Hanna Instruments HI 84430 User Manual

Instruction Manual
HI 84430
TOTAL ACIDITY
MINITITRATOR & pH METER
for Water Analysis
www.hannainst.com
1
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary
information for the correct use of the instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with
directives.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION................................................................................................. 4
GENERAL DESCRIPTION ........................................................................................................ 4
SPECIFICATIONS ................................................................................................................... 6
PRINCIPLE OF OPERATION .................................................................................................... 8
FUNCTIONAL AND PHYSICAL DESCRIPTION............................................................................. 9
TITRATOR STARTUP ........................................................................................................... 11
SETUP CONFIGURATION MENU .......................................................................................... 12
ELECTRODE PREPARATION .................................................................................................. 18
ELECTRODE CALIBRATION PROCEDURE................................................................................. 19
pH BUFFER TEMPERATURE DEPENDENCE ............................................................................ 23
PUMP TUBE INSTALLATION................................................................................................. 24
PURGE .............................................................................................................................. 24
PUMP CALIBRATION PROCEDURE ........................................................................................ 26
TITRATION PROCEDURE ..................................................................................................... 29
pH MEASUREMENT ............................................................................................................ 34
TEMPERATURE CALIBRATION PROCEDURE (for technical personnel only) ............................... 38
PC INTERFACE ................................................................................................................... 40
TROUBLESHOOTING GUIDE ................................................................................................ 44
ELECTRODE CONDITIONING & MAINTENANCE ...................................................................... 45
ACCESSORIES ..................................................................................................................... 46
WARRANTY........................................................................................................................47
32
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your Dealer. Each HI 84430 minititrator is supplied complete with:
HI 1131B pH electrode
HI 7071 Filling solution (30 mL)
HI 7662-M Temperature probe
HI 84430-50 Titrant low range (100 mL)
HI 84430-51 Titrant high range (100 mL)
HI 84430-55M Pump calibration solution (230 mL)
HI 84430-58 Additional reagent (30 mL)
HI 7061 Cleaning solution
HI 7001M pH 1.68 buffer solution (230 mL)
HI 7004M pH 4.01 buffer solution (230 mL)
HI 70083M pH 8.30 buffer solution (230 mL)
• Two 100 mL beakers
• Two tube sets with dispensing tip
• Stir bars (medium 2 pcs.)
• 12 Vdc power adapter
• Instruction manual
Note:Save all packing material until you are sure that the instrument works correctly.
Any defective item must be returned in its original packing.
GENERAL DESCRIPTION
The HI 84430 is an easy to use microprocessor-based automatic minititrator and pH meter designed for the rapid and accurate analysis of Total Titratable Acidity in water. By eliminating subjective factors including color indicators, errors in mathematical calculations or erratic titrant additions from the measurement, the HI 84430 makes Total Titratable acidity analysis precisely. This will quickly become a valuable analysis tool of waters and waste waters.
The instrument benefits from Hanna’s many years of experience as manufacturer of quality analytical instruments. A clear and well-designed user interface makes the instrument intuitive and simple to use. A dedicated HELP key aids in set-up, calibration, status and troubleshooting.
By simply pressing the START key, the HI 84430 automatically starts pump operation and titrates the sample to the end point. The HI 84430 has a simple and accurate peristaltic pump to ensure the best accuracy and repeatability. By performing pump calibration with the Hanna standard provided, the instrument accuracy is assured.
The instrument employs a powerful and effective built-in algorithm to analyze the pH response to determine the exact pH endpoint, then uses this to make the necessary calculations. The Titratable Acidity determination is instantaneously displayed in selected measurement units on the large dot matrix display. The instrument is ready for the next analysis immediately.
Other features:
• Log on demand up to 100 samples (50 for pH measurement; 50 for titration results)
• GLP feature, to view last calibration data for pH electrode and pump
• PC interface via USB connection
MEASUREMENT SIGNIFICANCE
Acidity is the quantitative capacity of the water sample to react with a strong base solution in order to obtain a pH end point value. Acids contribute to the corrosive capacity of the water, influence the chemical reaction rates and biological processes. The results of acidity measurement also reflect the quality of the waters (surface, drinking, waste waters) and are an essential monitoring device to define and control the pollution. Strong acids (such as mineral acids: HNO
, H2SO4, HCl) and weak acids (such as organic acids), carbon dioxide
3
content in water, or others acid components (some polyvalent cations such as aluminium, iron, manganesse, chromium, cadmium trivalent cations) can contribute to the acidity of water. According to the Standard Methods for the Examination of Waters and Wastewaters, Volume 20, three options exist to determine water acidity:
pH measurement using the meter in pH mode.
Strong acidity determination by titration of the water samples with alkaline reagent to a
3.7 pH end point (known as methyl orange acidity).
Total acidity determination by titration of the water samples with alkaline reagent to a
8.3 pH end point (known as phenolphthalein acidity).
Potentiometric end point determination, using a pH electrode, is more objective than visual end point determination, using color changing indicators. The HI 84430 is a potentiometric titrator. The volume of dispensing titrant necessary to reach the end point is used to calculate the water acidity, expressed in mg/L as CaCO
(most usual unit) or meq/L as CaCO3.
3
Interferences:
dissolved gases, lost or gained during sample storage or transport, can modify the acidity
of the samples;
the presence in the samples of: precipitates, solids suspensions, oils;
samples containing polyvalent cations (aluminum, iron, manganese, chromium) may
react slowly with the titrant causing a drifting end point;
To eliminate or diminish the interference effects, it is recommended:
to treat the samples that contain polyvalent cations with hydrogen peroxide in order to
increase the hydrolyzing rate of metal ions present.
to protect the samples against atmospheric action (store in an airtight vessel).
to avoid vigorous shaking or mixing: stir gently.
