Table of contents
4
Section 4Operation
For more information on System Setup (current outputs, relays, and
network interfaces), refer to the sc1500 User Manual.
The following tables show the sensor menu for the AN-ISE sc
probe, AISE sc probe and NISE sc probe. If a menu item does not
apply to all three probes, its applicability is explained in footnotes.
Footnote
AN-ISE sc probe is used to determine the ammonium and nitrate
concentration, as well as the potassium and chloride concentration.
Footnote
probe is used to determine the ammonium and potassium
concentration.
Footnote
probe is used to determine the nitrate concentration and chloride
concentration.
4.1Sensor diagnostics menu
SENSOR STATUS
AN-ISE sc or AISE sc or NISE sc
Error listShows all current error messages.
Warning listShows all current warnings.
English
1 indicates applicability to the AN-ISE sc probe. The
2 indicates applicability to the AISE sc probe. The AISE sc
3 indicates applicability to the NISE sc probe. The NISE sc
4.2Sensor setup menu
SELECT AN-ISE sc or AISE sc or NISE sc (if more than one sensor or analyzer is attached)
Take the grab sample
immediately for the lab
analysis.
1-point value correction
1
–N
NH
4
3
–N
1
NO
Enter the values for the
first point
The most recently used menu is displayed.
The currently active corrections are shown in Information.
1-point matrix correction for ammonium and nitrate
1-point matrix correction for ammonium
1-point matrix correction for nitrate
1-point matrix correction for ammonium and potassium
1-point matrix correction for nitrate and chloride
1-point matrix correction for ammonium, potassium, nitrate and chloride
Information window: When this window appears, the sample must be taken immediately
and then analyzed in the laboratory.
Perform 1-point value correction.
The most recently used menu is displayed.
The currently active corrections are shown in Information.
Select parameter for the 1-point value correction
Enter the values for the 1-point value correction
Note: The following example shows the input when using a AN-ISE sc probe for
ammonium. When using the AISE sc probe, the input is the same. When using the
NISE sc, it is only possible to input the nitrate and chloride values.
English5
Operation
4.2Sensor setup menu (Continued)
Calibration (continued)
AN-ISE SC NH
–NEnter the displayed ammonium value
4
AN-ISE SC KEnter the displayed potassium value
Lab value NH
–NEnter the laboratory ammonium value
4
Confirm the entry.Confirm the values entered
Correction resultDisplay the correction results
2-point value correctionPerform 2-point value correction
1
–N
NH
NO
4
3
–N
1
Select parameter for the 2-point value correction
Enter the values for the 2-point value correction (first point)
Enter the values for the
first point
Note: The following example shows the input when using a AN-ISE sc probe for
ammonium. When using the AISE sc probe, the input is the same. When using the
NISE sc, it is only possible to input the nitrate and chloride values.
AN-ISE SC NH
–NEnter the displayed ammonium value
4
AN-ISE SC KEnter the displayed potassium value
Lab value NH
–NEnter the laboratory ammonium value
4
Confirm the entry.Confirm the values entered
Enter the values for the 2-point value correction (second point).
Enter the values for the
second point.
Note: The following example shows the input when using a AN-ISE sc probe for
ammonium. When using the AISE sc probe, the input is the same. When using the
NISE sc, it is only possible to input the nitrate and chloride values.
AN-ISE SC NH
–NEnter the displayed ammonium value
4
AN-ISE SC KEnter the displayed potassium value
Lab value NH4–NEnter the laboratory ammonium value
Confirm the entry.Confirm the values entered
Correction resultDisplay the correction results
More correctionsOther matrix correction options
NoneNo More corrections is activated
2-point matrix correctionA 2-point matrix correction can be performed here
1
NH
4
1
NO
3
Concentration
measurement 1
Parameter selection for the 2-point matrix correction.
Saves the currently measured measurement for the first point
DateDisplays the date of the current correction of the first point
Concentration 1Entry and display of reference value for the first point
Concentration
measurement 2
Saves the currently measured measurement for the second point
DateDisplays the date of the current correction of the second point
Concentration 2Entry and display of the reference value for the second point
Previous correctionsSelection of one of the last corrections performed
Sensor codeThe sensor code can be activated or entered here
ActivateActivates the sensor code for the individual channels
1
+ K
NH
4
+ Cl
NO
3
+K + NO3+Cl
NH
4
1
Activate the sensor code for ammonium and potassium
Activate the sensor code for nitrate and chloride
1
Activate the sensor code for ammonium, potassium, nitrate and chloride
Reset calibrationActivates factory calibration
6English
4.2Sensor setup menu (Continued)
Calibration (continued)
Sensor codeEntry of the sensor code
Lab value correctionThe laboratory values of the last matrix correction can be changed
Enter the lab value.
Ammonium
Nitrate
Potassium
Chloride
1,2
1,3
1,2
1,3
Confirm the entry.Confirm the values entered
Correction resultDisplay the correction results
1,2
–N
NH
4
1,3
–N
NO
3
1,2
K+
1,3
Cl
InformationInformation on the matrix correction used per parameter
1,2
–N
NH
4
1,3
–N
NO
3
1,2
K+
1,3
Cl
Settings
Enter the laboratory values if 1-point matrix correction or 2-point matrix correction has
been selected
Entry of ammonium laboratory value
Enter the laboratory value for nitrate
Enter the laboratory value for potassium
Enter the laboratory value for chloride
Displays whether or not the ammonium correction was successful
Shows whether or not the nitrate correction was successful
Displays whether or not the potassium correction was successful
Displays whether or not the chloride correction was successful
Matrix correction used for ammonium
Matrix correction used for nitrate
Matrix correction used for potassium
Matrix correction used for chloride
Operation
Name
Enter or edit the name.
