Hach 910, 930, 920 Instrument Manual

Page 1
Catalog Number 4975
Sigma Flow Meter Models
910, 920, and 930
INSTRUMENT MANUAL
05/03 7ed
Page 2
Visit http: //www.hach.com
Page 3
Catalog Number 4975
Sigma Flow Meter Models
910, 920, and 930
INSTRUMENT MANUAL
eac 05/03 7ed
Page 4
Visit http: //www.hach.com
Page 5
Table of Contents
Safety Precautions .................................................................................................................................................... 5
Specifications ............................................................................................................................................................ 9
OPERATION ......................................................................................................................................................... 13
Section 1 Getting Started................................................................................................................................. 15
1.1 Items Required for Installing and Using the Flow Meter ................................................................................... 15
1.2 Optional Accessories for Use with the Flow Meter............................................................................................ 15
1.3 Planning for Success: Considerations for Installing Your Flow Meter ............................................................... 15
1.3.1 Choosing the Proper Site ........................................................................................................................ 15
1.3.2 Dealing with Difficult Sites ....................................................................................................................... 16
1.3.3 Using the Proper Batteries ...................................................................................................................... 16
1.3.4 Mounting the Flow Meter with the Proper Orientation ............................................................................. 16
1.3.5 Preventing Damage to the Flow Meter in a Harsh Environment .............................................................. 16
1.4 Choosing the Appropriate Meter and Sensor Combination .............................................................................. 16
1.4.1 Types of Sensors ..................................................................................................................................... 17
Section 2 Software and Communications ................................................................................................... 19
2.1 Required Software ............................................................................................................................................ 19
2.2 Programming and Analysis Options .................................................................................................................. 19
2.3 RS232 Serial Port (All Models) ......................................................................................................................... 19
2.4 Modem Communications (Models 920 and 930 Only) ...................................................................................... 21
2.5 Sampler Receptacle (Models 920 and 930 Only) ............................................................................................. 21
2.5.1 Sampler Receptacle Pin Descriptions ..................................................................................................... 21
2.5.2 Logging Sample Information on Models 920 and 930 from a Hach Sampler .......................................... 22
INSTALLATION AND MAINTENANCE............................................................................................................ 23
Section 3 Channel Installation Options........................................................................................................ 25
Section 4 Meter Installation............................................................................................................................. 31
4.1 Battery Information............................................................................................................................................ 31
4.1.1 Installing Batteries ................................................................................................................................... 31
4.1.2 Replacing Batteries in Models 910 and 920 ............................................................................................ 31
4.1.3 Replacing Batteries in Model 930............................................................................................................ 31
4.1.4 Battery-Life Estimates ............................................................................................................................. 32
4.2 Mounting the Flow Meter................................................................................................................................... 33
4.3 Connecting Sensors to the Flow Meter............................................................................................................. 34
4.3.1 Ultrasonic Level Sensor........................................................................................................................... 34
4.3.2 Level/Velocity Sensor .............................................................................................................................. 34
4.3.3 Rain Gauge ............................................................................................................................................. 35
Section 5 Maintenance ..................................................................................................................................... 37
5.1 Sensor Calibration............................................................................................................................................. 37
5.1.1 Velocity Sensor Calibration...................................................................................................................... 37
5.1.2 Submerged Sensor Calibration ............................................................................................................... 37
Page 3
GettingStartedTOC.fm Table of Contents
Page 6
Table of Contents
5.2 Desiccant Maintenance..................................................................................................................................... 37
5.2.1 Reusing Desiccant................................................................................................................................... 37
5.2.2 Replacing the Sensor Desiccant ............................................................................................................. 37
5.2.3 Replacing the Battery and Electronics Compartment Desiccant ............................................................. 38
5.3 Maintaining the O-Ring Gasket on the End Cap ............................................................................................... 38
5.4 Cleaning the Flow Meter ................................................................................................................................... 38
5.5 Storing the Flow Meter ...................................................................................................................................... 38
5.6 Maintenance in the Electronics Compartment .................................................................................................. 39
5.6.1 Fuses....................................................................................................................................................... 39
5.6.2 Memory Batteries .................................................................................................................................... 39
Section 6 Exploded View Drawings............................................................................................................... 41
GENERAL INFORMATION ................................................................................................................................45
Replacement Parts and Accessories ...................................................................................................................... 47
Contact Information for U.S.A. and Outside Europe ............................................................................................... 49
Contact Information for Europe............................................................................................................................... 50
Warranty ................................................................................................................................................................. 51
Page 4 Table of Contents
GettingStartedTOC.fm
Page 7