54
SPECIFICATIONS
Titrator Range Titratable acidity (low range)
: 15.0 - 500.0 mg/L as CaCO
mg/L meq/L : 0.3 - 10.0 meq/L as CaCO Titratable acidity (high range)
: 400 - 4000 mg/L as CaCO
mg/L meq/L : 8 - 80 meq/L as CaCO
3
3
Resolution Titratable acidity (low range):
0.1 mg/L
0.1 meq/L Titratable acidity (high range): 1mg/L
1meq/L Accuracy 5% of reading Titration method Acid-base titration (total acidity / strong acidity) Principle End point titration : 8.3 pH / 3.7 pH Pump debit 0.5 mL/min Stirring speed 600 rpm Log data Up to 50 samples
pH meter pH meter -2.0 to 16.0 pH / -2.00 to 16.00 pH
pH Resolution: 0.1 pH / 0.01 pH pH Accuracy: ± 0.01 pH pH Calibration: 1, 2 or 3 calibration points;
3 available buffers (1.68; 4.01; 8.30) Temperature manual or automatic from
compensation: -20 to 120 °C (-4 to 248 °F) Log data Up to 50 samples
Temperature Range -20.0 to 120.0 °C (-4.0 to 248.0 °F)
Resolution 0.1 °C Accuracy ±0.4 °C without probe error
Electrode HI 1131B (included)
3
3
Temperature Probe HI 7662-M (included) Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condensing Power supply 12 Vdc power adapter Dimensions 208 × 214 × 163 mm (8.2×8.4×6.4”) (with beaker) Weight 2200 g (77 oz.)
REQUIRED REAGENTS Code Description Quantity/Test
HI 84430 - 50 Titrant (Low Range) 1 mL HI 84430 - 51 Titrant (High Range) 1 mL HI 84430 - 55M Pump Calibration Solution 2 mL for low range
20 mL for high range
76
PRINCIPLE OF OPERATION
Water acidity determination is based on the neutralization of all acidic species contained in water samples by titration with base solutions (the titrant alkaline reagent):
+
H
+OH- H2O In an ideal solution, the end point of an acid titration corresponds stoichiometrically to the complete neutralization of the acids present. Water acidity results from strong and weak acids, and from the hydrolysis of salts. Due to these more complex mixtures, end points for water acidity are arbitrarily fixed on practical considerations. Unpolluted surface waters contain carbon dioxide as the main acidic component and titration of these samples corresponds to stoichiometric neutralization of carbonic acid (titration to 8.3 pH at 25°C degrees). Strong acidity of the polluted waters is determined by titration to 3.7 pH. If the samples are suspected to contain hydrolizable salts or polyvalent cations (such as chromium, iron, manganese), it is necessary to pretreat the samples with a HI 84430-58 reagent (see Sample preparation /Titration). A potentiometric acid-base titration employs an indicator electrode (such as a pH sensor) to establish the end point of the titration. The HI 84430 minititrator is designed to determine the acidity of waters by potentiometric titration in a low range (15-500mg/mL as CaCO Both Strong and Total acidity can be determined by the HI 84430, however the actual sample pH may limit your choice. If the measured pH is lower than 3.7 both total and strong acidity can be determined. If the sample pH is greater than 3.7 only total acidity can be determined. The HI 84430 utilizes two different titrant solutions to cover the entire measurement range. The titrants used have different concentrations, appropriate for each work range. Low Range measurements use HI 84430-50 titrant. High Range measurements use HI 84430-51 titrant. If the range is changed, the titrant and the tube set must be changed and the peristaltic pump must be purged and recalibrated. The standard solution (HI 84430-55M) is used for pump calibration for both working ranges, but the pump calibration solution volume used is different.
Titratable Acidity in waters, as performed on the HI 84430 minititrator, utilizes a simple sample preparation, a high quality peristaltic dosing pump for titrant, potentiometric endpoint detector and instantaneaous computations. To maintain the high precision of the titrator, a simple pump calibra­tion procedure is required. A range change requires different titrant and a pump recalibration. The pump calibration involves the analysis of a known volume of a known solution (standard provided) and compensates for changes in pump dosing that may occur due to many factors including tube stretching or aging. This procedure should be performed regularly.
) and in a high range (400-4000mg/mL as CaCO3).
3
FUNCTIONAL AND PHYSICAL DESCRIPTION
OVERHEAD VIEWOVERHEAD VIEW
OVERHEAD VIEW
OVERHEAD VIEWOVERHEAD VIEW
1. Titrant bottle
2. Graphic LCD
3. Keypad
4. Dedicated guide tube for Temperature Probe
5. Electrode holder
6. Dedicated guide tube for titrant tip
7. Peristaltic pump tube
REAR VIEWREAR VIEW
REAR VIEW
REAR VIEWREAR VIEW
8. Temperature probe
9. pH Electrode
10. Beaker
11. BNC electrode connector
12. Temperature probe socket
13. USB connector
14. Power switch
15. Power adapter connector
16. Peristaltic pump
FRONT VIEWFRONT VIEW
FRONT VIEW
FRONT VIEWFRONT VIEW
98
KEYPAD FUNCTION AND INDICATORS
ESC - used to leave the current screen and to return either to the previous screen or to the
main screen. When pressed while modifying a parameter within the SETUP menu, exits parameter without changing it.
/
- used to modify the parameters’ values, to scroll the information displayed while
viewing a help or to move between the options from the instrument’s SETUP
HELP - used to access/leave the instrument’s contextual help MENU - used to enter SETUP, Recall or GLP selection menu, while instrument is in pH or
Titration main screen STIR - used to start/stop the stirrer. Note: The stirer starts automatically during pump calibration and titration and cannot be stopped by pressing STIR key.
GUIDE TO INDICATORS
During the instrument’s operation a set of information is displayed on the LCD. Displayed icons:
Unstable reading. Stirrer on.
1
2
3
4
5
Pump running.
Parameter can be changed.
9
8
7
6
5. Functional key area
6. Indicates that the displayed value can be changed using ARROW keys
7. Temperature value (
°C, °F)
8. pH temperature compensation mode (Manual or Automatic)
9. Stirrer and reading status area
PERISTALTIC PUMP
Peristaltic pumps are self priming. Liquid never contacts the pump components. The titrant tubing is pressed along the rotating rollers of the pump. The rollers compress the tubing, driving the titrant to the dispensing tip.
TITRATOR STARTUP
This is a general outline of the steps required to make a titration. The following sections expand upon each section.
• Place the instrument on a flat table. Do not place the instrument in direct sun light.
• Connect the power adapter to the instrument.
• Turn the instrument ON using the power switch from the rear panel of the instrument.
• Set up the instrument. See the “Setup Configuration Menu” section for details. Pay attention to the Acidity Type.
• Connect the pH electrode and the temperature probe to the instrument.
• Calibrate the pH electrode. At least a single point calibration is necessary for titration.
• Place the peristaltic pump tube on the pump (inlet tube is connected with the reagent bottle, outlet tube is connected with the dosing tip). See the “Pump Tube Replacement” section for the procedure.
• Remove the reagent bottle cap and replace the bottle cap with the tubes. Place the reagent bottle in the appropriate place on the titrator top.
