Hach Accu4 User Manual

DOC023.97.80356
Accu4™ (T53, 8320)
01/2013, Edition 1
User Manual
用户手册
中文.......................................................................................................................................................................................................29
2
Table of contents
Specifications on page 3 Operation on page 18
General information on page 4 Maintenance on page 23
Installation on page 7 Troubleshooting on page 24
User interface and navigation
on page 17
Replacement parts and accessories
on page 27
Startup on page 18
Specifications
Specifications are subject to change without notice.
Model 8320 turbidity sensor
Specification Details
Operational:
Flow rate 0.19 to 26.5 LPM (0.05 to 7 GPM)
Ambient conditions 0 to 60 °C (32 to 140 °F)
Sample temperature range
0 to 60 °C (32 to 140 °F)
Pressure range 0–3.4 bar at 20 °C (0–50 psig at 68 °F)
Pressure drop 0.0001 bar at 0.36 LPM (0.0017 psig at 0.1 GPM)
Residence time 3.8 LPM (9.5 seconds at 1 GPM)
Air venting Integral bubble trap for 0.19 to 1.9 LPM (0.05 to
0.5 GPM) flows.
Installation of an external bubble trap and a restrictor valve on the sensor outlet is recommended for flows above 1.9 LPM (0.5 GPM) with air in the sample. Refer to Use an external bubble trap on page 27.
Mechanical:
Light sources Two near-infrared (880 nm wavelength) LEDs
Specification Details
Power requirements 5 V supplied by the controller to the light sources
and detectors
Process connections ½–in. NPT female standard; adaptable to 3/8–in. or
¼–in. NPT, barb or tube fittings
Wetted materials PVC, polycarbonate, polystyrene, PPO, nitrile and
Buna-N
Enclosure Molded fiber-glass-reinforced polyester (flame
retardant) with four brackets for surface mounting
Dimensions (W x D x H) 285 x 172 x 400 mm (11.21 x 6.79 x 15.75 in.)
Net weight 4.5 kg (10 lb) approximately
Certifications 15 year environmental friendly use period
Model T53 controller
Specification Details
Operational:
Display Graphic dot matrix LCD, 128 x 64 pixels with LED
backlighting; 13 mm (½–in.) main character height; 3 mm (1/8–in.) auxiliary information character height; menu screens contain up to six lines of text
Ambient conditions Operation: –20 to 60 °C (–4 to 140 °F); 0–95% relative
humidity, non-condensing
Storage: –30 to 70 °C (–22 to 158 °F); 0–95% relative humidity, non-condensing
Pollution degree 2
Installation category II
Protection class I
Altitude 2000 m (6561 ft) maximum
English 3
Specification Details
Sensor-to-controller distance
9 m (30 ft) maximum
1
Power requirements 90–130 VAC (115 VAC nominal) or 180–260 VAC
(230 VAC nominal), 50/60 Hz, 10 VA maximum
Relays Four electromechanical relays; SPDT (Form C)
contacts; 115/230 VAC, 5 A @ 30 VDC resistive
Analog outputs Two isolated 0.00–20.00 mA or 4.00–20.00 mA outputs
each with 0.004 mA (12–bit) resolution; up to 600 ohm load capacity
Mechanical:
Enclosure NEMA 4X; polycarbonate face panel, epoxy-coated
cast aluminum door and case with four 13-mm (½–in.) cable entry holes, nylon mounting bracket and stainless hardware
Mounting Panel, surface or pipe mount
Net weight 2.3 kg (5 lb)
Fuses One type T, 80 mA, 250 V slow-blow fuse for the 230 V
line power circuits, 5 mm x 20 mm; one type T, 100 mA, 250 V slow-blow fuse for the 115 V line power circuits, 5 mm x 20 mm
Certifications 15 year environmental friendly use period
Accu4 system performance:
Measurement range 0.000–100.0 NTU; auto-ranging and automatic decimal
point shift above 1.000 NTU and 10.00 NTU (same for other measurement units)
Measurement units NTU, TEF, FNU or FTU
Signal averaging 0 to 60 seconds
System accuracy ±2% of reading, all ranges
Sensitivity 0.001 NTU
Specification Details
Repeatability 0.1% of span or better
Temperature drift Zero and span: 0.01% of span per °C
1
Contact the manufacturer if a longer distance is necessary.
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
4 English
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol, if noted on the instrument, will be included with a danger or caution statement in the manual.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol indicates the presence of devices sensitive to Electro­static Discharge (ESD) and indicates that care must be taken to prevent damage with the equipment.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol indicates that the marked item requires a protective earth connection. If the instrument is not supplied with a ground plug on a cord, make the protective earth connection to the protective conductor terminal.
Certification
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numèrique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
English
5
1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the
interference.
5. Try combinations of the above.
Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoring requirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument.
The Accu4 low-range turbidimeter system includes a Model 8320 turbidity sensor and Model T53 controller. Refer to Figure 1 and
Figure 12 on page 17. The system design meets the International
Standards for Measurement of Turbidity (ISO 7027) for FNU and USEPA-approved GLI Method 2 for NTU.
This system is used to monitor the turbidity of potable water or filtered water.
Figure 1 Model 8320 turbidity sensor overview
1 Calibration hose and funnel 6 Detectors (2x)
2 Terminal board 7 Door latch
3 Light sources (2x) 8 Inlet (½–in. NPT female)
4 Flow chamber 9 3-way value, internal
5 Outlet (½–in. NPT female) 10 Sensor cable
Product components
Make sure that all components have been received. Refer to Figure 2 and Figure 3. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
6
English
Figure 2 Turbidity sensor components
1 Turbidity sensor 3 Mounting screws (4x)
2 Mounting brackets (4x)
Figure 3 Controller components
1 Controller 6 Flat washers, ¼–in. ID (8x)
2 Sealing gasket for panel mount,
Neoprene
7 Hex nuts, M6 (8x)
3 Mounting foot (2x) for panel mount 8 Pan head screws, M6 x 1.0 x
20 mm (4x)
4 Mounting bracket 9 Pan head screws, M6 x 1.0 x
100 mm for pipe mount (4x)
5 Ground screw, green 10 Pan head screws, M6 x 1.0 x
150 mm for panel mount (4x)
Installation
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
Installation guidelines
Install the turbidity sensor as near as possible to where the controller is to be installed. Contact the manufacturer if a distance longer than 9 m (30 ft) is necessary.
English
7
Make sure that the ambient conditions of the installation location of the turbidity sensor and controller are within specifications. Refer to
Specifications on page 3.
Install the turbidity sensor indoors or outdoors where there is little or no ambient vibration. Install the controller in a location that is:
• Clean and dry where there is little or no vibration
• As far as possible away from sources of vibration
• As near as possible to the water source
• Protected from corrosive fluids
Note: If exposed to direct sunlight, the operating temperature of the controller may increase above the specified limit. Direct sunlight may also decrease the display visibility. If necessary, use a sun shield (1000G3088-001).
Mechanical installation
Sensor mounting
Install the turbidity sensor on a flat surface. Refer to the illustrated steps in Figure 4. Mounting hardware is supplied by the user.
Figure 4 Sensor mounting
Controller mounting
Install the controller on a flat surface, panel or pipe (horizontal or vertical). Refer to the instructions supplied with the controller mounting hardware.
8
English
Electrical installation
W A R N I N G
Potential electrocution hazard. In order to maintain the NEMA/IP environmental ratings of the enclosure, use only conduit fittings and cable glands rated for at least NEMA 4X/IP66 to route cables in to the instrument.
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
• Put all wiring to the sensor and controller in ½-inch, grounded metal conduit to protect the cable from moisture, electrical noise or mechanical damage. If only shielded cables are used, applicable strain reliefs or cable grips are necessary. Accessory watertight cable grips (3H1091) and locknuts (3H1230) are available from the manufacturer.
• Make sure that the diameter of the cables used is 4.3–11.4 mm (0.17–
0.45 in.) so that the cable strain reliefs hold the cables securely when tightened.
• Do not put more than one cable in a cable strain relief.
• Close all unused cable entry holes with applicable plugs to keep out moisture.
• Close all unused cable strain reliefs with rods or cables to keep out moisture.
Electrostatic discharge (ESD) considerations
N O T I C E
Potential Instrument Damage. Delicate internal electronic components can be damaged by static electricity, resulting in degraded performance or eventual failure.
Refer to the steps in this procedure to prevent ESD damage to the instrument:
• Touch an earth-grounded metal surface such as the chassis of an instrument, a metal conduit or pipe to discharge static electricity from the body.
• Avoid excessive movement. Transport static-sensitive components in anti-static containers or packages.
• Wear a wrist strap connected by a wire to earth ground.
• Work in a static-safe area with anti-static floor pads and work bench pads.
Wiring overview
Figure 5 shows the wiring connections and fuses for the controller.
English 9
Figure 5 Wiring overview
1 Analog output connections (2x) 5 AC input power connections
2 RS232 connection (not used) 6 Fuses (2x)
3 TTL connection 7 Relay connections (4x)
4 Sensor cable connections 8 Grounding strip
General wiring procedure
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
Connect wiring to the controller as shown in the illustrated steps in
Figure 6. When wiring is complete, close the controller cover and tighten
all the cover screws in order to maintain the NEMA/IP environmental ratings of the enclosure.
10 English
Figure 6 General wiring procedure
Wiring for power
D A N G E R
Electrocution hazard. Protective Earth Ground (PE) connection is required.
D A N G E R
Electrical shock and fire hazards. Make sure to identify the local disconnect clearly for the conduit installation.
W A R N I N G
Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault Interrupt device must be used for connecting the equipment to its mains power source.
N O T I C E
Install the device in a location and position that gives easy access to the disconnect device and its operation.
Supply power to the controller with conduit or a power cable. Make sure that a circuit breaker with sufficient current capacity is installed in the power line. The circuit breaker size is based on the wire gauge used for installation.
For installation with conduit:
• Install a local disconnect for the controller within 3 m (10 ft) of the controller. Put a label on the disconnect that identifies it as the main disconnect device for the controller.
• Make sure that the power and safety ground service drops for the controller are 18–12 AWG.
• Connect equipment in accordance with local, state or national electrical codes.
• Connect the conduit through a conduit hub that holds the conduit securely and seals the enclosure when tightened.
For installation with a power cable, make sure that the power cable is:
• Less than 3 m (10 ft) in length
English
11
• Rated for at least 60 °C (140 °F) and applicable to the installation environment
• Not less than 18 AWG with applicable insulation colors for local code requirements
• A power cable with a three-prong plug (with ground connection) that is applicable to the supply connection
• Connected through a cable gland (strain relief) that holds the power cable securely and seals the enclosure when tightened
• Does not have a locking type device on the plug
Wiring procedure
Refer to Table 1, Figure 5 on page 10 and General wiring procedure on page 10 to connect the controller to power. Supply power to only one terminal (terminal 3 or 4). Connect line power using the standard three­wire connection arrangement.
Do not put the sensor cable (and interconnect cable, if used) in the same conduit with power wiring. Electrical noise may interfere with the sensor signal.
