Data Sheet
SENSOR MOUNTING
Ultrasonic Sensor Model PZ32T
25 ft (7.6 m) RG62AU
COAXIAL CABLE
50 ft (15 m) OPTIONAL
3/4 " NPT
~5"
(127 mm)
OVERALL
3.875"
98.4mm
1.75"
44.5 mm
TEFLON
ISOLATION
COUPLING
(SUPPLIED)
3/4 " NPT
SIDE
VIEW
END
VIEW
General Specifications
Maximum Range:
Minimum Range (Deadband):
Operating Frequency:
Beam Angle:
Operating Temperature:
Maximum Operating Pressure:
Sensor Face:
Sensor Housing:
Sensor Mounting:
Maximum Cable Length:
Optional Hazardous Rating:
32 ft. (10 m)
12" (305 mm)
42 KHz
8°
-40° to 150° F (-40° to 65° C)
20 psi (1.35 bar)
Teflon
PVC
3/4" NPT
500 ft (150 m)
CSA rated Intrinsically Safe Class I,
Groups C,D, Class II, Groups E,F,G
with optional Intrinsic Safety Barrier.
Note: Max Range reduced to 25 ft.
(7.8 m) with ISB option.
Notes:
1. Shield from direct sunlight (temperature probe in sensor)
2. Use the "Isolation Coupling" supplied for mounting.
3. Use with Option ISB Intrinsic Safety Barrier for installation in
hazardous locations.
4. Run Sensor cable in grounded metal conduit with no other
conductors.
5. Extend Sensor cable up to 500 ft (152 m) with RG62AU coaxial only.
6. Refer to the Instruction Manual for complete installation and wiring
instructions.
1 FT (300 mm)
HORIZONTAL FROM
SIDEWALL FOR EVERY
10 FT (3 m) VERTICAL
NOTE
:
2 FT (600 mm) FOR EVERY
10 FT (6 m) DEPTH IN
APPLICATIONS WITH
ROUGH SIDEWALLS,
LADDERS, REINFORCING
RINGS ETC.
Recommended Sensor Mounting
PREFERRED SENSOR
LOCATION
SENSOR FACE MUST
BE HORIZONTAL ±1°
MAX. LIQUID LEVEL
6" SCHEDULE 40
PIPE IF EXTENSION
REQUIRED