Hanna Instruments HI 8064 User Manual

Instruction Manual
HI 8064
Portable Thermohygrometer
1
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the neces­sary information for the correct use of the instrument. If you need additional technical support, do not hesitate to e-mail us at tech@hannainst.com This instrument is in compliance with the CE directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing ma­terial and examine it carefully to make sure that no damage has occurred during ship­ment. If there is any damage, immediately notify your Dealer.
sure that the instrument functions cor­rectly. All defective items must be re­turned in their original packing with the supplied accessories.
TABLE OF CONTENTS
Preliminary Examination ............................ 3
General Description.................................... 3
Specifications ............................................ 4
Functional Description................................ 4
Relative Humidity Calibration ...................... 5
Temperature Calibration ............................. 7
Taking measurements................................ 7
Relative Humidity Calibration Table............. 8
Warranty .................................................. 10
Hanna Products ....................................... 11
GENERAL DESCRIPTION
With a simple rotary switch measuring both relative humidity and temperature, HI 8064 is an easy-to-operate thermohygrometer specially dedicated to education. Just turn the dial from off to RH% for relative humidity or °C for temperature, and the large LCD will show the appropriate measured value in seconds. The probe handle contains the amplification circuitry to ensure an accurate and linear read­ing throughout the range. The probe tip is perforated to allow air circulation and thus guarantee a fast response time. To measure humidity electronically, a thin film polymer ca­pacitance is used. The probe handle is made of tough ABS material to withstand the wear and tear of everyday use. Weighing less than a pound with battery life of one hundred hours and a response time of less than half a minute, HI 8064 is ideal for use in schools, libraries, museums and horticulture.
2
3
SPECIFICATIONS
RH CALIBRATION
Range 10.0 to 95.0% RH
0.0 to 60.0°C
Resolution 0.1% RH / 0.1°C Accuracy ±2% RH / ±0.4°C Response Time 6 seconds for 95% accuracy RH Calibration Manual, 2 point, through
trimmers on the RH probe
Environment 0 to 50°C (32 to 122°F);
RH max 98% non-condensing
Battery Type 1 x 9V, alkaline Battery Life Approx. 100 hours of use Dimensions 180 x 83 x 40 mm
(7.1 x 3.3 x 1.6'')
Weight 180 g (6.3 oz.)
FUNCTIONAL DESCRIPTION
• OFF position: the instrument is OFF; any other knob position turn the instrument ON.
• ºC position: the meter is in temperature measurement mode.
• RH% position: the meter is operating in Relative Humidity measurement mode.
• Replacing battery: when the battery level is low and only a few hours of working life are left, the "V" symbol will appear on the display to warn the user that the battery needs to be replaced. Remove the battery compartment rear cover on the rear of the instrument and replace the battery with a new one, while paying attention to the correct polarity.
The HI 8064 thermohygrometer is factory cali- brated, but it is recommended to periodically control the calibration status of the meter.
Use the HI 7601 calibration kit composed of a container with two separate chambers, each one equipped with a threaded plug and three envelops containing the necessary salts.
Preparing the calibration solutions
• Pour approximately 26 cc of distilled water into a glass container.
• Immerse the container in a bath of ice and water, and shake briefly. Then add slowly the contents of an envelope containing LiCl to the container while continuing to shake.
When the salt has completely dissolved, add the contents of the second envelope.
• Allow the solution to cool, and then pour it into the chamber marked "RH 11.1%", making sure that no residue remains on the walls of the glass container.
Note: When not in use, carefully seal the
chamber as the LiCl solution is ex­tremely hygroscopic and tends to cap­ture the air humidity causing the solution to expand slowly until it overflows from the container.
• Pour approximately 26 cc of distilled water into the other chamber (marked "RH 75.4") and add the contents of the envelope contain­ing NaCl, while shaking the container to avoid any lump formation.
Note: When not in use, carefully seal the
container.
4
5
Calibration procedure
• Bring the calibration kit to a temperature of approximately 20 ºC.
• Remove the plug from the chamber contain­ing the LiCl solution and insert the probe.
• Remove the adhesive sticker which covers the calibration trimmer access holes.
• Wait for the reading to stabilize (this will take a few hours) and then turn the "Low humidity trimmer" (the one nearest to the probe con­nection cable) until the value of 11.1% RH will be displayed on the LCD.
• Remove the probe and tightly seal the cham­ber containing the LiCl solution.
• Remove the plug from the second chamber and insert the probe.
• Wait for the reading to stabilize and then note the displayed value.
• By using the Calibration Table (pages 8-9), find the displayed readout value on column "A" and the corresponding RH value of col­umn "B", to which the meter has to be brought by acting on the "High humidity trimmer".
• Adjust the "High humidity trimmer" (the one opposed to the connection cable) until the required value is displayed on the LCD.
• Finally adjust the "Low humidity trimmer" until the value of 75.4% RH will be displayed on the LCD.
TEMPERATURE CALIBRATION
The HI 8064 thermohygrometer is factory cali- brated and therefore usually it requires no fur­ther intervention by the user.
If recalibration is desired, proceed as follows:
• Remove the adhesive sticker which covers the calibration trimmer access holes.
• Insert the probe into a recipient in which air heated at a temperature of 50 ºC has been forced to circulate.
• Insert a thermometer with an accuracy of 0.1 ºC into the same recipient.
• Read the temperature value on the thermom­eter.
• Adjust the temperature calibration trimmer (the central trimmer) until the measured tem­perature value is displayed on the LCD.
TAKING MEASUREMENTS
• For a rapid response, expose the end of the humidity detector to an air current moving at more than 0.5 m/second.
In absence of air movement, the response can be speeded up by shaking the probe.
• If condensed drops form upon the surface of the humidity sensors, turn the instrument off and wait until this undesired water has com­pletely evaporated. To speed up the evapora­tion process, expose the humidity sensors to an air current.
• If the meter is used in dusty surroundings, the filter must be kept on at all time.
6
7
RELATIVE HUMIDITY
CALIBRATION TABLE
8
9
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workman-
ship and materials when used for their in­tended purpose and maintained according to the instructions. The probes are warranted for a period of six months. Damages due to accident, misuse, tamper­ing or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charge for repair or replacement. If the instrument is to be returned to Hanna Instruments, obtain a Return Goods Authorization from the Cus­tomer Service Department first and then send it with shipment cost prepaid. When shipping any instrument, make sure it is properly pack­aged for complete protection.
HANNA PRODUCTS
• CALIBRATION & MAINTENANCE SOLUTIONS
• CHEMICAL TEST KITS
• CHLORINE METERS
• CONDUCTIVITY/TDS METERS
• DISSOLVED OXYGEN METERS
• HYGROMETERS
• ION SPECIFIC METERS
• MAGNETIC STIRRERS
• Na/NaCl METERS
• pH/ORP/Na ELECTRODES
• PROBES (DO, µS/cm, RH, T, TDS)
• PUMPS
• REAGENTS
• SOFTWARE
• THERMOMETERS
• TITRATORS
• TRANSMITTERS
• TURBIDITY METERS
• Wide Range of Accessories
Most Hanna meters are available in the following formats:
• BENCH-TOP METERS
• POCKET-SIZE METERS
• PORTABLE METERS
• PRINTING/LOGGING METERS
• PROCESS METERS (Panel- and Wall-mounted)
• METERS FOR FOOD INDUSTRY
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify
the design, construction and appearance of its
products without advance notice.
10
For additional information, contact your dealer or the nearest Hanna Customer Service Center.
You can also e-mail us at tech@hannainst.com
11
SALES & TECHNICAL SERVICE
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
12
MAN8064R1 08/05