Dear Customer,
Thank you for choosing a Hanna Product.
Please read this instruction manual carefully before using the instru-
ment. If you need additional technical information, do not hesitate to
e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
WARRANTY
All Hanna Instruments meters are warranted for two years
against defects in workmanship and materials when used for their
intended purpose and maintained according to instructions. Probesare warranted for a period of six months. This warranty is
limited to repair or replacement free of charge. Damages due to
accident, misuse, tampering or lack of prescribed maintenance are not
covered. If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be
notified of the charges incurred. If the instrument is to be returned to
Hanna Instruments, first obtain a Returned Goods Authorization
number from the Customer Service department and then send it with
shipping costs prepaid. When shipping any instrument, make sure it
is properly packaged for complete protection.
CE DECLARATION OF CONFORMITY............................................ 14
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any noticeable damage, notify your Dealer or the nearest
Hanna office immediately.
Each meter is supplied with:
• Conductivity probe with 1 m (3.3') cable
• Calibration screwdriver (HI 933000 only)
• 9V battery and instruction manual
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any damaged or defective item
must be returned in its original packing materials together
with the supplied accessories.
GENERAL DESCRIPTION
HI 8033 and HI 933000 are some of the most complete and versatile
portable conductivity meters ever manufactured.
Designed with utmost precision and simplicity, these meters provide for
up to 3 or 4 measurement ranges.
The conductivity of a solution depends on the temperature and for
this reason measurements are carried out with reference to a
standard temperature of 25°C.
If the solution measured has a different temperature than 25°C,
compensation must be performed.
HI 8033 compensates for temperature manually, while HI 933000,
with a built-in temperature sensor and circuitry, automatically compensates for temperature changes.
The temperature coefficient is fixed at 2%.
All rights are reserved. Reproduction in whole or in part is prohibited without the
written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
Compensationwith ß = 2%/°C
Probe (included)HI 76301W with 1 m (3.3') cable
Environment0 to 50°C (32 to 122°F); RH max 95%
Battery Type1 x 9 V alkaline
Battery LifeApprox. 100 hours of continuous use
Dimensions185 x 82 x 47 mm (7.3 x 3.2 x1.9")
Weight270 g (9.5 oz.)
1) Liquid Crystal Display
2) ON/OFF key
3) Measurement range selection keys
4) Probe connector
Range0.0 to 199.9 / 0 to 1999 µS/cm
0.00 to 19.99 / 0.0 to 199.9 mS/cm
Resolution0.1 / 1 µS/cm ; 0.01 / 0.1 mS/cm
Accuracy (@ 20°C / 68°F) ±1% FS (excluding probe error)
Typical EMC Deviation±2% FS
CalibrationManual, 1 point through trimmer
TemperatureAutomatic, 10 to 40°C (50 to 104°F)
Compensationwith ß = 2%/°C
Probe (included)HI 76302W with 1 m (3.3') cable
Environment0 to 50°C (32 to 122°F); RH max 95%
Battery Type1 x 9 V alkaline
Battery LifeApprox. 100 hours of continuous use
Dimensions143 x 80 x 38 mm (5.6 x 3.2 x1.5")
Weight360 g (13 oz.)
54
OPERATIONAL GUIDE
CALIBRATION
• Each meter is supplied complete with a 9V battery. Slide off the
battery compartment cover on the back of the meter and install
the battery while paying attention to its polarity (see "Battery
Replacement" section for details).
• Connect the probe to the meter by aligning the pins with the
socket and pushing the plug in. With HI 933000, tighten the
threaded ring.
• Make sure that the meter has been calibrated before taking any
measurements (see "Calibration" section for details).
• Immerse the conductivity probe into the
sample, making sure that the holes on
the shaft are completely submerged.
If possible, use plastic beakers to minimize
any EMC interference.
• Tap the probe lightly on the bottom of the beaker to remove any
air bubbles which may be trapped inside the sleeve.
• Turn the instrument on by pressing the ON/OFF key or by setting
the rotary switch (HI 8033) to the desired measurement range.
• For HI 8033:
°C
Measure the solution temperature with a
ChecktempC or another accurate thermometer, and set the temperature knob
to the measured value (e.g. 20°C).