Up to 10 alphanumeric characters
UnitSelect either mg/L or ppm as the measurement unit
ParameterSelect NH
–N or NH4 and/or NO3–N or NO
4
3
TemperatureSelection of °C or °F as temperature unit
Offset temperature Enter a temperature offset
Response timeEntry of the response time (30 sec to 300 sec)
Datalog interval
Select the data log interval (OFF, 30 sec, 1 min, 2 min, 5 min, 10 min, 15 min and
30 min), 5 min is the factory setting
Select automatic potassium compensation:
On
K+ compensation
1,2
Off
0 = compensation OFF
0.1–2000 mg/L Cl = Fixed compensation value
K+ concentration
1,2
Only shown when K+ COMPENSATE is OFF
Select automatic chloride compensation:
On
Cl - compensation
1,3
Off
0 = compensation OFF
0.1–2000 mg/L Cl = Fixed compensation value
Cl - concentration
1,3
Only shown when Cl COMPNSATE is OFF
Reset configurationResets the configuration to the factory setting
Maintenance
Sensor informationInformation on the connected sensor
Sensor typeName of the connected sensor
English7
Operation
4.2Sensor setup menu (Continued)
Maintenance (continued)
NameSerial number or name of the measurement location
Serial numberSerial number of the connected sensor
Sensor typeInstrument designation of the connected sensor
SoftwareSoftware version
Calibration history
1,2
–N
NH
4
1,3
–N
NO
3
1,2
K+
1,3
Cl
SignalsSignals and measurement results of the individual measurement channels
Ammonium
Nitrate
Potassium
Chloride
1,2
1,3
1,2
1,3
Reference electrodeDisplays the signals and measurement results for the reference system
mV RAWDisplay the signals and measurement results for MV RAW
Impedance statusDisplay the signals and measurement results for impedance
TemperatureDisplay the signals and measurement results for temperature
MoistureDisplay the signals and measurement results for humidity
RFIDDisplay the signals and measurement results for RFID
Last calibrationShows the age of the last matrix correction
Ammonium
Nitrate
1,2
1,3
Service
Cartridge testPerform a sensor check with the test cartridge
Cartridge testDisplay whether the individual sensor channels are OK or not
MaintenanceDisplays whether or not DIAG/TEST is OK
Ground rodDisplays whether or not GNDROD is OK
Reference electrodeDisplays whether or not REF channel is OK
1,3
NO
3
1,2
NH
4
ORPDisplays whether or not ORP channel is OK
1,3
Cl-
1,2
K+
TemperatureDisplays whether or not temperature channel is OK
Cartridge replacementFollow the menu process
Cleaning...Follow the menu process
Data of the selected matrix correction and information about slope and offset of the
individual channels, for example
Matrix correction selected for ammonium
Matrix correction selected for nitrate
Matrix correction selected for potassium
Matrix correction selected for chloride
Display the signals and measurement results for ammonium
Shows the signals and measurement results for nitrate
Display the signals and measurement results for potassium
Display the signals and measurement results for chloride
Display the age of the last matrix correction for ammonium
Display the age of the last matrix correction for nitrate
Displays whether or not NO3 channel is OK
Displays whether or not NH4 channel is OK
Shows whether or not Cl channel is OK
Displays whether or not K channel is OK
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
8English
4.3Calibration/matrix correction
The four electrodes with the reference system of the compact
sensor cartridge were calibrated with one another at the factory
using special standard solutions (CARTICAL
membranes on the ion-selective electrodes are not 100% selective
due to other substances that may affect the measurement. Perform
a matrix correction (refer to 4.3.4, page 11) to compensate for other
ions present on the ISE electrodes.
Potassium has the largest interference effect on the ammonium
membrane, while chloride has the largest effect on the nitrate
membrane. The AN-ISE sc probe compensates for this problem
with the aid of a built-in potassium/chloride electrode.
When using the AISE sc probe, only the ammonium membrane and
the integrated potassium electrode are active.
When using the NISE sc probe, only the nitrate membrane and the
integrated chloride electrode are active.
Cross sensitivities between ammonium and potassium/nitrate are
automatically eliminated. Solids do not interfere with the
measurement. Due to matrix effects, correction and validation
cannot be performed with standard solutions. A matrix correction
can be carried out quickly and easily at any time.
Operation
TM
). However, the
4.3.1 Sensor code calibration
NOTICE
A matrix correction may only be performed if the sensor has been
immersed in the corresponding waste water matrix for over 12 hours.
This is the minimum time required to adapt the ISE membranes to the
waste water matrix.
The sensor code is a calibration code and is delivered with the
sensor cartridge certificate. It contains the factory calibration
described in section 3.12, page 9 for the sensor cartridge.
Instruments with automatic sensor code recognition
(LXG440.99.x000x) read this automatically and assume the
Cartrical calibration.
Instruments without automatic sensor code recognition
(LXG440.99.x001x) require the sensor code to be entered during
the initial setup and whenever a new sensor cartridge is activated. If
the sensor code certificate has been lost, carry out factory
calibration (under the sensor code menu) as a temporary solution.
After activating the code, the sensor is fully calibrated but not yet
adapted to the specific matrix of the relevant application on a waste
water treatment plant. At least 12 hours must elapse before a
matrix correction is performed to allow the cartridge to adapt to the
specific matrix.
English9
Operation
Proceed as follows to change the sensor code:
4.3.2 Matrix correction via LINK2SC
1. Select
CALIBRATION > MORE CORRECTIONS> SENSOR CODE >
ENTER
2. Enter the sensor code.
3. Press
meter for the cartridge is set to zero.
All old calibration data are now overwritten with the new calibration
data from the sensor code. The sensor code data is checked by the
system. If an error is indicated, check the sensor code and, if
necessary, enter the sensor code again.
The LINK2SC procedure offers a secure method of data exchange
between process probes and LINK2SC-compatible photometers
using an SD memory card or via a local area network (LAN). Two
different options are available:
a. The pure laboratory control measurement
b. A matrix correction that involves the measurement data
During a pure control measurement, the measurement data is
transferred from the probe to the photometer where it is then
archived together with the photometric reference data that has
been recorded.
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
ENTER to confirm and activate the sensor code. The day
generated in the laboratory being used to correct the probe
4.3.3 Matrix correction – manual
During a matrix correction, the reference data generated in the
laboratory is transferred to the probe to be used for the correction.
The matrix correction process requires operating steps to be
completed on the sc controller and on a LINK2SC-compatible
photometer.
Refer to the LINK2SC user manual for a detailed description of the
LINK2SC procedure.
When using the LINK2SC software, sections 4.3.3 and 4.3.4 are
not relevant.
ISE probes offer different options (refer to Table 1) for correcting
the sensor value with laboratory values (as a reference value).
The laboratory value of the water sample is entered as nitrate
nitrogen (NO
laboratory value replaces the prior value measured by the sensor.
–N) and/or as ammoniacal nitrogen (NH4–N). This
3
10English
Correction optionApplication
A 1-point matrix correction is the most commonly used correction option and performs a 1-point
matrix correction for ammonium and/or nitrate (4.3.4.1, page 11). It is advisable to perform a 1-point
matrix correction as the first correction. The 1-point matrix correction can be performed both with
1-point matrix
correction
1-point value
correction
2-point value
correction
2-point matrix
correction
Previous corrections
and without correction of the compensation electrodes (potassium or chloride); in most cases, it is
sufficient to perform it without correction. A correction featuring potassium and/or chloride is only
necessary if a high level of accuracy is required. With a 1-point matrix correction, a sample must be
taken when the correction is triggered and analyzed in the laboratory. The 1-point matrix correction is
activated when the laboratory value is entered.
1-point value correction (correction at one concentration point) corresponds to a 1-point matrix
correction with an alternative entry format.
Comparison values between the ISE probe and the laboratory can be collected over a period of
around a week with this correction. The correction can be performed at a later stage.
2-point value correction (correction at 2 different concentration points) should be performed if
dynamic concentration fluctuations are present over at least half a decadematrix correction or 1-point value correction does not achieve a sufficiently accurate result.
Comparison values between the ISE probe and the laboratory can be collected over a period of
around a week with this correction. The correction can be performed at a later stage.
