Hanna Instruments HI 83200-01 User Manual

Instruction Manual
HI 83200
Multiparameter Bench
Photometer
for Laboratories
www.hannainst.com
This Instrument is in
1
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the
instrument. This manual will provide you with the necessary information for the correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. This instrument is in compliance with
directives.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ............................................. 3
ABBREVIATIONS ............................................................... 3
GENERAL DESCRIPTION ..................................................... 3
SPECIFICATIONS .............................................................. 4
PRECISION AND ACCURACY ............................................... 4
PRINCIPLE OF OPERATION ................................................ 4
FUNCTIONAL DESCRIPTION ............................................... 6
NEED TO KNOW ............................................................... 7
TIPS FOR AN ACCURATE MEASUREMENT ........................... 8
HEALTH & SAFETY ..........................................................11
METHOD REFERENCE TABLE ............................................. 11
OPERATIONAL GUIDE ....................................................... 12
SETUP ........................................................................... 14
HELP MODE ................................................................... 16
SAMPLE PREPARATION .................................................... 17
ALUMINUM ....................................................................21
ALKALINITY .................................................................... 23
AMMONIA MR ................................................................ 25
AMMONIA LR ................................................................. 27
BROMINE ......................................................................29
CALCIUM ....................................................................... 31
FREE CHLORINE ............................................................. 33
TOTAL CHLORINE ............................................................ 36
CHLORINE DIOXIDE ......................................................... 39
CHROMIUM VI HR .......................................................... 42
CHROMIUM VI LR ........................................................... 44
COLOR OF WATER ........................................................... 46
COPPER HR ....................................................................48
COPPER LR .................................................................... 50
CYANIDE ........................................................................52
CYANURIC ACID .............................................................. 55
FLUORIDE ...................................................................... 57
CALCIUM HARDNESS ....................................................... 59
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
MAGNESIUM HARDNESS ................................................. 62
HYDRAZINE .................................................................... 65
IODINE .......................................................................... 67
IRON HR ........................................................................69
IRON LR ......................................................................... 71
MAGNESIUM .................................................................. 74
MANGANESE HR ............................................................. 76
MANGANESE LR ............................................................. 78
MOLYBDENUM ............................................................... 81
NICKEL HR ..................................................................... 84
NICKEL LR ...................................................................... 86
NITRATE ......................................................................... 89
NITRITE HR .................................................................... 91
NITRITE LR ..................................................................... 93
DISSOLVED OXYGEN ........................................................ 95
OZONE .......................................................................... 97
pH .............................................................................. 100
PHOSPHATE HR ............................................................ 102
PHOSPHATE LR ............................................................. 104
PHOSPHORUS ............................................................. 106
POTASSIUM HR ............................................................ 108
POTASSIUM MR ............................................................ 111
POTASSIUM LR ............................................................. 113
SILICA ......................................................................... 115
SILVER ........................................................................ 118
SULFATE ...................................................................... 121
ZINC............................................................................ 123
ERRORS AND WARNINGS ............................................... 125
DATA MANAGEMENT ...................................................... 126
STANDARD METHODS .................................................... 127
ACCESSORIES .............................................................. 128
WARRANTY .................................................................. 131
HANNA LITERATURE ...................................................... 131
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your Dealer.
Each Meter is supplied complete with:
• Four Sample Cuvettes and Caps
• Sample Preparation Kit (see page 17)
• Cloth for wiping cuvettes (4 pcs)
• Scissors
• AC/DC Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the instrument works correctly. Any defective item
must be returned in its original packing with the supplied accessories.
ABBREVIATIONS
°C: degree Celsius EPA: US Environmental Protection Agency °F: degree Fahrenheit g/L: grams per liter (ppt) HR: high range LR: low range mg/L: milligrams per liter (ppm) mL: milliliter MR: medium range μg/L: micrograms per liter (ppb) PAN: 1-(2-pyridylazo)-2-naphtol TPTZ: 2,4,6-tri-(2-pyridyl)-1,3,5-triazine
GENERAL DESCRIPTION
HI 83200 is a multiparameter bench photometer dedicated for Laboratory analysis. It can measure 45 different methods using specific liquid or powder reagents. The amount of reagent is precisely dosed to ensure maximum reproducibility.
HI 83200 bench photometer can be connected to a PC via an USB cable. The optional HI 92000
®
Windows
Compatible Software helps users manage all their results.
2 3
SPECIFICATIONS
Light Life Life of the instrument Light Detector Silicon Photocell Environment 0 to 50°C (32 to 122°F);
max 90% RH non-condensing
Power Supply external 12 Vdc power adapter
built-in rechargeable battery
Dimensions 235 x 200 x 110 mm (9.2 x 7.87 x 4.33") Weight 0.9 Kg
For specifications related to each method (e.g. range, precision, etc.) refer to the related measurement section.
PRECISION AND ACCURACY
Precision is how closely repeated measurements agree with each other. Precision is usually expressed as standard deviation (SD).
Accuracy is defined as the nearness of a test result to the true value. Although good precision suggests good accuracy, precise results can be inaccurate. The figure explains these definitions.
For each method, the precision is expressed in the related measurement section as standard deviation at a specific concentration value of the analite. The standard deviation is obtained with a single instrument
using a representative lot of reagents.
PRINCIPLE OF OPERATION
Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the matter and on the physical-chemical characteristics of substance according to the Lambert-Beer Law:
-log I/I A = ε
= ε
o
or
λ
c d
c d
λ
Where:
-log I/I
=Absorbance (A)
o
I
=intensity of incident light beam
o
I =intensity of light beam after absorption
ε
=molar extinction coefficient at wavelength λ
λ
c =molar concentration of the substance d =optical path through the substance
Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents.
Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements. The optical system of HI 83200 is based on special subminiature tungsten lamps and narrow-band interference filters to guarantee both high performance and reliable results.
Five measuring channels allow a wide range of tests.
Instrument block diagram (optical layout)
A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette. The optical path is fixed by the diameter of the cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity -I
or -I-.
o-
The photoelectric cell collects the radiation -I- that is not absorbed by the sample and converts it into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual measurement is performed.
The cuvette has a very important role because it is an optical element and thus requires particular attention. It is important that both the measurement and the calibration (zeroing) cuvette are optically identical to provide the same measurement conditions. Most of methods use the same cuvette for both, so it is important that measurements are taken at the same optical point. The instrument and the cuvette cap have special marks that must be aligned in order to obtain better reproducibility.
The surface of the cuvette must be clean and not scratched. This is to avoid measurement interference due to unwanted reflection and absorption of light. It is recommended not to touch the cuvette walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measurement phases, it is necessary to close the cuvette to prevent any contamination.
4 5
FUNCTIONAL DESCRIPTION
KEYPAD DESCRIPTION
INSTRUMENT DESCRIPTION
The keypad contains 8 direct keys and 3 functional keys with the following functions:
Press to perform the function displayed above it. The functions are screen related.
ESC
Press to exit the current screen.
Press to access the select method menu.
Press to move up in a menu or a help screen, to increment a set value, to access second level
functions.
Press to move down in a menu or a help screen, to decrement a set value, to access second
level functions.
Press to log the current reading.
RCL
Press to recall the log.
HELP
Press to display the help screen.
SETUP
Press to access the setup screen.
NEED TO KNOW
1) Cuvette Lid
2) Liquid Crystal Display (LCD).
3) Splash proof keypad.
4) ON/OFF power switch
5) Power input connector
6) USB connector
HI 83200 has a powerful interactive user support that assists the user during the analysis process. Each step in the measurement process is help supported. A tutorial mode is available in the Setup Menu.
6 7
TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure most accurate results.
• Color or suspended matter in large amounts may cause interference, they should be removed by treatment with active carbon and filtration.
• Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom of this convexity must be at the same level of the 10 mL mark.
COLLECTING AND MEASURING SAMPLES
• In order to measure exactly 0.5 mL of reagent with the 1 mL syringe: (a) push the plunger completely into the syringe and insert the tip into the solution. (b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark. (c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging
on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above the cuvette, push the plunger down into the syringe until the lower edge of the seal is exactly on the 0.5 mL mark. Now the exact amount of 0.5 mL has been added to the cuvette, even if the tip still contains some solution.
USING LIQUID AND POWDER REAGENTS
• Proper use of the dropper: (a) for reproducible results, tap the dropper on the table for several times and wipe the outside of the
dropper tip with a cloth.
(b) always keep the dropper bottle in a vertical position while dosing the reagent.
(a) (b)
• Proper use of the powder reagent packet: (a) use scissors to open the powder packet; (b) push the edges of the packet to form a spout; (c) pour out the content of the packet.
(c)(a) (b)
USING CUVETTES
• Proper mixing of the cuvette is done by shaking the cuvette, moving the cuvette up and down. The
(c)(a) (b)
movement may be gentle or vigorous. This mixing method is indicated with “shake gently” or “shake vigorously”, and one of the following icons:
shake gently shake vigorously
8 9
Pay attention to push the cuvette completely down in the holder.
HEALTH & SAFETY
The chemicals contained in the reagent kits may be hazardous if improperly handled. Read the
Material Safety Data Sheet (MSDS) before performing tests. Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully. Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water. If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.
METHOD REFERENCE TABLE
• In order to avoid reagent leaking and to obtain more accurate measure­ments, close the cuvette first with the supplied HDPE plastic stopper and then the black cap.
• Each time the cuvette is used, the cap must be tightened to the same degree.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside, and free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion.
• Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.
• It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement when possible (for most precise results follow the measurement procedures carefully).
• Discard the sample immediately after the reading is taken, or the glass might become permanently stained.
• All the reaction times reported in this manual are at 20°C (68°F). As a general rule of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).
REAGENT BLANK CORRECTION
• Some methods require a “reagent blank correction”. The blank and the sample are prepared exactly in the same way, only the blank is deionized water instead of sample. A blank cuvette may be used more than once: stability and storing conditions are described for each method in the related chapter.
INTERFERENCES
• In the method measurement section the most common interferences that may be present in an average wastewater matrix have been reported. It may be that for a particular treatment process other compounds do interfere with the method of analysis.
HI83200 - LABORATORIES
Method Method Page
description
1 Aluminum 21 2 Alkalinity 23 3 Ammonia MR 25 4 Ammonia LR 27
5 Bromine 29
6 Calcium 31 7 Free Chlorine 33 8 Total Chlorine 36
9 Chlorine Dioxide 39 10 Chromium VI HR 42 11 Chromium VI LR 44 12 Color of Water 46 13 Copper HR 48 14 Copper LR 50 15 Cyanide 52 16 Cyanuric Acid 55 17 Fluoride 57 18 Calcium Hardness 59 19 Magnesium Hardness 62 20 Hydrazine 65 21 Iodine 67 22 Iron HR 69 23 Iron LR 71
Method Method Page
description
24 Magnesium 74 25 Manganese HR 76 26 Manganese LR 78 27 Molybdenum 81 28 Nickel HR 84 29 Nickel LR 86 30 Nitrate 89 31 Nitrite HR 91 32 Nitrite LR 93 33 Dissolved Oxygen 95 34 Ozone 97 35 p H 100 36 Phosphate HR 102 37 Phosphate LR 104 38 Phosphorus 106 39 Potassium HR 108 40 Potassium MR 111 41 Potassium LR 113 42 Silica 115 43 Silver 118 44 Sulfate 121 45 Zinc 123
10 11
OPERATIONAL GUIDE
POWER CONNECTION AND BATTERY MANAGEMENT
The meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery. Note: Always turn the meter off before unplugging it to ensure no data is lost. When the meter switches ON, it verifies if the power supply adapter is connected. The battery icon on the
LCD will indicate the battery status:
- battery is charging from external adapter - battery fully charged (meter connected to AC/DC adapter)
After the desired method was selected, follow the measurement described in the related section.
Before performing a test you should read all the instructions carefully.
DATA MANAGEMENT
The instrument features a data log function to help you keep track of all your analysis. The data log can hold 200 individual measurements. Storing, viewing and deleting the data is possible using LOG and
..
RCL keys
.
..
Storing data
stored in a stack as a record with date and time stamps.
: You can store only a valid measurement. Press LOG
and the last valid measurement will be
- battery capacity (no external adapter) - battery Low (no external adapter)
- battery Dead (no external adapter)
METHOD SELECTION
Turn ON the instrument via the ON/OFF power switch.
The meter will perform an autodiagnostic test. During this test, the Hanna Instrument logo will appear on
the LCD. After 5 seconds, if the test was successful, the last selected method will appear on the display.
