Hanna Instruments HI 8043 User Manual

Instruction Manual
HI 8043
Portable
Dissolved Oxygen Meter
DO meter
HI 8043
STB
OFF
°C
O
2
ZERO
SLOPE
rum
°C
n
s
This Instrument is in
Compliance with the CE
Directives
1
Dear Customer, Thank you for choosing a Hanna Instruments
Product. Please read this instruction manual carefully
before using the instrument. This manual will provide you with all the neces-
sary information for the correct use of the instrument, as well as a more precise idea of its versatility in a wide range of applications.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with the directives.
TABLE OF CONTENTS
Preliminary Examination............................. 3
General Description .................................... 3
Functional Description Probe...................... 4
Functional Description Meter...................... 5
Specifications ............................................. 6
Initial Probe Preparation ............................. 7
Dissolved Oxygen Calibration ..................... 9
Temperature Calibration ............................ 15
Taking Measurements .............................. 16
Probe & Membrane Maintenance ............. 17
Battery Replacement ................................ 19
Accessories.............................................. 20
Warranty ................................................... 21
CE Declaration of Confirmity .................... 22
2
PRELIMINARY EXAMINATION
Remove the instrument from the packing ma­terial and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer.
Each meter is supplied with:
DO Meter
DO Probe (HI 76401)
Membrane cap (2 pcs)
O-ring (2 pcs)
Protective cap
30 mL electrolyte solution (HI 7041S)
Screwdriver for calibration purposes
1 x 9V battery
Instruction manual
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective items must be returned in the original packaging with the supplied accessories.
GENERAL DESCRIPTION
HI 8043 is a portable meter for Dissolved
Oxygen measurements. It is housed in a lightweight case, with an easy-to-read LCD. Dissolved oxygen and temperature measure­ments can be performed in the field (waste­water treatment, fish-farming, water analysis, etc.) as well as in the laboratory without compromising accuracy. Dissolved oxygen is indicated in mg/l (ppm) and this value is compensated for the tem­perature effect (ATC). The dissolved oxygen probe has a membrane covering the polarographic sensors and a built­in thermistor for temperature measurements and compensation. The thin permeable mem-
3
brane isolates the sensor elements from the sample tested, but allows oxygen to enter. A voltage is applied across the sensor, and the oxygen that has passed through the mem­brane reacts causing the current to flow, hence determining a reading. A convenient feature of the instrument is its standby mode which keeps the probe polarized and ready for instant measurements.
FUNCTIONAL DESCRIPTION PROBE
1. D.O. Probe
2. Protective Cap
3. Watertight Shielded Cable
4. Probe Body
5. O-Ring Seal
6. Silver Chloride Anode
7. Platinum Cathode (sensor)
8. Oxygen Permeable PTFE Membrane
9. Membrane Cap
4
FUNCTIONAL DESCRIPTION
1
3
DO meter
STB
OFF
HI 8043
°C
O
2
42 5
9
6 7
ZERO SLOPE °C
8
s
ins
rum n
1.Liquid Crystal Display
2."OFF" (to switch the meter off)
3."STB" (to switch the meter on and polarize the probe. Use this standby position with­out switching the meter off if you are going to take further measurements)
4."°C" Measurement
5."O2" Measurement
6.ZERO Calibration Trimmer
7.SLOPE Calibration Trimmer
8.°C Calibration Trimmer
9.Rotary Switch
5
SPECIFICATIONS
Range DO 0.00 to 19.99 mg/l
Resolution DO 0.01 mg/l
Accuracy DO ±1.5% of full scale
Typical EMC DO ±1.5% of full scale Deviation °C ±1°C
Calibration Manual on 2 points Temperature Automatic from 0 to 30°C
Compensation (32 to 86°F) Probe HI 76401 polarographic
Battery 9V / 100 hours of continuous use Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 650 g (23 oz.) with probe and
°C 0.0 to 50.0°C (32 to 122 °F)
°C 0.1°C
°C ±0.5°C
with 3 m cable
soft carrying case
6
INITIAL PROBE PREPARATION
Shipping
PROBE PREPARATION
All probes shipped from Hanna Instruments are dry. To hydrate the probe and prepare it for use, connect it to the meter and proceed as follows. 1 Remove the red and black
plastic cap. This cap is for
cap
shipping purposes and can be thrown away.
2 Wet the sensor by soak-
black
ing the bottom 2½ cm (1") of the probe in electrolyte
red
(HI 7041S) for 5 minutes.
3 Rinse the supllied mem-
brane (HI 76407A) with electrolyte while shaking it gently. Refill with clean electrolyte.
