Vivo 30
To ensure a successful transition from hospital to home, healthcare providers seek respiratory solutions that can be trusted, make life easier and enjoy dedicated personal support. Moreover, serviceability and cost of ownership must meet increasingly strict demands.
A device you can trust
The Vivo* ventilators are well trusted by patients and care givers as proven by more than 30,000 units sold worldwide. The Vivo 30 provides both non-invasive support in PSV, PCV and CPAP mode.
Thanks to eSync*, the highly responsive trigger technology, the Vivo 30 synchronizes smoothly with the patient’s breathing eort, leading to lower the work of breathing and higher the level of comfort.
To ensure excellent control of the treatment, there are ample ways to adapt the settings to the need of the individual patient. Apart from IPAP and EPAP, also rise time, trigger sensitivity and ramp timeoptional minimal and maximal Inspiratory Time can be set. A comprehensive alarm package provides security for patients and care givers.
Making life easier for patients and caregivers
The user interface and menu structure of the Vivo 30 are intuitive and consistent throughout the Vivo range. This makes the Vivo 30 easy to learn and easy to teach. It facilitates situations where multiple sta members work with the device. The (optional) integrated humidier is completely regulated through the ventilator and provides additional patient comfort. If data analysis or reports are required, data can easily be downloaded to a ash card. An external DC connection is available for additional power source.
Designed for the home environment
With its attractive Scandinavian design, low noise level and small footprint the Vivo 30 is highly suitable for the home environment. The Vivo 30 will allow for a secure and comfortable transition from hospital to home.
• Enhanced trigger
• Enhanced display and look
• Improved humidifier
Large screen and intuitive user interface
Integrated humidier Enhanced trigger
function
Minimal and Maximal Inspiratory Time
eSync*, synchronized breath delivery
Designed for use at Home or in Hospital
Large easy-to-read display with simple menu structure
and backlight
Large easy-to-use buttons with backlight
Built-in alarm clock
Designed for a wide range of masks
Low noise < 30dB(A)
3 Ventilation modes
2 Device modes
Comprehensive alarm package
Integrated Heated Humidier
Internal data/calendar memory
Stable unit with small footprint
Easy to carry
Worldwide mains power compatibility
Flexible DC power options
Technical Specications
Settings/Performance
Ventilation Modes • PSV (Pressure Support Ventilation)
• PCV (Pressure Control Ventilation)
• CPAP (Continuous Positive Airway Pressure)
Device Modes • Clinical
• Home
IPAP/EPAP 4 to 30 cmH2O / 2 to 20 cmH2O or IPAP -2 cmH2O
CPAP 4 to 20 cmH2O
Breath Rate 4 to 40 BPM
Inspiratory Time 0.3 to 5 s
Rise Time 1 to 9
Inspiratory Trigger (eSync) 1 to 9, O
Expiratory Trigger 1 to 9
Ramp Function 10 – 60 min
Humidication 1 to 9, (10 to 30 mgH2O/L)
Audible alarm 1 to 9
Maximum ow > 200 litres/min
Minimal Inspiratory Time O, 0.3 – 3.0 s
Maximal Inspiratory Time 0.3 – 3.0 s, O
Noise level (at 10 cmH2O constant pressure) Less than 30 dB(A)
Indications
Pressure; bar graph and digital data 0 to 30 cmH2O
Calculated tidal volume accuracy ± 20%
Leakage calculation accuracy ± 20%
Total Breath Rate BPM (Breaths Per Minute)
I/E Ratio 1 :10 to 10 :1
Alarms
Audible Alarm Level 45 to 85 dB(A)
Low/High pressure alarm, Low volume alarm, Low/High leakage alarm, Low external DC Warning, Low power alarm, Internal function failure alarms.
* Vivo by Breas and eSync are trademarks by General Electric Company.
© 2012 General Electric Company – All rights reserved. General Electric Company reserves the right to make changes in specications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE representative for the most current information. GE, the GE Monogram and imagination at work are trademarks of General Electric Company.
Power Supplies
Mains supply 100 to 240 V AC
External battery 12/24 V DC
Operating Conditions
Recommended mask leakage 20 to 40 litres/min at 10 cmH2O
Minimum mask leakage 12 litres/min at 4 cmH2O
Dimensions
W x H x D 185 x 230 x 227 mm
Weight 3.1 kg
Air outlet 22 mm male conical standard connector
Breas Medical AB · Företagsvägen 1 · SE-435 33 Mölnlycke · Sweden Phone +46 31 86 88 00 · Fax +46 31 86 88 10 · breasswe@ge.com · www.breas.com · www.gehealthcare.com
Breas Medical AB/EN/October 2012/DOC1222952, rev1