The 2-point matrix correction corresponds to a 2-point value correction, but uses an alternative entry format and is recommended if there is a dynamic process with a large
nitrate/ammonium fluctuation greater than at least half a decade
sample must be taken for both points when the correction is triggered and analyzed in the laboratory.
The 2-point matrix correction is activated when the laboratory value is entered.
Return to one of the last matrix and value corrections performed if a correction has not produced
a successful result.
Table 1 Correction options for ISE probes
Operation
1
and a 2-point
1
. With a 2-point matrix correction, a
1
Examples of half a decade: The concentration of nitrogen nitrate shifts between 1 and 5 mg NO3–N or
between 5 and 25 mg/L NO3–N. (conc2 = (conc1 x 10)/2)
4.3.4 Performing the matrix correction
Note: Take laboratory value measurements or reference values promptly
or, alternatively, take these from the stabilized sample. This will prevent
changes in sample concentration, as time is a factor in comparative tests.
Refer to 7.3 Validation accessories, Page 39 for recommended
laboratory measurement tests.
4.3.4.1 1-point matrix correction
Proceed as follows to perform 1-point matrix correction:
Calibration
Matrix correction
More corrections
Information
1. Select
CALIBRATION > 1-point matrix correction
2. Select
press
3. Select the parameters you wish to correct and confirm by
pressing
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
2-point matrix correction from the selection window and
Take the grab sample
immediately for the lab
analysis.
Calibration
Matrix correction
More corrections
Enter the lab value.
Information
The sensor saves the current values of the selected parameters at
this point.
4. Take a water sample immediately from the closest point
possible to the sensor. Filter the sample as quickly as possible
and carry out a prompt laboratory analysis of the selected
parameters, as the measurement value can change quickly.
When the laboratory value has been determined, proceed as
follows:
5. Select SENSOR MENU > AN-ISE SC or AISE SC or NISE SC
> CALIBRATION > Enter the lab value.
6. The laboratory values for the parameters can only be entered if
the 1-point matrix correction correction has been selected
beforehand. Once the laboratory values have been entered,
select Confirm the entry. to confirm.
When the entered laboratory value is confirmed, the matrix
correction is activated.
7. Once the correction is activated, the result Correction result is
shown.
4.3.4.2 1-point value correction
Calibration
Matrix correction
More corrections
Information
Note: This process must always be carried out in full to make sure the
matrix correction is completed successfully.
If a correction does not produce a successful result, calculations are made
with the previous correction.
The 1-point value correction offers the option of retrospectively
performing a matrix correction at one point.
1. Take several samples with different concentrations on various
days, preferably within one week. Analyze the samples in the
laboratory. During the time the samples are being taken, the
sample temperature should vary a maximum of 5 °C, as
temperature changes are not taken into account in the value
correction.
2. Make a note of the two values measured in the samples and
displayed for the parameters to be corrected (ammonium and
potassium values, or nitrate and chloride values)
3. Also note the laboratory values measured for ammonium or
nitrate.
These three values form the correction point.
4. From the values taken, select a correction point that lies in the
middle of the expected concentration range.
12English
Operation
5. Go to the sensor menu and select CALIBRATION > MATRIX
CORRECTION >
pressing
ENTER.
1-point value correction and confirm by
Enter value point
AN–ISE SC NH
AN–ISE SC K
Lab value NH
Confirm the entry.
–N
4
–N
4
4.3.4.3 2-point value correction
Calibration
Matrix correction
More corrections
Information
6. Select the parameter1 (NH
–N or NO3–N) that requires
4
correction.
Note: The example opposite shows the NH4-N and K correction of the
AN-ISE sc probe.
7. Enter the three values for the sought correction point and push
Confirm the entry. to activate the correction.
Correction result is shown.
Note: If a correction does not produce a successful result, calculations are
made with the previous correction.
After successful completion of value correction, the corrected value is
shown as the display value for ammonium or nitrate the next time the menu
is opened.
The 2-point value correction makes it possible to perform a
subsequent 2-point correction to achieve higher accuracy for a
larger concentration range.
Note: The 2-point value correction and 2-point matrix correction are
comparable from a calculation perspective.
1. Take several samples on various days with different
concentrations, preferably within a week, and perform an
analysis of the samples in the laboratory. During the time the
samples are being taken, the sample temperature should be
around a maximum of 5 °C, as temperature changes are not
taken into account in the value correction.
1Applies to AN-ISE sc
Note: The 2-point matrix correction concentrations should be within a
range greater than half a decade. The following formula can assist in the
calculation of the half decade:
Conc2>=
Conc1 x 10
2
2. Make a note of the two values measured with the sensor in the
samples and displayed for the parameters to be corrected
(ammonium and potassium values, or nitrate and chloride
values).
3. Also note the laboratory value measured for ammonium or
nitrate.
All three values form one of the two correction points.
4. Look for two correction points where the laboratory values are
at least half a decade apart and display typical operating
conditions for the installation.
English13
Operation
5. Go to the sensor menu and select CALIBRATION > MATRIX
CORRECTION >
ENTER.
6. Select the parameter
2-point value correction and confirm with
1
(NH4–N or NO3–N) that requires
correction.
Note: When using the AN-ISE sc probe, you can only correct one
parameter at a time. If both parameters need to be corrected, the
procedure must be performed again.
Enter the values for the
first point
AN–ISE SC NH
AN–ISE SC K
Lab value NH
Confirm the entry.
Enter the values for the
second point
AN–ISE SC NH
AN–ISE SC K
Lab value NH
Confirm the entry.
–N
4
–N
4
–N
4
–N
4
4.3.4.4 2-point matrix correction
Ammonium
Concentration
measurement 1
Date
Enter the lab value 1.
Concentration
measurement 2
Date
Enter the lab value 2.
7. Enter the three values for the first correction point and push
CONFIRM THE ENTRY.
Note: The example opposite shows the NH4-N and K correction of the
AN-ISE sc probe.
8. To activate the correction, enter the three values for the second
correction point and push
CONFIRM THE ENTRY.
Correction result is shown.
Note: If a correction does not produce a successful result, calculations are
made with the previous correction. After successful completion of value
correction, the corrected value is shown as the display value for ammonium
or nitrate the next time the menu is opened.
Proceed as follows to perform 2-point matrix correction:
1. Select
CALIBRATION > MORE CORRECTIONS
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
.
2. Select 2-point matrix correction from the selection window
and press
ENTER.
3. Select the Parameter requiring a 2-point matrix correction.
4. Select the point to be corrected.
5.
SELECT Concentration measurement 1 or Concentration
measurement 2
6. Take a water sample from the closest point possible to the
sensor. Filter this sample promptly and perform an immediate
laboratory analysis of the selected parameters. The
measurement value can change very quickly:
When the laboratory value has been determined, proceed as
follows:
7. Select
CALIBRATE > MORE CORRECTIONS>
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
2-point matrix
correction
8. Select the parameters to be corrected with the laboratory value
entry:
14English
Operation
9. Enter the laboratory reference value and confirm.
The 2-point matrix correction is activated when the entry is
confirmed for both points.