4 Gently tap the sides of the
membrane with your finger tip to ensure that no air bubbles remain trapped. To avoid damaging the mem­brane, do not tap the mem­brane directly on the bot­tom.
5 Make sure that the rubber O-ring sits prop-
erly inside the membrane cap.
6 With the sensor facing down, screw the
cap clockwise. Some electrolyte will over­flow.
When not in use, place the protective cap supplied in the kit with the meter.
7
PROBE POLARIZATION
When the probe is under polarization it is continuously fed with a voltage of approxi­mately 800 mV. Probe polarization is essential for stable mea­surements to be taken with the same recur­ring degree of accuracy. In the presence of the polarized probe, oxy­gen is continually "consumed" by passing through the sensitive diaphragm and dissolv­ing in the electrolyte solution contained in the probe. If this operation is interrupted, the electrolyte solution continues to be enriched with oxy­gen until it reaches an equilibrium with the surrounding solution. Whenever measurements are taken with a non-polarized probe, the oxygen level detected by the meter is both that of the tested sample as well as the oxygen present in the electro­lyte solution itself. By leaving the instrument in
OFF
STB
°C
O
the "STB" mode, which sig­nifies keeping the probe in a continual state of polariza­tion, the oxygen which is present in the electrolyte solution is progres­sively "consumed", and the following mea­surements detect only the quantity of oxygen present in the sample to be tested. Therefore, in order to keep the probe continu­ously polarized it is necessary to switch the meter to the "STB" mode and wait for at least 30 minutes. This should be done with the probe covered with the protective cap and filled with some electrolyte solution.
STB
Polarization is not maintained when the instrument is
OFF
°C
O
switched to the "OFF" posi­tion.
2
2
8
DISSOLVED OXYGEN CALIBRATION
The calibration should be verified:
- After approximately 20 hours of use.
- Whenever the electrolyte or membrane is
changed and after cleaning the electrode.
- After excessive use in aggressive solu-
tions.
Accessories:
- 200 cc of HI 7040 solution
- 1 bottle of HI 7041 solution
Procedure for calibration at sea level:
If you are starting from
the "OFF" position, switch
OFF
STB
°C
O
to "STB" and wait 30 min­utes for complete polar­ization of the probe be­fore proceeding to the cali­bration procedure.
2
ZERO CALIBRATION:
Switch from "STB" posi-
tion to "O2" position.
Remove the protective
cap and immerse the probe in the HI 7040 zero oxygen solution and wait for approximately 5 min­utes.
The indications given by
the instrument must fall to stabilization levels. If the probe has not been sufficiently polarized, the readout will continue to fluctuate.
9
OFF
STB
°C
O
2
In this case, switch back to "STB" position, place
OFF
STB
°C
O
the protective cap and wait for a few hours for complete polarization be­fore proceeding.
Using a small screwdriver, turn the Zero Cali-
bration screw until the display reads zero.
ZERO
SLOPE
°C
• If the zero point is
passed, "1" will appear on the display. If you can­not get zero reading the probe is probably defective, in which case you should check the membrane, electrode and electrolyte solution (see below).
Rinse the probe thoroughly with normal tap water after the zero calibration has been completed.
2
SLOPE CALIBRATION:
In order to perform a highly accurate calibration, take a B.O.D. bottle and then fill with water to a depth of 1 or 2 cm (½ or ¾"), then seal the bottle with the probe inside. If this calibration is to be car­ried out with the bottle opened to air rather than according to the conditions described above, the margin of error is approximately 0.1 ppm.
Switch the instrument to
"°C", wait for readings to stabilize and then note the temperature of the water.
10
OFF
STB
°C
O
2
Refer to Table 1 on page 10 for the corre-
sponding dissolved oxygen reading at that temperature.
• Switch to "O2" position
and adjust the "Slope"
OFF
STB
°C
O
trimmer until the correct reading is displayed; e.g. at 20.5°C the trimmer should be adjusted until display reads 9.08.
ZERO
SLOPE
°C
CORRECTION FOR ALTITUDE EFFECT
If the calibration is not performed at sea level a correction should be made for the differ­ence in altitude according to Table 2 on page 11; e.g. if calibrated at 20.5°C at an altitude of 300 m above sea level, the display should be adjusted to 9.08 x 0.96 = 8.72.
2
ZERO
SLOPE
°C
CORRECTION FOR SALINITY EFFECT
Another correction has to be performed if the sample presents significant salinity concen­trations. In such case, the reading has to be adjusted according to Table 3 on page 12; e.g. if calibrated at 20.5°C and the sample has a salinity content of 3 g/l, the display should be set at 9.08 - (3 x 0.0478) = 8.94.
