Hamilton MICROLAB 500, MICROLAB 540B, MICROLAB 510B, MICROLAB 511C, MICROLAB 531C User Manual

...
Artisan Technology Group is your source for quality
new and certied-used/pre-owned equipment
FAST SHIPPING AND
DELIVERY
TENS OF THOUSANDS OF IN-STOCK ITEMS
• EQUIPMENT DEMOS
HUNDREDS OF MANUFACTURERS
SUPPORTED
• LEASING/MONTHLY RENTALS
• ITAR CERTIFIED SECURE ASSET SOLUTIONS
SERVICE CENTER REPAIRS
Experienced engineers and technicians on staff at our full-service, in-house repair center
Instra
Remotely inspect equipment before purchasing with our interactive website at www.instraview.com
Contact us: (888) 88-SOURCE | sales@artisantg.com | www.artisantg.com
SM
REMOTE INSPECTION
View
WE BUY USED EQUIPMENT
Sell your excess, underutilized, and idle used equipment We also offer credit for buy-backs and trade-ins
www.artisantg.com/WeBuyEquipment
LOOKING FOR MORE INFORMATION?
Visit us on the web at www.artisantg.com for more information on price quotations, drivers, technical
specications, manuals, and documentation
R
USERS
MANUAL
®
MICROLAB 500B/C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The MICROLAB ® 500 Series:
MICROLAB 510B, 530B, and 540B
and MICROLAB 511C, 531C, and 541C
User’s Manual
R
THE M
EASURE OF EXCELLENCE
Part Number 69176 (Rev. D)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SM
.
B Hamilton Company Instrument Warranty
Hamilton Company warrants this equipment (except valves*) to be free of defects in material and workmanship for 12 months from the date of receipt. This warranty is extended to the buyer of record on the original purchase order to Hamilton Company. Hamilton Company or an authorized Hamilton representative agrees to repair or replace, at its option and free of charge to the buyer at a normal place of business or at a Hamilton repair facility, any part or parts that under proper and normal use prove to be defective during the warranty period.** Abuse, unauthorized replacement of parts, modifications, or adjustments made by other than Company or its assigned representatives voids this warranty.
This warranty gives you specific rights. No other warranties, expressed or implied, including implications of warranties of merchantability and fitness for a particular product, are made. Hamilton Company's liability on the sale of all products shall be limited to repair, replacement, or refund of price of any defective product.**
Hamilton Company endeavors to provide prompt and satisfactory service. *All Hamilton Company valves are warranted to be free of defects in material and
workmanship at the time of delivery. **Hamilton Company reserves the right to refuse to accept the return of any instrument or valve
that has been used with radioactive or microbiological substances, or any other material that may be deemed hazardous to employees of Hamilton Company.
© July 1999 by Hamilton Company GASTIGHT is a registered trademark of Hamilton Company. MICROLAB is a registered trademark licensed to Hamilton Company. SANI-CLOTH is a registered trademark of Professional Disposables, Inc.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Instruction to the User
This equipment has been tested and found to comply with the limits for a class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in an installation. This equipment generates, uses, and can radiate radio frequency energy, and if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
This equipment has been verified to comply with the limits for a class B computing device, pursuant to FCC Rules. In order to maintain compliance with the FCC regulations, shielded cables must be used with this equipment . Operation with non-approved equipment or unshielded cables is likely to result in interference to radio and TV reception. The user is cautioned that changes and modifications made to the equipment without the approval of the manufacturer could void the user’s authority to operate this equipment.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Contents
Figures and Tables..................................................................xi
Preface ....................................................................PR–1
About the MICROLAB 500 Series of Instruments.....................PR–2
Upgrading Your MICROLAB 500 System..............................PR–3
About This Manual.................................................................PR–4
Conventions Used in This Manual ...........................................PR–5
A Word About Single Syringe Instruments ...............................PR–5
Chapter 1 Getting Started..................................................1–1
MICROLAB 500 Parts Lists.......................................................1 –2
A Brief Introduction to the MICROLAB 500B..............................1 – 8
Drive Unit..............................................................................1–8
Power Cord Connector Receptacle ................................... 1–13
Hand Probe or Footswitch Connector Receptacle..............1–13
Fuse Box.........................................................................1–13
Communications Settings.................................................1–13
Power On/Off Switch and Power Indicator Light.................1–14
Step/Prime Switch...........................................................1–14
Valve Assembly...............................................................1–15
Syringe Drive Arms........................................................... 1–15
Controller Unit........................................................................1–16
Pipettors/Probes................................................................... 1–18