In order to select the desired method press METHOD and a screen with the available methods will appear.
Press   keys to highlight the desired method. Press Select.
Viewing and deleting
on the LIFO (last in, first out) scheme. Additionally, you can delete the data records all at once.
: You can view and delete the data log by pressing the RCL key. Deleting is based
CHEMICAL FORM
Chemical form conversion factors are pre-programmed into the instrument and are method specific. In order to view the displayed result in the desired chemical form press the  or  to access to the second level of functions and then press the Chem Frm functional key to toggle between the existing chemical forms for the selected method.
SPECIAL CONVERSIONS
For Magnesium and Calcium Hardness, special conversion factors can be used to convert the readings from mg/L to French degrees (°f), German degrees (°dH) and English degrees (°E) of hardness. This can be achieved by pressing the  or  to access the second level of functions and then press the Unit functional key to toggle between °f, °dH, °E and mg/L.
12 13
SETUPSETUP
SETUP
SETUPSETUP
In the Setup mode the instrument’s par amete rs can be changed. Some param eters affect the measuring sequence and others are general parameters that change the behavior or appearance of the instrument.
Press SETUP to enter the setup mode. Press ESC or SETUP to return to the main screen. A list of setup parameters will be displayed with currently configured settings. Press HELP for additional information.
Press the   new value as follows:
keys to select the parameter and select a
Backlight
Values: 0 to 8.
Press Modify functional key to access the backlight value. Use the WX functional keys or the   keys to increase/ decrease the value. Press Accept functional key to confirm or ESC to return to the setup
menu without saving the new value.
Contrast
Values: 0 to 20.
This option is used to set the display’s contrast. Press Modify functional key to change the display’s contrast. Use the WX functional keys or the   keys to increase/ decrease the value. Press Accept functional key to confirm the value or ESC to return to the
setup menu without saving the new value.
Date / Time
This option is used to set the instrument’s date and time. Press Modify functional key to change the date/time. Press the WX functional keys to highlight the value to be modified (year, month, day, hour, minute or second). Use the  keys to change the value. Press Accept functional key to confirm or ESC to return to the setup
without saving the new date or time.
Time format
Option: AM/PM or 24 hour.
Press the functional key to select the desired time format.
Date format
Press Modify functional key to change the Date Format. Use the   keys to select the desired format. Press Accept functional key to confirm or ESC to return to the setup
menu without saving the new format.
Language
Press the corresponding functional key to change the option. If the new selected language cannot be loaded, the previously selected language will be reloaded.
Tutorial
Option: Enable or Disable.
If enabled this option will provide the user short guides, related to the current screen. Press the functional key to enable/disable the tutorial mode.
Beeper
Option: Enable or Disable.
When enabled, a short beep is heard every time a key is pressed. A long beep alert sounds when the pressed key is not active or an error is detected.
Press the functional key to enable/disable the beeper.
14 15
Instrument ID
Option: 0 to 9999.
This option is used to set the instrument’s ID (identification number). The instrument ID is used while exchanging data with a PC. Press Modify functional key to access the instrument ID screen. Press the   keys in order to set the desired value. Press Accept functional key to confirm the value or ESC to return to the setup menu without saving the new value.
Meter information
Press “Select” functional key to view the Instrument model, firmware version, language version and instrument serial number.
Press ESC to return to the Setup mode.
HELP MODEHELP MODE
HELP MODE
HELP MODEHELP MODE
HELP MODE
SAMPLE PREPARATION
LIST OF MATERIAL
HI 83200 is supplied with the following accessories for sample preparation:
• 4 cuvettes (10 mL)
• 2 plastic beakers (100 and 170 mL)
• 1 graduated cylinder (100 mL)
• 1 syringe with screw rim (60 mL)
• 1 syringe (5 mL)
• 1 filter assembly
• 25 filter discs
• 1 spoon
• 2 pipettes
• Carbon powder packets (50 pcs)
• 1 Demineralizer Bottle with filter cap for about 12 liters of deionized water (depending on the hardness level of water to be treated)
SAMPLE PREPARATION PROCEDURE
Note: If the water sample to be analyzed is very turbid, let it stand in a beaker for a while until most
of the solid particles have settled. Then, use the pipette to transfer the supernatant solution to the other beaker and prepare the samples as described below. To prevent the displacement of the settled solids at the bottom of the beaker, do not induce air bubbles into the solution.
HI 83200 offers an interactive contextual help mode that assists the user at any time.
To access help screens press HELP. The instrument will display additional information related to the current screen. To read all available data, scroll the text using the   keys. Press Support functional key to access a screen with Hanna service centers and their contact details. Press Accessories functional key to access a page with instru- ment accessories. To exit support or accessories screens press ESC and the instrument will return to the previous help screen. To exit help mode just press HELP or ESC key again and the meter will display the last screen the user was in before entering help mode.
16 17
IRRIGATION WATER (LR):
• Measure 100 mL of sample with the graduated cylinder.
• If the solution contains some turbidity or color, pour it in the large 170mL beaker and add a powder packet of active carbon.
• Mix well using the spoon and then wait for 5 minutes.
• Remove the cap and fill the Demineralizer Bottle with tap water.
• Unscrew the filter assembly, put a filter disc inside and close it.
• Connect the filter assembly to the 60 mL syringe by the screw rim.
• Remove the syringe plunger and fill the syringe with the sample to be filtered. Pour the sample gently, trying to avoid transferring of the actived carbon to the syringe.
• Reinsert the plunger and filter the solution into the 100mL beaker by pushing gently the syringe plunger. The sample is now ready.
Note: Filter at least 40 mL of solution if all the four methods will be tested. If the solution is still turbid or colored, treat it again with a packet of active carbon. After use, throw the filter disc away and wash the syringe and the filter assembly well. Always use a new disc for another sample.
filter
disc
• Replace the cap and shake gently for
at least 2 minutes.
• Open the upper part of the Demineralizer Bottle cap and gently squirt the demineralized water into the cylinder, up to the 100 mL mark.
Note: The ion exchange resin contained in the Demineralizer Bottle is provided with an indicator substance. The indicator will change from green to blue when the resin has been exhausted and needs to be replaced.
• Pour the solution in the large 170mL beaker, replace the cap and invert several times to mix.
• If the solution contains some turbidity or color, add a powder packet of active carbon and follow the procedure described for Irrigation Water (LR).
NUTRIENTS SOLUTION (HR):
• Add 10 mL of sample to the graduated cylinder using the 5 mL syringe (twice).
NUTRIENTS SOLUTION (MR):
• Use the graduated cylinder to measure exactly 20 mL of sample.
18 19
Note: To measure exactly 5 mL of sample with the syringe, push the plunger completely into the syringe and insert the tip into the sample. Pull the plunger out until the lower edge of the seal is on the 5 mL mark of the syringe.
probable level of liquid
taken up by syringe
• Remove the cap and fill the Demineralizer Bottle with tap water.
• Replace the cap and shake gently for
at least 2 minutes.
ALUMINUM
SPECIFICATIONS
Range 0.00 to 1.00 mg/L Resolution 0.01 mg/L Accuracy ±0.02 mg/L ±4% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the aluminon method. The reaction between aluminum and reagents
causes a reddish tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93712A-0 Ascorbic acid 1 packet HI 93712B-0 Aluminon reagent 1 packet HI 93712C-0 Bleaching powder 1 packet
REAGENT SETS
HI 93712-01 Reagents for 100 tests HI 93712-03 Reagents for 300 tests
For other accessories see page 128.
• Open the upper part of the Demineralizer Bottle cap and squirt gently the demineralized water into the cylinder, up to the 100 mL mark.
• Pour the solution in the large 170mL beaker, replace the cap and invert several times to mix.
• If the solution contains some turbidity or color, add a powder packet of active carbon and follow the procedure described for Irrigation Water (LR).
20 21
MEASUREMENT PROCEDURE
• Select the
Method Selection
• Fill a graduated beaker with 50 mL of sample.
• Add the content of one packet of HI 93712A-0 Ascorbic acid and mix until completely dissolved.
• Add the content of one packet of HI 93712B-0 Aluminon reagent and mix until completely dissolved. This is the sample.
• Fill two cuvettes with 10 mL of sample each (up to the mark).
Aluminum
method using the procedure described in the
section (see page 12).
# 1
10 mL10 mL
# 2
Aluminum
• Add the content of one packet of HI 93712C-0 Bleaching powder to one of the two cuvettes. Replace the cap and shake vigorously until completely dissolved. This is the blank.
• Place the blank into the holder and close the lid.
• Press TIMER and the display will show the countdown prior to zeroing the blank. Alternatively wait for 15 minutes and then press ZERO. The display will show "-0.0-" when the meter is zeroed and ready for measurement.
• Remove the blank and insert the other cuvette into the instrument.
• Press the READ key and the meter will perform the reading. The instrument displays the results in mg/L of aluminum.
# 1
# 2
# 1
ALKALINITY
SPECIFICATIONS
Range 0 to 500 mg/L (as CaCO3) Resolution 5 mg/L Precision ±10 @ 100 mg/L Typical EMC ±5 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method
REQUIRED REAGENTS
Code Description Quantity/test
HI 93755-0 Alkalinity Indicator Reagent 1 packet
REAGENT SETS
HI 93755-01 Reagents for 100 tests HI 93755-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Method Selection
• Fill the cuvette with 10 mL of unreacted sample, up to the mark, and replace the cap.
Colorimetric Method. At different alkalinity levels a distinctive range of colors from yellow
to green and greenish blue will develop.
Alkalinity
method using the procedure described in the
section (see page 12).
10 mL
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of Al
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Iron above 20 mg/L Alkalinity above 1000 mg/L Phosphate above 50 mg/L Fluoride must be absent
Aluminum
22 23
2O3
• Place the cuvette into the holder and close the lid
.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Alkalinity
• Remove the cuvette.
• Add carefully the content of one packet of HI 93755-0 Alkalinity Indicator Reagent. Replace the cap and shake vigorously for 30 seconds. Note: Pay attention not to spill reagent otherwise full color development may be inhibited.
• Replace the cuvette into the holder and close the lid.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of alkalinity (CaCO
).
3
AMMONIA MEDIUM RANGE
SPECIFICATIONS
Range 0.00 to 10.00 mg/L Resolution 0.01 mg/L Accuracy ±0.05 mg/L ±5% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 420 nm Method Adaptation of the
Nessler method. The reaction between ammonia and reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93715A-0 First Reagent 4 drops (6 drops for seawater) HI 93715B-0 Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests HI 93715-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
Ammonia MR
Method Selection
the
and replace the cap.
section (see page 12).
ASTM Manual of Water and Environmental Technology, D1426-92
method using the procedure described in
,
Alkalinity
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
24 25
Ammonia MR
• Remove the cuvette.
AMMONIA LOW RANGE
• Add 4 drops of HI 93715A-0 First Reagent (6 drops for seawater analysis). Replace the cap and mix the solution.
• Add 4 drops of HI 93715B-0 Second Reagent (10 drops for seawater analysis). Replace the cap and mix the solution.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of ammonia nitrogen (NH
-N).
3
x 4
x 4
SPECIFICATIONS
Range 0.00 to 3.00 mg/L Resolution 0.01 mg/L Accuracy ±0.04 mg/L ±4% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 420 nm Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92
Nessler method. The reaction between ammonia and reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93700A-0 First Reagent 4 drops (6 drops for seawater) HI 93700B-0 Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93700-01 Reagents for 100 tests HI 93700-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
Ammonia LR
Method Selection
section (see page 12).
and replace the cap.
method using the procedure described in the
,
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of ammonia (NH
) and ammonium (NH
3
+
4
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic chloramines, sulfide, various aliphatic and aromatic amines.
Ammonia MR
26 27
• Place the cuvette into the holder and close the lid.
).
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Ammonia LR
• Remove the cuvette.
• Add 4 drops of HI 93700A-0 First Reagent (6 drops for seawater analysis). Replace the cap and mix the solution.
• Add 4 drops of HI 93700B-0 Second Reagent (10 drops for seawater analysis). Replace the cap and mix the solution.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of ammonia nitrogen (NH
-N).
3
x 4
x 4
BROMINE
SPECIFICATIONS
Range 0.00 to 8.00 mg/L Resolution 0.01 mg/L Accuracy ±0.08 mg/L ±3% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
18
th
edition
pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93716-0 DPD Reagent 1 packet
REAGENT SETS
HI 93716-01 Reagents for 100 tests HI 93716-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the
Bromine
method using the procedure described in the
Method Selection
section (see page 12).
mark) and replace the cap.