English15
Section 5Maintenance
5.3Replace the sensor cartridge
The sensor cartridge is replaced as described below and in
Figure 15 on page 17.
1. Replace the cartridge using menu item
2. Clean the probe and thoroughly dry the sensor cartridge and
3. Loosen the 4 socket head screws.
The sensor cartridge must point downward such that no water can run into
the probe adapter. Pay attention to the contacts between the probe and
the sensor cartridge. These contacts must remain dry.
AN-ISE SC or AISE SC or NISE SC > MAINTENANCE > SERVICE >
CARTRIDGE REPLACEMENT
probe adapter.
.
NOTICE
4. Pull the sensor cartridge out of the probe adapter and discard
the old sensor cartridge as per the applicable regulations.
5. Make sure that a new black gasket is installed every time the
sensor cartridge is replaced. Before the gasket is installed,
clean the surface that faces the cartridge and the groove for the
gasket.
6. Insert the new sensor cartridge into the probe adapter. Observe
the marker hole on the sensor cartridge flange and the probe
adapter.
7. Secure the sensor cartridge with the 4 socket head screws.
8. The sensor code (calibration data) is automatically read for
instruments with automatic recognition (LXG440.99.x000x). For
instruments with non-automatic recognition
(LXG440.99.x001x), enter the new sensor code manually (refer
to certificate).
.
16English
Figure 15 Replace the sensor cartridge
Maintenance
1Socket head screw5Union nut
2Sensor cartridge6Marker hole
3O-ring7Socket head wrench
4Sensor
English17
Section 6Troubleshooting
6.1Error messages
If the sensor is in an error state, the measurement value for this
sensor will flash on the display and the relay contacts and current
outputs associated with this sensor will be stopped. The errors are
described in Table 2.
Table 2 Error messages
Displayed errorsCauseResolution
NH4 mV is out of range!
1,2
Ammonium mV value exceeds the measurement range
K+ mV is out of range!
1,2
NO
mV is out of range!
3
1,3
Cl- mV is out of range!
1,3
Reference electrode 1
mV is out of range!
Reference electrode 2
mV is out of range!
Temperature is out of
range!
No cartridge was found.No sensor cartridge connected
Sensor codeSensor code calibration failed
Potassium mV value exceeds the measurement range
Nitrate mV value exceeds the measurement range
Chloride mV value exceeds the measurement range
REF1 reference value is out of measuring range
ORP electrode mV value is out of measuring range
Temperature value exceeds measurement range
Refer to 6.3.1 Troubleshooting
during operation, Page 21.
Connect the sensor cartridge;
refer to section 3.3, page 15.
Refer to 6.3.2 Troubleshooting
during calibration, Page 22
HumidityHumidity in the probeInform service engineer
-N concentration is
NH
4
1,2
high.
NH4-N concentration is
1,2
low.
NO
-N concentration is
3
1,3
high.
-N concentration is
NO
3
1,3
low.
K+ concentration is high.
1,2
Ammonium concentration value exceeds measuring range
Ammonium concentration value is below measuring range
Nitrate concentration value exceeds measurement range
Nitrate concentration value is below measurement range
Potassium concentration value exceeds measuring range
Refer to 6.3.1 Troubleshooting
during operation, Page 21.
K+ concentration is low.
1,2
Cl- concentration is high.
1,3
Cl- concentration is low.
1,3
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
Potassium concentration value is below measuring range
Chloride concentration value exceeds measuring range
Chloride concentration value below measurement range
18English
Troubleshooting
6.2Warnings
In the event of a sensor warning, all menus, relays and outputs
continue to function as normal but a warning symbol lights up.
Warnings may be used to activate a relay; users can set warning
levels to define the severity. Warnings are defined in Table 3.
Table 3 Warnings
Displayed warningsCauseResolution
RFID dataCartridge faulty, read process failed
mV is out of range!
NH
1,2
4
Ammonium mV value is close to measuring range limit
Replace cartridge, check probe
with test cartridge
K+ mV is out of range!
1,2
NO3 mV is out of range!
1,3
Cl- mV is out of range!
1,3
Reference electrode 1
mV is out of range!
Reference electrode 2
mV is out of range!
Temperature is out of
range!
The cartridge has
expired.
-N concentration is
NH
4
1,2
high.
NH4-N concentration is
1,2
low.
-N concentration is
NO
3
1,3
high.
NO
-N concentration is
3
1,3
low.
K+ concentration is high.
1,2
Potassium mV value is close to measuring range limit
Nitrate mV value is close to measurement range limit
Chloride mV value is close to measurement range limit
Refer to 6.3.1 Troubleshooting
during operation, Page 21.
1st reference value is close to limit
2nd reference value is close to limit
Temperature is close to limit
Sensor cartridge more than 1 year oldReplace the sensor cartridge
Ammonium concentration value exceeds measuring range
Ammonium concentration value is below measuring range
Nitrate concentration value exceeds measuring range
Nitrate concentration value is below measurement range
Potassium concentration value exceeds measuring range
See 6.3.1 Troubleshooting during
operation, Page 21.
K+ concentration is low.
1,2
Cl- concentration is high.
1,3
Cl- concentration is low.
1,3
Potassium concentration value is below measuring range
Chloride concentration value exceeds measuring range
Chloride concentration value is below measurement range
English19
Troubleshooting
Table 3 Warnings
Ammonium
OffsetAmmonium offset exceeds measurement range
SlopeAmmonium slope exceeds measurement range
Potassium
OffsetPotassium offset exceeds measurement range
SlopePotassium slope is out of measuring range
Nitrate
OffsetNitrate offset is out of measurement range
SlopeNitrate slope is out of measurement range
Chloride
OffsetChloride offset is out of measurement range
SlopeChloride slope is out of measurement range
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
1,2
1,2
Refer to 6.3.2 Troubleshooting
during calibration, Page 22.
20English
6.3Troubleshooting
6.3.1 Troubleshooting during operation
SymptomPossible causeCorrective measures
Troubleshooting
Incorrect
measurement values
Incorrect
measurement values
Unstable
measurement values
Calibration too old; calibration was not suitable for
the particular application; large change in the waste
water matrix
Cl- mV is out of range!
Reference electrode 1 mV is out of range!
Reference electrode 2 mV is out of range!
Temperature is out of range!
The cartridge has expired.Replace the sensor cartridge
Moisture at the contacts of the sensor cartridge
Moisture inside the measurement probe/faulty
sensor electronics
Check the sensor electronics by using the test
cartridge (section 7.2, page 39).
1Select Sensor menu > Maintenance > Service >
Cartridge test> Cartridge test. Press Enter
2If all channels are confirmed with OK, the
sensor electronics are operational:
Test cartridge
Ok
Enter
Potassium concentrations too high (e.g.: >700 mg/L
in the case of small ammonium concentrations) or
chloride concentrations too high (e.g.: >1000 mg/L
in the case of small nitrate concentrations)
Air bubbles, depth of immersion
Moisture at the contacts of the sensor cartridge
Sensor membrane damaged
Reference element damaged
1,3
1,3
Perform a suitable calibration.
