Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
Technical Support for
Customers
Tel. (800) 426 6287
Fax (401) 765 7575
Local Sales and Customer Service Office
HI 8424
Portable pH/mV/°C Meter
Printed in EUROPE
(ROMANIA)
www.hannainst.comMAN8424W R2 09/10
Dear Customer,
Thank you for choosing a HANNA instruments® product.
Please read this instruction manual carefully before using the instrument.
This manual will provide you with the necessary information for correct use of the
instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.comtech@hannainst.com
tech@hannainst.com or see the back cover for our worldwide contact list.
tech@hannainst.comtech@hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 8424 is guaranteed for two years against defects in workmanship and
materials when used for its intended purpose and maintained according to
instructions. Electrodes and probes are guaranteed for six months. This
warranty is limited to repair or replacement free of charge. Damage due to
accidents, misuse, tampering or lack of prescribed maintenance is not covered. If
service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial
number and the nature of the problem. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If the instruments are to
be returned to Hanna Instruments, first obtain a Returned Goods Authorization
number from the Technical Service department and then send it with shipping
costs prepaid. When shipping any instrument, make sure it is properly packed
for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent
of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
27
RECOMMENDATIONS FOR USERS
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residential
areas could cause unacceptable interferences to radio and TV equipment,
requiring the operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to electrostatic
discharges. Avoid touching this glass bulb at all times. During operation, ESD
wrist straps should be worn to avoid possible damage to the electrode by
electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade
the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 Vac or 60 Vdc.
To avoid damage or burns, do not perform any measurement in microwave
ovens.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to
make sure that no damage has occurred during shipment. If there is any
damage, immediately notify your dealer.
The meter is supplied complete with:
• HI 1230B pH electrode
• HI 7662 temperature probe
• pH 4.01 & pH 7.01 buffer solutions, 20 mL each
• HI 700661 cleaning solution, 20 mL sachet (2 pcs)
• 9V battery
• Instruction manual
Note: Save all packing material until you are sure that the instrument
functions correctly. Any defective items must be returned in the original
packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
HI 8424 is a portable microprocessor-based pH/mV/temperature meter.
It features an enhanced user interface, rainproof casing, battery percentage
indication, low battery detection, automatic shut-off, automatic calibration and
error codes to guide the user in calibration and troubleshooting.
The pH calibration procedure automatically recognizes three memorized buffer
values: pH 4.01, 7.01 and 10.01.
This instrument can also measure ORP. The resolution automatically switches
from 0.1 mV to 1 mV when readings exceed ±699.9 mV.
The user can set the following parameters by entering the setup mode through
the keyboard:
• enabling or disabling the auto-off feature
• enabling or disabling the acoustic signal
• selecting the temperature unit, °C or °F
26
3
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
1) BNC connector for pH or ORP electrodes
2) Liquid Crystal Display (LCD)
3) RANGE key, to select pH, mV or temperature range
4) Up Arrow key, to manually set the temperature value when no tempera-