• Purge the titrant.
• Calibrate the pump.
• Prepare the sample.
• Run a titration and log sample results.
1. Current time and instrument mode information (pH meter or Titrator)
2. pH electrode condition information
3. Main reading information
4. Instrument status information
1110
SETUP CONFIGURATION MENU
Calibration timeout
The titrator’s setup configuration menu may be accessed from the pH or titration screens by pressing the MENU key, then Setup.
A list of setup parameters will be displayed with currently configured setting. While in the setup menu it is possible to modify the instrument’s operation parameters. The
ARROW keys permit the user to scroll the setup parameters. Press HELP to view the contextual help. Press ESC to return to the main screen.
Concentration unit
Options: mg/L, meq/L.
Press the corresponding function key to change the option.
Acidity type
Options: Total LR (total acidity low range)
Total HR (total acidity high range) Strong LR (strong acidity low range) Strong HR (strong acidity high range)
Press the corresponding function key to change
the option. From your knowledge of expected concentrations, use the table below to determine which settings are appropriate:
epyTytidicARLlatoTRLgnortSRHlatoTRHgnortS
OCaCL/gmegnartnemerusaeM
3
OCaCL/qem
3
)M55-03448IH(noitulosnoitarbilacpmuP Lm2Lm2Lm02Lm02
deriuqertnartiT 05-03448IH05-03448IH15
)Lm(eziselpmaS 05050505
elpmasfoHpmuminiM 3.8<7.3<3.8<7.3<
noitartitfotniopdnE Hp3.8Hp7.3Hp3.8Hp7.
005-51005-510004-0040004-004
00.01-3.00.01-3.08-808-8
-03448IH15-03448IH
3
pH resolution
Temperature unit
Backlight
Options: Disabled or 1 to 7 days.
This option is used to set the number of days before the pH calibration expired warning message is flagged.
Press Modify to access the calibration timeout value modify parameter.
Use the ARROW keys in order to increase/decrease the value.
Press Accept to confirm or ESC to return to the setup menu without saving the new value.
Options: 0.1, 0.01. Press the displayed function key in order to change
the pH resolution.
Options:
°
C, °F.
Press the function key in order to change the temperature unit.
Options: 0 to 8. Press Modify to access the backlight level.
Use Total LR or Strong LR for waters acidity in the 15 - 500 mg/L range. Use Total HR or Strong HR for waters acidity in the 400 - 4000 mg/L range. Use the appropriate titrant for each range. Also, the pump must be recalibrated when the measurement range is changed.
1312
Use the ARROW keys or increase/decrease the displayed constrast.
Press Accept to confirm or ESC to return to the setup menu.
/ in order to
Time format
Option: AM/PM or 24 hours.
Press the functional key to select the new value.
Contrast
Date / Time
Option: 0 to 20.
This option is used to set the display’s contrast. Press Modify to change the display’s contrast.
Use the ARROW keys or increase/decrease the value.
Press Accept to confirm the value or ESC to return to the setup menu.
This option is used to set the instrument’s date and time.
Press Modify to change the date/time.
Press / to highlight the value to be modified (year, month, day, hour, minute or second). Use the ARROW keys to change the value.
Press Accept to confirm the new value or ESC to return to the setup.
/ in order to
Date format
Press Modify to change the Date Format.
Use the ARROW keys to select the desired format. Press Accept to confirm the value or ESC to
return to the setup menu.
Language
Press the corresponding function key to change the language.
If the new selected language cannot be loaded, the previously selected language will be reloaded.
If no language can be loaded at startup the instrument will work in the “safe mode”. In “safe mode” all the messages are displayed in English and tutorial and help information are not available.
Tutorial
This option is used to enable/disable tutorial mode. If enabled this option will provide the user short
guides on the screen. Press the function key to select this option.
1514
Beep On
Instrument ID
Press the function key to select the new option. When enabled, a short beep is heard every time a key is pressed or when the calibration can be confirmed.
A long beep alert sounds when the pressed key is not active or a wrong condition is detected while in calibration.
Option: 0 to 9999. This option is used to set the instrument’s ID. The
instrument ID is used while exchanging data with a PC.
Press Modify to access the instrument ID screen. Press ARROW in order to select the desired value
between 0 and 9999. Press Accept to confirm the value or ESC to return to the setup menu without saving the new value.
Meter information
Press Select to view the firmware version, language version, mV factory calibration date and time and temperature factory calibration date and time.
Press ESC to return to the Setup mode.
Baud rate
Options: 600, 1200, 2400, 4800, 9600, 19200, 38400 bps.
This parameter is used for setting the speed of the serial link between the instrument and a PC. Both must use the same setting.
Press Modify to access the baud rate screen. Use
ARROW to scroll through baud rate values and press Accept to confirm the selection or ESC to return to
the setup menu without saving the new value.
1716
ELECTRODE PREPARATION
ELECTRODE CALIBRATION PROCEDURE
PREPARATION PROCEDURE
Remove the electrode protective cap. DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and they will disappear when rinsed with distilled water. During transport tiny bubbles of air may have formed inside the glass bulb. The electrode cannot function properly under these conditions. These bubbles can be removed by "shaking down" the electrode as you would do with a glass thermometer. If the bulb is dry, soak the electrode in HI 70300 Storage Solution for at least one hour.
It is recommended to calibrate the instrument frequently, especially if high accuracy is required. The pH electrode should be recalibrated:
a) Whenever the pH electrode is replaced b) At least once a week c) After testing aggressive chemicals and after electrode is cleaned d) When high accuracy is required
e) If the pH calibration expired warning is displayed during measurement. Every time you calibrate the instrument clean the electrode (see the “pH Cleaning Procedure” section) and use fresh buffers.
PROCEDURE
A single, two or three-points calibration can be performed, using the three predefined buffers
1.68, 4.01 and 8.30 pH. For a single point calibration any of the three buffers may be used, but using 8.30 pH is recommended. Note: The HI 84430 will not accept other pH buffers for calibration.
Pour small quantities of selected buffer solutions into clean beakers. For accurate calibration
use two beakers for each buffer solution, the first one for rinsing the electrode and the second one for calibration.
• Put a magnetic stir bar in each beaker with the calibration buffer solution.
Remove the protective cap, open the fill hole and rinse the electrodes with some of the buffer
solution to be used for the first calibration point.
• Put the first beaker with calibration buffer in the beaker holder.
• Place the electrode holder on the top of the beaker and secure it by turning clockwise.
