330 days typical with a 15 minute recording interval, 1 level and 1 velocity, data download once
per week, 10 °C (50 °F) (also affected by site conditions)
User InterfaceOptically Isolated Interface to IBM compatible PC
Air IntakeAtmospheric pressure reference is desiccant protected.
Operating Temperature0 to 70 °C (32 to 158 °F)
Zero Stability: 0.015 m/s (<0.05 ft/s)
Accuracy: ± 2% of reading
1
Recommended Range: –1.52 to 6.10 m/s (–5 to 20 ft/s)
Method: Pressure transducer with stainless steel diaphragm
Accuracy (static)
2
:
±0.16% full scale ±1.5% of reading at constant temp (± 2.5 °C)
±0.20% full scale ±1.75% of reading from 0 to 30 °C (32 to 86 °F)
±0.25% full scale ±2.1% of reading from 0 to 70 °C (32 to 158 °F)
Velocity-Induced Depth Error: Compensated based on pipe geometry and flow velocity.
Depth Range:
Standard: 0–3 m (0–10 ft); Extended: 0–9 m (0–30 ft)
Specifications
Depth Compensated
Temperature
MaterialNoryl
0 to 70 °C (32 to 158 °F)
®
plastic outer shell with epoxy potting within
Power Consumption1.2 W @ 12 VDC
Cable MaterialUrethane sensor cable with air vent
Cable Lengths Available
Standard: 9, 15, 23, and 30.5 m (30, 50, 75, and 100 ft)
Custom: 30.75-76 m (101-250 ft) maximum
Cable Diameter0.91 cm (0.36 in.)
Dimensions2.3 cm H x 3.8 cm W x 13.5 cm L (0.9 in. H x 1.5 in. W x 5.31 in. L)
In-Pipe Ultrasonic Sensor
Operating Frequency75 kHz
Accuracy
Range
±0.027 ft for sensor to liquid distance between 2.5 inches and 5 ft at ±1 ft change in head from
calibration point, 20 °C still air, ideal target, 25 ft cable.
Distance from sensor to liquid: 0.2 inches (minimum) to 5 feet (maximum), @ 20 °C still air,
ideal target, 25 ft cable.
Resolution0.019 cm (0.0075 in.)
Operating Temperature
Range
–20 to 60 °C (–4 to 140 °F)
Storage Temperature–20 to 60 °C (–4 to 140 °F)
Temperature Error0.00005 meter/°C typical
MaterialStat-Kon A-E ABS Plastic
Cable Length7.6 m (25 ft) standard
Dimensions4.44 cm (1.75 in.) maximum diameter, 31.435 cm (12.375 in.) long
Mounting
Dedicated Mounting Rings, Permanent Mounting Bracket (installs directly to pipe wall),
Adjustable Mounting Band Kit.
Downlook Ultrasonic Sensor
5
Page 8
Specifications
Table 2 940 Flow Meter Specifications (continued)
Operating Frequency75 kHz
Beam Angle5°
Accuracy
Range10 ft
Operating Temperature–18 to 60 °C (0 to 140 °F)
Temperature Error
Resolution0.019 cm (0.0075 in.)
Material ABS housing with ABS acoustical window
Cable Length7.62 to 30.48 m (25 to 100 ft)
Dimensions15.24 cm x 5.7 cm (6 in. x 2.25 in.)
1
When the sensor is out of the water, the system may report velocity readings of up to 0.76 m/s due to radio
±15.24 cm or 3.048 m ±0.003 m (±0.01 ft), at 22 °C (72 °F), still air, 40–70% relative humidity,
from 6 in. to 10 ft
0.00005 m/°C (0.0001 ft/°F) (maximum error with compensated temperature range—per degree
of change.
frequency and interferences at frequencies of 140–170 MHz and 300 MHz with field strengths greater than 3 V/m.
2
For temperatures above 40 °C (104 °F) add ± 0.3 cm/°C (0.03 in./°F)
6
Page 9
Section 2General Information
2.1Safety Information
Read the entire manual before unpacking, setting up, or operating this instrument. Pay
particular attention to all danger and caution statements. Failure to do so could result in
serious injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is not impaired, do not use or
install this equipment in any manner other than what is specified in this manual.
2.1.1 Use of Hazard Information
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
Important Note: Information that requires special emphasis.
Note: Information that supplements points in the main text.
2.1.2 Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol, if noted on the instrument, will be
included with a danger or caution statement in the manual.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed off in European public disposal systems after 12
August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of life equipment to the producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or
electrocution exists and indicates that only individuals qualified to work with hazardous voltages should open the
enclosure or remove the barrier.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol, when noted on the product, indicates that the marked item can be hot and should not be touched
without care.
This symbol, when noted on the product, indicates the presence of devices sensitive to Electro-static Discharge and
indicates that care must be taken to prevent damage to them.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
7
Page 10
General Information
2.1.3 Hazardous Locations
DANGER
Although some Sigma products are designed and certified for installation in
hazardous locations as defined by the National Electrical Code, many Sigma
products are not suitable for use in hazardous locations. It is the responsibility of
the individuals who are installing the products in hazardous locations to determine
the acceptability of the product for the environment. Additionally, to ensure safety,
the installation of instrumentation in hazardous locations must be as per the
manufacturer's control drawing specifications. Any modification to the
instrumentation or the installation is not recommended and may result in life
threatening injury and/or damage to facilities.
DANGER
Bien que certains produits Sigma soient conçus et certifiés pour être installés dans
des endroits dangereux tels que définis par le National Electric Code, de nombreux
produits Sigma ne conviennent pas pour de tels endroits. Il relève de la
responsabilité des personnes qui placent les produits dans des endroits dangereux
de déterminer s'ils sont adaptés à cet environnement. En outre, à des fins de
sécurité, le placement de machines dans des endroits dangereux doit s'effectuer
dans le respect des consignes des schémas de contrôle du fabricant. Toute
modification apportée aux machines ou tout déplacement de celles-ci est
déconseillé, car susceptible de provoquer des accidents matériels et/ou corporels.
2.1.4 Intrinsically Safe Equipment
The 911 and 940 Flow Meters are listed as “Intrisically Safe” for Class 1, Div.1, Groups C,
and D Hazardous Locations. This means that the circuits within these flow meters are
designed to be incapable of producing a spark or thermal effect that could ignite a mixture
of flammable or combustible gases when properly installed in a Class 1, Division 1, Group
C, or Group D Hazardous Location. It does not mean that these flow meters are
“Explosion Proof”. If proper safety precautions are not followed, or if the equipment is not
installed properly, there is a serious potential for explosion. Be sure to review all safety
precautions, installation, and wiring practices throughout this manual prior to installing a
911 or 940 Flow Meter.
•It is important that the installer of an Intrinsically Safe system reference the
requirements of the authority having jurisdiction at the installation site. In the United
States, reference should be made to Article 504 of the National Electric Code and
ANSI/ISA RP 12.6. The 911 and 940 Flow Meters should be installed in accordance to
these requirements and in accordance with the approved control drawings in the
manual (Refer to section 4.1 on page 15).
•Wherever possible, associated apparatus should be mounted in a non-hazardous
location, closest to the hazardous location. This minimizes the length of the
intrinsically safe conductors within the non-hazardous location, thereby decreasing
the possibility of inadvertent connection to non-intrinsically safe energy to the
protected circuit. It is recommended that the associated apparatus be mounted and
installed in dust and moisture free enclosures. Dust and moisture are conductive
materials and may compromise the required minimum distance of 50 mm (2 in.)
between intrinsically safe and non-intrinsically safe conductors. The panel layout of
these enclosures should be constructed such that the separation of intrinsically safe
and non-intrinsically safe conductors is maximized.
8
Page 11
2.1.4.2 Intrinsically Safe Wiring Practices
DANGER
Intrinsically safe wiring must be separated from non-intrinsically safe wiring to
prevent the transferring of unsafe levels of energy to the hazardous area.
DANGER
Le câblage à sécurité intrinsèque doit être séparé du câblage à sécurité non
intrinsèque afin d'empêcher tout transfert de niveaux d'énergie non sûrs vers la
zone dangereuse.
The following practices are meant to be used as guidelines or recommendations only. For
specific rules or more detailed practiced, refer to the National Electric Code, Canadian
Electric Code, or ANSI/ISA-RP 12.6-1987 or subsequent.
Intrinsically safe wiring must be installed, maintained, and repaired with considerations for
the following:
Connections
Do not connect non-intrinsically safe wiring to intrinsically safe terminations of any
associated apparatus, usually identified with light blue.
Routing
Intrinsically safe wiring must enter or leave the non-hazardous location by the shortest and
most direct route.
General Information
Clearances
A clearance distance of at least 50 mm (2 in.) must be maintained between any
non-intrinsically safe conductors and intrinsically safe wiring or terminations.
The only exceptions allowed for minimum clearances are:
•All intrinsically safe circuit conductors are in Type MI or MC cables or
•All non-intrinsically safe circuit conductors are in raceways or Type MI or MC cables
where the sheathing or cladding is capable of carrying fault current to the ground.
Raceways
Raceways should be used to keep intrinsically safe wire and non-intrinsically safe wire
separated. Wire lacing or ties are considered acceptable methods. All raceways for the
intrinsically safe system are identified with permanently affixed labels with the wording
Intrinsic Safety Wiring or equivalent.
Conductor Identification
Intrinsically safe conductors must be identified, either by color coding with light blue
jacketed cable or by tagging, at regular intervals of every 7.62 m (25 ft), as identified by
NEC (National Electric Code).
Voltage Limitations
The electrical equipment for non-hazardous location must not contain a source voltage
greater than 250 V unless sufficient means have been employed to prevent the shorting of
a source voltage greater than 250 V onto the non-intrinsically safe terminals of the
associated apparatus.
9
Page 12
General Information
Multi-Conductor Cable
Conductors may be part of a multi-conductor cable provided that all conductors within the
cable are intrinsically safe. Each intrinsically safe conductor must have an insulation
thickness of 0.25 mm (0.010 in.) which is rated for the maximum temperature rating of the
approved device to which it is connected. The intrinsically safe wiring must be color coded
with light blue, if no other wiring is light blue or appropriately tagged.
Seals
A seal must be provided at the point the intrinsically safe wire passes between the
hazardous and non-hazardous locations to prevent the hazardous atmosphere from
entering the unprotected control room environment. The seal specification is that it must
not pass more than 198 cm3 (0.007 ft3) of air per hour at a pressure equivalent to 6 in.
(1493 Pa) of water. While this is the same specification for the sealing requirements of an
explosion proof system, it may not be necessary to employ those seals.
Grounding
The 911 and 940 Flow Meters and associated Opto-Isolated Interfaces do not require
grounding or bonding. Under no circumstances should any attempt be made to ground or
bond the 911 or 940 Flow Meters or Opto-Isolated Interfaces.
2.1.5 Confined Space Entry
Important Note: The following information is provided to guide users of 911 and 940 Flow
Meters on the dangers and risks associated with entry into confined spaces.
DANGER
Additional training in Pre-Entry Testing, Ventilation, Entry Procedures,
Evacuation/Rescue Procedures and Safety Work Practices is necessary to ensure
against the loss of life in confined spaces.
DANGER
Pour éviter les accidents mortels dans les espaces confinés, il faut organiser des
formations supplémentaires dans les matières suivantes: Contrôle avant entrée,
Ventilation, Procédures d'entrée, Procédures d'évacuation et de secours et
Méthodes de travail sûres.
On April 15, 1993, OSHA's final ruling on CFR 1910.146, Permit Required Confined
Spaces, became law. This standard directly affects more than 250,000 industrial sites in
the United States and was created to protect the health and safety of workers in
confined spaces.
2.1.5.1 Definition of Confined Space
Confined Space is any location or enclosure that presents or has the immediate potential
to present one or more of the following conditions:
•An atmosphere with less than 19.5% or greater than 23.5% oxygen and/or more than
10 ppm Hydrogen Sulfide (H2S)
•An atmosphere that may be flammable or explosive due to gases, vapors, mist, dust,
or fibers
10
•Toxic materials which, upon contact or inhalation, could result in injury, impairment of
health, or death
Confined spaces are not designed for human occupancy. They have restricted entry and
contain known or potential hazards.
Examples of confined spaces include manholes, stacks, pipes, vats, switch vaults, and
other similar locations.
Page 13
General Information
Important Note: Standard safety procedures must always be followed prior to entry into
confined spaces and/or locations where hazardous gases, vapors, mist, dust, or fibers
may be present.
Note: Before entering any confined space, check with the employer for procedures related to
confined space entry.
11
Page 14
General Information
12
Page 15
Section 3Getting Started
3.1Installing the Flow Meter-Considerations
3.1.1 Choosing the Proper Site
The accuracy of flow measurement depends on the suitability of individual monitoring
sites. Select sites that have normalized flow and minimal turbulence. Turbulence can
make it difficult to detect an average velocity in the flow stream. Obstructions, vertical
drops, pipe bends, and elbows can create turbulence. Table 3 contains suggestions for
preventing turbulence.
Table 3 Suggestions for Preventing Turbulence
Site ConditionSuggested Remedy
OutfallsAttach the sensor at least ten times the maximum expected level upstream of the outfall.