Concorde Push-button Hand Pipettor/Probe.....................1–18
Dual Push-button Hand Pipettor/Probe ............................. 1–18
v
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Disposable Tip Push-button Hand Pipettor/Probes ............ 1–19
Luer Lock Tip Push-button Pipette Hand Pipettor/Probe....1–19
Viscous Sample Push-button Hand Pipette
Pipettor/Probe.................................................................1–20
Footswitch..........................................................................1–20
Probe Button Functions ...................................................... 1–20
Chapter 2 Installing the MICROLAB 500 System ....................2–1
Overview of Installation Procedures...........................................2–3
Selecting a Location.................................................................2 – 3
Installing the Accessory Holder.................................................2 – 3
Selecting Communications Options.......................................... 2– 4
Installing Electrical Connections................................................2–6
Installing Valve Assemblies.......................................................2–8
Installing the Valve Assembly on the ML 510B/511C...............2–8
Installing Valve Assemblies on the
ML530B/531C and 540B/541C ............................................. 2– 9
Selecting, Installing, and Removing Syringes...........................2–10
Preparing Syringes for Installation........................................2–12
Installing Syringes...............................................................2–12
Removing Syringes.............................................................2–14
Selecting and Installing Tubing...............................................2–16
Selecting Tubing................................................................ 2–16
Installing Tubing..................................................................2–17
Chapter 3 Programming and Using the
MICROLAB 500B System.....................................3–1
Using the Controller Unit...........................................................3–3
Using the Arrow Keys............................................................3 –3
vi MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Using the Numeric Keypad....................................................3–4
Using the Run/Stop Key........................................................3 – 4
Using the Function Keys.......................................................3– 4
Editing Conventions............................................................. 3– 5
Powering on the MICROLAB 500..............................................3 – 6
The Main Menu........................................................................3 –7
Priming the MICROLAB 500.....................................................3–8
Preparing to Prime................................................................3–8
Using the Prime Function from the Main Menu......................3–11
Creating a New Method..........................................................3–13
Creating an Aliquot Dispense Method..................................3–17
Creating a Serial Dispense Method.......................................3–18
Creating a Dilution Method...................................................3–20
Creating a Pipette Method...................................................3–23
Creating a Titrate Method.....................................................3–25
Creating a Custom Method..................................................3–28
Running an Existing Method...................................................3 –30
Editing an Existing Method .....................................................3–32
Performing Manual Dilutions...................................................3–33
Performing Manual Dispenses................................................3–35
The Utilities Menu..................................................................3–37
Renaming an Existing Method.............................................3 –38
Deleting an Existing Method ................................................3–40
Copying a Method to a Custom Method................................3–41
Downloading a Method to the Drive unit................................3–42
Diagnostic and Configuration Menu......................................3–44
Changing the Default Language.......................................3–44
Contents vii
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setting the Baud Rate to the Drive Unit..............................3–45
Turning Caps Lock On and Off..........................................3–46
Performing the Display Test..............................................3–46
Performing the Keyboard Test .......................................... 3–46
Sample MICROLAB 510B Application Configurations...............3–47
Single Syringe Dispensing............................................ 3–47
Sample MICROLAB 530B Application Configurations...............3–48
Example 1: Dilutions......................................................3–48
Example 2: Dilutions......................................................3–49
Sample MICROLAB 540B Application Configurations...............3–50
Example 1: Dual Dispensing..........................................3–50
Example 2: Using the Dual Dispenser for
Single Dispensing......................................................... 3–50
Chapter 4 Caring for the MICROLAB 500..............................4–1
Deciding When to Clean the MICROLAB 500.............................4–2
Cleaning the Fluid Path of the MICROLAB 500..........................4 –2