Standard Methods for the Examination of Water and Wastewater,
, DPD method. The reaction between bromine and the reagent causes a
10 mL
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of ammonia (NH
) and ammonium (NH
3
+
4
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic chloramines, sulfide, various aliphatic and aromatic amines.
Ammonia LR
28 29
• Place the cuvette into the holder and close the lid.
).
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Bromine
• Remove the cuvette and add the content of one packet of HI 93716-0 DPD reagent. Replace the cap and shake gently for about 20 seconds to dissolve most of the reagent.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of bromine.
CALCIUM
SPECIFICATIONS
Range 0 to 400 mg/L Resolution 10 mg/L Accuracy ±10 mg/L ±5% of reading Light Source Tungsten lamp with narrow band interference filter @ 466 nm Method Adaptation of the Oxalate method.
REQUIRED REAGENTS
Code Description Quantity
- Buffer Reagent 4 drops HI 93752A-0 Ca Calcium Buffer Reagent 7 mL
BB
HI 93752
B-0 Ca Calcium Oxalate Reagent 1 mL
BB
REAGENT SETS
HI 937521-01 Reagents for 50 tests HI 937521-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure at page 17.
Calcium
• Select the
method using the procedure described in the
• Using the 5 mL syringe add exactly 3.00 mL of sample to the cuvette.
Method Selection
section (see page 12).
INTERFERENCES
Interference may be caused by: Chlorine, Iodine, Ozone, Oxidized forms of Chromium and Manganese. In case of water with hardness greater than 500 mg/L CaCO minutes after adding the reagent. In case of water with alkalinity greater than 250 mg/L CaCO
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
Bromine
30 31
, shake the sample for approximately 2
3
or acidity greater than 150 mg/L CaCO3,
3
3 mL of sample
• Add 4 drops of Buffer Reagent.
• Use the pipette to fill the cuvette up to the 10 mL mark with the HI 93752A-0 Calcium Buffer Reagent.
Calcium
• Replace the cap and invert several times to mix.
• Place the cuvette into the holder and close the lid.
FREE CHLORINE
FREE CHLORINEFREE CHLORINE
FREE CHLORINE
FREE CHLORINEFREE CHLORINE
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Using the 1 mL syringe, add exactly 1 mL of the HI 93752B-0 Calcium Oxalate Reagent. Replace the cap and invert the cuvette 10 times to mix.
• Press TIMER or wait for 5 minutes, then invert again the cuvette 10 times. Reinsert the cuvette into the instrument.
• Press READ to start the reading. The instrument displays the results in mg/L of Calcium.
Note: To ensure accurate results, perform the tests at room temperature, between 18°C and 28°C (65°F
to 83°F).
INTERFERENCES:
Interferences may be caused by: Acidity (as CaCO Alkalinity (as CaCO Magnesium (Mg
) above 1000 mg/L
3
) above 1000 mg/L
3
2+
) above 400 mg/L
SPECIFICATIONS
Range 0.00 to 2.50 mg/L Resolution 0.01 mg/L Accuracy ±0.03 mg/L ±3% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
EPA DPD method 330.5.
The reaction between free chlorine and the
DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER:
Code Description Quantity HI 93701-0 DPD 1 packet
LIQUID:
Code Description Quantity HI 93701A-F DPD1 Indicator 3 drops HI 93701B-F DPD1 Buffer 3 drops
REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid) HI 93701-01 Reagents for 100 tests (powder) HI 93701-03 Reagents for 300 tests (powder)
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the
• Place the cuvette into the holder and close the lid.
in the
Method Selection
Free Chlorine
method using the procedure described section (see page 12).
mark) and replace the cap.
10 mL
Calcium
32 33
Free Chlorine
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Reinsert the cuvette into the instrument.
• Remove the cuvette.
Powder reagents procedure
• Add the content of one packet of HI 93701 DPD reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes for seawater analysis).
• Wait for a minute to allow the undissolved reagent to precipitate and reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 1 minute and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of free chlorine.
• Press READ to start the reading. The instrument displays the results in mg/L of free chlorine.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese. In case of water with hardness greater than 500 mg/L CaCO
minutes after adding the powder reagent. In case of water with alkalinity greater than 250 mg/L CaCO
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
, shake the sample for approximately 2
3
or acidity greater than 150 mg/L CaCO3,
3
Liquid reagents procedure
• To an empty cuvette add 3 drops of HI 93701A-F DPD1 indicator and 3 drops of HI 93701B-F DPD1 buffer. Swirl gently to mix, and immediately add 10 mL of unreacted sample. Replace the cap and shake gently again.
34 35
x 3x 3
Free ChlorineFree Chlorine
TOTAL CHLORINE
SPECIFICATIONS
Range 0.00 to 3.50 mg/L Resolution 0.01 mg/L Accuracy ±0.03 mg/L ±3% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER:
Code Description Quantity HI 93711-0 DPD 1 packet
LIQUID:
Code Description Quantity HI 93701A-T DPD1 indicator 3 drops HI 93701B-T DPD1 buffer 3 drops HI 93701C DPD3 solution 1 drop
REAGENT SETS
HI 93701-T Reagents for 300 total chlorine tests (liquid) HI 93711-01 Reagents for 100 total chlorine tests (powder) HI 93711-03 Reagents for 300 total chlorine tests (powder)
For other accessories see page 128.
EPA DPD method 330.5.
The reaction between the chlorine and the
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
Powder reagents procedure
• Add 1 packet of HI 93711 DPD reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes for seawater analysis).
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of total chlorine.
MEASUREMENT PROCEDURE
• Select the scribed in the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
Total Chlorine
Total Chlorine
Method Selection
method using the procedure de-
section (see page 12).
36 37
10 mL
Liquid reagents procedure
• To an empty cuvette add 3 drops of HI 93701A-T DPD1 indicator, 3 drops of HI 93701B-T DPD1 buffer and 1 drop of HI 93701C DPD3 solution. Swirl gently to mix and
immediately add 10 mL of unreacted sample.
Replace the cap and shake gently again.
• Reinsert the cuvette into the instrument.
x 3
x 3
x 1
Total Chlorine
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of total chlorine.
Note: free and total chlorine have to be measured separately with fresh unreacted samples following the related procedure if both values are requested.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese. In case of water with hardness greater than 500 mg/L CaCO
minutes after adding the powder reagent. In case of water with alkalinity greater than 250 mg/L CaCO
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
, shake the sample for approximately 2
3
or acidity greater than 150 mg/L CaCO3,
3
CHLORINE DIOXIDE
SPECIFICATIONS
Range 0.00 to 2.00 mg/L Resolution 0.01 mg/L Accuracy ±0.10 mg/L ±5% of reading Typical EMC ±0.01 mg/L Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the Chlorophenol Red method. The reaction between chlorine dioxide and
REQUIRED REAGENT
Code Description Quantity HI 93738A-0 Reagent A 1 mL HI 93738B-0 Dechlorinating Reagent B 1 packet HI 93738C-0 Reagent C 1 mL HI 93738D-0 Reagent D 1 mL
REAGENT SETS
HI 93738-01 Reagents for 100 tests HI 93738-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the described in the
reagents causes a colorless to purple tint in the sample.
Chlorine Dioxide
Method Selection
method using the procedure
section (see page 12).
Total Chlorine
• Fill two graduated mixing cylinders (#1 & #2) up to the 25 mL mark with the sample.
• Add 0.5 mL of HI 93738A-0 Chlorine Dioxide Reagent to each cylinder (#1 & #2), close them and invert several times to mix.
#1 #2
38 39
#1
25 ml
#1 & #2
#2
25 ml
Chlorine Dioxide
• Add the content of one packet of HI 93738B-0 Dechlorinating Reagent to one of the two cylinders (#1), close and invert it several times until it is totally dissolved. This is the blank.
#1
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Add
precisely 0.5 mL of HI 93738C-0 Chlorine Dioxide Reagent to each cylinder (#1 & #2), close them
and invert several times to mix.
#1 & #2
#1
#2
• Add 0.5 mL of HI 93738D-0 Chlorine Dioxide Reagent to each cylinder (#1 & #2), close them and invert several times to mix. Cylinder #2 is the reacted sample.
#1 & #2
#1
• Fill a cuvette with 10 mL of the blank (#1) up to the mark and replace
#2
#1
the cap.
10 mL
• Place the blank (#1) into the holder and close the lid.
• Fill another cuvette with 10 mL of the reacted sample (#2) up to the mark and replace the cap.
#2
#2
10 mL
• Insert the sample into the instrument.
• Press READ and the meter will perform the reading. The instrument displays the results in mg/L of chlorine dioxide.
SAMPLING PROCEDURE
It is recommended to analyze chlorine dioxide samples immediately after collection. Chlorine dioxide samples must be stored in sealed dark glass bottle, with minimal head space. Excessive heat (above 25°C/78°F), agitation and exposure to light must be avoided.
INTERFERENCES
Interferences may be caused by strong oxidants.
Chlorine Dioxide
#1
40 41
Chlorine Dioxide
CHROMIUM VI HIGH RANGE
SPECIFICATIONS
Range 0 to 1000 µg/L Resolution 1 µg/L Accuracy ±5 µg/L ±4% of reading Typical EMC ±1 µg/L Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93723-0 Powder reagent 1 packet
REAGENT SETS
HI 93723-01 Reagents for 100 tests HI 93723-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette up to the mark with 10 mL of unreacted sample and
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is
Chromium VI HR
the
Method Selection
section (see page 12).
replace the cap.
zeroed and ready for measurement.
ASTM Manual of Water and Environmental Technology, D1687-92
method using the procedure described in
10 mL
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in µg/L of chromium VI.
,
• Press the  or  to access the second level of functions.
2-
• Press the Chem Frm functional key to convert the result in µg/L of Chromate (CrO
2-
(Cr
).
2O7
) and Dichromate
4
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed Iron above 1 ppm Mercurous and mercuric ions cause slight inhibition of the reaction.
• Remove the cuvette and add the content of one packet of HI 93723-0 reagent. Replace the cap and shake vigorously for about 10 seconds.
Chromium VI HR
42 43
Chromium VI HR
CHROMIUM VI LOW RANGE
SPECIFICATIONS
Range 0 to 300 µg/L Resolution 1 µg/L Accuracy ±1 µg/L ±4% of reading Typical EMC ±1 µg/L Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93749-0 Powder reagent 1 packet
REAGENT SETS
HI 93749-01 Reagents for 100 tests HI 93749-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette up to the mark with 10 mL of unreacted sample
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is
Chromium VI LR
the
Method Selection
section (see page 12).
and replace the cap.
zeroed and ready for measurement.
ASTM Manual of Water and Environmental Technology, D1687-92
method using the procedure described in
10 mL
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in µg/L of chromium VI.
,
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in µg/L of Chromate (CrO
2-
(Cr
).
2O7
2-
) and Dichromate
4
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed. Iron above 1 ppm Mercurous and mercuric ions cause slight inhibition of the reaction.
• Remove the cuvette and add the content of one packet of HI 93749-0 reagent. Replace the cap and shake vigorously for about 10 seconds.
Chromium VI LR
44 45
Chromium VI LR
COLOR OF WATER
SPECIFICATIONS
Range 0 to 500 PCU (Platinum Cobalt Units) Resolution 1 PCU Accuracy ±10 PCU ±5% of reading Typical EMC ± 1 PCU
Deviation Light Source Tungsten lamp with narrow band interference filter @ 420 nm Method Adaptation of the
18
th
edition
REQUIRED ACCESSORIES
0.45 µm membrane for true color measurement.
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill one cuvette up to the mark with deionized water and replace
• Place the blank (# 1) into the holder and close the lid.
Color of Water
Method Selection
the
section (see page 12).
the cap. This is the blank.
Standard Methods for the Examination of Water and Wastewater,
, Colorimetric Platinum Cobalt method.
method using the procedure described in
# 1
10 mL
# 1
• Fill the second cuvette up to the mark with unfiltered sample and replace the cap. This is the apparent color.
• Filter 10 mL of sample through a filter with a 0.45 µm membrane into the third cuvette, up to the 10 mL mark and replace the cap. This is the true color.
• Insert the apparent color cuvette (# 2) into the instrument and
# 2
close the lid.
• Press READ to start the reading.
• The meter displays the value of apparent color in PCU.