ture probe is connected
5) CAL/SELECT key, to enter calibration mode or select menu options
6) CFM/SETUP key, to confirm data or enter/exit setup menu
7) Down Arrow key, to manually set the temperature value when no
temperature probe is connected
8) HOLD key, to freeze the reading on display
9) ON/OFF key, to switch the instrument ON or OFF
10) RCA socket for temperature probe
pp
H ELECTRODE APPLICATIONH ELECTRODE APPLICATION
p
H ELECTRODE APPLICATION
pp
H ELECTRODE APPLICATIONH ELECTRODE APPLICATION
REFERENCE GUIDEREFERENCE GUIDE
REFERENCE GUIDE
REFERENCE GUIDEREFERENCE GUIDE
Application Electrodes *
1. AquariumHI1332B, HI1312S
2. Bath-waterHI1130B, HI1110S
3. BeerHI1131B, HI1111S
4. BreadHI2031B, FC200B, HI2020S, FC200S
5. CheeseFC200B, FC200S
6. Dairy productsFC911B, FC100B
7. Dirty waterHI1230B, HI1210S
8. EmulsionsHI1053B, HI1050S
9. EnvironmentHI1230B, HI1210S
10. FlasksHI1331B, HI1310S
11. Food industry general useFC911B, FC100B
12. FruitFC200B, FC220B, FC200S
13. Fruit juices, organicFC210B
14. Galvanizing waste solutionHI1130B, HI1110S
15. High purity waterHI1053B, HI1050S
16. HorticultureHI1053B, FC200B, HI1050S, FC200S
17. Laboratory general useHI1131B, HI1230B, HI1332B, HI1330B
18. LeatherHI1413B, HI1410S
19. Lemon juiceFC100B
20. MeatFC200B, HI2031B, FC200S, HI2020S
21. Micro plate sampling of less thanHI1083B
100 mL
22. Milk and YogurtFC210B
23. PaintsHI1053B, HI1050S
24. PaperHI1413B, HI1410S
25. Photographic chemicalsHI1230B, HI1210S
26. Quality controlHI1332B, HI1312S
27. SausagesFC200B, HI2031B, FC200S, HI2020S
28. Semi-solid productsHI2031B, HI2020S
29. SkinHI1413B, HI1410S
30. Soil samplesHI1230B, HI1210S
31. SolventsHI1043B, HI1040S
32. Strong acidHI1043B, HI1040S
33. Submersion applicationHI1130B, HI1110S
34. Surface measurementsHI1413B, HI1410S
35. Swimming poolHI1130B, HI2114P/2
36. Titrations with constantHI1131B, HI1111S
temperature range
37. Titrations with wide temperatureHI1131B, HI1111S
range
38. Very high humidityFC911B
39. Vials and test tubeHI1330B, HI1310S
40. Wine processingFC220B
HI1111S, HI1210S, HI1312S, HI1310S
B = BNC-type connectorS = Screw-type connector
* All electrodes ending with "B" are supplied with 1 m (3.3') cable and BNC Connector
4
25
120 mm
4.7"
12 mm
0.5"
HI 4430
"S"VERSION
HI 4430B / HI 4410S
CONNECT TO
SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
Plastic body, gel-filled, combination gold ORP electrode.
Use: general purpose.
Consult the HANNA General Catalog for a complete and wide selection of
electrodes.
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES (SCREW
TO BNC CONNECTOR)
HI 7855/1Extension cable 1 m (3.3') long
HI 7855/3Extension cable 3 m (9.9') long
HI 7855/5Extension cable 5 m (16.5') long
HI 7855/10 Extension cable 10 m (33') long
HI 7855/15 Extension cable 15 m (49.5') long
OTHER ACCESSORIES
HI 98501
HI 98502
ChecktempC
ChecktempF
pocket-size thermometer (–50.0 to 150.0°C)
pocket-size thermometer (–58.0 to 302.0°F)
HI 710015Shockproof rubber boot, blue
HI 710016Shockproof rubber boot, orange
HI 710022Spare protective case
HI 76405Electrode holder
HI 7662Temperature probe with 1 m (3.3') screened cable
HI 8427pH/ORP electrode simulator with 1 m (3.3') coaxial cable and
BNC connector
HI 931001 pH/ORP electrode simulator with LCD, 1 m (3.3') coaxial cable
with 3 memorized buffer values (pH 4.01, 7.01, 10.01)
Offset: ±1 pH; Slope: from 75 to 110%
Temperature Compensation
Automatic, -20 to 120°C (-4 to 248°F)
or manual without temperature probe
Probes (included)
HI 1230B double junction, gel-filled pH electrode
HI 7662 temperature probe
Battery Type1 x 9V
Battery LifeApprox. 150 hours of continuous use
Auto-offAfter 20 minutes of non-use or disabled (user-selectable)
Environment0 to 50°C (32 to 122°F); RH max 100%
Dimensions164 x 76 x 45 mm (6.5 x 3.0 x 1.8'')
Weight180 g (6.3 oz.)