Cleaning Syringes and Tubing..................................................4 – 3
Cleaning the Exterior of the MICROLAB 500..............................4–4
Chemical Compatibility..............................................................4–5
Storing the MICROLAB 500......................................................4 – 6
Replacing Batteries..................................................................4 – 6
Chapter 5 Troubleshooting the MICROLAB 500.................... 5 –1
Error Message Code Guide.......................................................5–2
Audible Messages.................................................................5–10
Troubleshooting Guide.......................................................... 5–11
Getting Technical Support ......................................................5–15
Returning Instruments for Repair.............................................5–16
viii MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Appendixes
Appendix A Technical Specifications for the MICROLAB 500........................A–1
Pin Outs for RS-232 and TTL Ports...........................................A –3
Appendix B Instrument Performance Test Reports.......................................B – 1
Appendix C Ordering Parts and Accessories for the MICROLAB 500.............C–1
Appendix D Chemical Compatibility of the MICROLAB 500............................ D –1
Appendix E Communication Protocols.........................................................E – 1
Hamilton Protocol 1/RNO+ Syntax Overview..........................E –1
Auto-addressing ................................................................E–2
Data Transfer Format ...........................................................E-4
Data String Components.....................................................E-5
DIN Protocol/BDZ+ Syntax..................................................... E–7
Hardwire-addressing..........................................................E – 8
Auto-addressing .................................................................E-8
Extablishing a Data Transfer Session....................................E-9
Terminating a Data Transfer Session................................... E-10
Data Transfer Session......................................................E–10
Data Block Format............................................................ E–10
Data String Components...................................................E-11
Broadcast Addressing ....................................................... E-16
Stored Methods................................................................E-16
Stored Method Definition..................................................E-16
Creating Stored Methods..................................................E-16
Stored Method Execution.................................................E-17
Appendix F Protocol Command Summary....................................................F –1
Channel Selection Commands.............................................F-2
Contents ix
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Initialization Commands .......................................................F-3
Syringe Positioning Commands ........................................... F-4
Valve Positioning Commands..............................................F-5
Timer and Digital I/O Commands...........................................F-6
Execution Commands ......................................................... F-7
Instrument Control Commands.............................................F-8
Syringe Parameter Change..................................................F-9
Valve Parameter Change...................................................F-10
Stored Method Parameter Change.....................................F-11
Instrument Information Requests.......................................F-11
Instrument Status Requests.............................................. F-13
Syringe Parameter Request...............................................F-19
Valve Parameter Request..................................................F-20
Timer and Digital I/O Requests...........................................F-21
Firmware Version Request.................................................F-22
ASCII Chart..........................................................................F-22
ML500B Default Environmental Parameters...........................F-26
ML500B Command Buffer .................................................... F-26
ML500B Methods ................................................................ F-27
ML500B Specific Protocol Information...................................F-27
Glossary ....................................................................GL–1
Index .....................................................................IN–1
x MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figures and Tables
Preface
Figure PR–1 The MICROLAB 500 Series of Instruments.........PR–2
Chapter 1
Figure 1–1 MICROLAB 500B System Components...............1– 3
Figure 1–2 MICROLAB 500B Shipping Kit Components........1 –5
Figure 1–3 Small Parts Kit.....................................................1–6
Figure 1–4 Tubing Kit........................................................... 1 –7
Figure 1–5 Front View of theML510B Single
Syringe Dispenser.............................................. 1 – 9
Figure 1–6 Front View of the ML530B Dual Syringe Diluter... 1–10 Figure 1–7 Front View of the ML540B Dual
Syringe Dispenser............................................ 1–11
Figure 1–8 Rear View of the ML510B, 530B, and 540B........1–12