Insert the temperature probe into the dedicated tube guide.
Immerse the pH electrode and the temperature probe approximately 2 cm (0.8”) into the buffer paying attention not to touch the stir bar.
To select Electrode calibration screen follow the next steps:
• From pH meter
• From Titrator
• The electrode calibration screen will be displayed.
screen press CAL function key then Electrode.
screen press CAL function key then Electrode.
Point 1 calibration
The 8.30 buffer will be selected by default. If necessary press the ARROW keys in order to
select a different buffer value.
The (unstable measurement) symbol will be shown on the display until the reading becomes stable.
1918
When the reading is stable and close to the selected buffer, the (unstable measurement)
symbol will disappear and the Confirm key will become active.
Press Confirm to confirm the calibration.
Press ESC to exit calibration.
The calibrated value will be shown on the display and the third expected buffer value will be
automatically selected.
After the second calibration point has been confirmed, press ESC to exit without performing the
third calibration point.
Point 3 calibration
• Remove the electrode holder with electrodes from the top of the beaker.
• Place the third buffer solution in a beaker and place in beaker holder. Rinse the probes in a
beaker with third buffer rinsing solution.
• Place the electrode holder (with electrode) in the beaker with third buffer and secure top by
locking. Press STIR.
ARROW
• If necessary press the
The (unstable measurement) symbol will be shown on the display until the reading becomes stable.
• When the reading is stable and close to the selected buffer, the
symbol will disappear and the Confirm key will become active.
keys in order to select correct buffer value.
(unstable measurement)
Notes:
Point 2 calibration
• Remove the electrode holder with electrodes from the top of the beaker.
To clear a previous calibration and to return to the default value, press Clear at any time after entering calibration mode. The “Calibration cleared” message will be shown for a few seconds on the display. If Clear is invoked during the first calibration point the instrument returns to the measurement mode.
• The Clear key is displayed only if a previous calibration exists.
The calibrated value will be shown on the display and the second expected buffer value will be displayed.
Place the second buffer into the beaker and place in beaker holder. Rinse the electrodes in a beaker containing the second buffer rinsing solution.
Place the electrode holder (with electrodes) on the top of the beaker, lock cap by turning.
If necessary press the ARROW keys in order to select a different buffer value. The (unstable measurement) symbol will be shown on the display until the reading becomes stable. When the reading is stable and close to the selected buffer, the (unstable measurement)
symbol will disappear and the Confirm key will become active.
Press Confirm to confirm the calibration.
• Press Confirm to confirm the calibration. The instrument stores the calibration value and returns to pH
meter/titrator calibration menu, where the date and time for the last pH calibration will be updated.
Notes:
A buffer confirmed during the calibration process is removed from the list of calibration buffers
available for further calibration points.
If the value measured by the instrument is not close to the selected buffer a “Wrong Buffer”
error message will be shown on the display.
Check if the correct buffer has been used or regenerate the pH electrodes by following the Cleaning Procedure (see the “pH Cleaning Procedure” section). If necessary change the buffer or the electrode.
2120
If the measured offset isn’t within the preset limits the meter will display the message “Buffer
Contaminated” alternatively with ”Electrode Dirty/Broken”.
If the computed slope isn’t within the preset limits the meter will display the message “Wrong
Slope”. If the slope is too high the symbol symbol
If the “Wrong Old Slope” error message is displayed, an inconsistency exists between the
will be displayed.
will be displayed. If the slope is too low the
current and the previous (old) calibration. Clear the calibration parameters by pressing Clear and proceed with calibration from the current calibration point. The instrument will keep all the
pH BUFFER TEMPERATURE DEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. During calibration the instrument will automatically calibrate to the pH value corresponding to the measured or set temperature.
During calibration the instrument will display the pH buffer value at 25
PMETSREFFUBHp
C°F°86.110.403.8
023 76.110.484.8
514 76.100.444.8
0105 76.100.414.8
5195 76.100.473.8
0286 86.100.433.8
5277 86.110.403.8
368 86.120.472.8
0
5359 96.130.442.8
04401 96.140.412.8
°C.
confirmed values during the current calibration point.
If the temperature reading is out of the defined temperature range of the buffer (0 to 45°C)
°
the “Wrong Buffer Temperature” error message will be displayed, and the symbol
C will
blink on the display. Calibration cannot be confirmed in this situation.
2322
PUMP TUBE INSTALLATION
To mount the new peristaltic pump tube set follow next steps:
• Position one peristaltic pump fixing ring on its location.
• Stretch the tube over the peristaltic pump rolls.
• Attach the second pump fixing ring on its location.
• Attach the tube to the reagent bottle. Choose which titrant is required for expected range. See table in Acidity Type section.
Note: Purge the peristaltic pump until drops of reagent appears on the dosing tip by pressing the
PURGE key from the titrator main screen.
To remove the tube of the peristaltic pump follow next steps: Caution: Purge line with water to remove titrant solution from tube.
• Detach the tubes system from the reagent bottle.
• Grasp one fixing ring of the peristaltic pump tube.
• Pull the tube until the fixing rings are taken out from their location.
• Remove the other side of the tube.
PURGE
Purging should be performed:
• whenever the tube of the peristaltic pump is replaced;
• whenever the titrant is changed or a new bottle of titrant is used;
• before starting a pump calibration;
• before starting a lot of titrations.
In order to start purging press the Purge key from the titrator main screen. The purging stops automatically after 5 minutes. To access the Purge key follow the next steps.
• From the instrument main screen (pH meter screen) press “Titrator” function key.
Press CAL to access the titration calibration menu where electrode and pump calibration may be accessed.
Press HELP to view the contextual help.
Press Continue or ESC to skip the message and enter titrator main screen.
Press Purge to begin a purge cycle.
The purging stops automatically after 5 minutes. To stop purging at any time and return to the main screen press ESC or Stop. During a purge, a countdown timer displays remaining time in 5 minute purge.
• Press Pause to interrupt the purge process.
• Press Pause or Stop (by pressing the corresponding function key in the purge screen)
• after the first drops of fresh titrant appear at the dosing tip
• in case an error condition is observed (empty titrant bottle, tubes or dosing tip disconnected, pump error)
• if you want to resume at a later time
If Pause is pressed the next screen is displayed:
The instrument will display the next screen if tutorial is enabled and if any of the following conditions exist:
- the meter hasn’t been calibrated in 8.30 pH buffer
- the pH calibration has expired
- a pump calibration hasn’t been performed or more than 3 days have passed since the last pump calibration
Press Resume to continue purging.