24
5
OPERATIONAL GUIDEOPERATIONAL GUIDE
110 mm
4.3"
12 mm
0.5"
HI 1413
150 mm
5.9"
12 mm
0.5"
"S"VERSION
HI 3131
120 mm
4.7"
12 mm
0.5"
HI 3230
"S"VERSION
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
FC 220
110 mm
4.3"
12 mm
0.5"
FC 911
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
INITIAL PREPARATION
• Remove the electrode protective cap before taking any measurements. If the
electrode has been left dry, soak the tip in HI 70300 storage solution for a few
hours or overnight to reactivate it.
• Connect the pH electrode to the BNC connector on the top of the instrument.
• Connect the temperature probe to the RCA connector. The temperature probe
can be used independently to take temperature measurements, or in
conjunction with the pH electrode to utilize the ATC capability of the meter.
• Turn the meter ON by pressing the ON/OFF key. The display shows all the
used segments for a few seconds (or as long as the button is held), followed
by the percentage indication of the remaining battery life, and then enters
normal measurement mode.
pH MEASUREMENTS
• To take a pH measurement simply submerge the
electrode tip (at least 4 cm/1½") and the temperature probe into the sample to be tested.
• Select the pH mode by pressing the RANGE key until
the display changes to pH.
• Stir gently and wait for the stability symbol (hourglass) to turn off. The display will
show the pH value automatically compensated for temperature.
• In order to take accurate pH measurements, make sure that the instrument
has been calibrated before use (see page 9).
• If measurements are taken in different samples successively, it is recommended to rinse the electrode thoroughly to avoid cross-contamination. After
cleaning, it is recommended to rinse the electrode with some of the sample to
be measured.
6
HI 3230B / HI 3210S
Plastic body, gel-filled, combination platinum ORP electrode.
Use: general purpose.
PVDF body, double junction, refillable, combination pH electrode.
Use: general purpose for food industry.
TEMPERATURE COMPENSATION
The meter is designed to compensate for temperature, as the response of the pH
electrode is directly affected by temperature.
Automatic Temperature Compensation (ATC shown on LCD)
To use the ATC feature, submerge the temperature probe into the sample as
close as possible to the electrode and wait for a few minutes. The displayed pH
reading is compensated for the temperature of the sample.
Manual Temperature Compensation (MTC shown on LCD)
If the temperature probe is not connected, it is possible to enter the temperature value manually.
• Record the sample temperature by using a ChecktempC (if you are
measuring temperature in °C, or ChecktempF for °F readings) or another
accurate thermometer.
• Press RANGE to select the temperature mode. The "°C" (or "°F") symbol will
blink to indicate that the temperature probe is not connected.
• Use the UP and DOWN keys to display and set
the sample temperature (e.g. 25 °C).
• Press RANGE to select the pH measurement mode and immerse the electrode
into the sample. The displayed pH reading will be temperature compensated
at the set value (in this case at 25 °C).
FC 200B/ FC 200S
PVDF body, single junction, conic, Viscolene, non-refillable, combination pH
electrode. Use: meat & cheese.
Oxidation Reduction Potential (ORP) measurements provide a quantification of
the oxidizing or reducing power of the sample tested.
• Connect the ORP electrode (optional) to the BNC connector.
• To enter the "mV" mode turn the instrument ON and press the RANGE key
until the display changes to mV.
• Submerge the ORP electrode tip (at least 4 cm / 1½") into the sample to be
tested and allow time for the reading to stabilize (hourglass symbol turns off).
• Measurements within the ±699.9 mV range
are displayed with 0.1 mV resolution, while
outside this range the resolution is 1 mV.
7
Notes:
120 mm
4.7"
5mm
0.2"
5mm DIA
0.2"
"S"VERSION
HI 1330
210 mm
8.25"
8 mm
0.3"
7.5mm DIA
0.29"
"S"VERSION
HI 1331
120 mm
4.7"
12 mm
0.5"
HI 1230
120 mm
12 mm
0.5"
5 mm
0.2"
3 mm
0.12"
3.0 mm DIA
0.12"
HI 1083
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S"VERSION
HI 1131
• To perform correct ORP measurements, the surface of the ORP electrode must
be clean and smooth.
