Hach AN-ISE sc, NISE sc, AISE sc Addendum To The Operator Manual

DOC023.98.90594

AN-ISE sc, AISE sc, NISE sc probe

Addendum—Operation with the SC1500 Controller
05/2017, Edition 1

Table of contents

English..................................................................................................................................................... 5
Deutsch.................................................................................................................................................. 23
Español.................................................................................................................................................. 41
Français ................................................................................................................................................. 59
Italiano ................................................................................................................................................... 79
Nederlands ............................................................................................................................................ 97
3
Table of contents
4

Section 4 Operation

For more information on System Setup (current outputs, relays, and network interfaces), refer to the sc1500 User Manual.
The following tables show the sensor menu for the AN-ISE sc probe, AISE sc probe and NISE sc probe. If a menu item does not apply to all three probes, its applicability is explained in footnotes.
Footnote AN-ISE sc probe is used to determine the ammonium and nitrate concentration, as well as the potassium and chloride concentration.
Footnote probe is used to determine the ammonium and potassium concentration.
Footnote probe is used to determine the nitrate concentration and chloride concentration.

4.1 Sensor diagnostics menu

SENSOR STATUS
AN-ISE sc or AISE sc or NISE sc
Error list Shows all current error messages. Warning list Shows all current warnings.
English
1 indicates applicability to the AN-ISE sc probe. The
2 indicates applicability to the AISE sc probe. The AISE sc
3 indicates applicability to the NISE sc probe. The NISE sc

4.2 Sensor setup menu

SELECT AN-ISE sc or AISE sc or NISE sc (if more than one sensor or analyzer is attached)
Calibration
Matrix correction options.
Matrix correction
None No Matrix corrections activated 1-point matrix correction 1-point matrix correction
+ NO3
NH
4
NH
4
NO
3
+ K
NH
4
+ Cl
NO
3
+K + NO3+Cl
NH
4
1
1,2
1,3
1,2
1,3
1
Take the grab sample immediately for the lab analysis.
1-point value correction
1
–N
NH
4
3
–N
1
NO
Enter the values for the first point
The most recently used menu is displayed. The currently active corrections are shown in Information.
1-point matrix correction for ammonium and nitrate 1-point matrix correction for ammonium 1-point matrix correction for nitrate 1-point matrix correction for ammonium and potassium 1-point matrix correction for nitrate and chloride 1-point matrix correction for ammonium, potassium, nitrate and chloride
Information window: When this window appears, the sample must be taken immediately and then analyzed in the laboratory.
Perform 1-point value correction. The most recently used menu is displayed.
The currently active corrections are shown in Information.
Select parameter for the 1-point value correction
Enter the values for the 1-point value correction
Note: The following example shows the input when using a AN-ISE sc probe for ammonium. When using the AISE sc probe, the input is the same. When using the NISE sc, it is only possible to input the nitrate and chloride values.
English 5
Operation
4.2 Sensor setup menu (Continued)
Calibration (continued)
AN-ISE SC NH
–N Enter the displayed ammonium value
4
AN-ISE SC K Enter the displayed potassium value Lab value NH
–N Enter the laboratory ammonium value
4
Confirm the entry. Confirm the values entered
Correction result Display the correction results
2-point value correction Perform 2-point value correction
1
–N
NH NO
4
3
–N
1
Select parameter for the 2-point value correction
Enter the values for the 2-point value correction (first point)
Enter the values for the first point
Note: The following example shows the input when using a AN-ISE sc probe for ammonium. When using the AISE sc probe, the input is the same. When using the NISE sc, it is only possible to input the nitrate and chloride values.
AN-ISE SC NH
–N Enter the displayed ammonium value
4
AN-ISE SC K Enter the displayed potassium value Lab value NH
–N Enter the laboratory ammonium value
4
Confirm the entry. Confirm the values entered
Enter the values for the 2-point value correction (second point).
Enter the values for the second point.
Note: The following example shows the input when using a AN-ISE sc probe for ammonium. When using the AISE sc probe, the input is the same. When using the NISE sc, it is only possible to input the nitrate and chloride values.
AN-ISE SC NH
–N Enter the displayed ammonium value
4
AN-ISE SC K Enter the displayed potassium value Lab value NH4–N Enter the laboratory ammonium value Confirm the entry. Confirm the values entered
Correction result Display the correction results
More corrections Other matrix correction options
None No More corrections is activated 2-point matrix correction A 2-point matrix correction can be performed here
1
NH
4
1
NO
3
Concentration measurement 1
Parameter selection for the 2-point matrix correction.
Saves the currently measured measurement for the first point
Date Displays the date of the current correction of the first point Concentration 1 Entry and display of reference value for the first point Concentration
measurement 2
Saves the currently measured measurement for the second point
Date Displays the date of the current correction of the second point
Concentration 2 Entry and display of the reference value for the second point Previous corrections Selection of one of the last corrections performed Sensor code The sensor code can be activated or entered here
Activate Activates the sensor code for the individual channels
1
+ K
NH
4
+ Cl
NO
3
+K + NO3+Cl
NH
4
1
Activate the sensor code for ammonium and potassium Activate the sensor code for nitrate and chloride
1
Activate the sensor code for ammonium, potassium, nitrate and chloride
Reset calibration Activates factory calibration
6 English
4.2 Sensor setup menu (Continued)
Calibration (continued)
Sensor code Entry of the sensor code
Lab value correction The laboratory values of the last matrix correction can be changed
Enter the lab value.
Ammonium Nitrate Potassium Chloride
1,2
1,3
1,2
1,3
Confirm the entry. Confirm the values entered
Correction result Display the correction results
1,2
–N
NH
4
1,3
–N
NO
3
1,2
K+
1,3
Cl
Information Information on the matrix correction used per parameter
1,2
–N
NH
4
1,3
–N
NO
3
1,2
K+
1,3
Cl
Settings
Enter the laboratory values if 1-point matrix correction or 2-point matrix correction has been selected
Entry of ammonium laboratory value Enter the laboratory value for nitrate Enter the laboratory value for potassium Enter the laboratory value for chloride
Displays whether or not the ammonium correction was successful Shows whether or not the nitrate correction was successful Displays whether or not the potassium correction was successful Displays whether or not the chloride correction was successful
Matrix correction used for ammonium Matrix correction used for nitrate Matrix correction used for potassium Matrix correction used for chloride
Operation
Name
Enter or edit the name.
Up to 10 alphanumeric characters Unit Select either mg/L or ppm as the measurement unit Parameter Select NH
–N or NH4 and/or NO3–N or NO
4
3
Temperature Selection of °C or °F as temperature unit Offset temperature Enter a temperature offset Response time Entry of the response time (30 sec to 300 sec)
Datalog interval
Select the data log interval (OFF, 30 sec, 1 min, 2 min, 5 min, 10 min, 15 min and
30 min), 5 min is the factory setting
Select automatic potassium compensation:
On
K+ compensation
1,2
Off
0 = compensation OFF
0.1–2000 mg/L Cl = Fixed compensation value
K+ concentration
1,2
Only shown when K+ COMPENSATE is OFF
Select automatic chloride compensation:
On
Cl - compensation
1,3
Off
0 = compensation OFF
0.1–2000 mg/L Cl = Fixed compensation value
Cl - concentration
1,3
Only shown when Cl COMPNSATE is OFF Reset configuration Resets the configuration to the factory setting
Maintenance
Sensor information Information on the connected sensor
Sensor type Name of the connected sensor
English 7
Operation
4.2 Sensor setup menu (Continued)
Maintenance (continued)
Name Serial number or name of the measurement location Serial number Serial number of the connected sensor Sensor type Instrument designation of the connected sensor Software Software version
Calibration history
1,2
–N
NH
4
1,3
–N
NO
3
1,2
K+
1,3
Cl
Signals Signals and measurement results of the individual measurement channels
Ammonium Nitrate Potassium Chloride
1,2
1,3
1,2
1,3
Reference electrode Displays the signals and measurement results for the reference system mV RAW Display the signals and measurement results for MV RAW Impedance status Display the signals and measurement results for impedance Temperature Display the signals and measurement results for temperature Moisture Display the signals and measurement results for humidity RFID Display the signals and measurement results for RFID
Last calibration Shows the age of the last matrix correction
Ammonium Nitrate
1,2
1,3
Service
Cartridge test Perform a sensor check with the test cartridge
Cartridge test Display whether the individual sensor channels are OK or not
Maintenance Displays whether or not DIAG/TEST is OK Ground rod Displays whether or not GNDROD is OK Reference electrode Displays whether or not REF channel is OK
1,3
NO
3
1,2
NH
4
ORP Displays whether or not ORP channel is OK
1,3
Cl-
1,2
K+
Temperature Displays whether or not temperature channel is OK Cartridge replacement Follow the menu process Cleaning... Follow the menu process
Data of the selected matrix correction and information about slope and offset of the individual channels, for example
Matrix correction selected for ammonium Matrix correction selected for nitrate Matrix correction selected for potassium Matrix correction selected for chloride
Display the signals and measurement results for ammonium Shows the signals and measurement results for nitrate Display the signals and measurement results for potassium Display the signals and measurement results for chloride
Display the age of the last matrix correction for ammonium Display the age of the last matrix correction for nitrate
Displays whether or not NO3 channel is OK Displays whether or not NH4 channel is OK
Shows whether or not Cl channel is OK Displays whether or not K channel is OK
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
8 English

