Dear Customer,
Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the
instrument. This manual will provide you with the necessary information for the correct use of the
instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
HANNA LITERATURE ........................................................ 67
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna
Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
2
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage
occurred during shipment, please notify your local Hanna Office.
Each Meter is supplied complete with:
• Three Sample Cuvettes and Caps
• Cloth for wiping cuvettes (1 pcs)
• Scissors
• AC/DC Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the instrument works correctly. Any defective item
must be returned in its original packing with the supplied accessories.
ABBREVIATIONS
EPA:US Environmental Protection Agency
°C:degree Celsius
°F:degree Fahrenheit
μg/L:micrograms per liter (ppb)
mg/L:milligrams per liter (ppm)
g/L:grams per liter (ppt)
mL:milliliter
HR:high range
MR:medium range
LR:low range
PAN:1-(2-pyridylazo)-2-naphtol
GENERAL DESCRIPTION
HI 83207 is a multiparameter bench photometer dedicated for Industrial Wastewater analysis. It measures
20 different methods using specific liquid or powder reagents. The amount of reagent is precisely dosed to
ensure maximum reproducibility.
HI 83207 bench photometer can be connected to a PC via an USB cable. The optional HI 92000
Windows® Compatible Software helps users manage all their results.
HI 83207 has a powerful interactive user support that assists the user during the analysis process.
Each step in the measurement process is help supported. A tutorial mode is available in the Setup Menu.
3
SPECIFICATIONS
Light LifeLife of the instrument
Light DetectorSilicon Photocell
Environment0 to 50°C (32 to 122°F);
max 90% RH non-condensing
Power Supplyexternal 12 Vdc power adapter
built-in rechargeable battery
Dimensions235 x 200 x 110 mm (9.2 x 7.87 x 4.33")
Weight0.9 Kg
For specifications related to each method (e.g. range, resolution, etc.) refer to the related measurement
section.
PRECISION AND ACCURACY
Precision is how closely repeated measurements agree
with each other. Precision is usually expressed as
standard deviation (SD).
Accuracy is defined as the nearness of a test result to
the true value.
Although good precision suggests good accuracy, precise
results can be inaccurate. The figure explains these
definitions.
For each method, the precision is expressed in the
related measurement section.
PRINCIPLE OF OPERATION
Absorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter.
When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or
crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the
matter and on the physical-chemical characteristics of substance according to the Lambert-Beer Law:
-log I/Io = ε
A = ε
c d
λ
or
c d
λ
4
Where:
-log I/I
= Absorbance (A)
o
Io= intensity of incident light beam
I = intensity of light beam after absorption
ε
= molar extinction coefficient at wavelength λ
λ
c= molar concentration of the substance
d= optical path through the substance
Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors
are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific
chemical reaction between sample and reagents.
Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a
narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of HI 83207 is based on special subminiature tungsten lamps and narrow-band
interference filters to guarantee both high performance and reliable results.
Four measuring channels allow a wide range of tests.
Instrument block diagram (optical layout)
A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and
beamed trough the sample contained in the cuvette. The optical path is fixed by the diameter of the
cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of
intensity Io or I.
The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an
electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to
display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual
measurement is performed.
The cuvette has a very important role because it is an optical element and thus requires particular attention. It
is important that both the measurement and the calibration (zeroing) cuvette are optically identical to provide the
same measurement conditions. Most methods use the same cuvette for both, so it is important that measurements
are taken at the same optical point. The instrument and the cuvette cap have special marks that must be
aligned in order to obtain better reproducibility.
The surface of the cuvette must be clean and not scratched. This is to avoid measurement interference due to
unwanted reflection and absorption of light. It is recommended not to touch the cuvette walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measurement phases,
it is necessary to cap the cuvette to prevent any contamination.
5
FUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION
1) Open Cuvette Lid
2) Indexing mark
3) Cuvette point
4) Liquid Crystal Display (LCD).
5) Splash proof keypad.
6) ON/OFF power switch
7) Power input connector
8) USB connector
6
KEYPAD DESCRIPTION
The keypad contains 8 direct keys and 3 functional keys with the following functions:
Press to perform the function displayed above it on the LCD.
ESC
Press to exit the current screen.
Press to access the select method menu.
Press to move up in a menu or a help screen, to increment a set value, to access second level
functions.
Press to move down in a menu or a help screen, to decrement a set value, to access second
level functions.
Press to log the current reading.
RCL
Press to recall the log.
HELP
SETUP
Press to display the help screen.
Press to access the setup screen.
7
TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure most accurate results.
• Color or suspended matter in large amounts may cause interference, and should be removed by
treatment with active carbon and filtration.
• Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom
of this convexity must be at the same level as the 10 mL mark.
COLLECTING AND MEASURING SAMPLES
• In order to measure exactly 0.5 mL of reagent with the 1 mL syringe:
(a) push the plunger completely into the syringe and insert the tip into the solution.
(b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark.
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging
on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above
the cuvette, push the plunger down into the syringe until the lower edge of the seal is exactly on
the 0.5 mL mark. Now the exact amount of 0.5 mL has been added to the cuvette, even if the
tip still contains some solution.
USING LIQUID AND POWDER REAGENTS
• Proper use of the dropper:
(a) for reproducible results, tap the dropper on the table for several times and wipe the outside of the
dropper tip with a cloth.
(b) always keep the dropper bottle in a vertical position while dosing the reagent.
(a)(b)
8
• Proper use of the powder reagent packet:
(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.
USING CUVETTES
• Proper mixing is very important for reproducibility of the measurements. The right way of mixing a
cuvette is specified for each method in the related chapter.
(a) invert the cuvette a couple of times or for a specified time: hold the cuvette in the vertical position. Turn
the cuvette upside-down and wait for all of the solution to flow to the cap end, then return the cuvette
to the upright vertical position and wait for all of the solution to flow to the cuvette bottom. This is one
inversion. The correct speed for this mixing technique is 10-15 complete inversions in 30 seconds.