• When not in use, the tip of the electrode should be kept moist (use HI 70300
storage solution) and safe from any mechanical stress which might cause
damage to the glass/platinum junction.
• Turn the instrument ON and press the RANGE key to select the temperature
mode.
• Make sure the temperature probe is connected to the meter.
• Dip the temperature probe into the sample, allow the reading to stabilize
(hourglass symbol turns off) and read the temperature value.
• Temperature measurements can be displayed in °C or °F units (see "Menu
selection" for details).
Notes:
• A blinking full scale value means that the reading is out of range.
• To freeze a reading on display while in measurement mode, press the HOLD key. The "HOLD"
tag will blink. The pH, mV and temperature
values are held, and the RANGE key can be
used to view the values. Press HOLD again to
return to normal mode.
• If enabled, keypresses are followed with an acoustic signal. A lower note
indicates that the key is not currently active.
• To save battery life, the meter is provided with an auto-off feature, which
turns the instrument off after 20 minutes of non-use. This feature can be
disabled by the user (see "Menu selection" for details).
HI 1131B / HI 1111S
Glass body, single junction, refillable, combination pH electrode.
Use: general purpose.
HI 70823.5 M KCl electrolyte solution, 4 x 30 mL bottle, for double
B = BNC CONNECTION PLUG + 1 m (3.3’) CABLE
S = SCREW-TYPE CONNECTOR
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S"VERSION
HI 1043
120 mm
4.7"
12 mm
0.5"
"S"VERSION
HI 1053
junction electrodes
HI 80823.5 M KCl electrolyte solution, 4 x 30 mL FDA bottle, for double
junction electrodes
HI 80931 M KCl+AgCl electrolyte solution, 4 x 30 mL FDA bottle, for
double junction electrodes
ORP SOLUTIONS
HI 7091MReducing pretreatment solution, 230 mL bottle
HI 7091LReducing pretreatment solution, 500 mL bottle
HI 7092MOxidizing pretreatment solution, 230 mL bottle
HI 7092LOxidizing pretreatment solution, 500 mL bottle
HI 7020MTest solution @200-275 mV, 230 mL bottle
HI 7020LTest solution @200-275 mV, 500 mL bottle
HI 7021MTest solution @240 mV, 230 mL bottle
HI 7021LTest solution @240 mV, 500 mL bottle
HI 7022MTest solution @470 mV, 230 mL bottle
HI 7022LTest solution @470 mV, 500 mL bottle
For better accuracy, frequent calibration of the instrument is recommended.
The instrument should be recalibrated for pH:
a) Whenever the pH electrode or temperature probe is replaced
b) At least once a week
c) After testing aggressive chemicals
d) When extreme accuracy is required
PREPARATION
Pour small quantities of pH 7.01 (HI 7007) and pH 4.01 (HI 7004) or pH
10.01 (HI 7010) buffer solutions into two clean beakers.
For accurate calibration use two beakers for each buffer solution, one for rinsing
the electrode tip, and one for calibration. In this way contamination of the
buffers is minimized.
For measurements in acidic samples, it is recommended to calibrate the meter
by using pH 7.01 (HI 7007) and pH 4.01 (HI 7004) buffers, while for
alkaline measurements use pH 7.01 (HI 7007) and pH 10.01 (HI 7010)
buffers.
PROCEDURE
• Connect the pH electrode and the temperature probe,
then switch the meter ON.
• Remove the electrode protective cap, rinse the electrode
tip with pH 7.01 solution, then immerse the pH
electrode and temperature probe into pH 7.01 buffer
solution; stir gently and wait a few minutes for the
electrode to stabilize and reach thermal equilibrium.
Note: The electrode should be submerged approximately 4
cm (1½") into the solution. The temperature probe
should be located close to the pH electrode.
• Press RANGE to display pH measurement.
• Press CAL to enter the calibration mode. The buffer value @25°C (77°F) and
the "pH" symbol will blink on the display.
