFICATIONS OF HI 93640
ins
rum n
°C
H
THERMOHYGROMETER
RANGE:
RH% 5 to 95
°C 0 to 60
°F 32 to 140
°F
HI93640
OFF
RH%
1
Range5.0 to 95.0% RH
0.0 to 60.0°C / 32 to 140°C
2
Resolution0.1% RH
0.1°C / 1°F
Accuracy±2% RH
3
s
e
±0.4°C / ±1°F (for 1 year excluding probe error)
Typical EMCDeviation
±3% RH / ±0.4°C / ±0.8°F
Environment 0 to 50°C (32 to 122°F);
V
ON
4
5
Battery Type / Life
Dimensions190 x 80 x 38 mm
6
RH max 98% non-condensing
1 x 9V / approx. 100 hours of use
(7.5 x 3.1 x 1.5")
Weight200 g (7.1 oz.)
7
°C
1) Detachable sintered cap
2) Built-in RH/temperature probe
3) LCD (Liquid Crystal Display)
4) ON/OFF key, to turn the meter on or off
5) RH% key, to display RH measurements
6) °F key, to display temperature in °F
7) °C key, to display temperature in °C
Display Symbols:
°Cto indicate temperature readings in °C
Hto indicate RH readings
Vto indicate that battery needs to be replaced
67
Page 5
OPERATIONAL GUIDE
CALIBRATION
INITIAL PREPARATION
Each meter is supplied complete with a 9V
battery. Slide off the battery compartment
cover on the back of the meter (see "Battery
Replacement" section), unwrap the battery and
install it.
To prepare HI 8564 for use, connect the RH
probe to the DIN socket on the top of the meter.
Turn the meter on by pressing the ON/OFF
button.
RELATIVE HUMIDITY MEASUREMENTS
• The end of the humidity detector should be
exposed to a current of air moving at 0.5 m
(20") per second or more.
• In the absence of air movement, the response
can be accelerated by moving the probe.
• The probe sensor must never come into
contact with water or other liquids.
If this happens, or if condensation causes
drops to form on the surface of the humidity
sensor, turn the instrument off and wait until
the liquid has evaporated completely. In order
to accelerate the evaporation process, expose the humidity sensor to a current of air.
For HI93640 only: for a faster response,
remove the sintered cap. Whenever the probe
is to be used in a dusty or smoky environment, the cap must be kept on at all times.
TEMPERATURE MEASUREMENTS
Simply press the °C or °F keys to display the
temperature reading.
When measuring temperature in degrees Centigrade, "°C" is displayed.
For any problem in taking measurements, please
contact the nearest HANNA Customer Service
Department.
89
All HANNA hygrometers have been
precalibrated at the factory.
HANNA instruments® uses state-of-the-art
thermal humidity chambers for this purpose.
It is generally recommended to have all hygrometers recalibrated at least once a year.
For an accurate annual recalibration, contact
your nearest Hanna Service Center.
You can also check the status of your HI8564
and perform a quick recalibration (with an
accuracy of ±5%) by using the Hanna minicalibration chambers HI7101.
The kits are composed of two
thermally isolated chambers,
each one equipped with a
Na Cl
threaded cap and three bottles
containing the appropriate
Li Cl
11.1%
CALIBRATION KIT
75.4%
precalibrated saturated salts
to produce a known RH value.
Preparing the Calibration Solutions (HI 8564
only)
O
• Pour approximately 26 cc of
2
H
distilled water into a glass
container.
26 cc
• Immerse this container into a
bath of ice and water and
shake briefly.
Li Cl
• Slowly add the contents of
a HI 7111 bottle containing
LiCl into the glass container
while continuing to shake.
Page 6
• When the salt has dissolved
completely, add all the content of the second bottle of
HI 7111.
• Allow the solution obtained
to cool, and then pour it into
the chamber marked
"RH 11.1%", making sure
that no residue remains on
the walls of the glass container.
Na Cl
Li Cl
75.4%
11.1%
CALIBRATION KIT
Li Cl
RH Calibration Procedure (HI 8564 only)
• Bring the calibration kit to a temperature of
approximately 20°C.
• Remove the cap from the
"RH 11.1%" chamber containing the LiCl solution and
insert the probe paying attention not to tip it into the
Na Cl
Li Cl
75.4%
11.1%
CALIBRATION KIT
liquid.
• Remove the adhesive sticker which covers
the calibration trimmer access holes.