Refer to 4.6 Calibration/matrix correction,
Page 25
Clean the sensor cartridge using a brush
and/or rinse the sensor cartridge with clean
water (without cleaning agents), and wipe the
sensor cartridge carefully with a soft, clean
cloth.
Clean all components (membranes/reference
electrode/temperature sensor).
Install the cleaning unit
Increase the cleaning interval
Check the sensor installation/
replace the sensor cartridge
Replace the sensor cartridge
Replace the sensor cartridge/check the
waste water temperature
Dry the contact with a cloth or paper
Check the black gasket for damage and
make sure it is in the correct position.
Screw the 4 socket head screws tight.
If the test cartridge data is not within this
range and/or if the test cartridge check is not
successful, contact the service department.
Switch off potassium/chloride compensation
(in the configuration menu - then potentially
enter a fixed value for potassium/chloride)
Check the sensor installation
Check the cleaning unit configuration
Dry the contacts with a cloth or paper.
Check the black gasket for damage and
make sure it is in the correct position.
Screw the 4 socket head screws tight
Check the sensor installation/
replace the sensor cartridge
English21
Troubleshooting
1 applies to AN-ISE sc
3 applies to NISE sc
6.3.2 Troubleshooting during calibration
SymptomPossible causeCorrective measures
Sensor codeSensor code entered incorrectly
Ammonium
Offset
Slopw
1, 2
Error during the last ammonium correction, sensor
cartridge too old, contaminated, faulty
Potassium
Offset
Slope
Nitrate
Offset
Slope
1,3
Error during the last potassium correction, sensor
cartridge too old, contaminated, faulty
Error during the last nitrate correction, sensor
cartridge too old, contaminated, faulty
Chloride
Offset
Slope
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
Error during the last chloride correction, sensor
cartridge too old, contaminated, faulty
Using the certificate, check whether the
sensor code was entered correctly.
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
22English
Kapitel 4Betrieb
Weitere Informationen zur Systemeinrichtung (Stromausgänge,
Relais und Netzwerkschnittstellen) entnehmen Sie bitte dem
sc1500 Benutzerhandbuch.
In den folgenden Tabellen wird das Sensormenü gemeinsam für
die AN-ISE sc Sonde, die AISE sc Sonde und die NISE sc Sonde
dargestellt. Gilt ein Menüpunkt nicht für alle drei Sonden
gemeinsam, so wird die Gültigkeit über Fußnoten dargestellt.
Die Fußnote
der AN-ISE sc Sonde wird die Ammonium und Nitrat-Konzentration
sowie die Kalium und Chlorid-Konzentration bestimmt.
Die Fußnote
der AISE sc Sonde wird die Ammonium und Kalium-Konzentration
bestimmt.
Die Fußnote 3 zeigt eine Gültigkeit für die NISE sc Sonde an. Mit
der NISE sc Sonde wird die Nitrat-Konzentration und
Chlorid-Konzentration bestimmt.
4.1Menü Sensor-Diagnose
SENSOR STATUS
AN-ISE sc bzw. AISE sc oder NISE sc
FEHLERZeigt alle aktuellen Fehlermeldungen an.
WARNUNGENZeigt alle aktuellen Warnungen an.
Deutsch
1 zeigt eine Gültigkeit für die AN-ISE sc Sonde an. Mit
2 zeigt eine Gültigkeit für die AISE sc Sonde an. Mit
4.2Menü Sensoreinstellungen
Wählen Sie AN-ISE sc oder AISE sc oder NISE sc (falls mehr als ein Sensor angeschlossen ist)
Nehmen Sie unverzüglich
die Einzelmessung für die
Laboranalyse vor.
1-Punkt-Wertekorrektur
1
–N
NH
4
3
–N
1
NO
In der Anzeige erscheint das zuletzt benutzte Menü.
In Information sieht man die zurzeit aktiven Korrekturen.
1 Punkt Matrixkorrektur für Ammonium und Nitrat
1-Punkt-Matrixkorrektur für Ammonium
1 Punkt Matrixkorrektur für Nitrat
1 Punkt Matrixkorrektur für Ammonium und Kalium
1 Punkt Matrixkorrektur für Nitrat und Chlorid
1 Punkt Matrixkorrektur für Ammonium, Kalium, Nitrat und Chlorid
Hinweisfenster: Die Probe muss bei Erscheinen dieses Fensters sofort genommen und
im Labor analysiert werden.
Ausführung der 1-Punkt-Wertekorrektur.
In der Anzeige erscheint das zuletzt benutzte Menü.
In Information sieht man die zurzeit aktiven Korrekturen.
Auswahl des Parameters für die 1-Punkt-Wertekorrektur
Deutsch23
Betrieb
4.2Menü Sensoreinstellungen (Fortsetzung)
Kalibrierung (Fortsetzung)
Eingabe der Werte der 1-Punkt-Wertekorrektur
Eingabe der Werte des
ersten Punktes
AN-ISE SC NH4–NEingabe des angezeigten Ammonium-Wertes
AN-ISE SC KEingabe des angezeigten Kalium-Wertes
Laborwert NH
–NEingabe des Labor-Ammonium-Wertes
4
Bestätigung der
Eingabe.
KorrekturergebnisAnzeige der Korrekturergebnisse
2-Punkt-WertekorrekturAusführung der 2-Punkt-Wertekorrektur
1
–N
NH
4
3
–N
1
NO
Eingabe der Werte des
ersten Punktes
AN-ISE SC NH4–NEingabe des angezeigten Ammonium-Wertes
AN-ISE SC KEingabe des angezeigten Kalium-Wertes
Laborwert NH
–NEingabe des Labor-Ammonium-Wertes
4
Bestätigung der
Eingabe.
Eingabe der Werte des
zweiten Punktes.
AN-ISE SC NH4–NEingabe des angezeigten Ammonium-Wertes
AN-ISE SC KEingabe des angezeigten Kalium-Wertes
Laborwert NH
–NEingabe des Labor-Ammonium-Wertes
4
Bestätigung der
Eingabe.
KorrekturergebnisAnzeige der Korrekturergebnisse
Weitere KorrekturenWahlmöglichkeit für weitere Matrixkorrekturen
KeineWeitere Korrekturen ist nicht aktiviert
2-Punkt-MatrixkorrekturHier kann eine 2-Punkt-Matrixkorrektur durchgeführt werden
1
NH4
1
NO3
Konzentrationsmessung
1
DatumAnzeige des Datums der aktuellen Korrektur des ersten Punktes
Konzentration 1Eingabe und Anzeige des Referenzwertes des ersten Punktes
Konzentrationsmessung
2
DatumAnzeige des Datums der aktuellen Korrektur des zweiten Punktes
Konzentration 2Eingabe und Anzeige des Referenzwertes des zweiten Punktes
Vorherige KorrekturenAuswahl einer der letzten durchgeführten Korrekturen
SensorcodeHier kann der Sensorcode aktiviert bzw. eingegeben werden
Hinweis: Das folgende Beispiel zeigt die Eingabe bei einer AN-ISE sc Sonde für
Ammonium. Bei der AISE sc Sonde ist die Eingabe dieselbe. Bei der NISE sc Sonde ist
es nur möglich, die Nitrat- und Chloridwerte einzugeben.