• Select the appropriate measurement range.
Note: If the display shows only a "1", the
meter is out of range. Select the next
(higher) range.
• For HI 933000:
Before taking any measurement, wait a couple of minutes for the
temperature sensor to reach thermal equilibrium with the sample.
When the temperature of the sample is lower than 20°C or higher
than 30°C, allow more time for the thermal equilibrium to be
achieved.
• After measurements, switch the instrument off and clean the
probe (see "Probe Maintenance" section for details).
The instrument should be re-calibrated at least once a month, or
after probe or battery replacement.
For more accurate results, it is recommended to use a calibration
solution with a conductivity value close to the range to be measured.
See the "Accessories" section for a wide selection of conductivity
solutions.
PROCEDURE FOR HI 8033
• Pour the conductivity calibration solution
(e.g. HI 7030) into a beaker while making sure that the holes on the probe are
completely submerged. If possible, use
plastic beakers to minimize any EMC interference.
• Immerse the conductivity probe and the
ChecktempC in the solution.
• Wait a couple of minutes for the thermal
equilibrium to be reached.
• Tap the probe on the bottom, then shake it while rotating to
make sure no air bubbles remain trapped in the sleeve.
• Measure the temperature of the calibration solution with the
ChecktempC, and adjust the °C knob accordingly.
• Turn the rotary knob to select the 19990 µS/cm range.
• Turn the k% calibration knob until the display shows the
conductivity reading at 25°C (see "Conductivity vs. temperature
chart "), e.g. 12880 µS/cm (= 12.88 mS/cm), and all
subsequent measurements will be compensated to 25°C (77°F).
• If you prefer to standardize the temperature compensation to
20°C (68°F), adjust the trimmer to read "11.67 mS" (see
"Conductivity vs. temperature chart "). All subsequent measurements will be compensated to 20°C.
• The calibration is now complete and the meter is ready for use.
HI 7030
°C
PROCEDURE FOR HI 933000
• Pour the conductivity calibration solution (e.g. HI 7030) into a
beaker while making sure that the holes on the probe are
completely submerged. If possible, use plastic beakers to minimize
any EMC interference.
76
• Immerse the conductivity probe in the
solution and wait a couple of minutes for
thermal equilibrium to be reached.
• Tap the probe on the bottom, then shake
it while rotating to make sure no air
bubbles remain trapped in the sleeve.
• Switch the instrument on by pressing ON/OFF, and select the
19.99 mS/cm range by pressing the corresponding range key.
• With the supplied screwdriver, adjust the calibration trimmer in
the battery compartment until the display shows "12.88 mS", i.e.
the solution conductivity value at 25°C. All subsequent measurements will be compensated to 25°C (77°F).
CONDUCTIVITY VERSUS
TEMPERATURE CHART
The conductivity of an aqueous solution is the measure of its ability to
carry an electrical current by means of ionic motion.
The conductivity invariably increases with increasing temperature.
It is affected by the type and number of ions in the solution and by
the viscosity of the solution itself. Both parameters are temperature
dependent. The dependency of conductivity on temperature is expressed as a relative change per degree Celsius at a particular
temperature, commonly as percent per °C.
For manual temperature compensation, refer to the following chart:
mS
9 VOLT BATTERY
• If you prefer to standardize the temperature compensation to
20°C (68°F), adjust the trimmer to read "11.67 mS" (see
"Conductivity vs. temperature chart "), and all subsequent measurements will be compensated to 20°C.
• The calibration is now complete and the meter is ready for use.
For instance, the conductivity values of the calibration solutions at
25°C are 12880 µS/cm, 1413 µS/cm or 5000 µS/cm when using
HI 7030, HI 7031 or HI 7039, respectively.
At 20°C, the values are 11670 µS/cm, 1278 µS/cm or
4523 µS/cm, respectively.
With the solutions at 30°C, the values are 14120 µS/cm, 1548 µS/cm
or 5479 µS/cm, respectively.
98
PROBE MAINTENANCE
BATTERY
HI 8033
HI 8633HI 8733HI 8734
SLIDE
OFF
BATTERY REPLACEMENT
After every series of measurements, rinse
the probe with tap water.