This mixing technique is indicated with “invert to mix” and the following icon:
(b) shaking the cuvette, moving the cuvette up and down. The movement may be gentle or vigorous.
This mixing method is indicated with “shake gently” or “shake vigorously”, and one of the following
icons:
shake gentlyshake vigorously
9
• Pay attention to push the cuvette completely down in the holder and to align the white point on the
cap to the indexing mark on the meter.
• In order to avoid reagent leaking and to obtain more accurate
measurements, close the cuvette first with the supplied HDPE plastic
stopper and then the black cap.
• Each time the cuvette is used, the cap must be tightened to the same
degree.
• Whenever the cuvette is placed into the measurement cell, it must be dry
outside, and free of fingerprints, oil or dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
• Shaking the cuvette can generate bubbles in the sample, causing
higher readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand too long after reagent is added. For
best accuracy, respect the timings described in each specific method.
• It is possible to take multiple readings in a row, but it is recommended
to take a new zero reading for each sample and to use the same cuvette
for zeroing and measurement when possible (for most precise results
follow the measurement procedures carefully).
• Discard the sample immediately after the reading is taken, or the glass might become permanently
stained.
• All the reaction times reported in this manual are at 25 °C (77 °F). In general, the reaction time
should be increased for temperatures lower than 20 °C (68 °F), and decreased for temperatures higher
than 25 °C (77 °F).
INTERFERENCES
• In the method measurement section the most common interferences that may be present in an average
wastewater matrix have been reported. It may be that for a particular treatment process other
compounds do interfere with the method of analysis.
10
HEALTH & SAFETY
• The chemicals contained in the reagent kits may be hazardous if improperly handled.
• Read the Material Safety Data Sheet (MSDS) before performing tests.
• Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions
carefully.
• Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.
• Waste disposal: for proper disposal of reagent kits and reacted samples, refer to the Material Safety
Data Sheet (MSDS).
The meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.
Note: Always turn the meter off before unplugging it to ensure no data is lost.
When the meter switches ON, it verifies if the power supply adapter is connected. The battery icon on the
LCD will indicate the battery status:
- battery is charging from external adapter- battery fully charged (meter connected to AC/DC adapter)
- battery capacity (no external adapter)- battery Low (no external adapter)
- battery Dead (no external adapter)
METHOD SELECTION
• Turn ON the instrument via the ON/OFF power switch.
• The meter will perform an autodiagnostic test. During this test, the Hanna Instrument logo will appear on
the LCD. After 5 seconds, if the test was successful, the last selected method will appear on the display.
• In order to select the desired method press the METHOD key and a screen with the available methods
will appear.
• Press the keys to highlight the desired method. Press Select.
12
• After the desired method is selected, follow the measurement described in the related section.
• Before performing a test you should read all the instructions carefully.
DATA MANAGEMENT
The instrument features a data log function to help you keep track of all your analysis. The data log can
hold 200 individual measurements. Storing, viewing and deleting the data is possible using the LOG
and
RCL keys
Storing data
stored with date and time stamps.
..
.
..
: You can store only a valid measurement. Press LOG
and the last valid measurement will be
Viewing and deleting
the last saved measurement. Additionally, you can delete the data records all at once.
: You can view and delete the data log by pressing the RCL key. You can only delete
CHEMICAL FORM
Chemical form conversion factors are pre-programmed into the instrument and are method specific. In order
to view the displayed result in the desired chemical form press or to access the second level function
and then press the Chem Frm key to toggle between the available chemical forms for the selected method.
13
SETUPSETUP
SETUP
SETUPSETUP
In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuring
sequence and others are general parameters that change the behavior or appearance of the instrument.
Press SETUP to enter the setup mode.
Press ESC or SETUP to return to the main screen.
A list of setup parameters will be displayed with currently
configured settings. Press HELP for additional information.
Press the keys to select the parameter and select a
new value as follows:
Backlight
Values: 0 to 8.
Press the Modify key to access the backlight value.
Use the W X functional keys or the keys to increase or
decrease the value.
Press the Accept key to confirm or ESC to return to the setup
menu without saving the new value.
Contrast
Values: 0 to 20.
This option is used to set the display’s contrast.
Press the Modify key to change the display’s contrast.
Use the W X functional keys or the keys to increase or
decrease the value.
Press the Accept key to confirm the value or ESC to return to the
setup menu without saving the new value.
14
Date / Time
This option is used to set the instrument’s date and time.
Press the Modify key to change the date/time.
Press the W X functional keys to highlight the value to be
modified (year, month, day, hour, minute or second). Use the
keys to change the value.
Press the Accept key to confirm or ESC to return to the setup
without saving the new date or time.
Time format
Option: AM/PM or 24 hour.
Press the functional key to select the desired time format.
Date format
Press Modify key to change the Date Format.
Use the keys to select the desired format.
Press the Accept key to confirm or ESC to return to the setup
menu without saving the new format.
Language
Press the corresponding functional key to change the option.
If the new selected language cannot be loaded, the previously
selected language will be reloaded.
Tutorial
Option: Enable or Disable.
If enabled this option will provide the user short guides, related
to the current screen.
Press the functional key to enable/disable the tutorial mode.
Beeper
Option: Enable or Disable.
When enabled, a short beep is heard every time a key is pressed.
A long beep alert sounds when the pressed key is not active or an
error is detected.
Press the functional key to enable/disable the beeper.
15
Instrument ID
Option: 0 to 9999.
This option is used to set the instrument’s ID (identification
number). The instrument ID is used while exchanging data with
a PC.
Press the Modify key to access the instrument ID screen. Press
the keys in order to set the desired value.
Press the Accept key to confirm the value or ESC to return to the
setup menu without saving the new value.
Meter information
Press the Select key to view the Instrument model, firmware
version, language version and instrument serial number.
Press ESC to return to the Setup mode.
