Hanna Instruments HI 8033 User Manual

Instruction Manual
HI 8033
HI 933000
Portable Multi-Range
Conductivity Meters
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instru-
ment. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Probes are warranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ...................................................... 3
GENERAL DESCRIPTION .............................................................. 3
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8033........ 4
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 933000.... 5
OPERATIONAL GUIDE ................................................................. 6
CALIBRATION ............................................................................. 7
CONDUCTIVITY VERSUS TEMPERATURE CHART ............................. 9
PROBE MAINTENANCE .............................................................. 10
BATTERY REPLACEMENT ............................................................ 11
ACCESSORIES ........................................................................... 12
CE DECLARATION OF CONFORMITY............................................ 14
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer or the nearest Hanna office immediately.
Each meter is supplied with:
• Conductivity probe with 1 m (3.3') cable
• Calibration screwdriver (HI 933000 only)
• 9V battery and instruction manual
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any damaged or defective item must be returned in its original packing materials together with the supplied accessories.
GENERAL DESCRIPTION
HI 8033 and HI 933000 are some of the most complete and versatile portable conductivity meters ever manufactured.
Designed with utmost precision and simplicity, these meters provide for up to 3 or 4 measurement ranges.
The conductivity of a solution depends on the temperature and for this reason measurements are carried out with reference to a standard temperature of 25°C.
If the solution measured has a different temperature than 25°C, compensation must be performed.
HI 8033 compensates for temperature manually, while HI 933000, with a built-in temperature sensor and circuitry, automatically com­pensates for temperature changes.
The temperature coefficient is fixed at 2%.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
32
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 8033
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 933000
1
mS
Conductivity meter
ppm
salinity
0-199,9
0-19,990
CaCO
3
OFF
HI 8033
µS
0-1999
µS
0-19,990
µS
2
°C
0
10
34
20
30
40
50
+
0
-
ins rum n
k%
15
10
5
5
10
15
s
1) Liquid Crystal Display
2) Rotary switch
3) Knob for manual temperature compensation
4) Calibration knob
Range 0.0 to 199.9 ; 0 to 1999 µS/cm
0.00 to 19.99 mS/cm 0 to 19990 ppm
Resolution 0.1 ; 1 µS/cm / 0.01 mS/cm / 10 ppm Accuracy (@20°C/68°F) ±1% FS (excluding probe error) Typical EMC Deviation ±2 % FS Calibration Manual, 1 point Temperature Manual, 0 to 50°C (32 to 122°F)
Compensation with ß = 2%/°C Probe (included) HI 76301W with 1 m (3.3') cable Environment 0 to 50°C (32 to 122°F); RH max 95% Battery Type 1 x 9 V alkaline Battery Life Approx. 100 hours of continuous use Dimensions 185 x 82 x 47 mm (7.3 x 3.2 x1.9") Weight 270 g (9.5 oz.)
1) Liquid Crystal Display
2) ON/OFF key
3) Measurement range selection keys
4) Probe connector
Range 0.0 to 199.9 / 0 to 1999 µS/cm
0.00 to 19.99 / 0.0 to 199.9 mS/cm
Resolution 0.1 / 1 µS/cm ; 0.01 / 0.1 mS/cm Accuracy (@ 20°C / 68°F) ±1% FS (excluding probe error) Typical EMC Deviation ±2% FS Calibration Manual, 1 point through trimmer Temperature Automatic, 10 to 40°C (50 to 104°F)
Compensation with ß = 2%/°C Probe (included) HI 76302W with 1 m (3.3') cable Environment 0 to 50°C (32 to 122°F); RH max 95% Battery Type 1 x 9 V alkaline Battery Life Approx. 100 hours of continuous use Dimensions 143 x 80 x 38 mm (5.6 x 3.2 x1.5") Weight 360 g (13 oz.)
54
OPERATIONAL GUIDE
CALIBRATION
• Each meter is supplied complete with a 9V battery. Slide off the battery compartment cover on the back of the meter and install the battery while paying attention to its polarity (see "Battery Replacement" section for details).
• Connect the probe to the meter by aligning the pins with the socket and pushing the plug in. With HI 933000, tighten the threaded ring.
• Make sure that the meter has been calibrated before taking any measurements (see "Calibration" section for details).
• Immerse the conductivity probe into the sample, making sure that the holes on the shaft are completely submerged.
If possible, use plastic beakers to minimize any EMC interference.
• Tap the probe lightly on the bottom of the beaker to remove any air bubbles which may be trapped inside the sleeve.
• Turn the instrument on by pressing the ON/OFF key or by setting the rotary switch (HI 8033) to the desired measurement range.
• For HI 8033:
°C
Measure the solution temperature with a ChecktempC or another accurate ther­mometer, and set the temperature knob to the measured value (e.g. 20°C).
