Dear Customer,
Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the
meter. This manual will provide you with the necessary information for the correct use of the instrument. If
you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
GENERAL DESCRIPTION .....................................................................................................................................................3
PRECISION AND ACCURACY ...............................................................................................................................................4
PRINCIPLE OF OPERATION ................................................................................................................................................ 4
TIPS FOR AN ACCURATE MEASUREMENT .......................................................................................................................... 8
HEALTH & SAFETY ..........................................................................................................................................................12
HELP MODE ...................................................................................................................................................................19
AMMONIA LOW RANGE ................................................................................................................................................... 20
AMMONIA HIGH RANGE .................................................................................................................................................. 22
CHLORINE, TOTAL .......................................................................................................................................................... 26
NITROGEN, TOTAL LR ...................................................................................................................................................... 30
NITROGEN, TOTAL HR ..................................................................................................................................................... 35
OXYGEN DEMAND, CHEMICAL LR ..................................................................................................................................... 40
OXYGEN DEMAND, CHEMICAL MR .................................................................................................................................... 43
OXYGEN DEMAND, CHEMICAL HR .................................................................................................................................... 46
PHOSPHORUS, TOTAL LR ................................................................................................................................................ 55
PHOSPHORUS, TOTAL HR ................................................................................................................................................62
ERRORS AND WARNINGS ................................................................................................................................................ 66
DATA MANAGEMENT ....................................................................................................................................................... 67
STANDARD METHODS ..................................................................................................................................................... 67
HANNA LITERATURE ....................................................................................................................................................... 69
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna
Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
2
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage
occurred during shipment, please notify your Local Hanna Office.
Each Meter is supplied complete with:
• Ten Sample Vials
• Cloth for wiping vials (1 pcs.)
• Scissors
• Instruction Manual
Note: Save all packing material until you are sure that the instrument works correctly. Any defective item
must be returned in its original packing with the supplied accessories.
ABBREVIATIONS
°C:degree Celsius
°F:degree Fahrenheit
COD:Chemical Oxygen Demand
EPA:US Environmental Protection Agency
µg/L:micrograms per liter (ppb)
mg/L:milligrams per liter (ppm)
g/L:grams per liter (ppt)
mL:milliliter
LR:low range
MR:medium range
HR:high range
GENERAL DESCRIPTION
HI 83214 is a multiparameter bench photometer. It measures 15 colorimetry based methods.
The reagents are in liquid or powder form and are supplied in bottles, ready-to-use vials or packets. The
amount of reagent is precisely dosed to ensure maximum repeatability.
HI 83214 bench photometer can be connected to a PC via an USB cable. Its software companion
is HI 92000 Windows® Compatible Software that helps the user to manage all the results.
HI 83214 has a powerful interactive user support that assists the user during the analysis process.
Each step in the method is help supported on the LCD. Help functions are accessed easily by pressing the HELP
key. A tutorial mode is also available.
3
SPECIFICATIONS
Light LifeLife of the instrument
Light DetectorSilicon Photocell
Environment0 to 50 °C (32 to 122 °F);
max 90% RH non-condensing
Power SupplyExternal 12 Vdc power adaptor
Dimensions235 x 212 x 143 mm (9.2 x 8.34 x 5.62")
Weight1.5 Kg (3.3 lb)
For specifications related to each method (e.g. range, resolution etc.) refer to the related measurement
section.
PRECISION AND ACCURACY
Precision is how closely repeated measurements agree
with each other. Precision is usually expressed as
standard deviation (SD).
Accuracy is defined as the nearness of a test result to
the true value.
Although good precision suggests good accuracy, precise
results can be inaccurate. The figure explains these
definitions.
For each method, the precision is expressed in the
related measurement section.
PRINCIPLE OF OPERATION
Absorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter.
When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or
crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the
matter and on the physical-chemical characteristics of substance according to the Lambert-Beer Law:
-log I/Io = ελ c d
or
A = ελ c d
4
Where:
-log I/I
=Absorbance (A)
o
Io=intensity of incident light beam
I =intensity of light beam after absorption
ελ=molar extinction coefficient at wavelength λ
c=molar concentration of the substance
d=optical path through the substance
Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors
are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific
chemical reaction between sample and reagents.