20
• The meter expects a pH 7.01 buffer.
9
• When the buffer value is recognized and the
reading is stable, an acoustic signal (if enabled) advises the user, the "pH" symbol stops
flashing, the hourglass indicator disappears
and the "CFM" tag starts blinking to indicate
that the value can be confirmed.
• Press CFM to store the first calibration point.
• The meter expects a pH 4.01 or 10.01 buffer.
• Rinse and immerse the pH electrode and the
temperature probe in pH 4.01 or pH 10.01
buffer (2nd calibration point) and stir gently.
• When the buffer value is recognized and the
reading is stable, an acoustic signal (if enabled) advises the user, the "pH" symbol stops
flashing, the hourglass indicator disappears
and the "CFM" tag starts blinking to indicate
that the value can be confirmed.
• Press CFM to store the second calibration point.
• The meter returns to normal mode.
The pH calibration is now complete; "CAL" and the
pH tags corresponding to the buffers used for
calibration are lit on the LCD.
Notes:
• If the buffer value is not recognized, after 12 seconds the meter will display
blinking dashes together with the "WRONG"
tag. Either the buffer solution is wrong or out
of specification and needs to be replaced or the
electrode is damaged.
• The meter will retain the calibration if the battery is removed.
• To quit calibration and keep previous data: press CAL after entering the
calibration mode and before the first point is accepted.
• To perform a single-point calibration: press CAL after the first point has been
confirmed.
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
pH CALIBRATION SOLUTIONS
HI 70004P pH 4.01 buffer solution, 20 mL sachet (25 pcs)
HI 7004MpH 4.01 buffer solution, 230 mL bottle
HI 7004LpH 4.01 buffer solution, 500 mL bottle
HI 8004LpH 4.01 buffer solution, 500 mL FDA bottle
HI 70007P pH 7.01 buffer solution, 20 mL sachet (25 pcs)
HI 7007MpH 7.01 buffer solution, 230 mL bottle
HI 7007LpH 7.01 buffer solution, 500 mL bottle
HI 8007LpH 7.01 buffer solution, 500 mL FDA bottle
HI 70010P pH 10.01 buffer solution, 20 mL sachet (25 pcs)
HI 7010MpH 10.01 buffer solution, 230 mL bottle
HI 7010LpH 10.01 buffer solution, 500 mL bottle
HI 8010LpH 10.01 buffer solution, 500 mL FDA bottle
STORAGE & CLEANING SOLUTIONS
HI 70300M Storage solution, 230 mL bottle
HI 80300M Storage solution, 230 mL FDA bottle
HI 70300L Storage solution, 500 mL bottle
HI 80300L Storage solution, 500 mL FDA bottle
HI 70000P Electrode rinsing solution, 20 mL sachet (25 pcs.)
HI 7061MGeneral cleaning solution, 230 mL bottle
HI 8061MGeneral cleaning solution, 230 mL FDA bottle
HI 7061LGeneral cleaning solution, 500 mL bottle
HI 8061LGeneral cleaning solution, 500 mL FDA bottle
HI 7073MProtein cleaning solution, 230 mL bottle
HI 8073MProtein cleaning solution, 230 mL FDA bottle
HI 7073LProtein cleaning solution, 500 mL bottle
HI 8073LProtein cleaning solution, 230 mL FDA bottle
HI 7074MInorganic cleaning solution, 230 mL bottle
HI 7074LInorganic cleaning solution, 500 mL bottle
HI 7077MOil & Fat cleaning solution, 230 mL bottle
HI 8077MOil & Fat cleaning solution, 230 mL FDA bottle
HI 7077LOil & Fat cleaning solution, 500 mL bottle
HI 8077LOil & Fat cleaning solution, 500 mL FDA bottle
If the temperature probe is not connected and manual temperature compensation is required, follow the procedure below:
• Press RANGE to select the temperature mode.
• Rinse the pH electrode and place it into the pH 7.01 buffer, stir briefly and
wait a few minutes to reach thermal equilibrium.