2524
After the 5 minutes purging interval has elapsed the “Completed” message is displayed. Another purge period can be initiated by pressing Restart or press Esc to return to main titrator screen.
PUMP CALIBRATION PROCEDURE
Note: The chemical reagents may be hazardous if improperly handled. Read Health and Safety Data Sheets before performing the test.
• From the titrator main screen press CAL.
Please set up the instrument before performing the pump calibration.
Select the acidity type in SETUP menu to correspond with expected sample:
TINU
L/gm 0.005ot0.510004ot004
L/qem 0.01ot3.008ot8
egnaRwoL
)elpmasLm05(
egnaRhgiH
)elpmasLm05(
The calibration of the pump must be performed each time the pump tube, the reagent bottle or the pH electrode is changed. It is recommended to perform the pump calibration before each set of titrations.
Verify: The electrode has been calibrated in 8.30 pH buffer.
Sample preparation:
Precisely measure the amount of HI 84430-55 Pump Calibration
Solution as indicated below and add to a clean beaker:
Low Range (Strong LR, Total LR) - 2 ml High Range (Strong HR, Total HR) -20 ml
Fill the beaker up to the 50 mL mark with the distilled or deionized water. Place the stir bar into the beaker and then place the beaker in the appropriate place on the instrument top.
• Place the electrode holder on the top of the beaker and secure it by turning clockwise.
• Immerse the pH electrode approximatively 2 cm (0.8”) into the standard paying attention not to touch the stir bar.
• Insert the temperature probe into the dedicated guide tube.
• Verify if the titrant correspondes to selected titration range. For Low Range (Strong LR, Total LR) use the HI 84430-50. For High Range (Strong HR, Total HR) use the HI 84430-51.
• Carefuly insert the dosing tip into the dedicated guide tube.
The instrument displays the date and time of the last electrode calibration, and the date and time of the last pump calibration, or calibration expired messages.
Press Pump.
• The next screen will be displayed.
Press Start.
After the pump calibration is started, on the upper right side of the display two animations will be
shown in order to indicate that the pump and the stirrer are working. On the lower right side of the display is shown the amount of time that has passed since the beginning of the calibration.
2726
After the pump calibration is complete a confirmation message is displayed for a few seconds,
then the instrument will return to the titrator calibration menu and will display the new time and date for the last pump calibration.
Notes:
The calibration of the pump is dependent on acidity range selected during SETUP.
If an erroneous situation is encountered during the calibration, an error message is
displayed and the calibration can be restarted by pressing Restart.
If the calibration doesn’t complete within 6 minutes the error message “Too much
standard” will be displayed and the calibration can be restarted by pressing Restart after a new standard is prepared.
TITRATION PROCEDURE
Verify: The instrument has been calibrated (pH and pump) before performing a lot of titrations.
An electrode calibration in at least one point (8.30 pH buffer) is recommended.
Sample preparation: For all measurement samples, pour 50 mL of sample into the beaker provided. Place the stir bar into the beaker and then place the beaker in the appropriate place on the instrument top.
Notes: • Water samples must be collected and stored in
capped botlles. Avoid disturbing any sediment in the sample.
• For samples containing metallic ions (iron, aluminium, chromium, manganesse) it is neccesary to add few drops (4-5) of an additional reagent (HI 84430-58) in order to increase the rate of hydrolysis at the metal ions present. Boil the solution 2-4 minute and cool to room temperature before titration.
• Place the electrode holder on the top of the beaker and secure it by turning clockwise.
• Insert the temperature probe into the dedicated guide tube.
• Immerse the pH electrode approximatively 2 cm (0.8”) into the sample to be tested paying attention not to touch the stir bar. Use O-Rings provided to secure the pH electrode in holder.
• Insert the dosing tip in the appropriate holder place.
2928
• From the titrator main screen press Titration. To enter titrator main screen from pH meter mode press Titrator and then Titration. If the tutorial is enabled this screen will be shown.
Press Start to begin the titration process.
After the titration is started two animations will be shown on the upper right side of the display
to indicate that the pump and the stirrer are running. On the lower right side of the display is shown the period of time since the titration has been started.
After the titration is complete, the concentration value is displayed in the selected unit.
Note: If the end-point is not reached or it is not recognized or the input reading is out of range,
an error message will be displayed. The titration can be restarted after a new sample is prepared by pressing Restart.
TIPS FOR AN ACCURATE MEASUREMENT
• Calibrate the instrument in 8.30 pH buffer solution at least once a day, before you start to perform measurements.
Purge the peristaltic pump to have the fresh titrant when starting a new calibration.
Calibrate the peristaltic pump daily before performing a set of analyses.
• Clean the electrode in order to remove the possible coating from bulb.
• Precision of the measurement can be improved by using volumetric pipets for standard and sample additions.
VIEW/DELETE TITRATOR RECORDED LOG DATA
Press MENU key while in Titrator main screen.
Press Unit to change the display unit.
Press LOG to record the concentration value into the instrument’s memory.
A message will be displayed for a few seconds indicating the amount of the free log space. 50 log samples can be recorded in the instrument’s memory. When the titrator free log space is under 12% the message will be shown permanently.
If the concentration is out of limits an exceeded range limit message will be displayed blinking
and the message “Out of range” will be shown. Another titration can be initiated by pressing Restart. Prepare a fresh sample and then press Restart.
Press Recall to access the titrator recorded data. The instrument will display a list of all the titration records stored in the titration log. Use the ARROW keys to scroll the stored records list. If the saved concentration was out of range the “!” symbol is displayed in front of the reading.
3130
Press Delete to enter record deleting mode. Press Delete All to enter all records deleting mode. Press More to view more information.
Press Unit to convert the result to other unit.
Use the ARROW keys when
is displayed to scroll between the log records.
Press ESC to return to the previous screen. If Delete was pressed the instrument will ask for confirmation.
Use the ARROW keys to focus on the record to be deleted. Press Confirm to delete the record or ESC to return to the previous screen. Deleting a record will renumber the list of records.
If Delete All was pressed the instrument will ask for confirmation.
TITRATOR GLP INFORMATION
Press MENU while in Titration mode and then GLP.
From this screen it is possible to select between viewing the electrode GLP or the pump GLP. Press GLP elec. to view the electrode’s last calibration parameters and date. Press GLP pump to view the pump’s last calibration time and date. If GLP elec. is pressed one of the next screens will be displayed.