• Select the appropriate measurement range.
Note: If the display shows only a "1", the
meter is out of range. Select the next (higher) range.
• For HI 933000:
Before taking any measurement, wait a couple of minutes for the temperature sensor to reach thermal equilibrium with the sample.
When the temperature of the sample is lower than 20°C or higher than 30°C, allow more time for the thermal equilibrium to be achieved.
• After measurements, switch the instrument off and clean the probe (see "Probe Maintenance" section for details).
The instrument should be re-calibrated at least once a month, or after probe or battery replacement.
For more accurate results, it is recommended to use a calibration solution with a conductivity value close to the range to be measured. See the "Accessories" section for a wide selection of conductivity solutions.
PROCEDURE FOR HI 8033
• Pour the conductivity calibration solution (e.g. HI 7030) into a beaker while mak­ing sure that the holes on the probe are completely submerged. If possible, use plastic beakers to minimize any EMC in­terference.
• Immerse the conductivity probe and the ChecktempC in the solution.
• Wait a couple of minutes for the thermal equilibrium to be reached.
• Tap the probe on the bottom, then shake it while rotating to make sure no air bubbles remain trapped in the sleeve.
• Measure the temperature of the calibration solution with the ChecktempC, and adjust the °C knob accordingly.
• Turn the rotary knob to select the 19990 µS/cm range.
• Turn the k% calibration knob until the display shows the conductivity reading at 25°C (see "Conductivity vs. temperature chart "), e.g. 12880 µS/cm (= 12.88 mS/cm), and all subsequent measurements will be compensated to 25°C (77°F).
• If you prefer to standardize the temperature compensation to 20°C (68°F), adjust the trimmer to read "11.67 mS" (see "Conductivity vs. temperature chart "). All subsequent measure­ments will be compensated to 20°C.
• The calibration is now complete and the meter is ready for use.
HI 7030
°C
PROCEDURE FOR HI 933000
• Pour the conductivity calibration solution (e.g. HI 7030) into a beaker while making sure that the holes on the probe are completely submerged. If possible, use plastic beakers to minimize any EMC interference.
76
• Immerse the conductivity probe in the solution and wait a couple of minutes for thermal equilibrium to be reached.
• Tap the probe on the bottom, then shake it while rotating to make sure no air bubbles remain trapped in the sleeve.
• Switch the instrument on by pressing ON/OFF, and select the
19.99 mS/cm range by pressing the corresponding range key.
• With the supplied screwdriver, adjust the calibration trimmer in the battery compartment until the display shows "12.88 mS", i.e. the solution conductivity value at 25°C. All subsequent measure­ments will be compensated to 25°C (77°F).
CONDUCTIVITY VERSUS
TEMPERATURE CHART
The conductivity of an aqueous solution is the measure of its ability to carry an electrical current by means of ionic motion.
The conductivity invariably increases with increasing temperature. It is affected by the type and number of ions in the solution and by
the viscosity of the solution itself. Both parameters are temperature dependent. The dependency of conductivity on temperature is ex­pressed as a relative change per degree Celsius at a particular temperature, commonly as percent per °C.
For manual temperature compensation, refer to the following chart:
mS
9 VOLT BATTERY
• If you prefer to standardize the temperature compensation to 20°C (68°F), adjust the trimmer to read "11.67 mS" (see "Conductivity vs. temperature chart "), and all subsequent mea­surements will be compensated to 20°C.
• The calibration is now complete and the meter is ready for use.
°C °F HI 7030 HI 7031 HI 7033 HI 7034 HI 7035 HI 7039
HI 8030 HI 8031 HI 8033 HI 8034 HI 8035 HI 8039
(µS/cm) (µS/cm) (µS/cm) (µS/cm) (µS/cm) (µS/cm)
0 32 7150 776 64 48300 65400 2760
5 41 8220 896 65 53500 74100 3180 10 50 9330 1020 67 59600 83200 3615 15 59 10480 1147 68 65400 92500 4063 16 60.8 10720 1173 70 67200 94400 4155 17 62.6 10950 1199 71 68500 96300 4245 18 64.4 11190 1225 73 69800 98200 4337 19 66.2 11430 1251 74 71300 100200 4429 20 68 11670 1278 76 72400 102100 4523 21 69.8 11910 1305 78 74000 104000 4617 22 71.6 12150 1332 79 75200 105900 4711 23 73.4 12390 1359 81 76500 107900 4805 24 75.2 12640 1386 82 78300 109800 4902
25 77 12880 1413 84 80000 111800 5000
26 78.8 13130 1440 86 81300 113800 5096 27 80.6 13370 1467 87 83000 115700 5190 28 82.4 13620 1494 89 84900 117700 5286 29 84.2 13870 1521 90 86300 119700 5383 30 86 14120 1548 92 88200 121800 5479 31 87.8 14370 1575 94 90000 123900 5575
For instance, the conductivity values of the calibration solutions at 25°C are 12880 µS/cm, 1413 µS/cm or 5000 µS/cm when using HI 7030, HI 7031 or HI 7039, respectively.
At 20°C, the values are 11670 µS/cm, 1278 µS/cm or 4523 µS/cm, respectively.
With the solutions at 30°C, the values are 14120 µS/cm, 1548 µS/cm or 5479 µS/cm, respectively.
98
PROBE MAINTENANCE
BATTERY
HI 8033 HI 8633 HI 8733 HI 8734
SLIDE
OFF
BATTERY REPLACEMENT
After every series of measurements, rinse the probe with tap water.
If a more thorough cleaning is required, remove the PVC sleeve and clean the probe with a cloth or a non-abrasive detergent. When reinserting the sleeve onto the probe, be sure that the sleeve is in the right direction with the holes towards the cable end.
After cleaning the probe, re-calibrate the instrument.
The probe body is made of PVC. For this reason it must never come into close contact with a heat source.
If the probe is exposed to high tempera­tures (above 50°C/122°F), the rings might become loose or detached, result­ing in a serious impairment of the probe.
In such cases, the probe has to be replaced.
When the battery becomes weak, a "V" will be displayed to warn the user that only a few hours of working life are left.
A low battery can result in unreliable measurements. It is recom­mended to replace it immediately.
Battery replacement must only take place in a non-hazardous area, and using an alkaline 9V battery.
Slide off the battery compartment cover (see figure below) and replace the 9V battery with a new one.
Make sure the battery contacts are tight and secure before replacing the cover.
HI 933000
9 VOLT BATTERY
SLIDE
OFF
1110
ACCESSORIES
A
A
CALIBRATION SOLUTIONS
HI 7030L 12880 µS/cm, 500 mL bottle HI 7030M 12880 µS/cm, 230 mL bottle HI 7031L 1413 µS/cm, 500 mL bottle HI 7031M 1413 µS/cm, 230 mL bottle HI 7033L 84 µS/cm, 500 mL bottle HI 7033M 84 µS/cm, 230 mL bottle HI 7034L 80000 µS/cm, 500 mL bottle HI 7034M 80000 µS/cm, 230 mL bottle HI 7035L 111800 µS/cm, 500 mL bottle HI 7035M 111800 µS/cm, 230 mL bottle HI 7039L 5000 µS/cm, 500 mL bottle HI 7039M 5000 µS/cm, 230 mL bottle HI 7032L 1382 ppm (mg/L), 500 mL bottle HI 7032M 1382 ppm (mg/L), 230 mL bottle HI 7036L 12.41 ppt (g/L), 500 mL bottle HI 7036M 12.41 ppt (g/L), 230 mL bottle
HI 8030L 12880 µS/cm, 500 mL FDA bottle HI 8031L 1413 µS/cm, 500 mL FDA bottle HI 8033L 84 µS/cm, 500 mL FDA bottle HI 8034L 80000 µS/cm, 500 mL FDA bottle HI 8035L 111800 µS/cm, 500 mL FDA bottle HI 8039L 5000 µS/cm, 500 mL FDA bottle
CONDUCTIVITY PROBES
HI 76301W 4-ring conductivity probe with 1 m (3.3') cable
19,5mm
DIA
0.77"
85mm
118.5mm
4.66"
3.34"
11.5mm
0.45"
DI
HI 76302W 4-ring conductivity probe with built-in temperature
sensor and 1 m (3.3') cable
19,5mm
DIA
0.77"
85mm
118.5mm
4.66"
3.34"
11.5mm
0.45"
DI
OTHER ACCESSORIES
ChecktempC Electronic thermometer (range: -50.0 to 150.0°C)
HI 710001 Soft carrying case HI 710007 Blue shockproof rubber boot for HI 933000 HI 710008 Orange shockproof rubber boot for HI 933000 HI 710009 Blue shockproof rubber boot for HI 8033 HI 710010 Orange shockproof rubber boot for HI 8033 HI 731326 Calibration screwdriver (20 pcs)
1312
CE DECLARATION OF CONFORMITY
HI 8033 and HI 933000
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interference to radio and TV equipment, requiring the operator to take all necessary steps to correct interferences.
The metal rings of the probe are sensitive to electrostatic discharges. Avoid touching these rings at all times. During calibration of the instruments, ESD wrist straps should be worn to avoid possible damage to the probe by electrostatic discharge. Use plastic beakers to minimize any EMC interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc.
To avoid damage or burns, do not perform any measurement in microwave ovens.
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66-2619-0708 • Fax. 66-2619-0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
MANCONDR3 05/05
Loading...