HELP MODEHELP MODE
HELP MODE
HELP MODEHELP MODE
HELP MODE
HI 83207 offers an interactive contextual help mode that assists the user at any time.
To access help screens press HELP.
The instrument will display additional information related to the
current screen. To read all available information, scroll the text
using the keys.
Press the Support key to access a screen with Hanna service
centers and their contact details.
Press the Accessories key to access a list of instrument reagents
and accessories.
To exit support or accessories screens press ESC and the
instrument will return to the previous help screen.
To exit help mode press the HELP or ESC key again and the
meter will return to the previously selected screen.
16
ALUMINUM
SPECIFICATIONS
Range0.00 to 1.00 mg/L
Resolution0.01 mg/L
Accuracy±0.02 mg/L ±4% of reading at 25 °C
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the aluminon method. The reaction between aluminum and reagents
causes a reddish tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93712A-0Ascorbic acid1 packet
HI 93712B-0Aluminon reagent1 packet
HI 93712C-0Bleaching powder1 packet
REAGENT SETS
HI 93712-01 Reagents for 100 tests
HI 93712-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Method Selection
Aluminum
method using the procedure described in the
section (see page 12).
• Fill a graduated beaker with 50 mL of sample.
• Add the content of one packet of HI 93712A-0 Ascorbic acid and mix
until completely dissolved.
• Add the content of one packet of HI 93712B-0 Aluminon reagent
and mix until completely dissolved. This is the sample.
• Fill two cuvettes with 10 mL of sample each (up to the mark).
17
10 mL
# 1
10 mL
# 2
Aluminum
• Add the content of one packet of HI 93712C-0 Bleaching
powder to one of the two cuvettes. Replace the cap and shake
vigorously until completely dissolved. This is the blank.
• Place the blank into the holder and close the lid.
# 1
• Press Timer and the display will show the countdown prior to
# 1
zeroing the blank. Alternatively wait for 15 minutes and then press
Zero. The display will show “-0.0-” when the meter is zeroed and
ready for measurement.
# 2
• Remove the blank and insert the other cuvette into the instrument.
• Press the Read key and the meter will perform the reading. The instrument displays the results in
mg/L of aluminum.
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of Al2O3.
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Iron above 20 mg/L, Alkalinity above 1000 mg/L, Phosphate above 50 mg/L, Fluoride must be absent.
Aluminum
18
FREE CHLORINE
FREE CHLORINEFREE CHLORINE
FREE CHLORINE
FREE CHLORINEFREE CHLORINE
SPECIFICATIONS
Range0.00 to 2.50 mg/L
Resolution0.01 mg/L
Accuracy±0.03 mg/L ±3% of reading at 25 °C
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the
DPD reagent causes a pink tint in the sample.
EPA DPD method 330.5.
The reaction between free chlorine and the
REQUIRED REAGENTS
POWDER:
CodeDescriptionQuantity
HI 93701-0DPD1 packet
LIQUID:
CodeDescriptionQuantity
HI 93701A-FDPD1 Indicator3 drops
HI 93701B-FDPD1 Buffer3 drops
REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder)
HI 93701-03 Reagents for 300 tests (powder)
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
in the
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
Free Chlorine
Method Selection
method using the procedure described
section (see page 12).
19
10 mL
Free Chlorine
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
Powder reagents procedure
• Add the content of one packet of HI 93701 DPD reagent. Replace the
cap and shake gently for 20 seconds (or 2 minutes for seawater
analysis).
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of free chlorine.
Liquid reagents procedure
• To an empty cuvette add 3 drops of HI 93701A-F DPD1
indicator and 3 drops of HI 93701B-F DPD1 buffer. Swirl
gently to mix, and immediately add 10 mL of unreacted
sample. Replace the cap and shake gently again.
Free Chlorine
20
• Reinsert the cuvette into the instrument.
• Press Read to start the reading. The instrument displays the results in mg/L of free chlorine.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO3, shake the sample for approximately
2 minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO3 or acidity greater than 150 mg/L CaCO3,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
21
Free Chlorine
TOTAL CHLORINE
SPECIFICATIONS
Range0.00 to 3.50 mg/L
Resolution0.01 mg/L
Accuracy±0.03 mg/L ±3% of reading at 25 °C
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the
DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER:
CodeDescriptionQuantity
HI 93711-0DPD1 packet
LIQUID:
CodeDescriptionQuantity
HI 93701A-TDPD1 indicator3 drops
HI 93701B-TDPD1 buffer3 drops
HI 93701CDPD3 solution1 drop
REAGENT SETS
HI 93701-T Reagents for 300 total chlorine tests (liquid)
HI 93711-01 Reagents for 100 total chlorine tests (powder)
HI 93711-03 Reagents for 300 total chlorine tests (powder)
For other accessories see page 65.
EPA DPD method 330.5.
The reaction between the chlorine and the
MEASUREMENT PROCEDURE
• Select the
described in the
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
Total Chlorine
Total Chlorine
Method Selection
method using the procedure
section (see page 12).
22
10 mL
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
Powder reagents procedure
• Add 1 packet of HI 93711 DPD reagent. Replace the cap and
shake gently for 20 seconds (or 2 minutes for seawater analysis).
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
and press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L of total
chlorine.
Liquid reagents procedure
• To an empty cuvette add 3 drops of HI 93701A-T
DPD1 indicator, 3 drops of HI 93701B-T DPD1 buffer and
1 drop of HI 93701C DPD3 solution. Swirl gently to mix
and immediately add 10 mL of unreacted sample.
Replace the cap and shake gently again.
• Reinsert the cuvette into the instrument.
23
Total Chlorine
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and 30 seconds and press Read. When the timer ends the meter will perform the
reading.
• The instrument displays the results in mg/L of total chlorine.