GLP contains a set of information regarding electrode calibration. The following items are included in electrode GLP: the time and date of the last calibration, offset, slope, electrode condition, calibration timeout and the calibration buffers. The buffers displayed in video inverse mode are from the previous calibration.
If a calibration hasn’t been performed the message “Not Calibrated” will be displayed.
Press Confirm to delete all the records or ESC to return to the previous screen.
If the titrator log is empty the message “No Records!” will be displayed.
If GLP pump is pressed, one of the next screens is displayed.
3332
The pump GLP displays the Time and Date of the last pump calibration. If a calibration hasn’t been performed the message “Not Calibrated” will be displayed.
pH MEASUREMENT
The HI 84430 may be used as a pH meter for direct measurements. Verify that the instrument has been calibrated before taking pH measurements. Set the instrument
to pH meter. At power up the instrument enters pH meter mode. From titrator mode press ESC until pH units are displayed.
If an electrode calibration hasn’t been performed, or the number of days exceedes calibration time out value set, the message “CAL DUE” will blink on the left side of the display (see Calibration timeout option in Setup for details).
If CAL DUE is displayed perform an electrode calibration. Place calibrated pH electrode into electrode holder. Open the fill hole. Rinse the pH tip with distilled or deionized water. Immerse the pH electrode (bottom 2 cm / 0.8”) in the sample and stir gently for a few seconds. For a faster response and to avoid cross-contamination of the samples, rinse the electrode tip with a few drops of the solution to be tested, before taking measurements. Press LOG to record sample pH in instrument pH log.
Press MENU to access the instrument’s menu.
Press LOG to memorize the current reading. A message indicating the free log space will be displayed for a few seconds.
In order to take pH measurements follow the next steps:
Submerge the pH bulb 2 cm (0,8’’) and the temperature probe into the sample to be tested
and stir gently. Allow time for the electrode to stabilize. When the reading becomes stable the
(unstable measurement) symbol will disappear.
If the pH reading is less then -2.00 pH or greater then 16.00 pH the closest full-scale value
(-2.00 pH or 16.00 pH) will be displayed blinking.
If measurements are taken successively in different samples, it is recommended to rinse the electrodes thoroughly with deionized water or distilled water and then with some of the next sample to prevent cross-contamination.
The pH measurement is affected by temperature. In order to have accurate pH measurements, the temperature effect must be compensated for. To use the Automatic Temperature Compensation (ATC) feature, connect and submerge the HI 7662-M temperature probe into the sample in the dedicated tubeguide and wait for a few seconds. The “ATC” message will be shown on the display. Automatic Temperature Compensation will provide pH corrected values at the temperature of measurements. If Manual Temperature Compensation (MTC) is desired, the temperature probe must be disconnected from the instrument. The default temperature of 25
temperature reading will be displayed preceded by the symbol
°
C (77 °F) or the last
and the “MTC” message.
Press HELP to view the contextual help, every time you need supplementary information. The help is customized for every situation that can appear during instrument usage.
Press STIR to start/stop the stirrer. Press Titrator to enter titrator mode. Press CAL to access the calibration menu.
The manually set temperature can now be adjusted with the ARROW keys (from -20.0 to 120.0
3534
°
C).
VIEW/DELETE RECORDED pH DATA
To view or delete previously logged pH records press MENU key while in pH meter screen.
Press Recall to access the pH recall. A list of records is stored in the pH log.
Use the ARROW keys to scroll the list of records. Press More to see detailed information about the highlighted record. Press Delete to enter record deleting mode. Press Delete All to enter all records deleting mode. If More is pressed a complete set of data is displayed.
If Delete All was pressed the instrument will ask for confirmation.
Press Confirm to delete all records or ESC to return to the previous screen without deleting. If the pH log is empty the message “No Records!” will be displayed.
pH METER GLP INFORMATION
The pH meter GLP screens references the last pH calibration data. GLP is an acronym for Good Laboratory Practice.
To view this information, press MENU key then GLP. A set of information regarding electrode calibration is displayed.
Use ARROW keys when
is displayed to scroll between the records.
If Delete was pressed the instrument will ask for confirmation.
Use the ARROW keys to focus on the record to be deleted. Press Confirm to delete the record or ESC to return to the previous screen without deleting. Deleting a record will renumber the list of records.
The following items are included in electrode GLP: the time and date of the last calibration, offset, slope, electrode condition, calibration timeout and the calibration buffers. The buffers displayed in video inverse mode are from the previous calibration.
If a calibration hasn’t been performed the message “Not Calibrated” will be displayed.
3736
TEMPERATURE CALIBRATION PROCEDURE (for technical personnel only)
All the instruments are factory calibrated for temperature. HANNA’s temperature probes are interchangeable and no temperature calibration is needed when
they are replaced. If the temperature measurements are inaccurate, temperature recalibration should be performed. For an accurate recalibration, contact your dealer or the nearest HANNA Customer Service Center,
or follow the instructions below.
Prepare a vessel containing ice and water and another one containing hot water (at a
temperature of around 50 temperature changes.
Use a calibrated thermometer with a resolution of 0.1 °C as a reference.
To enter user calibration screen press and hold down the ARROW keys simultaneously, then
power on the instrument. After a few seconds the User calibration screen is displayed.
Press Temp function key to enter temperature calibration.
Immerse the temperature probe in the vessel with ice and water as near as possible to the
reference thermometer. Allow a few seconds for the probe to stabilize.
Use the ARROW keys to set the calibration point value to that of the ice and water measured
by the reference thermometer.
The (unstable measurement) symbol will be shown on the display until the reading becomes
stable.
When the reading is stable and close to the selected calibration point, the (unstable
measurement) symbol will disappear and the Confirm key will become active.
°
C). Place insulation material around the vessels to minimize
Immerse the temperature probe in the second vessel as near as possible to the reference
thermometer. Allow a few seconds for the probe to stabilize.
Use the ARROW keys to set the calibration point value to that of the hot water, measured by
the reference thermometer.
The (unstable measurement) symbol will be shown on the display until the reading becomes stable.
When the reading is stable and close to the selected calibration point, the (unstable
measurement) symbol will disappear and the Confirm key will become active.
Press Confirm to confirm the calibration point. The instrument will return to the pH meter/
titrator main screen.
Note: If the reading is not close to the selected calibration point, the “Wrong” message will be
displayed. Change the temperature probe and restart the calibration.
If the temperature probe is disconnected or the measured temperature is out of the - 20 to120 the instrument will display “----”. The calibration point value can be changed using the ARROW keys.
