Cleaning
Wipe the outer surface of the pipette daily with a mild soap and water. In
case of accidental contamination of the pipette shaft, immediately remove
the ejector sleeve and lower body assembly (refer to Pipette Parts Diagrams).
Rinse the ejector sleeve and lower body assembly with a mild soap and
water followed by distilled water. Dry all components completely. Lightly
lubricate the piston (Lubricant, Part No. 5763-01). Hold the plunger button
down to engage the piston into the seal. After reattaching the lower body
assembly, attach the injector sleeve over the lower body C-ring.
Topic Action
Consistent Technique Pipette with a consistent rhythm, pressure, and speed
Tip Selection Use Hamilton pipette tips for guaranteed accuracy.
For information, visit www.hamiltoncompany.com
Tip Size and Fit Use proper size tip, firmly placed on pipette shaft
Pre-Rinse Tips Pre-rinse pipette tips for improved precision
Sample Aspiration Keep the disposable tip immersed in fluid during aspira-
tion. Do not let the plunger snap back to starting position.
Immersion Depth Maintain an immersion depth of 2 to 4 mm
Immersion Angle
Maintain an angle of <20º from vertical
Viscous Samples Aspirate slowly. If bubbles are observed, resample. Volume
errors may still occur. Refer to the Calibration section.
Acid Samples Pipetting strong acids and corrosive solutions is not
recommended. These liquids may damage the piston and
seal. To protect the pipette, use filter barrier pipette tips
and inspect piston frequently. For more information, visit
www.hamiltoncompany.com
High Vapor Pressure
Samples
Pipetting solutions with high vapor pressure is not recommended. These liquids may damage the piston and seal.
Sample Temperature Fluids at a temperature other than that for which the
pipette and pipette tips have been calibrated may result in
volume measurement errors.
Storage Store upright in stand. Do not lay the pipette on its side
with fluid in the tip. Fluid reaching the piston causes
contamination and possible corrosion.
Cleaning Piston Wipe piston with alcohol and a soft, lint-free cloth. Dry and
lightly lubricate the piston (Lubricant, Part No. 5763-01).
Seal Replacement Replace the TEFLON® PTFE seals every 3-6 months
depending on use and the samples aspirated
Performance Checks Check the accuracy and precision of your pipette every
3-6 months depending on use and the samples aspirated
Optimizing Performance