If a more thorough cleaning is required,
remove the PVC sleeve and clean the
probe with a cloth or a non-abrasive
detergent. When reinserting the sleeve
onto the probe, be sure that the sleeve
is in the right direction with the holes
towards the cable end.
After cleaning the probe, re-calibrate
the instrument.
The probe body is made of PVC. For this
reason it must never come into close
contact with a heat source.
If the probe is exposed to high temperatures (above 50°C/122°F), the rings
might become loose or detached, resulting in a serious impairment of the probe.
In such cases, the probe has to be
replaced.
When the battery becomes weak, a "V" will be displayed to warn the
user that only a few hours of working life are left.
A low battery can result in unreliable measurements. It is recommended to replace it immediately.
Battery replacement must only take place in a non-hazardous area,
and using an alkaline 9V battery.
Slide off the battery compartment cover (see figure below) and replace
the 9V battery with a new one.
Make sure the battery contacts are tight and secure before replacing
the cover.
HI 933000
9 VOLT BATTERY
SLIDE
OFF
1110
ACCESSORIES
A
A
CALIBRATION SOLUTIONS
HI 7030L12880 µS/cm, 500 mL bottle
HI 7030M12880 µS/cm, 230 mL bottle
HI 7031L1413 µS/cm, 500 mL bottle
HI 7031M1413 µS/cm, 230 mL bottle
HI 7033L84 µS/cm, 500 mL bottle
HI 7033M84 µS/cm, 230 mL bottle
HI 7034L80000 µS/cm, 500 mL bottle
HI 7034M80000 µS/cm, 230 mL bottle
HI 7035L111800 µS/cm, 500 mL bottle
HI 7035M111800 µS/cm, 230 mL bottle
HI 7039L5000 µS/cm, 500 mL bottle
HI 7039M5000 µS/cm, 230 mL bottle
HI 7032L1382 ppm (mg/L), 500 mL bottle
HI 7032M1382 ppm (mg/L), 230 mL bottle
HI 7036L12.41 ppt (g/L), 500 mL bottle
HI 7036M12.41 ppt (g/L), 230 mL bottle
HI 8030L12880 µS/cm, 500 mL FDA bottle
HI 8031L1413 µS/cm, 500 mL FDA bottle
HI 8033L84 µS/cm, 500 mL FDA bottle
HI 8034L80000 µS/cm, 500 mL FDA bottle
HI 8035L111800 µS/cm, 500 mL FDA bottle
HI 8039L5000 µS/cm, 500 mL FDA bottle
CONDUCTIVITY PROBES
HI 76301W 4-ring conductivity probe with 1 m (3.3') cable
19,5mm
DIA
0.77"
85mm
118.5mm
4.66"
3.34"
11.5mm
0.45"
DI
HI 76302W 4-ring conductivity probe with built-in temperature
sensor and 1 m (3.3') cable
19,5mm
DIA
0.77"
85mm
118.5mm
4.66"
3.34"
11.5mm
0.45"
DI
OTHER ACCESSORIES
ChecktempC Electronic thermometer (range: -50.0 to 150.0°C)
HI 710001Soft carrying case
HI 710007Blue shockproof rubber boot for HI 933000
HI 710008Orange shockproof rubber boot for HI 933000
HI 710009Blue shockproof rubber boot for HI 8033
HI 710010Orange shockproof rubber boot for HI 8033
HI 731326Calibration screwdriver (20 pcs)
1312
CE DECLARATION OF CONFORMITY
HI 8033 and HI 933000
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they
are used. Operation of these instruments in residential area could cause unacceptable interference to radio
and TV equipment, requiring the operator to take all necessary steps to correct interferences.
The metal rings of the probe are sensitive to electrostatic discharges. Avoid touching these rings at all times.
During calibration of the instruments, ESD wrist straps should be worn to avoid possible damage to the probe
by electrostatic discharge. Use plastic beakers to minimize any EMC interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC
performance. To avoid electrical shock, do not use these instruments when voltages at the measurement
surface exceed 24 Vac or 60 Vdc.
To avoid damage or burns, do not perform any measurement in microwave ovens.
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66-2619-0708 • Fax. 66-2619-0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and
Technical offices, please see www.hannainst.com
MANCONDR3 05/05
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.