Safety Precautions

Please read this entire manual before unpacking, setting up, or operating this instrument.
Pay particular attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that which is specified in this manual.
Use of Hazard Information
If multiple hazards exist, this manual will use the signal word (Danger, Caution, Note) corresponding to the greatest hazard.
DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury.
NOTE Information that requires special emphasis.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists and indicates that only individuals qualified to work with hazardous voltages should open the enclosure or remove the barrier.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol, when noted on the product, indicates that the marked item can be hot and should not be touched without care.
This symbol, when noted on the product, indicates the presence of devices sensitive to Electro-static Discharge and indicates that care must be taken to prevent damage to them.
This symbol, when noted on the product, identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
Page 5
Safety.fm Safety Precautions
Page 8
Safety Precautions
Hazardous Locations
The 910/920/930 Flow Meters are not approved for use in hazardous locations as defined in the National Electrical Code.
DANGER Although some Hach products are designed and certified for installation in hazardous locations as defined by the National Electrical Code, many Hach products are not suitable for use in hazardous locations. It is the responsibility of the individuals who are installing the products in hazardous locations to determine the acceptability of the product for the environment. Additionally, to ensure safety, the installation of instrumentation in hazardous locations must be per the manufacturer's control drawing specifications. Any modification to the instrumentation or the installation is not recommended and may result in life threatening injury and/or damage to facilities.
Confined Space Entry
The following information is provided to guide users of 950 Flow Meters on the dangers and risks associated with entry into confined spaces.
DANGER Additional training in Pre-Entry Testing, Ventilation, Entry Procedures, Evacuation/Rescue Procedures and Safety Work Practices is necessary to ensure against the loss of life in confined spaces.
On April 15, 1993, OSHA's final ruling on CFR 1910.146, Permit Required Confined Spaces, became law. This standard directly affects more than 250,000 industrial sites in the United States and was created to protect the health and safety of workers in confined spaces.
Definition of Confined Space
A Confined Space is any location or enclosure that presents or has the immediate potential to present one or more of the following conditions:
An atmosphere with less than 19.5% or greater than 23.5% oxygen and/or more than 10 ppm Hydrogen Sulfide (H
An atmosphere that may be flammable or explosive due to gases, vapors, mists, dusts, or fibers.
Toxic materials which upon contact or inhalation, could result in injury, impairment of health, or death.
Confined spaces are not designed for human occupancy. They have restricted entry and contain known or potential hazards.
Examples of confined spaces include manholes, stacks, pipes, vats, switch vaults, and other similar locations.
Standard safety procedures must always be followed prior to entry into confined spaces and/or locations where hazardous gases, vapors, mists, dusts, or fibers may be present. Before entering any confined space check with your employer for procedures related to confined space entry.
2
S).
FCC Requirements
Page 6 Safety Precautions
1. The Federal Communications Commission (FCC) has established Rules
which permit this device to be directly connected to the telephone
Safety.fm
Page 9
Safety Precautions
network. Standardized jacks are used for these connections. This equipment should not be used on party lines or coin lines.
2. If this device is malfunctioning, it may also be causing harm to the telephone network; this device should be disconnected until the source of the problem can be determined and until repair has been made. If this is not done, the telephone company may temporarily disconnect service.
3. The telephone company may make changes in its technical operations and procedures; if such changes affect the compatibility or use of this device, the telephone company is required to give adequate notice of the changes.
4. If the telephone company requests information on what equipment is connected to their lines, inform them of:
a. The telephone number that this unit is connected to,
b. The ringer equivalence number [1.4B]
c. The USOC jack required [RJ11C], and
d. The FCC Registration Number
Items (b) and (d) are indicated on the label. The ringer equivalence number (REN) is used to determine how many devices can be connected to your telephone line. In most areas, the sum of the RENs of all devices on any one line should not exceed five. If too many devices are attached, they may not ring properly.
Service Requirements
In the event of equipment malfunction, all repairs should be performed by Hach Company or an authorized agent. It is the responsibility of users requiring service to report the need for service to Hach Company. or to one of our authorized agents. Service can be facilitated throughout our office:
Hach Company 5600 Lindbergh Drive, P.O. Box 389 Loveland, CO 80539 Telephone: (970) 669-3050
Equipment Attachment Limitations
Notice: The Canadian Industry Canada label identifies certified equipment.
This certification means that the equipment meets certain telecommunications network protective, operational and safety requirements. The Department does not guarantee the equipment will operate to the user's satisfaction.
Before installing this equipment, users should ensure that it is permissible to be connected to the facilities of the local telecommunications company. The equipment must also be installed using an acceptable method of connection. In some cases, the company's inside wiring associated with a single line individual service may be extended by means of a certified connector assembly (telephone extension cord). The customer should be aware that compliance with the above conditions may not prevent degradation of service in some situations.
Repairs to certified equipment should be made by an authorized Canadian maintenance facility designated by the supplier. Any repairs or alterations
Page 7
Safety.fm Safety Precautions
Page 10
Safety Precautions
made by the user to this equipment, or equipment malfunctions, may give the telecommunications company cause to request the user to disconnect the equipment.
Users should ensure for their own protection that the electrical ground connections of the power utility, telephone lines and internal metallic water pipe system, if present, are connected together. This precaution may be particularly important in rural areas.
CAUTION
Users should not attempt to make such connections themselves, but should contact the appropriate electric inspection authority, or electrician, as appropriate.
The Load Number (LN) assigned to each terminal device denotes the percentage of the total load to be connected to a telephone loop which is used by the device, to prevent overloading. The termination on a loop may consist of any combination of devices subject only to the requirement that the total of the Load Numbers of all the devices does not exceed 100.
Page 8 Safety Precautions
Safety.fm
Page 11
Specifications are subject to change without notice.
Model 910
Dimensions 11.4 cm dia. x 44.8 cm L (4.5" dia. x 17.625" L)
Weight 3.54 kg (7.8 lbs.) with batteries
Enclosure Material PVC
Enclosure Rating NEMA 6P (IP67)
Operating Temperature Range -18 to 60 °C (0 to 140 °F)
Storage Temperature Range -40 to 60 °C (-40 to 140 °F)
Power Source One Energizer EN-529, Alkaline, 6 V dc battery

Specifications

Battery Life
User Interface PC
Monitoring Intervals 1, 2, 3, 5, 6, 10, 12, 15, 20, 30, 60 minutes
Program Memory Non-volatile, programmable flash, can be updated via RS232 port
Time-Based Accuracy ±1 second per day
Units of Measurement
Data Storage
Communications Serial connection to IBM compatible computer with Hach analysis software
Velocity Measurement Accuracy
60 days typical, with a 15-minute recording interval, one level and one velocity, one data download per week, at 10 °C (50 °F) (also affected by site conditions)
Level: in., m, cm, ft.
Flow: gps, gpm, gph, lps, lpm, lph, mgd, afd, cfs, cfm, cfh, cfd, m
Totalized Flow: gal., ft.³, acre-ft., L, m³
Capacity: 90 days of one level reading and one velocity reading at 15-minute
recording intervals
Data Types: Level and velocity
Storage Mode: Wrap or slate
RAM Memory: 128K
Method: Doppler ultrasonic
Transducer Type: Twin 1 MHz piezoelectric crystals
Typical minimum depth for velocity: 2 cm (0.8 in.)
Range: -1.52 to 6.10 m/s (-5 to 20 fps)
Zero Stability: 0.015 m/ss (<0.05 fp)
Accuracy: ±2% of reading
Operating Temperature: -18 to 60 °C (0 to 140 °F)
3
s, m3m, m3h, m3d
Page 9
Specs.fm Specifications
Page 12
Specifications
Model 920
Dimensions 16.8 cm dia. x 44.7 cm L (6.625" dia. x 17.625" L)
Weight 7.5 kg (16.5 lbs.) with batteries
Enclosure Material PVC
Enclosure Rating NEMA 6P (IP67)
Operating Temperature Range -18 to 60 °C (0 to 140 °F)
Storage Temperature Range -40 to 60 °C (-40 to 140 °F)
Power Source Two Energizer EN-529, Alkaline, 6 V dc batteries
Battery Life
User Interface PC
Monitoring Intervals 1, 2, 3, 5, 6, 10, 12, 15, 20, 30, 60 minutes
Program Memory Non-volatile, programmable flash, can be updated via RS232 port
Time-Based Accuracy ±1 second per day
Units of Measurement
Data Storage
Communications Serial connection to IBM compatible computer with Hach analysis software
Modem
Local Terminal RS232 at 19.2k baud
Velocity Measurement Accuracy
90 days typical, with a 15-minute recording interval, one level and one velocity, one data download per week, at 10 °C (50 °F) (also affected by site conditions)
Level: in., m, cm, ft.
Flow: gps, gpm, gph, lps, lpm, lph, mgd, afd, cfs, cfm, cfh, cfd, m
Totalized Flow: gal., ft.³, acre-ft., L, m³
Capacity: 240 days of two level readings and two velocity readings at a 15-minute
recording intervals
Data Types: Level, velocity, and rainfall
Storage Mode: Wrap or slate
RAM Memory: 128K
14400, V.32 bis,V.42, MNP2-4 error correction. V.42 bis, MNP5 data compression. MNP10EC Cellular Protocol
Method: Doppler ultrasonic
Transducer Type: Twin 1 MHz piezoelectric crystals
Typical minimum depth for velocity: 2 cm (0.8 in.)
Range: -1.52 to 6.10 m/s (-5 to 20 fps)
Zero Stability: 0.015 m/ss (<0.05 fp)
Accuracy: ±2% of reading
Operating Temperature: -18 to 60 °C (0 to 140 °F)
3
s, m3m, m3h, m3d
Page 10 Specifications
Specs.fm
Page 13
Model 930
Dimensions 21.9 x 60 cm (8.625" dia. x 23.625" L)
Weight 16.2 kg (35.7 lbs.) with batteries
Enclosure Material PVC
Enclosure Rating NEMA 6P (IP67)
Operating Temperature Range -18 to 60 °C (0 to 140 °F)
Storage Temperature Range -40 to 60 °C (-40 to 140 °F)
Power Source Six Energizer EN-529, Alkaline, 6 V dc batteries
Specifications
Battery Life
User Interface PC
Monitoring Intervals 1, 2, 3, 5, 6, 10, 12, 15, 20, 30, 60 minutes
Program Memory Non-volatile, programmable flash, can be updated via RS232 port
Time-Based Accuracy ±1 second per day
Units of Measurement
Data Storage
Output Conditions Set point on level, velocity, rainfall, flow, or flow rate of change
Sampler Output 6 to 12 V dc pulse, 100 mA max. at 500 ms duration
Communications Serial connection to PC computer with Hach analysis software
Modem
365 days typical, with a 15-minute recording interval, one level and one velocity, one data download per week, at 10 °C (50 °F) (also affected by site conditions)
Level: in., m, cm, ft.
Flow: gps, gpm, gph, lps, lpm, lph, mgd, afd, cfs, cfm, cfh, cfd, m
Totalized Flow: gal., ft.³, acre-ft., L, m³
Capacity: 175 days of three level readings, three velocity readings, and rainfall
measurments at a 15-minute recording intervals
Data Types: Level, velocity, and rainfall
Storage Mode: Wrap or slate
RAM Memory: 512K
14400, V.32 bis,V.42, MNP2-4 error correction. V.42 bis, MNP5 data compression. MNP10EC Cellular Protocol
3
s, m3m, m3h, m3d
Local Terminal RS232 at 19.2k baud
Method: Doppler ultrasonic
Transducer Type: Twin 1 MHz piezoelectric crystals
Typical minimum depth for velocity: 2 cm (0.8 in.)
Velocity Measurement Accuracy
Specs.fm Specifications
Range: -1.52 to 6.10 m/s (-5 to 20 fps)
Zero Stability: 0.015 m/ss (<0.05 fp)
Accuracy: ±2% of reading
Operating Temperature: -18 to 60 °C (0 to 140 °F)
Page 11
Page 14
Visit http: //www.hach.com
Page 15