ZERO
SLOPE
°C
11
Table 1
Dissolved Oxygen (ppm) in Fresh Water
as a function of Temperature (°C)
°C
0.0
0.1
0.2
0
14.62
14.58
1
14.23
14.19
2
13.84
13.80
3
13.48
13.45
4
13.13
13.10
5
12.80
12.77
6
12.48
12.45
7
12.17
12.14
8
11.87
11.84
9
11.59
11.33
11.08
10.83
10.60
10.37
10.15
9.95
9.74
9.54
9.35
9.17
8.99
8.83
8.68
8.53
8.38
8.22
8.07
7.92
7.77
7.63
7.50
7.40
7.30
7.20
7.10
7.00
6.90
6.80
6.70
6.60
6.50
6.40
6.30
6.20
6.10
6.00
5.90
5.80
5.70
5.60
11.56
11.31
11.06
10.81
10.58
10.35
10.13
9.93
9.72
9.52
9.33
9.15
8.97
8.82
8.67
8.52
8.36
8.21
8.06
7.91
7.76
7.62
7.49
7.39
7.29
7.19
7.09
6.99
6.89
6.79
6.69
6.59
6.49
6.39
6.29
6.19
6.09
5.99
5.89
5.79
5.69
5.59
10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50
14.54
14.15
13.77
13.41
13.06
12.74
12.42
12.11
11.81
11.54
11.28
11.03
10.78
10.55
10.33
10.11
9.91
9.70
9.50
9.31
9.13
8.96
8.80
8.65
8.50
8.35
8.19
8.04
7.89
7.74
7.60
7.48
7.38
7.28
7.18
7.08
6.98
6.88
6.78
6.68
6.58
6.48
6.38
6.28
6.18
6.08
5.98
5.88
5.78
5.68
5.58
0.3
14.50
14.11
13.73
13.38
13.03
12.70
12.39
12.08
11.79
11.51
11.26
11.01
10.76
10.53
10.30
10.09
9.89
9.68
9.48
9.30
9.12
8.94
8.79
8.64
8.49
8.33
8.18
8.03
7.88
7.73
7.59
7.47
7.37
7.27
7.17
7.07
6.97
6.87
6.77
6.67
6.57
6.47
6.37
6.27
6.17
6.07
5.97
5.87
5.77
5.67
5.57
0.4
14.46
14.07
13.70
13.34
13.00
12.67
12.36
12.05
11.76
11.49
11.23
10.98
10.74
10.51
10.28
10.07
9.87
9.66
9.46
9.28
9.10
8.93
8.77
8.62
8.47
8.32
8.16
8.01
7.86
7.71
7.58
7.46
7.36
7.26
7.16
7.06
6.96
6.86
6.76
6.66
6.56
6.46
6.36
6.26
6.16
6.06
5.96
5.86
5.76
5.66
5.56
0.5
14.43
14.03
13.66
13.31
12.97
12.64
12.33
12.02
11.73
11.46
11.21
10.96
10.72
10.49
10.26
10.05
9.84
9.64
9.44
9.26
9.08
8.91
8.76
8.61
8.46
8.30
8.15
8.00
7.85
7.70
7.57
7.45
7.35
7.25
7.15
7.05
6.95
6.85
6.75
6.65
6.55
6.45
6.35
6.25
6.15
6.05
5.95
5.85
5.75
5.65
5.55
0.6
14.39
14.00
13.62
13.27
12.93
12.61
12.29
11.99
11.70
11.43
11.18
10.93
10.69
10.46
10.24
10.03
9.82
9.62
9.43
9.24
9.06
8.89
8.74
8.59
8.44
8.28
8.13
7.98
7.83
7.69
7.55
7.44
7.34
7.24
7.14
7.04
6.94
6.84
6.74
6.64
6.54
6.44
6.34
6.24
6.14
6.04
5.94
5.84
5.74
5.64
5.54
0.7
14.35
13.96
13.59
13.24
12.90
12.58
12.26
11.96
11.67
11.41
11.16
10.91
10.67
10.44
10.22
10.01
9.80
9.60
9.41
9.22
9.04
8.88
8.73
8.53
8.53
8.27
8.12
7.97
7.82
7.67
7.54
7.43
7.33
7.23
7.13
7.03
6.93
6.83
6.73
6.63
6.53
6.43
6.33
6.23
6.13
6.03
5.93
5.83
5.73
5.63
5.53
0.8
14.31
13.92
13.55
13.20
12.87
12.54
12.23
11.93
11.65
11.38
11.13
10.89
10.65
10.42
10.19
9.99
9.78
9.58
9.39
9.21
9.03
8.86
8.71
8.56
8.41
8.25
8.10
7.95
7.80
7.66
7.53
7.42
7.32
7.22
7.12
7.02
6.92
6.82
6.72
6.62
6.52
6.42
6.32
6.22
6.12
6.02
5.92
5.82
5.72
5.62
5.52
0.9
14.27
13.88
13.52
13.17
12.83
12.51
12.20
11.90
11.62
11.36
11.11
10.85
10.62
10.39
10.17
9.97
9.76
9.56
9.37
9.19
9.01
8.85
8.70
8.55
8.40
8.24
8.09
7.94
7.79
7.64
7.51
7.41
7.31
7.21
7.11
7.01
6.91
6.81
6.71
6.61
6.51
6.41
6.31
6.21
6.11
6.01
5.91
5.81
5.71
5.61
5.51
12
Table 2
Correction for measurements
at different altitude
ALTITUDE (METERS)
Sea level
50 100 150 200 300 400 500 600 700 800 900
1000 1100 1200 1300 1400 1500 1600 1700 1800 1900
ATMOSPHERIC
PRESSURE
KPa
101.3
100.7
100.1
99.4
98.8