4.3 Calibration/matrix correction

The four electrodes with the reference system of the compact sensor cartridge were calibrated with one another at the factory using special standard solutions (CARTICAL membranes on the ion-selective electrodes are not 100% selective due to other substances that may affect the measurement. Perform a matrix correction (refer to 4.3.4, page 11) to compensate for other ions present on the ISE electrodes.
Potassium has the largest interference effect on the ammonium membrane, while chloride has the largest effect on the nitrate membrane. The AN-ISE sc probe compensates for this problem with the aid of a built-in potassium/chloride electrode.
When using the AISE sc probe, only the ammonium membrane and the integrated potassium electrode are active.
When using the NISE sc probe, only the nitrate membrane and the integrated chloride electrode are active.
Cross sensitivities between ammonium and potassium/nitrate are automatically eliminated. Solids do not interfere with the measurement. Due to matrix effects, correction and validation cannot be performed with standard solutions. A matrix correction can be carried out quickly and easily at any time.
Operation
TM
). However, the

4.3.1 Sensor code calibration

NOTICE
A matrix correction may only be performed if the sensor has been immersed in the corresponding waste water matrix for over 12 hours. This is the minimum time required to adapt the ISE membranes to the waste water matrix.
The sensor code is a calibration code and is delivered with the sensor cartridge certificate. It contains the factory calibration described in section 3.12, page 9 for the sensor cartridge.
Instruments with automatic sensor code recognition (LXG440.99.x000x) read this automatically and assume the Cartrical calibration.
Instruments without automatic sensor code recognition (LXG440.99.x001x) require the sensor code to be entered during the initial setup and whenever a new sensor cartridge is activated. If the sensor code certificate has been lost, carry out factory calibration (under the sensor code menu) as a temporary solution.
After activating the code, the sensor is fully calibrated but not yet adapted to the specific matrix of the relevant application on a waste water treatment plant. At least 12 hours must elapse before a matrix correction is performed to allow the cartridge to adapt to the specific matrix.
English 9
Operation
Proceed as follows to change the sensor code:

4.3.2 Matrix correction via LINK2SC

1. Select
CALIBRATION > MORE CORRECTIONS> SENSOR CODE > ENTER
2. Enter the sensor code.
3. Press
meter for the cartridge is set to zero.
All old calibration data are now overwritten with the new calibration data from the sensor code. The sensor code data is checked by the system. If an error is indicated, check the sensor code and, if necessary, enter the sensor code again.
The LINK2SC procedure offers a secure method of data exchange between process probes and LINK2SC-compatible photometers using an SD memory card or via a local area network (LAN). Two different options are available:
a. The pure laboratory control measurement
b. A matrix correction that involves the measurement data
During a pure control measurement, the measurement data is transferred from the probe to the photometer where it is then archived together with the photometric reference data that has been recorded.
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
ENTER to confirm and activate the sensor code. The day
generated in the laboratory being used to correct the probe

4.3.3 Matrix correction – manual

During a matrix correction, the reference data generated in the laboratory is transferred to the probe to be used for the correction.
The matrix correction process requires operating steps to be completed on the sc controller and on a LINK2SC-compatible photometer.
Refer to the LINK2SC user manual for a detailed description of the LINK2SC procedure.
When using the LINK2SC software, sections 4.3.3 and 4.3.4 are not relevant.
ISE probes offer different options (refer to Table 1) for correcting the sensor value with laboratory values (as a reference value).
The laboratory value of the water sample is entered as nitrate nitrogen (NO laboratory value replaces the prior value measured by the sensor.
–N) and/or as ammoniacal nitrogen (NH4–N). This
3
10 English
Correction option Application
A 1-point matrix correction is the most commonly used correction option and performs a 1-point matrix correction for ammonium and/or nitrate (4.3.4.1, page 11). It is advisable to perform a 1-point matrix correction as the first correction. The 1-point matrix correction can be performed both with
1-point matrix correction
1-point value correction
2-point value correction
2-point matrix correction
Previous corrections
and without correction of the compensation electrodes (potassium or chloride); in most cases, it is sufficient to perform it without correction. A correction featuring potassium and/or chloride is only necessary if a high level of accuracy is required. With a 1-point matrix correction, a sample must be taken when the correction is triggered and analyzed in the laboratory. The 1-point matrix correction is activated when the laboratory value is entered.
1-point value correction (correction at one concentration point) corresponds to a 1-point matrix correction with an alternative entry format.
Comparison values between the ISE probe and the laboratory can be collected over a period of around a week with this correction. The correction can be performed at a later stage.
2-point value correction (correction at 2 different concentration points) should be performed if dynamic concentration fluctuations are present over at least half a decade matrix correction or 1-point value correction does not achieve a sufficiently accurate result.
Comparison values between the ISE probe and the laboratory can be collected over a period of around a week with this correction. The correction can be performed at a later stage.
The 2-point matrix correction corresponds to a 2-point value correction, but uses an alternative entry format and is recommended if there is a dynamic process with a large nitrate/ammonium fluctuation greater than at least half a decade sample must be taken for both points when the correction is triggered and analyzed in the laboratory. The 2-point matrix correction is activated when the laboratory value is entered.
Return to one of the last matrix and value corrections performed if a correction has not produced
a successful result.
Table 1 Correction options for ISE probes
Operation
1
and a 2-point
1
. With a 2-point matrix correction, a
1
Examples of half a decade: The concentration of nitrogen nitrate shifts between 1 and 5 mg NO3–N or
between 5 and 25 mg/L NO3–N. (conc2 = (conc1 x 10)/2)