• Seal the chamber well when not in use, as
the LiCl solution is extremely hygroscopic
and tends to capture the humidity present in
the air causing the solution to expand in
volume and to overflow from the container.
H
2
O
• Pour approximately 12 cc of
distilled water into the other
12 cc
chamber marked "RH
75.4%".
• Add all the content of the
Li Cl
11.1%
CALIBRATION KIT
Na Cl
Na Cl
75.4%
HI 7121 bottle containing
NaCl while continuously
shaking the container to
avoid the formation of lumps.
Seal this container well when
Na Cl
Li Cl
75.4%
11.1%
CALIBRATION KIT
not in use.
The calibration kit needs
4 hours for stabilization.
12
9
3
4 h
6
ON
• Press the ON/OFF key to
OFF
switch the instrument on.
12
9
• Wait for the measurement to
stabilize (this takes about
3
4 h
6
4 hours).
• Turn the low humidity trimmer (see Functional Description) until a value of 0.0% is
seen on the readout (reading between 0.0%
and 1.0% RH is also acceptable).
H
• Remove the probe and tightly seal the chamber containing the LiCl solution.
1011
Page 7
• Remove the cap from the
"RH 75.4%" chamber containing the NaCl solution and
insert the probe.
• Wait for the measurement to
stabilize (approximately
4 hours)
Li Cl
11.1%
CALIBRATION KIT
12
9
6
Na Cl
75.4%
3
4 h
• Turn the high humidity trimmer until the
readout on the display is 64.3%.
H
• Wait for 1 hour and readjust if necessary.
BATTERY REPLACEMENT
Battery replacement must only take place in a
safe area and using the battery type specified in
this instruction manual.
FOR HI8564
When the battery becomes
weak, the "V" symbol is displayed on the LCD to warn the
user.
In order to replace the battery, simply slide off the cover
and replace the 9V battery
with a new one.
BATTERY
• Leaving the probe in the "RH 75.4%" chamber, adjust the low humidity trimmer until the
value of 75.4% is read.
H
• Now, the humidity calibration is complete.
Spare saturation salts are also available:
HI 7111/P (LiCl) for low range humidity, and
HI 7121/P (NaCl) for high range humidity.
1213
FOR HI93640
When the battery becomes weak, the "V" symbol is displayed on the LCD to warn the user.
HI93640 is provided with the advanced BEPS
(Battery Error Preventing System) technology, that turns the meter off when a low
power condition is detected. It is recommended
to replace the 9V battery with a new one.
When the battery need to
be replaced, simply slide off
the battery cover and install
9 VOLT BATTERY9 VOLT BATTERY
a new 9V battery.
Page 8
ACCESSORIES
HI 70601/2RH probe for HI8564 with built-
in temperature sensor and 2 m
(6.6') cable
HI 70601/5RH probe for HI8564 with built-
in temperature sensor and 5 m
(16.5') cable
HI 710001Soft carrying case for HI 8564
HI 710007Blue shockproof rubber boot
for HI 93640
HI 710008Orange shockproof rubber boot
for HI93640
HI 710009Blue shockproof rubber boot
for HI 8564
HI 710010Orange shockproof rubber boot
for HI8564
HI 710011Probe sintered cap for HI93640
HI 710031Rugged carrying case
HI 7101RH calibration chamber for
HI8564
HI 7102RH calibration chamber for
HI 93640
HI 7111/PSpare saturation LiCl salts for
low humidity calibration (15 g,
6 pcs)
HI 7121/PSpare saturation NaCl salts for
high range humidity calibration
(33 g, 6 pcs)
Hanna Instruments reserves the right to
modify the design, construction and appear-
ance of its products without advance notice.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely
suitable for the environment in which they are used.
Operation of these instruments in residential area could cause
unacceptable interferences to radio and TV equipments, requiring
the operator to take all necessary steps to correct interferences.
In HI 8564 the trimmers are sensitive to electrostatic discharge.
It is recommended to use an antistatic screwdriver during calibration.
Any variation introduced by the user to the supplied equipment
may degrade the instrument's EMC performance.
To avoid electrical shock, do not use these instruments when
voltages at the measurement surface exceed 24 Vac or 60 Vdc.
To avoid damages or burns, do not perform any measurement in
microwave ovens.
1415
Page 9
SALES & TECHNICAL SERVICE
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and
Technical offices, please see www.hannainst.com
16