Given that the absorption of a compound strictly depends on the wavelength of the incident light beam,
a narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize
measurements.
The optical system of HI 83214 is based on special subminiature tungsten lamps and narrow-band
interference filters to guarantee both high performance and reliable results.
Three measuring channels (at three different wavelengths) allow a wide range of tests.
Instrument block diagram (optical layout)
A microprocessor-controlled special tungsten lamp emits radiation which is first optically conditioned and
beamed through the sample contained in the vial. The optical path is fixed by the diameter of the vial.
Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity Io or I.
The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an
electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to
display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual
measurement is performed.
The vial has a very important role because it is an optical element and thus requires particular attention.
It is important that both the measurement and the calibration (zeroing) vials are optically identical to
provide the same measurement conditions. Most methods use the same vial for both, so it is important that
measurements are taken at the same optical point. Instrument measurement principle ensures this.
It is necessary that the surface of the vial is clean and not scratched. This is to avoid measurement
interference due to unwanted reflection and absorption of light. It is recommended not to touch the vial
walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring phases, it is
necessary to cap the vial to prevent any contamination.
5
FUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION
1) Vial holder
2) Liquid Crystal Display (LCD)
3) Keypad
4) Bottom lid
5) USB connector
6) ON/OFF power switch
7) Power supply connector
6
KEYPAD DESCRIPTION
ESC
RCL
HELP
SETUP
The keypad contains 8 direct keys and 3 functional keys with the following functions:
Press to perform the function displayed above it on the LCD.
Press to exit the current screen.
Press to access the method menu.
Press to move up in a menu or a help screen, to increment a set value, or to access second
level functions.
Press to move down in a menu or a help screen, to decrement a set value, or to access second
level functions.
Press to log the current reading.
Press to recall the log.
Press to display the help screen.
Press to access the setup screen.
7
TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure most accurate results.
COLLECTING AND MEASURING SAMPLES
• In order to add the exact amount of sample or liquid reagent to the reagent vials it is strongly
recommended to use the available Hanna automatic fixed-volume pipettes or class A laboratory pipettes
(symbolized like a generic pipette tip in the following method related chapters):
pipette codevolume
HI 731340200 µL
HI 7313411000 µL
HI 7313422000 µL
For correct use of the automatic pipette, please follow the related Instruction Sheet.
Alternatively, the optional HI 740142 1 mL graduated syringe or HI 740226 5 mL graduated syringe
can be used. For the correct use of the syringes, see instructions below.
1. In order to measure exactly 5 mL of reagent with the 5 mL syringe:
(a) push the plunger completely into the syringe and insert the tip into the solution
(b) pull the plunger up until the lower edge of the seal is exactly on the 5 mL mark
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging
on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above
the vial, push the plunger completely down into the syringe. Exactly 5 mL has been added to the vial
8
(c)
(a)(b)
Probable level of
liquid taken up
by syringe
2. In order to measure exactly 0.5 mL of reagent with the 1 mL syringe:
(a) push the plunger completely into the syringe and insert the tip into the solution
(b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging
on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above
the vial, push the plunger down into the syringe until the lower edge of the seal is exactly on the
0.5 mL mark. Now the exact amount of 0.5 mL has been added to the vial, even if the tip still
contains some solution
• Color or suspended matter in large amounts may cause interference, therefore, these should be removed
by treatment with active carbon and by prior filtration.
USING LIQUID AND POWDER REAGENTS
• Proper use of the dropper:
(a) to get good reproducible results, tap the dropper on the table for several times and wipe the
outside of the dropper tip with a cloth
(b) always keep the dropper bottle in a vertical position while dosing the reagent
(a)(b)
9
(b)
(c)
(a)
• Proper use of the powder reagent packet:
(a) use scissors to open the powder packet
(b) push the edges of the packet to form a spout
(c) pour out the content of the packet
USING VIALS
• Never insert hot vials in the instrument, the vial holder may become damaged.
• In order to avoid reagent leaking and to obtain most accurate results, it is recommended to close the
vial tightly with the supplied cap after addition of reagents or sample.