• Remove the cuvette, insert the true color cuvette (# 3) into the instrument and ensure that the notch on the cap is positioned securely into the groove.
10 mL
# 2
# 3
# 3
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank.
Color of Water
46 47
• Press READ to start the reading. The meter displays the value of true color in PCU.
Color of Water
COPPER HIGH RANGE
• Remove the cuvette.
SPECIFICATIONS
Range 0.00 to 5.00 mg/L Resolution 0.01 mg/L Accuracy ±0.02 mg/L ±4% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the
EPA method.
The reaction between copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93702-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests HI 93702-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is
Copper HR
method using the procedure described in the
Method Selection
section (see page 12).
and replace the cap.
zeroed and ready for measurement.
• Add the content of one packet of HI 93702-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of copper.
10 ml
INTERFERENCES
Interference may be caused by: Silver Cyanide For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted between 6 and 8.
Copper HR
48 49
Copper HR
COPPER LOW RANGE
• Remove the cuvette.
SPECIFICATIONS
Range 0 to 1000 μg/L Resolution 1 μg/L Accuracy ±10 μg/L ±5% of reading Typical EMC ±1 μg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the
EPA method.
The reaction between copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93747-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93747-01 Reagents for 100 tests HI 93747-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is
Copper LR
method using the procedure described in
Method Selection
the
section (see page 12).
mark) and replace the cap.
zeroed and ready for measurement.
• Add the content of one packet of HI 93747-0 Bicinchoninate. Replace the cap and shake gently for about 15 seconds.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of copper.
10 ml
INTERFERENCES
Interference may be caused by: Silver Cyanide For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted between 6 and 8.
Copper LR
50 51
Copper LR
CYANIDE
SPECIFICATIONS
Range 0.000 to 0.200 mg/L Resolution 0.001 mg/L Accuracy ±0.005 mg/L ±3% of reading Typical EMC Dev. ±0.001 mg/L Light Source Tungsten lamp with narrow band interference filter @ 610 nm Method Adaptation of the
18
th
edition
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93714A-0 Reagent A 1 spoon HI 93714B-0 Reagent B 1 packet HI 93714C-0 Reagent C 1 packet
REAGENT SETS
HI 93714-01 Reagents for 100 tests HI 93714-03 Reagents for 300 tests
For other accessories see page 128.
Standard Methods for the Examination of Water and Wastewater,
, Pyridine-Pyrazolone method. The reaction between cyanide and reagents
Note: Pay attention to the way the spoon is filled:
- do not press the powder;
- do not overfill it.
• Place the HDPE plastic stopper and cap
immediately, to prevent the escape of chlorine gas which is developed during the reaction, and shake gently for 30 seconds.
• Wait for 30 seconds leaving the cuvette tightly capped and undis­turbed, then add the content of one packet of HI 93714B-0 reagent and shake gently for 10 seconds.
• Immediately add the content of one packet of HI 93714C-0 reagent, replace the cap and shake vigorously for 20 seconds.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is
• Remove the cuvette and add 1 level spoon of HI 93714A-0 Cyanide
Cyanide
Cyanide
method using the procedure described in the
Method Selection
section (see page 12).
mark) and replace the cap.
zeroed and ready for measurement.
Reagent. Remember to close the reagent bottle immediately after use.
52 53
10 mL
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 25 minutes and press READ. When the timer ends the meter will perform the reading.
Note: Gently shake the cuvette 4 or 5 times during the first 20 minutes of the countdown prior to the measurement. Accuracy is not affected by undissolved reagent powder.
• The instrument displays the results in mg/L of cyanide.
Cyanide
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of Potassium Cyanide (KCN).
• Press the  or  to go back to the measurement screen. Note: for most accurate results perform the test at 20-25 °C.
INTERFERENCES
Interference may be caused by large amounts of turbidity that will cause high readings. Oxidizing (such as chlorine) or reducing agents (such as sulfide or sulfur dioxide) are known to interfere with
the measurement. Distillation will remove these. Samples with high pH values should be adjusted to approximately pH 7 before testing.
CAUTION: cyanides, their solutions, and hydrogen cyanide liberated by acids, are very poisonous.
CYANURIC ACID
SPECIFICATIONS
Range 0 to 80 mg/L Resolution 1 mg/L Accuracy ±1 mg/L ±15% of reading Typical EMC ±1 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the turbidimetric method. The reaction between cyanuric acid and the
REQUIRED REAGENTS
Code Description Quantity HI 93722-0 Powder reagent 1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests HI 93722-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the scribed in the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
reagent causes a white suspension in the sample.
Cyanuric Acid
method using the procedure de-
Method Selection
section (see page 12).
10 mL
Cyanide
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Fill a graduated beaker up to the 25 mL mark with the sample, add the content of one packet of HI 93722-0 reagent and swirl gently to mix.
54 55
Cyanuric Acid
• Fill a second cuvette with 10 mL of the reacted sample up to the
10 mL
mark. Replace the cap.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays concentration in mg/L of cyanuric acid.
FLUORIDE
SPECIFICATIONS
Range 0.00 to 2.00 mg/L Resolution 0.01 mg/L Accuracy ±5% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the
18
th
edition
causes a red tint in the sample.
REQUIRED REAGENT
Code Description Quantity HI 93729-0 SPADNS Reagent 4 mL
REAGENT SETS
HI 93729-01 Reagents for 100 tests HI 93729-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Fluoride
Method Selection
method using the procedure described in the
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
, SPADNS method. The reaction between fluoride and the liquid reagent
Cyanuric Acid
• Add 2 mL of HI 93729-0 SPADNS Reagent to two cuvettes.
• Fill one of the cuvettes with distilled water up to the mark, replace the cap and invert several times to mix.
• Fill the other cuvette with sample up to the mark, replace the cap and invert several times to mix.
• Place the cuvette with the reacted distilled water (# 1) into the holder and close the lid.
56 57
10 ml
# 1
2 ml
10 ml
# 2
# 1
Fluoride
• Press TIMER and the display will show the countdown prior to zeroing the blank or, alternatively, wait for two minutes and press ZERO. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
# 2
• Insert the other cuvette (# 2) with the reacted sample into the instrument.
• Press READ to start reading. The instrument displays the results in mg/L of fluoride.
CALCIUM HARDNESS
SPECIFICATIONS
Range 0.00 to 2.70 mg/L Resolution 0.01 mg/L Accuracy ±0.11 mg/L ±5% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
th
18
edition,
reddish-violet tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93720A-0 Ca & Mg indicator 0.5 mL HI 93720B-0 Alkali solution 0.5 mL HI 93720C-0 EGTA solution 1 drop
REAGENT SETS
HI 93720-01 Reagents for 100 tests HI 93720-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the dure described in the page 12).
Calcium Hardness
Standard Methods for the Examination of Water and Wastewater,
Calmagite method. The reaction between calcium and reagents causes a
method using the proce-
Method Selection
section (see
Note: For wastewater or seawater samples, before performing measurements, distillation is required. For most accurate results, use two graduated pipettes to deliver exactly 8 mL of distilled water and 8 mL of sample.
INTERFERENCES
Interferences may be caused by: Alkalinity (as CaCO Aluminum above 0.1 mg/L Iron, ferric above 10 mg/L Chloride above 700 mg/L Phosphate, ortho above 16 mg/L Sodium hexametaphosphate above 1.0 mg/L Sulfate above 200 mg/L Highly colored and turbid samples may require distillation Highly alkaline samples can be neutralized with nitric acid.
Fluoride
) above 5000 mg/L
3
58 59
• Rinse a graduated beaker several times with unreacted sample, before filling it to the 50 mL mark with the sample.
• Add 0.5 mL of HI 93720A-0 Calcium indicator solution and swirl to mix.
# 1 # 2
• Add 0.5 mL of HI 93720B-0 Alkali solution and swirl to mix. Use this solution to rinse 2 cuvettes before filling them up to the 10 mL mark.
Hardness Ca
• Add 1 drop of HI 93720C-0 EGTA solution to one cuvette (# 1), replace the cap and invert the cuvette several times to mix. This is the blank.
• Place the blank (# 1) into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank and insert the second cuvette (# 2) into the instrument.
• Press READ to start the reading. The instrument dis­plays concentration in mg/L of calcium hardness, as CaCO
.
3
# 1
# 1
# 2
• Press the Unit functional key to change the current measurement unit. The results can be converted to
French degrees (°f), German degrees (°dH) and English degrees (°E).
• Press the  or  to go back to the measurement screen. Note: This test will detect any calcium contamination in the beaker, measuring syringes or sample cells. To
test cleanliness, repeat the test multiple times until you obtain consistent results. Note: For better accuracy wash glassware with HCl 6N.
SAMPLE DILUTION
This meter is designed to determine low levels of hardness, typically found in water purification systems. When testing some other sources of water, it is not uncommon to come across levels of hardness that are greater than the range of this meter. This problem can be overcome through dilution. Dilutions must be performed with hardness-free water or the
readings will be erroneous. A dilution to reduce the level of hardness by a factor of one hundred is performed as follows:
• Fill a 1 mL syringe with the sample.
• Place the syringe in a 50 mL beaker, making sure that the beaker is clean and empty, and inject 0.5 mL into the beaker.
• Fill the beaker up to the 50 mL mark with hardness-free water.
INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of Calcium (Ca).
Hardness Ca
60 61
Hardness Ca
MAGNESIUM HARDNESS
SPECIFICATIONS
Range 0.00 to 2.00 mg/L Resolution 0.01 mg/L Accuracy ±0.11 mg/L ±5% of reading Typical EMC ±0.02 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
18
th
edition,
causes a reddish-violet tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93719A-0 Mg indicator 0.5 mL HI 93719B-0 Alkali solution 0.5 mL HI 93719C-0 EDTA solution 1 drop HI 93719D-0 EGTA solution 1 drop
REAGENT SETS
HI 93719-01 Reagents for 100 tests HI 93719-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the scribed in the
• Rinse a graduated beaker several times with unreacted sample, before filling it to the 50 mL mark with the sample.
Magnesium Hardness
Method Selection
Standard Methods for the Examination of Water and Wastewater,
EDTA colorimetric method. The reaction between magnesium and reagents
method using the procedure de-
section (see page 12).
10 mL 10 mL
• Fill both cuvettes up to the 10 mL mark.
# 1
# 2
• Add 1 drop of HI 93719C-0 EDTA solution to one cuvette (# 1), replace the cap and invert the cuvette several times to mix. This is the blank.
• Add 1 drop of HI 93719D-0 EGTA solution to the second cuvette (# 2), replace the cap and invert the cuvette several times to
# 2
# 1
mix. This is the sample.
# 1
• Place the blank (# 1) into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank (# 1), insert the sample (# 2) into the instrument, and close the lid.
# 2
• Add 0.5 mL of HI 93719A-0 Magnesium indicator solution, then swirl to mix.
• Add 0.5 mL of HI 93719B-0 Alkali solution and swirl to mix. Use this solution to rinse 2 cuvettes.
Hardness Mg
62 63
• Press READ to start the reading. The instrument displays concentration in mg/L of magnesium hardness, as CaCO
.
3
Hardness Mg
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of Magnesium (Mg).
• Press the Unit functional key to change the current measurement unit. The results can be converted to
French degrees (°f), German degrees (°dH) and English degrees (°E).
• Press the  or  to go back to the measurement screen.
Note: This test will detect any magnesium contamination in the beakers, measuring syringes or sample cells. To test cleanliness, repeat the test multiple times until you obtain consistent results.
SAMPLE DILUTION
This meter is designed to determine hardness typically found in water purification systems. In order to measure samples with high hardness, follow dilution procedure explained on page 61 (Ca Hardness).
INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
HYDRAZINE
SPECIFICATIONS
Range 0 to 400 µg/L Resolution 1 µg/L Accuracy ±4% of full scale Typical EMC ±2 µg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 420 nm Method Adaptation of the
D1385-88,
and the liquid reagent causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity HI 93704-0 Liquid Reagent 24 drops
REAGENT SETS
HI 93704-01 Reagents for 100 tests HI 93704-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill one cuvette up to the mark with 10 mL of distilled water.
in the
Method Selection
Hydrazine
method using the procedure described
section (see page 12).
ASTM Manual of Water and Environmental Technology, method
p-Dimethylaminobenzaldehyde method. The reaction between hydrazine
10 mL
# 1
Hardness Mg
• Place the cap, insert the cuvette # 1 into the holder and close the lid.