Vertical drops in the
channel floor
Elbows, sharp turns,
and “Y” connections
3.1.2 Dealing with Difficult Sites
Attach the sensor at least ten times the maximum expected level upstream of the vertical drop.
Attach the sensor at least ten times the maximum expected level downstream of the vertical drop.
Locate the sensor at least ten times the maximum expected level upstream of the obstruction.
Locate the sensor at least ten times the maximum expected level downstream of the obstruction.
Some sites may be difficult to monitor due to poor site conditions. The direction and speed
of particles in the flow stream contribute to the signal received by the velocity sensor. If the
turbulence near the measurement point is excessive, it may be difficult for the sensor to
determine the average velocity of the stream. The flow meters provide several unique
features to help deal with these problem sites.
3.1.3 Choosing the Appropriate Meter and Sensor Combination
Use Table 4 to determine which sensor to use. See Section 6 on page 43 for sensor
installation instructions.
Table 4 Flow Meter and Sensor Configurations
Flow Meter ModelLevel and Velocity Sensor Configurations
911 Meter plus one Submerged Area/Velocity Sensor or one Velocity-Only Sensor
940
Meter plus two Submerged Area/Velocity Sensors or
Meter plus one Submerged Area/Velocity Sensor and one In-Pipe/Downlook Ultrasonic Sensor.
13
Page 16
Getting Started
14
Page 17
Section 4Meter Installation
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.
DANGER
Seul un technicien qualifié peut effectuer les tâches d'installation décrites dans
cette section du manuel.
This section describes the ONLY approved installation instructions for the 911 and 940
Flow Meters. This section also describes battery installation and interface wiring for a
RS232, Modem, and Sampler connection.
4.1Approved 911 and 940 Installation Drawings
DANGER
Any installation or flow meter configuration not specifically detailed on the
following control drawings is not allowed. In all cases, the local authority having
jurisdiction shall have a final say.
DANGER
Il est interdit de procéder à toute installation ou configuration d'un débitmètre qui
n'est pas explicitement détaillée dans les schémas de contrôle de l'installation
ci-dessous. Dans tous les cas, c'est l'autorité locale responsable qui aura le
dernier mot.
Figure 1 on page 16 through Figure 13 on page 28 are approved control drawings. These
certified drawings detail the ONLY approved method of installing the 911 and 940 Flow
Meters. Additionally, these drawings list both part number, description, and the only
certified sensors, probes, and associated equipment to be used with these meters. Any
substitutions automatically void the Intrinsically Safe certification of the flow meter and
could result in fire or explosion.
15
Page 18
Meter Installation
1
2
3
4
APPROVED
REVISION
DESCRIPTION
REV
D
Jeff Archer
TO COMPUTER
09-561
10 FT
3.1 m
3-5v DC
L
-OR-
P/N 4738
P/N 4087
LOCATION
NONHAZARDOUS (CLASSIFIED)
3-5v DC
RS-232
INTERFACE
-OR-
OPTO-ISOLATED
150 FT.
MAX LENGTH
C
1000
TO SIGMA
2.0 m
6.56 FT
RS-232
-OR-
P/N 5254200
P/N 5254100
INTERFACE
OPTO-ISOLATED
150 FT.
MAX LENGTH
NON-
10 FT
3.1 m
P/N 4274
CLASSIFIED
SAMPLER
SAMPLER
INTERFACE
OPTO-ISOLATED
B
A
HACH COMPANY
5600 LINDBERGH DR.
LOVELAND, CO. 80539
911 BASE
DRAWING, CONTROL
TITLE:
DATE
12/2/05
S.D
S.D
NAME
THIRD ANGLE PROJECTION
ENGINEER
DRAWN
.25
.005
.01
.03
DIMENSIONS ARE IN INCHES
TOLERANCES:
.X =
.XX =
.XXX =
ANGLES =
UNLESS OTHERWISE SPECIFIED:
INTERPRET GEOMETRIC
TOLERANCING PER:
L
REV
911-V1
DWG. NO.
B
SIZE
ANSI Y14.5M 1994
SHEET 1 OF 1
1
WEIGHT:
SCALE: 1:2
2
3
DO NOT SCALE DRAWING
4
NOTICE--HACH COMPANY CLAIMS PROPIETARY RIGHTS IN THE INFORMATION DISCLOSED ON THIS DRAWING. IT IS ISSUED
IN CONFIDENCE FOR ENGINEERING INFORMATION ONLY AND MAY NOT, IN WHOLE OR IN PART, BE USED TO MANFACTURE
ANYTHING, WHETHER OR NOT SHOWN HEREON, REPRODUCED OR DISCLOSED TO ANYONE WITHOUT DIRECT PERMISSION
P/N 5150
MODEL 911
5
HAZARDOUS (CLASSIFIED) LOCATION
CLASS 1, DIV. 1, GROUP C,D
6
INTEGRAL
ASSEMBLY
IN LINE VENT
25 FT - 100 FT MAX
7.6m - 30.5m
7
8
IF THE ELECTRICAL PARAMETERS OF THE CABLE ARE UNKNOWN, THE FOLLOWING DEFAULT VALUES MAY BE USED.
CAPACITANCE - .2nf/METER, INDUCTANCE - 1uh/METER
CANADIAN ELECTRIC CODE, C22.1-94, SECTION 18 AND ANSI/ISA RP 12.6
1.
2. WIRING METHODS MUST BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE, ANSI/NFPA 70, ARTICLE 504,
D
PROBE
REF. TABLE 1
OR 2 FOR P/N
AREA VELOCITY
FLOWMETER
C
TABLE 1
REGULATORY AGENCY
TABLE 2
B
MATERIAL
FROM HACH COMPANY.
THIS DRAWING
APPROVAL REQUIRED
PRIOR TO CHANGING
A
5
6
7
8
16
Figure 1 Control Drawing 911-V1
Page 19
REVISION
REV
REVISED PER R-3142-05
E
1.
CONNECTION OF P/N 4080 FLASH CONVERTER TO 911 FLOW METER
DESCRIPTION
MUST BE MADE IN A NON HAZARDOUS LOCATION.
Meter Installation
APPROVED
P/N 5150
911
FLOW METER
NOTICE-HACH COMPANY CLAIMS PROPRIETARY
RIGHTS IN THE INFORMAT ION DISCLOSED ON THE
DRAWING. IT IS ISSUED IN C ONFIDENCE FOR ENGINEERING INFORMATION ONLY AND MAY NOT, IN
WHOLE OR IN PART, BE USED TO MANFACTURE
ANYTHING, WHETHER OR NOT SHOWN HEREON/
REPRODUCED OR DISCLOSED TO ANYONE WITHOUT
DIRECT PERMISSION FROM HACH COMPANY.
MATERIAL
NON HAZARDOUS (CLASSIFIED)
LOCATION
10 FT/
3.1m
P/N 4080
FLASH CONVERTER
UNLESS OTHER WISE SPECIFIED
DIMENSIONS ARE IN INCH ES
TOLERANCES:
.X =
.XX =
.XXX =
ANGLES = .2 5
1. IF THE ELECTRICAL PARAMETERS OF THE CABLE ARE UNKNOWN, THE FOLLOWING DEFAULT VALUES MAY BE USED:
MINIMUM BEND RADIUS IS 3 IN./7.6 cm.
2. WIRING METHODS MUST BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE, ANSI/NFPA 70, ARTICLE 504,
3. FOR MODEM INSTALLATION IN CONDUIT, MINIMUM CONDUIT DIAMETER IS 1/2 IN./1.3 cm.,
D
C
TABLE 1
B
THIS DRAWING
APPROVAL REQUIRED
PRIOR TO CHANGING
A
5
6
7
8
Figure 4 Control Drawing 940-V1
19
Page 22
Meter Installation
1
2
3
4
APPROVED
REVISION
DESCRIPTION
REV
D
Jeff Archer
TO COMPUTER
09-561
10 FT
3.1 m
3-5v DC
L
-OR-
P/N 4738
P/N 4087
LOCATION
NONHAZARDOUS (CLASSIFIED)
3-5v DC
RS-232
INTERFACE
-OR-
OPTO-ISOLATED
150 FT.
MAX LENGTH
C
1000
TO SIGMA
2.0 m
6.56 FT
RS-232
-OR-
P/N 5254200
P/N 5254100
INTERFACE
OPTO-ISOLATED
150 FT.
MAX LENGTH
NON-
10 FT
3.1 m
P/N 4274
CLASSIFIED
SAMPLER
SAMPLER
INTERFACE
OPTO-ISOLATED
B
P/N 4443
F.O. MODEM
F.O.
MODEM
P/N 4413
CUSTOMER DEFINABLE LENGTH
INTERFACE
NOTICE--HACH COMPANY CLAIMS PROPIETARY RIGHTS IN THE INFORMATION DISCLOSED ON THIS DRAWING. IT IS ISSUED
IN CONFIDENCE FOR ENGINEERING INFORMATION ONLY AND MAY NOT, IN WHOLE OR IN PART, BE USED TO MANFACTURE
A
L
REV
SHEET 1 OF 1
1
HACH COMPANY
5600 LINDBERGH DR.
LOVELAND, CO. 80539
940 ULTRASONIC
DRAWING - CONTROL,
TITLE:
DATE
12/05/05
S.D
S.D
NAME
THIRD ANGLE PROJECTION
ENGINEER
DRAWN
.25
.005
.01
.03
DIMENSIONS ARE IN INCHES
TOLERANCES:
.X =
.XX =
.XXX =
ANGLES =
UNLESS OTHERWISE SPECIFIED:
MATERIAL
ANYTHING, WHETHER OR NOT SHOWN HEREON, REPRODUCED OR DISCLOSED TO ANYONE WITHOUT DIRECT PERMISSION
FROM HACH COMPANY.
INTERPRET GEOMETRIC
940-V2
DWG. NO.
B
SIZE
TOLERANCING PER:
ANSI Y14.5M 1994
WEIGHT:
SCALE: 1:2
2
3
DO NOT SCALE DRAWING
4
P/N 4840
MODEL 940
5
HAZARDOUS (CLASSIFIED) LOCATION
CLASS 1, DIV. 1, GROUP C,D
6
INTEGRAL
ASSEMBLY
IN LINE VENT
25 FT - 100 FT MAX
7.6m - 30.5m
7
8
CAPACITANCE - .2nf/METER, INDUCTANCE - 1uh/METER
CANADIAN ELECTRIC CODE, C22.1-94, SECTION 18 AND ANSI/ISA RP 12.6
MINIMUM BEND RADIUS IS 3 IN./7.6 cm.
1. IF THE ELECTRICAL PARAMETERS OF THE CABLE ARE UNKNOWN, THE FOLLOWING DEFAULT VALUES MAY BE USED.
3. FOR MODEM INSTALLATION IN CONDUIT, MINIMUM CONDUIT DIAMETER IS 1/2 IN./1.3 cm.,
2. WIRING METHODS MUST BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE, ANSI/NFPA 70, ARTICLE 504,
D
PROBE
REF. TABLE 1
OR 2 FOR P/N
AREA VELOCITY
FLOWMETER
C
100 FT MAX
PROBE
P/N 4741
DEADBAND
75KHz ZERO
30.5m
B
TABLE 1
APPROVAL REQUIRED
REGULATORY AGENCY
TABLE 2
THIS DRAWING
PRIOR TO CHANGING
A
5
6
7
8
20
Figure 5 Control Drawing 940-V2
Page 23
REVISION
REV
REVISED PER R-3146-05
E
CONNECTION OF P/N 4080 FLASH CONVERTER TO 940 FLOW METER
1.
DESCRIPTION
MUST BE MADE IN A NON HAZARDOUS LOCATION.
Meter Installation
APPROVED
P/N 4840
940
FLOW METER
NOTICE-HACH COMPANY CLAIMS PROPRIETARY
RIGHTS IN THE INFORMAT ION DISCLOSED ON THE
DRAWING. IT IS ISSUED IN C ONFIDENCE FOR ENGINEERING INFORMATION ONLY AND MAY NOT, IN
WHOLE OR IN PART, BE USED TO MANFACTURE
ANYTHING, WHETHER OR NOT SHOWN HEREON/
REPRODUCED OR DISCLOSED TO ANYONE WITHOUT
DIRECT PERMISSION FROM HACH COMPANY.
MATERIAL
NON HAZARDOUS (CLASSIFIED)
LOCATION
10 FT/
3.1m
P/N 4080
FLASH CONVERTER
UNLESS OTHER WISE SPECIFIED
DIMENSIONS ARE IN INCH ES
TOLERANCES:
.X =
.XX =
.XXX =
ANGLES =
2. WIRING METHODS MUST BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE, ANSI/NFPA 70, ARTICLE 504,
1.
D
C
TABLE 1
B
MATERIAL
FROM HACH COMPANY.
THIS DRAWING
APPROVAL REQUIRED
PRIOR TO CHANGING
A
5
6
7
8
28
Figure 13 Control Drawing 940-V15
Page 31
4.2Mounting the flow meter
When mounting the flow meter, remember to:
•Mount the meter so that the connectors face down. The end of the meter that
contains the batteries should face up. If the connectors face up, they may corrode and
allow water to seep into the instrument. Refer to Figure 14 on page 29.
•When not in use, cover the connectors with their protective caps to prevent corrosion.