• Shaking the vial can generate bubbles in the sample, causing higher readings. To obtain accurate
measurements, remove such bubbles by swirling or by gently tapping the vial.
• Proper mixing is very important for reproducibility of the measurements. The right way of mixing a vial
is specified for each method in the related chapter.
(a) invert the vial a couple of times or for a specified time: hold the vial in the vertical position with the
cap up. Turn the vial upside-down and wait for all of the solution to flow to the cap end, then return
the vial to the upright vertical position and wait for all of the solution to flow to the vial bottom. This
is one inversion. The correct speed for this mixing technique is 10 complete inversions in 30 seconds.
This mixing technique is indicated with “invert to mix” and the following icon:
invert to mix
10
(b) shake the vial: move the vial up and down. The movement may be gentle or vigorous.
This mixing technique is indicated with “shake gently” or “shake vigorously” and one of the following icons:
shake gentlyshake vigorously
Pay attention to push the cuvette completely down in the holder.
• Whenever the vial is placed into the measurement cell, it must be dry
on the outside, and completely free of fingerprints, oil or dirt. Wipe it
thoroughly with HI 731318 or a lint-free cloth prior to insertion.
• Do not let the reacted sample stand too long after reagent is added.
For best accuracy, respect the timings described in each specific
method.
• It is possible to take multiple readings in a row, but it is recommended
to take a new zero reading for each sample and to use the same vial
for zeroing and measurement where possible (for most precise results follow the measurement procedures
carefully).
• All the reaction times reported in this manual are at 25 °C (77 °F). In general, the reaction time
should be increased for temperatures lower than 20 °C (68 °F), and decreased for temperatures higher
than 25 °C (77 °F).
11
DIGESTION
• Some analytical methods require digestion of the sample. Use the HI 839800 test tube heater for
digestion. The optional HI 740217 safety shield is strongly recommended. For correct use of the reactor
follow the Reactor Instruction Manual. At the end of the digestion period the vials are still hot, allow
the vials to cool to room temperature in the optional HI 740216 test tube cooling rack.
REAGENT BLANK CORRECTION
• Some methods require a “reagent blank correction”. The blank and the sample are prepared exactly
in the same way, with the only difference that for the blank deionized water is used instead of sample.
A blank vial may be used more than once; stability and storing conditions are described for each
method in the related chapter.
INTERFERENCES
• In the method related measurement sections we have reported the most common interferences that may
be present in an average wastewater matrix. It may be that for a particular treatment process other
compounds do interfere with the method of analysis.
HEALTH & SAFETY
• The chemicals contained in the reagent kits may be hazardous if improperly handled.
• Read the appropriate Material Safety Data Sheet (MSDS) before performing tests.
• Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions
carefully.
• Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.
• Reagent vial disposal: Reagents vials may contain different waste pollutants. After use dispose of the
reagent vials according to the local regulations.
HI 93711-03
5HI 93766-50Nitrate28
6HI 93767A-50Nitrogen, Total LR30
7HI 93767B-50Nitrogen, Total HR35
8HI 93754A-25Oxygen Demand, Chemical (COD) LR40
9HI 93754B-25Oxygen Demand, Chemical (COD) MR43
10HI 93754C-25Oxygen Demand, Chemical (COD) HR46
11HI 93758A-50Phosphorus, Reactive49
12HI 93758B-50Phosphorus, Acid Hydrolyzable51
13HI 93758C-50Phosphorus, Total LR55
14HI 93763A-50Phosphorus, Reactive HR59
15HI 93763B-50Phosphorus, Total HR62
13
VIALS IDENTIFICATION
Vials related to different parameters can be distinguished by means of the letter printed on the vial (beside the
Hanna logo) and the color of the printing:
MethodMethodCode specific letterLabel rectangle
reagent set codedescriptionon the labelcolor
HI 93764A-25Ammonia LRA LRwhite
HI 93764B-25Ammonia HRA HRgreen
HI 93701-01Chlorine, Free(no letter)white
HI 93701-03
HI 93711-01Chlorine, Total(no letter)white
HI 93711-03
HI 93766-50NitrateNwhite
HI 93767A-50Nitrogen, Total LRN LRgreen (digestion vial)
red (reagent vial)
HI 93767B-50Nitrogen, Total HRN HRred (digestion vial)
green (reagent vial)
HI 93754A-25Oxygen Demand,COD Ared
Chemical (COD) LR
HI 93754B-25Oxygen Demand,COD Bwhite
Chemical (COD) MR
HI 93754C-25Oxygen Demand,COD Cgreen
Chemical (COD) HR
HI 93758A-50Phosphorus, ReactiveP Rred
HI 93758B-50Phosphorus,P AHwhite
Acid Hydrolyzable
HI 93758C-50Phosphorus, Total LRP TLRred
HI 93763A-50Phosphorus, Reactive HRP RHRgreen
HI 93763B-50Phosphorus, Total HRP THRgreen
14
OPERATIONAL GUIDE
POWER CONNECTION
Connect the AC/DC adaptor that is supplied with the instrument.