°
C range
Press Confirm to confirm the calibration point.
The meter will be automatically move to the second calibration point, and will display 50 °C for
the buffer value.
3938
PC INTERFACE
Data transmission from the instrument to the PC can be done with the HI 92000 Windows compatible software (optional). HI 92000 also offers graphing and on-line help features.
Data can be exported to the most popular spreadsheet programs for further analysis. To connect your instrument to a PC use a standard USB connector. Make sure that your instrument
is switched off and plug one connector to the instrument’s USB socket and the other to the USB port of your PC.
If you are not using Hanna Instruments HI 92000 software, please see the following instructions.
SENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any terminal program. Start the terminal program and set the communication options as follows: 8, N, 1, no flow control.
COMMAND TYPES
To send a command to the instrument follow the next scheme:
<command prefix> <comand><CR>
where <command prefix> is the ASCII character having the code 16.
<command> is the command code.
Note: Either small or capital letters can be used. Upon reception of a command the instrument will answer with:
<STX><answer><checksum><ETX>
where the checksum is the bytes sum of the answer string, sent as 2 ASCII characters. All the answer messages are with ASCII characters.
COMMANDS
RAS Causes the instrument to send a complete set of readings in according with the
current meter mode. The answer string contains:
-Meter mode (2 chars): -00 pH meter (0.1 resolution)
-01 pH meter (0.01 resolution)
-02 titrator
-Meter status (2 chars) represents a 8 bit hexadecimal encoding.
-0
x
10: temperature probe connection
-0
x
01: new GLP data available
-0
x
02: new setup parameter
-Reading status (1 char): R-in range, O - over range, U - under range.
-pH reading with sign and decimal point (7 chars)
-Temperature, with sign and 2 decimals point, always in
-Checksum (2 chars) Note: If the meter is in titrator mode only, the meter status and the
checksum will be sent.
MDR Requests the instrument model name and firmware code (20 chars). GLP Requests te calibration data record.
The answer string contains:
-GLP status (1 char): represents a 4 bit hexadecimal enconding.
-0
x
1-pH calibration available
-0
x
2- pump calibration available
-pump calibration data (if available):
-calibration time, yymmddhhmmss (12 chars)
-pH calibration data (if available):
-the number of calibrated buffers (1 char)
-reserved (2 chars)
-the offset, with sign and decimal point (7 chars)
-the average of slopes, with sign and decimal point (7chars)
-the calibration time: yymmddhhmmss (12 chars)
-buffer information (for each buffer)
-type (1 char): O-standard
-status (1 char): N (new)-calibrated in last calibration O (old)-from an old calibration
-warning during calibration (2 chars): 00- no warning, 04- Clean Electrode warning
-buffer value, with sign and decimal point (7 chars)
-calibration time, yyymmddhhmmss (12 chars)
-electrode condition, with sign (3 chars). The “01” code means not calculated.
-Checksum (2 chars)
°
C (7 chars).
4140
PAR Requests the setup parameters setting.
The answer string contains:
-Instrument ID (4 chars)
-Calibration alarm time out (2 chars)
-Concentration unit (1 char): 0=mg/L, 1=meq/L
-Acidity type (1 char): 0=Total LR, 1=Total HR, 2=Strong LR, 3=Strong HR
-SETUP information (2 chars): 8 bit hexadecimal encoding
-0
x
01-beep ON (else OFF)
-0
x
04-degree Celsius (else degrees Fahrenheit)
-Backlight value (1 char)
-Contrast value (2 chars)
-Time format (1 char): 0=24 hours, 1=AM/PM
-Date format (1 char): 0-DD/MM/YYYY 1-MM/DD/YYYY 2-YYYY/MM/DD 3-YYYY-MM-DD 4-Mon DD, YYYY 5-DD-Mon-YYYY 6-YYYY-Mon-DD
-Language name (3 chars, “sfm” when the instrument is in safe mode)
-Checksum (2 chars)
-Overrage for the measured value: R-in range, O-over range, U-under range.
-Measured value (7 chars): for the pH meter log this is the measured pH value; for the titrator log this is the concentration value
-Temperature reading, with two decimal points (7 chars) (only for pH meter log)
-Calibration offset (7 chars) (only for pH meter log)
-Calibration slope (7 chars) (only for pH meter log)
-Concentration unit (1 char) (only for titrator log): 0=mg/L, 1=meq/L
-Acidity type (1 char) (only for titrator log): 0=Total LR, 1=Total HR, 2=Strong LR, 3=Strong HR
-Logged time (12 chars): yymmddhhmmss
-Temperature probe presence (1 char)
-Checksum (2 chars)
Notes:
-”Err8” is sent if the instrument is not in measurement mode
-”Err6” is sent if the requested range is not available
-”Err4” is sent if the requested set parameter is not available
-”Err3” is sent if the Log on demand is empty
-Invalid commands will be ignored.
NSLx Requests the number of logged samples (4 chars).
The command parameter (1 char):
-P-request for pH meter
-T-request for titrator
LODPxxx Requests the xxxth pH record logged data LODTxxx Requests the xxxth titrator record logged data LODPALL Requests all the stored records from the pH meter log LODTALL Requests all the stored records from the titrator log
The answer string for each record contains:
-Logged mode (2 chars): -00-pH meter (0.1 resolution)
-01-pH meter (0.01 resolution)
-02-titrator
4342
TROUBLESHOOTING GUIDE
SMOTPMYSSMOTPMYS
SMOTPMYSSMOTPMYSMELBORPMELBORP
SMOTPMYS
MELBORPMELBORPNOITULOSNOITULOS
MELBORP
NOITULOSNOITULOS
NOITULOS
edortcelenekorB
SMOTPMYSSMOTPMYS
SMOTPMYSSMOTPMYSMELBORPMELBORP
SMOTPMYS
.tfird
2-,edom
.gniknilb
evissecxe/esnopserwolS
pusetautculfgnidaeR
.)esion(nwoddna
deyalpsidsiHp
L
.)ylno.noitcennocelbaC
gnidaerHpnielihW
00.61ro00.
