Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you with the necessary
information for the correct use of the instrument. Please read it carefully before using the meter.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
HI 84184-01 - 115 Vac power supply
HI 84184-02 - 230 Vac power supply
WARRA (NTY
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the
copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
2
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any
damage occured during shipment, please notify your Dealer.
Each HI 84184 instrument is supplied complete with:
• Reagents set for 20 tests
• One 1000 µL automatic pipette
• Four plastic tips for the 1000 µL automatic pipette
• Two 50 mL beakers
• Two 20 mL beakers
• HI 61010 Fluoride half cell
• HI 5315 Reference half cell
• HI 7662-T Temperature probe
• Stir bar (2 pcs.)
• Power cable
• HI 7075 Refilling solution (4×30 mL)
• One 1 mL syringe
• One 1 mL pipette
• Instruction sheet
• Instruction manual
Note:Save all packing material until you are sure that the instrument works correctly.
Any defective item must be returned in its original packing.
GENERAL DESCRIPTION
The HI 84184 is a low cost, easy to use, fluoride ISE instrument that performs automatic analysis
with all the necessary calculations assuring to the user a simple and effective interface.
The instrument comes with a powerful and effective built-in algorithm to analyse the shape of the ISE
electrode response and to determine the reaction completion.
By simply pressing the Start key, the instrument performs automatic analysis, all the necessary
calculations and verifications. The result is immediately displayed in convenient units, then the
instrument is ready for another measurement.
SIGNIFICANCE OF USE
Fluoride ion (F-) comes from grapes used for wine and its usual concentration value varies between
0.1 and 2 mg/L. Higher levels of fluoride content in wine can derive from cryolite used by farmers
as an insecticide in the vine.
The HI 84184 ISE Fluoride meter measures the fluoride content in wine using an ion selective electrode.
The used method is double standard addition, and the result is readily displayed in F
3
-
mg/L (ppm).
SPECIFICATIONS
Range0.0 to 5.0 mg/L (ppm) F
Resolution0.1 mg/L (ppm) F
-
-
Accuracy±5% of reading (see note)
Sample volume50 mL
Temperature CompensationAutomatic from 0 to 80°C
ElectrodesHI 61010 Fluoride half cell
HI 5315 Reference half cell
Temperature ProbeHI 7662-T (included)
LOG feature50 samples
Stirring speed500 rpm
Environment0 to 50 °C (32 to 122 °F); max 95% RH non-condensing
Power supply220V/50Hz; 10VA
Dimensions208 x 214 x 163 mm (8.2 x 8.4 x 6.4") (with beaker)
Weight2200 g (77.6 oz.)
Note: Above 3 mg/L the instrument process information about the approximate fluoride content.
REQUIRED REAGENTS
CodeDescriptionQuantity/Test
HI 84184-0TISAB W/510 mL
HI 84184-1Standard 11 mL
HI 84184-2Standard 21 mL
PRINCIPLE OF OPERATION
The determination of fluoride concentration is made according to double standard addition method.
This technique enables an unknown concentration and the electrode’s slope factor to be determined
simultaneously.
The Automatic Temperature compensation is available using HI 7662-T temperature probe. By
setting the electrode isopotential point and enabling temperature compensation option in Setup
menu, the obtained concentration is temperature compensated.
The accuracy of the obtained result depends on the stability of the mV reading in every step of the
measurement.
Three selectable stability criteria can be selected in Setup menu. Select Medium or Accurate
criteria in order to have accurate analysis.
4
FUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION
FRONT PANEL
1)Graphic LCD
2)Functional keys
3)Keypad
4)Electrode holder
REAR PANEL
5)Fluoride half cell10)Temperature probe socket
6)Reference half cell11)BNC electrode connector
7)Temperature probe12)Power switch
8)Electrode holder13)Fuse
9)Reference socket14)Power cable connector
5
KEYPAD DESCRIPTION
1) ESC - to leave the current screen and to return either to the previous screen or to the
main screen. In the setup parameter modifying mode by pressing ESC the value is
not modified.
2)
3) HELP - to access/leave the instrument’s contextual help
4)
5) Functional Keys
- to modify a parameter value, to scroll the information from the instrument’s setup
- to modify a parameter value, to scroll the information from the instrument’s setup
START UP
• Place the instrument on a flat table. Do not place the instrument in direct sunlight.
• Connect the instrument to mains socket with ground connection and the correct voltage and
frequency. See the label on the instrument rear for this.
• Connect the HI 61010 fluoride half cell, the HI 5315 reference half cell and the HI 7662-T
temperature probe.
• Turn the instrument ON using the power switch from the rear panel of the instrument and
wait until main screen is displayed.
GUIDE TO DISPLAY CODES
During the instrument’s operation a series of information is displayed on the LCD.
1) current time information area
2) instrument status area
3) functional keys area
4) stability criteria
5) stirrer and reading status area