Bestätigung der eingegebenen Werte
Auswahl des Parameters für die 2-Punkt-Wertekorrektur
Eingabe der Werte der 2-Punkt-Wertekorrektur des ersten Punktes
Hinweis: Das folgende Beispiel zeigt die Eingabe bei einer AN-ISE sc Sonde für
Ammonium. Bei der AISE sc Sonde ist die Eingabe dieselbe. Bei der NISE sc Sonde ist
es nur möglich, die Nitrat- und Chloridwerte einzugeben.
Bestätigung der eingegebenen Werte
Eingabe der Werte der 2-Punkt-Wertekorrektur des zweiten Punktes.
Hinweis: Das folgende Beispiel zeigt die Eingabe bei einer AN-ISE sc Sonde für
Ammonium. Bei der AISE sc Sonde ist die Eingabe dieselbe. Bei der NISE sc Sonde ist
es nur möglich, die Nitrat- und Chloridwerte einzugeben.
Bestätigung der eingegebenen Werte
Auswahl des Parameters für die 2-Punkt-Matrixkorrektur.
Speichern des aktuell gemessenen Messergebnisses für den ersten Punkt
Speichern des aktuell gemessenen Messergebnisses für den zweiten Punkt
24Deutsch
4.2Menü Sensoreinstellungen (Fortsetzung)
Kalibrierung (Fortsetzung)
AktivierenAktivierung des Sensorcodes für die einzelnen Kanäle
1,2
1,2
1,3
1
1,2
Aktivierung des Sensorcodes für Ammonium und Kalium
1
Aktivierung des Sensorcodes für Nitrat und Chlorid
1
Aktivierung des Sensorcodes für Ammonium, Kalium, Nitrat und Chlorid
Aktivierung der Werkskalibrierung
Eingabe der Laborwerte, wenn 1-Punkt-Matrixkorrektur oder 2-Punkt-Matrixkorrektur
ausgewählt wurde
Eingabe Ammonium-Laborwertes
Eingabe des Nitrat-Laborwertes
Eingabe des Kalium-Laborwertes
Eingabe des Chlorid-Laborwertes
Anzeige, ob Ammonium-Korrektur erfolgreich war
Anzeige, ob Nitrat-Korrektur erfolgreich war
Anzeige, ob Kalium-Korrektur erfolgreich war
Anzeige, ob Chlorid-Korrektur erfolgreich war
Angewendete Matrixkorrektur für Ammonium
Angewendete Matrixkorrektur für Nitrat
Angewendete Matrixkorrektur für Kalium
Angewendete Matrixkorrektur für Chlorid
Eingabe bzw. Bearbeitung des Namens.
Bis zu 10 alphanumerische Zeichen
Auswahl des Intervalls für die Protokollierung der Daten (AUS, 30 Sek., 1 Min., 2 Min.,
5 Min., 10 Min., 15 Min. und 30 Min.), 5 Min. ist die Werkseinstellung
Auswahl der automatischen Kalium-Kompensation:
Ein
Aus
0 = Kompensation aus
0,1–2000 mg/L Cl = Festwertkompensation
NH4 + K
NO3 + Cl
NH4+K + NO3+Cl
Kalibrierung
zurücksetzen
SensorcodeEingabe des Sensorcodes
LaborwertkorrekturLaborwerte der letzten Matrixkorrektur können verändert werden
Eingabe des Laborwertes.
Ammonium
1,3
Nitrat
1,2
Kalium
Chlorid
1,3
Bestätigung der Eingabe.Bestätigung der eingegebenen Werte
KorrekturergebnisAnzeige der Korrekturergebnisse
NH4–N
NO3–N
1,2
K+
1,3
Cl
InformationenInformation über die angewendete Matrixkorrektur pro Parameter
NH4–N
NO3–N
K+
Cl
1,2
1,3
1,2
1,3
Einstellungen
Name
EinheitAuswahl von mg/L oder ppm als Messeinheit
ParameterAuswahl von NH4–N oder NH4 und/oder NO3–N oder NO
TemperaturAuswahl von °C oder °F als Temperatureinheit
Offset-Temperatur Eingabe eines Temperatur-Offsets
ReaktionszeitEingabe der Reaktionszeit (30 Sek. bis 300 Sek.)
Intervall für die
Datenprotokollierung
K+-Kompensation
Betrieb
3
K+-Konzentration
1,2
Anzeige nur, wenn K+-KOMPENSATION auf AUS
Deutsch25
Betrieb
4.2Menü Sensoreinstellungen (Fortsetzung)
Wartung (Fortsetzung)
Auswahl der automatischen Chlorid-Kompensation:
Ein
1,3
1,2
1,2
1,3
1,2
1,3
1,3
Aus
0 = Kompensation aus
0,1–2000 mg/L Cl = Festwertkompensation
Anzeige nur, wenn Cl-KOMPENSATION auf AUS
Setzt die Konfiguration auf die Werkseinstellung zurück
Daten der ausgewählten Matrixkorrektur und Information über z. B. Steilheit und Offset
der einzelnen Kanäle
Ausgewählte Matrixkorrektur für Ammonium
Ausgewählte Matrixkorrektur für Nitrat
Ausgewählte Matrixkorrektur für Kalium
Ausgewählte Matrixkorrektur für Chlorid
Anzeige der Signale und Messergebnisse für Ammonium
Anzeige der Signale und Messergebnisse für Nitrat
Anzeige der Signale und Messergebnisse für Kalium
Anzeige der Signale und Messergebnisse für Chlorid
Anzeige des Alters der letzten Matrixkorrektur von Ammonium
Anzeige des Alters der letzten Matrixkorrektur von Nitrat
Anzeige ob NO3-Kanal OK ist
Anzeige ob NH4-Kanal OK ist
Cl-Kompensation
Cl-Konzentration
Zurücksetzen auf
Werkseinstellung
Wartung
SensorinformationInformation über den angeschlossenen Sensor
SensortypName des angeschlossenen Sensors
NameSeriennummer oder Name des Messortes
SeriennummerSeriennummer des angeschlossenen Sensors
SensortypGerätebezeichnung des angeschlossenen Sensors
SoftwareSoftwareversion
Zurückliegende Kalibrierdaten
NH4–N
1,2
NO3–N
1,2
K+
1,3
Cl
SignaleSignale und Messergebnisse der einzelnen Messkanäle
Ammonium
1,3
Nitrat
1,2
Kalium
Chlorid
1,3
ReferenzelektrodeAnzeige der Signale und Messergebnisse des Referenzsystems
mV ROHAnzeige der Signale und Messergebnisse der mV Roh
ImpedanzstatusAnzeige der Signale und Messergebnisse der Impedanzen
TemperaturAnzeige der Signale und Messergebnisse der Temperatur
