Dear Customer,
Thank you for choosing a Hanna Instruments
Product.
Please read this instruction manual carefully
before using the instrument.
This manual will provide you with all the
necessary information for the correct use of
the instrument, as well as a precise idea of its
versatility in a wide range of applications.
These instruments are in compliance with
directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing
material and examine it to make sure that
no damage has occurred during shipping. If
there is any damage, notify your Dealer.
Note: Save all packing material until you
are sure that the instrument functions correctly. All defective items
must be returned in the original packing with the supplied accessories.
grometers measure both Temperature and
Relative Humidity.
HI93640 is a compact, easy to carry meter
with a convenient wrist-wrap. The built-in
thin film capacitance sensor assures accurate humidity measurements from 5.0 to
95.0% RH with a resolution of 0.1%.
The RH probe contains both the sensor and
the electronic circuitry necessary to amplify
the RH as well as the temperature readings. Another electronic circuit converts the
amplified signal into a digital readout in %
Relative Humidity.
Note: The RH probe sensor must never
come into contact with water or
other liquids.
HI8564 comes supplied with: HI70601/2,
detachable relative Humidity probe and a
9V battery.
HI93640 comes supplied with a 9V battery.
1
Page 3
1
4
3
2
5
6
7
FUNCTIONAL DESCRIPTION OF HI8564
1
2
3
6
5
4
FUNCTIONAL DESCRIPTION OF HI93640
7
8
9
10
11
12
13
°C
H
HI 8564
thermo
hygrometer
RH%
°F
RANGES
RH%
10....95
°C
0....60
°F
32....140
ins
rum n
ON
OFF
°C
V
s
e
1) DIN socket for Relative Humidity/Temp. Probe
2) Liquid Crystal Display
3) ON/OFF Key (to turn the meter on or off)
4) RH% Key (to display RH measurements)
5)OF Key (to display Temperature in Fahrenheit)
6)OC Key (to display Temperature in Centigrade)
7) Shielded Cable
8) Low RH Trimmer
9) Low Temperature Trimmer
10 ) High RH Trimmer
11) High Temperature Trimmer
12 ) Polypropylene Probe Body
13 ) Protective Cover (to prevent damage to sensor)
rum n
ON
OFF
RH%
°C
s
e
V
ins
°C
H
THERMOHYGROMETER
RANGE:
RH% 5 to 95
°C 0 to 60
°F 32 to 140
°F
HI93640
1) Detachable filter cap (HI936CAP)
2) Built-in RH/Temp. probe
3) Liquid Crystal Display
4) ON/OFF Key (to switch on/off)
5) RH% Key (to display RH measurements)
6)OF Key (to display Temperature in Fahrenheit)
7)OC Key (to display Temperature in Centigrade)
Display Indicators:
"OC" (to indicate Temperature readings in degrees Centigrade)
"H" (to indicate Relative Humidity readings)
"V" (to indicate that battery needs to be replaced)
2
Display Indicators:
"OC" (to indicate Temperature readings in degrees Centigrade)
"H" (to indicate Relative Humidity readings)
"V" (to indicate that battery needs to be replaced)
3
Page 4
SPECIFICATIONS OF HI 8564SPECIFICATIONS OF HI 8564
SPECIFICATIONS OF HI 8564
SPECIFICATIONS OF HI 8564SPECIFICATIONS OF HI 8564
SPECIFICATIONS OF HI 93640SPECIFICATIONS OF HI 93640
SPECIFICATIONS OF HI 93640
SPECIFICATIONS OF HI 93640SPECIFICATIONS OF HI 93640
RANGER H10.0 to 95.0%
°C0.0 to 60.0
°F32 to 140
RESOLUTION
R H0.1%
°C0.1
°F1
ACCURACY
(@20°C / 68°F) RH+2%
°C+0.4
°F+1
for one year, excluding probe error
TYPICAL EMC
DEVIATIONRH+2%
°C+0.5
°F+1
OPERATING0 to 50 oC (32 to 122 oF)
CONDITIONS98% RH max.