Figure 1–9 The Controller Unit............................................1–17
Figure 1–10 The Concorde, the Disposable Tip, and the Dual Hand
Pipettor/Probes ................................................1–19
Chapter 2
Table 1–1 MICROLAB 500 System Descriptions...................1–2
Table 1–2 MICROLAB 500B Shipping Kit............................. 1 – 4
Table 1–3 Small Parts Kit #35888 (For All Models).................1– 6
Table 1–4 Tubing Kit #35887 (For All Models)...................... 1– 7
Figure 2–1 Overview of Installation Procedures......................2 –2
Figure 2–2 Installing Electrical Connections...........................2 –7
Figure 2–3 Installing a Valve Assembly on the ML510B.......... 2– 8
Figures and Tables xi
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 2–4 Installing a Valve Assembly on the
ML530B and 540B..............................................2 –9
Figure 2–5 The TLL-type Dispenser/Diluent Syringe............ 2–11
Figure 2–6 The TLLX-type Dispenser/Diluent Syringe..........2–11
Figure 2–7 The DX-type Sample Syringe.............................2–11
Figure 2–8 Installing a Syringe ............................................2–13
Figure 2–9 Removing a Syringe..........................................2–15
Figure 2–10 ML510B Valve and Tubing Connections............ 2–18
Figure 2–11 ML530B Valve with a D Configuration Syringe.....2–19
Figure 2–12 ML530B Valve with Two TLLX or TLL Syringes ... 2–19
Figure 2–13 ML540B Valve and Tubing Connectors .............. 2–20
Figure 2–14 Using the Tubing Reducer with the
Dual Hand Probe...............................................2–21
Table 2–1 Communications Switches and Ports....................2–4
Table 2–2 Reagent/Diluent Syringes
(TLL and TLLX-types)........................................2–10
Table 2–3 Sample Syringes (DX-type)................................2–11
Table 2–4 Tubing Selection Guide.....................................2–16
Chapter 3
Figure 3–1 Using the MICROLAB 500B.................................3–1
Figure 3–2 Overview of MICROLAB 500B Functions..............3–2
Figure 3–3 The Controller Unit..............................................3–3
Figure 3–4 Priming the MICROLAB 500..............................3–10
Figure 3–5 The Utilities Menu.............................................3–37
Figure 3–6 Single Syringe Dispensing ................................3–47
Figure 3–7 Dilutions........................................................... 3–48
Figure 3–8 Serial Dilutions..................................................3–49
Figure 3–9 Dual Dispensing................................................ 3–50
Figure 3–10 Using the Dual Dispenser for Single Dispensing.. 3–51
xii MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 5
Appendix A
Appendix B
Appendix C
Table 3–1 Function Key Actions .......................................... 3– 4
Table 5-1 Error Message Code Guide..................................5–2
Table 5–2 Audible System Messages ................................5–10
Table 5–3 Troubleshooting Guide .....................................5–11
Table A–1 Technical Specifications for the MICROLAB 500...A–1
Table A–2 Accuracy and Precision.......................................A–2
Figure B–1 Sample Performance Test Report........................B–2
Table C–1 Reagent/Diluent Syringe Replacement Parts ........C–1
Table C–2 Sample Syringe Replacement Parts .....................C –2
Table C–3 Valve Assemblies................................................C–2
Table C–4 Tubing ...............................................................C– 2
Table C–5 Pipettors/Probes ...............................................C –3
Table C–6 Parts and Accessories......................................... C –3
Appendix D
Appendix E
Appendix F
Table D–1 Chemical Compatibility.........................................D–1
Table E–1 Protocol 1/RNO+ Control Characters....................E– 2
Table E–2 DIN Protocol/BDZ+ Control Characters.................E– 8
Table F-1 Channel Selection Commands.............................F–2
Table F-2 Initialization Commands........................................ F-3
Table F-3 Syringe Positioning Commands............................ F-4
Figures and Tables xiii
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table F-4 Valve Positioning Commands ...............................F-5
Table F-5 Timer and Digital I/O Commands............................F-6
Table F-6 Execution Commands.......................................... F-7
Table F-7 Instrument Control Commands.............................. F-8
Table F-8 Syringe Parameter Change................................... F-9
Table F-9 Valve Parameter Change.................................... F-10
Table F-10 Stored Method Parameter Change......................F-11
Table F-11 Instrument Information Requests ........................F-11
Table F-12 Instrument Status Request.................................F-13
Table F-13 Syringe Parameter Request................................ F-19
Table F-14 Valve Parameter Request...................................F-20
Table F-15 Timer and Digital I/O Requests ............................F-21
Table F-16 Firmware Version Request..................................F-22
Table F-17 ASCII Chart........................................................F-23
xiv MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Preface Welcome to the World of
Hamilton Precision Instruments
Congratulations on your purchase of a Hamilton MICROLAB 500 system. The Hamilton MICROLAB 500 is a versatile, semi­automatic, precision liquid processor. Various models of the MICROLAB 500 function as either single- or dual-syringe diluter/dispensers.