OPERATION

DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les Fiches de Données de Sécurité des Produits (FDSP) et se familiariser avec toutes les procédures de sécurité avant de manipuler tous les produits chimiques.
PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligrosa. Revise las fichas de seguridad de materiales y familiarícese con los procedimientos de seguridad antes de manipular productos químicos.
GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden. Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden Materialsicherheitsdatenblätter aufmerksam zu lesen.
PERICOLO
La manipolazione di campioni, standard e reattivi chimici può essere pericolosa. La preghiamo di prendere conoscenza delle Schede Techniche necessarie legate alla Sicurezza dei Materiali e di abituarsi con tutte le procedure di sicurezza prima di manipolare ogni prodotto chimico.
Page 13
OpStop.fm OPERATION
Page 16
Visit http: //www.hach.com
Page 17

Section 1 Getting Started

1.1 Items Required for Installing and Using the Flow Meter

Batteries
Use only Energizer EN-529, Alkaline, 6 V dc batteries (Cat. No. 3667).
Battery Compartment Desiccant
Use a desiccant cartridge to avoid moisture damage to the batteries and electronics. See Desiccant Maintenance on page 37 for more information.
Level and Velocity Sensors
Select the level or velocity sensors that best suit your application. For more information, see Choosing the Appropriate Meter and Sensor Combination on
page 16.
Mounting Apparatus
Select the appropriate Logger and Sensor mounting apparatus from
Replacement Parts and Accessories on page 47.
Programming and Data Retrieval Software
These flow meters require the use of InSight integrated sewer system management application. InSight is recommended for small- to medium-sized flow studies. For broader applications, Vision supports automated data collection with an unlimited number of sensors. See
Software and Communications on page 19 for more information.
®
or Vision®, a high-end,

1.2 Optional Accessories for Use with the Flow Meter

Rain Gauge
Connect your flow meter to the Hach tipping bucket rain gauge to record rainfall. See Rain Gauge on page 35 for more information.
Wastewater Sampler
Connect an optional Hach wastewater sampler to the flow meter. Models 920 and 930 can set the pace for an external sampler in proportion to flow. In storm water applications, these models can trigger the sampler in response to water level or rainfall.

1.3 Planning for Success: Considerations for Installing Your Flow Meter

1.3.1 Choosing the Proper Site

The accuracy of flow measurements greatly depends on the suitability of the monitoring site. Select sites that have normalized flow and minimal turbulence. Turbulence can make it difficult to detect an average velocity in the flow stream. Obstructions, vertical drops, pipe bends, and elbows can create turbulence and affect the accuracy of your measurements. Ta b le 1 contains suggestions for preventing turbulence.
Page 15
GettingStarted.fm Getting Started
Page 18
Section 1
Table 1 Suggestions for Preventing Turbulence
Site Condition Suggested Remedy
Outfalls Attach the sensor at least ten times the maximum expected level upstream of the outfall.
Vertical drops in the channel floor
Elbows, sharp turns, and “Y” connections

1.3.2 Dealing with Difficult Sites

1.3.3 Using the Proper Batteries

Attach the sensor at least ten times the maximum expected level upstream of the vertical drop.
Attach the sensor at least ten times the maximum expected level downstream of the vertical drop.
Locate the sensor at least ten times the maximum expected level upstream of the impediment.
Locate the sensor at least ten times the maximum expected level downstream of the impediment.
Some sites may be difficult to monitor due to poor site conditions. The direction and the speed of the particles in the flow stream contribute to the signal received by the velocity sensor. If turbulence near the measurement point is excessive, it may be difficult for the sensor to determine the average velocity of the stream. Hach loggers provide several unique features to help deal with these problem sites. See Compensating for Velocity Direction in your sensor manual.
Use only Energizer EN-529, Alkaline, 6 V dc batteries. These batteries are required to achieve the rated performance of the flow meter. Other batteries will produce unacceptable results. See Battery Information on page 31 for more information.

1.3.4 Mounting the Flow Meter with the Proper Orientation

You must mount the flow meter so that the connectors face down. The
end of the meter that contains the batteries should face up. If the connectors face up, they may corrode and allow water to seep into the instrument.

1.3.5 Preventing Damage to the Flow Meter in a Harsh Environment

The connectors on your flow meter have protective caps. Cover the connectors with the protective caps when not in use to prevent corrosion.