BATTERYOne 9V, 100 hours of continuous
use, with low battery indicator
DIMENSIONS
Meter185 x 82 x 45 mm
(7.3" x 3.2" x 1.8")
ProbeL:165 mm
D:25 mm
with 2 m cable
(L:6.5" D:1"
7' cable)
RANGER H5.0 to 95.0%
°C0.0 to 60.0
°F32 to 140
RESOLUTION
R H0.1%
°C0.1
°F1
ACCURACY
(@20°C / 68°F) RH+2%
°C+0.4
°F+1
for one year excluding probe error
TYPICAL EMC
DEVIATIONR H±3%
°C±0.4
°F±0.8
OPERATING0 to 50oC (32 to 122oF)
CONDITIONS98% RH max.
BATTERYOne 9V, 100 hours of continuous
use, with low battery indicator
DIMENSIONS190 x 80 x 38 mm
(7.5 x 3.1 x 1.5")
WEIGHT320 g (11.3 oz.)
WEIGHT
Meter315 g (11.1 oz.)
Probe200 g (7 oz.)
4
5
Page 5
°C
OPERATIONAL GUIDE
INSTRUMENT PREPARATION
Each meter is supplied complete with a 9V
battery. Slide off the battery compartment
cover on the back of the meter (see page 12),
unwrap the battery and install it.
To prepare HI8564 for use connect the RH
probe to the DIN socket on the top of the meter.
ON
OFF
Turn the
meter on by
pressing the
ON/OFF key.
ON
OFF
Using the HI8564 or HI93640 thermo-hygrometer is simple. However, the recommendations below should always be followed.
be removed. Whenever the probe is to be
used in a dusty or smoky environment, the
filter must be kept on at all times.
TEMPERATURE MEASUREMENTS
Simply press the °C or °F keys to take the
temperature readings directly from the same
probe.
Note: Press °C key to display
the temperature in °C.
°C
"°C" is displayed when measuring temperature in degrees
Centigrade.
EXAMPLE: A temperature
reading taken in degrees
Centigrade would read:
°C
RELATIVE HUMIDITY MEASUREMENTS
• The end of the humidity detector should be
exposed to a current of air moving at 0.5 m
(20") per second or more.
• In the absence of air movement, the response can be accelerated by moving the
probe.
• The probe sensor must never come into
contact with water or other liquids. If this
should happen, or if condensation causes
drops to form on the surface of the humidity
sensor, turn off the instrument and wait until
the liquid has evaporated completely. In
order to accelerate the evaporation process, the humidity sensor could be exposed
to a current of air.
For HI93640 only: in order to obtain a
faster response, the sintered filter should
6
Press °F key to read the same
°F
value in degrees Fahrenheit:
°F
If you experience any problems in taking
measurements, please contact your dealer or
the nearest Hanna Instruments Customer
Service Department.
7
Page 6
CALIBRATION
All Hanna hygrometers have been
precalibrated at the factory.
Hanna Instruments uses state-of-the-art thermal humidity chambers for this purpose.
It is generally recommended to have all
hygrometers recalibrated at least once a
year.
For an accurate annual recalibration, contact your nearest Hanna Service Center.
You can also check the status of your
HI8564 and perform a quick recalibration
(with an accuracy of ±5%) by using the
Hanna mini-calibration chambers HI 7101.
The kits are composed of two
l
a C
N
l
Li C
thermally isolated chambers,
each one equipped with a
75.4%
11.1 %
CALIBRATION KIT
threaded cap and three bottles
containing the appropriate precalibrated saturated salts to produce a known RH value.
Preparing the Calibration Solutions
(HI 8564 only):
O
2
H
• Pour approximately 26 cc of
distilled water into a glass
container.
26 cc
• Immerse this container into
a bath of ice and water and
shake briefly.
l
Li C
• Slowly add the contents of
a HI 7111 bottle containing
LiCl into the glass container
while continuing to shake.
l
• When the salt has dissolved
Li C
completely, add all the content of the second bottle of
HI 7111.