10
REFILLING ELECTROLYTE SOLUTIONS
HI 70713.5 M KCl+AgCl electrolyte solution, 4 x 30 mL bottle, for single
junction electrodes
HI 80713.5 M KCl+AgCl electrolyte solution, 4 x 30 mL FDA bottle, for
single junction electrodes
HI 70721 M KNO3 electrolyte solution, 4 x 30 mL bottle
19
For refillable electrodes, if the refill solution (electrolyte) is more than 2.5 cm (1")
below the fill hole, add the appropriate electrolyte solution.
MEASUREMENT
Rinse the electrode tip with distilled water, immerse it (4 cm / 1½") in the
sample and stir gently for a few seconds.
For a faster response and to avoid cross contamination of the samples, rinse the
electrode tip with the solution to be tested, before taking any measurements.
• Rinse the temperature probe of a ChecktempC (or ChecktempF) or another
accurate thermometer, and place it close to the pH electrode.
• Use the UP and DOWN arrow keys to manually adjust the temperature to
match the reference thermometer.
STORAGE PROCEDURE
To minimize clogging and ensure a quick response time, the glass bulb and the
junction should always be kept moist.
When not in use, store it with a few drops of HI 70300 storage solution in the
protective cap.
NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable used for the connection to the meter must
be intact and there must be no points of broken insulation on the cable or
cracks on the electrode stem or bulb. If any scratches or cracks are present,
replace the electrode. Rinse off any salt deposits with water.
Connectors must be perfectly clean and dry.
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
For refillable electrodes:For refillable electrodes:
Refill the electrode with fresh electrolyte (see the electrode’s specifications to select
the correct refilling solution). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
•
General
•
Protein
•
Inorganic
•
Oil/grease
IMPORTANT: After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water and soak it in HI 70300 storage
solution for at least 1 hour before taking measurements.
Soak in HI 7061 general cleaning solution for approximately 30 minutes.
Soak in HI 7073 protein cleaning solution for 15 min.
Soak in HI 7074 inorganic cleaning solution for 15
minutes.
Rinse with HI 7077 Oil & Fat cleaning solution for 1
minute.
• Follow the pH calibration procedure explained in the previous pages.
pp
H BUFFER TEMPERATURE DEPENDENCEH BUFFER TEMPERATURE DEPENDENCE
p
H BUFFER TEMPERATURE DEPENDENCE
pp
H BUFFER TEMPERATURE DEPENDENCEH BUFFER TEMPERATURE DEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions are affected
by temperature changes to a lesser degree than normal solutions. During
calibration the instrument will automatically calibrate to the pH value corresponding to the measured or set temperature.
During calibration the instrument will display the pH buffer value at 25°C.
TEMP pH VALUES
°C °F 4.01 7.01 10.01
0
32
4.01
7.13
10.32
5
41
4.00
7.10
10.24
10
50
4.00
7.07
10.18
15
59
4.00
7.04
10.12
20
68
4.00
7.03
10.06
25
77
4.01
7.01
10.01
30
86
4.02
7.00
35
95
4.03
6.99
40
104
4.04
6.98
45
113
4.05
6.98
50
122
4.06
6.98
55
131
4.07
6.98
60
140
4.09
6.98
65
149
4.11
6.99
70
158
4.12
6.99
75
167
4.14
7.00
80
176
4.16
7.01
85
185
4.17
7.02
90
194
4.19
7.03
95
203
4.20
7.04
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75
9.74
9.73
9.74
9.75
9.76
18
11
MENU SELECTIONMENU SELECTION
MENU SELECTION
MENU SELECTIONMENU SELECTION
While in normal measurement mode, press and hold the CFM/SETUP key for about
5 seconds until the meter enters the menu selection mode.
The following parameters can be set from the menu:
1. Auto-off feature: 20 minutes (default setting) or disabled;
2. Acoustic signal: enabled (default setting) or disabled;
3. Temperature unit: °C (default setting) or °F.
When entering the menu mode, the auto-off selection is entered. The LCD shows
in three subsequent screens "Auto", "OFF" and "20" to indicate that the 20
minutes selection is active, or "Auto", "OFF" and "no" if the feature is disabled.