Use the appropriate manhole support bracket/spanner bar. See Replacement Parts and
Accessories on page 67.
Meter Installation
Figure 14 Installation options
1Instrument Support Bracket (Cat. No. 5713000)4911 flow meter
2Manhole cover5Wall/Ladder Mount Suspension Bracket (Cat. No. 4874
and Cat. No. 4839)
3Suspension harness (Cat. No. 4920)6940 flow meter
29
Page 32
Meter Installation
4.3Installing the Batteries
Never short circuit, puncture, deform, or incinerate any of the battery packs used in the
911 or 940 flow meters. Proper precautions must be observed in the handling, shipping,
and disposal of battery packs.
4.3.1 Installing the Battery Pack in the 940
DANGER
Never install, remove, or charge batteries in a hazardous location.
DANGER
Vous ne devez jamais installer, retirer ou recharger les batteries dans un
environnement dangereux.
The steps mentioned below give the installation procedure of the battery pack in the 940.
1. Loosen the 5/16” socket head cap screw in the center of the bottom end cap.
2. Lift the end cap off.
3. Remove the two thumbscrews securing the battery retaining plate and unlatch the two
rubber hold down straps.
4. Place the battery pack (Cat. No. 6608500) inside the battery compartment (Figure 15).
5. Place the two rubber hold down straps across the battery to secure the battery in
place.
6. Attach the battery power connector to the flow meter power connector.
7. Both connectors are physically polarized.
8. Try to pull the two connectors apart gently, after placing the two connectors together,
to ensure that they are latched together, . Do not use excessive force or permanent
damage to the connectors may occur.
9. Reattach the battery retaining plate and install the two thumbscrews.
10. Replace the flow meter end cap and retighten the 5/16” socket head cap screw.
1
Figure 15 940 Flow Meter Battery Pack
1Battery Pack Location
30
Page 33
4.3.2 Installing the Battery Pack in the 911
DANGER
Use only exact replacement battery packs purchased directly from the factory or a
factory authorized outlet. Any unauthorized battery substitution will void the
intrinsically safe design and approval of the unit and could result in fire or
explosion.
DANGER
Utilisez exclusivement des batteries de rechange identiques que vous achetez
directement auprès de l'usine ou d'un point de vente agréé par celle-ci. Tout
remplacement non autorisé d'une batterie annule la conception à sécurité
intrinsèque et l'agréation de l'appareil, et peut provoquer un incendie ou une
explosion.
DANGER
Never install, remove, or charge batteries in a hazardous location.
DANGER
Vous ne devez jamais installer, retirer ou recharger les batteries dans un
environnement dangereux.
The entire battery assembly unscrews and separates to access the battery connector
(Cat. No. 5160-01). Follow the steps below to install the battery pack in the 911.
Meter Installation
1. Hold the 911 Flow Meter by the handle on the end opposite of the interface connectors
and twist counter-clockwise to open.
2. Remove the rechargable battery pack. Recharge or dispose
of the battery.
3. Reinstall the rechargable battery. Attach the battery connector to the center mounted
connector at the bottom of the 911 case (Figure 16).
Note: Both connectors are physically polarized.
4. Place the two connectors together. Gently try to pull the two connectors apart to
ensure that the connectors are securely latched together. Do not use excessive force
or permanent damage to the connectors may occur.
5. Tighten the screw of the battery pack assembly onto the flow meter.
1
2
Figure 16 911 Flow Meter Case and Battery Packs
1911 Flow Meter Case212 Amp-hour rechargeable gell cell battery pack.
31
Page 34
Meter Installation
4.3.3 Battery Compartment Desiccant
The battery compartment in both the 911 and 940 Flow Meters is kept dry with a small
desiccant cartridge to avoid moisture damage to the batteries and power circuitry
(Figure 17 and Figure 18).
When the beads are blue they can remove moisture from the enclosure. When they turn
pink, they cannot absorb any more
moisture. When the blue desiccant turns pink, replace the desiccant.
Figure 17 911 Case, Battery End with Inserted Desiccant
1
1Desiccant
Figure 18 940 Case, Battery End with Inserted Desiccant
1Desiccant
4.3.4 Estimating the Battery Life
When using the battery pack, battery life is based on:
•Number of sensors installed (more sensors decrease battery life).
Batteries for the 911 and 940 flow meters should last as listed above under the following
conditions:
•15 minute logging intervals
•1 level channel and 1 velocity channel logged
•Data downloaded once per week
•10 °C (50 °F) ambient temperature
•Also affected by site conditions
4.3.5 Recharging the Lithium Ion Battery Pack for Model 940
DANGER
Use only exact replacement battery packs purchased directly from the factory or a
factory authorized outlet. Any unauthorized battery substitution will void the
intrinsically safe design and approval of the unit and could result in fire or
explosion.
Meter Installation
DANGER
Utilisez exclusivement des batteries de rechange identiques que vous achetez
directement auprès de l'usine ou d'un point de vente agréé par celle-ci. Tout
remplacement non autorisé d'une batterie annule la conception à sécurité
intrinsèque et l'agréation de l'appareil, et peut provoquer un incendie ou une
explosion.
DANGER
Never install, remove or charge batteries in a hazardous location.
DANGER
Vous ne devez jamais installer, retirer ou recharger les batteries dans un
environnement dangereux.
To recharge the 940 Lithium Ion Battery Pack (6608500):
1. Move the 940 Flow Meter to a non-hazardous location.
2. Remove the battery pack from the 940 Flow Meter as described in section 4.3.1 on
page 30.
3. Inspect the battery pack and the battery connection for any damage. Do not attempt to
use or repair a damaged battery pack assembly.
4. Make sure that the charging voltage switch on the back of the charger is set to 16.4 V.
5. Plug the charger (6678000) into a standard 120 VAC wall outlet. The green power
LED will illuminate.
6. Connect the charger to the battery pack using the battery charger and battery pack
cables. The yellow charge LED indicator will illuminate. See Table 6 for the Lithium
Ion battery charging status. A completely discharged battery pack will charge in
approximately 8 to 12 hours.
33
Page 36
Meter Installation
Table 6 Battery pack charging status
Charge LED IndicatorBattery Condition
FlashingDeeply discharged or needs replacement
IlluminatedCharging
OffCharged
1
If the charge indicator LED does not stop flashing within 1 hour the battery pack is shorted and
should be disposed of properly.
7. Install the battery pack in the 940 Flow Meter as described in section 4.3.1 on
page 30.
4.3.6 Recharging the Battery Pack for Model 911
DANGER
Use only exact replacement battery packs purchased directly from the factory or a
factory authorized outlet. Any unauthorized battery substitution will void the
intrinsically safe design and approval of the unit and could result in fire or
explosion.
DANGER
Utilisez exclusivement des batteries de rechange identiques que vous achetez
directement auprès de l'usine ou d'un point de vente agréé par celle-ci. Tout
remplacement non autorisé d'une batterie annule la conception à sécurité
intrinsèque et l'agréation de l'appareil, et peut provoquer un incendie ou une
explosion.
1
DANGER
Never install, remove or charge batteries in a hazardous location.
DANGER
Vous ne devez jamais installer, retirer ou recharger les batteries dans un
environnement dangereux.
To recharge the 911 Lithium Ion Battery Pack (5160-01):
1. Move the 911 Flow Meter to a non-hazardous location.
2. Remove the battery pack from the 911 Flow Meter as described in section 4.3.2 on
page 31.
3. Inspect the battery pack and the battery connection for any damage. Do not attempt to
use or repair a damaged battery pack assembly.
4. Plug the charger (913) into a standard 120 volt wall outlet.
5. Connect the charger to the battery pack using the battery charger and battery pack
cables. A completely discharged battery pack will charge in approximately 8 to 12
hours. Do not overcharge.
6. Install the battery pack in the 911 Flow Meter as described in section 4.3.2 on
page 31.
34
Page 37
4.4Wiring the RS232 Interface
DANGER
All connections to the flow meter must be made in a non-hazardous location.
DANGER
Tous les raccordements au débitmètre doivent être effectués dans un
environnement qui ne présente aucun danger.
All RS232 interface wiring must follow the installation drawings in section 4.1 on page 15.
4.4.1 RS232 Serial Port
DANGER
The RS232 Opto-Isolated Interface must be located in a non-hazardous location.
DANGER
L'interface à isolation optique RS232 doit se trouver en un emplacement non
dangereux, conformément aux schémas de contrôle de l'installation.
Note: A PC or laptop must be used in a non-hazardous location.
The high speed RS232 serial port communicates with a PC via an optically isolated
interface (Figure 19 on page 35). The Opto-Isolated Interface (Cat. No. 4087) is used only
for data downloading and remote programming. This interface isolates the energy
potential in the PC from hazardous locations and from the 911 or 940 Flow Meters.
Meter Installation
4
Figure 19 RS232 Communications Setup
1Use “approved” seal to minimize passage of gases or
vapors from a hazardous location to a non-hazardous
location.
2RS232, Opto-Isolated Interface in a non-hazardous
location (Cat number 4087)
1
3PC compatible laptop or desktop located in a
non-hazardous location for downloading data
4940 or 911 Flow Meter in a hazardous location
2
3
DANGER
Route the cable through an approved seal to prevent migration of dangerous fumes
or vapors from the hazardous location to the non-hazardous location.
DANGER
Acheminez le câble à travers un joint approuvé afin d'empêcher la migration de
35
Page 38
Meter Installation
fumées ou vapeurs dangereuses de l'emplacement dangereux vers l'emplacement
non dangereux.
Follow the steps mentioned below to route the RS232 Serial Port
1. Attach the cable with the round connector from the Opto-Isolated Interface to the
connector on the 911/940 Flow Meter labelled RS232.
2. Attach the cable with a rectangular connector to the appropriate serial COM port on
the PC or laptop.
4.5Wiring the Sampler Interface on Model 940
DANGER
All connections to the flow meter must be made in a non-hazardous location.
DANGER
Tous les raccordements au débitmètre doivent être effectués dans un
environnement qui ne présente aucun danger.
All sampler interface wiring must follow the installation drawings in section 4.1 on page 15.
Refer to Table 7 and Figure 20.
Pin Letter
A
(+12 V dc)
B
(Ground)
C
Flow Pulse
Output
D
Sampler
Start
E
Event Input
F
Bottle
Number
Input
Table 7 Sampler Cable Connector Pin Assignments and Description
Wire
Color
White+12 V dc input—
Blue
Yellow
Black
Red
Green
Provides the ground line that is used in conjunction with the other signals
Used in conjunction with Pin B (ground) to signal a sampler that a
pre-determined amount of flow has accumulated.
Used to “wake up” a wastewater sampler when a level and/or rainfall set-point
condition is met so that it can begin sampling. Used in conjunction with Pin B
(ground), this line is normally allowed to float and is switched to ground (by
transistor) once the set-point conditions are met.
Confirms that a sample has been collected. The time and date of each water
sample appears on the data printout when downloaded using a Data Transfer
Unit or Sigma Support Software.
Used in conjunction with the “Event Input” signal described above. It tells the flow
meter which bottle was used and when a sample was taken. This information
appears in the data printout (see “Sample Times and Dates”) when downloaded
using a Data Transfer Unit or Support Software.
Bottle Number signal description
If the Program Complete Output (Pin F of the auxiliary connector) is disabled on
the sampler, then it is used to transmit the bottle number to the connected
Time B = 100 ms (50 ms HI 50 ms LO)
If a sample attempt fails, the first pulse width 'B' is 150 ms HI and
If Sample Distribution is programmed for multiple bottles per sample mode, only
the first bottle number of the set is transmitted via Pin F.
DescriptionRating
on this connector.
+8 V dc. Output
current is 100 mA
dc (max) with a
duration of 500 ms.
+24 V dc (max)
+12 V dc (max)
minimum 3 second
pulse.
device.
Time A = 200 ms
50 ms LO.
—
—
36
Page 39
Meter Installation
RED
ORANGE
YELLOW
GREEN
BLACK
BROWN
TO FLOW METER
DO NOT TIN WIRES PRIOR TO TERMINATION
Figure 20 Sampler Interface Wiring
4.6Sampler Receptacle (940 Only)
DANGER
Do not attempt to connect any sampler to the 940 Flow Meter without the use of the
Optically Isolated Sampler Interface (Cat. No. 4274). Failure to do so may result in
fire or explosion.
DANGER
Le raccordement d'un échantillonneur quelconque à un débitmètre 940 sans
utilisation de l'interface d'échantillonneur à isolation optique (No. Cat. 4274) annule
à la fois la garantie et l'agrément de sécurité intrinsèque du débitmètre, et peut
provoquer un incendie ou une explosion. Reportez-vous aux schémas de contrôle
de l'installation.
BLUE
WHITE
BLACK
YELLOW
GREEN
RED
T
O
S
A
M
P
L
E
R
The optional “sampler” port lets the user to connect a 940 flow meter to a Sigma
wastewater sampler (Figure 21). The option provides several features:
•Flow proportional sampling
•Storm water trigger, based on level, useful for EPA NPDES permitting program.
37
Page 40
Meter Installation
•Sample history logging.
Figure 21 Sampler Communications
1Use “approved” seal to minimize passage of gases or
vapors from hazardous location to non-hazardous
location.
