GE Healthcare
Engström Carestation
Breathing life into critical care
Features
• Simplified user interface
• Paramagnetic O2 sensing
• Non-Invasive ventilation (Optional)
• Secure access to central stations
• Sophisticated power management control
with battery backup
• Auxiliary pressure sensor
• Airway Resistance Compensation
™
Integrated Ventilation and Monitoring
• Advanced ventilation
• INview
• Plug and play modules
• Patient Spirometry
• Gas monitoring with metabolics and energy expenditure
• Optional use of proximal Neo Flow Sensor
with Neonatal ventilation
Exceptional Design
• Adaptable to your environment
• Flexible and moveable display
• Transferable module bay
• Quick-release expiratory valve
• Multiple trolley configurations
Aerogen Aeroneb® Pro
• Built-in advanced nebulization system
• Operated in-line or independently for infants
through adults
™
Suite: SpiroDynamics™ and FRC INview
Physical Specifi cations
Modes of Ventilation
Dimensions
Height: 44.5 cm/17.5 in (Display down)
67.5 cm/26.6 in (Display up)
Height
including cart: 122 cm/48 in (Display down)
145 cm/57.1 in (Display up)
Width: 38 cm/15 in
Depth: 36 cm/14 in
Weight: 31 kg/68.3 lb (not including cart);
76 kg/167.6 lb (including cart)
Display motion
Vertical tilt: 160° in raised position
60° in lowered position
Height
adjustment: 23 cm/9.1 in
Volume Controlled (VCV)
Pressure Controlled (PCV)
Pressure Controlled, Volume Guaranteed (PCV-VG)
Synchronized Intermittent Mandatory Ventilation,
Volume Controlled (SIMV-VC)
Synchronized Intermittent Mandatory Ventilation,
Pressure Controlled (SIMV-PC)
Synchronized Intermittent Mandatory Ventilation, Pressure
Controlled, Volume Guaranteed (SIMV-PCVG) (optional)
BiLevel Airway Pressure Ventilation (APRV capable)
BiLevel with Volume Guaranteed (BiLevel-VG) (optional)
Non-Invasive Ventilation (NIV) (optional); nCPAP available
with Neonatal option
Constant Positive Airway Pressure/Pressure Support
Ventilation (CPAP/PSV)
Apnea backup available in SIMV-VC, SIMV-PC, BiLevel,
SIMV-PCVG, BiLevel-VG, CPAP/PSV and VG-PS
(institutionally selectable defaults)
Key:
Available only when Adult patient type is selected
Available only when Pediatric patient type is selected
Available only when Neonatal patient type is selected
Note: Neonatal software is an optional feature. If not specified with the icon,
features listed in this specifications sheet apply to Adult/Pediatric units and
patient population selections.
Note: Ranges and Settings without an icon pertain to both Adult
and Pediatric patient types.
Volume Guarantee Pressure Support (VG-PS) available
with Neonatal option
Control and Ranges
Maximum
peak fl ow: 200 L/min
Flow: 0.2 to 30 L/min (0.004 to 0.5 L/sec)
2 to 90 L/min (0.04 to 1.5 L/sec)
2 to 160 L/min (0.04 to 2.6 L/sec)
Incremental
settings: 0.2 to 5 L/min (increments of 0.1 L/min)
5 to 30 L/min (increments of 0.5 L/min)
2 to 40 L/min (increments of 1 L/min)
40 to 90 L/min (increments of 5 L/min)
40 to 160 L/min (increments of 5 L/min)
FiO2: 21 to 100% O
2
Rate: 3 to 150 breaths per minute for VCV, PCV,
PCV-VG and BiLevel (increments of 1 breath
per minute)
3 to 120 breaths per minute for VCV, PCV,
PCV-VG and BiLevel (increments of 1 breath
per minute)
2 to 60 breaths per minute for SIMV-VC,
SIMV-PC, SIMV-PCVG, BiLevel-VG
(increments of 1 breath per minute)
1 to 60 breaths per minute for SIMV-VC,
SIMV-PC, SIMV-PCVG and BiLevel-VG
(increments of 1 breath per minute)
Control and Ranges (continued)