4.3.4 Performing the matrix correction

Note: Take laboratory value measurements or reference values promptly or, alternatively, take these from the stabilized sample. This will prevent changes in sample concentration, as time is a factor in comparative tests.
Refer to 7.3 Validation accessories, Page 39 for recommended laboratory measurement tests.
4.3.4.1 1-point matrix correction
Proceed as follows to perform 1-point matrix correction:
Calibration Matrix correction More corrections Information
1. Select
CALIBRATION > 1-point matrix correction
2. Select
press
3. Select the parameters you wish to correct and confirm by
pressing
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
2-point matrix correction from the selection window and
ENTER.
ENTER.
.
Selection options for AN-ISE sc: NH4 + NO3; NH4; NO3; NH4 + K; NO3 + Cl; NH4 + K NO3 + Cl
Selection options for AISE sc: NH4; NH4 + K
English 11
Operation
Selection options for NISE sc: NO3; NO3 + Cl
1-point matrix correction
Take the grab sample immediately for the lab analysis.
Calibration Matrix correction More corrections Enter the lab value. Information
The sensor saves the current values of the selected parameters at this point.
4. Take a water sample immediately from the closest point
possible to the sensor. Filter the sample as quickly as possible and carry out a prompt laboratory analysis of the selected parameters, as the measurement value can change quickly.
When the laboratory value has been determined, proceed as follows:
5. Select SENSOR MENU > AN-ISE SC or AISE SC or NISE SC > CALIBRATION > Enter the lab value.
6. The laboratory values for the parameters can only be entered if
the 1-point matrix correction correction has been selected beforehand. Once the laboratory values have been entered, select Confirm the entry. to confirm.
When the entered laboratory value is confirmed, the matrix correction is activated.
7. Once the correction is activated, the result Correction result is shown.
4.3.4.2 1-point value correction
Calibration Matrix correction More corrections Information
Note: This process must always be carried out in full to make sure the matrix correction is completed successfully. If a correction does not produce a successful result, calculations are made with the previous correction.
The 1-point value correction offers the option of retrospectively performing a matrix correction at one point.
1. Take several samples with different concentrations on various days, preferably within one week. Analyze the samples in the laboratory. During the time the samples are being taken, the sample temperature should vary a maximum of 5 °C, as temperature changes are not taken into account in the value correction.
2. Make a note of the two values measured in the samples and displayed for the parameters to be corrected (ammonium and potassium values, or nitrate and chloride values)
3. Also note the laboratory values measured for ammonium or nitrate.
These three values form the correction point.
4. From the values taken, select a correction point that lies in the middle of the expected concentration range.
12 English
Operation
5. Go to the sensor menu and select CALIBRATION > MATRIX
CORRECTION >
pressing
ENTER.
1-point value correction and confirm by
Enter value point AN–ISE SC NH AN–ISE SC K Lab value NH Confirm the entry.
–N
4
–N
4
4.3.4.3 2-point value correction
Calibration Matrix correction More corrections Information
6. Select the parameter1 (NH
–N or NO3–N) that requires
4
correction.
Note: The example opposite shows the NH4-N and K correction of the AN-ISE sc probe.
7. Enter the three values for the sought correction point and push Confirm the entry. to activate the correction.
Correction result is shown.
Note: If a correction does not produce a successful result, calculations are made with the previous correction. After successful completion of value correction, the corrected value is shown as the display value for ammonium or nitrate the next time the menu is opened.
The 2-point value correction makes it possible to perform a subsequent 2-point correction to achieve higher accuracy for a larger concentration range.
Note: The 2-point value correction and 2-point matrix correction are comparable from a calculation perspective.
1. Take several samples on various days with different concentrations, preferably within a week, and perform an analysis of the samples in the laboratory. During the time the samples are being taken, the sample temperature should be around a maximum of 5 °C, as temperature changes are not taken into account in the value correction.
1Applies to AN-ISE sc
Note: The 2-point matrix correction concentrations should be within a range greater than half a decade. The following formula can assist in the
calculation of the half decade:
Conc2 >=
Conc1 x 10
2
2. Make a note of the two values measured with the sensor in the samples and displayed for the parameters to be corrected (ammonium and potassium values, or nitrate and chloride values).
3. Also note the laboratory value measured for ammonium or nitrate.
All three values form one of the two correction points.
4. Look for two correction points where the laboratory values are at least half a decade apart and display typical operating conditions for the installation.
English 13
Operation
5. Go to the sensor menu and select CALIBRATION > MATRIX
CORRECTION > ENTER.
6. Select the parameter
2-point value correction and confirm with
1
(NH4–N or NO3–N) that requires
correction.
Note: When using the AN-ISE sc probe, you can only correct one parameter at a time. If both parameters need to be corrected, the procedure must be performed again.
Enter the values for the first point
AN–ISE SC NH AN–ISE SC K Lab value NH Confirm the entry.
Enter the values for the second point
AN–ISE SC NH AN–ISE SC K Lab value NH Confirm the entry.
–N
4
–N
4
–N
4
–N
4
4.3.4.4 2-point matrix correction
Ammonium Concentration
measurement 1 Date Enter the lab value 1. Concentration
measurement 2 Date Enter the lab value 2.
7. Enter the three values for the first correction point and push
CONFIRM THE ENTRY.
Note: The example opposite shows the NH4-N and K correction of the AN-ISE sc probe.
8. To activate the correction, enter the three values for the second
correction point and push
CONFIRM THE ENTRY.
Correction result is shown.
Note: If a correction does not produce a successful result, calculations are made with the previous correction. After successful completion of value correction, the corrected value is shown as the display value for ammonium or nitrate the next time the menu is opened.
Proceed as follows to perform 2-point matrix correction:
1. Select
CALIBRATION > MORE CORRECTIONS
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
.
2. Select 2-point matrix correction from the selection window
and press
ENTER.
3. Select the Parameter requiring a 2-point matrix correction.
4. Select the point to be corrected.
5.
SELECT Concentration measurement 1 or Concentration measurement 2
6. Take a water sample from the closest point possible to the
sensor. Filter this sample promptly and perform an immediate laboratory analysis of the selected parameters. The measurement value can change very quickly:
When the laboratory value has been determined, proceed as follows:
7. Select
CALIBRATE > MORE CORRECTIONS>
SENSOR MENU > AN-ISE SC or AISE SC or NISE SC >
2-point matrix
correction
8. Select the parameters to be corrected with the laboratory value
entry:
14 English
Operation
9. Enter the laboratory reference value and confirm.
The 2-point matrix correction is activated when the entry is confirmed for both points.
English 15