2Optically Isolated Sampler Interface (Cat. No. 4274) in a
non-hazardous location
3Sampler located in a non-hazardous location
4940 Flow Meter in a hazardous location
4.6.1 Connecting the Sampler Cable to a 940 Flow Meter
Note: The sampler must be located in a non-hazardous location.
1. Connect the Optically Isolated Sampler Interface (Cat. No. 4274) between the 940
sampler receptacle and the samplers auxiliary receptacle.
2. Program the sampler for Special Output/Sample Output.
4.6.2 Connecting a Non-Sigma Sampler to a 940 Flow Meter
DANGER
Do not attempt to connect any sampler to the 940 Flow Meter without the use of the
Optically Isolated Sampler Interface (Cat. No. 4274). Failure to do so may result in
fire or explosion.
DANGER
Le raccordement d'un échantillonneur quelconque à un débitmètre 940 sans
utilisation de l'interface d'échantillonneur à isolation optique (No. Cat. 4274) annule
à la fois la garantie et l'agrément de sécurité intrinsèque du débitmètre, et peut
provoquer un incendie ou une explosion. Reportez-vous aux schémas de contrôle
de l'installation.
38
The Optically Isolated Sampler Interface can also be used with samplers produced by
other manufacturers. In many cases a non-Sigma sampler may be connected to the
Optically Isolated
Sampler Interface without any modifications or alterations
If a non-Sigma sampler is not compatible with the connector on the end of the sampler
interface cable it may be necessary to remove the existing cable from the sampler
interface and use the other manufacturer’s sampler cable.
Page 41
Meter Installation
Note: It is the installers responsibility to ensure compatibility and proper connection of a non-Sigma
sampler to the Optically Isolated Sampler Interface.
1. Remove the cover from the Optically Isolated Sampler Interface.
2. Disconnect all the wires from the terminals labeled J4 through J6. Do not remove the
wires from terminals labeled J1 through J3.
3. Loosen the compression nut on the sampler cable connector and remove the cable
from the interface.
4. Insert the other manufacturer’s sampler cable end into the cable connector of the
Optically Isolated Sampler Interface.
39
Page 42
Meter Installation
40
Page 43
Section 5Software and Communications
5.1Required Software
Programming and data retrieval are performed via an IBM compatible personal computer
(computer must have a serial port, USB is not supported) using an Optically Isolated
Interface with one of the following software packages:
•InSight® Data Analysis Software is a program for small- to medium -sized flow
monitoring jobs.
•Flo-Center Integrated Sewer System Management Software provides high-end,
automated data collection and automated batch report processing, using an unlimited
number of meters.
•Modicon Modbus Protocol (940 Only)
•Flo-Ware for Windows (911 Only)
5.2Programming and Analysis Options
•Built-in flow equations
•Storage of 116,000 data points in battery backed RAM memory
•Storm water sampling trigger option (940 only)
•Flow proportional output option (940 only)
For details on programming the 911 or 940 Flow Meter, refer to the
software manual.
41
Page 44
Software and Communications
42
Page 45
Section 6Sensor Installation
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.
DANGER
Seul un technicien qualifié peut effectuer les tâches d'installation décrites dans
cette section du manuel.
DANGER
All connections to the flow meter must be made in a non-hazardous location.
DANGER
Tous les raccordements au débitmètre doivent être effectués dans un
environnement qui ne présente aucun danger.
6.1Downlook Ultrasonic Sensor
The downlook ultrasonic sensor is mounted over the flow stream at the proper location for
level measurement for the primary device. See Appendix A on page 75 for more
information. The transducer emits a pulse of sound at a high frequency and awaits for the
echo to return from the surface of the water. The time it takes for this echo to return is
relative to the distance between the transducer and the surface water. As the level in the
flow stream increases, the time it takes for the echo to return to the transducer decreases
(distance is shorter).
By continuously transmitting these pulses and timing the returning echoes, the sensor
measures the level of the liquid on the flow stream. After measuring the level, the
microprocessor converts the level reading to a flow rate based on the user-defined
characteristics of the primary device.
6.1.1 Installing the Downlook Ultrasonic Sensor
Note: Do not suspend the sensor by the cable. Only use appropriate mounting hardware.
Site selection can greatly enhance system performance. Several important factors should
be considered when installing the downlook ultrasonic sensor at any primary device.
•Locate the ultrasonic sensor at the proper head measurement point for that primary
device.
•Determine the appropriate sensor height.
•Suspend the ultrasonic sensor over the center of the flow stream where there is the
least surface turbulence.
•Mount the sensor to a stable, secure location, isolated from vibration.
Invisible Range
The 940 Flow Meter is equipped with an invisible range (adjustable deadband feature) to
prevent false echoes from tops of channel walls, ladder rungs, shelves, etc. A user
selected range is defined which is invisible to the flow meter. Extend the invisible range to
where it meets or overlaps the highest expected level in the channel. Refer to Figure 22
on page 44. A gap of at least 5 cm (2 in.) should be left between the invisible range and
the highest expected level. The minimum distance must be at least 29 cm (11 in.) The
43
Page 46
Sensor Installation
sensor is “blind” to anything closer than the deadband and stops reading level when the
distance is shorter than that.
4
1
2
3
Figure 22 Invisible Range
1Maximum Range3Highest expected water level
2Minimum Distance (must be at least 11 in. (29 cm)4Invisible Range (set to ignore tops of channel walls)
Beam Angle
The beam angle from the bottom of the ultrasonic sensor spreads out at an angle of 5° for
the 75 kHz sensor as it travels away from the sensor. Therefore, if the sensor is mounted
too high above a narrow channel, the beam may be wider than the channel when it
reaches the bottom of the channel. This can cause false echoes from the tops of the
channel walls rather than from the water surface. To prevent false echoes, determine the
appropriate sensor height so that the entire beam falls within the channel and does not
strike any obstructions.
Beam Width Formulas
The maximum beam width for a given distance from the sensor can be calculated as
follows: . Refer to Figure 23 and Table 8 on page 45.
0.087 distance from sensor
1
2
Figure 23 Beam Width as Distance from Transducer Increases
1Distance from Transducer2Beam Width
Protecting the Sensor
Keep the face of the sensor free of accumulating grease and dirt. Since the 940 Flow
Meter listens for the returning echo, a coated sensor will not be able to accurately detect
the echo and may provide inaccurate level measurement.
44
Page 47
Sensor Installation
Table 8 Beam Angle Spread by Distance
Distance from Target (in.)75 KHz Beam Width (in.) w/ Horn
11 (27.9 cm)0.96 (2.4 cm)
12 (30.5 cm)1.05 (2.67 cm)
13 (33.0 cm)1.14 (2.9 cm)
14 (35.6 cm)1.2 (3.1 cm)
15 (38.1 cm)1.31 (3.3 cm)
16 (40.6 cm)1.4 (3.6 cm)
17 (43.2 cm)1.5 (3.8 cm)
18 (45.7 cm)1.6 (4.0 cm)
19 (48.3 cm)1.7 (4.2 cm)
20 (50.8 cm)1.75 (4.5 cm)
21 (53.3 cm)1.83 (4.6 cm)
22 (55.9 cm)1.9 (4.9 cm)
23 (58.4 cm)2.0 (5.1 cm)
24 (61.0 cm)2.1 (5.3 cm)
36 (91.4 cm)3.1 (8.0 cm)
48 (1.2 m)4.2 (10.6 cm)
60 (1.5 m)5.2 (13.3 cm)
72 (1.8 m)6.3 (16.0 cm)
84 (2.1 m)7.3 (18.6 cm)
96 (2.4 m)8.4 (21.3 cm)
108 (2.7 m)9.4 (24.0 cm)
6.1.1.1 Mounting the Ultrasonic Sensor
Note: To prevent mechanical vibration ensure that the supplied isolation washers are installed as
shown in Figure 24 on page 46.
Several brackets are available for mounting the sensor to a fixed location such as a wall or
floor. See Figure 24 on page 46 and Figure 25 on page 47. For mounting options see
Section 8 on page 67. Each bracket utilizes the 1.9 cm (¾ in.) NPT connector on the
sensor to secure the sensor to the bracket. Always level the sensor using a level. The
maximum water level must be below the deadband to ensure accurate readings.
45
Page 48
Sensor Installation
1
4
2
3
This surface must be level.
Figure 24 Permanent Bracket and Isolation Washers
1Wall/Rail Mount Bracket (Cat. No. 2974)3Downlook Ultrasonic Sensor
2Isolation Washers (Cat. No. 6820)41 inch x 0.25 Hex Nuts (Cat. No. 1429)
RS485 Time Out
(Flow meter did not receive
data within the specified
time)
Recalibrate the sensor. Look for unusual temperatures or the inability for a new calibrated
level to be read.
Try a different sensor on the same flow meter to rule out instrument problems, and try a
different flow meter with the same sensor to rule out sensor problems.
Check for excessive foam on the water surface. Foam may cause sound waves to be
absorbed rather than reflected.
Check the sensor height. See Figure 22 on page 44.
Check the sensor cable for cuts or nicks.
Check the connection between the flow meter and the sensor.
Make sure that the sensor is level. The sensor must be level for proper return of the signal.
Try shielding the transducer from convection currents.
Repeat the temperature calibration procedure. If the unit reads extreme high or low
temperatures, the sensor may have a bad temperature.
Try a different sensor on the same flow meter to rule out instrument problems, and try a
different flow meter with the same sensor to rule out sensor problems.
Check to see if convection components between the sensor target are present. If these
components are present, install a shield around the sound beam to eliminate temperature
variation due to convection currents. The system is designed with averaging routines to help
mitigate this problem.
Check for excessive foam and oil. While the 940 Flow Meter is designed with an AGC
(Automatic Gain Control) scheme to reduce these factors, it is recommended that sites
without these anomalies be selected.
Use the invisible range to select targets beyond peripheral obstructions. Care must be taken
to set the range higher than the expected level.
Try moving the transducer farther from the liquid.
Check under the transducer for obstructions. This includes the front and sides
of the transducer.
Check the face of the transducer for a coating. This is rare, but cleaning the transducer face
may correct the problem. If this is a constant problem due to site conditions, coat the face of
the sensor with a thin film of silicone grease to keep debris from clinging.
Make sure that the proper rubber isolation washers are used on the mounting bracket. The
transducer can resonate against steel mounting rails. See section 6.1.1.1 on page 45.
If the CPU has trouble communicating with the ultrasonic board, RS485 time outs may occur.
Wait a few minutes and see if the trouble has stopped. If it continues, a problem with the
ultrasonic, velocity, or CPU board may exist.
If the logging interval is 1 or 2 minutes, or if conditions are poor, or if the problem continues
indefinitely, try increasing the logging interval to capture more signals.
When using a combination of area velocity and ultrasonic sensors, this condition may occur
more frequently if site conditions are less than ideal. This is not necessarily an indication of a
defective sensor. This condition may occur occasionally.
Check to see if pins are shorted. Do not short pins on receptacle to avoid damage.
48
Page 51
6.2In-Pipe Ultrasonic Sensor
The in-pipe ultrasonic sensor is used in pipes where level measurement near the top of
the pipe is desired. The sensor reads the level until the liquid reaches the bottom of
the sensor housing. The in-pipe sensor is not recommended for weir or flume applications
due to limited range, but may be desirable in some applications.
Sensor Installation
Figure 26 Side View of In-Pipe
1Pipe Ceiling5Internal Deadband (18.3 cm (7.2 inches))
2Distance from Sensor (Range: 0.2 in. to 5 ft)6Pipe Floor
345° Deflector7Minimum distance to reflecting obstruction (46 cm
(18 in.))
4Ultrasonic Sensor8Reflecting Obstruction
6.2.1 Mounting the In-Pipe Sensor
•Mount the sensor over the center of the flow stream where the surface turbulence is
minimized (Figure 27 on page 50).
•Mount the sensor 46 cm (18 in.) away from obstructions located in front of the in-pipe
sensor to prevent inaccurate liquid level readings (Figure 26).
•Level the sensor using the built-in bubble level.
•Ensure the isolation gasket is in place and the mounting bracket thumbscrews are
finger-tight to avoid sensor ringing.
•Install the sensor within 1.5 m (5 ft) of the lowest expected level (the maximum range
of the sensor).
49
Page 52
Sensor Installation
Figure 27 End View of In-Pipe
1Sensor5Wastewater
2Bracket Thumbscrew6Vibration Isolation Gasket
3Bubble Level7Mounting Clip
4Sound Waves8Pipe (size may vary)
6.2.2 Angling the Beam
The narrow beam of sound that emanates from the bottom of the in-pipe ultrasonic sensor
spreads out at an angle of ±12° (-10 dB) as it travels away from the sensor. This means
that if the sensor is mounted too high above a narrow channel, the beam may be too wide
when it reaches the bottom of the channel. This may cause false echoes from the sides on
the channel walls.
6.2.3 Calibrating the In-Pipe Ultrasonic Sensor
Calibrate the in-pipe sensor each time the sensor is installed at a new site. Calibrate the
in-pipe via one of two methods; Liquid Depth or Sensor Height. Each method has its own
advantages and disadvantages. Liquid Depth calibration is the recommended calibration
method; use the sensor height method only when Liquid Depth calibration is not an
option. An Invisible Range can also be set which allows the transducer to ignore
reflections from obstructions between the sensor and the water surface, such as ladder
rungs, channel side walls, etc.