Minimum rate: 2 to 60 breaths per minute for VG-PS
(increments of 1 breath per minute)
0 to 60 breaths per minute for CPAP/PSV
and 0 to 40 breaths per minute for NIV
(increments of 1 breath per minute)
Inspiratory/
expiratory ratio: 1:199 to 40:1 in BiLevel
1:9 to 4:1 (ventilator setting)
1:79 to 60:1 in BiLevel
Tidal volume
range: 2 to 350 mL
20 to 2000 mL
Incremental
settings: 2 to 50 mL (increments of 0.5 mL)
50 to 100 mL (increments of 1 mL)
100 to 350 mL (increments of 5 mL)
For VCV, PCV-VG, SIMV-VC, SIMV-PCVG,
VG-PS and BiLevel-VG
20 to 50 ml (increments of 0.5 ml)
50 to 100 ml (increments of 1 ml)
100 to 300 ml (increments of 5 ml)
300 to 1000 mL (increments of 25 mL)
1000 to 2000 mL (increments of 50 mL)
For VCV, PCV-VG, SIMV-VC, SIMV-PCVG
and BiLevel-VG
Patient weight: 0.25 to 1 kg (increments of 0.01 kg)
1 to 7 kg (increments of 0.1 kg)
0.5 to 2 lb (increments of 0.02 lb)
2 to 15 lb (increments of 0.2 lb)
5 to 15 kg (increments of 0.5 kg)
15 to 100 kg (increments of 1 kg)
100 to 200 kg (increments of 2 kg)
10 to 34 lb (increments of 1 lb)
34 to 220 lb (increments of 2 lb)
220 to 440 lb (increments of 5 lb)
Inspiratory
pressure
(P
) range: 1 to 98 cm H2O (increments of 1 cm H2O)
insp
P
: 1 to 98 cm H2O (increments of 1 cm H2O)
high
: Off, 1 to 50 cm H2O (increments of 1 cm H2O)
P
low
Pressure limit
(P
) range: 7 to 100 cm H2O for VCV and SIMV-VC
limit
(increments of 1 cm H2O)
Max. inspiratory
pressure
(P
) limit: 7 to 100 cm H2O (increments of 1 cm H2O)
max
9-100 cm H2O (increments of 1 cm H2O)
in NIV and nCPAP
PEEP: Off, 1 to 50 cm H2O (increments of 1 cm H2O)
2-15 cm H2O (increments of 1 cm H2O)
in nCPAP
2-20 cm H2O (increments of 1 cm H2O) in NIV
Inspiratory time: 0.1 to 10 sec
0.1 to 1 sec (increments of 0.01)
1 to 4 sec (increments of 0.1)
4 to 10 sec (increments of 0.25)
0.25 to 15 sec
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.10)
4 to 15 sec (increments of 0.25)
T
: 0.1 to 10 sec
high
0.1 to 1 sec (increments of 0.01)
1 to 4 sec (increments of 0.1)
4 to 10 sec (increments of 0.25)
0.25 to 15 sec
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.1)
4 to 15 sec (increments of 0.25)
T
: 0.25 to 18 sec
low
0.25 to 1 sec (increments of 0.01)
1 to 4 sec (increments of 0.1)
4 to 18 sec (increments of 0.25)
0.25 to 18 sec
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.1)
4 to 18 sec (increments of 0.25)
T
: 0.1 to 0.8 sec (increments of 0.01)
supp
0.25 to 4 sec for NIV
0.25 to 1 sec (increments of 0.05)
1 to 4 sec (increments of 0.1)
Expiratory time: 0.25 to 59.75 sec
0.25 to 29.9 sec Invasive vent modes
0.5 to 59.75 sec for NIV
Rise time: 0 to 500 ms of inspiratory period for either
fl ow or pressure depending on the mode
selected. Active in VCV, PCV, PCV-VG, SIMV-VC,
SIMV-PC, SIMV-PCVG, BiLevel-VG, NIV and
BiLevel (increments of 50 ms)
PSV rise time: 0 to 500 ms of inspiratory period for pressure
supported breaths only. Active in SIMV-VC,
SIMV-PC, SIMV-PCVG, BiLevel, BiLevel-VG,
CPAP/PSV and VG-PS (increments of 50 ms)
Trigger window: 0 to 80% of expiration time (increments of 5%)
Flow trigger: 0.2 to 1 L/min (increments of 0.05 L/min)
1 to 3 L/min (increments of 0.1 L/min)
3 to 9 L/min (increments of 0.5 L/min)
Pressure trigger: –10 to –3 cm H2O (increments of 0.5 cm H2O)
–3 to –0.25 cm H2O (increments of 0.25 cm H2O)
Bias fl ow rate: 2 to 15 L/min (increments of 0.5 L/min)
for nCPAP
2 to 10 L/min (increments of 0.5 L/min)
8 to 20 L/min for NIV (increments of 0.5 L/min)