Dear Customer, Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for correct opera­tion. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e­mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has oc­curred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Cen­ter.
The meters are supplied with 12 VDC power adapter. Note: Conserve all packing material until the in-
strument has been observed to function correctly because any defective item must be returned in its original packing.
These Instruments are in
Compliance with the CE
Directives
HI 150 is an indicator with a relay Output and built-in LED Alarm Signal for the monitoring of temperature.
Constructed in a compact casing and with a molded eye, HI 150 can be installed on the QC bench or wherever necessary to control the ambient tem­perature of food chain, from the storage room to the point of delivery.
The housing has been completely sealed against vapors and humidity.
The measurement range is from -20 to 60ºC with an excellent ±0.5ºC accuracy. The stainless steel encapsulated sensor is positioned right on the outer body of this thermometer to provide a rapid response.
In order to ensure that temperature control is never interrupted, HI 150 runs permanently on a 12 VDC power supply. Temperature is constantly displayed on a large LCD so that the operator can check it at any time.
HI 150 also offers a unique alarm control. Any value from 0 to 50.0ºC (HI 150-00) or -20 to 20ºC (HI 150-10) can be adjusted as the setpoint through a trimmer located on the back of the meter. Should the temperature vary from this user­selectable setpoint, the red LED blinks and the relay is activated to warn of an abnormality or switch on a system. The built-in LED is well visible from a distance.
This alarm relay is a cost-saving feature with a tremendous potential that can help you in auto­mating your temperature control. It can also be used as a forewarning that the refrigeration or air conditioning unit is malfunctioning, or that a door has been left open, or point to any other mishaps, common in a busy workplace.
HI 150 is a convenient and inexpensive way to ensure that temperature never varies too much from acceptable levels. It frees workers from hav­ing to manually check the temperature and immediately signals if and when an abnormality arises.
HI150-00HI150-00
HI150-00
HI150-00HI150-00
Range -20.0 to 60.0ºC Resolution 0.1ºC Accuracy ±0.5ºC Setpoint Selectable from 0.0 to 50.0 ºC Hysteresis ±0.5ºC around setpoint Alarm LED blinks and alarm relay activated
when measured temperature is lower than the setpoint
Relay Through contacts inside the meter Sensor Incorporated, encapsulated stainless
steel
Environment -20 to 60ºC; RH 100% Power supply External 12 VDC (included) Dimensions 86 x 94 x 33mm (3.4 x 3.7 x 1.3”) Weight 150g (5.2oz.)
HI150-10HI150-10
HI150-10
HI150-10HI150-10
Range -20.0 to 60.0ºC Resolution 0.1ºC Accuracy ±0.5ºC Setpoint Selectable from -20.0 to 20.0 ºC Hysteresis ±0.5ºC around setpoint Alarm LED blinks and alarm relay activated
when measured temperature is higher than the setpoint
Relay Through contacts inside the meter Sensor Incorporated, encapsulated stainless
steel
Environment -20 to 60ºC; RH 100% Power supply External 12 VDC (included) Dimensions 86 x 94 x 33 mm (3.4 x 3.7 x 1.3”) Weight 150g (5.2oz.)
1. Molded eye
2. Alarm LED
3. Access for relay contacts
4. 12VDC power adapter (included)
5. Power supply connector
6. Sensor
7. Liquid Cristal Display
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
TAKING MEASUREMENTS
• Turn the meter on by connecting the 12 VDC power adapter to the meter and to the mains.
• A value will appear on the LCD to indicate the temperature. Wait for a few minutes for the display to stabilize.
ADJUSTING THE SETPOINT
With the HI 150, you can select your own setpoint and be alerted with a visual LED alarm when an abnormal situation arises.
• Unscrew and remove the rear panel and gasket seal to access the MEASURE/SET switch.
• Move the switch to the left (SET Mode).
• With a small screwdriver adjust the setpoint trimmer to dis­play the desired value in the 0 to 50 ºC or -20 to 20ºC range (based on the model).
• Make sure the switch is moved back to the right (MEASURE Mode).
RELAY CONNECTIONRELAY CONNECTION
RELAY CONNECTION
RELAY CONNECTIONRELAY CONNECTION
• To utilize the relay function open the casing as explained for setpoint calibration.
After wiring close the casing properly. The relay wires should pass through the access hole.
• Whenever the temperature reading reaches the setpoint ( considering the hysteresis ), the alarm LED blinks and relay is activated.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential areas could cause unac­ceptable interferences to radio and TV equipment.
The metal band at the end of the probe is sensitive to electrostatic dis­charges. Avoid touching this metal band at all times.
During operation, ESD wrist straps should be worn to avoid possible dam­age to the probe by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may de­grade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 VAC or 60 VDC.
Use plastic beakers to minimize any EMC interferences. To avoid damages or burns, do not perform any measurement in micro-
wave ovens.
CE DECLARATIONCE DECLARATION
CE DECLARATION
CE DECLARATIONCE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
OF CONFORMITYOF CONFORMITY
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 150-00/10 thermometer is warranted for one year
against defects in workmanship and materials when used for their intended purpose and maintained ac­cording to instructions. The sensor are warranted for a period of one year also. This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service de­partment and then send it with shipment costs prepaid. When shipping any instrument, make sure it is prop­erly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
• Replace the rear panel and the gasket, ensuring the unit is properly closed.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.