6.2.3.1 Calibrating the Temperature
The speed of sound in air varies with the temperature of air. The in-pipe sensor is
equipped with temperature compensation to help eliminate the effect of temperature
variation under normal site conditions. Enter the ambient air temperature at the transducer
location. For optimum results, allow enough time (100 minutes) to ensure that the sensor
is at equilibrium with the surrounding air temperature.
50
Page 53
6.2.3.2 Calibrating the Liquid Depth
Liquid depth calibration requires knowing the level or the depth of the liquid in the channel
that is contributing to the flow. Liquid depth calibration is the recommended calibration
method for the in-pipe sensor. Take a physical measurement of the liquid depth and enter
the value into the application software. For a dry channel enter 0 depth.
6.2.3.3 Calibrating the Sensor Height
Sensor height calibration is generally used when access to the primary device is difficult
(such as confined space entry in a manhole) or when there is no liquid flow during
installation of the flow meter. This calibration method requires knowing the distance
between the zero flow point and the bottom of the sensor. In a round pipe the zero flow
point is typically the invert or bottom of the pipe. Further, compensation is required for the
invisible range (internal deadband) in the sensor housing. Measurement uncertainty
increases to 1.28 cm (0.042 ft) for a ±30 cm (±1 ft) change in level from the calibration
point. Use this method ONLY if the Liquid Depth is not an option.
Measure the distance from the bottom of the sensor to the zero flow point. Add 18.3 cm
(7.2 in.) to the measured distance to obtain the total zero flow distance. Enter the total zero
flow distance value into the application software.
6.2.3.4 Invisible Range/Deadband
Sensor Installation
Note: When programming the invisible range, 18.3 cm (7.2 in.) must be added to the desired range
to compensate for the internal deadband distance between the sensor, the reflector, and the bottom
of the sensor housing.
The 940 flow meter is equipped with an invisible range feature to prevent false echoes
from top of channel walls, ladder rungs, shelves, etc. A user-selected range is defined that
is invisible to the flow meter. Do not extend the invisible range to where it meets or
overlaps the highest expected level in the channel.
Measure the distance between the bottom of the sensor housing and the object that is to
be excluded from the level measurement. Add 18.3 cm (7.2 in.) to the measured distance
to obtain the total invisible range (deadband) distance. Enter the total invisible range value
into the application software.
6.2.4 Protecting the In-Pipe Ultrasonic Sensor
Keep the sensor and the reflector free of grease and dirt. Since the logger “listens” for the
relatively faint sound of the returning echo, a heavily coated sensor will not be able to
detect the echo well and may not provide accurate level measurement.
51
Page 54
Sensor Installation
6.2.5 Troubleshooting the In-Pipe Sensor
During surcharged conditions, the ultrasonic level gives random depth readings that need
to be edited in the application software.
Symptoms of this hydraulic condition occur when the level appears erratic within 5.7 cm
(2.25 in.) of the pipe diameter. The readings first appear to be flat-lined. Later, the erratic
readings (that in some cases exceed the pipe diameter) occur.
If the hydraulics are frequently in surcharge conditions, use a submerged level sensor
instead of an in-pipe ultrasonic sensor.
6.3Installing the Submerged Area/Velocity Sensor
Read the instructions carefully before installing the sensor. Failure to do so could result in
serious injury to the operator or damage to the equipment.
6.3.1 Zeroing the Submerged Area/Velocity Sensor
The sensor has been factory-calibrated and compensated for temperature. The sensor
needs to be zeroed during each installation, but does not require calibration. The sensor
should be zeroed when moving it from one flow meter or sample to another.
6.3.2 Important Installation Guidelines for Submerged Area/Velocity Sensor
•Do not install more than one sensor at a time in pipes less than 61 cm (24 inches).
Multiple sensors in smaller pipes can create turbulent or accelerated flows near the
sensors that may cause inaccurate measurements.
•Mount the sensor as close as possible to the bottom of the pipe invert to accurately
measure low velocity levels.
•Do not monitor flows in the invert of the manhole itself. The best sensor location is 3 to
5 times the sewer diameter/height upstream of the invert.
•Locate monitoring sites as far as possible from inflow junctions to avoid interference
caused by combined flows.
•Avoid sites that contain flow obstacles within 2 to 4 pipe diameters in front of the
sensor installation (rocks, stones, pipe joints, valve stems, etc.) as these contribute to
turbulence and generate high speed flows in the immediate vicinity of the obstruction.
•Avoid any sites with slow moving flows that encourage the build-up of silt in the invert
or channel. Excessive silting around the sensor may inhibit the Doppler signal and
decrease sensor accuracy, and may affect depth measurement accuracy.
•Avoid sites with deep, rapid flows that make it physically difficult or dangerous to
install the sensor.
•Avoid sites with high velocity, low-depth flows. Splash-over and excessive turbulence
might be present around the sensor and data may be inaccurate.
52
Page 55
6.3.2.1 Proper Strain Relief of Submerged Area/VelocitySensor Cable
Attach the desiccant hub to the instrument handle to provide a strain relief for the sensor
cable and connector (Figure 28).
Sensor Installation
Figure 28 Proper Strain Relief
1Flow Meter 3Carabineer
2Desiccant Hub 4Lanyard
6.3.3 Connecting the Submerged Area/Velocity Sensor to the Mounting Bands
Important Note: If using an oil-filled sensor, replenish the oil prior to mounting the sensor
to a mounting band. Refer to section 7.8.3 on page 62 for oil replenishment instructions, if
applicable.
1. Attach the sensor to the mounting band. Mounting bands come with pre-drilled holes
for direct mounting of the sensor to the band.
2. Route the cable along the edge of the band and fasten the cable to the mounting band
with nylon wire ties (Figure 29 on page 54) to reduce the likelihood of debris collecting
on the cable and the mounting band. The cable should exit the tied area at, or near
the top of the pipe to keep it out of the flow stream.
Note: If there is a large amount of silt at the bottom of the pipe, rotate the band until the sensor is out
of the silt (Figure 30 on page 54), assuring that the sensor remains below the minimum expected
water level at all times. The silt should not be disturbed and must be measured frequently.
53
Page 56
Sensor Installation
4
3
2
Figure 29 Attaching the Sensor to the Mounting Band
1Nylon Wire Ties 3Sensor
2Screws (2)4Mounting band
1
1
4
2
3
Figure 30 Avoiding Silt when Mounting the Sensor
1Pipe2Sensor3Silt4Water
6.3.4 Placing the Submerged Area/Velocity Sensor and the Mounting Band into the Pipe
Point the angle-face of the sensor into the flow. The manufacturer recommends placing
the sensor with the arrow pointing at the flow (Figure 31 on page 55). For other mounting
configurations, see the appropriate flow meter manual.
Slide the mounting band as far as possible into the pipe to eliminate drawdown effects
near the end of the pipe. Locate the sensor at the bottom-most point in the channel. If
excessive silt is present on the bottom of the pipe, rotate the band in the pipe until the
sensor is out of the silt.
54
Page 57
Manhole
Velocity Sensor
Upstream PositionNormal
Sensor Installation
Manhole
Velocity Sensor
Downstream Position
Figure 31 Placing the Sensor into the Flow
55
Page 58
Sensor Installation
56
Page 59
Section 7Maintenance
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.
DANGER
Seul un technicien qualifié peut effectuer les tâches d'installation décrites dans
cette section du manuel.
DANGER
All maintenance activities related to flow meter and flow meter accessories must be
performed in a non-hazardous location.
DANGER
Toutes les opérations de maintenance relatives au débitmètre et à ses accessoires
doivent être exécutées dans un environnement qui ne présente aucun danger.
7.1Maintaining the Battery Compartment Desiccant
The desiccant material is a silica gel indicator. When the gel is saturated with moisture, the
beads turn from blue to pink. To rejuvenate the beads for reuse, remove them from the
assembly. Heat the beads in an oven at 100 to 180 °C (212 to 350 °F) until the beads turn
blue again. Discard and replace the beads if they do not turn blue after heating.
7.1.1 Replacing the Battery Compartment Desiccant
A small desiccant cartridge in the battery compartment (Figure 17 on page 32 and
Figure 18 on page 32) prevents moisture damage to the batteries and power circuitry.
Table 9 contains replacement part numbers.
If the normally blue desiccant beads turn pink, replace the cartridge or remove the
cartridge end cap and recharge or replace the desiccant material.
A small clip holds the desiccant cartridge in place. Pull the cartridge straight out of the clip
to remove it.
Table 9 Replacement Desiccant
DescriptionPart Number
Model 911 Battery Compartment Desiccant Cartridge4933
Model 940 Battery Compartment Desiccant Cartridge4967
Bulk Desiccant Refill (1.5 lb.)3624
7.2Maintaining the O-Ring Gasket on the End Cap
Note: Do not use petroleum jelly to lubricate the O-rings.
Two O-ring gaskets are installed in each flow meter end cap. The end caps maintain the
water tight seal on the flow meter. Use care when the end caps are removed—do not soil,
cut, or nick the gaskets. Replace them immediately if any physical damage is apparent.
Use a light coating of O-ring lubricant to help maintain a water-tight seal. Table 10 on page
57 lists replacement O-ring gaskets.
Table 10 Replacement O-rings
DescriptionPart Number
Model 911 End Cap O-ring (2 required)4912
Model 940 End Cap O-ring (2 required)4807
57
Page 60
Maintenance
7.3Cleaning the Flow Meters
Clean the outside of the logger case with warm water and mild detergent. Do not use
solvents or harsh cleaners to clean the logger. Do not use high pressure washing
equipment to clean the case.
7.4Storing the Flow Meters
Remove all batteries from the logger when storing for long periods (more than 3 months).
Store the sensors in a dry area. Make sure that all desiccant materials in the probes are
kept fresh (blue) at all times.
7.5Cleaning the Connectors
Coat connectors with lithium or di-electric grease to prevent corrosion.
7.6Maintaining the Electronics Compartment
7.6.1 Fuses
All internal fuses are self-resetting and require no maintenance. There are no other
user-serviceable components inside the electronics compartment.
7.6.2 Memory Batteries
An internal battery is provided to power RAM memory as well as the real time clock and
associated circuitry. The memory battery keeps the programmed settings as well as the
logged data whenever the main power fails or is removed during transport, storage, or
replacement. The internal memory battery is located in the electronic compartment and is
not user replaceable.
The internal memory battery has a life expectancy in excess of five years with the main
battery installed. If the main battery is removed for an extended period of time, the life of
the internal memory
battery will be substantially shorter. The user can see a low memory battery indicator
using the appropriate application software. See the software manual for details.
58
Page 61
7.7Replacing the Batteries in the RS232 Interface
DANGER
Never replace batteries in a hazardous location.
DANGER
Ne jamais remplacer les piles dans une zone dangereuse.
The Opto-Isolated RS232 interface uses a quantity of industrial quality Alkaline “C” cells.
Under normal circumstances the batteries should last for more than 1,000 downloads
based on an average download time of 5 minutes. The batteries need to be changed once
a year due to the limited shelf life of any type of battery.
1. Remove the cover from the RS232 interface (Figure 32).
2. Remove the old batteries. Install fresh batteries into the RS232 interface, making
certain the polarity is correct.
Maintenance
Figure 32 Replacing RS232 Batteries
7.8Maintaining the Submerged Area/Velocity Sensor
7.8.1 Cleaning the Transducer Port
Clean the transducer port when:
•A drift in the readings is observed.
•Level data are missing or incorrect but velocity data are valid.
•Excessive silt has deposited between the transducer and its protective cover.
59
Page 62
Maintenance
7.8.2 Cleaning the Submerged Area/Velocity Sensor (Oil-filled and Non-oil)
Important Note: DO NOT interchange an oil-filled protective cover plate with a non-oil
cover plate. This adversely affects level readings. It is possible to convert one type of
sensor to the other using the Oil Probe Conversion Kit (Cat. No. 7730000), refer to the Oil
Probe Conversion Kit Instruction Sheet, Cat. No. 7730089 for more information.
Important Note: When cleaning the transducer, use the most gentle technique possible.
Do not use sharp or pointed object to remove sediment from the face of the transducer. If
the transducer is nicked or dented, it breaks!
1. Soak the sensor in soapy water
Note: Do not soak the sensor in bleach. Bleach permanently damages the sensor. Refer to
Table 11 for acceptable cleaning solutions.
2. Remove the screws from the protective cover (Figure 33 and Figure 34 on page 61).
3. Remove the cover and gasket.
4. Swirl the sensor carefully in an appropriate cleaning solution to remove soil. Use a
spray or squeeze bottle to wash away heavier deposits.
5. Clean the gasket and the cover. Replace the gasket (Cat. No. 7722000) if it is torn or
damaged. Level readings are adversely affected if the gasket is damaged or not
installed.
6. Reattach the gasket and the cover (note orientation in Figure 34 on page 61). Tighten
the screws until the gasket starts to compress.
7. If using an oil-filled sensor, continue to follow the procedure mentioned in
section 7.8.3 on page 62.