97.6
96.4
95.2
94.0
92.8
91.7
90.5
89.4
88.3
87.2
86.1
85.0
84.0
82.9
81.9
80.9
79.9
CORRECTION
FACTOR
1.00
0.99
0.99
0.98
0.98
0.96
0.95
0.94
0.93
0.92
0.90
0.89
0.88
0.87
0.86
0.85
0.84
0.83
0.82
0.81
0.80
0.79
13
Table 3
Correction for salinity effect
Quantity to be
°C
subtracted per mg/l
of NaCl
0 1 2 3 4 5 6 7 8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
0.0892
0.0861
0.0830
0.0802
0.0779
0.0749
0.0724
0.0701
0.0678
0.0657
0.0637
0.0618
0.0599
0.0582
0.0565
0.0549
0.0533
0.0519
0.0505
0.0491
0.0478
0.0466
0.0454
0.0442
0.0431
0.0421
°C
26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50
Quantity to be
subtracted per mg/l of
NaCl
0.0410
0.0400
0.0391
0.0382
0.0373
0.0364
0.0356
0.0348
0.0341
0.0333
0.0326
0.0319
0.0312
0.0306
0.0299
0.0293
0.0287
0.0281
0.0275
0.0270
0.0265
0.0259
0.0254
0.0249
0.0244
14
TEMPERATURE CALIBRATION
The temperature should be calibrated at least every 3 months or whenever the readings are in doubt, using the following procedure:
• Immerse the bottom 60 mm (2.4") of the probe in a beaker of water.
Switch the instrument to °C.
STB
°C
Agitate the water for at
OFF
O
2
least 10 minutes until ther­mal equilibrium between the probe and water is achieved.
°
• Check the temperature
C
of the water using ChecktempC or a ther­mometer with a resolution of 0.1°C (e.g. 20.0°C).
Adjust the °C trimmer until the display reads the thermometer temperature value.
ZERO
SLOPE
°C
Note: It is important to ensure thermal
equilibrium between probe and the water as 1 degree of temperature difference entails an error of ±3% in dissolved oxygen measurement.
15
TAKING MEASUREMENTS
Make sure the meter has been calibrated and the protective cap has been removed.
Immerse the tip of the probe in the sample to be tested.
For accurate dissolved oxygen measurements a water move­ment of 0.3 m/sec is required at a minimum. This is to en­sure that the oxygen-depleted membrane surface is con­stantly replenished. A moving stream will provide adequate circulation. To quickly check if the water speed is sufficient, wait for the reading to be stable and then move the D.O. probe. If the reading is still stable, the measurement con­ditions are right, while if the reading increases the water movement is too low.
During field measurements, this condition may be met by manually agitating the probe. Ac­curate readings are not possible while the liquid is at rest.
During laboratory measurements, the use of a magnetic stirrer to ensure a certain velocity in the fluid is recommended. In this way, errors due to the diffusion of the oxygen present in the air in the solution are reduced to a minimum.
At all times, the time necessary for thermal equilibrium to occur between the probe and the sample must be allowed (a few minutes for temperature difference of several degrees).
16
PROBE & MEMBRANE MAINTENANCE
The oxygen probe body is made of reinforced plastic for maximum durability. A thermistor temperature sensor provides tem­perature measurements of the sample tested. It is always recommended that the protective cap be kept on the probe when the probe is not in use to provide protection against dam­age and dirt.