Section 5 Maintenance

5.3 Replace the sensor cartridge

The sensor cartridge is replaced as described below and in
Figure 15 on page 17.
1. Replace the cartridge using menu item
2. Clean the probe and thoroughly dry the sensor cartridge and
3. Loosen the 4 socket head screws.
The sensor cartridge must point downward such that no water can run into the probe adapter. Pay attention to the contacts between the probe and the sensor cartridge. These contacts must remain dry.
AN-ISE SC or AISE SC or NISE SC > MAINTENANCE > SERVICE > CARTRIDGE REPLACEMENT
probe adapter.
.
NOTICE
4. Pull the sensor cartridge out of the probe adapter and discard
the old sensor cartridge as per the applicable regulations.
5. Make sure that a new black gasket is installed every time the
sensor cartridge is replaced. Before the gasket is installed, clean the surface that faces the cartridge and the groove for the gasket.
6. Insert the new sensor cartridge into the probe adapter. Observe
the marker hole on the sensor cartridge flange and the probe adapter.
7. Secure the sensor cartridge with the 4 socket head screws.
8. The sensor code (calibration data) is automatically read for
instruments with automatic recognition (LXG440.99.x000x). For instruments with non-automatic recognition (LXG440.99.x001x), enter the new sensor code manually (refer to certificate).
.
16 English
Figure 15 Replace the sensor cartridge
Maintenance
1 Socket head screw 5 Union nut 2 Sensor cartridge 6 Marker hole 3 O-ring 7 Socket head wrench 4 Sensor
English 17

Section 6 Troubleshooting

6.1 Error messages

If the sensor is in an error state, the measurement value for this sensor will flash on the display and the relay contacts and current outputs associated with this sensor will be stopped. The errors are described in Table 2.
Table 2 Error messages
Displayed errors Cause Resolution
NH4 mV is out of range!
1,2
Ammonium mV value exceeds the measurement range
K+ mV is out of range!
1,2
NO
mV is out of range!
3
1,3
Cl- mV is out of range!
1,3
Reference electrode 1 mV is out of range!
Reference electrode 2 mV is out of range!
Temperature is out of range!
No cartridge was found. No sensor cartridge connected
Sensor code Sensor code calibration failed
Potassium mV value exceeds the measurement range
Nitrate mV value exceeds the measurement range
Chloride mV value exceeds the measurement range
REF1 reference value is out of measuring range
ORP electrode mV value is out of measuring range
Temperature value exceeds measurement range
Refer to 6.3.1 Troubleshooting
during operation, Page 21.
Connect the sensor cartridge; refer to section 3.3, page 15.
Refer to 6.3.2 Troubleshooting
during calibration, Page 22
Humidity Humidity in the probe Inform service engineer
-N concentration is
NH
4
1,2
high. NH4-N concentration is
1,2
low. NO
-N concentration is
3
1,3
high.
-N concentration is
NO
3
1,3
low. K+ concentration is high.
1,2
Ammonium concentration value exceeds measuring range
Ammonium concentration value is below measuring range
Nitrate concentration value exceeds measurement range
Nitrate concentration value is below measurement range
Potassium concentration value exceeds measuring range
Refer to 6.3.1 Troubleshooting
during operation, Page 21.
K+ concentration is low.
1,2
Cl- concentration is high.
1,3
Cl- concentration is low.
1,3
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
Potassium concentration value is below measuring range
Chloride concentration value exceeds measuring range
Chloride concentration value below measurement range
18 English
Troubleshooting

6.2 Warnings

In the event of a sensor warning, all menus, relays and outputs
continue to function as normal but a warning symbol lights up.
Warnings may be used to activate a relay; users can set warning levels to define the severity. Warnings are defined in Table 3.
Table 3 Warnings
Displayed warnings Cause Resolution
RFID data Cartridge faulty, read process failed
mV is out of range!
NH
1,2
4
Ammonium mV value is close to measuring range limit
Replace cartridge, check probe with test cartridge
K+ mV is out of range!
1,2
NO3 mV is out of range!
1,3
Cl- mV is out of range!
1,3
Reference electrode 1 mV is out of range!
Reference electrode 2 mV is out of range!
Temperature is out of range!
The cartridge has expired.
-N concentration is
NH
4
1,2
high. NH4-N concentration is
1,2
low.
-N concentration is
NO
3
1,3
high. NO
-N concentration is
3
1,3
low. K+ concentration is high.
1,2
Potassium mV value is close to measuring range limit
Nitrate mV value is close to measurement range limit
Chloride mV value is close to measurement range limit
Refer to 6.3.1 Troubleshooting
during operation, Page 21.
1st reference value is close to limit
2nd reference value is close to limit
Temperature is close to limit
Sensor cartridge more than 1 year old Replace the sensor cartridge
Ammonium concentration value exceeds measuring range
Ammonium concentration value is below measuring range
Nitrate concentration value exceeds measuring range
Nitrate concentration value is below measurement range
Potassium concentration value exceeds measuring range
See 6.3.1 Troubleshooting during
operation, Page 21.
K+ concentration is low.
1,2
Cl- concentration is high.
1,3
Cl- concentration is low.
1,3
Potassium concentration value is below measuring range
Chloride concentration value exceeds measuring range
Chloride concentration value is below measurement range
English 19
Troubleshooting
Table 3 Warnings
Ammonium Offset Ammonium offset exceeds measurement range Slope Ammonium slope exceeds measurement range Potassium Offset Potassium offset exceeds measurement range Slope Potassium slope is out of measuring range Nitrate Offset Nitrate offset is out of measurement range Slope Nitrate slope is out of measurement range Chloride Offset Chloride offset is out of measurement range Slope Chloride slope is out of measurement range
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
1,2
1,2
Refer to 6.3.2 Troubleshooting
during calibration, Page 22.
20 English

