GE Healthcare HA 20 Vivo Brochure

HA 20 – Heated Humidier
The HA 20* is an integrated heated humidier, especially designed for use with the iSleep* and Vivo* 30/40 devices. To meet increasing demands of customers regarding comfort and security, the HA 20 has now been further enhanced.
The HA 20 is truly integrated with the iSleep and Vivo 30/40 devices, both in design and function. When connected to an iSleep or Vivo 30/40 device, all HA 20 operations are set and monitored through the menu of the device. The HA 20 consists of a heating unit and a separate water chamber. The water chamber is easy to connect and to remove through the quick-release handle. No separate power cords are required because power is provided through the iSleep or Vivo 30/40 device.
What’s new?
The current enhancements help further increase handling comfort and reducing the risk of water entering the device.
1. A smart water trap has been inserted into the lid to reduce the risk of water entering the device
2. A clear indication of the maximum water level has been added to the container
3. A new lling hole is added to facilitate lling the HA 20.
4. A simple leaet with instructions for use is provided in the HA 20 (remove before use).
Now further enhanced!
Please remember to always remove the HA 20 from the device and use the lling hole to ll it. Do not surpass the maximum indicated water level and remove / empty the HA 20 before transporting the device.
Part numbers
The part numbers for the HA 20 have remained unchanged:
2220XX Complete HA 20 White (where XX is the country code)
2230XX Complete HA 20 Silver (where XX is the country code)
004861 Water chamber
Heated humidication
Accessory to iSleep and Vivo 30/40
Activated and controlled through the iSleep and Vivo
30/40 device
The HA 20 is operated through the iSleep/Vivo 30/40 device
No separate power cable required
Smart water trap
Convenient lling hole and clear water indication
Easy to clean
Technical Specications
Power Supplies
Power supply 10-30 VDC, max 40 W.
Environmental Conditions
Operating temperature 5 to 40ºC (41 to 104ºF)
Storage and transport temperature -20 to +60ºC (-4 to 140ºF)
Ambient pressure range 700 to 1060 mbar
Humidity 10% to 95%, non-condensing
Dimensions
W x H x D 150 x 140 x 100 mm
Weight 425 g
Air outlet 22 mm conical standard connector
Water chamber volume 300 ml
Performance
Gas leakage at max. operating pressure < 5 ml/min
Warm up time for delivered gas temperature Less than 60 minutes to reach the set temperature from starting temperature of 23+/-2ºC
Breas Medical AB reserves the right to make changes to the contents of this Product Specication without any prior notication.
* HA20 and iSleep and Vivo are trademarks of General Electric Company
© 2013 General Electric Company – All rights reserved. General Electric Company reserves the right to make changes in specications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE represen­tative for the most current information. GE, the GE Monogram and imagination at work are trademarks of General Electric Company.
Breas Medical AB · Företagsvägen 1 · SE-435 33 Mölnlycke · Sweden Phone +46 31 86 88 00 · Fax +46 31 86 88 10 · breasswe@ge.com · www.breas.com
Date 20130207 DOC1280974