Note: Free and total chlorine have to be measured separately with fresh unreacted samples following
the related procedure if both values are requested.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO3, shake the sample for approximately
2 minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO3 or acidity greater than 150 mg/L CaCO3,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Total Chlorine
24
COLOR OF WATER
SPECIFICATIONS
Range0 to 500 PCU (Platinum Cobalt Units)
Resolution1 PCU
Accuracy±10 PCU ±5% of reading at 25 °C
Typical EMC± 1 PCU
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 420 nm
MethodAdaptation of the
18th edition
REQUIRED ACCESSORIES
0.45 μm membrane for true color measurement.
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
the
• Fill one cuvette up to the mark with deionized water and replace
the cap. This is the blank.
• Place the blank (# 1) into the holder and close the lid.
Color of Water
Method Selection
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
, Colorimetric Platinum Cobalt method.
method using the procedure described in
# 1
10 mL
# 1
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank.
25
Color of Water
• Fill the second cuvette up to the mark with unfiltered sample
and replace the cap. This is the apparent color.
• Filter 10 mL of sample through a filter with a 0.45 μm
membrane into the third cuvette, up to the 10 mL mark and
replace the cap. This is the true color.
10 mL
# 2
# 3
• Insert the apparent color cuvette (# 2) into the instrument
# 2
and close the lid.
• Press Read to start the reading.
• The meter displays the value of apparent color in PCU.
• Remove the cuvette, insert the true color cuvette (# 3) into the
instrument and ensure that the notch on the cap is positioned
# 3
securely into the groove.
• Press Read to start the reading. The meter displays the value of true color in PCU.
Color of Water
26
COPPER HIGH RANGE
SPECIFICATIONS
Range0.00 to 5.00 mg/L
Resolution0.01 mg/L
Accuracy±0.02 mg/L ±4% of reading at 25 °C
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
MethodAdaptation of the
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93702-0Bicinchoninate1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Method Selection
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Copper HR
method using the procedure described in the
section (see page 12).
EPA method.
The reaction between copper and the bicinchoninate
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The meter will show “-0.0-”” when the meter is zeroed and ready for
measurement.
27
Copper HR
• Remove the cuvette.
• Add the content of one packet of HI 93702-0 Bicinchoninate.
Replace the cap and shake gently for about 15 seconds.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 45 seconds and press Read. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of copper.
INTERFERENCES
Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted between
6 and 8.
Copper HR
28
COPPER LOW RANGE
SPECIFICATIONS
Range0 to 1000 μg/L
Resolution1 μg/L
Accuracy±10 μg/L ±5% of reading at 25 °C
Typical EMC±1 μg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
MethodAdaptation of the
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 95747-0Bicinchoninate1 packet
REAGENT SETS
HI 95747-01 Reagents for 100 tests
HI 95747-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
the
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Copper LR
Method Selection
method using the procedure described in
section (see page 12).
EPA method.
The reaction between copper and the bicinchoninate
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
29
Copper LR
• Remove the cuvette.
• Add the content of one packet of HI 95747-0 Bicinchoninate.
Replace the cap and shake gently for about 15 seconds.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 45 seconds and press Read. When the timer ends the meter will perform the reading.
• The instrument displays the results in
μμ
μg/L of copper.
μμ
INTERFERENCES
Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted between
6 and 8.
Copper LR
30
FLUORIDE
SPECIFICATIONS
Range0.00 to 2.00 mg/L
Resolution0.01 mg/L
Accuracy±0.03 mg/L ±3% of reading at 25 °C
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
MethodAdaptation of the
18th edition
causes a red tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93729-0SPADNS Reagent4 mL
REAGENT SETS
HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Fluoride
Method Selection
method using the procedure described in the
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
, SPADNS method. The reaction between fluoride and the liquid reagent
• Add 2 mL of HI 93729-0 SPADNS Reagent to two cuvettes.
• Fill one of the cuvettes with distilled water up to the mark,
replace the cap and invert several times to mix.
• Fill the other cuvette with sample up to the mark, replace the
cap and invert several times to mix.
• Place the cuvette with the reacted distilled water (# 1) into
the holder and close the lid.
31
# 1
# 1
2 mL
10 mL10 mL
# 2
Fluoride
• Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively, wait
for two minutes and press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
# 2
• Remove the cuvette.
• Insert the other cuvette (# 2) with the reacted sample into the
instrument.
• Press Read to start reading. The instrument displays the results in mg/L of fluoride.
Note: For wastewater or seawater samples, before performing measurements, distillation is required.
For most accurate results, use two graduated pipettes to deliver exactly 8 mL of distilled water and
8 mL of sample.
INTERFERENCES
Interferences may be caused by:
Alkalinity (as CaCO3) above 5000 mg/L
Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Chloride above 700 mg/L
Phosphate, ortho above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.
Fluoride
32
MANGANESE HIGH RANGE
SPECIFICATIONS
Range0.0 to 20.0 mg/L
Resolution0.1 mg/L
Accuracy±0.2 mg/L ±3% of reading at 25 °C
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the
18th edition,
a pink tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93709A-0Citrate1 packet
HI 93709B-0Sodium periodate1 packet
REAGENT SETS
HI 93709-01 Reagents for 100 tests
HI 93709-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the
Manganese HR
the
Method Selection
mark) and replace the cap.
method using the procedure described in
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
Periodate method. The reaction between manganese and reagents causes
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add the content of one packet of HI 93709A-0 Citrate reagent.
Replace the cap and invert to mix with gently movements for
2 minutes.
33
Manganese HR
• Add the content of one packet of HI 93709B-0 Sodium
Periodate reagent. Replace the cap and invert to mix with
gently movements for 2 minutes.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and 30 seconds and press Read. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of manganese.
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of potassium permanganate (KMnO4) and
permanganate (MnO4).
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Calcium above 700 mg/L
Chloride above 70000 mg/L
Iron above 5 mg/L
Magnesium above 100000 mg/L.