The MICROLAB 500 functions on the principal of liquid/liquid displacement. At the heart of each MICROLAB 500 system is a highly efficient, precision stepper motor drive that is combined with world-famous Hamilton GASTIGHT syringes. The result is a precise and accurate instrument that is very easy to set up and use.
With proper care and maintenance, your new MICROLAB 500 system will serve you faithfully. To learn about the proper care and maintenance of your investment, please take the time to read this manual. Also, please read the warranty information that appears on the copyright page in this manual and on the separate warranty sheet that is included in your MICROLAB 500 shipping kit.
The Hamilton Company thanks you for purchasing this Hamilton product. Welcome to the world of Hamilton precision instruments!
PR–1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
About the MICROLAB 500 Series of Instruments
All of the MICROLAB 500 systems feature four common pipette modes: fill, dispense, auto-refill, and prime. The systems are capable of performing accurate and precise transfer pipetting, and of performing automated dilutions up to 1:25,000. The instruments can also dispense up to 50 mL per cycle. Figure PR–1 provides an overview of the MICROLAB 500 Series of Instruments.
Figure PR–1 The MICROLAB 500 Series of Instruments
The Microlab 500 Series of Instruments
Single Syringe Instruments
Controller without memory:
• Microlab 501A dispenser
Controller with memory:
• Microlab 510B diluter/dispenser
Computer-controlled:
• Microlab 511C diluter/dispenser
Dual Syringe Instruments
Controller without memory:
• Microlab 503A diluter
• Microlab 504A dispenser
Controller with memory:
• Microlab 530B diluter/dispenser
• Microlab 540B dispenser
Computer-controlled:
• Microlab 531C diluter/dispenser
• Microlab 541C dispenser
PR–2 MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The MICROLAB 500 Series of instruments consists of three different lines of diluter/dispensers.
Controller without memory: Use the controller unit to manually enter methods.
This controller unit does not have memory, so methods cannot be stored.
Controller with memory: This controller unit has memory, allowing you to
program and store your own custom methods. Or, use the controller unit to run manual methods. Any instrument that uses this controller can also be computer controlled.
Computer controlled: No controller unit is included; use a computer to operate
the instrument.
The MICROLAB 510B/511C, 530B/531C, and 540B/541C are described in this manual.
The MICROLAB 510B/511C is a single syringe dispenser designed for single
precision dispensing applications.
The MICROLAB 530B/531C is a dual syringe diluter designed for repetitive
dilution applications.
The MICROLAB 540B/541C is a dual syringe dispenser designed for precision
dispensing applications that require more than one liquid to be dispensed at a time.
Upgrading Your MICROLAB 500 System
Both the MICROLAB 510B/511C and 530B/531C systems can be upgraded to MICROLAB 540B/541C systems. For upgrade information or for information about purchasing any of the MICROLAB 500 models, contact your authorized Hamilton sales representative or contact Hamilton Company.
In the United States:
Hamilton Company, P.O. Box 10030, Reno, Nevada 89520–0012 Telephone Numbers (in the USA and Canada): Technical/Customer Service 1–800–648–5950 8 a.m. to 5 p.m. PST Instrument Service 1–800–527–5269
Preface PR–3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Outside the USA and Canada:
+1–775–858–3000
Fax Number: +1–775–856–7259
In Switzerland:
Hamilton Bonaduz AG, Ch–7402, P.O. Box 26, Bonaduz, Switzerland Telephone Number: +41–81–660–60–60 Fax Number: +41–81–660–60–70
About This Manual
This manual provides technical information about the MICROLAB 510B/511C, 530B/531C, and 540B/541C, and is divided into chapters that cover the following topics:
Chapter 1, Getting Started, provides an overview of the MICROLAB 500 system, including a complete parts list and a brief description of the system components.
Chapter 2, Installing the MICROLAB 500, describes how to set up the system.