To replace the membrane or refill it with electrolyte, proceed as follows:
Remove the protective
cap by gently twisting and pulling it off the body of the probe (see fig. 1).
Unscrew the membrane
TWIST
AND
PULL
by turning it counter­clockwise with the
fig. 1
other hand (see fig.2)
• Wet the sensor by
soaking the bottom 2½ cm (1") of the probe in electrolyte (HI 7041S) for 5 minutes.
Rinse the new mem-
brane (HI 76407A) sup­plied with the meter with electrolyte while shaking it gently. Refill with clean electrolyte.
Gently tap the sides of
the membrane with your finger tip to ensure that no air bubbles remain trapped. Do no directly tap the bottom with your finger as this will dam­age the membrane.
17
Make sure that the rubber O-ring is seated
properly inside the membrane cap.
With the sensor facing down, screw the
membrane cap clockwise. Some electro­lyte will overflow.
The Platinum cathode (#8 in the Functional Description at page 2) should always be bright and untarnished. If it is tarnished or stained, which could be due to contact with certain gases or extended use with a loose or dam­aged membrane, the cathode should be cleaned. You can use a clean lint-free card­board or cloth. Rub the cathode very gently side to side 4-5 times. This will be enough to polish and remove any stains without dam­aging the platinum tip. Afterwards, rinse the probe with deionized or distilled water and install a new membrane cap using fresh elec­trolyte and follow the steps above. Recalibrate the instrument.
Important: in order to have accurate and stable measurements, it is important that the surface of the membrane is in perfect condi­tion. This semipermeable membrane isolates the sensor elements from the environment but allows oxygen to enter. If any dirt is observed on the membrane, rinse carefully with distilled or deionized water. If any imper­fections still exist, or any damage is evident (such as wrinkles or tears - holes), the mem­brane should be replaced. Make sure that the O-Ring is properly seated in the membrane cap.
18
BATTERY REPLACEMENT
When the battery is rundown "V" is displayed on the Liq-
V
uid Crystal Display to warn the user.
This is to inform the user that the display will be shut-off after about 4 hours of use to prevent erroneous measurements being taken due to low voltage.
Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual (see page 17).
In order to replace rundown batteries, simply slide the cover off and replace the 9V battery with a new one.
BATTERY
19
ACCESSORIES
BATT9/P 9V battery (10 pcs) HI 7040M Zero Oxygen calibration solu-
tion, 230 mL
HI 7040L Zero Oxygen calibration solu-
tion, 460 mL
HI 7041S Refilling Electrolyte solution,
30 mL
HI 76401 D.O. probe with 3 meters (10')
cable
HI 76407A/P 5 spare membranes HI 76501/P Screwdrivers, length 90 mm for
calibration purposes (20 pcs)
20
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workman-
ship and materials when used for their in­tended purpose and maintained according to the instructions. The probes are warranted for a period of six months.
Damages due to accident, misuse, tamper­ing or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charge for repair or replacement. If the instrument is to be returned to Hanna Instruments, obtain a Return Goods Authorization from the Cus­tomer Service Department first and then send it with shipment cost prepaid. When shipping any instrument, make sure it is properly pack­aged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and appear­ance of its products without advance notice.
21
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which it is used.
Operation of this instrument in residential area could cause unacceptable interfer­ences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
22
HANNA PRODUCTS
• CALIBRATION AND MAINTENANCE SOLUTIONS
• CHEMICAL TEST KITS
• CHLORINE METERS
• CONDUCTIVITY/TDS METERS
• DISSOLVED OXYGEN METERS
• HYGROMETERS
• ION SPECIFIC METERS
• MAGNETIC STIRRERS
• Na/NaCl METERS
• pH/ORP/Na ELECTRODES
• PROBES (DO, µS/cm, RH, T, TDS)
• PUMPS
• REAGENTS
• SOFTWARE
• THERMOMETERS
• TITRATORS
• TRANSMITTERS
• TURBIDITY METERS
• Wide Range of Accessories
Most Hanna meters are available in the fol-
lowing formats:
• BENCH-TOP METERS
• POCKET-SIZE METERS
• PORTABLE METERS
• PRINTING/LOGGING METERS
• PROCESS METERS
(Panel- and Wall-mounted)
• WATERPROOF METERS
• METERS FOR FOOD INDUSTRY
For additional information, contact your dealer or the nearest Hanna Customer Service Cen­ter.
You can also e-mail us at:
tech@hannainst.com
23
Stay in Touch with the Manufacturer!
24
MAN8043R1
07/96
Loading...