6.3 Troubleshooting

6.3.1 Troubleshooting during operation

Symptom Possible cause Corrective measures
Troubleshooting
Incorrect measurement values
Incorrect measurement values
Unstable measurement values
Calibration too old; calibration was not suitable for the particular application; large change in the waste water matrix
Severely contaminated membranes and/or reference electrode
Sensor membrane damaged Reference element damaged
mV is out of range!
NO
3
Cl- mV is out of range! Reference electrode 1 mV is out of range! Reference electrode 2 mV is out of range!
Temperature is out of range!
The cartridge has expired. Replace the sensor cartridge
Moisture at the contacts of the sensor cartridge
Moisture inside the measurement probe/faulty sensor electronics
Check the sensor electronics by using the test cartridge (section 7.2, page 39).
1 Select Sensor menu > Maintenance > Service >
Cartridge test> Cartridge test. Press Enter
2 If all channels are confirmed with OK, the
sensor electronics are operational: Test cartridge Ok Enter
Potassium concentrations too high (e.g.: >700 mg/L in the case of small ammonium concentrations) or chloride concentrations too high (e.g.: >1000 mg/L in the case of small nitrate concentrations)
Air bubbles, depth of immersion
Moisture at the contacts of the sensor cartridge
Sensor membrane damaged Reference element damaged
1,3
1,3
Perform a suitable calibration. Refer to 4.6 Calibration/matrix correction,
Page 25
Clean the sensor cartridge using a brush and/or rinse the sensor cartridge with clean water (without cleaning agents), and wipe the sensor cartridge carefully with a soft, clean cloth.
Clean all components (membranes/reference electrode/temperature sensor).
Install the cleaning unit Increase the cleaning interval
Check the sensor installation/ replace the sensor cartridge
Replace the sensor cartridge
Replace the sensor cartridge/check the waste water temperature
Dry the contact with a cloth or paper Check the black gasket for damage and
make sure it is in the correct position. Screw the 4 socket head screws tight.
If the test cartridge data is not within this range and/or if the test cartridge check is not successful, contact the service department.
Switch off potassium/chloride compensation (in the configuration menu - then potentially enter a fixed value for potassium/chloride)
Check the sensor installation Check the cleaning unit configuration
Dry the contacts with a cloth or paper. Check the black gasket for damage and
make sure it is in the correct position. Screw the 4 socket head screws tight
Check the sensor installation/ replace the sensor cartridge
English 21
Troubleshooting
1 applies to AN-ISE sc 3 applies to NISE sc

6.3.2 Troubleshooting during calibration

Symptom Possible cause Corrective measures
Sensor code Sensor code entered incorrectly
Ammonium Offset
Slopw
1, 2
Error during the last ammonium correction, sensor cartridge too old, contaminated, faulty
Potassium Offset
Slope
Nitrate Offset
Slope
1,3
Error during the last potassium correction, sensor cartridge too old, contaminated, faulty
Error during the last nitrate correction, sensor cartridge too old, contaminated, faulty
Chloride Offset
Slope
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc
Error during the last chloride correction, sensor cartridge too old, contaminated, faulty
Using the certificate, check whether the sensor code was entered correctly.
Repeat the correction. Use the previous correction. Clean or replace the sensor cartridge.
Repeat the correction. Use the previous correction. Clean or replace the sensor cartridge.
Repeat the correction. Use the previous correction. Clean or replace the sensor cartridge.
Repeat the correction. Use the previous correction. Clean or replace the sensor cartridge.
22 English

Kapitel 4 Betrieb

Weitere Informationen zur Systemeinrichtung (Stromausgänge, Relais und Netzwerkschnittstellen) entnehmen Sie bitte dem sc1500 Benutzerhandbuch.
In den folgenden Tabellen wird das Sensormenü gemeinsam für die AN-ISE sc Sonde, die AISE sc Sonde und die NISE sc Sonde dargestellt. Gilt ein Menüpunkt nicht für alle drei Sonden gemeinsam, so wird die Gültigkeit über Fußnoten dargestellt.
Die Fußnote der AN-ISE sc Sonde wird die Ammonium und Nitrat-Konzentration sowie die Kalium und Chlorid-Konzentration bestimmt.
Die Fußnote der AISE sc Sonde wird die Ammonium und Kalium-Konzentration bestimmt.
Die Fußnote 3 zeigt eine Gültigkeit für die NISE sc Sonde an. Mit der NISE sc Sonde wird die Nitrat-Konzentration und Chlorid-Konzentration bestimmt.

4.1 Menü Sensor-Diagnose

SENSOR STATUS
AN-ISE sc bzw. AISE sc oder NISE sc
FEHLER Zeigt alle aktuellen Fehlermeldungen an. WARNUNGEN Zeigt alle aktuellen Warnungen an.
Deutsch
1 zeigt eine Gültigkeit für die AN-ISE sc Sonde an. Mit
2 zeigt eine Gültigkeit für die AISE sc Sonde an. Mit