1.4 Choosing the Appropriate Meter and Sensor Combination

Use Tab l e 2 to determine which sensors to use.
Table 2 Flow Meter and Sensor Configuration Options
Flow Meter Model Level and Velocity Sensor Configuration Options
910 Meter plus one Submerged Depth/Velocity Sensor or one Velocity Only Sensor
Meter plus two Submerged Depth/Velocity Sensors
920
930 Meter plus up to three Submerged Depth/Velocity Sensors and up to two Ultrasonic 75 kHz Sensors
or Meter plus one Submerged Depth/Velocity Sensor and one Ultrasonic 75 kHz Sensor
Page 16 Choosing the Appropriate Meter and Sensor Combination
GettingStarted.fm
Page 19

1.4.1 Types of Sensors

Section 1
Submerged Depth/Velocity Sensors
Submerged Depth/Velocity sensors can measure level and velocity simultaneously. A submerged pressure transducer measures level. Velocity is measured with sound waves, using the Doppler principle.
Wafer Velocity Sensor
The Wafer Velocity Sensor (Cat. No. 88005 or 88006) is an extremely low­profile velocity sensor. It does not measure level. The streamlined shape of the wafer probe allows velocity measurement in very low-flow conditions. When used in conjunction with a level sensor (such as the In-Pipe Ultrasonic Level Sensor), you can calculate flow.
In-Pipe Ultrasonic Sensor
Use the In-Pipe Ultrasonic Level Sensor (Cat. No. 3702-01 or 3702-02) in pipes where level measurement to the top of the pipe is desired. This sensor has no effective deadband and will read the level until liquid reaches the bottom of the sensor housing. The entire deadband is contained horizontally within the sensor body.
Ultrasonic Sensor (Downlook)
Downlook ultrasonic sensors are available with different beam angles, beam spread, and deadbands. Use the 75 KHz sensor with the 920 or 930 Flow Meter.
Page 17
GettingStarted.fm Choosing the Appropriate Meter and Sensor Combination
Page 20
Visit http: //www.hach.com
Page 21

Section 2 Software and Communications

2.1 Required Software

You can program or retrieve data from your flow meter via modem, DTU II, or direct connection to the RS232 serial port using a PC and one of the following software packages:
InSight small- to medium-sized flow monitoring jobs.
Vision high-end, automated data collection and automated batch report processing, using an unlimited number of meters. Vision is a Microsoft Windows
®
Data Analysis Software is a Microsoft Windows® program for
®
Integrated Sewer System Management Software provides
®
program.

2.2 Programming and Analysis Options

Built-in flow equations
Pager activation upon alarm condition
2-way modem and serial communications
Storage of 116,000 data points in battery-backed RAM Memory
Storm water sampling trigger
Rainfall measurement
Flow proportional sampler output
High, low, rate of change and trouble alarms
For details on programming your flow meter, refer to your software manual.

2.3 RS232 Serial Port (All Models)

The high speed RS232 serial port communicates with a PC or Data Transfer Unit (DTU) at speeds from 1200 to 19,200 baud as shown in Figure 1.
The optional extension cable (Cat. No. 3358) includes a built-in ladder hook and is designed to connect to a cable (Cat. No. 1727) near the surface for easy reach in manholes. See Figure 2.
Page 19
SoftwareCommunications.fm Software and Communications
Page 22
Section 2
Figure 1 Communications with the Flow Meter, DTU, and PC
1
2
3
1. DTU to Meter Cable (Cat. No. 1726) 2. DTU to PC Cable (Cat. No. 3513) 3. Meter to PC Cable (Cat. No. 1727)
Figure 2 Extension Cable with Ladder Hook
1
2
1. Cable, Flow Meter to PC (Cat. No. 1727) 2. Extension Cable with Ladder Hook (Cat. No. 3358
Page 20 RS232 Serial Port (All Models)
SoftwareCommunications.fm
Page 23
Figure 3 Serial Port Pins
Section 2
Table 1Serial Port Pin Descriptions
Pin Letter Description Wire Color
A B+ (used to power DTU) White
B Ground Blue
CDSRYellow
D RCD Black
EDTRRed
FTXDGreen
B
C
A
F
E
D

2.4 Modem Communications (Models 920 and 930 Only)

CAUTION Use care when making modem connections as high voltage may be present on the phone wires!
Figure 4 Modem Receptacle Pins
Models 920 and 930 can be configured with a 14,400 baud, cellular capable internal telephone modem (Cat. No. 4872).
These advanced, very low power modems let you communicate with Hach loggers over long distances using public telephone lines or a cellular phone.
The Model 920 and 930 modems communicate at speeds from 300 to 14,400 baud. Connect the red and green wires from the logger modem connector to the red and green wires provided by the telephone company.
A
B
Pin Letter Designation Color Description
A Tip Green Ground
B Ring Red Signal
Table 2Modem Receptacle Pin Description

2.5 Sampler Receptacle (Models 920 and 930 Only)

2.5.1 Sampler Receptacle Pin Descriptions

The optional “sampler” port lets you connect Model 920 or 930 to a Hach wastewater sampler. This option provides several features:
Flow proportional sampling
Stormwater trigger, based on level and/or rate of rainfall. Useful for the
USEPA NPDES permitting program.
Sample history logging
Page 21
SoftwareCommunications.fm Modem Communications (Models 920 and 930 Only)
Page 24
Section 2
Table 3Sampler Receptacle Pin Description
Pin Letter
A White Not used.
B
Ground
C
Flow Pulse
Output
D
Sampler
Start
E
Event Input
F
Bottle
Number
Input
Wire
Color
Blue Provides the ground line that is used in conjunction with the other signals on this connector.
Ye l l o w
Black
Red
Green
Description
Used in conjunction with Pin B (ground) to tell a sampler that a pre-determined amount of flow has accumulated.
Rating: +8 V dc. Output current is 100 mA dc (max.) with a duration of 500 ms.
Used to “wake up” a wastewater sampler when a level and/or rainfall set-point condition is met so that it can begin sampling. Used in conjunction with Pin B (ground), this line is normally allowed to float and is switched to ground (by transistor) once the set-point conditions are met.
Rating: +24 V dc (max.)
This signal is received from a Hach wastewater sampler. It confirms that a sample has been collected. The time and date of each water sample appears on the data printout when downloaded using a Data Transfer Unit or Hach PC software. Rating: +12 V dc (max.) minimum 3 sec. pulse.
This signal is received from a Hach wastewater sampler. It is used in conjunction with the “Event Input” signal described above. It tells the flow meter which bottle was used and when a sample was taken. This information will appear in the data printout (see “Sample Times and Dates”) when downloaded using a Data Transfer Unit or Hach Support Software.
Bottle Number signal description
If the Program Complete Output (pin F of the aux. connector) is disabled on the sampler, then it is used to transmit the bottle number to the connected device.
Time A = 200 msec Time B = 100 msec (50 ms HI 50 ms LO)
Note: If a sample attempt fails, the first pulse width 'B' is 150 msec HI and 50 msec LO.
Note: If Sample Distribution is programmed for multiple bottles per sample mode, only the first
bottle number of the set is transmitted via pin-F.
Cable Required
Multi-Purpose Half Cable Assembly (Cat. No. 941): 3 m (10') long, 6-pin connector on one end, tinned wire leads on the other end. For use with non-Hach samplers.
or Multi-Purpose Full Cable Assembly (Cat. No. 940): 3 m (10') long, 6-pin connector on both ends. Custom sizes: 7.6 m lengths (25 ft.) and custom sizes are also available.