• Press the Blank function key to start adjusting the light level. The display will show “Blank Done” when the meter is ready to take a zero measurement.
• Fill a second cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
64 65
# 1
10 mL
# 2
Hydrazine
• Add 12 drops of the HI 93704-0 reagent to each cuvette. Replace the caps and shake gently to mix.
• Place the blank (#1) into the holder and close the lid.
• Press TIMER and the display will show the countdown prior to zeroing the blank. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank.
• Insert the cuvette with the reacted sample (# 2) into the instrument and close the lid.
• Press READ to start the reading. The instrument displays concentration in µg/L of hydrazine.
# 1
x 12
# 2
x 12
# 1
# 2
IODINE
SPECIFICATIONS
Range 0.0 to 12.5 mg/L Resolution 0.1 mg/L Accuracy ±0.1 mg/L ±5% of reading Typical EMC ±0.1 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
18
th
edition
tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93718-0 DPD Reagent 1 packet
REAGENT SETS
HI 93718-01 Reagents for 100 tests HI 93718-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the
Iodine
method using the procedure described in
the
Method Selection
section (see page 12).
the mark) and replace the cap.
meter is zeroed and ready for measurement.
Standard Methods for the Examination of Water and Wastewater,
, DPD method. The reaction between iodine and the reagent causes a pink
10 mL
INTERFERENCES
Interference may be caused by: Highly colored samples Highly turbid samples Aromatic amines
Hydrazine
• Remove the cap and add the content of one packet of HI 93718-0 DPD reagent. Replace the cap and shake gently for about 30 seconds to dissolve most of the reagent.
66 67
Iodine
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays concentration in mg/L of iodine.
INTERFERENCES
Interference may be caused by: Bromine, Chlorine, Ozone, Oxidized forms of Chromium and Manganese. In case of water with hardness greater than 500 mg/L CaCO
minutes after adding the reagent. In case of water with alkalinity greater than 250 mg/L CaCO
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
, shake the sample for approximately 2
3
or acidity greater than 150 mg/L CaCO3,
3
IRON HIGH RANGE
SPECIFICATIONS
Range 0.00 to 5.00 mg/L Resolution 0.01 mg/L Accuracy ±0.04 mg/L ±2% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
The reaction between iron and reagents causes an orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93721-0 Powder Reagent 1 packet
REAGENT SETS
HI 93721-01 Reagents for 100 tests HI 93721-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to
• Place the cuvette into the holder and close the lid.
Iron HR
method using the procedure described
Method Selection
in the
section (see page 12).
the mark) and replace the cap.
EPA Phenantroline method 315B,
for natural and treated waters.
10 mL
Iodine
• Press ZERO key. The display will show “-0.0-” the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93721-0 reagent. Replace the cap and shake until dissolu­tion is complete.
68 69
Iron HR
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and press READ. When the timer ends the meter will perform the reading.
• The instrument displays concentration in mg/L of iron.
INTERFERENCES
Interference may be caused by: Molybdate Molybdenum above 50 ppm Calcium above 10000 ppm (as CaCO Magnesium above 100000 ppm (as CaCO Chloride above 185000 ppm.
)
3
)
3
IRON LOW RANGE
SPECIFICATIONS
Range 0 to 400 µg/L Resolution 1 µg/L Accuracy ±10 µg/L ±8% of reading Typical EMC ±1 µg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a
REQUIRED REAGENTS
Code Description Quantity HI 93746-0 TPTZ Reagent 2 packets
REAGENT SETS
HI 93746-01 Reagents for 50 tests HI 93746-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the the
Method Selection
• Fill one graduated mixing cylinder up to the 25 mL mark with deionized water.
• Add the content of one packet of HI 93746-0 TPTZ reagent, close the cylinder and shake seconds. This is the blank.
violet tint in the sample.
Iron LR
method using the procedure described in
section (see page 12).
vigorously for 30
25 ml
9
30"
12
3
6
Iron HR
• Fill a cuvette with 10 mL of the blank up to the mark and replace the cap.
• Place the cuvette into the holder and close the lid.
70 71
10 mL
Iron LR
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Fill another graduated mixing cylinder up to the 25 mL mark with the sample.
• Add the content of one packet of HI 93746-0 TPTZ reagent, close the cylinder and shake
vigorously for 30 seconds. This is the
25 ml
9
30"
12
3
6
reacted sample.
• Fill a cuvette with 10 mL of the reacted sample up to the mark and replace the cap.
INTERFERENCES
Interference may be caused by: Cadmium above 4.0 mg/L
3+
Chromium Chromium Cobalt above 0.05 mg/L Copper above 0.6 mg/L Cyanide above 2.8 mg/L Manganese above 50.0 mg/L Mercury above 0.4 mg/L Molybdenum above 4.0 mg/L Nickel above 1.0 mg/L Nitrite ion above 0.8 mg/L Sample pH should be between 3 and 4 to avoid developed color to fade or turbidity formation.
above 0.25 mg/L
6+
above 1.2 mg/L
• Insert the sample into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 30 seconds and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in µg/L of iron.
Iron LR
72 73
10 mL
Iron LR
MAGNESIUM
SPECIFICATIONS
Range 0 to 150 mg/L Resolution 5 mg/L Accuracy ±5 mg/L ±3% of reading Light Source Tungsten lamp with narrow band interference filter @ 466 nm Method Adaptation of the Calmagite method
.
• Remove the cuvette.
• Using the other 1 mL syringe, add to the cuvette exactly 0.5 mL of sample.
Note: Do not mix up the two syringes!
• Replace the cap and invert several times to mix.
0.5 mL
of sample
REQUIRED REAGENTS
Code Description Quantity
AA
HI 93752
A-0 Mg Magnesium Buffer Reagent 1 mL
AA BB
HI 93752
B-0 Mg Magnesium Indicator Reagent 9 mL
BB
REAGENT SETS
HI 937520-01 Reagents for 50 tests HI 937520-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure on page 17.
• Select the
Method Selection
• Using one 1 mL syringe add exactly 1.00 mL of HI 93752A-0 Mg Buffer reagent to the cuvette and use the pipette to fill the cuvette up to the 10 mL mark with the HI 93752B-0 Mg Indicator reagent.
• Replace the cap and invert several times to mix.
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Magnesium
method using the procedure described in the
section (see page 12).
1 mL
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 15 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of Magnesium (Mg).
INTERFERENCES
Interferences may be caused by: Acidity (as CaCO Alkalinity (as CaCO
Calcium (Ca Iron must be absent Aluminum must be absent Copper must be absent
) above 1000 mg/L
3
) above 1000 mg/L
3
2+
) above 200 mg/L
Magnesium
74 75
Magnesium
MANGANESE HIGH RANGE
SPECIFICATIONS
Range 0.0 to 20.0 mg/L Resolution 0.1 mg/L Accuracy ±0.2 mg/L ±3% of reading Typical EMC ±0.1 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
18
th
edition,
a pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93709A-0 Citrate 1 packet HI 93709B-0 Sodium periodate 1 packet
REAGENT SETS
HI 93709-01 Reagents for 100 tests HI 93709-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the described in the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
Manganese HR
Method Selection
Standard Methods for the Examination of Water and Wastewater,
Periodate method. The reaction between manganese and reagents causes
method using the procedure
section (see page 12).
10 mL
• Add the content of one packet of HI 93709B-0 Sodium Periodate reagent. Replace the cap and shake gently until completely dissolved.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 1 minute and 30 seconds and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of manganese.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of potassium permanganate (KMnO permanganate (MnO
).
4
) and
4
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add the content of one packet of HI 93709A-0 Citrate reagent. Replace the cap and shake gently until completely dissolved.
Manganese HR
76 77
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Calcium above 700 mg/L Chloride above 70000 mg/L Iron above 5 mg/L Magnesium above 100000 mg/L.
Manganese HR
MANGANESE LOW RANGE
SPECIFICATIONS
Range 0 to 300 µg/L Resolution 1 µg/L Accuracy ±2 µg/L ±3% of reading Typical EMC ±1 µg/L Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the PAN Method. The reaction between manganese and the reagents
causes an orange tint in the sample.
REQUIRED REAGENT
Code Description Quantity HI 93748A-0 Ascorbic acid 2 packets HI 93748B- 0 Alkaline-cyanide sol. 0.40 mL HI 93748C-0 0.1% PAN indicator 2 mL HI 93703-51 Dispersing Agent 4-6 drops (only when necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for 50 tests HI 93748-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill one cuvette with 10 mL of deionized water (up to the mark).
Method Selection
the
Manganese LR
section (see page 12).
method using the procedure described in
10 mL
• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent solution to each cuvette, replace the caps and shake gently.
• Add 1 mL of the HI 93748C-0 0.1% PAN indicator solution to each cuvette, replace the caps and shake gently.
• Place the cuvette with the reacted deionized water (blank) into the holder and close the lid.
• Press TIMER and the display will show the countdown prior to zeroing the blank. Alternatively wait for 2 minutes and then press ZERO. The display will show "-0.0-" when the meter is zeroed and ready for measurement.
• Fill a second cuvette with 10 mL of sample (up to the mark).
• Add the content of one packet of HI 93748A-0 Ascorbic acid to each cuvette, replace the caps and shake gently until completely dissolved.
Manganese LR
78 79
10 mL
• Insert the second cuvette with the reacted sample into the instrument.
Manganese LR
• Press READ to start the reading. The instrument displays the results in µg/L of manganese.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of potassium permanganate (KMnO
permanganate (MnO
).
4
• Press the  or  to go back to the measurement screen.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings, add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity is removed
INTERFERENCES
Interference may be caused by: Aluminum above 20 mg/L Cadmium above 10 mg/L Calcium above 200 mg/L as CaCO Cobalt above 20 mg/L Copper above 50 mg/L Iron above 10 mg/L Lead above 0.5 mg/L Magnesium above 100 mg/L as CaCO Nickel above 40 mg/L Zinc above 15 mg/L.
3
3
) and
4
MOLYBDENUM
SPECIFICATIONS
Range 0.0 to 40.0 mg/L Resolution 0.1 mg/L Accuracy ±0.3 mg/L ±5% of reading Typical EMC ±0.1 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 420 nm Method Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity HI 93730A-0 Reagent A 1 packet HI 93730B-0 Reagent B 1 packet
.
HI 93730C-0 Reagent C 1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests HI 93730-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the described in the
12).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
Molybdenum
Method Selection
method using the procedure
section (see page
Manganese LR
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
80 81 Molybdenum
• Fill one graduated mixing cylinder up to the 25 mL mark with the sample.
• Add the content of one packet of HI 93730A-0 molybdenum reagent, close the cylinder and invert it several times until completely dissolved.
• Add the content of one packet of HI 93730B-0 molybdenum reagent to the cylinder, close and invert it several times until completely dissolved.
• Add the content of one packet of HI 93730C-0 molybdenum reagent to the cylinder, close and shake it vigorously.
• Fill an empty cuvette with 10 mL of sample up to the mark and replace the cap.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for five minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of molybdenum.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of molybdate (MoO molybdate (Na
MoO4).
2
) and sodium
4
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Aluminum above 50 mg/L Chromium above 1000 mg/L Copper above 10 mg/L Iron above 50 mg/L Nickel above 50 mg/L Nitrite, as NO Sulfate above 200 mg/L Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents.
2
• Insert the cuvette into the instrument.
Molybdenum
82 83
Molybdenum
NICKEL HIGH RANGE
SPECIFICATIONS
Range 0.00 to 7.00 g/L Resolution 0.01 g/L Accuracy ±4% of reading Typical EMC ±0.02 g/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the photometric method. The reaction between nickel and the reagent
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93726-0 Powder reagent 1 packet
REAGENT SETS
HI 93726-01 Reagents for 100 tests HI 93726-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette up to the mark with 10 mL of unreacted
Nickel HR
method using the procedure described
in the
Method Selection
section (see page 12).
sample and replace the cap.
10 mL
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 1 minute and press READ. When the timer ends the meter will perform the reading.
• The instrument displays concentration in g/L of nickel.
INTERFERENCES
Interference may be caused by copper.
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93726-0 reagent. Replace the cap and shake gently until completely dissolved.