FeuchteAnzeige der Signale und Messergebnisse der Feuchte
RFIDAnzeige der Signale und Messergebnisse des RFID
Letzte KalibrierungAnzeige des Alters der letzten Matrixkorrektur
Ammonium
1,3
Nitrat
Wartung
PatronentestDurchführung einer Sensorüberprüfung mit Hilfe der Testkartusche
PatronentestAnzeige ob die einzelnen Kanäle des Sensors OK sind
WartungAnzeige ob DIAG/TEST OK ist
ErdungsstabAnzeige ob GNDROD (ERDSTB) OK ist
ReferenzelektrodeAnzeige ob REF-Kanal OK ist
NO3
NH4
ORPAnzeige ob ORP-Kanal OK ist
26Deutsch
4.2Menü Sensoreinstellungen (Fortsetzung)
Wartung (Fortsetzung)
1,3
Cl-
1,2
K+
TemperaturAnzeige ob Temperatur-Kanal OK ist
PatronenaustauschFolgen Sie dem menügeführten Prozess
Reinigung...Folgen Sie dem menügeführten Prozess
1
gilt für AN-ISE sc
2
gilt für AISE sc
3
gilt für NISE sc
Anzeige ob CL-Kanal OK ist
Anzeige ob K-Kanal OK ist
4.3Kalibrierung/Matrixkorrektur
Die 4 Elektroden mit Referenzsystem der kompakten
Sensorkartusche wurden mit speziellen Standardlösungen
werkseitig aufeinander kalibriert (CARTICAL
der ionenselektiven Elektroden sind aber aufgrund anderer Stoffe,
die die Messung beeinträchtigen können, nicht zu 100
Führen Sie eine Matrixkorrektur durch (siehe
Einflüsse anderer Ionen auf die ISE Elektroden zu kompensieren.
Betrieb
TM
). Die Membranen
% selektiv.
4.3.4, Seite 29), um
Den größten Störeinfluss auf die Ammoniummembran übt Kalium
aus, bei der Nitratmembran ist es Chlorid. Die AN-ISE
sc Sonde
gleicht dieses Problem mit Hilfe einer integrierten Kalium- bzw.
Chloridelektrode ab.
Bei der AISE sc Sonde sind nur die Ammoniummembran und die
integrierte Kaliumelektrode aktiv.
Bei der NISE sc Sonde sind nur die Nitratmembran und die
integrierte Chloridelektrode aktiv.
Querempfindlichkeiten zwischen Ammonium und Kalium bzw.
Nitrat und Chlorid werden automatisch eliminiert. Die Messung wird
nicht durch Feststoffe gestört. Aufgrund von Matrixeffekten kann
die Korrektur und Validierung nicht mit Standardlösungen erfolgen.
Es lässt sich jederzeit auf einfache und schnelle Weise eine
Matrixkorrektur durchführen.
HINWEIS
Eine Matrixkorrektur darf erst durchgeführt werden, wenn der Sensor
mehr als 12 Stunden in die entsprechende Abwassermatrix eingetaucht
war. 12 Stunden ist die mindestens erforderliche Zeit, um die ISE
Membranen an die Abwassermatrix anzupassen.
4.3.1 Sensorcode-Kalibrierung
Der Sensorcode ist ein Kalibier-Code und wird mit dem Zertifikat
der Sensorkartusche ausgeliefert. Er enthält die unter Kapitel 3.12,
Seite 9 beschriebene Werkskalibrierung für die Sensorkartusche.
Deutsch27
Betrieb
Bei Geräten mit automatischer Sensorcodeerkennung
(LXG440.99.x000x) wird dieser automatisch eingelesen und die
Cartrical-Kalibrierung übernommen.
Bei Geräten ohne automatische Sensorcodeerkennung
(LXG440.99.x001x) muss bei Aktivierung einer neuen
Sensorkartusche, sowie während der Erstinbetriebnahme, der
Sensorcode eingegeben werden. Bei Verlust des
Sensorcodezertifikats bitte als vorübergehende Lösung
Werkskalibrierung (unter Sensorcode Menü) durchführen.
Nach Aktivierung des Codes ist der Sensor vollständig kalibriert,
aber noch nicht auf die spezielle Matrix der jeweiligen Applikation
auf einer Kläranlage angepasst. Führen Sie nach frühestens 12
Stunden eine Matrixkorrektur durch, um die Kartusche an die
spezielle Matrix anzupassen.
Gehen Sie wie folgt vor, um den Sensorcode zu ändern:
4.3.2 Matrixkorrektur über LINK2SC
1. Wählen Sie
NISE SC \> KALIBRIERUNG\> WEITERE KORREKTUREN\>
SENSORCODE \> EINGABE
2. Geben Sie den Sensorcode ein.
3. Bestätigen Sie mit
Der Tageszähler für die Cartridge wird auf Null gesetzt.
Alle alten Kalibrierdaten werden mit den neuen Kalibrierdaten aus
dem Sensorcode überschrieben. Das System prüft die Daten des
Sensorcodes. Falls ein Fehler angezeigt wird, überprüfen Sie den
Sensorcode und geben Sie ihn ggf. erneut ein.
Das LINK2SC-Verfahren bietet einen sicheren Datenaustausch
zwischen Prozess-Sonden und LINK2SC-fähigen Photometern mit
Hilfe einer SD-Speicherkarte oder über ein lokales Netzwerk (LAN).
Es wird zwischen zwei Möglichkeiten unterschieden:
a. der reinen Labor-Kontrollmessung und
b. einer Matrix-Korrektur, bei der die im Labor erzeugten
Messdaten zur Korrektur der Sonde eingesetzt werden
können.
Bei einer reinen Kontrollmessung werden die Messdaten der
Sonde zum Photometer übertragen und dort zusammen mit den
photometrisch ermittelten Referenzdaten archiviert.
SENSORMENÜ \> AN-ISE SC oder AISE SC oder
EINGABE, um den Sensorcode zu aktivieren.
Bei einer Matrix-Korrektur werden die im Labor erzeugten
Referenzdaten an die Sonde übermittelt und dort für die Korrektur
verwendet.
Der Ablauf einer Matrix-Korrektur umfasst Bedienschritte am sc
Controller und an einem LINK2SC-fähigen Photometer.
Eine genaue Beschreibung des LINK2SC-Verfahrens entnehmen
sie bitte der LINK2SC Bedienungsanleitung.
Bei Nutzung der Software LINK2SC sind die Kapitel 4.3.3 und 4.3.4
überflüssig.