Chapter 3, Using the MICROLAB 500, provides step-by-step instructions for using the system. It also provides sample applications.
Chapter 4, Caring for the MICROLAB 500, describes everyday maintenance techniques.
Chapter 5, Troubleshooting, contains tables that list system messages and their meanings, and common problems and their solutions.
The Appendixes provide detailed information, such as technical specifications, ordering information, etc.
The Glossary defines terms used in this manual.
The Index provides a quick-reference to the topics described in this manual.
PR–4 MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Conventions Used in This Manual
Throughout this manual symbols are used to call your attention to various kinds of information.
A Warning! Information that is essential for avoiding personal
injury is flagged with the International Warning Symbol and appears like this in the text.
Important Information that is essential for avoiding damage to
equipment appears like this in the text.
Note: Interesting information or information that can help
improve system performance appears like this in the text.
System messages and prompts that appear on the controller unit’s display screen are shown in courier font. Items that are selected and user input appear in
boldface courier.
A
A Word About Single Syringe Instruments
Throughout this manual you will see references to multiple syringes, volumes, speeds, and to the right-side controls. All screen examples show the use of dual syringe instruments. If you are using a MICROLAB 510B/511C single syringe dispenser, please disregard these references. The right-side functions are not available on single syringe instruments.
Any operational differences between the single and the dual syringe instruments are called out in the text.
Preface PR–5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter 1 Getting Started
This chapter provides a brief overview of the MICROLAB 500 system. Information in this chapter includes:
• MICROLAB 500 parts lists
• MICROLAB 500 components – the drive unit – the controller unit – hand pipettors/probes
All MICROLAB 500 instruments come with everything you need to start using the system, with the exception of syringes. You must separately purchase syringes for use with the MICROLAB 500 systems. For complete lists of syringes, accessories, and replacement parts for the MICROLAB 500, see Appendix C, Ordering Parts and Accessories for the MICROLAB 500.
Note: Contact your local delivery company if you
notice any visual damage to the MICROLAB 500 shipping package or to its contents. Also, you may want to save the shipping container in case you ever need to return the instrument for
service.
1–1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
MICROLAB 500 Parts Lists
This section includes complete parts lists for the MICROLAB 500 systems. After you unpack your MICROLAB 500, check to see that you have received all parts before attempting to set up the system. The parts lists are presented in four separate tables with four corresponding figures.
Table 1–1 lists the programmable models in the MICROLAB 500 series, their components, and each component’s part number. Figure 1–1 shows these components.
Table 1–1 MICROLAB 500 System Descriptions
Model* Part # Drive
MICROLAB 510B 115V
MICROLAB 510B 220V
MICROLAB 530B 115V
MICROLAB 530B 220V
MICROLAB 540B 115V
MICROLAB 540B 220V
ML510115 35890 35893 69176 (English)
ML510220 35890 35893 same as above 355010 35792
ML530115 35891 35893 same as above 6541000 35793
ML530220 35891 35893 same as above 355010 35793
ML540115 35892 35893 same as above 6541000 35794
ML540220 35892 35893 same as above 355010 35794
Unit
Part #
Controller
Unit &
Cord
Part #
Manual Part # Power
69182 (French)
69180 (German)
69186 (Portuguese)
69188 (Spanish)
Shipping
Cord
6541000 35792
Kit
Part #
* All MICROLAB 500C models come with the drive unit, the manual, a power cord, and a valve. The MICROLAB 500 C does not come with a controller or a shipping kit.
1–2 MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 1–1 MICROLAB 500 System Components (items shown are
R
R
not to scale)
R
R
1
2
3
4
5
6
7
8
9
.
STEP
PRIME
POWER
Dual Drive Unit
Part # 35891 or 35892
(Microlab 530B/531C or 540B/541C)
0
Controller Unit and Cord
Part # 35893
(not with 500C models)
Stop
Run
R
R
PRIME
POWER
Single Drive Unit
Part # 35890
(Microlab 510B/511C)
Power Cord
Part # 6541000 or 355010
Shipping Kit Part # 35792 (Microlab 510B) Part # 35793 (Microlab 530B) Part # 35794 (Microlab 540B)
STEP
R
Manual
Part # 69176
Chapter 1 Getting Started 1–3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 1–2 lists the components that make up the MICROLAB 500B Shipping Kit (shown as a box in Figure 1–1). Figure 1–2 shows these components.
