GE Healthcare Vivo 40 Brochure

Vivo 40 with Target Volume
To ensure a successful transition from hospital to home, healthcare providers seek respiratory solutions that can be trusted, make life easier and enjoy dedicated personal support. Moreover, serviceability and cost of ownership must meet increasingly strict demands.
A device you can trust
The Vivo ventilators are well trusted by patients and care givers as proven by more than 30,000 units sold worldwide. The Vivo 40* provides both invasive and non-invasive support in PSV, PCV and CPAP mode. The adult and pediatric modes allow Vivo 40 to be used by a wide range of patients.
Thanks to eSync*, the highly responsive trigger technology, the Vivo 40 synchronizes smoothly with the patient’s breathing eort, leading to lower the work of breathing and higher the level of comfort.
The Target Volume feature aims for a stable tidal volume by auto­matically adapting the inspiratory pressure to the lung mechanics of the patient. Target Volume combines the comfort and technical advantages of pressure ventilation with the ease of use of volume ventilation.
To provide security to patients and caregivers, the Vivo 40 contains a comprehensive set of alarms and an internal battery for independent use up to three hours. An external DC connection is available as an additional power source.
“Vivo 40 is easy to use, secure and effective”.
Dr Fernando Rua Hospital de Santo António, Porto (Portugal)
• Enhanced trigger
• Enhanced display and look
• Improved humidification
“Vivo 40 is a pleasant device, easy to operate. The interface is practical since it has no submenus – you directly reach the set-up you want. Moreover, the screen is quite large enabling you to see the settings and the patient parameters at the same time.”
Dr Christine Cheval, Centre Hospitalier, Hyeres (France)
Making life easier for patients and caregivers
The user interface and menu structure of the Vivo 40 are intuitive and consistent throughout the Vivo range. This makes the Vivo 40 easy to learn and easy to teach. It facilitates situations where multiple sta members work with the device. The optional humidier is completely regulated through the ventilator and provides additional patient comfort. If data analysis or reports are required, data can easily be downloaded to a ash card.
For paedriatic (>10 kg) and adult use.
Target Volume stabilizes tidal volume delivery
Appropriate for invasive and non-invasive use
eSync feature for synchronized breath delivery
Designed for use at Home or in the Hospital
Large, easy-to-read display with simple menu structure
Designed for a wide range of masks
Oers 3 Ventilation modes, 2 Device modes,
2 Patient modes (adult/pediatric)
Large screen and intuitive user interface
Designed for hospital and home environment
Low noise <30dB(A)
Extensive alarm package
Internal 3 hour battery
Integrated Heated Humidier
Internal data/calendar memory
Stable device with small footprint
Large, easy-to-use buttons with backlight
Worldwide mains power compatibility
Flexible DC power options
Integrated humidier Enhanced trigger
function
Technical Specications
Settings/Performance
Ventilation Modes • PSV (Pressure Support Ventilation)
• PCV (Pressure Control Ventilation)
• PSV +Target Volume
• PCV +Target Volume
• CPAP (Continuous Positive Airway Pressure)
Patient Modes • Adult
• Paediatric (>10 kg)
Device Modes • Clinical
• Home
IPAP/EPAP 4 to 40 cmH2O / 2 to 20 cmH2O or IPAP -2 cmH2O
CPAP 4 to 20 cmH2O
Breath Rate 4 to 40 BPM
Inspiratory Time 0.3 to 5 s
Rise Time 1 to 9
Inspiratory Trigger (eSync) 1 to 9, O
Expiratory Trigger 1 to 9
Target Volume 200-1500 ml
Expiratory Trigger 1 to 9
Ramp function setting 10 – 60 min, O (PCV, PSV and CPAP mode)
Humidication setting 1 to 9 (10 to 30 mgH2O/L)
Audible alarm setting 1 to 9
Maximum ow > 200 litres/min
Minimal Inspiratory Time O, 0.3 – 3.0 s
Maximal Inspiratory Time 0.3 – 3.0 s, O
Noise level (at 10 cmH2O constant pressure) Less than 30 dB(A)
Indications
Pressure; bar graph and digital data 0 to 40 cmH2O
Calculated tidal volume accuracy ± 20%
Leakage calculation accuracy ± 20%
Total Breath Rate BPM (Breaths Per Minute)
I/E Ratio 1 :10 to 10 :1
Alarms
Audible Alarm Level 45 to 85 dB(A)
Low pressure alarm, High pressure alarm, Low volume alarm, Low breath rate alarm, High breath rate alarm, Low leakage alarm, High leakage alarm, Low External DC War­ning, Low internal battery warning, Low power alarm, Internal function failure alarms.
* Vivo by Breas and eSync are trademarks by General Electric Company.
© 2012 General Electric Company – All rights reserved. General Electric Company reserves the right to make changes in specications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE representative for the most current information. GE, the GE Monogram and imagination at work are trademarks of General Electric Company.
Breas Medical AB · Företagsvägen 1 · SE-435 33 Mölnlycke · Sweden Phone +46 31 86 88 00 · Fax +46 31 86 88 10 · breasswe@ge.com · www.breas.com · www.gehealthcare.com
Power Supplies
Mains supply 100 to 240 V AC
Internal battery / External battery Capacity 3.8 Ah / 12/24 V DC
Operating Conditions
Recommended mask leakage 20 to 40 litres/min at 10 cmH2O
Minimum mask leakage 12 litres/min at 4 cmH2O
Dimensions
W x H x D 185 x 240 x 227 mm
Weight 4.0 kg
Air outlet 22 mm male conical standard connector
Breas Medical AB/EN/October 2012/DOC1222972, rev1