Dear Customer, Thank you for choosing a Hanna Instruments
Product. Please read this instruction manual carefully
before using the instrument. This manual will provide you with all the
necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications.
These instruments are in compliance with directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.
Note: Save all packing material until you
are sure that the instrument func­tions correctly. All defective items must be returned in the original pack­ing with the supplied accessories.
TABLE OF CONTENTS
Preliminary Examination ........................... 1
General Description .................................. 1
Functional Description of HI 8564.............2
Functional Description of HI 93640...........3
Specifications of HI 8564 .......................... 4
Specifications of HI 93640 ........................ 5
Operational Guide ..................................... 6
Calibration ................................................. 8
Battery Replacement .............................. 12
Shockproof Rubber Boots ....................... 13
Accessories ............................................ 14
Warranty ................................................. 15
CE Declaration of Conformity ................. 17
GENERAL DESCRIPTION
HI 8564 and HI93640 portable Thermo-Hy-
grometers measure both Temperature and Relative Humidity.
HI93640 is a compact, easy to carry meter with a convenient wrist-wrap. The built-in thin film capacitance sensor assures accu­rate humidity measurements from 5.0 to
95.0% RH with a resolution of 0.1%. The RH probe contains both the sensor and
the electronic circuitry necessary to amplify the RH as well as the temperature read­ings. Another electronic circuit converts the amplified signal into a digital readout in % Relative Humidity.
Note: The RH probe sensor must never
come into contact with water or other liquids.
HI8564 comes supplied with: HI70601/2, detachable relative Humidity probe and a 9V battery.
HI93640 comes supplied with a 9V battery.
1
Page 3
1
4
3
2
5 6
7
FUNCTIONAL DESCRIPTION OF HI8564
1
2
3
6
5
4
FUNCTIONAL DESCRIPTION OF HI93640
7
8
9 10 11
12
13
°C H
HI 8564 thermo hygrometer
RH%
°F
RANGES
RH%
10 .... 95
°C
0 .... 60
°F
32 .... 140
ins
rum n
ON
OFF
°C
V
s
e
1) DIN socket for Relative Humidity/Temp. Probe
2) Liquid Crystal Display
3) ON/OFF Key (to turn the meter on or off)
4) RH% Key (to display RH measurements)
5)OF Key (to display Temperature in Fahrenheit)
6)OC Key (to display Temperature in Centigrade)
7) Shielded Cable
8) Low RH Trimmer
9) Low Temperature Trimmer 10 ) High RH Trimmer
11) High Temperature Trimmer 12 ) Polypropylene Probe Body 13 ) Protective Cover (to prevent damage to sensor)
rum n
ON OFF
RH%
°C
s
e
V
ins
°C
H
THERMOHYGROMETER
RANGE: RH% 5 to 95 °C 0 to 60 °F 32 to 140
°F
HI93640
1) Detachable filter cap (HI936CAP)
2) Built-in RH/Temp. probe
3) Liquid Crystal Display
4) ON/OFF Key (to switch on/off)
5) RH% Key (to display RH measurements)
6)OF Key (to display Temperature in Fahrenheit)
7)OC Key (to display Temperature in Centigrade)
Display Indicators: "OC" (to indicate Temperature readings in degrees Centigrade) "H" (to indicate Relative Humidity readings) "V" (to indicate that battery needs to be replaced)
2
Display Indicators:
"OC" (to indicate Temperature readings in degrees Centigrade) "H" (to indicate Relative Humidity readings) "V" (to indicate that battery needs to be replaced)
3
Page 4
SPECIFICATIONS OF HI 8564SPECIFICATIONS OF HI 8564
SPECIFICATIONS OF HI 8564
SPECIFICATIONS OF HI 8564SPECIFICATIONS OF HI 8564
SPECIFICATIONS OF HI 93640SPECIFICATIONS OF HI 93640
SPECIFICATIONS OF HI 93640
SPECIFICATIONS OF HI 93640SPECIFICATIONS OF HI 93640
RANGE R H 10.0 to 95.0%
°C 0.0 to 60.0
°F 32 to 140
RESOLUTION
R H 0.1%
°C 0.1
°F 1
ACCURACY (@20°C / 68°F) RH +2%
°C +0.4
°F +1
for one year, excluding probe error
TYPICAL EMC DEVIATION RH +2%
°C +0.5
°F +1
OPERATING 0 to 50 oC (32 to 122 oF) CONDITIONS 98% RH max.