2.5.2 Logging Sample Information on Models 920 and 930 from a Hach Sampler

To log sample information in Models 920 or 930 from Hach samplers:
1. Connect the appropriate cable (Cat. No. 940 or 540) to the “Sampler” receptacle on the flow meter and the “Auxiliary” receptacle on the sampler.
2. Program the sampler for Special Output/Sample Output. See the appropriate sampler manual for details.
Page 22 Sampler Receptacle (Models 920 and 930 Only)
SoftwareCommunications.fm
Page 25

INSTALLATION AND MAINTENANCE

DANGER
Some of the following manual sections contain information in the form of warnings, cautions and notes that require special attention. Read and follow these instructions carefully to avoid personal injury and damage to the instrument. Only personnel qualified to do so, should conduct the installation/maintenance tasks described in this portion of the manual.
DANGER
Certains des chapitres suivants de ce mode d’emploi contiennent des informations sous la forme d’avertissements, messages de prudence et notes qui demandent une attention particulière. Lire et suivre ces instructions attentivement pour éviter les risques de blessures des personnes et de détérioration de l’appareil. Les tâches d’installation et d’entretien décrites dans cette partie du mode d’emploi doivent être seulement effectuées par le personnel qualifié pour le faire.
PELIGRO
Algunos de los capítulos del manual que presentamos contienen información muy importante en forma de alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas instrucciones a fin de evitar accidentes personales y daños al instrumento. Las tareas de instalación y mantenimiento descritas en la presente sección deberán ser efectuadas únicamente por personas debidamente cualificadas.
GEFAHR
Einige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von Warnungen, Vorsichtsmaßnahmen oder Anmerkungen, die besonders beachtet werden müssen. Lesen und befolgen Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder Schäden am Gerät zu vermeiden. In diesem Abschnitt beschriebene Installations- und Wartungsaufgaben dürfen nur von qualifiziertem Personal durchgeführt werden.
PERICOLO
Alcune parti di questo manuale contengono informazioni sotto forma d’avvertimenti, di precauzioni e di osservazioni le quali richiedono una particolare attenzione. La preghiamo di leggere attentivamente e di rispettare quelle istruzioni per evitare ogni ferita corporale e danneggiamento della macchina. Solo gli operatori qualificati per l’uso di questa macchina sono autorizzati ad effettuare le operazioni di manutenzione descritte in questa parte del manuale.
Page 23
InstMaintStop.fm INSTALLATION AND MAINTENANCE
Page 26
Visit http: //www.hach.com
Page 27

Section 3 Channel Installation Options

Figures 5 through 11 show how to install Models 910, 920, or 930 for various applications.
Figure 5 shows Model 910 installed in a manhole with a single level/velocity
sensor stored in the small-diameter pipe.
Figure 5 Small-Diameter Pipes Down to Street or Building Collectors
910
Page 25
ChannelOptions.fm Channel Installation Options
Page 28
Section 3
Figure 6 shows Model 930 with three level/velocity sensors that will determine
an average reading.
Models 910/920/930 do not require profiling to establish average velocity, allowing setup in a dry channel (Figure 7). Rapid signal processing and temperature compensation accurately record the change from dry to wet conditions.
Figure 6 Installation in a Large Diameter Pipe, Trunk, Interceptor, or Tunnel
930
Figure 7 Installation Setup in a Dry Channel
Page 26 Channel Installation Options
920
ChannelOptions.fm
Page 29
Figure 8 shows a storm vault accepting flow from three locations. A single
Model 930 monitors level, velocity, and flow in all three pipes and sends data via modem to a central location.
Figure 8 Installation for Multiple Source Monitoring at a Single Site
930
Section 3
Page 27
ChannelOptions.fm Channel Installation Options
Page 30
Section 3
When a Combined Sewer Overflow (CSO) is conducive to accurate flow measurement, a Model 920 can measure depth and velocity in the primary channel and measure discharge in the overflow channel with ±5% accuracy. See Figure 9.
Figure 9 Installation for Combined Sewer Overflow (CSO) Applications
920
Page 28 Channel Installation Options
ChannelOptions.fm
Page 31
Section 3
In some instances, the overflow in some CSO applications may be too turbulent for accurate measurement. Figure 10 shows Model 920 with two depth/velocity probes that will measure flow upstream and downstream of the outfall to determine the volume and duration of the overflow.
Figure 10 Installation for Combined Sewer Overflow (CSO) Applications with Turbulence in the Overflow
920
Page 29
ChannelOptions.fm Channel Installation Options
Page 32
Section 3
920
920
920
Custody transfer or interagency billing demand redundant measurement for chain of custody. In Figure 11, one Model 920 provides ultrasonic measurement of flow in a Parshall flume while also measuring flow upstream using the Continuity Equation (Q=AV). This ensures no loss of data in a submerged flow condition.
Figure 11 Installation for Custody Transfer or a Billing Arrangement
920
Page 30 Channel Installation Options
ChannelOptions.fm
Page 33

Section 4 Meter Installation

4.1 Battery Information

Use only Energizer®* EN-529, alkaline, 6 V dc batteries. These batteries are required to achieve the rated performance of the flow meter. Other batteries will produce unacceptable results. Non-alkaline or non-industrial grade batteries will damage the battery compartment and may void the warranty.
Energizer EN-529 batteries are readily available and provide sufficient power to operate the flow meter for extended periods. Use this model battery to obtain the rated operating duration and to safeguard the logger circuitry. Replacement batteries are available from Hach Company (Cat. No. 3667).

4.1.1 Installing Batteries

CAUTION Use only alkaline batteries in this product. Other types of batteries can result in safety hazards.
Note: Do not soil or damage the two
rubber O-rings on the end cap while removing or replacing the end cap. The O-rings provide a watertight seal for the battery compartment. Replace them if they are lost or damaged.
Opening the Model 910 and 920 Battery Compartments
1. Grasp the logger by the handle opposite of the interface connectors and
twist that end counter-clockwise to open.
2. Unscrew the end cap.
Opening the Model 930 Battery Compartment
1. Turn the winged fastener counter-clockwise to loosen it (it does not come
all the way off).
2. Pull the end cap of the 930 straight up and off of the logger.

4.1.2 Replacing Batteries in Models 910 and 920

Model 910 uses one 6 V dc alkaline battery. Model 920 uses two 6 V dc alkaline batteries. Install the fresh battery(s) into the battery compartment, terminal-end first. It does not matter which way the battery is rotated when it is inserted. The battery holder will accept the batteries regardless of polarity.
Use two fresh batteries each time in Model 920. Do not mix fresh batteries with used ones.

4.1.3 Replacing Batteries in Model 930

Model 930 uses six 6 V dc alkaline batteries in a removable battery pack. The battery pack is secured inside the battery compartment with two captive, winged fasteners. See Figure 12.
Use a complete set of fresh batteries each time. Do not mix fresh batteries with used ones.
Note: A full battery pack is heavy. Use care when removing it.
* Energizer is a registered trademark of Eveready Battery Company, Inc.
MeterInstallation.fm Meter Installation
1. Loosen the two captive fasteners.
2. Carefully lift the pack straight up to remove it.
3. Unplug the battery pack power cable from the instrument.
4. Remove the battery retainer plate by loosening the two fasteners located
on the side of the pack.
Page 31
Page 34
Section 4
5. Remove the old batteries. Install the fresh batteries into the battery pack
compartment, terminal-end first. It does not matter which way the battery is rotated when it is inserted. The battery holder will accept the batteries regardless of polarity.
6. Replace the retainer plate and tighten the fasteners by hand until they are secure. This applies the proper compression to the battery terminals. Use care not to over-tighten the fasteners.
7. Reattach the power cable to the flow meter in the bottom of the battery compartment. The cable is keyed and will only fit in the proper orientation.
8. Insert the battery pack into the battery compartment and tighten the two captive fasteners.
9. Replace the instrument cap.
Figure 12 Battery Pack Compartment
1
Eveready Energizer
EN 529 Alkaline
2
1. Loosen to remove batteries (after removing the battey pack).
2. Loosen to remove the battery pack from the meter.
3. Remove power cable.
Eveready Energizer
EN 529 Alkaline
3

