Instruction manual
HI 83730
ISM
peroxide value
of edible oils
www.hannainst.com
Warranty
HI 83730HI 83730
HI 83730 is warranted for two years against defects in workmanship
HI 83730HI 83730
and materials when used for its intended purpose and maintained
according to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report
the model number, date of purchase, serial number and the
nature of the failure. If the repair is not covered by the warranty,
you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first
obtain a Returned Goods Authorization Number from the Customer
Service Department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged
for complete protection.
To validate your warranty, fill out and return the enclosed warranty
card within 14 days from the date of purchase.
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide
you with the necessary information for the correct use of the
instrument. Please read it carefully before using the meter. If you
need additional technical information, do not hesitate to e-mail
us at tech@hannainst.com.
This instrument is in compliance with
directives.
Preliminary examination
Please examine this product carefully. Make sure that the instrument
is not damaged. If any damage occurred during shipment, please
notify your Dealer.
Each HI 83730 Ion Selective Meter is supplied complete with:
• reagents for 10 tests;
• four graduated 1 mL syringes, scissors, tissue for wiping
vials;
• four 1,5V AA batteries and AC Adapter
• instruction manual and rigid carrying case
Note: save all packing material until you are sure that the
instrument works correctly. Any defective item must be
returned in its original packing.
General description
The HI 83730 is an auto-diagnostic portable microprocessor
meter that benefits from Hanna’s years of experience as a
manufacturer of analytical instruments. It has an advanced
optical system based on a special tungsten lamp and a narrow
band interference filter that allows most accurate and repeatable
readings. All instruments are factory calibrated.
The auto-diagnostic feature of this meter ensures always optimal
measurement conditions to ensure most precise readings. The
light level is automatically adjusted each time a zero-measurement
is made, and the temperature of the lamp is controlled to avoid
overheating.
Significance and use
Peroxides are the primary products of oil oxidation. Their identification gives useful informations about oil conservation and
HI 83730HI 83730
rancidity.
HI 83730 allows fast and simple analyses of perox-
HI 83730HI 83730
ides in oil in accordance with the EC 2568/91 method.
Oil peroxides content
< 10 meq O2/kg excellent conservation
10-15 meq O2/kg good conservation
< 10 meq O2/kg refined oil
> 20 meq O2/kg rancid oil
IST83730 12/05
Specifications
RangeRange
Range 0.0 to 25.0 meq O
RangeRange
ResolutionResolution
Resolution 0.5 meq O
ResolutionResolution
AccuracyAccuracy
Accuracy ± 0.5 meq O
AccuracyAccuracy
Light sourceLight source
Light source tungsten lamp with narrow band
Light sourceLight source
interference filter @ 466 nm
MethodMethod
Method adaptation of EC 2568/91 method
MethodMethod
and following amendments
Light detectorLight detector
Light detector silicon photocell
Light detectorLight detector
EnvironmentEnvironment
Environment 0 to 50 °C (32 to 122°F); max 95% RH
EnvironmentEnvironment
Power supplyPower supply
Power supply 4x1.5 V AA or adapter 12 Vdc
Power supplyPower supply
Auto shut-offAuto shut-off
Auto shut-off after 15 min of non use
Auto shut-offAuto shut-off
Weight/dimensionsWeight/dimensions
Weight/dimensions 512 g; 224 x 87 x 77 mm
Weight/dimensionsWeight/dimensions
Required reagentsRequired reagents
Required reagents
Required reagentsRequired reagents
Code Description Quantity/test
HI 83730A-0 Peroxide reagent A 1 vial
HI 83730B-0 Peroxide reagent B 1 packet
/kg
2
/kg
2
/kg
2
Functional description
1) Lid
2) Cuvet holder
3) Liquid Crystal Display (LCD)
4) ON/OFF key, to turn the
meter on and off
5) ZERO key, to zero the meter
6) TIMER key, to active a
countdown
7) READ key, to perform
measurement
8) Power socket 12V to
20 V dc 2.5 Watt
Chemical reaction
The reaction between sample and reagent causes a color variation,
proportional to the peroxide content expressed in meq O
2
/kg.
Guide to display codes
This prompt appears for a few seconds
each time the instrument is turned ON.
These prompts indicate the type of power
supply: “Line” (if the external power
supply is used) or the battery level.
Indicates that the instrument is in a
ready state and waiting for the next
command (Timer or Zero).
After TIMER is pressed, a blinking
hourglass icon appears and the display
shows a 6 minutes countdown.
Indicates that the meter is performing a
zero measurement. The light intensity is
automatically re-adjusted if necessary.
The instrument is zeroed and a
measurement can be made.
Indicates that the meter is making a
measurement.