BATTERY One 9V, 100 hours of continuous
use, with low battery indicator
DIMENSIONS
Meter 185 x 82 x 45 mm
(7.3" x 3.2" x 1.8")
Probe L:165 mm
D:25 mm
with 2 m cable
(L:6.5" D:1"
7' cable)
RANGE R H 5.0 to 95.0%
°C 0.0 to 60.0 °F 32 to 140
RESOLUTION
R H 0.1%
°C 0.1 °F 1
ACCURACY (@20°C / 68°F) RH +2%
°C +0.4 °F +1
for one year excluding probe error
TYPICAL EMC DEVIATION R H ±3%
°C ±0.4 °F ±0.8
OPERATING 0 to 50oC (32 to 122oF) CONDITIONS 98% RH max.
BATTERY One 9V, 100 hours of continuous
use, with low battery indicator
DIMENSIONS 190 x 80 x 38 mm
(7.5 x 3.1 x 1.5")
WEIGHT 320 g (11.3 oz.)
WEIGHT
Meter 315 g (11.1 oz.) Probe 200 g (7 oz.)
4
5
Page 5
°C
OPERATIONAL GUIDE
INSTRUMENT PREPARATION
Each meter is supplied complete with a 9V battery. Slide off the battery compartment cover on the back of the meter (see page 12), unwrap the battery and install it.
To prepare HI8564 for use connect the RH probe to the DIN socket on the top of the meter.
ON
OFF
Turn the meter on by pressing the ON/OFF key.
ON
OFF
Using the HI8564 or HI93640 thermo-hy­grometer is simple. However, the recommen­dations below should always be followed.
be removed. Whenever the probe is to be used in a dusty or smoky environment, the filter must be kept on at all times.
TEMPERATURE MEASUREMENTS
Simply press the °C or °F keys to take the temperature readings directly from the same probe.
Note: Press °C key to display the temperature in °C.
°C
"°C" is displayed when mea­suring temperature in degrees Centigrade.
EXAMPLE: A temperature reading taken in degrees Centigrade would read:
°C
RELATIVE HUMIDITY MEASUREMENTS
The end of the humidity detector should be
exposed to a current of air moving at 0.5 m (20") per second or more.
In the absence of air movement, the re­sponse can be accelerated by moving the probe.
The probe sensor must never come into contact with water or other liquids. If this should happen, or if condensation causes drops to form on the surface of the humidity sensor, turn off the instrument and wait until the liquid has evaporated completely. In order to accelerate the evaporation pro­cess, the humidity sensor could be exposed to a current of air.
For HI93640 only: in order to obtain a faster response, the sintered filter should
6
Press °F key to read the same
°F
value in degrees Fahrenheit:
°F
If you experience any problems in taking measurements, please contact your dealer or the nearest Hanna Instruments Customer Service Department.
7
Page 6
CALIBRATION
All Hanna hygrometers have been precalibrated at the factory. Hanna Instruments uses state-of-the-art ther­mal humidity chambers for this purpose. It is generally recommended to have all hygrometers recalibrated at least once a year.
For an accurate annual recalibration, con­tact your nearest Hanna Service Center.
You can also check the status of your HI8564 and perform a quick recalibration (with an accuracy of ±5%) by using the Hanna mini-calibration cham­bers HI 7101.