The user can toggle the selection by pressing CAL/SELECT or move to the next
step with CFM/SETUP.
The following selection is the acoustic signal, which is displayed on two
subsequent screens: "bEEP", "OFF" when the feature is disabled, and "bEEP",
"On" when the feature is enabled.
Press CAL/SELECT to toggle the selection and CFM/SETUP to move to next step.
At this point it is possible to set the temperature unit, by selecting "Unit", "°C"
or "Unit", "°F".
Press CAL/SELECT to toggle the selection and CFM/SETUP to exit the menu
selection mode and return to normal measurement mode.
12
Only for refillable electrodes; must be open while measuring.
PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with
electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass bulb. The
electrode cannot function properly under these conditions. These bubbles can be
removed by "shaking down" the electrode as you would do with a glass
thermometer.
If the bulb and/or junction are dry, soak the electrode in HI 70300 storage
solution for at least one hour.
17
TROUBLESHOOTING
mm
V CALIBRATIONV CALIBRATION
m
V CALIBRATION
mm
V CALIBRATIONV CALIBRATION
SymptomProblemSolution
Slow reponse orDirty pH electrodeSoak the electrode tip
excessive driftin HI 7061 solution for
30 minutes
Reading fluctuatesClogged/dirty junctionSoak the electrode tip
up and down (noise)or low electrolyte levelin warm HI 7082
(refillable electrodes)solution for one hour,
then rinse it with
distilled water
(refill with fresh
electrolyte if necessary)
Blinking full scaleReading is out of range
value
Blinking "°C" (or "°F")Temperature probe is not connected or broken
"WRONG" &Calibration errorCheck buffer solution or
"Er1" and Er2"EPROM errorContact your dealer or
messagesHANNA Service Center
HI 8424NEW has been accurately precalibrated for mV range at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for mV
readings at least once a year.
Contact your dealer or the nearest HANNA instruments® Customer Service
Center for more information.
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
HI 8424NEW has been accurately precalibrated for temperature at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for tempera-
ture at least once a year.
Contact your dealer or the nearest HANNA instruments® Customer Service
Center for more information.
Note: For field applications, it is always recommended to keep a conditioned spare elec-
trode handy. When anomalies cannot be resolved with simple maintenance, change
the electrode and recalibrate the meter.
The meter displays the remaining battery percentage when turned on. When the level is
below 5%, the battery symbol on the bottom
left of the LCD blinks to indicate a low battery condition.
If the battery level is low enough to cause erroneous readings, the Battery
Error Prevention System (BEPS) turns the meter
off. When the battery
needs to be replaced, remove the cover on the rear of the meter and replace
the rundown battery with a new
one, while paying attention to
the correct polarity. Reattach
the back making sure that the
gasket is in place and tighten
the 3 screws to ensure a good
seal.
Replacement should take
place in a non-hazardous area
using a 9V alkaline battery.
TAGS & SYMBOLS
• pH, mV, °C, °FMeasurement unit of the selected mode
• ATCIndicates Automatic Temperature Compensation (in pH
or temperature mode)
• MTCIndicates Manual Temperature Compensation (in pH
or temperature mode)
• HOLDBlinks when in Hold mode. Reading frozen on LCD.
The user can scroll through the three ranges by
pressing RANGE
• CALIn pH calibration mode, or in pH mode when the
meter is calibrated
• CFMBlinks in pH calibration mode when the meter is ready
to confirm a value
• WRONGDuring pH calibration, when the meter does not
recognize the pH buffer
• %At startup, when showing the percentage of the
remaining battery life
•In pH mode, when meter was calibrated with pH 7.01
buffer
•In pH mode, when meter was calibrated with pH 4.01
buffer
•In pH mode, when meter was calibrated with pH
10.01 buffer
• (hourglass symbol): When reading is not stable
• (battery symbol):At startup, if remaining battery life is below 5%
14
15
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.