Table 11 Cleaning Solutions
AcceptableUnacceptable
Dish Detergent and WaterConcentrated Bleach
Window CleanerKerosene
Isopropyl AlcoholGasoline
Dilute AcidsAromatic Hydrocarbons
60
Page 63
Maintenance
5
Figure 33 Removing the Protective Cover (Non-Oil Sensor)
1Screws (#6–32 x 5/16) 4Pressure Transducer
2 Protective Cover5Sensor
3Gasket
1
2
3
4
1
6
Figure 34 Removing the Protective Cover (Oil-filled Sensor)
1Screws (#6–32 x 5/16) 4Gasket
2 Protective Cover5Pressure Transducer
3Screw, set, #2-566Sensor
2
3
4
5
61
Page 64
Maintenance
7.8.3 Replenishing the Oil
The manufacturer recommends inspecting the oil in the sensor for large air bubbles during
the customer-scheduled service duty cycle, and prior to every installation. Small bubbles
(less than ¼-in. diameter) of air within the oil do not affect performance. Larger bubbles
may minimize the anti-fouling benefit of the oil.
1. Remove the yellow tape on the sensor, if the sensor is new.
2. Remove any debris from the sensor.
3. Load the oil cartridge into the dispensing gun (Figure 35).
4. Twist the feed tube onto the cartridge and attach the syringe tip to the feed tube
(Figure 36 on page 63).
5. Press the dispenser gun handle to purge any air bubbles from the syringe tip.
6. Remove the set screw in the transducer cover with the supplied 0.035 hex wrench.
Retain the set screw.
7. Insert the syringe tip slowly into the set screw hole and dispense the oil (Figure 37 on
page 63).
Note: While dispensing the oil, hold the probe at an angle to allow the air to be pushed out the
side port (Figure 37 on page 63).
Important Note: Slowly insert the syringe tip and do not dispense oil during insertion
or damage to the transducer may occur if too much pressure is applied. Continue to
dispense the oil until all the air bubbles are removed.
8. Continue to dispense the oil while removing the syringe from the set screw hole to
prevent air bubbles. Replace the set screw until it is flush with the transducer cover
and remove any excess oil around the screw hole or on the sensor.
9. Clean the entire probe and place a piece of electrical tape over the side port to
prevent oil from leaking out. Remove the tape from the sensor prior to zeroing and
installing the sensor.
1
2
Figure 35 Loading the Cartridge into the Dispensing Gun
1Dispensing Gun2Silicone Oil Cartridge
62
Page 65
Figure 36 Attaching the Feed Tube and Syringe
Maintenance
Figure 37 Oil Replenishment
1Sensor3Side port
2Set screw4Syringe
7.8.4 Changing the Submerged Area/Velocity Sensor Desiccant
The desiccant canister contains beads of silica gel which ensure proper orientation of the
pressure transducer. When the beads are blue, they can remove moisture from the air.
When they are pink, they are saturated and cannot absorb any more moisture from the air,
and they must be replaced immediately.
Important Note: When the beads begin to turn pink, replace or recondition the beads.
Permanent damage to the sensor may occur if the desiccant is not maintained. Never
operate the sensor without the proper desiccant. When rejuvenating beads, remove them
from the canister and heat at 100–180 °C (212–350 °F) until the beads turn blue. If the
beads do not turn blue, replace them with new beads. Do not heat the canister.
63
Page 66
Maintenance
7.8.5 Replacing the Desiccant
Note: Replacing the desiccant does not require that the desiccant container be removed from the
desiccant box.
1. Twist the bottom end-cap using a slight twisting motion until its slots align with the
retaining clips (Figure 38).
2. Remove the end cap gently by grasping it and pulling it straight out.
3. Pour the desiccant beads out of the canister.
4. Hold the canister up to the light and inspect the hydrophobic filter.
•If there is a small, dim light spot while looking through the hole, the filter is in good
condition. If there is a bright light spot, the filter is probably torn. Replace the filter.
•If the desiccant beads are completely saturated with water or the filter has been
saturated with water or grease, replace the filter.
Note: If the filter needs to replaced, complete the Replacing the Hydrophobic Filter
(section 7.8.5.1 on page 65), then continue with step 5.
5. Refill the canister tube with blue desiccant beads (Cat. No. 3624). Inspect the O-ring
(Cat. No. 5252) on the bottom cap for cracks, pits, or evidences of leakage. Replace if
necessary.
Note: Applying O-ring grease to the new or dry O-rings improves the ease of insertion, sealing,
and life span of the O-ring.
6. Make sure that the O-ring is clean and free of dirt or debris before replacing the end
cap.
7. Reinstall the end cap.
Figure 38 Removing the Bottom End Cap
1End Cap2Desiccant Container
64
Page 67
7.8.5.1 Replacing the Hydrophobic Filter
A single Teflon® hydrophobic filter (Cat. No. 3390) is installed in the top of the canister to
prevent liquid from entering the vent tube.
For best performance and to avoid grease buildup on the filter during submergence or
surcharge conditions, hang the canister vertically so that the end facing the sensor points
downward.
1. Disconnect the tubing from the top of the desiccant canister.
2. Unscrew the hex-head tubing nipple from the top of the canister and discard the old
filter.
3. Discard any remnants of Teflon tape from the nipple’s threads. Reapply two turns of
Teflon tape (Cat. No. 10851-45) to the threads, pulling the tape into the threads until it
conforms to the shape of the threads.
4. Place a new filter over the hole. Make sure that the smooth side of the filter faces the
inside of the canister.
5. Place the threaded nipple on top of the filter.
6. With a slight pressure, press the filter into the hole with the nipple threads and begin
threading the nipple into the hole. The filter deflects upward and feed completely into
the thread until it disappears. The filter must rotate with the nipple as it is threaded into
the cap. If it does not, it is torn. Start over with a new filter.
Maintenance
7. Inspect the installation. A small, dim light spot should be visible in the upper cap when
held up to the light. A bright spot indicates a torn filter. Start over with a new filter.
12
Figure 39 Replacing the Hydrophobic Filter
1Filter, smooth side down2Hex-head tubing nipple3Finished assembly
Manhole Spanner Bars, 18-27 in., used with suspension cable9542
Manhole Spanner Bars, 28-48 in., used with suspension cable9557
Suspension cable for hanging flow meter4920
User Manual4979
8.2Sensors
DescriptionPart Number
In-Pipe/Ultrasonic Level Sensor4741
75 kHz Ultrasonic Sensor5233-05
Submerged Area Velocity Sensors
I.S. Oil-Filled Submerged AV Sensor, 0–10 ft range, 30 ft cable, with connector88064-030
I.S. Oil-Filled Submerged AV Sensor, 0–10 ft range, 50 ft cable, with connector88064-050
I.S. Oil-Filled Submerged AV Sensor, 0–10 ft range, 75 ft cable, with connector88064-075
I.S. Oil-Filled Submerged AV Sensor, 0–10 ft range, 100 ft cable, with connector88064-100
I.S. Oil-Filled Submerged AV Sensor, 0–30 ft range, 30 ft cable, with connector88074-030
I.S. Oil-Filled Submerged AV Sensor, 0–30 ft range, 50 ft cable, with connector88074-050
I.S. Oil-Filled Submerged AV Sensor, 0–30 ft range, 75 ft cable, with connector88074-075
I.S. Oil-Filled Submerged AV Sensor, 0–30 ft range, 100 ft cable, with connector88074-100
I.S. Standard Submerged AV Sensor, 0–10 ft range, 30 ft cable, with connector88065-030
I.S. Standard Submerged AV Sensor, 0–10 ft range, 50 ft cable, with connector88065-050
I.S. Standard Submerged AV Sensor, 0–10 ft range, 75 ft cable, with connector88065-075
I.S. Standard Submerged AV Sensor, 0–10 ft range, 100 ft cable, with connector88065-100
I.S. Standard Submerged AV Sensor, 0–30 ft range, 30 ft cable, with connector88075-030
I.S. Standard Submerged AV Sensor, 0–30 ft range, 50 ft cable, with connector88075-050
I.S. Standard Submerged AV Sensor, 0–30 ft range, 75 ft cable, with connector88075-075
I.S. Standard Submerged AV Sensor, 0–30 ft range, 100 ft cable, with connector88075-100
8.3Sensor Mounting Rings and Bands
DescriptionPart Number
Submerged Pressure/Velocity Mounting Rings for 6–24 in. Pipes
6 inches1361
8 inches1362
67
Page 70
Replacement Parts and Accessories
8.3Sensor Mounting Rings and Bands (continued)
DescriptionPart Number
10 inches1363
12 inches1364
15 inches1365
18 inches1366
21 inches1353
24 inches1370
Submerged Pressure/Velocity Mounting Bands for 15 to 42 in. Pipes
15 inches9706100
18 inches9706200
21 inches9706300
24 inches9706400
27 inches9706500
30 inches9706600
33 inches9706700
36 inches9706800
42 inches9706900
15 to 42 inches3766
In-Pipe Mounting Rings for 6 to 12 in. Pipes
6 inches4021
8 inches4022
10 inches4023
12 inches4024
In-Pipe Sensor Mounting Bands for 15 to 42 in. Pipes
15 inches9706100
18 inches9706200
21 inches9706300
24 inches9706400
24 inches9706500
30 inches9706600
33 inches9706700
36 inches9706800
42 inches9706900
15 to 42 inches3766
8.4Ultrasonic Sensor Mounting Hardware
Description
Permanent Wall Mount Bracket2974
Adjustable for Floor or Wall2904
Tripod w/ Mounting Bracket for Sensor9538
Cable Straightener2883
Cable Grip (Requires #2883)3183
68
Part Number
Page 71
Replacement Parts and Accessories
8.5Mounting Plate Hardware for Directly Mounting to Pipe Wall
Sensor Type
Submerged Pressure4939
In-Pipe3875
Part Number
69
Page 72
Replacement Parts and Accessories
70
Page 73
Section 9Contact Information
Ordering information for the U.S.A.
By Mail:
Hach Company
By Telephone:
(800) 368-2723
By Fax:
301-874-8459
Information Required
•Hach account number (if available)•Billing address
•Your name and phone number•Shipping address
•Purchase order number•Catalog number
•Brief description or model number•Quantity
Ordering Information for Outside the U.S.A. and Europe
Hach maintains a worldwide network of dealers and distributors. To locate the representative nearest
you, send an e-mail to: hachflowsales@hach.com or visit ww.hachflow.com.
Technical and Customer Service Department personnel are eager to answer questions about our
products and their use. In the U.S.A., call 1-800-635-1230. Outside the U.S.A. and Europe, send E-mail
to hachflowservice@hach.com or call 1-301-874-5599.
Repair Service
Authorization must be obtained from Hach Company before sending any items for repair.
To send the monitor to the factory for repair:
1. Identify the serial number of the monitor unit.
2. Record the reason for return.
3. Call the Customer Service Department (1-800-368-2723) and get a Service Request Number (SRN) and
shipping label.
4. Use the shipping label provided and ship the equipment in the original packaging if possible.
Note: Do not ship manuals, computer cables, or other parts with the unit unless they are required for repair.
5. Make sure the equipment is free from foreign debris and is clean and dry before shipping. Sensors returned
without cleaning will be charged a fee.
6. Write the SRN number on the shipping box.
7. Make sure that all return shipments are insured.
8. Address all shipments to:
Hach Company
5600 Lindbergh Drive - North Dock
Loveland, Colorado, 80539-0389 U.S.A.
Attn: SRN#XXX
71
Page 74
Section 10Contact information for Europe
For technical support, repair service and ordering information please refer to the contact
information below.
For all countries except France, Spain and Great Britain:
Flow-Tronic
RUE J.H. COOL 19a
B-4840 Welkenraedt
Belgium
Ph: +-32-87-899797 or 899799
Fx: +-32-87-899790
Email: info@flow-tronic.com
www.flow-tronic.com
For France, Spain and Great Britain:
France
HACH LANGE FRANCE S.A.S.33
Rue du Ballon 93165 Noisy-le-Grand
Telephone: ++33 (0)1 48 15 68 70
Fax.: ++33 (0)1 48 15 80 00
Email: info@hach-lange.fr
www.hach-lange.fr
Spain
HACH LANGE, S.L.U
C/ Larrauri, 1C, 2ª Pl.
48160 Derio, Bizkaia
Telephone: 902 131441 94 6573388
Fax: 94 6573397
E-mail: info@hach-lange.es
www.hach-lange.es
Great Britain:
HACH LANGE LTD
Pacific Way
Salford
Manchester
M50 1DL
Telephone: 0 161 872 1487
Fax.: 0 161 872 7324
Email: sales@hach-lange.co.uk
www.hach-lange.co.uk
72
Page 75
Section 11Limited Warranty
Hach Company warrants its products to the original purchaser against any defects that are due to faulty
material or workmanship for a period of one year from date of shipment unless otherwise noted in the
product manual.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its
option, it will repair or replace the defective product or refund the purchase price excluding original
shipping and handling charges. Any product repaired or replaced under this warranty will be warranted
only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable
components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned
without authorization from Hach Company.