The kits are composed of two
l
a C
N
l
Li C
thermally isolated chambers, each one equipped with a
75.4%
11.1 %
CALIBRATION KIT
threaded cap and three bottles containing the appropriate precalibrated satu­rated salts to produce a known RH value.
Preparing the Calibration Solutions (HI 8564 only):
O
2
H
• Pour approximately 26 cc of distilled water into a glass container.
26 cc
• Immerse this container into a bath of ice and water and shake briefly.
l
Li C
• Slowly add the contents of a HI 7111 bottle containing LiCl into the glass container while continuing to shake.
l
• When the salt has dissolved
Li C
completely, add all the con­tent of the second bottle of HI 7111.
• Allow the solution obtained to cool, and then pour it into the chamber marked "RH 11.1%", making sure that no residue remains on the walls of the glass con­tainer.
Na Cl
Li Cl
75.4%
11.1%
IT
K
N
IO
T
A
R
IB
L
A
C
• Seal the chamber well when not in use, as the LiCl solution is extremely hygroscopic and tends to capture the humidity present in the air causing the solution to expand in volume and to overflow from the container.
H
2
O
• Pour approximately 12 cc of distilled water into the other
12 cc
chamber marked "RH
l
75.4%".
• Add all the content of the
Na Cl
N
l
i C
L
5
7
%
.1
11
CALIBRATION KIT
C
a
%
.4
HI 7121 bottle containing NaCl while continuously shaking the container to avoid the formation of lumps. Seal this container well when
N
l
i C
L
5
7
%
.1
11
CALIBRATION KIT
l
C
a
%
.4
not in use.
The calibration kit needs 4 hours for stabilization.
12
9
3
4 h
6
8
9
Page 7
RH Calibration Procedure (HI 8564 only):
• Bring the calibration kit to a temperature of approximately 20°C.
• Remove the cap from the "RH 11.1%" chamber con­taining the LiCl solution and insert the probe paying at­tention not to tip it into the
a Cl
N
Li Cl
75.4%
11.1%
IT
K
N
IO
T
A
R
IB
AL
C
liquid.
• Remove the adhesive sticker which covers the calibration trimmer access holes.
ON
• Press the ON/OFF key to
OFF
switch the instrument on.
12
9
• Wait for the measurement to stabilize (this takes about
3
4 h
6
4 hours).
• Turn the low humidity trimmer (see Func­tional Description) until a value of 0.0% is seen on the readout (reading between 0.0% and 1.0% RH is also acceptable).
• Remove the cap from the "RH 75.4%" chamber con­taining the NaCl solution and insert the probe.
• Wait for the measurement to stabilize (approximately 4 hours)
9
l
C
a
N
l
i C
L
%
.4
5
7
%
.1
11
IT
K
N
IO
T
A
R
LIB
A
C
12
3
4 h
6
• Turn the high humidity trimmer until the readout on the display is 64.3%.
H
• Wait for 1 hour and readjust if necessary.
• Leaving the probe in the "RH 75.4%" cham­ber, adjust the low humidity trimmer until the value of 75.4% is read.
H
• Remove the probe and tightly seal the cham­ber containing the LiCl solution.
10
H
• Now, the humidity calibration is complete. Spare saturation salts are also available:
HI 7111/P (LiCl) for low range humidity, and HI 7121/P (NaCl) for high range humidity.
11
Page 8
HI93640
RH%
°F °C
°C
OFF
ON
THERMOHYGROMETER
RANGE: RH% 5 to 95 °C 0 to 60 °F 32 to 140
BATTERY REPLACEMENT
9 VOLT BATTERY
SHOCKPROOF RUBBER BOOTS
Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual (see page 14).
FOR HI93640 ONLY:
When the battery is rundown "V" is displayed on the LCD to warn the user.
HI93640 employs the ad­vanced "BEPS" technology. BEPS will detect a low power condition and will shut the dis­play off. If the ON/OFF but­ton is pressed again, the meter will shut off after a few seconds. When this occurs, replace the 9V battery with a new one.