MELBORPMELBORPNOITULOSNOITULOS
MELBORP
.edortceleHpytriD
.noitcnujytrid/deggolC
leveletylortcelewo
sedortceleHpelballifer(
.egnarfotuognidaeR
1607 03rofnoitulosgninaelc
se
tunim. llifhtiwllifeR
.noitulos
1607 03rofnoitulosgninaelc
.noitulos
retempacevitcetorpyfirevdna .ffosi
retempacevitcetorpyfirevdna .ffosieh
NOITULOSNOITULOS
NOITULOS
nipitedortceleehtkaoS IH
nipitedortceleehtkaoS IH
llifhtiw
llifeRsetunim
oitcennocelbackcehCotn
oitcennocelbackcehCotn
tfoytilauqehtkcehC
.sedortceleehtnaelC.elpmas
.noitulosllifhserfhtiwllifeR
egnaR tnemurtsnieht
ELECTRODE CONDITIONING AND MAINTENANCE
STORAGE PROCEDURE
To assure a quick response time, the glass bulb should be kept moist and not allowed to dry out.
ninoitartitaretfA hgiH
roL/gm0004syalpsid
ehthtiw(qem008
.gniknilb)tinudetceles
retemehtputratstA
ANNAHehtsyalpsid
ltnenamrepogol
.y
.deyalpsid
siegassem"xxrorrE"
.elpmas
.egnar
.s
tontnemurtsnI
gnorW.detarbilac
fotuonoitartnecnoC
.kcutssisyekehtfoenO
relanretnI
.ror
ehtnaelc/kcehC
sedortcele
.noitaraperp
etarbilaceR.
pmup(tnemurtsnieht
eracesU.)Hpdna
elpmasgnirud
rodraobyekehtkcehC
.rodnevehttcatnoc
dnaretemehtfforewoP
.niaganotirewopneht ,stsisreprorreehtfI .rodnevehttcatnoc
Replace the solution in the protective cap with a few drops of HI 70300 or HI 80300 Storage
tonseodretemehT
cca
reffubHpehttpe
.noitarbilacrofnoitulos
noitarbilacpmupehT
d
emrofrepebt'nac
nekorBHp.edortcele
.edortceleHpnekorB
ignibutyfireV
.gnibutpmupnekorB
detanimatnocrognorW
.noitulosnoitarbilacpmup
fnidetarbilac
roedortceleehtecalpeR
.rodnevehttcatnoc
dnatcatnis
.degrupnehwdessapnoitulos
noitarbilacpmupehtkcehC
siedortceleyfireV.noitulos
.sreffubHphser
,dradnatsrehtonaeraperP
tnartithserfevahotegrup
.noitarbilacehttratserdna
Solution. Follow the Preparation Procedure section before taking measurements.
Note:NEVER STORE THE pH ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect the electrodes and the cables. The cable used for connection to the instrument must be intact and there must be no broken insulation on the cable or cracks on the electrode stem or bulb. Connectors must be perfectly clean and dry. If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water.
tehT
."CTM"syalpsid
A woL
egnaR tnemurtsnieht
egnaR tnemurtsnieht
les
ninoitartitaretf
ehthtiw(L/qem51
ninoitartitaretfA hgiH
ehthtiw(L/qem8
sieborperutarepme retemehttub,detcennoc
putesreteMrofgnorw
roL/gm005syalpsid
.gniknilb)tinudetceles
roL/gm004syalpsid
.gniknilb)tinudetce
.elpmas
.egnar
.elpmas
.egnar
.dezylanaelpmasgnorW
.eborperutarepmetnekorB.eborperutarepmetecalpeR
PUTES othgiHdnaegnaR
f
otuonoitartnecnoC
rofgnorwderugifnocreteM
fotuonoitartnecnoC
egnahc.tnartitdnaemirP
lacer
PUT
ES dnaegnaRwoLot
.egnaResUgniruderac
.)pmup(
.tnartitegnahcdnaemirP
itaraperpelpmas
rednuepytytidicaegnahC
tnemurtsniehtetarbi
rednuepytytidicaegnahC
tnemurtsniehtetarbilacer
woLtceleS.)Hpdnapmup(
.no
pH CLEANING PROCEDURE
General
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled or deionized water and soak the electrode in HI 70300 or HI 80300 Storage Solution for at least 1 hour before use.
Soak in Hanna HI 7061 or HI 8061 General Cleaning Solution for approximately ½ hour.
Recalibrate electrode before taking measurements.
4544
ACCESSORIES
WARRANTY
REAGENTS
HI 84430-50 Titrant solution for low range (100mL) HI 84430-51 Titrant solution for high range (100mL) HI 84430-55M Pump calibration solution (230mL) HI 84430-58 Additional reagent (30mL) HI 84430-70 Reagents kit for low and high range (about 150 titrations) HI 84430-71 Reagents kit for low range (about 150 titrations)
HI 84430-72 Reagents kit for high range (about 150 titrations)
pH CALIBRATION SOLUTIONS
HI 7001M Buffer solution pH 1.68 (230mL) HI 7004M Buffer solution pH 4.01 (230mL) HI 7008M Buffer solution pH 8.30 (230mL) HI 7001M-6 Buffer solution pH 1.68 (6 x 230mL) HI 7004M-6 Buffer solution pH 4.01 (6 x 230mL)
HI 70083M-6 Buffer solution pH 8.30 (6 x 230mL)
ELECTRODES
HI 1131B pH Electrode HI 7662-M Temperature probe
ELECTRODE FILL SOLUTION
HI 7071 Filling solution for HI 1131B (30 mL)
ELECTRODE STORAGE SOLUTION
HI 70300M Storage Solution, 230 mL bottle
CLEANING SOLUTION
HI 7061M Electrode Cleaning Solution, 230 mL bottle
OTHER ACCESSORIES
HI 70483M Tube set with cap and tip for titrant bottle HI 731319 Stir bar 25 x 7 mm (10 pcs) HI 731342 2000μL fixed volume pipette HI 731352 Tip for 2000μL fixed pipette (4pcs) HI 731341 1000μL fixed volume pipette HI 731351 Tip for 1000μL fixed pipette (25pcs)
HI 84430 is warranteed for two years against defects in workmanship and materials when used for its
intended purpose and maintained according to the instructions. This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your dealer. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
RECOMMENDATION FOR USERS
Before using this product, make sure that it is entirely suitable for your specific application and for the environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipments, this requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument EMC performance.
To avoid damages or burns, do not put the instrument in microwave ovens. For yours and the instrument safety do not use or store the instrument in hazardous environments.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products
without advance notice.
4746
SALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTS
SALES AND TECHNICAL SERVICE CONTACTS
SALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (210) 4584.2941 • Fax (210) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and a complete list of Sales and Technical offices, please see
www.hannainst.com
.
48
MAN84430R1
03/08
Loading...