4.1.4 Battery-Life Estimates

Page 32 Battery Information
When using Energizer EN-529, alkaline, 6 V dc batteries, battery life is based on:
Number of sensors installed
Recording intervals (longer intervals increase battery life)
Number of channels logged in memory
MeterInstallation.fm
Page 35
Section 4
Temperature (colder temperatures decrease battery life)
Site hydraulics (second order effect caused by excessive turbulence)
The chart below provides typical expected battery life for each logger model:
Table 4Estimated Battery Life
Model Battery Life in Days
910 60
920 90
930 365
You can access a real-time battery voltage check via InSight Select the 'Current Status' display. The chart below shows the estimated dc voltage readings when approximately 30 days of battery life remain.
Table 5Battery Voltage when a 30-Day Battery Capacity Remains
Model Battery Voltage (dc)
910 3.8
920 9.0
930 9.0

4.2 Mounting the Flow Meter

Important Note: Before using the flow meter, insert the port/vent plug into the
top hole of the unit with a
DANGER Not for use in hazardous locations where combustible gases may be present.
When mounting the flow meter, remember to:
Mount the meter so that the connectors face down. The end of the meter that contains the batteries should face up. If the connectors face up, they may corrode and allow water to seep into the instrument. See Figure
13.
When not in use, cover the connectors with their protective caps to prevent corrosion.
7
/16” wrench.
®
software.
Use the appropriate manhole support bracket/spanner bar. See Replacement
Parts and Accessories on page 47.
Page 33
MeterInstallation.fm Mounting the Flow Meter
Page 36
Section 4
Figure 13 Hanging the Meter Properly
1. Point connectors down.

4.3 Connecting Sensors to the Flow Meter

1

4.3.1 Ultrasonic Level Sensor

One 75 kHz Ultrasonic Sensor can be attached to each 'U-SONIC' receptacle. Loggers that use more than one ultrasonic sensor have ultrasonic receptacles labeled 'U-SONIC A', 'U-SONIC B', etc. Each sensor may be assigned in the software as the primary sensor or as one or more secondary sensors. The primary sensor is used for all flow calculations in InSight software. Vision software allows the use of any sensor in flow calculation.
Figure 14 and Tab l e 6 describe the pins on all of the sensor receptacles.
Figure 14 Ultrasonic Sensor Receptacle Pins
Table 6 Ultrasonic Sensor Receptacle Pin Descriptions
D
A C
B
Pin Letter Description Wire Color
A Temperature(+) Red
B Temperature (-) Black
C Ultrasonic (+) Shield
D Ultrasonic (-) Clear

4.3.2 Level/Velocity Sensor

Velocity receptacles A, B, and C accept Submerged Area/Velocity sensors, Velocity-Only probes, or any combination of the two.
®
Page 34 Connecting Sensors to the Flow Meter
Figure 15 and Tab l e 7 describe the pins on all of the velocity receptacles.
MeterInstallation.fm
Page 37
Figure 15 Velocity Receptacle Pins
A
B
H
JK
C
DE
G
F

4.3.3 Rain Gauge

An optional Hach tipping bucket rain gauge may be connected to the 'RAIN GA.' connector of Model 920 and 930 flow meters. Each tip of the bucket in the rain gauge transmits a dry contact closure to the flow meter. Each dry contact closure represents 0.01" (0.0254 cm) of rainfall. A constant dc voltage is supplied on Pin C of the rain gauge receptacle. The dry contact closure in the rain gauge returns a voltage pulse into Pin A with each tip.
Section 4
Table 7 Velocity Receptacle Pin Description
Pin Letter Description Wire Color
A +12 V dc Red
B Ground Green
C Velocity receive (ground) B/W shield
D Velocity receive (+) B/W center
E Velocity transmit (ground) Black shield
F Velocity transmit (+) Black center
G Depth (-) Black
H Depth (+) White
J
K
Not used NA
Figure 16 and Tab l e 8 describe the pins on all of the rain gauge receptacles.
Figure 16 Rain Gauge Receptacle Pins
A
F
E
B
C
D
Pin Letter*
A Tip Signal / 0.01" (2.54 cm) of rainfall
C Source Voltage
* Pins B, D, E and F are not used in the rain gauge connector.
Table 8 Rain Gauge Receptacle Pin Description
Description
Page 35
MeterInstallation.fm Connecting Sensors to the Flow Meter
Page 38
Visit http: //www.hach.com
Page 39

Section 5 Maintenance

5.1 Sensor Calibration

5.1.1 Velocity Sensor Calibration

The velocity sensor does not require calibration. The transmit frequency is fixed by a highly accurate quartz crystal-controlled frequency generator.

5.1.2 Submerged Sensor Calibration

Calibrating the submerged sensor synchronizes the meter electronics with the unique characteristics of each individual probe. In addition, the calibration compensates for a drift in the output of the sensor that may occur as the materials in the sensor age. To ensure optimum accuracy, the manufacturer recommends calibrating the submerged sensor when:
The sensor is first used.
Installing a new or different sensor on a flow meter or input receptacle.
The difference between the level reading of the flow meter and the
independent verification (measurement with a dipstick and ruler) is increasing.
Note: The data is constant if the difference between the level reading of the flow meter
and the independent verification is constant; recalibration is not required.
Note: Errors can occur with the level reading of the flow meter and the independent
verification. Errors are caused by variation in site conditions and measurement abilities. These errors may cause slight variations in the difference, therefore, not indicating a true change in the difference.

5.2 Desiccant Maintenance

5.2.1 Reusing Desiccant

The desiccant material is a silica gel indicator. When the gel is saturated with moisture, the beads turn from blue to pink. To rejuvenate the beads for reuse, remove them from the assembly. Heat the beads in an oven at 100 to 180 °C (212 to 350 °F) until the beads turn blue again. Discard and replace the beads if they do not turn blue after heating.

5.2.2 Replacing the Sensor Desiccant

Note: Before calibrating or installing the Submerged Sensor it is absolutely essential that the red plastic protective cap be removed from the atmospheric reference port of the desiccant cartridge.
The air dryer assembly is an integral part of the velocity probe cable. A small diameter tube is contained within the sensor cable to supply a reference port from the logger to the transducer in the sensor. The reference air passes through the in-line desiccant where all moisture is removed.
To remove or replace the sensor desiccant:
1. Remove the fill plug from the assembly.
2. Remove the pink desiccant beads and allow the assembly to air dry.
3. Fill the assembly with fresh (blue) desiccant.
4. Replace the fill plug.
Page 37
Maintenance.fm Maintenance
Page 40
Section 5

5.2.3 Replacing the Battery and Electronics Compartment Desiccant

A small desiccant cartridge in the battery compartment prevents moisture damage to the batteries and power circuitry. The electronics compartment desiccant cartridges should be changed whenever you service the electronics (see section 5.6 on page 39) to assure a moisture free environment inside the enclosure. Tabl e 9contains replacement part numbers.
If the normally blue desiccant beads turn pink, replace the cartridge or remove the cartridge end cap and recharge or replace the desiccant material. See Reusing Desiccant on page 37.
A small clip holds the desiccant cartridge in place. Pull the cartridge straight out of the clip to remove it.
Table 9Replacement Desiccant
Description Part Number
Model 910 Battery Compartment Desiccant Cartridge 4933
Model 920 Battery Compartment Desiccant Cartridge 4933
Model 930 Battery Compartment Desiccant Cartridge 4936
Electronics Compartment Desiccant (2 cartridges are required for Model 930)
Bulk Desiccant Refill (1.5 lb.) 3624
4933