Nickel HR
84 85
Nickel HR
NICKEL LOW RANGE
SPECIFICATIONS
Range 0.000 to 1.000 mg/L Resolution 0.001 mg/L Accuracy ±0.010 mg/L ±7% of reading Typical EMC ±0.001 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the PAN method. The reaction between nickel and the reagents causes
an orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93740A-0 Phthalate-phosphate 2 packets HI 93740B-0 0.3% PAN indicator 2 mL HI 93740C-0 EDTA 2 packets HI 93703-51 Dispersing Agent 4-6 drops (only when necessary, see note)
REAGENT SETS
HI 93740-01 Reagents for 50 tests HI 93740-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Nickel LR
method using the procedure described in
the
Method Selection
section (see page 12).
Note: for best results perform your tests between 20-24°C.
• Press TIMER and the display will show a countdown prior to adding reagent C or, alternatively, wait for 15 minutes. Add one packet of HI 93740C-0 EDTA reagent to each beaker, cap and swirl to mix until completely dissolved.
• Fill one cuvette up to the mark with 10 mL of the blank.
10 mL
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
10 mL
• Fill a second cuvette up to the mark with 10 mL of the reacted sample.
• Fill one graduated beaker with 25 mL of deionized water (blank) and another one with 25 mL of sample.
• Add the content of one packet of HI 93740A-0 Phthalate­phosphate reagent to each beaker. Cap and swirl gently until the reagent is dissolved.
Note: If sample contains iron (Fe
3+
), it is important that all powder be dissolved completely before continuing with follow­ing step.
• Add 1 mL of HI 93740B-0 0.3% PAN solution to each beaker, cap and swirl to mix.
Nickel LR
86 87
• Insert the second cuvette into the instrument.
• Press READ to start the reading. The instrument dis­plays the results in mg/L of nickel.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings, add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity is removed.
Nickel LR
INTERFERENCES
Interference may be caused by:
2+
must not be present
Co
2+
must not be present
Fe
3+
above 32 mg/L
Al
2+
above 1000 mg/L (as CaCO3)
Ca
2+
above 20 mg/L
Cd
above 8000 mg/L
Cl
3+
above 20 mg/L
Cr
6+
Cr
above 40 mg/L
2+
above 15 mg/L
Cu
above 20 mg/L
F
3+
above 10 mg/L
Fe
+
above 500 mg/L
K
2+
above 400 mg/L
Mg
2+
above 25 mg/L
Mn
6+
above 60 mg/L
Mo
+
Na
above 5000 mg/L
2+
above 20 mg/L
Pb
2+
above 30 mg/L
Zn
NITRATE
SPECIFICATIONS
Range 0.0 to 30.0 mg/L Resolution 0.1 mg/L Accuracy ±0.5 mg/L ±10% of reading Typical EMC ±0.1 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the cadmium reduction method. The reaction between nitrate and the
reagent causes an amber tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93728-0 Powder reagent 1 packet
REAGENT SETS
HI 93728-01 Reagents for 100 tests HI 93728-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the the
• Using the pipette, fill the cuvette with 6 ml of sample, up to half of its height, and replace the cap.
Nitrate
method using the procedure described in
Method Selection
section (see page 12).
6 mL
Nickel LR
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93728-0 reagent.
88 89
Nitrate
• Replace the cap and immediately shake vigorously up and down for exactly 10 seconds. Continue to mix by inverting the cuvette gently for 50 seconds, while taking care not to induce air bubbles. Powder will not com­pletely dissolve. Time and way of shaking could sensitively affect the measurement.
• Reinsert the cuvette into the instrument, taking care not to shake it.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 4 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of nitrate-nitrogen.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of nitrate (NO
NITRITE HIGH RANGE
SPECIFICATIONS
Range 0 to 150 mg/L Resolution 1 mg/L Accuracy ±4 mg/L ±4% of reading Typical EMC ±1 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the Ferrous Sulfate method. The reaction between nitrite and the
reagent causes a greenish-brown tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93708-0 Powder reagent 1 packet
REAGENT SETS
HI 93708-01 Reagents for 100 tests HI 93708-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette up to the mark with 10 mL of unreacted
-
).
3
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the
Nitrite HR
method using the procedure described
Method Selection
in the
section (see page 12).
sample and replace the cap.
meter is zeroed and ready for measurement.
10 mL
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Ammonia and amines, as urea and primary aliphatic amines Chloride above 100 ppm Chlorine above 2 ppm Copper Iron(III) Strong oxidizing and reducing substances Sulfide must be absent
Nitrate
90 91
• Remove the cuvette.
• Add the content of one packet of HI 93708-0 reagent. Replace the cap and shake gently until completely dissolved.
Nitrite HR
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 10 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of nitrite.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of nitrogen-nitrite (NO nitrite (NaNO
).
2
-N) and sodium
2
NITRITE LOW RANGE
SPECIFICATIONS
Range 0.00 to 0.35 mg/L Resolution 0.01 mg/L Accuracy ±0.02 mg/L ±4% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
EPA Diazotization method 354.1
the reagent causes a pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93707-0 Powder reagent 1 packet
REAGENT SETS
HI 93707-01 Reagents for 100 tests HI 93707-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette up to the mark with 10 mL of unreacted
Nitrite LR
method using the procedure described
Method Selection
in the
section (see page 12).
sample (up to the mark) and replace the cap.
. The reaction between nitrite and
10 mL
• Press the  or  to go back to the measurement screen.
Nitrite HR
92 93
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add the content of one packet of HI 93707-0 reagent. Replace the cap and shake gently for about 15 seconds.
Nitrite LR
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of nitrite.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of nitrogen-nitrite (NO nitrite (NaNO
).
2
-N) and sodium
2
DISSOLVED OXYGEN
SPECIFICATIONS
Range 0.0 to 10.0 mg/L Resolution 0.1 mg/L Accuracy ±0.4 mg/L ±3% of reading Typical EMC ± 0.1 mg/L Deviation Light Source Tungsten lamp with narrow band interference filter @ 420 nm Method Adaptation of the
th
18
edition,
and the reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93732A-0 Reagent A 5 drops HI 93732B-0 Reagent B 5 drops HI 93732C-0 Reagent C 10 drops
REAGENT SET
HI 93732-01 Reagents for 100 tests HI 93732-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Dissolved Oxygen
Method Selection
section (see page 12).
Standard Methods for the Examination of Water and Wastewater,
Azide modified Winkler method. The reaction between dissolved oxygen
method using the procedure described in the
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by the following ions: ferrous, ferric, cupric, mercurous, silver, antimonious, bismuth, auric, lead, metavanadate and chloroplatinate. Strongly reducing and oxidizing reagents. High levels of nitrate (above 100 mg/L) could yield falsely high readings due to a minute amount of
reduction to nitrite that could occur at these levels.
Nitrite LR
94 95
• Fill one 60 mL glass bottle completely with the unreacted sample.
• Replace the cap and ensure that a small part of the sample spills over.
• Remove the cap and add 5 drops of HI 93732A-0 and 5 drops of HI 93732B-0.
• Add more sample, to fill the bottle completely. Replace the cap again and ensure that a part of the sample spills over. This is to make sure that no air bubbles have been trapped inside, which could alter the reading.
• Invert several times the bottle. The sample becomes orange-yellow and a flocculent agent will appear.
x 5
Dissolved Oxygen
x 5
• Let the sample stand and the flocculent agent will start to settle.
• After approximately 2 minutes, when the upper half of the bottle becomes
x 10
limpid, add 10 drops of HI 93732C-0.
• Replace the cap and invert the bottle until the settled flocculent dissolves completely. The sample is ready for measurement when it is yellow and
completely limpid.
10 mL
• Fill the cuvette up to the mark with 10 mL of the unreacted (original) sample, and replace the cap. This is the blank.
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Fill another cuvette up to the mark with 10 mL of the reacted sample and replace the cap.
• Reinsert the cuvette into the instrument.
• Press READ to start the reading. The instrument will display the results in mg/L of dissolved oxygen.
OZONE
SPECIFICATIONS
Range 0.00 to 2.00 mg/L Resolution 0.01 mg/L Precision ±0.03 mg/L @ 1.00 mg/L Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Colorimetric DPD Method. The reaction between ozone and the DPD reagent causes a
REQUIRED REAGENTS
Code Description Quantity/test
HI 93757-0 DPD Powder Reagent 1 packet HI 93703-52-0 Glycine Powder (Optional Reagent) 1 packet
REAGENT SETS
HI 93757-01 Reagents for 100 tests HI 93757-03 Reagents for 300 tests HI 93703-52 Glycine Powder, Optional Reagent for 100 tests
For other accessories see page 128.
IMPORTANT NOTE: Chlorine is a strong interferent for ozone determination. If the sample is suspected to contain chlorine residues (free or total chlorine), please follow the alternative measurement procedure described below:
• Perform the Standard Measurement Procedure and take note of the reading:
• On a fresh sample perform the Additional Measurement Procedure and take note of the reading:
• Subtract reading mg/L (O
STANDARD MEASUREMENT PROCEDURE
• Select the in the
• Fill the cuvette with 10 mL of unreacted sample, up to the mark, and replace the cap.
pink tint in the sample.
B
from reading A to obtain the ozone concentration in mg/L:
) =
value A
3
Ozone
Method Selection
value B
method using the procedure described
section (see page 12).
value A
.
value B
.
10 mL
.
INTERFERENCES
Interferences may be caused by reducing and oxidizing materials.
Dissolved Oxygen
96 97
• Place the cuvette into the holder and close the lid.
Ozone
• Press ZERO key. The display will show “-0.0-” the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add the content of one packet of HI 93757-0 Ozone Reagent. Replace the cap and shake gently for 20 seconds.
• Replace the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add the content of one packet of the optional reagent HI 93703-52-0 Glycine Powder. Replace the cap and shake gently until completely dissolved.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ. When the timer ends the meter will perform the reading.
• The instrument displays concentration in mg/L of ozone (chlorine free samples only).
ADDITIONAL MEASUREMENT PROCEDURE
For samples containing chlorine
Ozone
• Select the the
Method Selection
• Fill the cuvette with 10 mL of unreacted sample, up to the mark, and replace the cap.
• Place the cuvette into the holder and close the lid.
method using the procedure described in
section (see page 12).
10 mL
• Add the content of one packet of HI 93757-0 Ozone Reagent. Replace the cap and shake gently for 20 seconds.
• Replace the cuvette into the holder and close the lid.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ. When the timer ends the meter will perform the reading.
• The instrument displays a concentration value refering to chlorine interference. Subtract this value from the reading from the Standard Measurement Procedure: this will be the concentration in mg/L of
ozone in the sample.
INTERFERENCES
Interference may be caused by: Bromine, Chlorine Dioxide, Iodine. Alkalinity above 250 mg/L CaCO To resolve this, neutralize the sample with diluted HCl. In case of water with hardness greater than 500 mg/L CaCO
minutes after adding the powder reagent.
will not reliably develop the full amount of color or it may rapidly fade.
3
, shake the sample for approximately 2
3
98 99
OzoneOzone
pH
SPECIFICATIONS
Range 6.5 to 8.5 pH Resolution 0.1 pH Accuracy ±0.1 pH Typical EMC ±0.1 pH
Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow
REQUIRED REAGENTS
Code Description Quantity HI 93710-0 Phenol Red Indicator 5 drops
REAGENT SETS
HI 93710-01 Reagents for 100 pH tests HI 93710-03 Reagents for 300 pH tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the pH method using the procedure described in the
Method Selection
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
to red tint in the sample.
10 mL
section (see page 12).
• Reinsert the cuvette into the instrument.
• Press the READ key to start the reading. The instrument displays the pH value.
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add 5 drops of HI 93710-0 Phenol Red Indicator. Replace the cap and mix the solution.
pH
100 101
x 5
pH
PHOSPHATE HIGH RANGE
SPECIFICATIONS
Range 0.0 to 30.0 mg/L Resolution 0.1 mg/L Accuracy ±1 mg/L ±4% of reading Typical EMC Dev. ±0.1 mg/L Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
th
18
edition,
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93717A-0 Molybdate 10 drops HI 93717B-0 Reagent B 1 packet
REAGENT SETS
HI 93717-01 Reagents for 100 tests HI 93717-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the
in the
Method Selection
Phosphate HR
method using the procedure described
section (see page 12).
mark) and replace the cap.
Standard Methods for the Examination of Water and Wastewater,
Amino Acid method. The reaction between phosphate and reagents
10 mL
• Add the content of one packet of HI 93717B-0 Phosphate HR Reagent B to the cuvette. Replace the cap and shake gently until completely dissolved.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 5 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of phosphate (PO
3-
).
4
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of phosphorus (P) and phosphorus pentoxide (P
2O5
).