When the battery is exhausted or the instru­ment cannot be switched on, simply slide the battery cover off from the rear of the instrument and replace with the 9V battery with a new one.
HI8564
BATTERY
FOR HI8564 ONLY:
When the battery is run down "V" is displayed on the LCD to warn the user.
In order to replace the bat­tery, simply slide the cover off and replace it with a new 9V battery.
HI93640
9 VOLT BATTERY
SLIDE
OFF
12
These rubber boots for HI93640 are specially made to prolong the life of your meter and to prevent damages due to accidental dropping.
This rubber boot measures 155 x 90 x 45 mm (6.1 x 3.5 x 1.8") and is available in two different colors:
HI710008 = orange HI710007 = blue.
Select the proper color according to your preference.
13
Page 9
ACCESSORIES
WARRANTY
HI70601/5 Detachable interchangeable
TFPC RH & temp. probe for
HI8564 with 5 m (16.5') cable
HI 710001 Soft carrying case (for HI 8564
only)
HI 710007 Blue rubber boot (for HI93640
only)
HI 710008 Orange rubber boot (for
HI93640 only)
HI710011 Spare sintered cap for pro-
tection (for HI93640 only)
HI 710031 Rugged carrying case HI7101 Calibration chambers for
HI 8564, complete with O-ring
and pre-calibrated saturated salts
HI7111/P Spare saturation salts 6 x 15g
bottles of LiCl for low range humidity calibration
HI7121/P Spare saturation salts 6 x 33g
bottles of NaCl for high range humidity calibration
HI 721310 9V battery (10 pcs) MANRHR2 Instruction manual
All Hanna Instruments meters are war­ranted for two years against defects in
workmanship and materials when used for their intended purpose and maintained ac­cording to the instructions. The probes are warranted for a period of six months.
Damages due to accidents, misuse, tam­pering or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be noti­fied of the charge for repair or replacement.
If the instrument is to be returned to Hanna Instruments, obtain a Return Goods Autho­rization from the Customer Service Depart­ment first and then send it with shipment cost prepaid. When shipping any instrument, make sure it is properly packaged for com­plete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
14
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and ap­pearance of its products without ad­vance notice.
15
Page 10
OTHER PRODUCTS FROM HANNA
Chemical Test Kits
Conductivity and TDS Meters
Dissolved Oxygen Meters
CE DECLARATION OF CONFORMITY
Dosing Pumps
Electrodes (pH and ORP)
Hygrometers
Ion Specific Meters
Magnetic Stirrers
pH Meters
Printing and Logging Meters
Process Instrumentation
Solutions and Accessories
Testers (pH, conductivity, temperature,
pNa, TDS, ORP)
Thermometers
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl V.le delle industrie 12 35010 Ronchi di Villafranca (PD) ITALY
herewith certify that the thermo-hygrometers
have been tested and found to be compliant with the following regulations:
IEC 801-2 Electrostatic Discharge IEC 801-3 RF Radiated EN 55022 Radiated, Class B
Date of Issue: 12-10-1995
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
In HI8564 the trimmers are sensitive to Electrostatic Discharge. It is recommended to use an antistatic screwdriver during calibration.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
HI 8564 HI 93640
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
16
17
Page 11
HANNA LITERATUREHANNA LITERATURE
HANNA LITERATURE
HANNA LITERATUREHANNA LITERATURE
HANNA LITERATURE
Hanna publishes a wide range of catalogs and hand­books for an equally wide range of applications. The reference literature currently covers areas such as:
• Water Treatment
• Process
• Swimming Pools
• Agriculture
• Food
• Laboratory
• Thermometry
and many others. New reference material is constantly being added to the library.
For these and other catalogs, handbooks and leaflets, contact your dealer or the Hanna Customer Service Center nearest to you. To find the Hanna Office in your vicinity, check our home page at www.hannainst.com.
http://www.hannainst.com
MANRHR2
03/00
Loading...