Limitations
This warranty does not cover:
•Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or undeclared),
terrorism, civil strife or acts of any governmental jurisdiction
•Damage caused by misuse, neglect, accident or improper application or installation
•Damage caused by any repair or attempted repair not authorized by Hach Company
•Any product not used in accordance with the instructions furnished by Hach Company
•Freight charges to return merchandise to Hach Company
•Freight charges on expedited or express shipment of warranted parts or product
•Travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its
products. All implied warranties, including without limitation, the warranties of merchantability and fitness
for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in
your state the above limitation may not apply to you. This warranty gives you specific rights, and you
may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person
is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive
remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in
no event shall Hach Company be liable for any incidental or consequential damages of any kind for
breach of warranty or negligence.
73
Page 76
74
Page 77
Appendix A Working with Primary Devices
A.1 Working with Primary Devices and Sensor Operation
Installing the sensor in any primary device requires familiarity with the proper location for
head measurement in that particular primary device. Just as the proper installation of the
primary device itself is critical to obtaining optimum flow measurement accuracy, the
appropriate location of the submerged or ultrasonic sensor also greatly affects flow
measurement accuracy.
Always consult the manufacturer of the primary device, whenever possible, for details
concerning the proper location of the sensor.
A.2 Setting an Offset (For Use in a Weir)
When installing a submerged sensor behind a weir, the user may need to “offset” the
actual reading from the submerged sensor to compensate for the difference in height
between the submerged sensor and the crest of the weir. This allows the user to place the
sensor at an arbitrary height in the weir, as long as it is located below the crest of the weir
(and at the proper distance upstream from the weir plate). See Figure 40 on page 76 for
details on head measurement locations in a weir.
In the text example below:
•The distance between the crest of the weir (bottom of the ‘V’) and the submerged
sensor is 4 in. (10 cm).
•The distance between the surface of the water and the submerged sensor is 6 in. (15
cm).
•The level of water which is contributing to flow is 2 in. (5 cm) (water flowing over the
weir).
Entering 2 in. (5.1 cm) using the LEVEL ADJUST key provides the desired offset. As the
water level falls to the bottom of the ‘V” the flow meter reads zero level and zero flow. If the
level falls below the crest of the weir, due to evaporation or other reason, the level reads a
negative number and the flow remains at zero.
The key to proper level setting is to adjust the level (using the LEVEL ADJUST key) to the
level that is contributing to flow. In a round pipe the level contributing to flow is the distance
from the surface of the water to the invert (bottom) of the pipe. In a flume, the level
contributing to flow is the distance from the surface of the water to the floor of the flume.
A.3 Types of Primary Devices
Use the primary device illustrations that follow as general guides for proper head
measurement location in commonly used primary devices. Refer to Figure 41 on page 77
for Parshall Flume, Figure 42 on page 78 for Palmer-Bowlus Flume (with Integral
Approach), Figure 43 on page 79 for Leopold-Lagco Flume, Figure 44 on page 80 for
H-Flume and Figure 45 on page 80 for Round Pipes. Contact the primary device
manufacturer for more details.
75
Page 78
Working with Primary Devices
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaa
Submerged
Probe
FLOW
Maximum
Head Height
Submerged
H
2H
Minimum
Crest
Height
Probe
Figure 40 Weir
Head
Measurement
Point
FLOW
Minimum Distance from
4H
Weir Plate
Draw down
Crest
Weir Plate
76
Page 79
Working with Primary Devices
A Length of Converging Section
2/3A Length to Stilling Well
Stilling Well
(Optional)
FLOW
Submerged
Probe
Submerged
Probe
Submerged Flow
Free Flow
Head Measurement Point
TOP VIEW
ELEVATION
H
Figure 41 Parshall Flume
77
Page 80
Working with Primary Devices
Submerged Probe
(Preferred Head Measurement Point)
FLOW
Top View
Preferred Head Measurement Point
Upstream Depth
Upper Transition
Water Surface
FLOW
"D"
Conduit Diameter
2
Throat
Small jump should occur here.
Elevation View
Figure 42 Palmer-Bowlus Flume (with Integral Approach)
Lower
Transition
Downstream Depth
78
Page 81
Submerged Probe
(Preferred Head Measurement Point)
Working with Primary Devices
Upper Transition
Preferred Head Measurement Point
FLOW
Water Surface
FLOW
Top View
Lower TransitionThroat
Small jump should occur here.
End ViewElevation View
Upstream Depth
"D"
Conduit Diameter
2
Figure 43 Leopold-Lagco Flume
Downstream Depth
79
Page 82
Working with Primary Devices
Approach
Section
FLOW
FLOW
Inlet Bulkhead
Elevation View
Head Measurement Section
4
3
Pipe Stub
Figure 44 H-Flume
Top View
Top View
W
1
2
H
T
WP
D
FLOW (Q)
A
d
ΔY
FLOW (Q)
ΔX
Elevation View
Figure 45 Round Pipes
80
Page 83
Appendix B Channel Installation Options
Figures 46 through 50 show how to install the 911 and 940 Flow Meters in various
applications.
Figure 46 shows Model 911 installed in a manhole with a single level/velocity sensor
stored in the small-diameter pipe.
Figure 46 Small-Diameter Pipes Down to Street or Building Collectors
81
Page 84
Channel Installation Options
The 940 and the 911 Flow Meters do not require profiling to establish average velocity,
allowing setup in a dry channel (Figure 47). Rapid signal processing and temperature
compensation accurately record the change from dry to wet conditions.
Figure 47 Installation Setup in a Dry Channel
82
Page 85
Channel Installation Options
Figure 48 shows a storm vault accepting flow from two locations. The 940 Flow Meter
monitors level, velocity, and flow in both pipes and sends data via modem to a central
location.
Figure 48 Installation for Multiple Source Monitoring at a Single Site
83
Page 86
Channel Installation Options
When a Combined Sewer Overflow (CSO) is conducive to accurate flow measurement,
the 940 Flow Meter can measure depth and velocity in the primary channel and measure
discharge in the overflow channel. See Figure 49. In some instances, the overflow in
some CSO applications may be too turbulent for accurate measurement.
Figure 49 Installation for Combined Sewer Overflow (CSO) Applications
84
Page 87
Channel Installation Options
Custody transfer or interagency billing demand redundant measurement for chain of
custody. In Figure 50, the 940 Flow Meter provides ultrasonic measurement of flow in a
Parshall flume while also measuring flow upstream using the Continuity Equation (Q=AV).
This ensures no loss of data in a submerged flow condition.
Figure 50 Installation for Custody Transfer or a Billing Arrangement
85
Page 88
Channel Installation Options
86
Page 89
Appendix C Batteries and Chargers
DANGER
Never install, remove or charge batteries in a hazardous location.
DANGER
Vous ne devez jamais installer, retirer ou recharger les batteries dans un
environnement dangereux.
C.1 Lead-Acid (Gel Cell) Batteries for 911 Flow Meter
The manufacturer’s lead-acid batteries are designed to prevent electrolyte leakage from
the terminals or case. The electrolyte is suspended in a gel, which ensures safe, efficient
operation in any position. Gel cells are classified as “Non-Spillable” and meet all
requirements of the International Air Transport Association.
Maintenance
The manufacturer’s lead-acid batteries are maintenance-free.
DANGER
The use of chargers other than those provided by the manufacturer for use with
these batteries is not recommended and may damage the battery or shorten its life.
DANGER
Il est déconseillé d'utiliser des chargeurs autres que ceux fournis par le fabricant
pour ces batteries, car ils risquent de les endommager et de réduire leur vie utile.
Charge
The manufacturer’s lead-acid cells are designed to be fully charged in 22 to 24 hours
using their lead-acid battery charger. Do not exceed 24 hours or life of the battery may be
damaged or shortened. The charge rate is 500 mA dc. The LED is lit, indicating the battery
is charging. The battery is fully charged when the LED indicator turns off.
Temperature
At higher temperatures, the electrical capacity that can be taken out of a battery increases.
At lower temperatures, the electrical capacity that can be taken out of a battery decreases.
However, excessive heat ruins batteries. Avoid placing batteries near heat sources of any
kind. To maximize battery life, operate the battery at an ambient temperature of 20 °C (70
°F). The permissible operating temperature range is -15 to 50 °C (5 to 120 °F), however,
use in the 5 to 35 °C (47 to 95 °F) temperature range is recommended.
Disposal
•911 Flow Meter: The battery pack assembly for the 911 Flow Meters are
rechargeable/disposable. Follow local applicable disposal guidelines for
gel–electrolyte type batteries. The entire battery housing is meant to be disposed
of after the service life of the battery. Do not attempt to open the case to gain
access to the batteries, there are no user serviceable parts inside.
Storage
Store lead-acid batteries in a cool, dry place. Their low self-discharge rate and excellent
charging characteristics permit storage for up to one year without loss of efficiency or
appreciable deterioration of battery performance.
At room temperature the self-discharge rate of lead-acid batteries is approximately 3% of
rated capacity per month. The self-discharge rate varies as a function of ambient storage
temperature.
Figure 51 shows the storage self-discharge characteristics of lead-acid batteries at
various ambient temperatures. Table 12 shows recommended storage times for lead-acid
batteries.
Storage TemperatureMaximum Recommended Storage Time
0 to 20 °C12 months
21 to 30 °C9 months
31 to 40 °C5 months
41 to 50 °C2.5 months
C.2 Rechargeable Lithium Ion Battery for 940 Flow Meter
0°C
10 °C
20 °C
30 °C
40 °C
A Lithium Ion rechargeable battery pack (6608500) is available for the 940 Flow Meter
only. This is an alternative to the lead-acid and primary metallic Lithium
(non-rechargeable) battery offered as an option for the 940 Flow Meter.
The Lithium Ion rechargeable battery pack is designed to allow up to 500
charge/discharge cycles and is shipped with a charger designed specifically for this
battery pack. The Lithium Ion charger has specific plug wiring to prevent use with the
primary non-rechargeable metallic Lithium pack or the lead-acid pack.
Do not attempt to use the Lithium Ion battery pack charger with the primary
non-rechargeable metallic Lithium pack or the lead-acid pack. Do not attempt to use the
lead-acid pack charger with the Lithium Ion battery pack.
DANGER
The use of chargers other than those provided by the manufacturer for use with
these batteries is not recommended and may damage the battery or shorten battery
life.
DANGER
Il est déconseillé d'utiliser des chargeurs autres que ceux fournis par le fabricant
pour ces batteries, car ils risquent de les endommager et de réduire leur vie utile.
Maintenance
This battery pack requires no maintenance other than charging.
Charge
Inspect the Lithium Ion battery pack and the battery connection for damage before use or
charging. Do not attempt to use or repair a damaged battery pack assembly.
88
Page 91
Batteries and Chargers
Use only the charger provided for the Lithium Ion pack (6678000). The pack can be
recharged at any time regardless of charge level. Total discharge will have a negative
impact on the number of charge/recharge cycles allowable and should be avoided. The
Lithium Ion battery can be fully charged in 8 to 12 hours using the charger provided. The
charger will terminate the charge automatically and float-charge the battery until it is
removed from the charger. The charge rate is 600 mA DC.
When the charger is plugged into a suitable AC voltage source the power LED will
illuminate. When the battery pack is connected the charge LED will illuminate yellow. A
flashing yellow charge LED indicates that the battery is deeply discharged but the charge
cycle is starting. If the flashing yellow charge LED does not stop flashing within 1 hour the
battery pack is shorted and should be disposed of properly. The battery is fully charged
when the yellow charge LED light goes off.
Temperature
Excessive heat ruins batteries. At higher temperatures, the electrical capacity that can be
taken out of a battery increases. At lower temperatures, the electrical capacity that can be
taken out of a battery decreases. Avoid placing batteries near heat sources of any kind. To
maximize battery life, operate the battery at an ambient temperature of 20 °C (70 °F). The
permissible operating temperature range is -15 to 50 °C (5 to 120 °F). Use in the 5 to 35
°C (47 to 95 °F) temperature range is recommended.
Disposal
The Lithium Ion battery pack assembly for the 940 Flow Meter is rechargeable. Follow
local applicable disposal guidelines for Lithium Ion type batteries. The entire battery
housing is meant to be disposed of after the service life of the battery. Do not attempt to
open the case to gain access to the batteries. There are no user serviceable parts inside.
Batteries and Chargers
The 940 Flow Meter may use two different types of Lithium-based battery. These batteries
require special disposal requirements and must be treated as hazardous material. Air
cargo is the only acceptable air shipment permitted. The batteries must be individually
wrapped and boxed separately. Do not puncture, incinerate or expose to water. If local
government does not allow for the disposal of these batteries contact the Hach Company
for proper disposal arrangements.
Storage
Store all battery types in a cool, dry place. A low self-discharge rate and excellent charging
characteristics permit storage of Lithium Ion batteries for up to one year without loss of
efficiency or appreciable deterioration of performance. At room temperature the
self-discharge rate of Lithium Ion batteries is approximately 1% of rated capacity per
month.
89
Page 92
Batteries and Chargers
90
Page 93
Appendix D SCADA-Modbus® System Guidelines
D.1 Introduction to SCADA - Modbus Communications
Use this section as a guide when using the Modbus ASCII protocol to communicate
directly with the 940 Flow Meter via an RS232 or modem connection.
This guide assumes that the user has a working knowledge of Supervisory Control and
Data Acquisition (SCADA), its components, and the different topologies used to construct
the communications network. Because a basic understanding of the Modbus ASCII
protocol is necessary, the key pieces of the protocol will be described.