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add 10 drops of HI 93717A-0 Molybdate reagent.
Phosphate HR
102 103
x 10
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Sulfide Chloride above 150000 mg/L ) Calcium above 10000 mg/L as CaCO Magnesium above 40000 mg/L as CaCO Ferrous iron above 100 mg/L
3
3
Phosphate HR
PHOSPHATE LOW RANGE
SPECIFICATIONS
Range 0.00 to 2.50 mg/L Resolution 0.01 mg/L Accuracy ±0.04 mg/L ±4% of reading Typical EMC Dev. ±0.01 mg/L Light Source Tungsten lamp with narrow band interference filter @ 610 nm Method Adaptation of the Ascorbic Acid method. The reaction between phosphate and the
reagent causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93713-0 Powder reagent 1 packet
REAGENT SETS
HI 93713-01 Reagents for 100 tests HI 93713-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the scribed in the
• Rinse, cap and shake the cuvette several times with unreacted sample. Fill the cuvette with 10 mL of sample up to the mark and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93713-0 reagent. Replace the cap and shake gently (for about 2 minutes) until the powder is completely dissolved.
Phosphate LR
Method Selection
method using the procedure de-
section (see page 12).
10 mL
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of phosphate (PO
3-
).
4
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of phosphorus (P) and phosphorus pentoxide (P
2O5
).
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Iron above 50 mg/L Silica above 50 mg/L Silicate above 10 mg/L Copper above 10 mg/L Hydrogen sulfide, arsenate, turbid sample and highly buffered samples also interfere.
Phosphate LR
104 105
Phosphate LR
PHOSPHORUS
SPECIFICATIONS
Range 0.0 to 15.0 mg/L Resolution 0.1 mg/L Accuracy ±0.3 mg/L ±4% of reading Typical EMC Dev. ±0.2 mg/L Light Source Tungsten lamp with narrow band interference filter @ 525 nm Method Adaptation of the
th
18
edition,
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93706A-0 Molybdate 10 drops HI 93706B-0 Amino Acid Powder 1 packet
REAGENT SETS
HI 93706-01 Reagents for 100 tests HI 93706-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the scribed in the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
Phosphorus
method using the procedure de-
Method Selection
Standard Methods for the Examination of Water and Wastewater,
Amino Acid method. The reaction between phosphate and reagents
section (see page 12).
10 mL
• Add the content of one packet of HI 93706B-0 Phosphorus Reagent B (Amino Acid) to the cuvette. Replace the cap and shake gently until completely dissolved.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 5 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L of phosphorus (P).
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of phosphate (PO pentoxide (P
2O5
).
3-
) and phosphorus
4
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add 10 drops of HI 937O6A-0 Molybdate reagent.
Phosphorus
106 107
x 10
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interference may be caused by: Sulfide Chloride above 150000 mg/L Calcium above 10000 mg/L as CaCO Magnesium above 40000 mg/L as CaCO Ferrous iron above 100 mg/L
3
3
Phosphorus
POTASSIUM HIGH RANGE
SPECIFICATIONS
Range 20 to 200 mg/L Resolution 5 mg/L Accuracy ±30 mg/L ±7% of reading Typical EMC ±5 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 610 nm Method Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93750A-0 Potassium Reagent 6 drops HI 93750B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests HI 93750-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENT SOLUTIONS (HR) procedure at page 19.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93750B-0 reagent. Replace the cap and
gently mix for
one minute by slowly turning the cuvette upside down.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L (ppm) of potas­sium (K).
• Select the described in the
Potassium HR
method using the procedure
Method Selection
section (see page 12).
• Fill the cuvette with 10 mL of sample, up to the mark.
• Add six drops of HI 93750A-0, replace the cap and swirl the solution.
• Place the cuvette into the holder and close the lid.
Potassium HR
108 109
x 6
10 mL
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of potassium oxide (K
• Press the  or  to go back to the measurement screen.
• For ULTRA HIGH RANGE samples: follow the procedure described at page 110.
O)
2
Potassium HR
INTERFERENCES
Interferences may be caused by: Ammonium above 10 ppm Calcium above 10000 ppm as CaCO
3
Chloride above 12000 ppm Magnesium above 8000 ppm as CaCO
3
Sodium above 8000 ppm
POTASSIUM ULTRA HIGH RANGE
For samples containing more than 200 ppm of Potassium: follow the sample preparation procedure described at page 19 for NUTRIENTS SOLUTION (HR). Then add to the graduated cylinder 20 mL of the prepared sample (for HR) and fill the cylinder with demineralized water from the Demineralizer Bottle up to the 100 mL mark.
Follow the MEASUREMENT PROCEDURE at page 108. Read the result in mg/L of Potassium on the display and multiply the reading by 5 to obtain the actual
concentration of Potassium.
POTASSIUM MEDIUM RANGE
SPECIFICATIONS
Range 10 to 100 mg/L Resolution 2.5 mg/L Accuracy ±15 mg/L ±7% of reading Typical EMC ±2.5 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 610 nm Method Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93750A-0 Potassium Reagent 6 drops HI 93750B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests HI 93750-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENTS SOLUTION (MR) procedure at page 18.
Potassium HR
• Select the described in the
Potassium MR
method using the procedure
Method Selection
section (see page 12).
• Fill the cuvette with 10 mL of sample, up to the mark.
• Add six drops of HI 93750A-0, replace the cap and swirl the solution.
• Place the cuvette into the holder and close the lid.
110 111
10 mL
x 6
Potassium MR
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93750B-0 reagent. Replace the cap and
gently mix for
one minute by slowly turning the cuvette upside down.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L (ppm) of potassium (K).
POTASSIUM LOW RANGE
SPECIFICATIONS
Range 0.0 to 20.0 mg/L Resolution 0.5 mg/L Accuracy ±3.0 mg/L ±7% of reading Typical EMC ±0.5 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 610 nm Method Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93750A-0 Potassium Reagent 6 drops HI 93750B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests HI 93750-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure at page 17.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of potassium oxide (K
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interferences may be caused by: Ammonium above 10 ppm Calcium above 10000 ppm as CaCO Chloride above 12000 ppm Magnesium above 8000 ppm as CaCO Sodium above 8000 ppm
Potassium MR
3
3
112 113
• Select the
Potassium LR
described in the
O).
2
• Fill the cuvette with 10 mL of sample, up to the mark.
• Add 6 drops of HI 93750A-0 Potassium Reagent, re-
method using the procedure
Method Selection
section (see page 12).
10 mL
x 6
place the cap and swirl the solution.
• Place the cuvette into the holder and close the lid.
Potassium LR
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93750B-0 reagent. Replace the cap and
gently mix for
one minute by slowly turning the cuvette upside down.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays the results in mg/L (ppm) of potassium (K).
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of potassium oxide (K
O).
2
SILICA
SPECIFICATIONS
Range 0.00 to 2.00 mg/L Resolution 0.01 mg/L Accuracy ±0.03 mg/L ±3% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 610 nm Method Adaptation of the
Heteropoly Blue method. The reaction between silica and reagents causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity HI 93705A-0 Molybdate 6 drops HI 93705B-0 Citric acid 1 packet HI 93705C-0 Amino acid 1 packet
REAGENT SETS
HI 93705-01 Reagents for 100 tests HI 93705-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Method Selection
• Fill the cuvette with 10 mL of unreacted sample (up to the mark).
• Add 6 drops of HI 93705A-0 Molybdate reagent. Replace the cap and swirl the solution.
Silica
method using the procedure described in the
section (see page 12).
ASTM Manual of Water and Environmental Technology, D859,
x 6
10 mL
• Press the  or  to go back to the measurement screen.
INTERFERENCES
Interferences may be caused by: Ammonium above 10 ppm Calcium above 10000 ppm as CaCO Chloride above 12000 ppm Magnesium above 8000 ppm as CaCO Sodium above 8000 ppm
Potassium LR
3
3
114 115
• Wait for 4 minutes, add the content of one packet of HI 93705B-0 Citric acid reagent and shake until it is completely dissolved.
• Wait for 1 minute. This is the blank.
1’
12
9
3
6
Silica
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of HI 93705C-0 Amino acid reagent and shake until it is completely dissolved.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement. Alternatively, wait for exactly 3 minutes and press READ. When the timer ends the meter will perform the reading. The instrument displays concentration in mg/L of silica (SiO
).
2
INTERFERENCES
Interference may be caused by: Phosphate above 60 mg/L Phosphate above 75 mg/L Sulfide and high concentration of iron Eliminate color and turbidity interferences by zeroing the meter with the original water sample.
• Press the  or  to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of Silicon (Si).
• Press the  or  to go back to the measurement screen.
Silica
116 117
Silica
SILVER
SPECIFICATIONS
Range 0.000 to 1.000 mg/L Resolution 0.001 mg/L Accuracy ±0.005 mg/L ±10% of reading Typical EMC ± 0.001 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm. Method Adaptation of the PAN method. The reaction between silver and reagents causes an
orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93737A-0 Buffer Reagent A 1 mL HI 93737B-0 Buffer Reagent B 1 mL HI 93737C-0 Indicator Reagent C 2 mL HI 93737D-0 Fixing Reagent D 2 mL HI 93703-51 Dispersing Agent 4-6 drops
(only when necessary, see note)
REAGENT SETS
HI 93737-01 Reagents for 50 tests HI 93737-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill two graduated beakers with 25 mL of sample.
Silver
method using the procedure described in
Method Selection
the
section (see page 12).
Note: for best results perform your tests between 20-24°C.
# 1
# 2
• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker (the sample) and swirl gently to mix.
# 2
• Press TIMER and the display will show the countdown prior to adding reagent C or, alternatively, wait for 2 minutes. Then add exactly 1.0 mL of HI 93737C-0 Indicator reagent to each beaker and swirl.
# 2
# 1
• Press TIMER or, alternatively, wait for 2 minutes. Then, in both cases, add 1.0 mL of HI 93737D-0 Fixing reagent to each beaker and swirl.
# 2
# 1
• Then press TIMER or, alternatively, wait for 2 minutes to allow reaction to complete.
• Add 1.0 mL of HI 93737A-0 Buffer reagent to one beaker (the blank) and swirl gently to mix.
Silver
118 119
# 1
10 mL
• Fill one cuvette up to the mark with 10 mL of the blank.
• Place the cuvette into the holder and close the lid.
10 mL
# 1
# 1
Silver
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
10 mL
• Fill a second cuvette up to the mark with 10 mL of the reacted sample.
# 2
SULFATE
SPECIFICATIONS
Range 0 to 150 mg/L Resolution 5 mg/L Accuracy ±5 mg/L ±3% of reading Light Source Tungsten lamp with narrow band interference filter @ 466 nm Method Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension
is measured.
• Insert the second cuvette into the instrument.
# 2
• Press READ to start the reading. The instrument displays the results in mg/L of silver.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings, add 2-3 drops of Dispersing agent (HI 93703-51) to each cuvette and swirl until turbidity is completely dissolved.
INTERFERENCES
Interference may be caused by:
3+
Al
above 30 mg/L Fe2+ above 1.5 mg/L
2+
above 1000 mg/L as CaCO3Fe3+ above 10 mg/L
Ca
2+
Cd
above 20 mg/L K+ above 500 mg/L
-
Cl
above 8000 mg/L Mn2+ above 25 mg/L
2+
above 1.5 mg/L Mg2+ above 1000 mg/L as CaCO
Co Cr3+ above 20 mg/L Na+ above 5000 mg/L
6+
Cr
above 40 mg/L Ni2+ above 1.5 mg/L
2+
above 15 mg/L Pb2+ above 20 mg/L
Cu
-
F
above 20 mg/L Zn2+ above 30 mg/L
3
REQUIRED REAGENTS
Code Description Quantity HI 93751-0 Indicator reagent 1 packet
REAGENT SETS
HI 93751-01 Reagents for 100 tests HI 93751-03 Reagents for 300 tests
For other accessories see page 17.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure on page 17.
Sulfate
• Select the in the
• Fill a cuvette with 10 mL of unreacted sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
method using the procedure described
Method Selection
section (see page 12).
10 mL
Silver
• Remove the cuvette.
120 121
Sulfate
ZINC
• Add the content of one packet of HI 93751-0 Indicator reagent.
12
9
6
• Replace the cap and shake gently for 1 minute.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 5 minutes and press READ. When the timer ends the meter will perform the reading.
• The instrument displays concentration in mg/L of Sulfate (SO
2-
).
4
INTERFERENCES
Interferences may be caused by: Calcium (as CaCO
Chloride (as Cl Magnesium (as MgCO Silica (as SiO
Color or suspended matter in large amounts will interfere: suspended matter should be removed by previous filtration.
Organic matter in large amounts may impede the precipitation of barium sulfate.
) above 20000 mg/L
3
-
) above 40000 mg/L
) above 10000 mg/L
3
) above 500 mg/L
2
1’
3
SPECIFICATIONS
Range 0.00 to 3.00 mg/L Resolution 0.01 mg/L Accuracy ±0.03 mg/L ±3% of reading Typical EMC ±0.01 mg/L
Deviation Light Source Tungsten lamp with narrow band interference filter @ 575 nm Method Adaptation of the
18
th
edition,
Standard Methods for the Examination of Water and Wastewater,
Zincon method. The reaction between zinc and the reagents causes an
orange to a dark violet tint in the sample.
REQUIRED REAGENT
Code Description Quantity HI 93731A-0 Zinc Reagent 1 packet HI 93731B-0 Cyclohexanone 0.5 mL
REAGENT SETS
HI 93731-01 Reagents for 100 tests HI 93731-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
• Fill one graduated mixing cylinder up to the 20 mL
• Add the content of one packet of HI 93731A-0 Zinc
Zinc
method using the procedure described in
Method Selection
the
section (see page 12).
mark with the sample.
reagent, close the cylinder and invert several times to mix until completely dissolved.
20 mL
Sulfate
• Fill one cuvette with 10 mL of the reacted sample up to the mark.
122 123
10 mL
Zinc
• Place the cap and insert the cuvette into the instrument and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
ERRORS AND WARNINGS
The instrument shows clear warning messages when erroneous conditions appear and when measured values are outside the expected range. These messages are described below.
No Light: The light source is not functioning properly.
• Remove the cuvette and add 0.5 mL of HI 93731B-0 Cyclohexanone to the cuvette.
Note: To prevent any contamination from the polycarbon­ate cap, prior to replacing it, close the sample cuvette with the supplied HDPE plastic stopper.
• Replace the cap and mix the sample for 15 seconds.
• Insert the sample into the instrument.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of zinc.
INTERFERENCES
Interference may be caused by: Aluminum above 6 mg/L Cadmium above 0.5 mg/L Copper above 5 mg/L Iron above 7 mg/L Manganese above 5 mg/L Nickel above 5 mg/L
Zinc
124 125
Light Leak: There is an excess amount of ambient light reaching the detector.
Inverted cuvettes: The sample and the zero cuvettes are inverted.
Battery Low: The battery capacity is lower than 10%.
Light Low: The instrument cannot adjust the light level. Please check
that the sample does not contain any debris.
Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette.
DATA MANAGEMENT
STANDARD METHODS
The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software.
Description Range Method
Aluminum 0.00 to 1.00 mg/L Aluminon Alkalinity 0 to 500 mg/L Colorimetric Ammonia MR 0.00 to 10.00 mg/ L Nessler Ammonia LR 0.00 to 3.00 mg / L Nessler Bromine 0.00 to 8.00 mg/L DPD Calcium 0 to 400 mg/L Oxalate Chlorine, Free 0.00 to 2.50 mg / L DPD Chlorine, Total 0.00 to 3.50 mg / L DPD Chlorine Dioxide 0.00 to 2.00 mg/L Chlorophenol Red Chromium VI HR 0 to 1000 μg/L Diphenylcarbohydrazide Chromium VI LR 0 to 300 μg/L Diphenylcarbohydrazide Color of Water 0 to 500 PCU Colorimetric Platinum Cobalt Copper HR 0.00 to 5.00 mg/L Bicinchoninate Copper LR 0 to 1000 μg/L Bicinchoninate Cyanide 0.000 to 0.200 mg/L Pyridine-Pyrazolone Cyanuric Acid 0 to 80 mg/L Turbidimetric Fluoride 0.00 to 2.00 mg/L SPADNS Calcium Hardness 0.00 to 2.70 mg/L Colorimetric Magnesium Hardness 0.00 to 2.00 mg/L Colorimetric Hydrazine 0 to 400 μg /L p-Dimethylaminobenzaldehyde Iodine 0.0 to 12.5 mg/L DPD Iron HR 0.00 to 5.00 mg/L Phenantroline Iron LR 0 to 400 μg/L TPTZ Magnesium 0 to 150 mg/L Calmagite Manganese HR 0.0 to 20.0 mg/L Periodate Oxidation Manganese LR 0 to 300 μg/L PAN Molybdenum 0.0 to 40.0 mg/L Mercaptoacetic Acid Nickel HR 0.00 to 7.00 g/L Photometric Nickel LR 0.000 to 1.000 mg/L PAN Nitrate 0.0 to 30.0 mg / L Cadmium Reduction Nitrite HR 0 to 150 mg/L Ferrous Sulfate Nitrite LR 0.00 to 0.35 mg/L Diazotization Oxygen, Dissolved 0.0 to 10.0 mg/L Winkler Ozone 0.00 to 2.00 mg/L DPD pH 6.5 to 8.5 pH Phenol Red Phosphate HR 0.0 to 30.0 mg/L Amino Acid Phosphate LR 0.00 to 2.50 mg/L Ascorbic Acid Phosphorus 0.0 to 15.0 mg / L Amino Acid Potassium HR 20 to 200 mg/L Turbidimetric Potassium MR 10 to 100 mg/L Turbidimetric Potassium LR 0.0 to 20.0 mg/L Turbidimetric Silica 0.00 to 2.00 mg/L Heteropoly Blue Silver 0.000 to 1.000 mg/L PAN Sulfate 0 to 150 mg/L Turbidimetric Zinc 0.00 to 3.00 mg/L Zincon
126 127
ACCESSORIES
REAGENT SETS HI 93700-01 100 ammonia LR tests HI 93700-03 300 ammonia LR tests HI 93701-01 100 free chlorine tests (powder) HI 93701-03 300 free chlorine tests (powder) HI 93701-F 300 free chlorine tests (liquid) HI 93701-T 300 total chlorine tests (liquid) HI 93702-01 100 copper HR tests HI 93702-03 300 copper HR tests HI 93704-01 100 hydrazine tests HI 93704-03 300 hydrazine tests HI 93705-01 100 silica tests HI 93705-03 300 silica tests HI 93706-01 100 phosphorus tests HI 93706-03 300 phosphorus tests HI 93707-01 100 nitrite LR tests HI 93707-03 300 nitrite LR tests HI 93708-01 100 nitrite HR tests HI 93708-03 300 nitrite HR tests HI 93709-01 100 manganese HR tests HI 93709-03 300 manganese HR tests HI 93710-01 100 pH tests HI 93710-03 300 pH tests HI 93711-01 100 total chlorine tests (powder) HI 93711-03 300 total chlorine tests (powder) HI 93712-01 100 aluminum tests HI 93712-03 300 aluminum tests HI 93713-01 100 phosphate LR tests HI 93713-03 300 phosphate LR tests HI 93714-01 100 cyanide tests HI 93714-03 300 cyanide tests HI 93715-01 100 ammonia MR tests HI 93715-03 300 ammonia MR tests HI 93716-01 100 bromine tests HI 93716-03 300 bromine tests HI 93717-01 100 phosphate HR tests HI 93717-03 300 phosphate HR tests HI 93718-01 100 iodine tests HI 93718-03 300 iodine tests HI 93719-01 100 Mg hardness tests HI 93719-03 300 Mg hardness tests HI 93720-01 100 Ca hardness tests HI 93720-03 300 Ca hardness tests
HI 93721-01 100 iron HR tests HI 93721-03 300 iron HR tests HI 93722-01 100 cyanuric acid tests HI 93722-03 300 cyanuric acid tests HI 93723-01 100 chromium VI HR tests HI 93723-03 300 chromium VI HR tests HI 93726-01 100 nickel HR tests HI 93726-03 300 nickel HR tests HI 93728-01 100 nitrate tests HI 93728-03 300 nitrate tests HI 93729-01 100 fluoride tests HI 93729-03 300 fluoride tests HI 93730-01 100 molybdenum tests HI 93730-03 300 molybdenum tests HI 93731-01 100 zinc tests HI 93731-03 300 zinc tests HI 93732-01 100 dissolved oxygen tests HI 93732-03 300 dissolved oxygen tests HI 93737-01 50 silver tests HI 93737-03 150 silver tests HI 93738-01 100 chlorine dioxide tests HI 93738-03 300 chlorine dioxide tests HI 93740-01 50 nickel LR tests HI 93740-03 150 nickel LR tests HI 93746-01 50 iron LR tests HI 93746-03 150 iron LR tests HI 93747-01 100 copper LR tests HI 93747-03 300 copper LR tests HI 93748-01 50 manganese LR tests HI 93748-03 150 manganese LR tests HI 93749-01 100 chromium VI LR tests HI 93749-03 300 chromium VI LR tests HI 93755-01 100 alkalinity tests HI 93755-03 300 alkalinity tests HI 937521-01 50 calcium tests HI 937521-03 150 calcium tests HI 937520-01 50 magnesium tests HI 937520-03 150 magnesium tests HI 93757-01 100 ozone tests HI 93757-03 300 ozone tests HI 93703-52-2 Glycine Powder, Optional Reagent for 100 tests HI 93750-01 100 potassium HR tests HI 93750-03 300 potassium HR tests HI 93751-01 100 sulfate tests HI 93751-03 300 sulfate tests
128 129
OTHER ACCESSORIES HI 731318 cloth for wiping cuvettes (4 pcs) HI 731321 glass cuvettes (4 pcs) HI 731325W new cap for cuvette (4 pcs) HI 740034 cap for 100 mL beaker (6 pcs) HI 740036 100 mL plastic beaker (6 pcs) HI 740038 60 mL glass bottle and stopper HI 740142 1 mL graduated syringe HI 740143 1 mL graduated syringe (6 pcs) HI 740144 pipette tip (6 pcs) HI 740157 plastic refilling pipette (20 pcs) HI 740220 25 mL glass cylinders with caps (2 pcs) HI 740223 170 mL plastic beaker HI 740224 170 mL plastic beakers (12 pcs) HI 740225 60 mL graduated syringe HI 740226 5 mL graduated syringe HI 740227 filter assembly HI 740228 filter discs (25 pcs) HI 740229 100 mL graduated cylinder HI 740230 230 mL demineralized water HI 92000 Windows compatible software HI 920013 PC connection cable HI 93703-50 cuvette cleaning solution (230 mL) HI 93703-54 dried resin (100 g) HI 93703-55 activated carbon (50 pcs)
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions.
This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact your dealer. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipments, this requiring the operator to take all necessary steps to
correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid damages or burns, do not put the instrument in microwave ovens. For yours and the instrument safety do not use or store the instrument in hazardous
environments.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
HANNA LITERATURE
Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as:
Water Treatment
Process
Swimming Pools
Agriculture
Food
Laboratory
and many others. New reference material is constantly being added to the library. For these and other catalogs, handbooks and leaflets contact your dealer or the Hanna Customer Service Center nearest to you. To find the Hanna Office in your vicinity, check our home page at www.hannainst.com.
130 131
SALES AND TECHNICAL SERVICE CONTACTS
Tel. (03) 9769.0666 • Fax (03) 9769.0699
e-mail: hannains@hannainst.com.au
China:
Australia:
Tel. (10) 88570068 • Fax (10) 88570060
e-mail: hannachina@vip.sina.com
Egypt:
Tel. & Fax (02) 2758.683
e-mail: hannaegypt@go.com.eg
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
e-mail: hannager@aol.com
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
e-mail: hannagr@otenet.gr
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
e-mail: transit@dnet.net.id
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
e-mail: sales@hanna.co.jp
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
e-mail: mccoyhan@chollian.net
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
e-mail: hannamal@tm.net.my
Norway:
Tel. (23) 3811.00 • Fax (23) 3811.01
e-mail: hanna@hannainst.no
Singapore:
Tel. 6296.7118 • Fax 6291.6906
e-mail: hannaap@pacific.net.sg
Tel. (011) 615.6076 • Fax (011) 615.8582
South Africa:
e-mail: hannasa@mweb.co.za
MAN83200 05/08
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
e-mail: salesteam@hannainst.co.uk
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
e-mail: sales@hannainst.com
For additional Technical Support in your local language, see www.hannainst.com
132
Loading...