4.2 Menü Sensoreinstellungen

Wählen Sie AN-ISE sc oder AISE sc oder NISE sc (falls mehr als ein Sensor angeschlossen ist)
Kalibrierung
Auswahlmöglichkeit für Matrixkorrektur.
Matrixkorrektur
Keine Keine Matrixkorrekturen aktiviert 1-Punkt-Matrixkorrektur 1-Punkt-Matrixkorrektur
+ NO3
NH
4
NH
4
NO
3
+ K
NH
4
+ Cl
NO
3
+K + NO3+Cl
NH
4
1
1,2
1,3
1,2
1,3
1
Nehmen Sie unverzüglich die Einzelmessung für die Laboranalyse vor.
1-Punkt-Wertekorrektur
1
–N
NH
4
3
–N
1
NO
In der Anzeige erscheint das zuletzt benutzte Menü. In Information sieht man die zurzeit aktiven Korrekturen.
1 Punkt Matrixkorrektur für Ammonium und Nitrat 1-Punkt-Matrixkorrektur für Ammonium 1 Punkt Matrixkorrektur für Nitrat 1 Punkt Matrixkorrektur für Ammonium und Kalium 1 Punkt Matrixkorrektur für Nitrat und Chlorid 1 Punkt Matrixkorrektur für Ammonium, Kalium, Nitrat und Chlorid
Hinweisfenster: Die Probe muss bei Erscheinen dieses Fensters sofort genommen und im Labor analysiert werden.
Ausführung der 1-Punkt-Wertekorrektur. In der Anzeige erscheint das zuletzt benutzte Menü.
In Information sieht man die zurzeit aktiven Korrekturen.
Auswahl des Parameters für die 1-Punkt-Wertekorrektur
Deutsch 23
Betrieb
4.2 Menü Sensoreinstellungen (Fortsetzung)
Kalibrierung (Fortsetzung)
Eingabe der Werte der 1-Punkt-Wertekorrektur
Eingabe der Werte des ersten Punktes
AN-ISE SC NH4–N Eingabe des angezeigten Ammonium-Wertes AN-ISE SC K Eingabe des angezeigten Kalium-Wertes Laborwert NH
–N Eingabe des Labor-Ammonium-Wertes
4
Bestätigung der Eingabe.
Korrekturergebnis Anzeige der Korrekturergebnisse
2-Punkt-Wertekorrektur Ausführung der 2-Punkt-Wertekorrektur
1
–N
NH
4
3
–N
1
NO
Eingabe der Werte des ersten Punktes
AN-ISE SC NH4–N Eingabe des angezeigten Ammonium-Wertes AN-ISE SC K Eingabe des angezeigten Kalium-Wertes Laborwert NH
–N Eingabe des Labor-Ammonium-Wertes
4
Bestätigung der Eingabe.
Eingabe der Werte des zweiten Punktes.
AN-ISE SC NH4–N Eingabe des angezeigten Ammonium-Wertes AN-ISE SC K Eingabe des angezeigten Kalium-Wertes Laborwert NH
–N Eingabe des Labor-Ammonium-Wertes
4
Bestätigung der Eingabe.
Korrekturergebnis Anzeige der Korrekturergebnisse
Weitere Korrekturen Wahlmöglichkeit für weitere Matrixkorrekturen
Keine Weitere Korrekturen ist nicht aktiviert 2-Punkt-Matrixkorrektur Hier kann eine 2-Punkt-Matrixkorrektur durchgeführt werden
1
NH4
1
NO3
Konzentrationsmessung 1
Datum Anzeige des Datums der aktuellen Korrektur des ersten Punktes Konzentration 1 Eingabe und Anzeige des Referenzwertes des ersten Punktes Konzentrationsmessung
2 Datum Anzeige des Datums der aktuellen Korrektur des zweiten Punktes
Konzentration 2 Eingabe und Anzeige des Referenzwertes des zweiten Punktes Vorherige Korrekturen Auswahl einer der letzten durchgeführten Korrekturen Sensorcode Hier kann der Sensorcode aktiviert bzw. eingegeben werden
Hinweis: Das folgende Beispiel zeigt die Eingabe bei einer AN-ISE sc Sonde für Ammonium. Bei der AISE sc Sonde ist die Eingabe dieselbe. Bei der NISE sc Sonde ist es nur möglich, die Nitrat- und Chloridwerte einzugeben.
Bestätigung der eingegebenen Werte
Auswahl des Parameters für die 2-Punkt-Wertekorrektur
Eingabe der Werte der 2-Punkt-Wertekorrektur des ersten Punktes
Hinweis: Das folgende Beispiel zeigt die Eingabe bei einer AN-ISE sc Sonde für Ammonium. Bei der AISE sc Sonde ist die Eingabe dieselbe. Bei der NISE sc Sonde ist es nur möglich, die Nitrat- und Chloridwerte einzugeben.
Bestätigung der eingegebenen Werte
Eingabe der Werte der 2-Punkt-Wertekorrektur des zweiten Punktes.
Hinweis: Das folgende Beispiel zeigt die Eingabe bei einer AN-ISE sc Sonde für Ammonium. Bei der AISE sc Sonde ist die Eingabe dieselbe. Bei der NISE sc Sonde ist es nur möglich, die Nitrat- und Chloridwerte einzugeben.
Bestätigung der eingegebenen Werte
Auswahl des Parameters für die 2-Punkt-Matrixkorrektur.
Speichern des aktuell gemessenen Messergebnisses für den ersten Punkt
Speichern des aktuell gemessenen Messergebnisses für den zweiten Punkt
24 Deutsch
4.2 Menü Sensoreinstellungen (Fortsetzung)
Kalibrierung (Fortsetzung)
Aktivieren Aktivierung des Sensorcodes für die einzelnen Kanäle
1,2
1,2
1,3
1
1,2
Aktivierung des Sensorcodes für Ammonium und Kalium
1
Aktivierung des Sensorcodes für Nitrat und Chlorid
1
Aktivierung des Sensorcodes für Ammonium, Kalium, Nitrat und Chlorid
Aktivierung der Werkskalibrierung
Eingabe der Laborwerte, wenn 1-Punkt-Matrixkorrektur oder 2-Punkt-Matrixkorrektur ausgewählt wurde
Eingabe Ammonium-Laborwertes Eingabe des Nitrat-Laborwertes Eingabe des Kalium-Laborwertes Eingabe des Chlorid-Laborwertes
Anzeige, ob Ammonium-Korrektur erfolgreich war Anzeige, ob Nitrat-Korrektur erfolgreich war Anzeige, ob Kalium-Korrektur erfolgreich war Anzeige, ob Chlorid-Korrektur erfolgreich war
Angewendete Matrixkorrektur für Ammonium Angewendete Matrixkorrektur für Nitrat Angewendete Matrixkorrektur für Kalium Angewendete Matrixkorrektur für Chlorid
Eingabe bzw. Bearbeitung des Namens. Bis zu 10 alphanumerische Zeichen
Auswahl des Intervalls für die Protokollierung der Daten (AUS, 30 Sek., 1 Min., 2 Min., 5 Min., 10 Min., 15 Min. und 30 Min.), 5 Min. ist die Werkseinstellung
Auswahl der automatischen Kalium-Kompensation:
Ein Aus
0 = Kompensation aus 0,1–2000 mg/L Cl = Festwertkompensation
NH4 + K NO3 + Cl NH4+K + NO3+Cl Kalibrierung
zurücksetzen
Sensorcode Eingabe des Sensorcodes
Laborwertkorrektur Laborwerte der letzten Matrixkorrektur können verändert werden
Eingabe des Laborwertes.
Ammonium
1,3
Nitrat
1,2
Kalium Chlorid
1,3
Bestätigung der Eingabe. Bestätigung der eingegebenen Werte
Korrekturergebnis Anzeige der Korrekturergebnisse
NH4–N NO3–N
1,2
K+
1,3
Cl
Informationen Information über die angewendete Matrixkorrektur pro Parameter
NH4–N NO3–N K+ Cl
1,2
1,3
1,2
1,3
Einstellungen
Name
Einheit Auswahl von mg/L oder ppm als Messeinheit Parameter Auswahl von NH4–N oder NH4 und/oder NO3–N oder NO Temperatur Auswahl von °C oder °F als Temperatureinheit Offset-Temperatur Eingabe eines Temperatur-Offsets Reaktionszeit Eingabe der Reaktionszeit (30 Sek. bis 300 Sek.) Intervall für die
Datenprotokollierung
K+-Kompensation
Betrieb
3
K+-Konzentration
1,2
Anzeige nur, wenn K+-KOMPENSATION auf AUS
Deutsch 25
Betrieb
4.2 Menü Sensoreinstellungen (Fortsetzung)
Wartung (Fortsetzung)
Auswahl der automatischen Chlorid-Kompensation:
Ein
1,3
1,2
1,2
1,3
1,2
1,3
1,3
Aus 0 = Kompensation aus 0,1–2000 mg/L Cl = Festwertkompensation
Anzeige nur, wenn Cl-KOMPENSATION auf AUS
Setzt die Konfiguration auf die Werkseinstellung zurück
Daten der ausgewählten Matrixkorrektur und Information über z. B. Steilheit und Offset der einzelnen Kanäle
Ausgewählte Matrixkorrektur für Ammonium Ausgewählte Matrixkorrektur für Nitrat Ausgewählte Matrixkorrektur für Kalium Ausgewählte Matrixkorrektur für Chlorid
Anzeige der Signale und Messergebnisse für Ammonium Anzeige der Signale und Messergebnisse für Nitrat Anzeige der Signale und Messergebnisse für Kalium Anzeige der Signale und Messergebnisse für Chlorid
Anzeige des Alters der letzten Matrixkorrektur von Ammonium Anzeige des Alters der letzten Matrixkorrektur von Nitrat
Anzeige ob NO3-Kanal OK ist Anzeige ob NH4-Kanal OK ist
Cl-Kompensation
Cl-Konzentration Zurücksetzen auf
Werkseinstellung
Wartung
Sensorinformation Information über den angeschlossenen Sensor
Sensortyp Name des angeschlossenen Sensors Name Seriennummer oder Name des Messortes Seriennummer Seriennummer des angeschlossenen Sensors Sensortyp Gerätebezeichnung des angeschlossenen Sensors Software Softwareversion
Zurückliegende Kalibrierdaten
NH4–N
1,2
NO3–N
1,2
K+
1,3
Cl
Signale Signale und Messergebnisse der einzelnen Messkanäle
Ammonium
1,3
Nitrat
1,2
Kalium Chlorid
1,3
Referenzelektrode Anzeige der Signale und Messergebnisse des Referenzsystems mV ROH Anzeige der Signale und Messergebnisse der mV Roh Impedanzstatus Anzeige der Signale und Messergebnisse der Impedanzen Temperatur Anzeige der Signale und Messergebnisse der Temperatur Feuchte Anzeige der Signale und Messergebnisse der Feuchte RFID Anzeige der Signale und Messergebnisse des RFID
Letzte Kalibrierung Anzeige des Alters der letzten Matrixkorrektur
Ammonium
1,3
Nitrat
Wartung
Patronentest Durchführung einer Sensorüberprüfung mit Hilfe der Testkartusche
Patronentest Anzeige ob die einzelnen Kanäle des Sensors OK sind
Wartung Anzeige ob DIAG/TEST OK ist Erdungsstab Anzeige ob GNDROD (ERDSTB) OK ist Referenzelektrode Anzeige ob REF-Kanal OK ist NO3 NH4 ORP Anzeige ob ORP-Kanal OK ist
26 Deutsch
4.2 Menü Sensoreinstellungen (Fortsetzung)
Wartung (Fortsetzung)
1,3
Cl-
1,2
K+
Temperatur Anzeige ob Temperatur-Kanal OK ist Patronenaustausch Folgen Sie dem menügeführten Prozess Reinigung... Folgen Sie dem menügeführten Prozess
1
gilt für AN-ISE sc
2
gilt für AISE sc
3
gilt für NISE sc
Anzeige ob CL-Kanal OK ist Anzeige ob K-Kanal OK ist