This section guides the users through the setup process by describing key points that
need to be addressed for successful communication. This section will not outline specific
implementation details of any particular Man Machine Interface (MMI) or controller,
although examples may reference certain manufacturers for illustrative purposes. The
description of the Modbus ASCII protocol is provided for reference only and is not
intended as a tutorial. The scope of this section is limited to the description of Modbus
ASCII as it pertains to the 940 Flow Meter.
Modbus, an open protocol, determines how each instrument knows its device address,
recognize a message addressed to it, determine the type of action to be taken, and extract
any data or other information contained in the message. The flow meter and Man Machine
Interface (MMI) communicate using a master-slave technique in which only the master
can initiate queries to a slave (940). The 940 is always considered the slave, never a
master. The master can address individual 940 Flow Meters or can broadcast a message
to instruments within its scope. Responses are never returned to broadcast queries from
the master. The Modbus protocol establishes the format for the master’s query by placing
into it the device address, a function code defining the requested action, any data to be
sent, and an error-checking field. The flow meter’s response message is constructed
using the Modbus format which confirms the action to be taken, any data to be returned,
and an error checking field.
D.2 ASCII Transmission Mode
The 940 Flow Meter is designed to communicate on standard Modbus networks using
Modbus ASCII. In ASCII mode, messages start with a colon ‘:’, and end with a ‘carriage
return-line feed’ pair. The allowable characters transmitted for all fields are hexadecimal
0–9, and A–F. When a message is transmitted over a Modbus ASCII communication link,
each character or byte is sent in the order of Least Significant Bit to Most Significant Bit. A
typical message frame looks like the following:
The address field of an ASCII message frame, ranging from 0 to 247 decimals, consists of
two characters that represent the slave address. Individual slaves are assigned values
between 1 and 247. A master addresses a slave by putting the slave’s address in the
address field of the message frame. When the slave sends its response, it places its own
address in the address field of the message frame to let the master know which slave is
responding.
END
(HEX)
91
Page 94
SCADA-Modbus® System Guidelines
D.4 Function Field
The function code field of an ASCII message frame, ranging from 1 to 255 decimals,
consists of two characters that represent the type of action the master is requesting from
the slave. Of these functions, the 940 Flow Meter currently supports function 3 (Read
Holding Registers). When a message is sent from the master to a slave device, the
function field tells the slave what kind of action to perform. For example, this may include
reading the channel values of Level and Velocity. When the slave responds to the master,
it echoes the function code field to indicate a normal response. In the event of an error,
such as parity error, LRC error, or a request that cannot be handled, the slave will not
respond and the master will eventually process a time-out condition.
D.5 Data Field
The data field of an ASCII message frame consists of n pairs of ASCII characters that
represent data sent to or from a slave device (flow meter). The data field contained in the
master request contains additional information that is required by the slave before any
action takes place. This may include channel register addresses, the number of registers
to read, and the actual byte count in the data field. For example, if a master requests that
the flow meter read the current status of a group of channels (function code 03), the data
field specifies the starting register and how many registers are to be read. If no error
occurs, the data field of the response from the meter to the master contains the data
requested.
D.6 LRC Field
The LRC field of an ASCII message frame consists of two ASCII characters that provide
an additional level of error checking to verify the integrity of the communication media.
The LRC field is one byte that contains an 8-bit binary value. The LRC value is calculated
by the transmitting device, which appends the LRC to the end of the message. The
receiving device recalculates the LRC and compares it against the LRC value of the
incoming message. If the two values are not equal, an error condition occurs. The LRC is
calculated by adding together successive 8-bit bytes of the message, discarding any
carries, and then complementing the result. The LRC is calculated by summing all values
in the ASCII message except for the leading ‘colon’ and ending <CR><LF>.
D.7 Communication Parameters
To successfully communicate with the 940 Flow Meter using Modbus ASCII, the
communication parameters of the master device must be set at 7 bits, Even Parity, and 1
Stop bit. The baud rate may be configured to any value offered by the 940 Flow Meter.
With the exception of baud rate, the communication parameters must not vary from this
format.
D.8 User Memory Customizing
The most familiar component of existing SCADA networks today is the Programmable
Logic Controller (PLC). Because the network integrator is most familiar with this type of
device, the flow meter emulation of an existing PLC simplifies the process of integrating
the manufacture’s instrumentation into the SCADA network. Modbus ASCII uses a
referencing system to identify the various types of memory inputs and outputs. Each
reference number has a leading digit that identifies its data type (discrete input, discrete
output, register input, register output) followed by a string of digits that indicates its
location in RAM (Table 13).
92
Page 95
SCADA-Modbus® System Guidelines
Table 13 Modbus ASCII Memory Input/Output Referencing System
Reference IndicatorReference TypeMeaning
0xxxxDiscrete Output or CoilBinary
1xxxxDiscrete InputBinary
3xxxxInput RegisterReal
4xxxxOutput Holding RegisterReal
6xxxxExtended Memory RegisterReal
The memory data is stored in 16-bit words. Within the predefined function codes of the
Modbus ASCII protocol, the data fields are subject to interpretation by the device
manufacturer. For example, the 940 Flow Meter places temperature information in
registers 40001-40002.
D.9 Modbus ASCII Function Codes Supported
Currently, the 940 Flow Meter is capable of a read-only function to retrieve channel and
total flow information. All data addresses in the Modbus ASCII message are referenced to
zero. Therefore, a reference to holding register 40001 is addressed as register 0000. The
function code field specifies the type of register accessed; therefore, the 4XXXX is implicit.
Function 03: Read Holding Registers
Reads the register (4X reference) contents of the 940 Flow Meter as defined in the tables
that follow.
The Modbus ASCII query must take the form shown below that specifies the starting
register and number of registers to be read:
Start
‘:’
Slave
Address
Function
(03)
Start
Address
High
Start
Address
Low
No. of Pts.
High
No. of.
Pts. Low
LRC<CR><LF>
For example, to read the level channel of the 940 Flow Meter, the query must be as
indicated in Table 18.
Table 18 Channel Query to Read Level (Example)
Start‘:’
Slave Address01
Function03
Starting Address High00
Starting Address Low06
No. of Registers High00
No. of Registers Low02
LRCF4
Stop<CR><LF>
The master queries the flow meter using a Read Holding Registers request, which implies
a 4XXXX register reference, to slave device address 01. The message requests data from
holding registers 40007–40008 to obtain the level information, which requires two
registers to store the floating point value. Note that registers are referenced from zero in
the data field.
95
Page 98
SCADA-Modbus® System Guidelines
D.11 Response
The 940 Flow Meter responds with the following transmission reflecting a level reading of
15.0 inches:
Table 19 Transmission Response that Reflects a 15 in. Level Reading
Start‘:’
Slave Address01
Function03
Byte Count04
Data High00
Data Low00
Data High41
Data Low70
LRC47
Stop<CR><LF>
The flow meter response echoes the address and function code, which indicates that no
problems exist in the communication between the master and 940. The ‘Byte Count’ field
specifies how many 8-bit data items are being returned in the data field. With Modbus
ASCII, this is one-half the actual count of ASCII characters transmitted in the data portion
of the response. The contents of 40007 are shown as two byte values of 00 00 hex, and
the contents of register 40008 are shown as two byte values 41 70 hex. Together, these
values represent the floating point IEEE representation of the level status.
D.12 940 Flow Meter Response Time
As a result of time lags associated with data acquisition, instrumentation could
conceivably take up to 12 seconds to respond to a SCADA RS232 request. Therefore, the
SCADA system must be designed to accommodate this potential communication lag. For
example, in a Wonderware® application running a Modbus ASCII DDE server, the COM
port reply time-out must be set to 12 seconds. This is the amount of time that the meter is
given to reply to Modbus queries via this serial port.
Communication Handshaking
The 940 Flow Meter contains minimal communication handshaking. For the meter to
identify an RS232 connection from an outside source, and to keep the RS232 hardware
active once communicating, the Data Terminal Equipment (DTE) must assert and hold
high the DTR line of the DB9 connector (DSR of meter). The 940 Flow Meter does not
support RTS/CTS hardware handshaking. Note that DTE must be capable of handling a
12-second maximum response lag.
PinDescriptionPinDescriptionPinDescription
1
Pin 1Data Carrier Detect (DCD)
Pin 2Received Data (RD)Pin 5Signal Ground (SG)Pin 8Clear to Send (CTS)
Pin 3Transmitted Data (TD)Pin 6Data Set Ready (DSR)Pin 9Ring Indicator
1
Not used.
Pin 4Data Terminal Ready (DTR)Pin 7Request to Send (RTS)
96
Page 99
SCADA-Modbus® System Guidelines
D.13 Complications with Floating Point Values
The manufacturer’s implementation of the Modbus protocol was based on the idea that we
would enable our flow meters to emulate a Modicon
we follow the exact same format that Modicon uses for the storing and processing floating
point numbers. Additionally, the Modbus protocol does not define how floating point values
are packed (stored) into the internal memory addresses or “Registers” of the flow meter.
When integrating the Modbus-capable flow meters, be aware that these meters store and
process floating point numbers in the exact same format as the Modicon Compact 984
PLC.
All current models of Modicon PLCs, including the Compact 984, pack two bytes of data
into each register. This alone presents no problems. Unsigned two-byte (16-bit) integer
values in the range of 0 to 65535 can be stored and retrieved from these registers without
any problems or complications. The complications arise when the stored value is a floating
point value, which by IEEE definition, require 4 bytes (32 bits). The IEEE standard for
floating point values states in part that the 8 most significant bits represent the exponent
and the remaining 23 bits (plus one assumed bit) represent the mantissa and the sign of
the value.
Since a data “word” consists of two bytes, a floating point value is represented by two data
words. Because a single Modicon register consists of one word (or 2 bytes), two
consecutive Modicon registers are needed to store one floating point value.
The representation of a floating point value can be broken down into a “High Order” and a
“Low Order” word. Additionally, each word can be broken down into a high order byte and
a low order byte.
®
, Compact 984 PLC. Consequently,
Table 20 and Table 21 depict how a IEEE floating point value is usually represented and
how the Modicon stores a floating-point value.
The complications arise because Modicon doesn't store floating point values in this
standard (IEEE) format. Modicon stores floating point values the opposite way with the
“Low-order” word in the first register and the “High-order” word in the second register.
Table 20 IEEE Floating Point Representation
First Register (i.e., 4001)Second Register (i.e., 4002)
High Word, High ByteHigh Word, Low ByteLow Word, High ByteLow Word, Low Byte
Table 21 Floating Point Values Representation
First Register (i.e., 4001)Second Register (i.e., 4002)
Low Word, High ByteLow Word, Low ByteHigh Word, High ByteHigh Word, Low Byte
Since the Modbus protocol doesn't define how floating point values are handled or stored,
some Modbus-capable servers incorrectly use the normal, “High word — Low word”
format for converting the Modbus message response to the client application. Since
Modicon stores the floating point values in the opposite order, the result is that the client
application may receive an incorrect representation of the floating point value from the
Modbus-capable server.
97
Page 100
SCADA-Modbus® System Guidelines
D.14 Port Expanders and Protocol Converters
In some situations, there may not be a Modbus ASCII port available for use with the 940
Flow Meter. A good example might be where there is a need to install a flow meter at a
remote pump site that already has a single Modbus line connected to a PLC that is used to
control the pumps.
Port expanders are available from third party manufacturers; these allow several Modbus
slave devices to be connected to a single Modbus Master device. Typically, a single port
expander has 3–5 separate Modbus ports on it. Depending on the manufacturer, the user
may be able to configure each of these ports for different communications parameters. In
essence, not only does this type of port expander allow multiple slave devices to be
connected to a single Modbus master device, but it can also be configured to convert
incompatible communications parameters such as Modbus ASCII to RTU (or vice versa),
baud rate, parity, stop bits, etc.
In addition to the port expanders mentioned above, other protocol converters from
third-party manufacturers can be used to convert other industrial protocols to Modbus
ASCII.
AEG Schneider Automation. Modicon ladder logic block library user guide. North
Andover, MA: AEG Schneider Automation, Inc. 1996.
D.16 Troubleshooting Tips
Problem: 940 Flow Meter responds to some Modbus messages but not all
Response: Check the Register Addresses
The flow meter only responds to valid Modbus message requests. If a Modbus message
sent to the flow meter asks for stored register addresses for values that are outside of the
address range currently supported by the meter, the meter ignores the request.
The flow meter currently only supports register addresses 40001 through 40083.
Consequently, a request to read the value in any register address greater then 40083 will
be ignored. If a range of registers is requested and that range includes register addresses
greater then 40083, the entire request will be ignored
Response: Check the number of registers being polled
Additionally, the 940 Flow Meter checks all Modbus messages to see if the correct number
of registers is requested for the type of data being returned. The meter ignores the request
if the number of registers requested does not coincide with the correct number of registers
needed to accurately display the data. For example, Velocity is a floating point value
stored in register 40009–40010. Because all floating point values require two registers,
the meter would ignore a request to read just the data in register 40009, yet it would
respond correctly to a request to read the data stored in both registers 40009 AND 40010.
Consequently, if the meter received a single request to read both Level 40007–40008 and
Velocity 40009–40010, the request would have to be for an even number of registers for
the meter to respond.
98