5.3 Maintaining the O-Ring Gasket on the End Cap

Two O-ring gaskets are installed in each end cap of the flow meter. They maintain the water tight seal on the flow meter. Use care when the end caps are removed—do not soil, cut, or nick the gaskets. Replace them immediately if any physical damage is apparent. A light coating of O-ring lubricant is sufficient to maintain a water-tight seal. Table 10lists replacement O-ring gaskets.
Note: Do not use petroleum jelly to lubricate the O-rings.
Table 10Replacement O-rings
Description Part Number
Model 910 End Cap O-ring 4912
Model 920 End Cap O-ring 4863
Model 930 End Cap O-ring 4807

5.4 Cleaning the Flow Meter

Clean the outside of the logger case with warm water and mild detergent. Do not use solvents or harsh cleaners to clean the logger. Do not use high pressure washing equipment to clean the case.

5.5 Storing the Flow Meter

Remove the 6 volt alkaline batteries from the logger when storing for long periods (more than 3 months). Store the sensors in a dry area. Make sure that all desiccant materials in the probes are kept fresh (blue) at all times.
Page 38 Maintaining the O-Ring Gasket on the End Cap
Maintenance.fm
Page 41

5.6 Maintenance in the Electronics Compartment

Important Note: The electronics compartment is only serviceable by a qualified electronics technician familiar with static sensitive CMOS components. Permanent damage to electronic circuitry may result from improper handling. Always wear a wrist grounding strap when handling CMOS electronic components.
To open the electronics compartment, remove the four pins that hold the electronics compartment end cap in place. The pins are located under a protective tape on the outside surface of the logger near the end cap. The end cap pulls straight off after all pins are removed.

5.6.1 Fuses

All internal fuses are self-resetting and require no maintenance. There are no other user-serviceable components inside the electronics compartment

5.6.2 Memory Batteries

An internal battery pack that contains two, 1.5 V dc alkaline “C” batteries powers RAM memory, the real time clock, and associated circuitry. The batteries also prevent the loss of programmed settings and logged data if the main power fails or is removed.
Section 5
Model 910, 920 and 930 memory batteries are located inside the electronics compartment. The two cells will provide many years of trouble-free operation. Use only alkaline replacement batteries.
You can set a low memory battery indicator using InSight Software. See the software manual for details.
®
or Vision®
Page 39
Maintenance.fm Maintenance in the Electronics Compartment
Page 42
Visit http: //www.hach.com
Page 43

Section 6 Exploded View Drawings

Model 910
4902
4912
3667
4907
4913
2 Req’d
4933
4914
4918
3652 Apply to inside surface of 4901 (battery end)
4901
4912 2 Req’d
8644 4 Req’d
Drawings.fm Exploded View Drawings
4933
Page 41
Page 44
Section 6
Model 920
4852
4863
2 Req’d
3667
2 Req’d
4933
4857
4862
4933
3652 Apply to inner wall of 4851 case (battery end)
4851
4871
4863
2 Req’d
8644
4 Req’d
4856
Page 42 Exploded View Drawings
Drawings.fm
Page 45
Model 930
4801
4807
2 Req’d
4805
Section 6
3667 6 Req’d
4936
4804
4870
4802
4847
8644
4 Req’d
4807 2 Req’d
4933 2 Req’d
4803
Page 43
Drawings.fm Exploded View Drawings
Page 46
Visit http: //www.hach.com
Page 47

GENERAL INFORMATION

At Hach Company, customer service is an important part of every product we make.
With that in mind, we have compiled the following information for your convenience.
Page 45
GenInfo.fm GENERAL INFORMATION
Page 48
Visit http: //www.hach.com
Page 49

Replacement Parts and Accessories

Logger Mounting Accessories
Description Model Part Number
Ladder Hanger, for use with 4915 and 4873 Suspension Bracket ....................................... 920/930................. 4874
Manhole Support Bracket/Spanner 45–68 cm (18–28”) (for use with Cat. No. 4920)... 910/920/930................. 9542
Manhole Support Bracket/Spanner, 71–122 cm (28–48”) (for use with Cat. No. 4920) 910/920/930................. 9557
Manhole Support Bracket/Spanner, 107–183 cm (42–72”)
(for use with Cat. No. 4920) .......................................................................................... 910/920/930................. 9556
Suspension Cable ....................................................................................................................... 910................. 4920
Wall/Ladder Mount Suspension Bracket ..................................................................................... 920................. 4915
Wall/Ladder Mount Suspension Bracket ..................................................................................... 930................. 4873
In-Pipe Ultrasonic Sensor Mounting Accessories
Description Model Part Number
Mounting Band Set for 15–42” pipes............................................................................. 910/920/930................. 3766
Includes:
(1) Extension Band ...................................................................................................................................... 1473
(2) Extension Bands .................................................................................................................................... 1525
(3) Extension Bands ..................................................................................................................................... 1759
(1) Extension Band ....................................................................................................................................... 1318
(1) Scissors Jack Assembly ......................................................................................................................... 3719
Mounting Clip (for use with Cat. No. 3766) .......................................................................... 920/930................. 3868
Mounting Clip, permanent (screws to channel wall)............................................................. 920/930................. 3875
Mounting Ring, 6”, with integral in-pipe sensor mounting clip
and velocity sensor mounting holes. .................................................................................... 920/930................. 4021
Mounting Ring, 8”, with integral in-pipe sensor mounting clip
and velocity sensor mounting holes. .................................................................................... 920/930................. 4022
Mounting Ring, 10”, with integral in-pipe sensor mounting clip
and velocity sensor mounting holes. .................................................................................... 920/930................. 4023
Mounting Ring, 12”, with integral in-pipe sensor mounting clip
and velocity sensor mounting holes. .................................................................................... 920/930................. 4024
Submerged Area Velocity Sensor Mounting Accessories
Description Model Part Number
Insertion Tool for street-level installation of mounting rings. ......................................... 910/920/930................. 9574
Mounting Ring for 6" diameter pipe (requires Cat. No. 3263) ....................................... 910/920/930................. 1361
Mounting Ring for 8" diameter pipe (requires Cat. No. 3263) ....................................... 910/920/930................. 1362
Mounting Ring for 10" diameter pipe (requires Cat. No. 3263) ..................................... 910/920/930................. 1363
Mounting Ring for 12" diameter pipe (sensor mounts directly to band) ........................ 910/920/930................. 1364
Mounting Ring for 15" diameter pipe (sensor mounts directly to band) ........................ 910/920/930................. 1365
Mounting Ring for 18" diameter pipe (sensor mounts directly to band) ........................ 910/920/930................. 1366
Mounting Ring for 20–21" diameter pipe (sensor mounts directly to band) .................. 910/920/930................. 1353
Mounting Ring for 24" diameter pipe (sensor mounts directly to band) ........................ 910/920/930................. 1370
Sensor Mounting Clip for use with mounting bands...................................................... 910/920/930................. 3263
Mounting Plate, wall mount sensor ............................................................................... 910/920/930................. 4939
Manhole Support Bracket/Spanner Bar
Description Part Number
“A” Dimension Range
18–24 in........................................................................................................................................................ 9542
28–48 in........................................................................................................................................................ 9557
3.6–6 ft.......................................................................................................................................................... 9556
Page 47
Parts.f m Replacement Parts and Accessories
Page 50
Replacement Parts and Accessories
Mounting Band Selection Chart
Mounting Band Selection1
A
Pipe Diameter
8" (20.32 cm)0010
10" (25.4 cm)1010
12" (30.48 cm)0110
15" (38.1 cm)0210
18" (45.72 cm)0120
21" (53.34 cm)0220
24" (60.96 cm)0130
27" (68.58 cm)1011
30" (76.2 cm)1111
33" (83.2 cm)1021
36" (91.44 cm)1121
42" (1.06 m)1131
45" (1.14 m)1112
48" (1.21 m)1022
1 In addition to the band segments shown below, a complete mounting band assembly also requires one AV Sensor Mounting Clip (Cat. No. 3263), and one Scissors Jack Assembly (Cat. No. 3719).
Cat. No. 1473
6.25" (15.85 cm) long, adds 2" (5.08 cm) to
band diameter
Cat. No. 1525
9.5" (24.13 cm) long, adds 3" (7.62 cm) to
band diameter
Cat. No. 1759
19" (48.26 cm) long,
adds 6"(15.24 cm) to
band diameter.
Cat. No. 1318
50.25" (1.27 cm) long, adds 16" (40.64 cm)
to band diameter
Page 48 Replacement Parts and Accessories
Parts.fm
Page 51