Note: Always turn the meter off before unplugging it to ensure no data is lost.
METHOD SELECTION
• Turn the instrument ON via the ON/OFF power switch.
• The meter will perform an autodiagnostic test. During this test, the Hanna Instrument logo will appear
on the LCD. After 5 seconds, if the test was successful, the last method used will appear on the display.
• In order to select the desired method press the METHOD key and the screen with the list of methods
will appear.
• Press the s t keys to highlight the desired method. Press Select.
• After the desired method is selected, follow the measurement procedure described in the related chapter.
• Before performing a test carefully read all the instructions related to the selected method.
15
DATA MANAGEMENT
You can keep track of your results using the instrument’s data log function. Up to 200 individual
measurements can be stored in the data log. Storing, viewing and deleting the data is possible using
LOG,RCL, tor s keys
Storing data
measurement will be stored with date and time stamps.
: You can store only a valid measurement. To store a valid measurement press LOG. The
..
.
..
Viewing and deleting
saved measurement. Additionally, you can delete the data records all at once.
: You can view and delete the data log by pressing the RCL key. You can delete the last
CHEMICAL FORM
Chemical form conversion factors are pre-programmed into the instrument and are method specific. In order
to view the displayed result in the desired chemical form press s or t to toggle between the available
chemical forms for the selected method.
16
SETUP
In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuring
sequence and others are general parameters that change the behavior or appearance of the instrument.
Press SETUP to enter the setup mode.
Press ESC or SETUP to return to the main screen.
A list of setup parameters will be displayed with currently
configured settings. Press HELP for additional information.
Press the s t keys to select a parameter and depending
on the parameter type, select the new value as follows:
Backlight
Values: 0 to 8.
Press the Modify key to access the backlight value.
Use the ⊳ u functional keys or the st keys to increase or
decrease the value.
Press the Accept key to confirm or ESC to return to the setup menu
without saving the new value.
Contrast
Values: 0 to 20.
This option is used to set the display’s contrast.
Press the Modify key to change the display’s contrast.
Use the ⊳ u functional keys or the st keys to increase or
decrease the value.
Press the Accept key to confirm the value or ESC to return to the
setup menu without saving the new value.
17
Date / Time
This option is used to set the instrument’s date and time.
Press the Modify key to change the date/time.
Press the ⊳u functional keys to highlight the value to be
modified (year, month, day, hour, minute or second). Use the
st keys to change the value.
Press the Accept key to confirm or ESC to return to the setup
without saving the new date or time.
Time format
Option: AM/PM or 24 hour.
Press the functional key to select the desired time format.
Date format
Press the Modify key to change the Date Format.
Use the s t keys to select the desired format.
Press the Accept key to confirm or ESC to return to the setup menu
without saving the new format.
Language
Press the corresponding key to change the language.
If the new language cannot be loaded, the previously selected
language will be reloaded.
Tutorial
Option: Enable or Disable.
If enabled this option will provide the user short guides related to
the current issue, on the screen.
Press the functional key to enable/disable the tutorial mode.
18
Beeper
Option: Enable or Disable.
When enabled, a short beep is heard every time a key is pressed.
A long beep alert sounds when the pressed key is not active or an
error condition is detected.
Press the functional key to enable/disable the beeper.
Instrument ID
Option: 0000 to 9999.
This option is used to set the instrument’s ID (identification number).
The instrument ID is used while exchanging data with a PC.