Manganese HR
34
MANGANESE LOW RANGE
SPECIFICATIONS
Range0 to 300 μg/L
Resolution1 μg/L
Accuracy±10 μg/L ±3% of reading at 25 °C
Typical EMC±1 μg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
MethodAdaptation of the PAN Method. The reaction between manganese and the reagents
causes an orange tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93748A-0Ascorbic acid2 packets
HI 93748B-0Alkaline-cyanide sol.0.40 mL
HI 93748C-00.1% PAN indicator2 mL
HI 93703-51Dispersing Agent4-6 drops (only when necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
page 12).
Manganese LR
method using the procedure described in the
Method Selection
section (see
• Fill one cuvette with 10 mL of deionized water (up to the mark).
• Fill a second cuvette with 10 mL of sample (up to the mark).
• Add the content of one packet of HI 93748A-0 Ascorbic acid to
each cuvette, replace the caps and shake gently until completely
dissolved.
35
10 mL
10 mL
Manganese LR
• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent
solution to each cuvette, replace the caps and invert gently
to mix for about 30 seconds.
• Add 1 mL of the HI 93748C-0 0.1% PAN indicator
solution to each cuvette, replace the caps and shake
gently.
• Add 3 drops of Dispersing Agent (HI 93703-51) to each
cuvette, replace the caps and invert gently to mix for
about 30 seconds.
• Place the cuvette with the reacted deionized water (blank)
into the holder and close the lid.
• Press Timer and the display will show the countdown prior to zeroing the blank. Alternatively wait for
2 minutes and then press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Manganese LR
36
• Insert the second cuvette with the reacted sample into the instrument.
• Press Read to start the reading. The instrument displays the results in
μμ
μg/L of manganese.
μμ
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of potassium permanganate (KMnO4) and
permanganate (MnO4).
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO
Nickel above 40 mg/L
Zinc above 15 mg/L.
3
3
37
Manganese LR
MOLYBDENUM
SPECIFICATIONS
Range0.0 to 40.0 mg/L
Resolution0.1 mg/L
Accuracy±0.3 mg/L ±5% of reading at 25 °C
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 420 nm
MethodAdaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93730A-0Reagent A1 packet
HI 93730B-0Reagent B1 packet
HI 93730C-0Reagent C1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
described in the
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
Molybdenum
Method Selection
method using the procedure
section (see page 12).
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Molybdenum
38
• Fill one graduated mixing cylinder up to the 25 mL
mark with the sample.
• Add the content of one packet of HI 93730A-0
molybdenum reagent, close the cylinder and invert
it several times until completely dissolved.
• Add the content of one packet of HI 93730B-0
molybdenum reagent to the cylinder, close and
invert it several times until completely dissolved.
• Add the content of one packet of HI 93730C-0
molybdenum reagent to the cylinder, close and
shake it vigorously.
25 mL
• Fill an empty cuvette with 10 mL of sample up to
the mark and replace the cap.
• Insert the cuvette into the instrument.
39
10 mL
Molybdenum
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for five minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in mg/L of molybdenum.
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of molybdate (MoO4) and sodium molybdate
(Na2MoO4).
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 50 mg/L
Chromium above 1000 mg/L
Copper above 10 mg/L
Iron above 50 mg/L
Nickel above 50 mg/L
Nitrite, as NO2¯
Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents.
Molybdenum
40
NICKEL HIGH RANGE
SPECIFICATIONS
Range0.00 to 7.00 g/L
Resolution0.01 g/L
Accuracy±0.07 ±4% of reading at 25 °C
Typical EMC±0.02 g/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
MethodAdaptation of the photometric method. The reaction between nickel and the reagent
causes a blue tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93726-0Powder reagent1 packet
REAGENT SETS
HI 93726-01 Reagents for 100 tests
HI 93726-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette up to the mark with 10 mL of unreacted
Nickel HR
method using the procedure described in the
sample and replace the cap.
Method Selection
section (see page 12).
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
41
Nickel HR
• Remove the cuvette and add the content of one packet of
HI 93726-0 reagent. Replace the cap and shake gently until
completely dissolved.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and press Read. When the timer ends the meter will perform the reading.
• The instrument displays concentration in g/L of nickel.
INTERFERENCES
Interference may be caused by copper.
Nickel HR
42
NICKEL LOW RANGE
SPECIFICATIONS
Range0.000 to 1.000 mg/L
Resolution0.001 mg/L
Accuracy±0.010 mg/L ±7% of reading at 25 °C
Typical EMC±0.001 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
MethodAdaptation of the PAN method. The reaction between nickel and the reagents causes
an orange tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93740A-0Phthalate-phosphate2 packets
HI 93740B-00.3% PAN indicator2 mL
HI 93740C-0EDTA2 packets
HI 93703-51Dispersing Agent4-6 drops (only when necessary, see note)
REAGENT SETS
HI 93740-01 Reagents for 50 tests
HI 93740-03 Reagents for 150 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
described in the
Note: For best results perform your tests between 20-24°C.
Nickel LR
method using the procedure
Method Selection
section (see page 12).
• Fill one graduated beaker with 25 mL of deionized
water (blank) and another one with 25 mL of sample.
• Add the content of one packet of HI 93740A-0
Phthalate-phosphate reagent to each beaker. Cap
and swirl gently until the reagent is dissolved.
Note: If sample contains iron (Fe3+), it is important
that all powder be dissolved completely before continuing
with following step.
• Add 1 mL of HI 93740B-0 0.3% PAN solution to each
beaker, cap and swirl to mix.
43
Nickel LR
• Press Timer and the display will show a countdown prior to adding reagent C or, alternatively, wait
for 15 minutes. Add one packet of HI 93740C-0 EDTA reagent to each beaker, cap and swirl to mix
until completely dissolved.
• Fill one cuvette up to the mark with 10 mL of the blank.
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Fill a second cuvette up to the mark with 10 mL of the reacted sample.
10 mL
• Insert the second cuvette into the instrument.
• Press Read to start the reading. The instrument displays the results in mg/L of nickel.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking
readings, add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity
is removed.