4.3 Kalibrierung/Matrixkorrektur

Die 4 Elektroden mit Referenzsystem der kompakten Sensorkartusche wurden mit speziellen Standardlösungen werkseitig aufeinander kalibriert (CARTICAL der ionenselektiven Elektroden sind aber aufgrund anderer Stoffe, die die Messung beeinträchtigen können, nicht zu 100 Führen Sie eine Matrixkorrektur durch (siehe Einflüsse anderer Ionen auf die ISE Elektroden zu kompensieren.
Betrieb
TM
). Die Membranen
% selektiv.
4.3.4, Seite 29), um
Den größten Störeinfluss auf die Ammoniummembran übt Kalium aus, bei der Nitratmembran ist es Chlorid. Die AN-ISE
sc Sonde gleicht dieses Problem mit Hilfe einer integrierten Kalium- bzw. Chloridelektrode ab.
Bei der AISE sc Sonde sind nur die Ammoniummembran und die integrierte Kaliumelektrode aktiv.
Bei der NISE sc Sonde sind nur die Nitratmembran und die integrierte Chloridelektrode aktiv.
Querempfindlichkeiten zwischen Ammonium und Kalium bzw. Nitrat und Chlorid werden automatisch eliminiert. Die Messung wird nicht durch Feststoffe gestört. Aufgrund von Matrixeffekten kann die Korrektur und Validierung nicht mit Standardlösungen erfolgen. Es lässt sich jederzeit auf einfache und schnelle Weise eine Matrixkorrektur durchführen.
HINWEIS
Eine Matrixkorrektur darf erst durchgeführt werden, wenn der Sensor mehr als 12 Stunden in die entsprechende Abwassermatrix eingetaucht war. 12 Stunden ist die mindestens erforderliche Zeit, um die ISE Membranen an die Abwassermatrix anzupassen.

4.3.1 Sensorcode-Kalibrierung

Der Sensorcode ist ein Kalibier-Code und wird mit dem Zertifikat der Sensorkartusche ausgeliefert. Er enthält die unter Kapitel 3.12,
Seite 9 beschriebene Werkskalibrierung für die Sensorkartusche.
Deutsch 27
Betrieb
Bei Geräten mit automatischer Sensorcodeerkennung (LXG440.99.x000x) wird dieser automatisch eingelesen und die Cartrical-Kalibrierung übernommen.
Bei Geräten ohne automatische Sensorcodeerkennung (LXG440.99.x001x) muss bei Aktivierung einer neuen Sensorkartusche, sowie während der Erstinbetriebnahme, der Sensorcode eingegeben werden. Bei Verlust des Sensorcodezertifikats bitte als vorübergehende Lösung Werkskalibrierung (unter Sensorcode Menü) durchführen.
Nach Aktivierung des Codes ist der Sensor vollständig kalibriert, aber noch nicht auf die spezielle Matrix der jeweiligen Applikation auf einer Kläranlage angepasst. Führen Sie nach frühestens 12 Stunden eine Matrixkorrektur durch, um die Kartusche an die spezielle Matrix anzupassen.
Gehen Sie wie folgt vor, um den Sensorcode zu ändern:

4.3.2 Matrixkorrektur über LINK2SC

1. Wählen Sie
NISE SC \> KALIBRIERUNG\> WEITERE KORREKTUREN\> SENSORCODE \> EINGABE
2. Geben Sie den Sensorcode ein.
3. Bestätigen Sie mit
Der Tageszähler für die Cartridge wird auf Null gesetzt.
Alle alten Kalibrierdaten werden mit den neuen Kalibrierdaten aus dem Sensorcode überschrieben. Das System prüft die Daten des Sensorcodes. Falls ein Fehler angezeigt wird, überprüfen Sie den Sensorcode und geben Sie ihn ggf. erneut ein.
Das LINK2SC-Verfahren bietet einen sicheren Datenaustausch zwischen Prozess-Sonden und LINK2SC-fähigen Photometern mit Hilfe einer SD-Speicherkarte oder über ein lokales Netzwerk (LAN). Es wird zwischen zwei Möglichkeiten unterschieden:
a. der reinen Labor-Kontrollmessung und
b. einer Matrix-Korrektur, bei der die im Labor erzeugten
Messdaten zur Korrektur der Sonde eingesetzt werden können.
Bei einer reinen Kontrollmessung werden die Messdaten der Sonde zum Photometer übertragen und dort zusammen mit den photometrisch ermittelten Referenzdaten archiviert.
SENSORMENÜ \> AN-ISE SC oder AISE SC oder
EINGABE, um den Sensorcode zu aktivieren.
Bei einer Matrix-Korrektur werden die im Labor erzeugten Referenzdaten an die Sonde übermittelt und dort für die Korrektur verwendet.
Der Ablauf einer Matrix-Korrektur umfasst Bedienschritte am sc Controller und an einem LINK2SC-fähigen Photometer.
Eine genaue Beschreibung des LINK2SC-Verfahrens entnehmen sie bitte der LINK2SC Bedienungsanleitung.
Bei Nutzung der Software LINK2SC sind die Kapitel 4.3.3 und 4.3.4 überflüssig.
28 Deutsch

4.3.3 Matrixkorrektur – Manuell

Tabelle 4 Korrekturmöglichkeiten für die ISE Sonden
Korrekturmöglichkeit Anwendung
Eine 1-Punkt-Matrixkorrektur ist die am häufigsten angewendete Korrekturmöglichkeit und führt eine 1-Punkt-Matrixkorrektur für Ammonium und/oder Nitrat durch (4.3.4.1, Seite 29). Es wird empfohlen, eine 1-Punkt-Matrixkorrektur als erste Korrektur durchzuführen. Die
1-Punkt-Matrixkorrek tur
1-Punkt-Wertekorrek tur
2-Punkt-Wertekorrek tur
2-Punkt-Matrixkorrek tur
1-Punkt-Matrixkorrektur kann mit oder ohne Korrektur der Kompensationselektroden (Kalium oder Chlorid) durchgeführt werden, wobei ohne in den meisten Fällen ausreichen wird. Eine Korrektur mit Kalium und/oder Chlorid braucht nur durchgeführt zu werden, wenn eine höhere Genauigkeit erforderlich ist. Bei einer 1-Punkt-Matrixkorrektur muss mit Auslösen der Korrektur eine Probe genommen und im Labor analysiert werden. Mit der Eingabe der Laborwerte wird die 1-Punkt-Matrixkorrektur aktiv.
Die 1-Punkt-Wertekorrektur (Korrektur an einem Konzentrationspunkt) entspricht einer 1-Punkt-Wertekorrektur mit einer alternativen Eingabeform.
Bei dieser Korrektur können Vergleichswerte zwischen ISE Sonde und Labor über einen Zeitraum von etwa einer Woche gesammelt werden. Zu einem späteren Zeitpunkt kann dann die Korrektur durchgeführt werden.
Die 2-Punkt-Wertekorrektur (Korrektur an zwei verschiedenen Konzentrationspunkten) sollte durchgeführt werden, wenn dynamische Konzentrationsfluktuationen über mindestens
½ Dekade 1-Punkt-Wertekorrektur kein genügend genaues Ergebnis erzielt.
Bei dieser Korrektur können Vergleichswerte zwischen ISE Sonde und Labor über einen Zeitraum von etwa einer Woche gesammelt werden. Zu einem späteren Zeitpunkt kann dann die Korrektur durchgeführt werden.
Die 2-Punkt-Matrixkorrektur entspricht einer 2-Punkt-Wertekorrektur, verwendet aber eine alternative Eingabeform und wird empfohlen, wenn ein dynamischer Prozess mit einer großen Nitrat/Ammonium-Schwankung über mindestens ½ Dekade 2-Punkt-Matrixkorrektur muss mit Auslösen der Korrektur für beide Punkte eine Probe genommen und im Labor ausgewertet werden. Mit der Eingabe der Laborwerte wird die 2-Punkt-Matrixkorrektur aktiv.
1
und einer 2-Punkt-Matrixkorrektur vorhanden sind oder eine
Betrieb
Bei den ISE Sonden gibt es verschiedene Möglichkeiten (siehe
Tabelle 4), den Sensorwert mit Laborwerten (als Referenzwert) zu
korrigieren.
Der Laborwert der Wasserprobe wird als Nitratstickstoff (NO3–N) und/oder als Ammoniumstickstoff (NH Laborwert ersetzt den zuvor vom Sensor gemessenen Wert.
1
vorliegt. Bei einer
–N) eingegeben. Dieser
4
Vorherige Korrekturen
1
Beispiele einer ½ Dekade: Die Nitratstickstoffkonzentration bewegt sich zwischen 1 und 5 mg NO3–N oder
zwischen 5 und 25 mg/L NO
Rückkehr zu einer der letzten durchgeführten Matrix- und Wertekorrekturen, falls eine Korrektur nicht zu einem Erfolg geführt hat.
–N. (Konz2 = (Konz1 x 10)/2)
3

4.3.4 Durchführung der Matrixkorrektur

Hinweis: Führen Sie die Messungen der Laborwerte bzw. Referenzwerte mit den Küvetten-Tests zeitnah oder aus der stabilisierten Probe durch. So verhindern Sie Änderungen der Probenkonzentration, da Vergleichsuntersuchungen Zeit empfindlich sind.
Unter 7.3 Validierungszubehör, Seite 41 finden Sie empfohlene Tests für Labor Messungen.
4.3.4.1 1-Punkt-Matrixkorrektur
Gehen Sie wie folgt vor, um eine 1-Punkt-Matrixkorrektur durchzuführen:
Deutsch 29
Betrieb
Kalibrierung Matrixkorrektur Weitere Korrekturen Informationen
1-Punkt-Matrixkorrekt ur
Nehmen Sie unverzüglich die Einzelmessung für die Laboranalyse vor.
1. Wählen Sie SENSORMENÜ \> AN-ISE SC oder AISE SC oder
NISE SC \> KALIBRIERUNG \> 1-Punkt-Matrixkorrektur
2. Wählen Sie
drücken Sie
2-Punkt-Matrixkorrektur im Auswahlfenster und
EINGABE.
.
3. Wählen Sie die Parameter aus, die Sie korrigieren möchten
und bestätigen Sie mit
EINGABE.
Auswahlmöglichkeiten für AN-ISE sc: NH4 + NO3; NH4; NO3; NH4 + K; NO3 + Cl; NH4 + K + NO3 + Cl
Auswahlmöglichkeiten für AISE sc: NH4; NH4 + K
Auswahlmöglichkeiten für NISE sc: NO3; NO3 + Cl
Der Sensor hat zu diesem Zeitpunkt die aktuellen Werte der ausgewählten Parameter gespeichert.
4. Entnehmen Sie sofort und so nah wie möglich am Sensor eine
Wasserprobe. Filtrieren Sie diese schnellst möglich und führen Sie zeitnah eine Laboranalyse der ausgewählten Parameter durch, da sich der Messwert schnell ändern kann.
Kalibrierung Matrixkorrektur Weitere Korrekturen Eingabe des
Laborwertes. Informationen
4.3.4.2 1-Punkt-Wertekorrektur
Kalibrierung Matrixkorrektur Weitere Korrekturen Informationen
Wenn Sie den Laborwert bestimmt haben, gehen Sie wie folgt vor:
5. Wählen Sie SENSORMENÜ \> AN-ISE SC oder AISE SC oder NISE SC \> KALIBRIERUNG \> Eingabe des Laborwertes.
6. Sie können nur die Laborwerte für die Parameter eingeben,
wenn vorher die 1-Punkt-Matrixkorrektur ausgewählt wurde. Wählen Sie nach der Eingabe der Werte Bestätigung der Eingabe. zum Bestätigen aus.
Wenn Sie den eingegebenen Laborwert bestätigen, ist die Matrixkorrektur aktiviert.
7. Nach der Aktivierung der Korrektur wird das Ergebnis Korrekturergebnis angezeigt.
Hinweis: Dieser Prozess muss immer vollständig durchlaufen werden, um eine Matrixkorrektur erfolgreich abzuschließen. Sollte eine Korrektur nicht zu einem erfolgreichen Ergebnis führen, wird mit der vorherigen Korrektur weitergerechnet.
Die 1-Punkt-Wertekorrektur bietet die Möglichkeit, nachträglich eine Matrixkorrektur an einem Punkt durchzuführen.
1. Nehmen Sie mehrere Proben an verschiedenen Tagen, möglichst innerhalb von einer Woche mit unterschiedlichen Konzentrationen. Führen Sie eine Analyse der Proben im Labor durch. Während des Probenahmezeitraumes sollte die
30 Deutsch