Table 1 AC power wiring information
Input voltage Terminal Description
Color—North America
Color—EU
115 VAC, single phase
3 Hot
Black Brown
230 VAC, single phase or split phase
4 Hot
115 or 230 VAC
2 Neutral (N) White Blue
Protective
Earth (PE) grounding strip
Green Green with
yellow stripe
Connect the sensor cable
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
Refer to General wiring procedure on page 10, Figure 7 and Table 2 to connect the sensor cable to the controller.
Cut the sensor cable to the shortest applicable length before it is connected to the controller to prevent unnecessary electrical noise that may interfere with the sensor signal. Keep the cable away from motors or other equipment that may give off electrical or magnetic fields.
Directly connect the sensor to the controller to remove potential problems caused by wet environments and a junction box.
12 English
Figure 7 Connect the sensor cable
Table 2 Sensor cable wiring information
Terminal block Terminal
number
Wire color Terminal
number
Wire color
TB-1 12 Red 18 Brown
13 Violet 19 Grey
14 Green 20 Blue
15 White 21 Black
16 Yellow 22 Orange
Connect the inner shield wires to the grounding strip.
Connect the analog outputs (optional)
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
The controller contains two isolated 0.00-20.00 mA or 4.00-20.00 mA analog outputs (Output 1 and Outputs 2) that represent the measured turbidity. Use the analog outputs for analog signaling or to control external devices. Refer to Figure 5 on page 10 and General wiring
procedure on page 10 to connect the analog outputs. Refer to Configure the analog outputs on page 19 for configuration information.
Make connections with twisted-pair shielded wire and connect the shield to the grounding strip in the controller.
• Do not connect the shield at both ends of the cable.
• Use of non-shielded cable may result in radio frequency emission or susceptibility levels higher than allowed.
• Maximum loop resistance is 600 ohms.
Notes:
• The analog outputs are isolated from the other electronics, but are not isolated from each other.
English
13
• The analog outputs are self-powered. Do not connect to a load with voltage that is independently applied.
• The analog outputs cannot be used to supply power to a 2-wire (loop­powered) transmitter.
Connect the relays (optional)
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
W A R N I N G
Potential Electrocution Hazard. Power and relay terminals are designed for only single wire termination. Do not use more than one wire in each terminal.
W A R N I N G
Potential fire hazard. Do not daisy-chain the common relay connections or jumper wire from the mains power connection inside the instrument.
C A U T I O N
Fire hazard. Relay loads must be resistive. Always limit current to the relays with an external fuse or breaker. Obey the relay ratings in the Specifications section.
The controller contains four electromechanical relays (Relay A, B, C and D) that are opened or closed by the measured turbidity or diagnostics. Refer to Configure the relays on page 19. The relay outputs are not powered.
Refer to General wiring procedure on page 10, Figure 8 and Table 3 to connect the relays to a control or alarm device.
The relay terminals accept 18–12 AWG wire (as determined by load application). Wire gauge less than 18 AWG is not recommended.
When switching large inductive loads (e.g., motors and pumps) or currents higher than 5 A, use an auxiliary relay to extend the controller relay life.
Use the relays at either all high voltage (greater than 30 V-RMS and
42.2 V-PEAK or 60 VDC) or all low voltage (less than 30 V-RMS and
42.2 V-PEAK, or less than 60 VDC). Do not configure a combination of both high and low voltage.
Figure 8 Typical wiring configuration
1 Relay A 5 Control or alarm device, non-
inductive load switching
2 Relay C 6 Control or alarm device, inductive
load switching
3 Capacitor (0.1 µF, 600 V) 7 100–230 VAC
4 Resistor (100 ohms, 1 watt) 8 Phase 2 or N
Table 3 Relay wiring information
NC COM NO
Normally closed Common Normally open
14 English
Connect the TTL input (optional)
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.
The controller has one TTL input that lets the user hold or transfer the analog outputs and the relays. Refer to Figure 5 on page 10. The TTL input feature function is the same as the output state (HOLD, XFER or ACTIVE) selected during the last calibration.
HOLD—The analog outputs and relays are held at their present values/on-off states.
XFER (transfer)—The analog output values and relay states are changed to the selected Set Transfer values/states. Refer to
Configure the analog outputs on page 19 and Configure the relays
on page 19.
ACTIVE—The analog outputs and relays represent the measured turbidity. The TTL input is disabled.
To apply a TTL hold or transfer, locally or remotely connect the two TTL terminals. When this connection is broken, the applied hold or transfer releases.
Note: The TTL input feature is disabled during calibration and system diagnostic tests that are started by the user.
Plumbing
Plumb the sensor
D A N G E R
Fire hazard. This product is not designed for use with flammable liquids.
N O T I C E
When metal fittings are used, do not over tighten them onto the plastic inlet or outlet fitting of the sensor or damage to the plastic fitting may occur.
Install the sensor in an "in line" process configuration using a closed loop piping arrangement or in a sample bypass line configuration using an open drain piping arrangement. Refer to Figure 9 and Figure 10.
Procedure notes:
• Bushings may be used to decrease to a smaller tubing size. Barb or tube fittings may be installed.
• Use larger sample tubing for applications in which sediment buildup commonly occurs. A faster flow rate helps flush sediment through the sensor.
• Use Teflon tape to seal the inlet and outlet connections. Do not use pipe dope or other liquid sealants.
English 15
Figure 9 Closed-loop piping arrangement
1 Outlet (½–in. NPT female) 6 Process output
2 To open atmosphere (to vent for
calibration)
7 3-way valve (½–in. IPS)
3 Isolation ball valve (½–in. IPS) 8 Inlet (½–in. NPT female)
4 Process output 9 3-way valve (½–in. IPS)
5 Isolation ball valve (½–in. IPS)
Figure 10 Sample bypass line configuration
1 Outlet (½–in. NPT female) 4 Isolation ball valve (½–in. IPS)
2 Isolation ball valve (½–in. IPS) 5 Process input
3 To open drain 6 Inlet (½–in. NPT female)
Connect the sample stream
Install the sample line into a larger process pipe to minimize interference from air bubbles or pipeline bottom sediment. A sample line that goes into the center of a process pipe is best.
Figure 11 shows examples of good and bad methods of sample line
installation into a process pipe.
Keep the sample line as short as possible to decrease analysis delay. Sediment can collect in long sample lines.
16
English
Figure 11 Sampling methods
1 Air 2 Sample flow
User interface and navigation
User interface
Figure 12 shows the display and keypad of the Model T53 controller. Table 4 gives the key functions.
Figure 12 Display and keypad
1 Display 2 Keypad
Table 4 Key functions
Key Function
MENU Go to the Main Menu screen and cancel the procedure
to change a setting
ESC Go back one menu level or cancel the procedure to
change a setting
ENTER Confirm a menu selection or confirm changes to a
setting
RIGHT and LEFT arrows
Toggle between the two Measure screens or change the cursor position
UP and DOWN arrows
Select a menu, or change a setting or value
Display description
Figure 13 shows the Measure screen. To go to the Measure screen,
push MENU and then ESC.
English
17
Note: If the keypad is not used within 30 minutes, except during calibration or while specific controller Test/Maintenance functions are used, the display automatically goes to the Measure screen.
From the Measure screen, push the RIGHT or LEFT arrow to toggle between the two Measure screens. The second Measure screen is for Service use only.
To adjust the display contrast, push and hold ENTER and simultaneously push the UP or DOWN arrow while the Measure screen shows.
Figure 13 Measure screen
1 Relay status
1
3 Analog Output 1 value (mA)
3
2 Turbidity reading
2
1
The relays that are energized show. For example, if Relay A and C are energized, "RELAY: A C" shows. The relay status flashes when a relay overfeed timer has expired.
2
If the reading is not within the measurement range of the controller, a series of "+" or "–" shows.
3
Push the DOWN arrow to see the Analog Output 2 value (mA).
Startup
Select the language
Select the display language (default = English).
1. Push MENU and select CONFIGURE>LANGUAGE.
2. Use the UP and DOWN arrows to select the language.
3. Push ENTER.
Start the flow
Start the flow of sample through the flow chamber.
1. Turn the 3-way valve handle of the sensor to the OPERATE position.
2. If the recommended isolation valve is used downstream of the
sensor outlet, open the isolation valve slowly to start the flow.
3. Measure the flow at the outlet. Make sure that the flow is within the flow range shown in Specifications on page 3.
Operation
Configuration
Configure the sensor
1. Push MENU and select CONFIGURE>SENSOR.
2. Select an option.
Option Description
SELECT UNITS
Select the measurement units shown—NTU (default), TEF, FNU or FTU.
SET FILTER Set the amount of time for measurement averaging—0 to
60 seconds (30 seconds = default).
For example, if the value is set to 30 seconds, the turbidity value shown is updated every 30 seconds and the value is the average of the turbidity values during the previous 30 seconds.
18 English
Option Description
PULSE SUPPRESS
Set the pulse suppression feature to on or off. ON—Pulse suppression is enabled. Pulse suppression prevents unstable readings caused by external interference (e.g., entrained gas bubbles in the sample and electromagnetic interference). Pulse suppression identifies a pulse change (reading spike) as a temporary upset and does not show most of the pulse change. OFF—Pulse suppression is disabled.
ENTER NOTE Select the alphanumberic text that is shown on the top line
of the Measure screen (default = ACCU4). The text can be a maximum of 8 characters. The selected digit is in square brackets. Push the UP or DOWN arrow to change the value of the selected digit. Push the RIGHT arrow to select the next digit.
3. Push ENTER to confirm.
Configure the analog outputs
If the analog outputs of the controller are used, configure the analog outputs. The two analog outputs of the controller represent the measured turbidity.
1. Push MENU and select CONFIGURE>SET OUTPUT 1 (or 2).
2. Select a configuration option.
Option Description
SET 4mA (or 0mA) VALUE
Select the turbidity value that represents 4 mA (or 0 mA) on the analog output—0.00 (default) to 100.00.
SET 20mA VALUE
Select the turbidity value that represents 20 mA on the analog output—0.00 to 100.00 (default).
Note: If the same turbidity values are set for 0/4 mA and 20 mA, the analog outputs automatically go to and stay at 20 mA.
SET TRANSFER
Set the transfer value of the analog output—4 (or 0) to 20 mA (default = 20 mA). During calibration, the analog outputs can be held, transferred or kept active. When transfer is selected, the analog outputs are held at the transfer value.
Option Description
SET FILTER Set the amount of time for analog output averaging—
0 (default) to 60 seconds.
For example, if the value is set to 30 seconds, the value of the analog outputs is updated every 30 seconds and the value is the average of the analog output values during the previous 30 seconds.
Note: The analog output filter time setting is added to the sensor filter time setting which increases the time delay between measurement updates.
SCALE 0mA/4mA
Set the analog output range of the analog output. 0mA— 0 to 20 mA. 4mA (default)—4 to 20 mA.
3. Push ENTER to confirm.
Configure the relays
If the relay connections of the controller are used, configure the relays. The four relays of the controller are energized or de-energized (opened or closed) by the measured turbidity or diagnostics.
1. Push MENU and select CONFIGURE>SET RELAY (A, B, C or D).
2. Select a configuration option.
Option Description
SET FUNCTION
Set the relay to function as a control, alarm or status relay. ALARM (default)—a relay with separate high and low alarm points and deadbands that operates in response to the measured turbidity. CONTROL—a relay with phasing, setpoint, deadband and overfeed timer that operates in response to the measured turbidity. STATUS—a relay that operates only in response to selected system diagnostic conditions.
English 19
Option Description
SET TRANSFER
Set the transfer state of the relay—DE-ENERGIZED (default) or ENERGIZED. During calibration, the relays can be held, transferred or kept active. When transfer is selected, the relays are held at the transfer state.
ACTIVATION This menu changes depending on the selected function of
the relay (alarm relay, control relay or status relay). Refer to SET FUNCTION.
A table for each relay type follows. Refer to the applicable table for the configuration options.
Alarm relay
Option Description
LOW ALARM Set the low alarm value—0.00 (default) to 100.00. The
relay is energized when the measured turbidity is lower than the low alarm value.
HIGH ALARM Setsthe high alarm value—0.00 to 100.00 (default). The
relay is energized when the measured turbidity is higher than the high alarm value.
LOW DEADBAND
Set the measured turbidity range in which the relay stays on after the measured turbidity increases above the low alarm value—0.00 (default) to 10.00.
HIGH DEADBAND
Set the measured turbidity range in which the relay stays on after the measured value decreases below the high alarm value—0.00 (default) to 10.00.
OFF DELAY Set the delay time for the relay to switch off—0 (default)
to 300 seconds.
ON DELAY Set the delay time for the relay to switch on—0 (default)
to 300 seconds.
Control relay
Option Description
PHASE Set the phase for the relay to high or low. HIGH (default)
—The relay is energized when the measured turbidity is higher than the setpoint. LOW—The relay is energized when the measured turbidity is lower than the setpoint.
SET SETPOINT Set the setpoint (measured turbidity) value at which the
relay is energized—0.00 to 100.00 (default).
DEADBAND Set the measured turbidity range in which the relay stays
on—0.00 (default) to 10.00. For a high phase relay, this is the range after the measured value decreases below the setpoint. For a low phase relay, this is the range after the measured value increases above the setpoint.
OVERFEED TIMER
Set the time to limit how long the relay can stay on—
0.0 (default) to 999.9 minutes. For more details on overfeed timer operation, refer to Overfeed timer for
control relays on page 20.
OFF DELAY Set the delay time for the relay to switch off—0 (default) to
300 seconds.
ON DELAY Set the delay time for the relay to switch on—0 (default) to
300 seconds.
Status relay
Option Description
SET STATUS TYPE
Set the status relay type. ALL (default)—The relay is energized when there is a "Fail" or a "Chamber Unknown" condition or a "Chamber Dirty" condition. WARN—The relay is energized only when there is a "Chamber Dirty" condition. FAIL—The relay is energized only when there is a "Fail" or a "Chamber Unknown" system condition. Refer to
Troubleshooting on page 24 for the list of conditions.
3. Push ENTER to confirm.
Overfeed timer for control relays
The overfeed timer setting prevents a condition that keeps the measured turbidity higher than the setpoint or deadband setting (e.g., damaged
20
English
sensor or a process upset) from keeping a control relay switched on continuously. The overfeed timer setting limits how long the relay and its connected control element stay on independent of the conditions.
When a relay overfeed timer is used and it has expired, the relay status flashes on the Measure screen until the overfeed timer is reset. Push Menu and select TEST/MAINT>OVERFEED RESET to reset the overfeed timer.
Reset the configuration
To set the configuration to the factory defaults:
1. From the Measure screen, push MENU and select TEST/MAINT>RESET CONFIG.
2. Push ENTER to confirm.
Note: Push ESC to not set the configuration to the factory defaults.
Calibration
Primary calibration and sample calibration
Before initial use and every 3 months, do a primary calibration or sample calibration to meet USEPA regulations. Regardless of the type of calibration used, the steps are the same. The only difference is the solution used.
Primary calibration—20 NTU or 40 NTU primary standard (formazin
suspension or StablCal)
Note: The USEPA requires 40 NTU primary standard to comply with Method 2.
Sample calibration—A sample with a known turbidity value
Prepare 4000 NTU formazin suspension (primary calibration only)
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
Items to collect:
• Distilled water
• Volumetric flask (4x)
*
• Reagent-grade hydrazine sulfate, (NH2)2H2SO4, 5 grams
• Reagent-grade hexamethylenetetramine, (CH2)6N4, 50 grams
1. Pour some distilled water through a 0.2 micron filter into each flask. Rinse each flasks at least two times with the filtered water
2. Fill one flask with at least 1000 mL of the filtered water.
3. Solution 1: In a second flask, dissolve 5.00 grams of reagent-grade
hydrazine sulfate in filtered water. Dilute the solution to 500 mL with filtered water. Gently swirl to mix.
4. Solution 2: In a third flask, dissolve 50.00 grams of reagent-grade hexamethylenetetramine in filtered water. Dilute the solution to 500 mL with filtered water. Gently swirl to mix.
5. In a fourth flask, mix 50.0 mL of Solution 1 and 50.0 mL of Solution 2 to make a 4000 NTU stock turbidity suspension. Gently swirl to mix.
6. Let the mixture stabilize for 24 hours at 25 °C, ±3 °C (77 °F, ±5.4 °F).
7. Pour the stock suspension into an amber glass (or other UV-light
blocking) bottle for storage. When kept at room temperature and in
*
For the best results, prepare formazin solution with Class A laboratory glassware.
English 21
the dark, the stock suspension stays stable for a maximum of one year.
8. Prepare 20 NTU or 40 NTU standard suspension.
a. 20 NTU standard suspension: Dilute 10 mL of 4000 NTU stock
suspension to 2000 mL with filtered water. Gently swirl to mix.
b. 40 NTU standard suspension: Dilute 20 mL of 4000 NTU stock
suspension to 2000 mL with filtered water. Gently swirl to mix.
Clean and fill the flow chamber
1. Clean the flow chamber. Refer to Clean the flow chamber
on page 23.
2. Push MENU and select CALIBRATE>SENSOR>PRIMARY CAL (primary calibration) or SAMPLE CAL (sample calibration).
3. Select an option.
Option Description
HOLD OUTPUTS Hold the analog outputs and relays at their present
values/on-off states.
XFER OUTPUTS Transfer the analog outputs and relays to set
values/states. Refer to SET TRANSFER setting in
Configure the analog outputs on page 19 and Configure the relays on page 19.
ACTIVE OUTPUTS
Let the analog outputs and relays represent the measured turbidity.
4. Push ENTER to confirm.
5. Pour the sample of known value or prepared standard suspension
(20 NTU or 40 NTU) into the calibration funnel until the suspension can be seen in the clear sight tube.
6. Hold the calibration funnel below the sensor height to drain the sensor.
7. Do step 5 and 6 again to rinse the internal tubing of the sensor.
8. Pour the sample of known value or prepared standard suspension
(20 NTU or 40 NTU) into the calibration funnel until the flow chamber is full.
9. Put the calibration funnel in the funnel holder.
Enter the value of the sample or standard suspension
1. From the “PRIMARY (or SAMPLE) CAL, SAMPLE READY?” screen, push ENTER.
2. Let the reading stabilize which can take as much as 15 minutes, then push ENTER.
Note: If the reading is not stable, "PLEASE WAIT" shows.
3. When "PRIMARY (or SAMPLE) CAL?" shows, enter the known value of the sample or the standard suspension used. Push ENTER to confirm. "PRIMARY (or SAMPLE) CAL: CONFIRM CAL OK?" shows when the calibration is complete.
Note: If “CONFIRM FAILURE” shows, the calibration was not successful. Push ENTER, then select CAL: EXIT to cancel the calibration or CAL: REPEAT to do
the calibration again.
4. Push ENTER to save the calibration. "Confirm Active?" shows.
Note: Push ESC to not save the calibration.
Return to operation
1. Remove the calibration funnel from the holder and hold it below the
sensor height to drain the sensor. Discard the used calibration solution in accordance with local, state and federal disposal regulations.
2. When the sensor is drained, put the calibration funnel in the funnel holder.
3. Start the flow of sample through the flow chamber. Refer to Start the
flow on page 18.
4. Push ENTER. "CONFIRM ACTIVE?" shows.
5. When the reading that shows corresponds to the actual typical
process value, push ENTER to release the analog outputs and relays to active outputs.
22
English
Calibrate the analog outputs
If the analog output connections of the controller are used, calibrate the analog outputs as necessary. The analog outputs are factory-calibrated. The adjustment range for analog output calibration is ±2 mA.
Note: If an analog output is configured to be 0–20 mA, 4 mA and 20 mA (not the 0 mA) are calibrated.
1. Push MENU and select CALIBRATE>CAL OUTPUTS>CAL OUTPUT 1 (or 2).
2. Select a calibration option.
Option Description
CAL OUT 1 4 mA
Set the count value that represents 4 mA on the analog outputs—550 to 910. Use a calibrated digital multimeter to measure the actual minimum value supplied at the analog output connections of the controller. Adjust the count value until the analog outputs are 4.00 mA.
CAL OUT 1 20 mA
Set the count value that represents 20 mA on the analog outputs—3460 to 3820. Use a calibrated digital multimeter to measure the actual maximum value supplied at the analog output connections of the controller. Adjust the count value until the analog outputs are 20.00 mA.
3. Push ENTER to confirm.
Reset the calibration
To set the calibration to the factory defaults:
1. From the Measure screen, push MENU and select TEST/MAINT>RESET CAL.
2. Push ENTER to confirm.
Note: Push ESC to not set the calibration to the factory defaults.
Enable or disable the passcode
Enable the passcode to prevent unauthorized changes to the configuration settings and calibrations. The passcode is disabled by default. The passcode is 3456 and cannot be changed.
1. Push MENU and select CONFIGURE>SET PASSCODE.
2. Push the UP or DOWN arrow to select the option.
Option Description
Disabled The passcode does not have to be entered to change the
configuration settings, do calibrations or go to the TEST/MAINT screens.
Enabled The configuration settings can be seen but not changed until the
passcode is entered. The CALIBRATE and TEST/MAINT screens cannot be seen until the passcode is entered.
3. Push ENTER to confirm.
Maintenance
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap solution and then wipe the instrument dry.
Clean the flow chamber
Clean the flow chamber to remove unwanted material (e.g., dirt, sand) from the flow chamber.
1. Drain the flow chamber.
a. Turn the internal 3-way valve handle to the CALIBRATE position. b. Remove the funnel from the holder and hold it below the sensor
height to drain any sample that is still in the sensor.
c. When the sensor is drained, put the funnel in the funnel holder.
2. Remove the flow chamber cover. Turn the cover 1/8 turn counter-
clockwise to unlock it, then lift the cover off.
English
23
3. Remove the baffle assembly from the flow chamber. It is "keyed" with a slot and guide rib on the left side to fit only one way in the flow chamber.
4. Wipe the interior surface of the flow chamber and the light source and detector windows with a soft, clean, damp cloth.
5. Rinse the baffle assembly with clean water.
6. Install the baffle assembly in the flow chamber.
7. Make sure that the O-ring on the flow chamber cover has silicone
lubricant.
8. Set the flow chamber cover on the flow chamber. Turn the cover 1/
8
turn clockwise to install.
Replace the fuses
D A N G E R
Electrocution hazard. Remove the instrument from the mains power to change the fuses. Disconnect the power cord or turn off the breaker at the panel.
D A N G E R
Fire hazard. Use the same type and current rating to replace fuses.
1. After power is removed from the controller, open the controller. Refer to General wiring procedure on page 10.
2. Remove the blown fuse and replace it with a new fuse. Refer to
Figure 5 on page 10 for the fuse locations. Refer to Replacement parts and accessories on page 27 for the replacement fuse kit.
3. Close the controller. Turn the four screws to tighten.
Replace a light source or detector
D A N G E R
Electrocution hazard. Remove all power from the instrument and relay connections before this maintenance task is started.
Refer to the replacement instructions supplied with the new light source or detector. Make sure to apply silicone lubricant to the O-rings on the new light source or detector before it is installed.
Calibrate the instrument after a light source or detector is replaced. Refer to Primary calibration and sample calibration on page 21.
Note: The manufacturer recommends that the two light sources (or two detectors) be replaced at the same time.
Take out of operation
For long-term storage, take the instrument out of operation.
1. Clean the flow chamber. Refer to Clean the flow chamber on page 23.
2. Dry the flow chamber with a clean, dry, soft cloth.
3. Remove power to the controller.
Troubleshooting
When the system diagnostics identify a condition that is not typical, the Measure screen flashes "Warning Check Status".
Push MENU and select TEST/MAINT>STATUS to see the system status messages. Push ENTER repeatedly to scroll through the status messages. Refer to Table 5 for descriptions of the status messages.
24
English
Contact technical support if the troubleshooting steps below do not correct the problem.
Table 5 Status messages
Message Description Solution
ANALYZEROKThe controller is
operating correctly.
ANALYZER FAIL
The controller is not operating correctly.
Contact technical support.
Source 1 (or 2)OKThe light source is
operating correctly.
SOURCE 1 (or
2) FAIL
The light source is not operating correctly.
Make sure that the light source wires are connected correctly. Replace the light source if necessary. Refer to Replace a light
source or detector on page 24.
DETECT (or 2)OKThe detector is
operating correctly.
DETECT 1 (or
2) FAIL
The detector is not operating correctly.
Make sure that the detector wires are connected correctly. Replace the detector if necessary. Refer to
Replace a light source or detector
on page 24.
CHAMBER OK The sensor is operating
correctly.
CHAMBER DIRTY
The flow chamber contains unwanted material.
Clean the flow chamber. Refer to
Clean the flow chamber
on page 23.
Table 5 Status messages (continued)
Message Description Solution
CHAMBER UNKNOWN
The sensor cable wiring is not connected or not correctly connected to the controller.
Make sure that the sensor cable wires are connected correctly. If necessary, disconnect and examine the sensor cable for continuity and internal shorts with an ohmmeter.
A light source or detector is not operating correctly.
Make sure that the light source and detector wires are connected correctly. If a source or detector failure occurred, replace it. Refer to
Replace a light source or detector
on page 24.
The measurement signal includes electrical noise.
Make sure that the sensor cable is not near motors or other equipment that may make electrical or magnetic fields.
RLY (A, B, C, D) ACTIVE
The relay is energized.
Control relay: The measured turbidity is higher than the setpoint.
Alarm relay: The measured turbidity is higher than the high alarm or lower than the low alarm setpoint.
Status relay: A selected system diagnostic condition has been identified.
English 25
Table 5 Status messages (continued)
Message Description Solution
RLY (A, B, C, D) INACTIVE
The relay is de­energized.
Control relay: The measured turbidity is less than the setpoint.
Alarm relay: The measured turbidity is between the low and high alarm setpoints.
Status relay: A selected system diagnostic condition has not been identified.
RLY (A, B, C, D) TIMEOUT
The relay is de­energized.
Control relay: The overfeed timer has expired and must be manually reset.
Push MENU and select TEST/MAINT>OVERFEED RESET to reset the overfeed timer.
RLY (A, B, C, D) COUNTING
The relay is energized.
Control relay: The time on the overfeed timer is counting down, but has not expired.
System diagnostics
1. From the Measure screen, push MENU and select TEST/MAINT.
2. Select a diagnostic option.
Option Description
CHECK CAL/HOLD
Hold the analog outputs and relays at their present values/on-off states for a maximum of 30 minutes.
Option Description
OUTPUT 1 (or 2) Set the analog value that is present at the analog output
—0 (or 4) to 20 mA (default = 4.05 mA). Use this option to confirm operation of the connected devices.
RELAY (A, B, C,D)Set the state of the relay—ENERGIZED or DE-
ENERGIZED. Use a continuity meter on the NO and NC relay output terminals to confirm the relay status selected.
SIM SENSOR Set the simulated turbidity measurement value—0.00 to
100.00 (default = 0.35 NTU). Use this option to confirm that the analog outputs and relays respond correctly to the simulated turbidity measurement.
3. Push ENTER or ESC to stop the test.
Remove bubbles from the flow cell
Two types of air or gas bubbles can be present in the sample:
• Bubbles that cannot be seen—made when air is absorbed into liquid.
This can occur naturally or be the result of human intervention such as when ozone is injected into water to act as a disinfectant. Bubbles that cannot be seen in the sample form and can be seen in the flow chamber.
• Bubbles that can be seen—made when liquid and air are mixed. For
example, when water flows through a pipe that contains air or when water falls.
Decrease the sensor outlet flow
Remove bubbles that cannot be seen in the sample but can be seen in the flow chamber. To do this, decrease the sensor outlet flow to apply pressure to the sample while it is being measured.
1. Keep the flow rate through the sensor within the specified limits. Refer to Specifications on page 3.
2. Install a flow restrictor in the sensor outlet tube.
3. Adjust the flow restrictor until the flow is slightly decreased.
4. Close the 3-way valve to stop flow through the sensor.
26
English
5. Drain the flow chamber.
a. Turn the internal 3-way valve handle to the CALIBRATE position. b. Remove the funnel from the holder and hold it below the sensor
height to drain any sample that is still in the sensor.
c. When the sensor is drained, put the funnel in the funnel holder.
6. Slowly open the 3-way valve to let the flow chamber fill with the
sample.
Use an external bubble trap
Remove air bubbles that can be seen from the sample with an external bubble trap. Refer to Replacement parts and accessories on page 27. The manufacturer recommends the use of an external bubble trap for flow rates higher than 1.9 LPM (0.5 GPM).
1. Install the external bubble trap vertically (air vent up) and upstream of the sensor. Refer to the instructions supplied with the external bubble trap.
2. Make sure that the air vent is open. Bubbles will stay in the liquid when the air vent is closed.
3. Periodically remove the bubble trap drain plug to let any collected sediment sludge drain.
Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Replacement parts
Description Quantity Item no.
Model T53 controller:
Fuse kit (one 80 mA fuse and one 100 mA fuse)
1 1000G3315-101
Mounting hardware kit, includes: 1 1000G3228-101
Flat washer, ¼-in. ID 8 8H1346
Ground screw, green 1 8H1275-038
Hex nut, M6 8 8H1116
Mounting bracket 1
1000C4F3217-101
Mounting foot 2 1000B4F3222
Pan head screw, M6 x 1.0 x 20 mm 4 8H1216-075
Pan head screw, M6 x 1.0 x 100 mm for pipe mount
4 8H1217-400
Pan head screw, M6 x 1.0 x 150 mm for panel mount
4 8H1217-600
Sealing gasket for panel mount, Neoprene
1
1000A4F3249-101
Model 8320 turbidity sensor:
Baffle assembly 1 8220G1060
Detector assembly with O-ring set, includes two detectors
1 8220G1330-204
Flow chamber cover 1 8220C2F1076
Internal 3-way valve for sensor with attached tubing
1 8320A1074-001
Light source assembly with O-ring set, includes two light sources
1 8220G1320-204
O-ring, flow chamber 1 5H1288
English 27
Replacement parts and accessories (continued)
Description Quantity Item no.
O-ring, light source or detector 1 5H1251
Retainer clip for light source or detector 1 8220F1078
Accessories
Description Quantity Item no.
Model 8320 turbidity sensor:
Bubble trap, head regulator 1 4668000
Lubricant, silicone 2 g 6M0920
Flask, class A, 2 L 1 2489454
Formazin, 4000 NTU 500 mL 2461-49
StablCal, 40 NTU 1 L 27463-53
StablCal, 40 NTU 1 gallon 27463-56
Tensette, Pipet, 1–10 mL 1 1970010
28 English
目录
规格 第 29 操作 第 42
基本信息 第 30 维护 第 46
安装 第 33 故障排除 第 47
用户界面及导航 第 40 更换部件与附件 第 49
启动 第 41
规格
产品规格如有变化,恕不另行通知。
8320 型浊度传感器
规格 详细信息
工作参数:
流速 0.19 26.5 LPM (0.05 7 GPM)
环境条件 0 60 °C (32 140 °F)
采样温度范围 0 60 °C (32 140 °F)
压力范围
0–3.4 bar (20 °C) (0–50 psig (68 °F))
压降 0.36 LPM 0.0001 bar (0.1 GPM 0.0017 psig)
停留时间 3.8 LPM1 GPM 9.5 秒)
通气 0.19 1.9 LPM0.05 0.5 GPM)流速下集成除泡器。
流速高于 1.9 LPM (0.5 GPM) 时样品中含有空气,建议在传 感器出口处安装一个外部除泡器和一个限流阀。请参阅 使用
外部除泡器 第 48。
机械:
光源 两个近红外(880 nm 波长)LED
电源要求 由控制器向光源和检测器供应 5 V 电源
规格 详细信息
工艺连接 ½–in. NPT 内螺纹标准接头;配套 3/8 in 或 ¼ in NPT 倒钩接
头或管接头
过流材料 PVC、聚碳酸酯、聚苯乙烯、PPO、丁腈橡胶
外壳 模制玻璃纤维增强聚酯(阻燃型),带有四个表面安装支架
尺寸(宽 x x 高)
285 x 172 x 400 mm (11.21 x 6.79 x 15.75 in.)
净重 4.5 kg (10 lb)
认证 15 年环保使用期限
T53 型控制器
规格 详细信息
工作参数:
显示屏 图形点阵式 LCD128 x 64 像素,LED 背光照明;13 mm (½
in) 主字符高度;3 mm (1/8 in)辅助信息字符高度;菜单屏幕含 有多达六行文本
环境条件 工作环境:–20 60 °C–4 140 °F);0–95% 相对湿度,
不凝结
存储环境:–30 至 70 °C(–22 158 °F);0–95% 相对湿度, 不凝结
污染程度
2
安装类别
II
保护等级
I
海拔 最高 2000 m (6561 ft)
传感器与控制器 的间距
最高 9 m (30 ft)
1
电源要求 90–130 VAC(115 VAC 标称电压)或 180–260 VAC(230 VAC
标称电压),50/60 Hz,最大 10 VA
中文
29
规格 详细信息
继电器 四个机电继电器;SPDTC 型)触点;115/230 VAC5 A @
30 VDC 阻抗
模拟输出端 两个 0.00–20.00 mA 4.00–20.00 mA 隔离输出端,分辨率各
0.004 mA12 位);负载能力可达 600 Ω
机械:
外壳 NEMA 4X;聚碳酸酯面板,环氧涂层铸铝门和壳体,带四个
13 mm (½ in) 电缆进孔、尼龙固定支架及不锈钢安装件
安装 板装、面装或管装
净重
2.3 kg (5 lb)
保险丝 一个为 T 80 mA/250 V 慢断保险丝,适用于 230 V 线性电源
电路,尺寸 5 mm x 20 mm;一个为 T 100 mA/250 V 慢断 保险丝,适用于 115 V 线性电源电路,尺寸 5 mm x 20 mm
认证 15 年环保使用期限
Accu4 系统性能:
测量范围 0.000–100.0 NTU;高于 1.000 NTU 10.00 NTU 时自动调节
量程和自动移动小数点(其他测量装置同样如此)
测量装置 NTUTEFFNU FTU
信号平均 0 60
系统精度 所有量程:±2% × 读数
灵敏度
0.001 NTU
重复性 量程的 0.1% 或更好
温度漂移 零点和跨度:0.01% × 跨度/°C
1
如果需要更大间距请联系制造商。
基本信息
对于因本手册中的任何不足或遗漏造成的直接、间接、特别、附带或结果 性损失,制造商概不负责。制造商保留随时更改本手册和手册中描述的产
品的权利,如有更改恕不另行通知或承担有关责任。修订版可在制造商的 网站上找到。
安全信息
注 意
对于误用和滥用造成的产品损坏,制造商概不负责,包括但不限于:直接、附带和 间接的损坏,并且对于适用法律允许的最大程度的损坏也不承担任何责任。用户唯 一的责任是识别重大应用风险和安装适当的系统,以在设备可能出现故障时保护流 程。
请在拆开本设备包装、安装或使用本设备前,完整阅读本手册。特别要注 意所有的危险警告和注意事项。否则,可能会对操作者造成严重的人身伤 害,或者对设备造成损坏。
确保设备提供的保护没有受损。请勿以本手册指定方式之外的其它方式使 用或安装本设备。
危险信息使用
危 险
表示潜在的或紧急的危险情况,如果不加以避免,将会导致死亡或严重伤害。
警 告
表示潜在或非常危险的情形,如不避免,可能导致严重的人身伤亡。
警 告
表示潜在的危险情形,可能导致一定程度的人身伤害。
注 意
表明如不加以避免则会导致仪器损坏的情况。需要特别强调的信息。
30
中文
警告标签
请阅读贴在仪器上的所有标签和标记。如未遵照这些安全标签的指示操 作,则可能造成人身伤害或仪器损坏。仪器上如有标志,则手册中会提供 危险或小心说明。
这是安全警报标志。请遵守此标志后面的所有安全信息,以避免可能造 成的伤害。如果仪器上有此标志,则请参见仪器手册,了解操作或安全 信息。
此标志指示存在电击和/或电死危险。
此标志指示存在静电释放 (ESD) 敏感的设备,且必须小心谨慎以避 免设备损坏。
使用此符号标记的电气设备在 2005 8 12 日后,不能通过欧洲公 共垃圾系统进行处理。为遵守欧洲地区和国家法规(欧盟指令 2002/96/EC),欧洲电气设备使用者现在必须将废弃或到期的设备送还 制造商进行处理,使用者不必支付任何费用。
注: 如果退回产品是为了进行再循环,请联系设备生产商或供应商,索取如何退回 使用寿命到期的设备、生产商提供的电源附件以及所有辅助部件的说明,以便进行 适当处理。
仪表上如有此标志,则指示保险丝或者限流装置的位置。
此标志指示标记的项目需要保护性接地连接。如果仪器的电绳没有随 附接地式插头,需确保保护导体端子连接了保护接地连接。
认证
加拿大无线电干扰产生设备法规(Canadian Radio Interference- Causing Equipment Regulation),IECS-003A 类:
制造商支持测试记录留存。 此 A 类数字设备符合加拿大干扰产生设备法规的所有要求。
Cet appareil numèrique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC 15 部分,“A”类限制 制造商支持测试记录留存。该设备符合 FCC 规定第 15 部分的要求。设
备操作满足以下两个条件:
1. 本设备不会造成有害干扰。
2. 本设备必须接受任何接收到的干扰,包括可能导致意外操作的干扰。
若未经负责出具符合声明的一方明确同意擅自对本设备进行改动或改装, 可能会导致取消用户操作该设备的权限。本设备已经过测试,符合 FCC 规 定第 15 部分中确定的 A 类数字设备限制。这些限制专门提供当设备在商 业环境下工作时针对有害干扰的合理保护。该设备产生、使用和放射无线 电射频能量,如果不按照说明手册的要求对其进行安装和使用,可能会对 无线电通讯造成有害干扰。本设备在居民区工作时可能会产生有害干扰, 这种情况下用户须自行承担费用消除这种干扰。以下方法可用于减少干扰 问题:
1. 断开设备的电源,以便确证它是干扰源与否。
2. 如果设备与遭受干扰的仪器连接到相同的插座,将设备连接到其他插
座。
3. 将设备从接受干扰的仪器边上移开。
4. 重新定位受干扰仪器的接收天线。
5. 同时尝试以上多项措施。
产品概述
危 险
化学或生物危害。如果该仪器用于监测处理过程和/或监测有法规限制以 及有与公众健康、公众安全、食品或饮料制造或加工相关监测要求的化 学药物添加系统时,仪器的使用者有责任了解并遵守所有适用的法规, 并且要建立适当的机制,确保在仪器发生故障的时候也不会违法这些法 规。
Accu4 低量程浊度计系统含有一个 8320 型浊度传感器和一个 T53 型控制 器。请参阅 1 12 41。系统设计满足国际标准:FNU:浊度测 量法 (ISO 7027) NTUUSEPA 批准的 GLI 方法 2
该系统用于监控饮用水或过滤水的浊度。
中文
31
1 8320 型浊度传感器概述
1 校准软管和漏斗 6 检测器(2 个)
2 接线板 7 门锁
3 光源(2 个) 8 入口(½ in NPT 内螺纹)
4 流动槽 9 内部三通阀
5 出口 (½ in NPT 内螺纹) 10 传感器线缆
产品组件
确保已收到所有组件。请参阅 2 3。如有任何物品丢失或损坏,请 立即联系制造商或销售代表。
2 浊度传感器组件
1 浊度传感器 3 安装螺钉(4 个)
2 固定支架(4 个)
32
中文
3 控制器组件
1 控制器 6 平垫圈,¼ inID8 个)
2 面板安装的密封圈,氯丁橡胶 7 六角螺母,M68 个)
3 面板安装用固定支脚(2 个) 8 盘头螺钉,M6 x 1.0 x 20 mm4 个)
4 固定支架 9 管装用盘头螺钉,M6 x 1.0 x
100 mm4 个)
5 地脚螺钉,绿色 10 板装用盘头螺钉,M6 x 1.0 x
150 mm4 个)
安装
警 告
多种危险。只有合规的专业人员才能从事文件中本部分所述的任务。
安装指南
尽可能靠近控制器的安装位置来安装浊度传感器。如果距离超过 9 m (30 ft),需要联系制造商。
确保浊度传感器和控制器的安装位置的环境条件处于规格范围内。请参阅
规格 第 29。
将浊度传感器安装在环境振动较小或无振动的室内或室外。控制器的安装 位置需要:
清洁、干燥,振动较小或无振动
尽可能远离振动源
尽可能靠近水源
防止接触腐蚀性流体
注: 如果在阳光下直晒,控制器的工作温度可能超过指定限值。阳光直晒可能降低显 示屏的可见度。必要时使用一个遮阳板
(1000G3088-001)
机械安装
安装传感器
将浊度传感器安装在一个平面上。请参阅 4 所示步骤。由用户提供安 装硬件。
中文
33
4 安装传感器
控制器安装
将控制器安装一个平面、面板或管子上(水平或垂直安装)。请参阅控制 器安装件随附的说明书。
电气安装
警 告
可能存在电击致命危险。为了保持外壳的 NEMA/IP 环境防护等级,仅限 使用至少为 NEMA 4X/IP66 防护等级的导线管配件和电缆衬垫,以将电 缆接入仪器。
警 告
可能存在电击致命危险。进行任何电气连接时,请务必断开仪器的电源。
将传感器和控制器的所有接入线放入一根 ½ 英寸的接地金属导管中,以 防电缆遭受湿气、电气噪声或机械损坏。 如果只使用屏蔽电缆,需要使用合适的应力消除装置或电缆夹。防水电 缆夹 (3H1091) 和锁紧螺母 (3H1230) 附件可从制造商处购得。
确保所用电缆的直径为 4.3–11.4 mm (0.17–0.45 in)。该直径能使电缆 应力消除装置牢固固定所拧紧的电缆。
切勿将多根电缆放在一个电缆应力消除装置中。
用合适的塞子封住所有未使用的电缆进孔,以防湿气进入。
用杆或电缆封住所有未使用的电缆应力消除装置,以防湿气进入。
静电放电 (ESD) 注意事项
注 意
可能导致仪器损坏。静电会损害精密的内部电子组件,从而导致仪器性 能降低或最终出现故障。
请参阅此流程中的步骤以防止 ESD 损坏仪器:
接触接地金属表面(如仪器外壳、金属导管或管道),以将静电排出机 壳外。
避免过度移动。运送静电敏感的组件时,请使用抗静电容器或包装。
配戴连接到接地线缆的腕带。
使用抗静电地板垫和工作台垫,以使工作区具备静电安全性。
布线概述
5 所示是控制器的接线头和保险丝。
34
中文
5 布线概述
1 模拟输出接头(2 个) 5 交流输入电源接头
2 RS232 接头(未用) 6 保险丝(2 根)
3 TTL 接头 7 继电器接头(4 个)
4 传感器电缆接头 8 接地片
总布线流程
警 告
可能存在电击致命危险。进行任何电气连接时,请务必断开仪器的电源。
按照 6 所示步骤连接控制器的接线。完成布线时,关上控制器盖,然后 拧紧所有的盖螺钉,以便保持 NEMA/IP 的环境封闭等级。
6 总布线流程
中文
35
电源接线
危 险
电击致命危险。需要连接保护接地地线。
危 险
电击和火灾危险。务必为管道安装清晰标识本地隔断开关。
警 告
可能存在电击致命危险。如果此设备在户外或在可能潮湿的场所使用, 则必须使用防高压触电装置将此设备连接到其电源。
注 意
请将设备安装在便于切断设备开关和其操作的场所和位置。
利用导管或电源电缆为控制器供电。确保电源线中安装一个电流能力足够 的断路器。断路器的规格取决于装置所用的线号。
对于含有导管的装置:
在控制器 3 m (10 ft) 范围内为控制器安装一个本地断路器。在断路器上 贴上一个标签,将其标为控制器的主断路装置。
确保控制器的电源和安全接地架空引入线为 18–12 AWG
按照当地、州或国家的电气规范连接设备。
通过一个导管连接头来连接导管,以便牢固地固定导管,并在拧紧后密
封外壳。
对于含有电源电缆的装置,确保电源电缆:
长度小于 3 m (10 ft)
额定工作温度至少为 60 °C (140 °F),并且适用于安装环境
线号不小于 18 AWG,采用符合当地规范要求的绝缘色彩
供电接头采用三芯插头式(带有接地端)电源电缆
通过电缆接头(应力消除装置)进行连接,以便牢固地固定电源电缆并
在拧紧时密封外壳
插头上不含锁紧装置
接线程序 请参阅 1 5 第 35 及总布线流程 第 35 连接控制器的电源。只能为一 个端子(端子 3 4)供电。利用标准的三线连接方式连接线电源。
切勿使传感器电缆(及可能使用的互连电缆)与电源线位于同一导管中。 因为电气噪声可能干扰传感器信号。
1 交流电源接线信息
输入电压 终端 说明
色彩北美 色彩欧盟
115 VAC,单 相
3
火线
黑色 棕色
230 VAC,单 相或分相
4
火线
115 230 VAC
2
中性线 (N) 白色 蓝色
保护接地 (PE)片绿色 带有黄色条纹
的绿色
连接传感器电缆
警 告
可能存在电击致命危险。进行任何电气连接时,请务必断开仪器的电源。
请参阅总布线流程 35 7 2 将传感器电缆连接到控制器。 传感器电缆连接到控制器之前先切到最短的适用长度,以防不必要的电气
噪音可能干扰传感器信号。使电缆远离可能产生电场或磁场的电动机或其 他设备。
直接将传感器连接到控制器可消除潮湿环境和接线盒造成的潜在问题。
36
中文
7 连接传感器电缆
2 传感器电缆接线信息
端子区 端子编号 线颜色 端子编号 线颜色
TB-1 12
红色
18
棕色
13
紫罗兰色
19
灰色
14
绿色
20
蓝色
15
白色
21
黑色
16
黄色
22
橙色
将内部屏蔽线连接到接地片。
连接模拟输出(可选)
警 告
可能存在电击致命危险。进行任何电气连接时,请务必断开仪器的电源。
控制器含有两个独立的 0.00-20.00 mA 或 4.00-20.00 mA 模拟输出端(输 出端 1 和输出端 2),用来输出所测浊度。模拟输出端用于输出模拟信号 或控制外部装置。请参阅图 5 第 35 和总布线流程 第 35 连接模拟输出端。 配置信息请参阅配置模拟输出 第 42。
用双绞屏蔽线进行连接,并将屏蔽线连接到控制器的接地片。
请勿连接电缆两端的屏蔽罩。
使用非屏蔽电缆可能会导致射频发射或磁化级别高于所允许的范围。
最大环路电阻为 600 欧姆。
注意:
模拟输出端与其他的电子装置隔离开,但自身不会相互隔离。
模拟输出端为自供电式。请勿与独立供电的负载相连。
模拟输出端不能用来为双线(环路供电)变送器供电。
中文
37
连接继电器(可选)
警 告
可能存在电击致命危险。进行任何电气连接时,请务必断开仪器的电源。
警 告
可能存在电击致命危险。电源和中继器的端子仅适用于单线端接。请勿 在每个端子连接多根电线。
警 告
可能存在火灾危险。日常不要束缚公共继电器连接或仪器内电源连接的 跳线。
警 告
火灾危险。继电器负载必须为阻性负载。始终限制具有外部保险丝或断 路器的继电器的电流。遵守规格部分的继电器额定值。
控制器含有四个机电继电器(继电器 A、B、C 和 D),通过所测浊度或诊 断进行打开或关闭。请参阅 配置继电器 第 43。继电器输出端不供电。
请参阅总布线流程 35 8 3 将继电器连接到控制器或报警装置。 继电器终端可用 18–12 AWG 电线(取决于负载情况)。不推荐使用线号
小于 18 AWG 的电线。 切换大型电感负载(比如电动机和泵等)或电流高于 5 A 时,使用一个辅
助继电器可延长控制器继电器的寿命。 继电器可在所有的高电压(高于 30 V-RMS 42.2 V-PEAK 或高于
60 VDC)或所有的低电压(低于 30 V-RMS 42.2 V-PEAK,或低于 60 VDC)下使用。切勿配置高电压与低电压的组合。
8 典型的布线配置
1 继电器 A 5 控制器或报警装置,非电感负载切换
2 继电器 C 6 控制器或报警装置,电感负载切换
3 电容器 (0.1 µF, 600 V) 7 100–230 VAC
4 电阻器(100 Ω, 1 W 8 2 相或 N
3 继电器接线信息
NC COM
常闭 常用 常开
连接 TTL 输入端(可选)
警 告
可能存在电击致命危险。进行任何电气连接时,请务必断开仪器的电源。
38
中文
控制器有一个 TTL 输入端能让用户保持或转换模拟输出端和继电器。请 参阅 5 35TTL 输入端的功能与最后一次校准时选择的输出状态
HOLDXFER ACTIVE)完全相同。
HOLD(保持)模拟输出端和继电器保持在当前值/-关状态。
XFER(转换)— 模拟输出值和继电器的状态变换为选定的设置转换 /状态。请参阅配置模拟输出 42 配置继电器 43
ACTIVE(启用)模拟输出端和继电器用于表示所测浊度。TTL 输入
端被禁用。
为了应用 TTL 的保持或转换功能,需要现场或远程连接两个 TTL 端子。断 开该连接时,所应用的保持或转换功能将被释放。
注: 用户启动校准及系统诊断测试时,
TTL
输入功能将被禁用。
装设管道
密封传感器
危 险
火灾危险。此产品不得与易燃液体一同使用。
注 意
使用金属接头时,切勿过度拧紧到传感器的塑料入口或出口接头上,否则可能损坏 塑料接头。
利用闭环管道布置将传感器安装在管路内工艺配置中,或是利用明渠管道 布置将传感器安装在样品旁路配置中。请参阅 9 10
流程注释:
可以使用衬套配合小尺寸管。可以安装倒钩接头或管接头。
对于沉积物较多的应用场合,使用大尺寸取样管。较快的流速有利于冲
洗沉积物穿过传感器。
利用特氟龙密封带密封入口和出口。切勿使用管道粘合剂或其他液体密 封胶。
9 闭环管道布置
1 出口 (½ in NPT 内螺纹) 6 工艺输出
2 接大气(排气校准) 7 三通阀 (½ in IPS)
3 隔离球阀 (½ in IPS) 8 入口(½ in NPT 内螺纹)
4 工艺输出 9 三通阀 (½ in IPS)
5 隔离球阀 (½ in IPS)
中文
39
10 采样旁路配置
1 出口 (½ in NPT 内螺纹) 4 隔离球阀 (½ in IPS)
2 隔离球阀 (½ in IPS) 5 工艺输入
3 通向明渠 6 入口(½ in NPT 内螺纹)
连接样品流
将取样管线安装到更大的工艺管道以最大程度地降低来自气泡或者管底沉 积物的干扰。工艺管道中心的取样管线最佳。
11 显示了在工艺管道中的取样管线安装的好与不好的方法的示例。
使取样管线尽可能地短以降低分析延迟。长采样管会聚集沉积物。
11 取样方法
1 空气 2 试样流
用户界面及导航
用户界面
12 所示是 T53 型控制器的显示屏和键盘。 4 介绍了键的功能。
40
中文
12 显示屏和键盘
1 显示屏 2 键盘
4 键功能
功能
MENU(菜单) 进入主菜单屏幕,并取消设置更改流程
Esc
返回一个菜单级,或取消设置更改流程
ENTER(输入) 确认菜单的选择,或确认设置的更改
RIGHT(向右)和 LEFT(向左)箭头键在两个测量屏幕之间切换,或更改光标位置
UP(向上)和 DOWN(向下)箭头 选择一个菜单,或更改设置或数值
显示屏说明
13 显示测量屏幕。如需进入测量屏幕时,先按 MENU(菜单)键,后按
ESC(退出)键。
注: 如果
30
分钟内未使用键盘,除了校准期间或是使用特定的控制器测试/维护功能
时,显示屏将自动进入测量屏幕。
在测量屏幕中按 RIGHT(向右)或 LEFT(向左)箭头键可在两个测量屏 幕之间进行切换。第二个测量屏幕仅在维修时使用。
如需调整显示屏对比度,可在显示测量屏幕时按住 ENTER 键的同时按 UP
(向上)或 DOWN(向下)箭头键。
13 测量屏幕
1 继电器状态
1
3 模拟输出端 1 的数值 (mA)
3
2 浊度读数
2
1
显示通电的继电器。比如,如果继电器 A C 通电,则显示 "RELAY: A C"。如 果继电器过载定时器超时,继电器的状态指示灯将会闪烁。
2
如果读数未处于控制器的测量范围内,则会显示一系列 "+" "–"
3
DOWN 箭头键可查看模拟输出端 2 的数值 (mA)
启动
选择语言
选择显示语言(默认为英语)。
1. MENU(菜单)并选择 CONFIGURE>LANGUAGE(配置>语言)。
2. 利用 UP(向上)和 DOWN(向下)箭头键选择语言。
3. ENTER
中文
41
开始流动
使样品开始流过流动槽。
1. 将传感器的三通阀手柄转到操作位置。
2. 如果在传感器出口的下游使用推荐的隔离阀,缓慢打开隔离阀使液体流
动。
3. 测量出口流速。确保流速处于 规格 29 所示流速范围内。
操作
配置
配置传感器
1. MENU(菜单)并选择 CONFIGURE>SENSOR(配置>传感器)。
2. 选择一个选项。
选项 说明
SELECT UNITS(选择
单位)
选择所示测量单位 — NTU(默认)、TEFFNU FTU
SET FILTER
(设置过滤器)
设定测量值平均时间 — 060 秒(30 = 默认值)。 比如,如果将数值设为 30 秒,则所示浊度值每 30 秒更新一
次,且该数值是前 30 秒内浊度值的平均值。
脉冲抑制 将脉冲抑制功能设为打开或关闭。ON(打开)启用脉冲抑
制。脉冲抑制可防止外部干扰(比如样品中截留的气泡和电磁 干扰等)造成读数不稳定。脉冲抑制将一个脉冲变化(读数尖 峰)视为一个临时的扰动,并且不显示大多数脉冲变化。OFF
(关闭)禁用脉冲抑制。
输入注释 选择测量屏幕顶行显示的字母数字文本(默认 = ACCU4)。
该文本最长 8 个字符。所选数字位于方括号中。按 UP(向上)DOWN(向下)箭头键更改所选数字的值。按 RIGHT(向 右)箭头键选择下一个数字。
3. ENTER 确认。
配置模拟输出
如果使用控制器的模拟输出,则应配置模拟输出。控制器的两个模拟输出 代表所测浊度。
1. MENU 并选择 CONFIGURE>SET OUTPUT 1(或 2)。
2. 选择一个配置选项。
选项 说明
SET 4mA (or 0mA) VALUE(设定 4mA
(或 0mA) 值)
选择模拟输出中代表 4 mA(或 0 mA)的浊度值 — 0.00
(默认)~100.00
SET 20mA VALUE
(设定 20 mA 值)
选择模拟输出中代表 20 mA 的浊度值 — 0.00100.00
(默认)。
注: 如果为
0/4 mA 和 20 mA
设置相同的浊度值,则模
拟输出会自动进入并保持在
20 mA
SET TRANSFER
(设定转换)
设定模拟输出转换值 — 4(或 0)20 mA(默认 = 20 mA)。校准期间,可以保持、转换模拟输出或使其 保持活动状态。选择转换时,模拟输出将保持在转换值。
SET FILTER(设置 过滤器)
设定模拟输出平均时间 — 0(默认)~60 秒。 比如,如果该值设为 30 秒,则模拟输出值每 30 秒更新
一次,并且该值是前 30 秒内模拟输出平均值。
注: 模拟输出过滤器的时间设置添加在传感器过滤器的 时间设置中,将会增加两次测量更新之间的延时。
SCALE 0 mA/4 mA
(范围 0 mA/4 mA
设定模拟输出的范围。0mA — 020 mA4mA(默 认)— 420 mA
3. ENTER 确认。
42
中文
配置继电器
如果使用控制器的继电器接头,需要配置继电器。控制器的四个继电器通 过所测浊度或诊断实现通电或断电(开或关)。
1. MENU 并选择 CONFIGURE>SET RELAYABC D)。
2. 选择一个配置选项。
选项 说明
SET FUNCTION
(设定功能)
将继电器设为控制、报警或状态继电器。ALARM(报警)
(默认)— 该继电器具有单独的报警上下限和死区,响应所
测浊度进行操作。CONTROL(控制)该继电器具有状 态、设定值、死区及过载定时器,响应所测浊度进行操作。 STATUS(状态) — 该继电器响应所选的系统诊断状况进 行操作。
SET TRANSFER
(设定转换)
设定继电器的转换状态断电(默认)或通电。校准期 间,继电器可以保持或转换状态,或保持启用。选择转换 时,继电器将保持在转换状态。
ACTIVATION(激活)该菜单随所选继电器功能(报警继电器、控制继电器或状
态继电器)而变。请参阅设置功能。 每种继电器都有一个介绍表格。请参阅适用表格了解配置
选项。
报警继电器
选项 说明
LOW ALARM(低位警报)设定报警下限 — 0.00(默认)至 100.00。所测浊度
低于报警下限时继电器通电。
HIGH ALARM(高位警报)设定报警上限 — 0.00 100.00(默认)。所测浊
度高于报警上限时继电器通电。
LOW DEADBAND(低 位死区)
设定所测浊度增加到高于报警下限之后使继电器保 持接通的浊度范围 — 0.00(默认)至 10.00
HIGH DEADBAND(高 位死区)
设定所测浊度值降到低于报警上限之后使继电器保 持接通的浊度范围 — 0.00(默认)至 10.00
OFF DELAY(关闭延时) 设定继电器的关闭延时 — 0(默认)至 300 秒。
ON DELAY(打开延时) 设定继电器的打开延时 — 0(默认)至 300 秒。
控制继电器
选项 说明
PHASE(状态) 设定继电器处于高位或低位。高位(默认)所测浊度
值高于设定值时继电器通电。低位所测浊度值低于 设定值时继电器通电。
SET SETPOINT(设 定点)
设定使继电器通电的设定值(所测浊度)— 0.00
100.00(默认)。
DEADBAND(死区) 设定使继电器保持接通的所测浊度范围 — 0.00(默认)
10.00。对于高位继电器,是指所测浊度值降至低于设 定值之后的浊度范围。对于低位继电器,是指所测浊度 值升至高于设定值之后的浊度范围。
OVERFEED TIMER
(过载定时器)
设定继电器保持接通的限制时间 — 0.0(默认)至
999.9 分钟。有关过载定时器操作的详情,请参阅控制继
电器的过载定时器 第 43。
OFF DELAY(关闭 延时)
设定继电器的关闭延时 — 0(默认)至 300 秒。
ON DELAY(打开延时)设定继电器的打开延时 — 0(默认)至 300 秒。
状态继电器
选项 说明
SET STATUS TYPE(设定状
态类型)
设定继电器的状态类型。ALL(所有)(默认)出现故 障流动槽未知流动槽变脏状况时,继电器通电。
WARN(警告)出现流动槽变脏状况时,继电器通电。 FAIL(故障)出现故障流动槽未知状况时,继电器通
电。请参阅故障排除 47 了解状况列表。
3. ENTER 确认。
控制继电器的过载定时器 过载定时器设置用于防止所测浊度高于设定值或死区设置(比如因为传感 器损坏或工艺扰动)的情况使控制继电器保持连续接通。过载定时器设置 可限制继电器及其所连控制元件保持接通的时间,不受其他情况影响。
如果使用继电器过载定时器并且出现超时,继电器的状态指示灯将在测量 屏幕上闪烁,直到重置过载定时器为止。按 Menu(菜单)并选择
中文
43
TEST/MAINT>OVERFEED RESET(测试/维护>过载重置)以重置过载定 时器。
重置配置
将配置设为出厂默认值:
1. 在测量屏幕中按 MENU(菜单)并选择 TEST/MAINT>RESET CONFIG (测试/维护>重置配置)。
2. ENTER 确认。
注: 按
ESC
取消将配置设为出厂默认值。
校准
初级校准和样品校准
首次使用之前以及每使用 3 个月时,都要按照 USEPA 规定执行一次初级 校准或样品校准。无论使用哪种类型的校准,其步骤完全相同。唯一的区 别在于所用溶液。
初级校准 — 20 NTU 40 NTU 初级标准液(Formazin 悬浮液或 StablCal
注:
USEPA 要求 40 NTU
初级标准液符合方法 2。
样品校准利用浊度值已知的样品进行校准。
制备 4000 NTU Formazin 悬浮液(仅用于初级校准)
警 告
化学品暴露风险。遵守实验室安全规程,穿戴适用于所处理化学品的所 有个人防护装备。有关安全协议,请参阅当前安全数据表 (MSDS/SDS)
需收集的物品:
蒸馏水
容量瓶(4 个)
*
试剂级硫酸肼,(NH2)2H2SO4,5 克
试剂级六亚甲基四胺,(CH2)6N4,50 克
1. 通过 0.2 微米过滤器向每个容量瓶倒入一些蒸馏水。用过滤水清洗每 个容量瓶至少两次
2. 为一个容量瓶加入至少 1000 mL 过滤水。
3. 溶液 1:在第二个容量瓶中用过滤水溶解 5.00 克试剂级硫酸肼。用过
滤水将溶液稀释到 500 mL。轻轻晃动混匀。
4. 溶液 2:在第三个容量瓶中用过滤水溶解 50.00 克试剂级六亚甲基四 胺。用过滤水将溶液稀释到 500 mL。轻轻晃动混匀。
5. 在第四个容量瓶中,混合 50.0 mL 溶液 1 50.0 mL 溶液 2,以制取 4000 NTU 浊度悬浮母液。轻轻晃动混匀。
6. 让混合液在 25 °C ±3 °C (77 °F ±5.4 °F) 下稳定 24 小时。
7. 将悬浮母液倒入琥珀玻璃(或抗紫外线)瓶中存放。悬浮母液可在室温
阴暗处稳定存放长达 1 年。
8. 制取 20 NTU 40 NTU 标准悬浮液。
a. 20 NTU 标准悬浮液:用过滤水将 10 mL 4000 NTU 悬浮母液稀释
2000 mL。轻轻晃动混匀。
b. 40 NTU 标准悬浮液:用过滤水将 20 mL 4000 NTU 悬浮母液稀释
2000 mL。轻轻晃动混匀。
清洗和填充流动槽
1. 清洗流动槽。请参阅 清洗流动槽 第 46。
2. MENU 并选择 CALIBRATE>SENSOR>PRIMARY CAL(初级校准)
SAMPLE CAL(示例校准)。
3. 选择一个选项。
选项 说明
HOLD OUTPUTS(保持 输出)
使模拟输出端和继电器保持在当前数值/开-关状态。
*
为获得最佳结果,需要利用 A 级实验室玻璃器皿制备 Formazin 溶液。
44
中文
选项 说明
XFER OUTPUTS(转换 输出)
转换模拟输出端和继电器以设置数值/状态。请参阅
配置模拟输出 第 42 和 配置继电器 第 43 中的“设置
转换
ACTIVE OUTPUTS(启 用输出)
让模拟输出端和继电器代表所测浊度。
4. ENTER 确认。
5. 将已知数值的样品或制备的标准悬浮液(20 NTU 40 NTU)倒入校
准漏斗中,直到在透明视管中可以看见悬浮液为止。
6. 使校准漏斗低于传感器高度以排放传感器。
7. 再次执行步骤 56 以冲洗传感器的内管。
8. 将已知数值的样品或制备的标准悬浮液(20 NTU 40 NTU)倒入校
准漏斗中,直到流动槽充满悬浮液为止。
9. 将校准漏斗放在漏斗座中。
输入样品或标准悬浮液的数值
1. “PRIMARY (or SAMPLE) CAL, SAMPLE READY?”(主要(或样品) 校准,是否已准备样品?)屏幕中按 ENTER 键。
2. 让读数稳定下来,约需 15 分钟,然后按 ENTER 键。
注: 如果读数不稳定,将显示
"PLEASE WAIT"
(请等候)。
3. 显示 "PRIMARYor SAMPLECAL?"(主要(或样品)校准?)时, 输入所用样品或标准悬浮液的已知数值。按 ENTER 确认。完成校准 时,将显示 "PRIMARY(or SAMPLECAL: CONFIRM CAL OK?"(主 要(或样品)校准:是否确认校准正确?)。
注: 如果显示
“CONFIRM FAILURE”
(校准失败),说明校准未成功。先按
ENTER
键,然后选择
CAL: EXIT
(校准:退出)以取消校准,或选择
CAL: REPEAT
(校
准:重复)重新校准。
4. ENTER 保存校准。随后出现 "Confirm Active?"(是否确认激活?)。
注: 按
ESC
不保存校准。
恢复操作
1. 从漏斗座上取下校准漏斗,放在传感器高度以下,以便排放传感器。按 照当地、州及联邦处理规定丢弃用过的校准溶液。
2. 排放传感器时,将校准漏斗放在漏斗座中。
3. 使样品开始流过流动槽。请参阅开始流动 42
4. ENTER。随后显示 "CONFIRM ACTIVE?"(确认激活?)"Confirm
Active?"(确认激活?)。
5. 当显示的读数与实际的典型工艺值相对应时,按 ENTER 释放模拟输出 端和继电器以启用输出端。
校准模拟输出
如果使用控制器的模拟输出接口,应按需要校准模拟输出。模拟输出经过 出厂校准。模拟输出校准值的调整范围为 ±2 mA
注: 如果模拟输出配置为
0–20 mA
,则对
4 mA 和 20 mA(非 0 mA
)进行校准。
1. MENU 并选择 CALIBRATE>CAL OUTPUTS>CAL OUTPUT 1(或 2)。
2. 选择一个校准选项。
选项 说明
CAL OUT 1 4 mA
设置模拟输出中代表 4 mA 的计数值 — 550910。用一个经 过校准的数字万用表测量控制器模拟输出接口提供的最小实际 值。调整计数值,直到使模拟输出为 4.00 mA
CAL OUT 1 20 mA
设置模拟输出中代表 20 mA 的计数值 — 34603820。用一个 经过校准的数字万用表测量控制器模拟输出接口提供的最大实 际值。调整计数值,直到使模拟输出为 20.00 mA
3. ENTER 确认。
重置校准
将校准设置为出厂默认值:
1. 在测量屏幕中按 MENU(菜单)并选择 TEST/MAINT>RESET CAL(测/维护>重置校准)。
2. ENTER 确认。
注: 按
ESC
取消将校准设为出厂默认值。
中文
45
启用或禁用密码
启用密码以防未授权更改配置设置和校准。密码默认为禁用。密码为 3456,无法更改。
1. MENU(菜单)并选择 CONFIGURE>SET PASSCODE(配置< 码)。
2. UP(向上)或 DOWN(向下)箭头键选择选项。
选项 说明
禁用 无需输入密码便能更改配置设置、进行校准或进入 TEST/MAINT(测
/维护)屏幕。
启用 除非输入密码,否则只能查看配置设置但不能更改。输入密码才能显示
CALIBRATE(校准)和 TEST/MAINT(测试/维护)屏幕。
3. ENTER 确认。
维护
警 告
多种危险。只有合规的专业人员才能从事文件中本部分所述的任务。
清洁仪器
使用湿布和温和的肥皂液清洁仪器的外部,然后擦干。
清洗流动槽
清洗流动槽,以除去流动槽中的异物(如灰尘、沙粒等)。
1. 排空流动槽。
a. 将内部 3 通阀手柄转到 CALIBRATE 位置。 b. 从漏斗座上取下漏斗,并使其低于传感器高度,以排放传感器中残
留的任何样品。
c. 排放传感器之后,将漏斗放在漏斗座中。
2. 拆下流动槽盖。逆时针转动腔盖 1/8 圈解锁,然后提起腔盖。
3. 从流动槽中拆下挡板组件。挡板组件通过槽和导肋键入左侧,只能单
向配合在流动槽中。
4. 用一块干净的软湿布擦净流动槽的内表面以及光源和检测器的窗口。
5. 用干净的水冲洗挡板组件。
6. 将挡板组件安装在流动槽中。
7. 确保流动槽盖中的 O 型圈涂上有机硅润滑剂。
8. 将流动槽盖安装在流动槽上。顺时针转动腔盖 1/8 圈安装到位。
更换保险丝
危 险
电击致命危险。断开仪器电源以更换保险丝。拔下电源线或关闭面板上 的断路器。
危 险
火灾危险。使用类型和额定电流相同的保险丝对原来的保险丝进行更 换。
1. 切断控制器电源之后,打开控制器。请参阅总布线流程 35
2. 取下熔断的保险丝,更换新保险丝。请参阅 5 35 了解保险丝的位
置。请参阅更换部件与附件 49 了解备用保险丝套件。
3. 盖上控制器。拧紧四个螺钉。
更换光源或检测器
危 险
电击致命危险。在开始此维护任务前,断开仪器的所有电源和继电器连 接。
请参阅新的光源或检测器随附的更换说明书。安装新的光源或检测器之 前,确保在其 O 型圈上涂上有机硅润滑剂。
更换光源或检测器之后,校准仪器。请参阅初级校准和样品校准 44
注: 制造商建议同时更换两个光源(或两个检测器)。
46
中文
停止操作
长期存放时,需要使仪器停止操作。
1. 清洗流动槽。请参阅清洗流动槽 46
2. 用一块清洁、干燥的软布擦干流动槽。
3. 切断控制器的电源。
故障排除
当系统诊断发现非典型状况时,测量屏幕会闪烁显示警告检查状态。 按 MENU(菜单)并选择 TEST/MAINT>STATUS(测试/维护>状态)以
查看系统状态消息。重复按 ENTER 滚动显示状态消息。请参阅 5 了解 状态消息的描述。
如果下列故障排除步骤无法解决问题,请联系技术支持人员。
5 状态消息
信息 说明 解决方案
ANALYZER OK(分 析仪正常)
控制器运行正确。
ANALYZER FAIL
(分析仪出故障)
控制器运行不正确。请联系技术支持部门。
Source 1 (or 2) OK
(光源 1(或 2)正
常)
光源运行正确。
SOURCE 1 (or 2) FAIL(光源 1(或 2)
出故障)
光源运行不正确。 确保光源正确接线。必要时更换光源。
请参阅更换光源或检测器 46
DETECT 1 (or 2) OK(检测器 1(或 2)
正常)
检测器运行正确。
DETECT 1 (or 2) FAIL(检测器 1(或 2)出故障)
检测器运行不正确。确保检测器正确接线。必要时更换检测
器。请参阅更换光源或检测器 46
5 状态消息 (续)
信息 说明 解决方案
CHAMBER OK(流 动槽正常)
传感器运行正确。
CHAMBER DIRTY
(流动槽变脏)
流动槽含有异物。 清洗流动槽。请参阅清洗流动槽
46
CHAMBER UNKNOWN(流动槽
未知)
传感器电缆未接 线,或是未正确连 接到控制器。
确保传感器电缆正确接线。必要时断开 传感器电缆,用一个欧姆计检查电缆的 连续性、内部是否短路。
光源或检测器运 行不正确。
确保光源和检测器正确接线。如果光源 或检测器出现故障,进行更换。请参阅
更换光源或检测器 第 46。
测量信号包括电 气噪声。
确保传感器电缆远离可能产生电场或磁 场的电动机或其他设备。
RLY (A, B, C, D) ACTIVE(继电器
ABCD)启
用)
继电器通电。 控制继电器:所测
浊度高于设定值。 报警继电器:所测
浊度高于高警报 设定值,或低于低 警报设定值。
状态继电器:已发 现一个选定的系 统诊断状况。
RLY (A, B, C, D) INACTIVE(继电器
ABCD)禁
用)
继电器断电。 控制继电器:所测
浊度低于设定值。 报警继电器:所测
浊度介于高警报 与低警报设定值 之间。
状态继电器:未发 现选定的系统诊 断状况。
中文
47
5 状态消息 (续)
信息 说明 解决方案
RLY (A, B, C, D) TIMEOUT(继电器
ABCD)超
时)
继电器断电。 控制继电器:过载
定时器已超时,必 须手动重置。
MENU(菜单)并选择
TEST/MAINT>OVERFEED RESET
(测试/维护>过载重置)以重置过载定时
器。
RLY (A, B, C, D) COUNTING(继电器
ABCD)正在
计数)
继电器通电。 控制继电器:过载
定时器正在倒计 时,但未超时。
系统诊断
1. 在测量屏幕中按 MENU(菜单)并选择 TEST/MAINT(测试/维护)。
2. 选择一个诊断选项。
选项 说明
CHECK CAL/HOLD
(检查校准/保持)
使模拟输出端和继电器在当前数值/开-关状态下保持最 长 30 分钟。
OUTPUT 1 (or 2)(输 出端 1(或 2))
设定模拟输出端输出的模拟值 — 0(或 4)至 20 mA
(默认 = 4.05 mA)。该选项可用来确认所连装置的操
作。
RELAY (A, B, C, D)
(继电器(ABC
D))
设定继电器的状态通电或断电。在常开和常闭继电 器输出端上利用一个导通测试计确认所选继电器的状 态。
SIM SENSORSIM 传感器)
设定所模拟的浊度测量值 — 0.00 100.00(默认 =
0.35 NTU)。该选项用来确认模拟输出端和继电器是 否正确响应所模拟的浊度测量值。
3. 按 ENTERESC 停止测试。
除去流动槽中的气泡
样品中会出现两种类型的气泡:
一种是不可见气泡 - 因液体吸收空气而形成。这属于自然情况,或是人 为干预所致,比如向水中注入臭氧作为消毒剂。样品中形成的不可见气 泡在流动槽中可见。
另一种是可见气泡 - 因气液混合所致。比如水流经含有空气的管道时或 水降落时。
降低传感器的出口流量
消除样品中不可见但是流动槽中可见的气泡。为此需要降低传感器的出口 流量,以便在测量样品时对样品加压。
1. 使流体通过传感器时的流速保持在指定限值内。请参阅规格 29
2. 将一个限流器安装在传感器出口管中。
3. 调整限流器,直到使流量略微降低。
4. 关闭三通阀截断流向传感器的流体。
5. 排空流动槽。
a. 将内部 3 通阀手柄转到 CALIBRATE(校准)位置。 b. 从漏斗座上取下漏斗,并使其低于传感器高度,以排放传感器中残
留的任何样品。
c. 排放传感器之后,将漏斗放在漏斗座中。
6. 缓慢打开三通阀,使流动槽充满样品。
使用外部除泡器
利用外部除泡器除去样品中可见的气泡。请参阅更换部件与附件 49。 制造商建议为高于 1.9 LPM (0.5 GPM) 的流速使用一个外部除泡器。
1. 将外部除泡器垂直安装在传感器的上游(向上排气)。请参阅外部除泡
器随附的说明书。
2. 确保排气孔畅通。关闭排气孔时,气泡会留在液体中。
3. 定期取下除泡器排放塞,以便排出所收集的任何沉积淤泥。
48
中文
更换部件与附件
警 告
人身伤害危险。使用未经批准的部件可能造成人身伤害、仪器损坏或设 备故障。本部分中的更换部件均经过制造商的批准。
注: 一些销售地区的产品和物品数量可能有所不同。请与相关分销商联系或参阅公司 网站上的联系信息。
更换部件
说明 数量 物品编号
T53 型控制器:
保险丝套件(一根 80 mA 保险丝和一根 100 mA 保险丝)
1 1000G3315-101
安装件套件,包括:
1 1000G3228-101
平垫圈,¼ in ID
8 8H1346
地脚螺钉,绿色
1 8H1275-038
六角螺母,M6
8 8H1116
固定支架
1
1000C4F3217-101
固定支脚
2 1000B4F3222
盘头螺钉,M6 x 1.0 x 20 mm
4 8H1216-075
管装用盘头螺钉,M6 x 1.0 x 100 mm
4 8H1217-400
板装用盘头螺钉,M6 x 1.0 x 150 mm
4 8H1217-600
面板安装的密封圈,氯丁橡胶
1
1000A4F3249-101
8320 型浊度传感器:
挡板组件
1 8220G1060
检测器组件,带一套 O 型圈,含两个检测器
1 8220G1330-204
更换部件与附件 (续)
说明 数量 物品编号
流动槽盖
1 8220C2F1076
传感器内部三通阀,带连接管
1 8320A1074-001
光源组件,带一套 O 型圈,含两个光源
1 8220G1320-204
流动槽 O 型圈
1 5H1288
光源或检测器 O 型圈
1 5H1251
光源或检测器固定夹
1 8220F1078
配件
说明 数量 物品编号
8320 型浊度传感器:
头部调节器除泡器
1 4668000
有机硅润滑剂
2 g 6M0920
容量瓶,A 级,2 L
1 2489454
Formazin, 4000 NTU 500 mL 2461-49
StablCal, 40 NTU 1L 27463-53
StablCal, 40 NTU
1 加仑
27463-56
Tensette,吸量管,1–10 mL
1 1970010
中文
49
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
©
Hach Company/Hach Lange GmbH, 2013. All rights reserved. Printed in U.S.A.
Loading...