Range0.0 to 30.0 mg/L
Resolution0.1 mg/L
Accuracy±0.5 mg/L ±10% of reading at 25 °C
Typical EMC±0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the cadmium reduction method. The reaction between nitrate and the
reagent causes an amber tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93728-0Powder reagent1 packet
REAGENT SETS
HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
the
• Using the pipette, fill the cuvette with 6 ml of sample, up to
half of its height, and replace the cap.
Nitrate
method using the procedure described in
Method Selection
section (see page 12).
6 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette and add the content of one packet of
HI 93728-0 reagent.
Nitrate
46
• Replace the cap and immediately shake vigorously up
and down for exactly 10 seconds. Continue to mix by
inverting the cuvette gently for 50 seconds, while
taking care not to induce air bubbles. Powder will not
completely dissolve. Time and way of shaking could
sensitively affect the measurement.
• Reinsert the cuvette into the instrument, taking care
not to shake it.
• Press Timer and the display will show the countdown prior
to the measurement or, alternatively, wait for 4 minutes and
30 seconds and press Read. When the timer ends the
meter will perform the reading. The instrument displays
the results in mg/L of nitrate-nitrogen.
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of nitrate (NO3¯).
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Ammonia and amines, as urea and primary aliphatic amines
Chloride above 100 ppm
Chlorine above 2 ppm
Copper
Iron(III)
Strong oxidizing and reducing substances
Sulfide must be absent
47
Nitrate
DISSOLVED OXYGEN
SPECIFICATIONS
Range0.0 to 10.0 mg/L
Resolution0.1 mg/L
Accuracy±0.4 mg/L ±3% of reading at 25 °C
Typical EMC± 0.1 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 420 nm
MethodAdaptation of the
18th edition,
and the reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93732A-0Reagent A5 drops
HI 93732B-0Reagent B5 drops
HI 93732C-0Reagent C10 drops
REAGENT SET
HI 93732-01 Reagents for 100 tests
HI 93732-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Dissolved Oxygen
the
Method Selection
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
Azide modified Winkler method. The reaction between dissolved oxygen
method using the procedure described in
• Fill one 60 mL glass bottle completely with the unreacted sample.
• Replace the cap and ensure that a small part of the sample spills
over.
• Remove the cap and add 5 drops of HI 93732A-0 and 5 drops of
HI 93732B-0.
• Add more sample, to fill the bottle completely. Replace the cap
again and ensure that a part of the sample spills over. This is to
make sure that no air bubbles have been trapped inside, which
could alter the reading.
• Invert several times the bottle. The sample becomes orange-yellow
and a flocculent agent will appear.
Dissolved Oxygen
48
• Let the sample stand and the flocculent agent will start to settle.
• After approximately 2 minutes, when the upper half of the bottle becomes limpid,
add 10 drops of HI 93732C-0.
• Replace the cap and invert the bottle until the settled flocculent dissolves completely.
The sample is ready for measurement when it is yellow and completely limpid.
• Fill the cuvette up to the mark with 10 mL of the unreacted
(original) sample, and replace the cap. This is the blank.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the
meter is zeroed and ready for measurement.
10 mL
• Remove the cuvette.
• Fill another cuvette up to the mark with 10 mL of the reacted
sample and replace the cap.
• Reinsert the cuvette into the instrument.
• Press Read to start the reading. The instrument will display the results in mg/L of dissolved oxygen.
INTERFERENCES
Interferences may be caused by reducing and oxidizing materials.
49
Dissolved Oxygen
pH
SPECIFICATIONS
Range6.5 to 8.5 pH
Resolution0.1 pH
Accuracy±0.1 pH at 25 °C
Typical EMC±0.1 pH
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the Phenol Red method. The reaction with the reagent causes a yellow
to red tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93710-0Phenol Red Indicator5 drops
REAGENT SETS
HI 93710-01 Reagents for 100 pH tests
HI 93710-03 Reagents for 300 pH tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the pH method using the procedure described in the
Method Selection
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
section (see page 12).
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
pH
50
• Remove the cuvette and add 5 drops of HI 93710-0 Phenol
Red Indicator. Replace the cap and mix the solution.
• Reinsert the cuvette into the instrument.
• Press the Read key to start the reading. The instrument displays the pH value.
51
pH
PHOSPHATE HIGH RANGE
SPECIFICATIONS
Range0.0 to 30.0 mg/L
Resolution0.1 mg/L
Accuracy±1 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.1 mg/L
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the
18th edition,
causes a blue tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93717A-0Molybdate10 drops
HI 93717B-0Reagent B1 packet
REAGENT SETS
HI 93717-01 Reagents for 100 tests
HI 93717-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
in the
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Phosphate HR
Method Selection
method using the procedure described
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
Amino Acid method. The reaction between phosphate and reagents
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add 10 drops of HI 93717A-0 Molybdate reagent.
Phosphate HR
52
• Add the content of one packet of HI 93717B-0 Phosphate
HR Reagent B to the cuvette. Replace the cap and shake
gently until completely dissolved.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of phosphate (PO
3
¯).
4
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of phosphorus (P) and phosphorus pentoxide
(P2O5).
• Press or to return to the measurement screen.
INTERFERENCES
Sulfide
Chloride above 150000 mg/L )
Calcium above 10000 mg/L as CaCO
Magnesium above 40000 mg/L as CaCO
Ferrous iron above 100 mg/L
3
3
53
Phosphate HR
PHOSPHATE LOW RANGE
SPECIFICATIONS
Range0.00 to 2.50 mg/L
Resolution0.01 mg/L
Accuracy±0.04 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.01 mg/L
Light SourceTungsten lamp with narrow band interference filter @ 610 nm
MethodAdaptation of the Ascorbic Acid method. The reaction between phosphate and the
reagent causes a blue tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93713-0Powder reagent1 packet
REAGENT SETS
HI 93713-01 Reagents for 100 tests
HI 93713-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
described in the
• Rinse, cap and shake the cuvette several times with unreacted
sample. Fill the cuvette with 10 mL of sample up to the
mark and replace the cap.