Contact Information for U.S.A. and Outside Europe

Ordering Information for the U.S.A.
By Telephone:
(800) 635-4567
By Fax:
(970) 461-3915
Ordering information by E-mail:
orders@hach.com
Information Required
Hach account number (if available)
Your name and phone number
Purchase order number
Brief description or model number
By Mail:
Hach Company P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A
Billing address
Shipping address
Catalog number
Quantity
Ordering Information for Outside the U.S.A. and Europe
Hach Company maintains a worldwide network of dealers and distributors. To locate the representative nearest you, send E-mail to intl@hach.com or visit www.hach.com.
Technical Support
Technical and Customer Service Department personnel are eager to answer questions about our products and their use. In the U.S.A., call 1-800-635-4567. Outside the U.S.A. and Europe, send E-mail to intltech@hach.com.
Repair Service
Authorization must be obtained from Hach Company before sending any items for repair. Please contact the Hach Company Service Center serving your location.
Hach Company P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A.
Telephone: 1-800-635-4567 or (970) 669-3050
Fax: (970) 669-2932
Page 49
Contact Information for U.S.A. and Outside Europe
Page 52

Contact Information for Europe

For technical support, repair service, and ordering information please refer to the contact information below for your specific country.
Belgium
Lange Group Ragheno Business Center 2 Motstraat 54 B-2800 Mechelen Tel.: ++32/(0)15 42/ 35 00 Fax: ++32/(0)15 41/ 61 20 email: info@langegroup.be
Germany (TCS & Service)
Dr. Bruno Lange GmbH & CO. KG Willstätterstr. 11 D-40549 Düsseldorf Tel.: ++49/(0)211/52 88-0 Fax: ++49/(0)211/52 88-143 email: kundenservice@drlange.de www.drlange.com
Spain
NEURTEK. M.A-LANGE GROUP c/ Araba, 45. Apdo. 220 E-20800 ZARAUZ Tel.: ++34 943 894.379 Fax: ++ 34 943 130.241 email: info@drlange.com www.drlange.com
Denmark
Dr. Lange Danmark A/S Jernhoolmen 34-40 DK-2650 Hvidovre Tel.: ++45/(0)36 77/ 2911 Fax: ++45/(0)36 77/ 4911 email: drlange@drlange.dk www.drlange.de
Germany (Ordering)
Dr. Bruno Lange GmbH & CO. KG Königsweg 10 D-14163 Berlin Tel.: ++49/(0)30/80 98 60 Fax: ++49(0)30/80 98 62 70 www.drlange.de
France
Polymetron/Dr. Lange Division Lab 33 rue du ballon F-93165 Noisy le Grand Tel.: ++33/(0)1/ 48 15 68 70 Fax: ++33/(0)1/ 48 15 68 79 email: info@drlange.fr www.drlange.fr
Poland
Dr. Lange Sp. z o.o. ul. Opolska 143a PL-52-013 Wroclaw Tel.: ++48/(0)71/ 3 42 10 -81/-83 Fax: ++48/(0)71/ 3 42 10 79 email: mail@drlange.pl www.drlange.de
Austria
Dr. Bruno Lange Ges.m.b.H. Industriestraße 12 A - 3200 Obergrafendorf Tel.: ++43/(0)2747 - 7412 Fax: ++43/(0)2747 - 4218 email: lange@via.at www.drlange.de
Italy
Dr. Bruno Lange s.r.I. via Riccione, 14 IT-20156 Milano Tel.: ++39/(0)2 39 23 14 1 Fax: ++33/(0)2 39 23 14 39 email: info@drlange.it www.drlange.it
Great Britain
Dr. Lange (UK) Ltd. Lennox Road GB-Basingstoke Hampshire RG22 4AP Tel.: ++44/(0)12 56/ 33 34 03 Fax: ++44/(0)12 56/ 33 07 24 email: info@drlange.co.uk www.drlange.co.uk
Sweden
Dr. Lange AB LAB Products Box 9008 S-40091 Götebord Tel.: ++46-31-28 50 50 Fax: ++46-31-28 50 39 email: info@lange.se www,lange.se
The Netherlands
Dr. Lange Benelux Postbus 6299 NL-4000 HG Tiel Tel.: ++31/(0)344/ 63 11 30 Fax: ++31/(0)344/ 63 11 50 email: info@langegroup.nl www.langegroup.nl
Sweden
Dr. Lange AB PROCESS Products Box 410 S-13525 Tyresö Tel.: ++46-87-98 05 00 Fax: ++46-87-42 17 24 email: info@lange.se www.lange.se
Switzerland
Dr. Bruno Lange AG Juchstrasse 1 CH-8604 Hegnau Tel.: ++41/(0)1/ 9 45 66 10 Fax: ++41/(0)1/ 9 45 66 76 email: info@drlange.ch www.drlange.ch
Page 50 Contact Information for Europe
Page 53

Warranty

Hach Company warrants this product to the original purchaser against any defects that are due to faulty material or workmanship for a period of one year from date of shipment.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price, excluding original shipping and handling charges. Any product repaired or replaced under this warranty will be warranted only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned without authorization from Hach Company.
Limitations
This warranty does not cover:
Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or undeclared), terrorism, civil strife or acts of any governmental jurisdiction
Damage caused by misuse, neglect, accident or improper application or installation
Damage caused by any repair or attempted repair not authorized by Hach Company
Any product not used in accordance with the instructions furnished by Hach Company
Freight charges to return merchandise to Hach Company
Freight charges on expedited or express shipment of warranted parts or product
Travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty or negligence.
Page 51
Warranty
Page 54
Loading...