Press the Modify key to access the instrument ID screen. Press the
st keys in order to set the desired value.
Press the Accept key to confirm the value or ESC to return to the
setup menu without saving the new value.
Meter information
Press the Select key to view the instrument model, firmware
version, language version and instrument serial number.
Press ESC to return to the Setup mode.
HELP MODE
HI 83214 offers an interactive contextual help mode that assists the user at any moment.
To access the help screens press HELP.
The instrument will display additional information related to the
current screen. To read all the available information, scroll the text
using the s t keys.
Press the Support key to access a screen with Hanna service
centers and their contact details.
Press the Accessories key to access a list of instrument reagents
and accessories.
To exit support or accessories screens press ESC and the instrument
will return to the previous help screen.
To exit help mode press HELP key again and the meter will return
to the previously selected screen.
19
AMMONIA LOW RANGE
SPECIFICATIONS
Range0.00 to 3.00 mg/L (as NH3-N)
Resolution0.01 mg/L
Accuracy±0.10 mg/L or ±5 % of reading @ 25 °C, whichever is greater
Light SourceTungsten lamp with narrow band interference filter @ 420 nm
MethodAdaptation of the
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
CodeDescriptionQ.ty/testQ.ty/set
HI 93764A-0*Reagent Vial1 vial25 vials
HI 93764-0Nessler Reagent4 drops1 bottle
*
Reagent Vial identification
Note: Store the unused vials in a cool and dark place.
REAGENT SET
HI 93764A-25Reagents for 25 tests
For other accessories see page 68.
MEASUREMENT PROCEDURE
• Select the
section (see page 15).
Ammonia LR
ASTM Manual of Water and Environmental Technology, D1426-92
: A LR, white label.
method following one of the procedures described in the “Method Selection”
,
• Remove the cap from a Reagent Vial.
• Add exactly 5.0 mL of sample to the vial, while keeping the
vial at a 45-degree angle.
• Replace the cap and mix by inverting the vial a couple of
times. This is the blank.
• Place the vial into the holder and push it completely down.
Ammonia LR
20
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
4
measurement.
• Remove the vial.
• Remove the cap and add 4 drops of HI 93764-0 Nessler
Reagent.
• Replace the cap tightly and mix by inverting the vial a couple
of times. This is the sample.
• Place the vial into the holder and push it completely down.
• Press Timer and the display will show the countdown prior to the
measurement and the message “Reaction Time”. Alternatively, wait
for 3 minutes and 30 seconds and press Read. The instrument will
perform the reading.
• The instrument displays concentration in mg/L of ammonia nitrogen (NH3-N). Press s or t to
convert the result in mg/L ammonia (NH3).
INTERFERENCES
Interference may be caused by:
• Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine or urea above 10 ppm.
To eliminate these interferences distillation is required.
• Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone above 0.1 %.
To eliminate these interferences distillation is required.
• Sulfide: may cause turbidity.
21
Ammonia LR
AMMONIA HIGH RANGE
SPECIFICATIONS
Range0 to 100 mg/L (as NH3-N)
Resolution1 mg/L
Accuracy±1 mg/L or ±5 % of reading @ 25 °C, whichever is greater
Light SourceTungsten lamp with narrow band interference filter @ 420 nm
MethodAdaptation of the
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
CodeDescriptionQ.ty/testQ.ty/set
HI 93764B-0*Reagent Vial1 vial25 vials
HI 93764-0Nessler Reagent4 drops1 bottle
*
Reagent Vial identification
Note: Store the unused vials in their container in a cool and dark place.
REAGENT SET
HI 93764B-25Reagents for 25 tests
For other accessories see page 68.
MEASUREMENT PROCEDURE
• Select
• Remove the cap from a Reagent Vial.
Ammonia HR
(see page 15).
method following one of the procedures described in the “Method Selection” section
ASTM Manual of Water and Environmental Technology, D1426-92
: A HR, green label.
,
• Add exactly 1.0 mL of sample to the vial, while keeping the
vial at a 45-degree angle.
• Replace the cap and mix by inverting the vial a couple of times. This is the blank.
• Place the vial into the holder and push it completely down.
Ammonia HR
22
Loading...
+ 50 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.