Phosphate LR
Method Selection
method using the procedure
section (see page 12).
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette and add the content of one packet of
HI 93713-0 reagent. Replace the cap and shake gently (for
about 2 minutes) until the powder is completely dissolved.
Phosphate LR
54
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 3 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays concentration in mg/L of phosphate (PO
3
¯).
4
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of phosphorus (P) and phosphorus pentoxide
(P2O5).
• Press or to return to the measurement screen.
Range0.0 to 15.0 mg/L
Resolution0.1 mg/L
Accuracy±0.3 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.2 mg/L
Light SourceTungsten lamp with narrow band interference filter @ 525 nm
MethodAdaptation of the
18th edition,
causes a blue tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93706A-0Molybdate10 drops
HI 93706B-0Amino Acid Powder1 packet
REAGENT SETS
HI 93706-01 Reagents for 100 tests
HI 93706-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
the
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Phosphorus
Method Selection
method using the procedure described in
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
Amino Acid method. The reaction between phosphate and reagents
10 mL
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add 10 drops of HI 937O6A-0 Molybdate reagent.
Phosphorus
56
• Add the content of one packet of HI 93706B-0 Phosphorus
Reagent B (Amino Acid) to the cuvette. Replace the cap and
shake gently until completely dissolved.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of phosphorus (P).
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of phosphate (PO
3
¯) and phosphorus pentoxide
4
(P2O5).
• Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO
Magnesium above 40000 mg/L as CaCO
Ferrous iron above 100 mg/L
3
3
57
Phosphorus
SILVER
SPECIFICATIONS
Range0.000 to 1.000 mg/L
Resolution0.005 mg/L
Accuracy±0.020 mg/L ±5% of reading at 25 °C
Typical EMC± 0.001 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm.
MethodAdaptation of the PAN method. The reaction between silver and reagents causes an
orange tint in the sample.
REQUIRED REAGENTS
CodeDescriptionQuantity
HI 93737A-0Buffer Reagent A1 mL
HI 93737B-0Buffer Reagent B1 mL
HI 93737C-0Indicator Reagent C 2 mL
HI 93737D-0Fixing Reagent D2 mL
HI 93703-51Dispersing Agent4-6 drops
(only when necessary, see note)
REAGENT SETS
HI 93737-01 Reagents for 50 tests
HI 93737-03 Reagents for 150 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Note: for best results perform your tests between 20-24°C.
Silver
method using the procedure described in the
Method Selection
section (see page 12).
• Fill two graduated beakers with 25 mL of sample.
• Add 1.0 mL of HI 93737A-0 Buffer reagent to one beaker (the
blank) and swirl gently to mix.
Silver
58
# 1# 2
# 1
• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second
beaker (the sample) and swirl gently to mix. Press Timer and the
display will show the countdown prior to adding reagent C or,
alternatively, wait for 2 minutes.
• Then add exactly 1.0 mL of HI 93737C-0 Indicator reagent to each
beaker and swirl. Press Timer or, alternatively, wait for
2 minutes.
• Then, in both cases, add 1.0 mL of HI 93737D-0 Fixing reagent
to each beaker and swirl. Press Timer or, alternatively, wait for
2 minutes.
10 mL
# 2
# 2# 1
• Fill one cuvette up to the mark with 10 mL of the blank.
• Add 3 drops of Dispersing Agent (HI 93703-51), replace the cap
and invert gently to mix for about 10 seconds.
• Place the cuvette into the holder and close the lid.
59
# 2# 1
10 mL
# 1
# 1
Silver
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Fill a second cuvette up to the mark with 10 mL of the reacted sample.
10 mL
# 2
• Add 3 drops of Dispersing Agent (HI 93703-51),
replace the cap and invert gently to mix for about
10 seconds.
# 2
• Insert the second cuvette into the instrument.
• Press Read to start the reading. The instrument displays the results in mg/L of silver.
Range0.00 to 3.00 mg/L
Resolution0.01 mg/L
Accuracy±0.03 mg/L ±3% of reading at 25 °C
Typical EMC±0.01 mg/L
Deviation
Light SourceTungsten lamp with narrow band interference filter @ 575 nm
MethodAdaptation of the
18th edition,
orange to a dark violet tint in the sample.
REQUIRED REAGENT
CodeDescriptionQuantity
HI 93731A-0Zinc Reagent1 packet
HI 93731B-0Cyclohexanone0.5 mL
REAGENT SETS
HI 93731-01 Reagents for 100 tests
HI 93731-03 Reagents for 300 tests
For other accessories see page 65
MEASUREMENT PROCEDURE
• Select the
the
Zinc
method using the procedure described in
Method Selection
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
Zincon method. The reaction between zinc and the reagents causes an
• Fill one graduated mixing cylinder up to the 20 mL mark
with the sample.
• Add the content of one packet of HI 93731A-0 Zinc
reagent, close the cylinder and invert several times to mix
until completely dissolved.
• Fill one cuvette with 10 mL of the reacted sample up to
the mark.
61
20 mL
10 mL
Zinc
• Place the cap and insert the cuvette into the instrument
and close the lid.
• Press the Zero key. The display will show “-0.0-” when
the meter is zeroed and ready for measurement.
• Remove the cuvette and add 0.5 mL of HI 93731B-0
Cyclohexanone to the cuvette.
Note: To prevent any contamination from the polycarbonate
cap, prior to replacing it, close the sample cuvette with
the supplied HDPE plastic stopper.
• Replace the cap and mix the sample for 15 seconds.
• Insert the sample into the instrument.
• Press Timer and the display will show the countdown
prior to the measurement or, alternatively, wait for
3 minutes and 30 seconds and press Read. When the
timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of zinc.