Table 1–2 MICROLAB 500B Shipping Kit
Shipping Kit*
Part #
MICROLAB
510B
35792
MICROLAB
530B
35793
MICROLAB
540B
35794
Valve Part # Hand Probe Part
#
HV Valve
35825
Concorde
Probe 35529
Diluter
Valve
35844
Dispenser
Valve
35842
Concorde
Probe 35529
Dual Hand
Probe 35767
Small Parts Kit
Part #
35888 35887
35888 35887
35888 35887
Tubing Kit
Part #
2 items
*The MICROLAB 500C is standard with only the valve. All other parts must be ordered separately.
1–4 MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 1–2 MICROLAB 500B Shipping Kit Components (items
R
shown are not to scale)
Small Parts Kit
Part # 35888
Concorde Probe
(Microlab 510B/511C or 530B/531C)
Part # 35529
Tubing Kit
Part # 35887
Dual Hand Probe
Part # 35767
(Microlab 540B/541C)
HV Valve
Part # 35825
(Microlab 510B/511C)
Dispenser Valve
Part # 35842
Diluter Valve Part # 35844
Chapter 1 Getting Started 1–5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 1–3 lists the components that make up the MICROLAB 500B Small Parts Kit (shown as a box in Figure 1–2). Figure 1–3 shows these components.
Table 1–3 Small Parts Kit #35888 (For All 500B Models)
Tubing Clips
Part #
230010
2 items
1 AMP Fuses
Part #
1524-01
2 items
Accessory
Holder
Part #
35783
1 item
Screws
Part #
16500
2 items
Figure 1–3 Small Parts Kit (items shown are not to scale)
(2) Tubing Clips
Part # 230010
Tubing Reducers
Part # 35770
Accessory
Holder
Part # 35783
Tubing
Reducers
Part #
35770
2 items
(2) Screws
Part # 16500
(2) Fuses
Part # 1524-01
1–6 MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 1–4 lists the components that make up the MICROLAB 500 Tubing Kit (shown as a box in Figure 1–2). Figure 1–4 shows these components.
Table 1–4 Tubing Kit #35887 (For All 500B Models)
12 ga. x 650 mm
Fill Tubing (not
tapered)
Part #
18 ga. x 650 mm
Fill Tubing (not
tapered)
Part #
12 ga. x 900 mm
Dispense Tubing
(tapered)
Part #
18 ga. x 900 mm
Dispense Tubing
(tapered)
Part #
240000* 240010* 240360* 240130*
* M6 threaded hubs are used on all tubing.
Figure 1–4 Tubing Kit (items shown are not to scale)
12 ga. Fill Tubing
Part # 240000
12 ga. Dispense Tubing
Part # 240360
18 ga. Fill Tubing
Part # 240010
18 ga. Dispense Tubing
Part # 240130
For complete lists of syringes, accessories, and replacement parts for the MICROLAB 500, see Appendix C, Ordering Parts and Accessories for the MICROLAB 500.
Chapter 1 Getting Started 1–7
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
A Brief Introduction to the MICROLAB 500B
The MICROLAB 510B, 530B, and 540B systems each consist of three basic units. These units include:
a drive unit
a controller unit
a hand probe This section briefly describes these units and the individual components that
comprise each unit. See Chapter 2, Installing the MICROLAB 500, for complete installation instructions; see Chapter 3, Using the MICROLAB 500, for complete usage instructions.
Drive Unit
The drive unit is the heart of each MICROLAB 500 system. The drive unit contains a precision drive motor, the syringe drive arms, the valve assembly, the power switches, and the connector receptacles. These features allow you to control other sub-assemblies, and together they create a very versatile and functional instrument.
Figure 1– 5 shows the front view of the MICROLAB 510B single syringe diluter/dispenser. In this figure, the controller unit rests on top of the drive unit and a syringe is attached to the syringe drive arm.
A Warning! This warning label appears on the front panel of the
ML500. It indicates that a pinch hazard exists when the syringe drive is moving.
1–8 MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual