Specifications are subject to change without notice.
SpecificationDetails
Dimensions, 9525 Degas System (W x H x D)559 x 813 x 305 mm (22 x 32 x 12 in.)
Dimensions, 9525 DCCP System (W x H x D) (Optional)826 x 813 x 241 mm (32.5 x 32 x 9.5 in.)
Weight36 kg (80 lb)
Power requirementsUS: 110 to 120 VAC, 1 phase, 50/60 Hz
Main supply voltage fluctuation±10% of nominal voltage
Power consumption1.6 kVA
Pollution Degree2
Installation CategoryII
Altitude RequirementsStandard 2000 m (6,562 ft.)
Recommended operating temperature23 to 27 °C (75 to 80 °F)
Ambient temperature range2 to 50 °C (36 to 122 °F)
Storage temperature0 to 49 °C (32 to 120 °F)
Relative humidity0% to 85% RH
8315 Conductivity sensorCell constant, k: 0.01 cm
sc200/9500 Controller power requirements100 to 240 VAC ±10%, 50/60 Hz
Controller enclosure ratingIP66/NEMA 4X
Sample flow100 to 150 mL/min; 6 to 9 L/hr (1.5 to 2.4 gph)
Sample inlet/outlet connections1/4 in. tube fitting
Cooling water consumption3.8 to 11.4 L/min (1 to 3 gpm)
Degas system certificationsCE, UL, CSA
EU: 220 to 240 VAC, 1 phase, 50/60 Hz
-1
Measuring range: 0.01 to 200 μS/cm
Sensor accuracy: <2%
English 3
Section 3 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
3.1 Safety information
N O T IC E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
3.1.1 Use of hazard information
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
3.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public
disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to
the user.
D A N GE R
W A R NI N G
C A U TI O N
N O T IC E
4English
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety
information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol indicates the need for protective eye wear.
This symbol indicates that the marked item can be hot and should not be touched without care.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
3.1.3 Certification
C A U TI O N
This equipment is not intended for use in residential environments and may not provide adequate protection to
radio reception in such environments.
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing
Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC
Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment. This equipment has been tested
and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the user will be required to
correct the interference at their expense. The following techniques can be used to reduce
interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the
interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect
the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
3.2 Product overview
The 9525sc degas system removes CO2 and measures degassed conductivity from the sample
stream. The 9525sc degas system boils the sample and distills the liquid through the crossover
system to remove CO2 gas from the sample stream. The unit removes CO2 gas through the steam
vent, and distills the flow through the sample cooler. The unit then measures the conductivity of the
sample.
Use the 9523sc cation conductivity panel to measure specific, cation and degassed conductivity. The
combined system is known as the 9525sc degas cation conductivity (DCCP) system. Refer to
English
5
Figure 1. Refer to the 9523sc user instructions to add the 9523sc cation conductivity panel to the
degas system.
Figure 1 Product overview
1 Temperature controller power switch7 Sample cooler
2 Power indicator light8 Heater
3 Degas (boiler) condenser9 Heat shield
4 Hach sc200 or Polymetron 9500 controller10 Temperature Controller
5 8315 conductivity sensor11 Electrical enclosure
6 Flow chamber
3.3 Product components
Make sure that all components have been received. Refer to Figure 2. If any items are missing or
damaged, contact the manufacturer or a sales representative immediately.
6
English
Figure 2 Product components
1 Sensor, controller and 9525 documentation3 9525sc Degas system assembly
2 Ferrite collar (used with the 220 volt unit only)
3.4 Customer-supplied parts
Before installation, make sure that all of the components that follow are available. Refer to Figure 3:
English
7
Figure 3 Customer-supplied parts
1 Sample plumbing parts7 3/8 in. NPT tube fittings (x2)
2 Needle valve/rotameter8 Tubing (align the size of the inlet/outlet fittings)
3 Tubing9 Electrical parts
4 Shutoff valve10 Electrical cable, 3.31 mm2 (12 AWG) (US)
5 Cooling water plumbing parts11 Electrical cable, 2.08 mm2 (14 AWG) (EU)
6 Shutoff valves (x2)
Section 4 Installation
D A N GE R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
4.1 Mechanical installation
W A R NI N G
Personal injury hazard. Instruments or components are heavy. Use assistance to install or move.
Select an installation location. Most panels are installed on a metal framing system or equivalent
mounting structures. Before installation, examine the location. The location must have the items that
follow:
• Sufficient structural integrity in the floor and wall
• A sufficient anchor location
• Sufficient operation and maintenance space. Refer to Figure 4
• Cooling water available
• Access to a floor drain
• A flat installation surface at the correct height
• Sufficient space above the unit to prevent condensation above the steam vent
• Sufficient light for operation and maintenance
For optimum equipment installation, the location must have the properties that follow:
• No mechanical vibration
8
English
• No media that can cause corrosion
• No strong heat radiation
• An ambient temperature near 23 to 27 °C (73 to 80 °F) with minimal variation
Use sufficient support to keep gravity and thermal loads on the connections and vibration to a
minimum. Refer to Figure 4.
Figure 4 Mounting dimensions
4.2 Electrical installation
D A N GE R
Electrocution hazard. Always remove power to the instrument before making electrical connections.
Supply power to the instrument with conduit or a power cable. Make sure that a circuit breaker with
sufficient current capacity is installed in the power line. The circuit breaker size is based on the wire
gauge used for installation.
For installation with conduit:
• Install a local disconnect for the instrument within 3 m (10 ft) of the instrument. Put a label on the
disconnect that identifies it as the main disconnect device for the instrument.
English
9
• Make sure that the power and safety ground service drops for the instrument are a minimum of
3.31 mm2 (12 AWG) (for 110 V) and a minimum of 2.08 mm2 (14 AWG) (for 220 V) and the wire
insulation is rated for 300 VAC or higher and 60 °C (140 °F) minimum.
• Connect equipment in accordance with local, state or national electrical codes.
• Connect the conduit through a conduit hub that holds the conduit securely and seals the enclosure
when tightened.
• If metal conduit is used, make sure that the conduit hub is tightened so that the conduit hub
connects the metal conduit to safety ground.
For installation with a power cable, make sure that the power cable is:
• Less than 3 m (10 ft) in length
• Rated sufficient for the supply voltage and current
• Rated for a minimum of 60 °C (140 °F) and applicable to the installation environment
• Not less than 3.31 mm2 (12 AWG) with applicable insulation colors for local code requirements
• A power cable with a three-prong plug (with ground connection) that is applicable to the supply
connection
• Connected through a PG11 or 3/4 inch cable gland (strain relief) that holds the power cable
securely and seals the enclosure when tightened
• Does not have a locking type device on the plug
• Near to the instrument with easy access by personnel
Table 1 Wiring information—AC power
TerminalDescriptionColor—North AmericaColor—EU
LHot or phase (L)BlackBrown
NNeutral (N)WhiteBlue
GProtective earth ground
(G)
GreenYellow with green stripe
Connect conduit or a power cord
Connect the controller to power with conduit or a power cord. Make connections at the same
terminals, regardless of the type of wire used. Refer to Table 1 to connect conduit or a power cord.
Put each wire into the applicable terminal until the insulation is flush against the connector with no
bare wire seen. Pull lightly after insertion to make sure that there is a tight connection. Refer to
Figure 5.
1. Loosen the four captive screws.
2. Carefully open the electrical cover.
3. Use a knockout punch driver to punch a hole in the top of the NEMA electrical enclosure.
4. Attach a power cord strain relief connector to the hole.
5. Attach the power cord to the strain relief connector.
6. Connect the live (L), neutral (N), and ground (G) wires as shown in Figure 5.
7. Close the electrical cover.
8. Tighten the screws.
10
English
Figure 5 Connect electrical wiring
4.3 Plumbing
4.3.1 Connect the plumbing
Note: Clean and deburr all of the connecting pipe and tubing before installing.
4.3.2 Connect the cooling water
Note: Adjust the cooling water flow at the cooler outlet.
The cooling water must have the properties that follow:
• An approximate flow rate of 3.78 liters/minute (1 gallon/minute)
English
11
• An approximate temperature of 25 °C (77 °F)
• Clean with no suspended solids and a low hardness
• Low in chlorides because chlorides can cause corrosion in the stainless steel coil and shell
Connect the cooling water as follows:
1. Use a 3/8-inch NPT (national pipe thread) connection to connect the tubing to the cooling water
outlet at the top of the cooler. Refer to Figure 6. Make sure that the outlet drains into a floor drain.
2. Use a 3/8-inch NPT connection to connect tubing to the cooling water inlet at the bottom of the
cooler. As an alternative, connect the tubing to the cooling water inlet at the top of the cooler.
Figure 6 Cooling water plumbing
1 Cooling water inlet (alternative)3 Shutoff valve
2 Cooling water outlet4 Cooling water inlet
4.3.3 Connect the sample lines
W A R NI N G
Burn hazard. Obey safe handling protocols during contact with steam.
Make sure that the samples supplied to the instrument are at a temperature of 25 °C ± 1 °C (77 °F ±
2 °F). Make sure the pressure is controlled by a back pressure regulator or a head cup to keep a
constant sample flow rate.
Install the needle valve/rotameter on the sample inlet before the degas unit. If the 9523sc cation
conductivity panel is used, install the needle valve/rotameter on the sample inlet before the cation
conductivity panel.
Install the flow cells in an orientation that keeps the water level near the crossover section of the
degas chamber. The sample outlet from the panel must be level with the crossover section of the
degas chamber.
Incorrect release of steam from the degas chamber or incorrect flow cell locations can have an
unwanted effect on the performance of the system.
12
English
Complete the steps that follow to connect the sample line. Refer to Figure 7.
1. Use a 1/4-inch NPT connection to connect the sample line to the sample inlet.
2. Use a 1/4-inch NPT connection to connect the sample outlet line to the sample outlet. The outlet
must drain into a floor drain.
3. Unroll the condensate drain tube.
4. Push the condensate drain tube through the hole in the rear of the unit. Make sure that the outlet
drains into a floor drain.
5. Remove the red cap from the steam vent.
6. Remove the red cap from the vacuum break.
7. Make sure that the tubing does not touch the hot surfaces.
Burn hazard. The heater surface and surrounding area gets hot. Do not touch.
W A R NI N G
Burn hazard. Obey safe handling protocols during contact with hot liquids.
14 English
Notes:
• The power indicator light will start and stop with the heater.
• For units installed in a small shelter or prefabricated structure, it is recommended that the steam
vent is pointed to the outer side of the structure.
The heater temperature is factory-programmed at 108 °C (226 °F) at the factory and the sample flow
rate must be between 100 and 150 ml/minute (6 and 9 L/hr). Decrease the flow rate if there is no
vapor at the steam vent. Droplets of hot boiling water at the steam vent shows that the distilled water
is too hot. If the distilled water becomes too hot, decrease the set temperature or increase the flow
rate within the specified range.
Refer to Figure 9 on page 16.
1. Set the water flow to the cooler to on.
2. Examine cooler for leaks at the two cooler inlets and the outlet.
3. Examine the outlet tube for correct drainage.
4. Set the sample flow to on.
5. Adjust the sample flow to 100 to 150 ml/minute.
6. Look for leaks and for correct drainage.
7. Set the temperature controller power switch to on. Refer to Figure 8. The power light comes on.
8. Make sure that the display on the controller is on.
9. Monitor the temperature controller. The temperature will increase continuously to approximately
106 °C.
Note: The temperature controller may not reach 106 °C, based on local air pressure and
temperature. The heater will heat the sample to the boiling point of water. If there is a constant
column of water vapor from the steam vent, the unit is operating correctly.
Figure 8 Set the temperature controller to on
English 15
Section 6 Operation
6.1 User interface—Temperature controller
Figure 9 Temperature controller
1 Left display5 Up arrow key
2 Output activity6 Infinity key
3 Right display7 Advance key
4 Down arrow key
When flow to the unit is stopped, the heater will become too hot. An alarm prevents the heater
operation until the user restores flow and stops the alarm on the controller. Push to enter the
alarm acknowledgement mode. Use and to scroll through the alarm responses. Select "CLr"
to stop the alarm.
6.1.1 Configure the heater alarm
Set the flow through the unit to a range of 100 to 150 ml/minute (6 to 9 L/hr) to make sure that the
heater stays filled with water. Set the flow before power is applied to the unit to prevent damage to
the heater. If cooling water is not supplied to the instrument, remove all cells downstream of the
heater and flow cells. For a description of the temperature controller, refer to Table 2.
By default, the temperature meter is set to show the actual temperature on the left display and the
maximum temperature on the right display. Do not change the factory-set defaults. Refer to the
Watlow EZ-ZONE PM Express User's Guide (available online) for all of the user instructions.
16
English
Table 2 Temperature controller
1 2
Display/KeyNameDescription
Left displayShows the actual temperature.
In the Operations or Setup menu,
shows the process value or set
point data.
Output 1 and Output 2, and Right
display
Up/Down arrow keysSelects new data when Advance
Infinity keyStops alarms.
Advance keyMoves forward through parameter
Output 1 and 2 shows which output
is set to on.
Right display shows the maximum
temperature. In the Operations or
Setup menu, shows the menu
display.
key is pushed.
Moves forward or backward through
software menus and parameters.
Starts and stops the timer.
Push to back up one level.
Push and hold for two seconds to
go back to the Operations menu.
prompts.
6.1.2 Configure the control mode
Push to scroll through the options on the Operations menu. Push to go back to the default
display. For a description of the Operation menu options, refer to Table 3.
To navigate the Operations menu, follow these steps:
1. Push to select a menu.
2. Push or to find an option.
3. Push to select the option.
4. Push and hold for two seconds to go back to the default display.
Table 3 Operations menu options
Menu optionNameDescription
Control modeShows the current control mode:
Off, Auto or Manual.
C.M1
Heat powerShows the current heat output level.
h.Pr1
English 17
Table 3 Operations menu options (continued)
Menu optionNameDescription
AutotuneShows the current status of
Aut1
Idle set pointConfigures a set point that can be
idS1
Autotune: No or Yes.
triggered by an event state.
6.1.3 Configure the setup mode
Push to scroll through the options on the Setup menu. Anywhere in the Setup menu, push and
hold for two seconds to go back to the default display. For a description of the Setup menu
options, refer to Table 4.
To navigate the Setup menu, follow these steps:
1. Push and hold or to show the Setup menu.
2. Push to select a menu.
3. Push and hold for two seconds to go back to the default display.
Table 4 Setup menu options
Menu optionNameDescription
OperationsNavigates to the operations menu.
oPEr
Ain
i.Er
i.CA
18English
Analog input value Shows the process value.
Input error latching Sets error latching to on or off. If latching is on, the user must
manually clear the errors.
Calibration offsetOffsets the input indication to adjust for lead wire resistance or
other factors that cause the input indication to change from the
actual process value.
Section 7 Maintenance
7.1 Prepare for storage
C A U TI O N
Burn hazard. Turn off power and allow the instrument to cool before this procedure.
D A N GE R
Electrocution hazard. Remove power from the instrument before doing maintenance or service
activities.
To prepare for storage or set the system to off, refer to Figure 10.
1. Set the heater to off.
2. Remove power from the heater.
3. Close the sample inlet valve to stop the sample flow.
4. Close the cooling water inlet valve to stop the water flow to the cooler.
5. Close the cooling water outlet isolation valve.
Figure 10 Set the system to off
English 19
7.2 Open the electrical enclosure
D A N GE R
Electrocution hazard. Remove power from the instrument before doing maintenance or service
activities.
To open the electrical enclosure, refer to Figure 11:
1. Loosen the four captive screws on the electrical enclosure.
2. Carefully open the electrical cover.
Figure 11 Open electrical enclosure
20 English
7.3 Look for leaks
Fire hazard. Use the same type and current rating to replace fuses.
Electrocution hazard. Always remove power to the instrument before making electrical connections.
Burn hazard. Obey safe handling protocols during contact with hot liquids.
Burn hazard. The heater surface and surrounding area gets hot. Do not touch.
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
D A N GE R
D A N GE R
W A R NI N G
W A R NI N G
C A U TI O N
W A R NI N G
Personal injury hazard. The sample line contains water under high water pressure that can burn skin if
hot. Qualified personnel must remove the water pressure and wear personal protective equipment
during this procedure.
Before each use, examine all of the connectors, valves, and other components for leaks. Repair
leaks as necessary.
7.4 Replace a fuse
Pre-requisites:
• Refer to the steps in Prepare for storage on page 19 to prepare for the maintenance procedure.
• Refer to the steps in Open the electrical enclosure on page 20 to open the electrical enclosure.
D A N GE R
Fire hazard. Use the same type and current rating to replace fuses.
D A N GE R
Electrocution hazard. Remove power from the instrument before doing maintenance or service
activities.
English21
Replace a fuse when it is blown. Refer to Figure 12.
Figure 12 Replace a fuse
7.5 Clean the analyzer electrodes
Regularly clean or replace the analyzer electrodes. Replace the electrodes that show low
performance. Refer to the Hach 8315 Polymetron sensor documentation for instructions.
7.6 Replace the heater
D A N GE R
Electrocution hazard. Remove power from the instrument before doing maintenance or service
activities.
C A U TI O N
Burn hazard. Turn off power and allow the instrument to cool before this procedure.
To replace the heater, refer to Figure 13 and the steps that follow:
22
English
Figure 13 Electrical wiring diagram
1 0.33 mm2 (22 AWG) yellow temperature controller
wire
2 3.31 mm2 (12 AWG) (US) or 2.08 mm2 (14 AWG)
(EU) red neutral (N) power connection wire
3 3.31 mm2 (12 AWG) (US) or 2.08 mm2 (14 AWG)
(EU) red SSR-1 power terminal wire
4 Zip ties9 0.33 mm2 (22 AWG) red temperature controller wire
5 Electrical enclosure
6 Ground (green) wire
7 Heater unit
8 Ferrite collars (used with the 220 volt unit only)
1. Prepare for the maintenance procedure. Refer to the steps in Prepare for storage on page 19.
2. Open the electrical enclosure. Refer to the steps in Open the electrical enclosure on page 20.
3. Carefully cut the zip ties that hold the four heater wires in the electrical enclosure. Refer to
Figure 13.
4. Disconnect the two red 3.31 mm2 (12 AWG) (US) or 2.08 mm2 (14 AWG) (EU) wires from the
SSR-1 power terminal and the neutral (N) power connection. Refer to Figure 13.
5. Disconnect the red and yellow 0.33 mm2 (22 AWG) wires from the temperature controller. Refer
to Figure 13.
6. Loosen the two screws that attach the heat shield to the unit. Refer to Figure 14.
7. Carefully lift the heat shield off the unit. Refer to Figure 14.
8. Remove the four screws from the bottom of the heater. Remove the bottom heater cover. Refer to
Figure 14.
9. Remove the electrical retainer nut in the heater housing. Refer to Figure 14.
10. Loosen the screw and remove the ground (green) wire in the heater housing. Refer to Figure 13.
English
23
11. Use a wrench to open the pipe union at the bottom of the reboil chamber. Refer to Figure 14.
Note: There can be water in the pipe.
12. Carefully remove the heater from the unit.
13. Put the new heater on the reboil chamber.
14. Use a wrench to close the pipe union at the bottom of the reboil chamber. Refer to Figure 14.
15. Remove the four screws from the bottom of the new heater. Remove the bottom heater cover.
Refer to Figure 14.
16. For the 220-volt unit, wind the yellow and red 0.33 mm2 (22 AWG) wires around a ferrite bead
three times. Refer to Figure 13.
17. For the 220-volt unit, wind each red 2.08 mm2 (14 AWG) (EU) wire around a ferrite bead three
times. Refer to Figure 13.
18. Push the yellow and red wires through the electrical retainer nut. Refer to Figure 14.
19. Carefully push the yellow and red wires up through the electrical conduit. Use an electrician draw
wire if necessary.
20. Cut the wires to the correct length.
21. Carefully push the green (ground) wire down through the electrical conduit. Refer to Figure 14.
Use an electrician draw wire if necessary.
22. Attach the green (ground) wire to one of the screws in the heater unit. Refer to Figure 13.
23. Put the bottom heater cover on. Tighten the four screws. Refer to Figure 14.
24. Connect the four heater wires in the electrical enclosure. Refer to Figure 13.
25. Close the electrical enclosure cover.
26. Tighten the screws on the electrical enclosure.
27. Connect the sample lines. Refer to Connect the sample lines on page 12.
28. Put the heat shield on the unit. Tighten the two screws. Refer to Figure 14.
29. Set the unit to on. Refer to the steps in Startup on page 14.
Burn hazard. Turn off power and allow the instrument to cool before this procedure.
To replace the sample cooler, refer to Figure 15 and the steps that follow:
1. Prepare for the maintenance procedure. Refer to the steps in Prepare for storage on page 19.
2. Remove the tubing from the cooling water outlet at the top of the cooler.
3. Remove the tubing from the cooling water inlet at the top or bottom of the cooler.
4. Loosen the sample inlet and outlet tube fittings above the cooler.
5. Remove the mounting hardware on the rear of the unit.
English
25
6. Carefully remove the cooler from the panel.
7. Put the new cooler on the panel.
8. Attach and tighten the mounting hardware on the rear of the unit. Use a wrench to tighten the
nuts.
9. Attach tube fittings above the cooler.
10. Refer to the steps in Connect the cooling water on page 11 to connect the cooling water.
11. Refer to the steps in Startup on page 14 to set the unit to on.
Figure 15 Replace the cooler
1 Sample inlet tube fittings4 Cool water inlet tube and fitting
2 Cool water outlet tube and fitting5 Mounting hardware
3 Sample cooler6 Sample outlet tube fittings
7.8 Replace the temperature controller
D A N GE R
Electrocution hazard. Remove power from the instrument before doing maintenance or service
activities.
To replace the temperature controller, refer to Figure 16 and the steps that follow:
26
English
1. Complete the steps in Prepare for storage on page 19.
2. Open the electrical enclosure. Refer to the steps in Open the electrical enclosure on page 20.
3. Disconnect the two temperature controller connectors.
4. On the left and right sides of the temperature controller, disengage the two tabs.
5. Hold the temperature controller above and below the front. Pull the unit straight out.
6. Push the new temperature controller into the controller housing.
7. Connect the two temperature controller connectors.
8. On the left and right sides of the temperature controller, connect the two tabs.
9. Close the electrical enclosure cover.
10. Tighten the screws on the electrical enclosure.
11. Set the unit to on. Complete the steps in Startup on page 14.
Figure 16 Replace the temperature controller
English 27
Section 8 Troubleshooting
ProblemPossible causeSolution
High temp alarm is active.There is no flow or the flow is less
The instrument does not accept
temperature settings.
Sample heater is not heating.The fuse is blown.1. Examine the continuity of the
Unit is heating but cannot reach
boiling point.
Temperature will not increase to
106 °C.
Temperature meter will not come
on.
than 100 ml/minute.
The settings are corrupted.1. The correct settings are
The heater is defective.Replace the heater on page 22.
The flow rate is too high.Adjust the flow rate between
The heater is defective.Replace the heater on page 22.
The boiling point of water is
dependant on the atmospheric
pressure.
The fuse is blown.Examine the continuity of the fuses.
1. Examine the flow rate. Make sure
that the flow is more than
100 ml/minute.
2. Remove power to the heater,
then set the power switch to on after
the flow is restored.
3. Make sure that the temperature
set point is between 106 and 108 °C
(223 and 226 °F)
programmed at the factory. If issues
continue, make sure that the
settings are correct.
2. If the settings are corrupted, send
the temperature controller back to
the factory.
fuses. Replace a fuse on page 21 if
blown.
2. If fuses continue to blow, the
heater likely has a short and must
be replaced.
100 and 150 ml/minute.
The heater increases the
temperature of the sample to the
boiling point of water, based on
local air pressure and temperature.
If there is a constant column of
water vapor from the steam vent,
the unit is operating correctly.
Replace a fuse on page 21 if blown.
Section 9 Replacement parts and accessories
W A R NI N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Instruments
DescriptionItem no.
Hach 9525sc system9525.99.0010
Polymetron 9525sc system9525.99.0020
28 English
DescriptionItem no.
Hach 9525sc DCCP system9525.99.10CC
Polymetron 9525sc DCCP system9525.99.20CC
Replacement parts
DescriptionItem no.
Degas system heater, 120 V9525.99.7030
Degas system heater, 240 V9525.99.7031
Fuse, temperature controller, 1/2 A (FU-1)9525.99.7071
Fuse, heater 120 V, 15 A (FU-2)9525.99.7070
Fuse, heater 240 V, 8 A (FU-2)9525.99.7072
Temperature controller9525.99.7050
Degas system sample cooler9525.99.7060
Degas system extension panel kit9525.99.7040
Accessories
DescriptionItem no.
Flow meter with 1/4 in. OD tubing4643600
Flow meter, 0 to 16 L/hr with adjustable valve, 9186 694=000=001
English 29
Kapitel 1 Inhaltsverzeichnis
Technische Daten auf Seite 30
Allgemeine Informationen auf Seite 31
Installation auf Seite 36
Inbetriebnahme auf Seite 42
Betrieb auf Seite 44
Wartung auf Seite 47
Fehlerbehebung auf Seite 58
Ersatzteile und Zubehör auf Seite 58
Kapitel 2 Technische Daten
Änderungen vorbehalten.
Technische DatenDetails
Abmessungen , 9525 Entgasungssystem (B x H x T)559 mm x 813 mm x 305 mm (22 x 32 x 12 Zoll)
Abmessungen, 9525 DCCP-System (B x H x T) (optional)826 mm x 813 mm x 241 mm (32.5 x 32 x 9.5 Zoll)
Gewicht36 kg (80 Pfund)
StromversorgungUS: 110 bis 120 VAC, 1 Phase, 50/60 Hz
Spannungsschwankung der Hauptversorgung±10 % der Nennspannung