INTERFERENCES
Interference may be caused by:
Aluminum above 6 mg/L
Cadmium above 0.5 mg/L
Copper above 5 mg/L
Iron above 7 mg/L
Manganese above 5 mg/L
Nickel above 5 mg/L
Zinc
62
ERRORS AND WARNINGS
The instrument shows clear warning messages when erroneous conditions appear and when measured values are
outside the expected range. These messages are described bellow.
No Light: The light source is not functioning properly.
Light Leak: There is an excess amount of ambient light reaching the
detector.
Inverted cuvettes: The sample and the zero cuvettes are inverted.
Battery Low: The battery capacity is lower than 10%.
Light Low: The instrument cannot adjust the light level. Please check
that the sample does not contain any debris.
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
63
DATA MANAGEMENT
The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software.
STANDARD METHODS
DescriptionRangeMethod
Aluminum0.00 to 1.00 mg/LAluminon
Chlorine, Free0.00 to 2.50 mg/ LDPD
Chlorine, Total0.00 to 3.50 mg/LDPD
Color of Water0 to 500 PCUColorimetric Platinum Cobalt
Copper HR0.00 to 5.00 mg/LBicinchoninate
Copper LR0 to 1000 μg/LBicinchoninate
Fluoride0.00 to 2.00 mg/LSPADNS
Manganese HR0.0 to 20.0 mg/LPeriodate Oxidation
Manganese LR0 to 300 μg/LPAN
Molybdenum0.0 to 40.0 mg/LMercaptoacetic Acid
Nickel HR0.00 to 7.00 g/LPhotometric
Nickel LR0.000 to 1.000 mg/LPAN
Nitrate0.0 to 30.0 mg/LCadmium Reduction
Oxygen, Dissolved0.0 to 10.0 mg/LWinkler
pH6.5 to 8.5 pHPhenol Red
Phosphate HR0.0 to 30.0 mg/LAmino Acid
Phosphate LR0.00 to 2.50 mg/LAscorbic Acid
Phosphorus0.0 to 15.0 mg/LAmino Acid
Silver0.000 to 1.000 mg/LPAN
Zinc0.00 to 3.00 mg/LZincon
Windows® is registered Trademark of "Microsoft Co."
64
ACCESSORIES
REAGENT SETS
HI 93701-01 100 free chlorine tests (powder)
HI 93701-03 300 free chlorine tests (powder)
HI 93701-F300 free chlorine tests (liquid)
HI 93701-T300 total chlorine tests (liquid)
HI 93702-01 100 copper HR tests
HI 93702-03 300 copper HR tests
HI 93706-01 100 phosphorus tests
HI 93706-03 300 phosphorus tests
HI 93709-01 100 manganese HR tests
HI 93709-03 300 manganese HR tests
HI 93710-01 100 pH tests
HI 93710-03 300 pH tests
HI 93711-01 100 total chlorine tests (powder)
HI 93711-03 300 total chlorine tests (powder)
HI 93712-01 100 aluminum tests
HI 93712-03 300 aluminum tests
HI 93713-01 100 phosphate LR tests
HI 93713-03 300 phosphate LR tests
HI 93717-01 100 phosphate HR tests
HI 93717-03 300 phosphate HR tests
HI 93726-01 100 nickel HR tests
HI 93726-03 300 nickel HR tests
HI 93728-01 100 nitrate tests
HI 93728-03 300 nitrate tests
HI 93729-01 100 fluoride tests
HI 93729-03 300 fluoride tests
HI 93730-01 100 molybdenum tests
HI 93730-03 300 molybdenum tests
HI 93731-01 100 zinc tests
HI 93731-03 300 zinc tests
HI 93732-01 100 dissolved oxygen tests
HI 93732-03 300 dissolved oxygen tests
HI 93737-01 50 silver tests
HI 93737-03 150 silver tests
HI 93740-01 50 nickel LR tests
HI 93740-03 150 nickel LR tests
HI 93748-01 50 manganese LR tests
HI 93748-03 150 manganese LR tests
HI 95747-01 100 copper LR tests
HI 95747-03 300 copper LR tests
65
OTHER ACCESSORIES
HI 731318cloth for wiping cuvettes (4 pcs)
HI 731321glass cuvettes (4 pcs)
HI 731325Wnew cap for cuvette (4 pcs)
HI 740034cap for 100 mL beaker (6 pcs)
HI 740036100 mL plastic beaker (6 pcs)
HI 74003860 mL glass bottle and stopper
HI 7401421 mL graduated syringe
HI 7401431 mL graduated syringe (6 pcs)
HI 740144pipette tip (6 pcs)
HI 740157plastic refilling pipette (20 pcs)
HI 74022025 mL glass cylinders with caps (2 pcs)
HI 740223170 mL plastic beaker
HI 740224170 mL plastic beakers (12 pcs)
HI 74022560 mL graduated syringe
HI 7402265 mL graduated syringe
HI 740227filter assembly
HI 740228filter discs (25 pcs)
HI 740229100 mL graduated cylinder
HI 740230230 mL demineralized water
HI 92000Windows compatible software
HI 920013PC connection cable
HI 93703-50 cuvette cleaning solution (230 mL)
HI 93703-54 dried resin (100 g)
HI 93703-55 activated carbon (50 pcs)
66
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials
when used for its intended purpose and maintained according to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number
from the Customer Service Department and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of
purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, this requiring the operator to take all necessary steps to
correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not put the instrument in microwave ovens. For yours and the instrument safety do not use or store the instrument in hazardous
environments.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products
without advance notice.
HANNA LITERATURE
Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The
reference literature currently covers areas such as:
• Water Treatment
• Process
• Swimming Pools
• Agriculture
• Food
• Laboratory
and many others. New reference material is constantly being added to the library.
For these and other catalogs, handbooks and leaflets contact your dealer or the Hanna Customer Service
Center nearest to you. To find the Hanna Office in your vicinity, check our home page at www.hannainst.com.
67
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA