Hanna Instruments HI 8043 User Manual

Instruction Manual
HI 8043
Portable
Dissolved Oxygen Meter
DO meter
HI 8043
STB
OFF
°C
O
2
ZERO
SLOPE
ins
rum
°C
n
1
PRELIMINARY EXAMINATION
Dear Customer, Thank you for choosing a Hanna Instruments
Product. Please read this instruction manual carefully
before using the instrument. This manual will provide you with all the neces-
sary information for the correct use of the instrument.
If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com
This instrument is in compliance with the CE directives.
TABLE OF CONTENTS
Preliminary Examination............................. 3
General Description .................................... 3
Functional Description (Probe) ................... 4
Functional Description (Meter).................... 5
Specifications ............................................. 6
Probe Initial Preparation ............................. 7
Dissolved Oxygen Calibration ..................... 9
Temperature Calibration ............................ 15
Taking Measurements .............................. 1 6
Probe & Membrane Maintenance ............. 17
Battery Replacement ................................ 19
Accessories.............................................. 20
Warranty ................................................... 21
CE Declaration of Confirmity .................... 22
Remove the instrument from the packing ma­terial and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.
Each meter is supplied complete with:
HI 76401 DO probe
Spare membrane with O-ring (2 pcs)
Protective cap
HI 7041S electrolyte solution (30 mL)
Calibration screwdriver
9V battery (1 pc)
Instructions
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective items must be returned in the original packaging with the supplied accessories.
GENERAL DESCRIPTION
HI 8043 is a portable meter for Dissolved
Oxygen measurements. It is housed in a lightweight case, with an easy-to-read LCD. Dissolved oxygen and temperature measure­ments can be performed in the field (waste­water treatment, fish-farming, water analysis, etc.) as well as in the laboratory without compromising accuracy. Dissolved oxygen is indicated in mg/L (ppm) and the reading is compensated for the tem­perature effect (ATC). The dissolved oxygen probe is provided with a membrane for covering the polarographic sensors, and features a built-in thermistor for temperature measurements and compensa­tion.
2
3
The thin permeable membrane isolates the sensor elements from the sample, but allows oxygen to enter. A voltage is applied across the sensor, and the oxygen that has passed through the mem­brane reacts causing the current to flow, hence determining a reading. A convenient feature of the instrument is its standby mode which keeps the probe polarized and ready for instant measurements.
FUNCTIONAL DESCRIPTION (METER)
1
FUNCTIONAL DESCRIPTION (PROBE)
1. D.O. Probe
2. Protective Cap
3. Watertight Shielded Cable
4. Probe Body
5. O-Ring Seal
6. Silver Chloride Anode
7. Platinum Cathode (sensor)
8. Oxygen Permeable Membrane
9. Membrane Cap
3
DO meter
STB
OFF
HI 8043
°C
O
2
42 5
9
6 7
ZERO SLOPE °C
8
s
ins
rum n
1. Liquid Crystal Display
2. "OFF" (to switch the meter off)
3. "STB" (to switch the meter on and polarize the probe. Use this standby position with­out switching the meter off if you are going to take further measurements)
4. "°C" Measurement
5. "O2" Measurement
6. ZERO Calibration Trimmer
7. SLOPE Calibration Trimmer
8. °C Calibration Trimmer
9. Rotary Switch
4
5
SPECIFICATIONS
PROBE INITIAL PREPARATION
Range 0.00 to 19.99 mg/L O
0.0 to 50.0°C (32 to 122 °F)
Resolution 0.01 mg/L O
0.1°C
Accuracy ±1.5% F.S. (O2)
±0.5°C
Typical EMC ±1.5% F.S. (O2) Deviation ±1°C
Calibration Manual, 2 point (zero and slope) Temperature Automatic, 0 to 30°C
Compensation (32 to 86°F) Probe HI 76401 polarographic
with 3 m (10') cable (included)
Battery Type 1 x 9V Life approx. 100 hours of continuous use
Environment 0 to 50°C (32 to 122°F); RH max 95% Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 240 g (8.4 oz.)
2
2
PROBE PREPARATION
All DO probes are shipped dry. To hydrate the probe and prepare it for use, connect it to the meter and proceed as follows.
1. Remove the red & black
Shipping
cap
plastic cap. This cap is only used for protecting the
black
probe during shipment, and can be thrown away.
red
2. Wet the sensor by soaking the bottom
(2.5 cm/1") of the probe in HI 7041S elec­trolyte solution for 5 minutes.
3. Rinse the supllied membrane (HI 76407A)
with some electrolyte while shaking it gen­tly. Refill with clean electrolyte.
4. Gently tap the sides of the
membrane with your finger tip to ensure that no air bubbles remain trapped in­side. To avoid damaging the membrane, do not tap the membrane directly on the bottom.
5. Make sure that the rubber
O-ring sits properly inside the membrane cap.
6. With the sensor facing
down, screw the cap clock­wise. Some electrolyte will overflow.
When the probe is not used, protect the membrane with the supplied cap.
6
7
PROBE POLARIZATION
When the probe is under polarization it is continuously fed with a voltage of approxi­mately 800 mV. Probe polarization is essential for stable mea­surements to be taken with the same degree of accuracy. With polarized probe, oxygen is continually "consumed" by passing through the sensitive diaphragm and dissolving in the electrolyte solution contained in the probe. If this operation is interrupted, the electrolyte solution continues to be enriched with oxy­gen until it reaches an equilibrium with the surrounding solution. If measurements are taken with a non-polar­ized probe, the oxygen level detected by the meter is the sum of the oxygen dissolved in the tested sample and that of the electrolyte solution itself.
STB
By leaving the instrument in the "STB" mode (i.e. keep-
OFF
°C
O
ing the probe in a continual state of polarization), the oxy­gen of the electrolyte solu­tion is progressively "con­sumed", and the following measurements de­tect only the quantity of oxygen dissolved in the sample to be tested. Therefore, in order to keep the probe continu­ously polarized it is necessary to switch the meter to the "STB" mode and wait for at least 30 minutes. This should be done with the probe covered with the protective cap and filled with some electrolyte solution.
Polarization is not maintained when the instrument is
OFF
STB
°C
O
switched OFF.
DISSOLVED OXYGEN CALIBRATION
The calibration should be verified:
• after approximately 20 hours of use
• whenever the electrolyte or membrane is
changed and after cleaning the electrode
• after use in aggressive solutions.
Accessories:
• 200 cc of HI 7040 solution
• 1 bottle of HI 7041 solution
Calibration at sea level:
Switch the meter ON, set
the knob to the "STB"
OFF
STB
°C
O
2
position and wait 30 min­utes for a complete polar­ization of the probe.
2
ZERO CALIBRATION
Switch from "STB" to "O2"
position.
OFF
STB
°C
O
2
Remove the membrane
protective cap, immerse the probe in HI 7040 zero oxygen solution and wait for approx. 5 minutes.
The indications given by the instrument
must fall to stabilization levels. If the probe has not been sufficiently polarized, the readout will continue to fluctuate. In this case, switch back to "STB" position, place
OFF
STB
°C
O
2
the protective cap and
2
wait for a few hours for complete polarization be­fore proceeding.
8
9
Using a small screwdriver, turn the "Zero"
SLOPE
ZERO
°C
calibration trimmer until the display reads zero.
ZERO
SLOPE
°C
• If the zero point is passed, "1" will appear on the display. If you cannot get zero reading, the probe is probably defective. Check the mem­brane, electrode and electrolyte solution.
After the zero calibration has been completed, rinse the probe thoroughly with tap water.
SLOPE CALIBRATION
In order to perform a highly accurate calibration, take a B.O.D. bottle and fill with water to a depth of 1-2 cm (½ - ¾"). Seal the bottle with the probe inside. If this calibration is carried out with the bottle opened to air rather than according to the conditions described above, the margin of error is approximately 0.1 ppm.
Switch the instrument to "°C", wait for reading to
OFF
STB
°C
O
stabilize and note the temperature of the water.
See Table 1 on page 12 for the DO reading corre­sponding at the noted
STB
temperature.
• Switch to "O2" position
OFF
°C
O
and adjust the "Slope" trimmer until the correct reading is displayed.
E.g. at 20.5°C the trimmer should be ad­justed to read 9.08.
ZERO
SLOPE
°C
ALTITUDE COMPENSATION
If the calibration is not performed at sea level, a correction should be made for the altitude effect according to Table 2 on page 13. E.g. at 20.5°C and at an altitude of 300 m above sea level, the "Slope" trimmer should be adjusted to read 9.08 x 0.96 = 8.72.
SALINITY COMPENSATION
An additional correction has to be performed if the sample presents significant salinity con­centrations. In such case, the reading has to be adjusted according to Table 3 on page 14. E.g. at 20.5°C and if the sample has a salin­ity content of 3 g/L, the "Slope" trimmer should be adjusted to read 9.08 - (3 x 0.0478) = 8.94.
2
ZERO
SLOPE
°C
2
10
11
Table 1
Dissolved Oxygen (ppm) in Fresh Water
as a function of Temperature (°C)
Table 2
Correction for measurements
at different altitude
°C
0.0
0.1
0.2
0
14.62
14.58
1
14.23
14.19
2
13.84
13.80
3
13.48
13.45
4
13.13
13.10
5
12.80
12.77
6
12.48
12.45
7
12.17
12.14
8
11.87
11.84
9
11.59
11.33
11.08
10.83
10.60
10.37
10.15
9.95
9.74
9.54
9.35
9.17
8.99
8.83
8.68
8.53
8.38
8.22
8.07
7.92
7.77
7.63
7.50
7.40
7.30
7.20
7.10
7.00
6.90
6.80
6.70
6.60
6.50
6.40
6.30
6.20
6.10
6.00
5.90
5.80
5.70
5.60
11.56
11.31
11.06
10.81
10.58
10.35
10.13
9.93
9.72
9.52
9.33
9.15
8.97
8.82
8.67
8.52
8.36
8.21
8.06
7.91
7.76
7.62
7.49
7.39
7.29
7.19
7.09
6.99
6.89
6.79
6.69
6.59
6.49
6.39
6.29
6.19
6.09
5.99
5.89
5.79
5.69
5.59
10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50
14.54
14.15
13.77
13.41
13.06
12.74
12.42
12.11
11.81
11.54
11.28
11.03
10.78
10.55
10.33
10.11
9.91
9.70
9.50
9.31
9.13
8.96
8.80
8.65
8.50
8.35
8.19
8.04
7.89
7.74
7.60
7.48
7.38
7.28
7.18
7.08
6.98
6.88
6.78
6.68
6.58
6.48
6.38
6.28
6.18
6.08
5.98
5.88
5.78
5.68
5.58
0.3
14.50
14.11
13.73
13.38
13.03
12.70
12.39
12.08
11.79
11.51
11.26
11.01
10.76
10.53
10.30
10.09
9.89
9.68
9.48
9.30
9.12
8.94
8.79
8.64
8.49
8.33
8.18
8.03
7.88
7.73
7.59
7.47
7.37
7.27
7.17
7.07
6.97
6.87
6.77
6.67
6.57
6.47
6.37
6.27
6.17
6.07
5.97
5.87
5.77
5.67
5.57
0.4
14.46
14.07
13.70
13.34
13.00
12.67
12.36
12.05
11.76
11.49
11.23
10.98
10.74
10.51
10.28
10.07
9.87
9.66
9.46
9.28
9.10
8.93
8.77
8.62
8.47
8.32
8.16
8.01
7.86
7.71
7.58
7.46
7.36
7.26
7.16
7.06
6.96
6.86
6.76
6.66
6.56
6.46
6.36
6.26
6.16
6.06
5.96
5.86
5.76
5.66
5.56
0.5
14.43
14.03
13.66
13.31
12.97
12.64
12.33
12.02
11.73
11.46
11.21
10.96
10.72
10.49
10.26
10.05
9.84
9.64
9.44
9.26
9.08
8.91
8.76
8.61
8.46
8.30
8.15
8.00
7.85
7.70
7.57
7.45
7.35
7.25
7.15
7.05
6.95
6.85
6.75
6.65
6.55
6.45
6.35
6.25
6.15
6.05
5.95
5.85
5.75
5.65
5.55
0.6
14.39
14.00
13.62
13.27
12.93
12.61
12.29
11.99
11.70
11.43
11.18
10.93
10.69
10.46
10.24
10.03
9.82
9.62
9.43
9.24
9.06
8.89
8.74
8.59
8.44
8.28
8.13
7.98
7.83
7.69
7.55
7.44
7.34
7.24
7.14
7.04
6.94
6.84
6.74
6.64
6.54
6.44
6.34
6.24
6.14
6.04
5.94
5.84
5.74
5.64
5.54
0.7
14.35
13.96
13.59
13.24
12.90
12.58
12.26
11.96
11.67
11.41
11.16
10.91
10.67
10.44
10.22
10.01
9.80
9.60
9.41
9.22
9.04
8.88
8.73
8.53
8.53
8.27
8.12
7.97
7.82
7.67
7.54
7.43
7.33
7.23
7.13
7.03
6.93
6.83
6.73
6.63
6.53
6.43
6.33
6.23
6.13
6.03
5.93
5.83
5.73
5.63
5.53
0.8
14.31
13.92
13.55
13.20
12.87
12.54
12.23
11.93
11.65
11.38
11.13
10.89
10.65
10.42
10.19
9.99
9.78
9.58
9.39
9.21
9.03
8.86
8.71
8.56
8.41
8.25
8.10
7.95
7.80
7.66
7.53
7.42
7.32
7.22
7.12
7.02
6.92
6.82
6.72
6.62
6.52
6.42
6.32
6.22
6.12
6.02
5.92
5.82
5.72
5.62
5.52
0.9
14.27
13.88
13.52
13.17
12.83
12.51
12.20
11.90
11.62
11.36
11.11
10.85
10.62
10.39
10.17
9.97
9.76
9.56
9.37
9.19
9.01
8.85
8.70
8.55
8.40
8.24
8.09
7.94
7.79
7.64
7.51
7.41
7.31
7.21
7.11
7.01
6.91
6.81
6.71
6.61
6.51
6.41
6.31
6.21
6.11
6.01
5.91
5.81
5.71
5.61
5.51
ALTITUDE (METERS)
Sea level
50 100 150 200 300 400 500 600 700 800 900
1000 1100 1200 1300 1400 1500 1600 1700 1800 1900
ATMOSPHERIC
PRESSURE
KPa
101.3
100.7
100.1
99.4
98.8
97.6
96.4
95.2
94.0
92.8
91.7
90.5
89.4
88.3
87.2
86.1
85.0
84.0
82.9
81.9
80.9
79.9
CORRECTION
FACTOR
1.00
0.99
0.99
0.98
0.98
0.96
0.95
0.94
0.93
0.92
0.90
0.89
0.88
0.87
0.86
0.85
0.84
0.83
0.82
0.81
0.80
0.79
12
13
Correction for salinity effect
SLOPE
ZERO
°C
Quantity to be
°C
subtracted per mg/l
of NaCl
0 1 2 3 4 5 6 7 8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
0.0892
0.0861
0.0830
0.0802
0.0779
0.0749
0.0724
0.0701
0.0678
0.0657
0.0637
0.0618
0.0599
0.0582
0.0565
0.0549
0.0533
0.0519
0.0505
0.0491
0.0478
0.0466
0.0454
0.0442
0.0431
0.0421
Table 3
°C
26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50
Quantity to be
subtracted per mg/l of
NaCl
0.0410
0.0400
0.0391
0.0382
0.0373
0.0364
0.0356
0.0348
0.0341
0.0333
0.0326
0.0319
0.0312
0.0306
0.0299
0.0293
0.0287
0.0281
0.0275
0.0270
0.0265
0.0259
0.0254
0.0249
0.0244
TEMPERATURE CALIBRATION
The temperature scale should be calibrated at least every 3 months or whenever the readings are in doubt.
• Immerse the bottom 60 mm (2.4") of the probe in a beaker containing wa­ter.
• Switch the knob to the "°C" position and agitate
OFF
STB
°C
O
2
the water for at least 10 minutes to reach the ther­mal equilibrium between the probe and water.
Using a
ChecktempC
or
°C
another thermometer with a resolution of 0.1°C, check the temperature of the water (e.g. 20.0°C).
Adjust the "°C" trimmer until the display shows the noted temperature.
Note: It is important to ensure the thermal
equilibrium between probe and water, because 1 degree of temperature difference entails an error of ±3% in dissolved oxygen measurement.
14
15
TAKING MEASUREMENTS
PROBE & MEMBRANE MAINTENANCE
Make sure the meter has been calibrated and the protective cap has been removed.
Immerse the tip of the probe in the sample to be tested.
For accurate dissolved oxygen measurements a water move­ment of at least 0.3 m/sec is required.
This movement will ensure that the oxygen­depleted membrane surface is constantly re­plenished. A moving stream will provide ad­equate circulation.
To quickly check if the water speed is suffi­cient, wait for stable reading and then move the D.O. probe. If the reading remains stable, measurement conditions are correct. If the reading increases, the water movement is too low.
During field measurements, this condition may be met by manually agitating the probe. Ac­curate readings are not possible while the liquid is at rest.
During laboratory measurements, the use of a magnetic stirrer is recommended. In this way errors due to the diffusion of ambient oxygen in the solution are minimized.
Anyway, the time necessary for reaching the thermal equilibrium between probe and sample must be allowed (a few minutes for a tem­perature difference of several degrees).
The D.O. probe body is made of reinforced plastic for maximum durability. A thermistor temperature sensor provides tem­perature measurements and compensation. When not in use, it is always recommended to protect the probe against damage and dirt with the supplied cap.
To replace the membrane or refill it with electrolyte, proceed as follows:
Remove the protective
cap by gently twisting and pulling it as sshown in figure 1.
Unscrew the membrane
by turning it counter­clockwise (see fig.2)
TWIST
AND
PULL
fig. 1
• Wet the sensor by
soaking the bottom (2.5 cm / 1") of the probe in HI 7041S electrolyte solution for 5 minutes.
Rinse the new mem-
brane (HI 76407A, sup- plied with the meter) with some electrolyte while shaking gently. Refill with clean elec­trolyte.
Gently tap the sides of
the membrane with your finger to ensure that no air bubbles remain trapped inside. Do no directly tap the bottom as this will damage the membrane.
Make sure that the rubber O-ring is seated
properly inside the membrane cap.
16
17
With the sensor facing down, screw the membrane cap clockwise. Some electro­lyte will overflow.
BATTERY REPLACEMENT
The Platinum cathode (#8 in the Functional Description at page 4) should always be bright and untarnished. If it is tarnished or stained, which could be due to contact with certain gases or extended use with a loose or dam­aged membrane, the cathode should be cleaned. Use a lint-free cardboard or cloth and rub the cathode very gently side to side 4-5 times. This will be enough to polish and remove any stains without damaging the plati­num tip. Afterwards, rinse the probe with deionized or distilled water and install a new membrane cap using fresh electrolyte. Recalibrate the instrument.
Important: In order to have accurate and stable measurements, it is important that the surface of the membrane is in perfect condi­tion. This semi-permeable membrane isolates the sensor elements from the environment but allows oxygen to enter. If any dirt is observed on the membrane, rinse carefully with distilled or deionized water. If any imper­fection still exists, or any damage is evident (such as wrinkles or tears), the membrane should be replaced. Make sure that the O­Ring is properly seated in the membrane cap.
When the battery level is low, a "V" will be displayed
V
to warn the user that the battery should be replaced.
The meter will still work for approximately 4 hours, then the display will be shut-off to avoid erroneous readings.
Battery replacement must only take place in a safe area and using the battery type speci­fied in this instruction manual.
Slide off the battery cover on the rear of the meter and replace the 9V battery with a new one.
18
BATTERY
19
ACCESSORIES
WARRANTY
HI 7040M Zero oxygen solution, 230 mL HI 7040L Zero oxygen solution, 500 mL HI 7041S Refilling electrolyte solution,
30 mL
HI 710001 Soft carrying case HI 710009 Shockproof rubber boot, blue HI 710010 Shockproof rubber boot, orange HI 76401 D.O. probe with 3 m (10') cable HI 76407A/P Spare membrane (5 pcs)
All Hanna Instruments meters are warranted for two years against defects in workman-
ship and materials when used for their in­tended purpose and maintained according to the instructions. The probes are warranted for a period of six months.
Damages due to accident, misuse, tamper­ing or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charge for repair or replacement. If the instrument is to be returned to Hanna Instruments, obtain a Return Goods Authorization from the Cus­tomer Service Department first and then send it with shipment cost prepaid. When shipping any instrument, make sure it is properly pack­aged for complete protection.
20
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to
modify the design, construction and appear-
ance of its products without advance notice.
21
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which it is used.
Operation of this instrument in residential area could cause unacceptable interfer­ences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
22
23
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
24
MAN8043R2 08/05
Loading...