28Deutsch
4.3.3 Matrixkorrektur – Manuell
Tabelle 4 Korrekturmöglichkeiten für die ISE Sonden
Korrekturmöglichkeit Anwendung
Eine 1-Punkt-Matrixkorrektur ist die am häufigsten angewendete Korrekturmöglichkeit und führt
eine 1-Punkt-Matrixkorrektur für Ammonium und/oder Nitrat durch (4.3.4.1, Seite 29). Es wird
empfohlen, eine 1-Punkt-Matrixkorrektur als erste Korrektur durchzuführen. Die
1-Punkt-Matrixkorrek
tur
1-Punkt-Wertekorrek
tur
2-Punkt-Wertekorrek
tur
2-Punkt-Matrixkorrek
tur
1-Punkt-Matrixkorrektur kann mit oder ohne Korrektur der Kompensationselektroden (Kalium oder
Chlorid) durchgeführt werden, wobei ohne in den meisten Fällen ausreichen wird. Eine Korrektur mit
Kalium und/oder Chlorid braucht nur durchgeführt zu werden, wenn eine höhere Genauigkeit
erforderlich ist. Bei einer 1-Punkt-Matrixkorrektur muss mit Auslösen der Korrektur eine Probe
genommen und im Labor analysiert werden. Mit der Eingabe der Laborwerte wird die
1-Punkt-Matrixkorrektur aktiv.
Die 1-Punkt-Wertekorrektur (Korrektur an einem Konzentrationspunkt) entspricht einer
1-Punkt-Wertekorrektur mit einer alternativen Eingabeform.
Bei dieser Korrektur können Vergleichswerte zwischen ISE Sonde und Labor über einen Zeitraum
von etwa einer Woche gesammelt werden. Zu einem späteren Zeitpunkt kann dann die Korrektur
durchgeführt werden.
Die 2-Punkt-Wertekorrektur (Korrektur an zwei verschiedenen Konzentrationspunkten) sollte
durchgeführt werden, wenn dynamische Konzentrationsfluktuationen über mindestens
½ Dekade
1-Punkt-Wertekorrektur kein genügend genaues Ergebnis erzielt.
Bei dieser Korrektur können Vergleichswerte zwischen ISE Sonde und Labor über einen Zeitraum
von etwa einer Woche gesammelt werden. Zu einem späteren Zeitpunkt kann dann die Korrektur
durchgeführt werden.
Die 2-Punkt-Matrixkorrektur entspricht einer 2-Punkt-Wertekorrektur, verwendet aber eine alternative Eingabeform und wird empfohlen, wenn ein dynamischer Prozess mit einer großen
Nitrat/Ammonium-Schwankung über mindestens ½ Dekade
2-Punkt-Matrixkorrektur muss mit Auslösen der Korrektur für beide Punkte eine Probe genommen
und im Labor ausgewertet werden. Mit der Eingabe der Laborwerte wird die 2-Punkt-Matrixkorrektur
aktiv.
1
und einer 2-Punkt-Matrixkorrektur vorhanden sind oder eine
Betrieb
Bei den ISE Sonden gibt es verschiedene Möglichkeiten (siehe
Tabelle 4), den Sensorwert mit Laborwerten (als Referenzwert) zu
korrigieren.
Der Laborwert der Wasserprobe wird als Nitratstickstoff (NO3–N)
und/oder als Ammoniumstickstoff (NH
Laborwert ersetzt den zuvor vom Sensor gemessenen Wert.
1
vorliegt. Bei einer
–N) eingegeben. Dieser
4
Vorherige
Korrekturen
1
Beispiele einer ½ Dekade: Die Nitratstickstoffkonzentration bewegt sich zwischen 1 und 5 mg NO3–N oder
zwischen 5 und 25 mg/L NO
Rückkehr zu einer der letzten durchgeführten Matrix- und Wertekorrekturen, falls eine Korrektur nicht zu einem Erfolg geführt hat.
–N. (Konz2 = (Konz1 x 10)/2)
3
4.3.4 Durchführung der Matrixkorrektur
Hinweis: Führen Sie die Messungen der Laborwerte bzw. Referenzwerte
mit den Küvetten-Tests zeitnah oder aus der stabilisierten Probe durch. So
verhindern Sie Änderungen der Probenkonzentration, da
Vergleichsuntersuchungen Zeit empfindlich sind.
Unter 7.3 Validierungszubehör, Seite 41 finden Sie empfohlene
Tests für Labor Messungen.
4.3.4.1 1-Punkt-Matrixkorrektur
Gehen Sie wie folgt vor, um eine 1-Punkt-Matrixkorrektur
durchzuführen:
Deutsch29
Betrieb
Kalibrierung
Matrixkorrektur
Weitere Korrekturen
Informationen
1-Punkt-Matrixkorrekt
ur
Nehmen Sie
unverzüglich die
Einzelmessung für die
Laboranalyse vor.
1. Wählen Sie SENSORMENÜ \> AN-ISE SC oder AISE SC oder
Der Sensor hat zu diesem Zeitpunkt die aktuellen Werte der
ausgewählten Parameter gespeichert.
4. Entnehmen Sie sofort und so nah wie möglich am Sensor eine
Wasserprobe. Filtrieren Sie diese schnellst möglich und führen
Sie zeitnah eine Laboranalyse der ausgewählten Parameter
durch, da sich der Messwert schnell ändern kann.
Kalibrierung
Matrixkorrektur
Weitere Korrekturen
Eingabe des
Laborwertes.
Informationen
4.3.4.2 1-Punkt-Wertekorrektur
Kalibrierung
Matrixkorrektur
Weitere Korrekturen
Informationen
Wenn Sie den Laborwert bestimmt haben, gehen Sie wie folgt vor:
5. Wählen Sie SENSORMENÜ \> AN-ISE SC oder AISE SC oder
NISE SC \> KALIBRIERUNG \> Eingabe des Laborwertes.
6. Sie können nur die Laborwerte für die Parameter eingeben,
wenn vorher die 1-Punkt-Matrixkorrektur ausgewählt wurde.
Wählen Sie nach der Eingabe der Werte Bestätigung der Eingabe. zum Bestätigen aus.
Wenn Sie den eingegebenen Laborwert bestätigen, ist die
Matrixkorrektur aktiviert.
7. Nach der Aktivierung der Korrektur wird das Ergebnis
Korrekturergebnis angezeigt.
Hinweis: Dieser Prozess muss immer vollständig durchlaufen werden, um
eine Matrixkorrektur erfolgreich abzuschließen.
Sollte eine Korrektur nicht zu einem erfolgreichen Ergebnis führen, wird mit
der vorherigen Korrektur weitergerechnet.
Die 1-Punkt-Wertekorrektur bietet die Möglichkeit, nachträglich
eine Matrixkorrektur an einem Punkt durchzuführen.
1. Nehmen Sie mehrere Proben an verschiedenen Tagen,
möglichst innerhalb von einer Woche mit unterschiedlichen
Konzentrationen. Führen Sie eine Analyse der Proben im Labor
durch. Während des Probenahmezeitraumes sollte die
30Deutsch
Loading...