• Allow the solution obtained
to cool, and then pour it into
the chamber marked
"RH 11.1%", making sure
that no residue remains on
the walls of the glass container.
Na Cl
Li Cl
75.4%
11.1%
IT
K
N
IO
T
A
R
IB
L
A
C
• Seal the chamber well when not in use, as
the LiCl solution is extremely hygroscopic
and tends to capture the humidity present in
the air causing the solution to expand in
volume and to overflow from the container.
H
2
O
• Pour approximately 12 cc of
distilled water into the other
12cc
chamber marked "RH
l
75.4%".
• Add all the content of the
Na Cl
N
l
i C
L
5
7
%
.1
11
CALIBRATION KIT
C
a
%
.4
HI 7121 bottle containing
NaCl while continuously
shaking the container to
avoid the formation of lumps.
Seal this container well when
N
l
i C
L
5
7
%
.1
11
CALIBRATION KIT
l
C
a
%
.4
not in use.
The calibration kit needs
4 hours for stabilization.
12
9
3
4 h
6
8
9
Page 7
RH Calibration Procedure (HI 8564 only):
• Bring the calibration kit to a temperature of
approximately 20°C.
• Remove the cap from the
"RH 11.1%" chamber containing the LiCl solution and
insert the probe paying attention not to tip it into the
a Cl
N
Li Cl
75.4%
11.1%
IT
K
N
IO
T
A
R
IB
AL
C
liquid.
• Remove the adhesive sticker which covers
the calibration trimmer access holes.
ON
• Press the ON/OFF key to
OFF
switch the instrument on.
12
9
• Wait for the measurement to
stabilize (this takes about
3
4 h
6
4 hours).
• Turn the low humidity trimmer (see Functional Description) until a value of 0.0% is
seen on the readout (reading between 0.0%
and 1.0% RH is also acceptable).
• Remove the cap from the
"RH 75.4%" chamber containing the NaCl solution and
insert the probe.
• Wait for the measurement to
stabilize (approximately
4 hours)
9
l
C
a
N
l
i C
L
%
.4
5
7
%
.1
11
IT
K
N
IO
T
A
R
LIB
A
C
12
3
4 h
6
• Turn the high humidity trimmer until the
readout on the display is 64.3%.
H
• Wait for 1 hour and readjust if necessary.
• Leaving the probe in the "RH 75.4%" chamber, adjust the low humidity trimmer until the
value of 75.4% is read.
H
• Remove the probe and tightly seal the chamber containing the LiCl solution.
10
H
• Now, the humidity calibration is complete.
Spare saturation salts are also available:
HI 7111/P (LiCl) for low range humidity, and
HI 7121/P (NaCl) for high range humidity.
11
Page 8
HI93640
RH%
°F°C
°C
OFF
ON
THERMOHYGROMETER
RANGE:
RH% 5 to 95
°C 0 to 60
°F 32 to 140
BATTERY REPLACEMENT
9 VOLT BATTERY
SHOCKPROOF RUBBER BOOTS
Battery replacement must only take place in a
non hazardous area using the battery types
specified in this instruction manual (see
page 14).
FOR HI93640 ONLY:
When the battery is rundown
"V" is displayed on the LCD to
warn the user.
HI93640 employs the advanced "BEPS" technology.
BEPS will detect a low power
condition and will shut the display off. If the ON/OFF button is pressed again, the
meter will shut off after a few seconds.
When this occurs, replace the 9V battery
with a new one.
When the battery is exhausted or the instrument cannot be switched on, simply slide
the battery cover off from the rear of the
instrument and replace with the 9V battery
with a new one.
HI8564
BATTERY
FOR HI8564 ONLY:
When the battery is run down
"V" is displayed on the LCD to
warn the user.
In order to replace the battery, simply slide the cover
off and replace it with a new
9V battery.
HI93640
9 VOLT BATTERY
SLIDE
OFF
12
These rubber boots for HI93640 are specially
made to prolong the life of your meter and to
prevent damages due to accidental dropping.
This rubber boot measures 155 x 90 x 45 mm
(6.1 x 3.5 x 1.8") and is available in two
different colors:
HI710008 = orange
HI710007 = blue.
Select the proper color according to your
preference.
13
Page 9
ACCESSORIES
WARRANTY
HI70601/5Detachable interchangeable
TFPC RH & temp. probe for
HI8564 with 5 m (16.5') cable
HI 710001Soft carrying case (for HI 8564
only)
HI 710007Blue rubber boot (for HI93640
only)
HI 710008Orange rubber boot (for
HI93640 only)
HI710011Spare sintered cap for pro-
tection (for HI93640 only)
HI 710031Rugged carrying case
HI7101Calibration chambers for
HI 8564, complete with O-ring
and pre-calibrated saturated
salts
HI7111/PSpare saturation salts 6 x 15g
bottles of LiCl for low range
humidity calibration
HI7121/PSpare saturation salts 6 x 33g
bottles of NaCl for high range
humidity calibration
HI 7213109V battery (10 pcs)
MANRHR2Instruction manual
All Hanna Instruments meters are warranted for two years against defects in
workmanship and materials when used for
their intended purpose and maintained according to the instructions. The probes arewarranted for a period of six months.
Damages due to accidents, misuse, tampering or lack of prescribed maintenance
are not covered. This warranty is limited to
repair or replacement free of charge.
If service is required, contact the dealer
from whom you purchased the instrument.
If under warranty, report the model number,
date of purchase, serial number and the
nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charge for repair or replacement.
If the instrument is to be returned to Hanna
Instruments, obtain a Return Goods Authorization from the Customer Service Department first and then send it with shipment
cost prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return
the enclosed warranty card within 14 days
from the date of purchase.
14
All rights are reserved. Reproduction in
whole or in part is prohibited without the
written consent of the copyright owner.
Hanna Instruments reserves the right to
modify the design, construction and appearance of its products without advance notice.
15
Page 10
OTHER PRODUCTS FROM HANNA
•Chemical Test Kits
•Conductivity and TDS Meters
•Dissolved Oxygen Meters
CE DECLARATION OF CONFORMITY
•Dosing Pumps
•Electrodes (pH and ORP)
•Hygrometers
•Ion Specific Meters
•Magnetic Stirrers
•pH Meters
•Printing and Logging Meters
•Process Instrumentation
•Solutions and Accessories
•Testers (pH, conductivity, temperature,
pNa, TDS, ORP)
•Thermometers
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl
V.le delle industrie 12
35010 Ronchi di Villafranca (PD)
ITALY
herewith certify that the thermo-hygrometers
have been tested and found to be compliant with the following regulations:
IEC 801-2Electrostatic Discharge
IEC 801-3RF Radiated
EN 55022Radiated, Class B
Date of Issue: 12-10-1995
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which
they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and
TV equipments, requiring the operator to take all necessary steps to correct interferences.
In HI8564 the trimmers are sensitive to Electrostatic Discharge. It is recommended to use an
antistatic screwdriver during calibration.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC
performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface
exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
HI 8564 HI 93640
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
16
17
Page 11
HANNA LITERATUREHANNA LITERATURE
HANNA LITERATURE
HANNA LITERATUREHANNA LITERATURE
HANNA LITERATURE
Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications.
The reference literature currently covers areas such
as:
• Water Treatment
• Process
• Swimming Pools
• Agriculture
• Food
• Laboratory
• Thermometry
and many others. New reference material is constantly
being added to the library.
For these and other catalogs, handbooks and leaflets,
contact your dealer or the Hanna Customer Service
Center nearest to you. To find the Hanna Office in
your vicinity, check our home page at
www.hannainst.com.
http://www.hannainst.com
MANRHR2