Batteries voltage is getting low and the
batteries need to be replaced.
Indicates that the batteries are dead and
must be replaced. After this message
appears, the instrument is switched off.
Error messages
Change the batteries and restart the meter.
The meter has lost its configuration. Contact
your dealer or the nearest Hanna Customer
Service Center.
“Light high”: there is too much light to
perform a measurement. Please check the
preparation of the zero cuvet.
“Light low”: there is not enough light to
perform a measurement. Please check the
preparation of the zero cuvet.
“No Light”: the lamp is not working because
of a malfunction. Contact your dealer or the
nearest Hanna Customer Service Center.
“Inverted”: the sample and the zero cuvet
are inverted.
The sample absorbs less light than the zero
reference. Check the procedure and make
sure you use the same cuvet for reference
(zero) and measurement.
A flashing value of the maximum
concentration indicates an over range
condition. The concentration of the sample
is beyond the programmed range: dilute the
sample and measure again.
CE declaration of conformity
Recommendations for Users
Before using these products, make sure that they are entirely
suitable for your specific application and for the environment in
which they are used. Operation of these instruments may cause
unacceptable interferences to other electronic equipments, this
requiring the operator to take all necessary steps to correct
interferences. Any variation introduced by the user to the supplied
equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not put the instrument in
microwave ovens. For yours and the instrument safety do not use
or store the instrument in hazardous environments.
General tips for an accurate measurement
The instructions listed below should be carefully followed during
testing to ensure best accuracy.
• In order to measure exactly 1 mL of oil:
(a) Push the plunger completely into the syringe.
(b) Insert the syringe in the oil and push the plunger up and
down twice to rinse
the plunger up until the lower edge of the seal is exactly
on the 0.0 mL mark.
(c) Take out the syringe and clean the outside of the syringe
tip. Then, keeping the syringe in vertical position above
the vial, push the plunger completely down into the
syringe.
• Proper use of the powder reagent packet:
(a) use scissors to open the powder pocket;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.
• In order to avoid reagent leaking and to obtain more
accurate measurements, it is recommended to close the cuvet
very well with the cap.
• Whenever the cuvet is placed into the
measurement cell, it must be dry
outside, and completely free of
fingerprints, oil or dirt. Wipe it throughly
with HI 731318 or a lint-free cloth
prior to insertion.
• The official method EC 2568/91 recommends to work at room
temperature, between 15 and 25 °C.
and eliminate air bubbles; then pull
Instructions
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Remove the cap from a vial of HI 83730A-0 Peroxide reagent.
• Use the graduated syringe to add exactly 1 mL of oil. For a
correct use of the syringe, please see the paragraph "General
tips for an accurate measurement".
• Add the sample to the vial and replace the cap.
• Mix by inverting the vial twice.
• Turn on the meter by pressing the ON/OFF key. When the
LCD displays “---”, it is ready.
• Place the vial into the instrument.
• Press the ZERO key and “----” will
blink on the display.
• After a few seconds the display will
show “-0.0-”. The meter is zeroed
and ready for measurement. Remove
the vial.
• Remove the cap from the vial and
add one packet of HI 83730B-0
Peroxide reagent.
• Replace the cap and press TIMER to
start the countdown.
• Mix
VIGOROUSLY for 1 minute.
• Insert the vial into the instrument
and wait for
Note: 30 seconds before countdown finishes, invert the vial twice.
5 minutes.
• When the countdown finishes, the meter makes the
reading. If you have not used the TIMER key, press READ
to make the measure. In both cases the display will show
“----” during the measurement.
• The meter directly displays the peroxide value in meq O
/kg
2
on the LCD.
Notes:
• To convert the reading to mmol O
/kg multiply the reading
2
by 0.5.
• To convert the reading to mg O
/kg multiply the reading
2
by 8.
Battery replacement
Battery replacement must only take
place in a non-hazardous area. The
blinking “
” will appear when the
batteries power gets low.
When batteries are completely
discharged, “0% bAtt” will appear and
after two seconds the instrument is
switched off.
Remove the battery cover from the
bottom of the instrument and change
the old batteries with 4 fresh 1.5V
batteries, paying attention to the correct
polarity.
Replace the cover.
Accessories
HI 83730-20 reagents kit (21 tests)
HI 93703-50 cleaning solution, 230 mL
HI 740216 test tube cooling rack
HI 740142P 1 mL graduated syringe, 10 pcs.
HI 731318 tissue for wiping cuvets, 4 pcs.
HI 710005 voltage adapter from 115V to 12 Vdc
HI 710006 voltage adapter from 230V to 12 Vdc
Health and safety
The chemicals contained in this kit may be hazardous if improperly
handled. Read the relevant Health and Safety Data Sheet before
performing this test.