PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
Each meter is supplied with battery and instructions
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
GENERAL DESCRIPTION
These instruments are specially designed to meet the needs
of simple continuous monitoring of temperature.
The housing has been completely sealed against vapors and
humidity with IP54 rating.
You can simply hang the meter right above the sample to be
tested for continuous measurement.
The probe is easy to clean and requires little maintenance.
Two models are available:
• HI 146-00 displays temperature in Centigrade degrees
• HI 146-01 displays temperature in Fahrenheit degrees.
A protected switch allows a quick check of the meter accuracy
at any time.
FUNCTIONAL DESCRIPTION
1. Molded eye
2. Probe
3. Liquid Crystal Display
SPECIFICATIONS
HI 146-00
Range -50.0 to 150.0°C
Resolution 0.1°C
Accuracy ±0.3°C (-20 to 90°C), ±0.5°C (outside)
Typical EMC Deviation ±0.3°C
HI 146-01
Range -58.0 to 302°F
Resolution 0.1°F (-58.0 to 199.9°F), 1°F(200 to 302°F)
Accuracy ±0.5°F (-4 to 194°F), ±1°F (outside)
Typical EMC Deviation ±0.5°F
Common Specifications
Probe Stainless steel probe with 2 m (6.6’) cable
Battery Type 1 x 1.5V AA alkaline
Battery Life approx. 5 years
Casing IP54
Environment 0 to 50°C (32 to 122°F); RH max 95%
Dimensions 86 x 110 x 43 mm (3.4 x 4.3 x 1.7’’)
Weight 150 g (5.3 oz.)
OPERATIONAL GUIDE
OPERATION
• Dip the probe in the sample.
An immersion level of 4 cm (1 ½") is
recommended.
• Do NOT immerse the probe handle.
• Do NOT wrench the probe or put heavy
objects on the cable. If the cable is
damaged, the meter needs to be replaced.
• Wait for a few seconds for the display
to stabilize.
• Variations in readings or fading display can indicate a weak battery.
CALIBRATION CHECK SWITCH
Remove the back cover and move the internal switch to the
“TEST” position.
The display should show 0.0±0.3°C (32.0°F±0.5°F).
Contact your dealer or the nearest Hanna Service Center for
recalibration, if necessary.
To return to the measurement mode, move back the switch
to “MEAS” position. Replace the back cover and tighten the
screws.
BATTERY REPLACEMENT
When the display fades, remove the back cover and replace
the 1.5V AA battery with a new one, while paying attention to its polarity. Close the meter.
Battery should only be replaced in a safe area using the
battery type specified in this instruction manual.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make
sure that they are entirely suitable
for the environment in which they
are used. Operation of these instruments in residential areas could
cause unacceptable interferences to
radio and TV equipment. Any variation introduced by the user to the
supplied equipment may degrade the
instrument’s EMC performance. To
avoid electrical shock, do not use
these instruments when voltages at
the measurement surface exceed 24
Vac or 60 Vdc. To avoid damages
or burns, do not perform any measurement in microwave ovens.
WARRANTY
HI146 is warranted for two years against defects in workmanship and materials when used
for its intended purpose and maintained according to instructions. The probe is warranted
for a period of six months.
This warranty is limited to repair or replacement free of charge. Damages due to accident,
misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under
warranty, report the model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer Service department and then send it with
shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for
complete protection.
IST146R2 05/05