2030434-006 Rev. D EasyOne-CS Field Service Manual Page 1
Contents
1 Manual Purpose .............................................................................................................................. 2
2 Intended Audience .......................................................................................................................... 3
3 Introduction ................................................................................................................................... 3
4 Operation Principle ......................................................................................................................... 3
5 Technical Details ............................................................................................................................ 4
5.1 EasyOne-CS Spirometer Handheld Unit ........................................................................................... 4
5.2 Screen-Connector ........................................................................................................................... 5
5.3 Block Diagram ................................................................................................................................ 6
5.4 Printed Circuit board....................................................................................................................... 6
6 Specifications ................................................................................................................................. 7
6.1 EasyOne-CS Model 2001 Spirometer .............................................................................................. 7
6.2 EasyOne-CS Model 2010 Screen-Connector ................................................................................... 8
7 Replacement Parts List .................................................................................................................... 8
7.1 Manuals ......................................................................................................................................... 8
7.2 Accessories .................................................................................................................................... 9
7.3 Spare Parts .................................................................................................................................. 10
7.4 Exchange Unit .............................................................................................................................. 12
8 Checkout Procedure ...................................................................................................................... 12
9 Hidden Keys ................................................................................................................................. 13
10 Device Error Codes ........................................................................................................................ 14
11 EasyOne-CS Device Software Update .............................................................................................. 15
11.1 Introduction ................................................................................................................................. 15
11.2 Validated Upgrade Steps .............................................................................................................. 15
11.3 Preparation .................................................................................................................................. 15
11.4 Unpacking of Upgrade Package .................................................................................................... 15
11.5 Installation of EasyFlash PC Software ............................................................................................ 15
11.6 EasyOne-CS Device Software Update Procedure ............................................................................ 17
12 EasyWare PC Software Installation ................................................................................................. 18
13 EasyOne-CS Lithium Battery Replacement ...................................................................................... 20
13.1 Background .................................................................................................................................. 20
13.2 Environment ................................................................................................................................. 21
13.3 Preparation (save database) ......................................................................................................... 21
13.4 Battery exchange procedure ......................................................................................................... 22
13.5 Restart EasyOne-CS ...................................................................................................................... 23
13.6 Restore database ......................................................................................................................... 23
14 EasyOne-CS board with sensor and housing parts Replacement ....................................................... 23
14.1 Background .................................................................................................................................. 23
15 Appendix ...................................................................................................................................... 25
15.1 Schematic Diagrams ..................................................................................................................... 25

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 2
2030434-006 Rev. D
2030434-006 Rev. C
Chapter 7.1 with new content
Chapter 7.2 with V2.9.5
Chapter 9 with additional comment for 1-2-5
Chapter 11.2 with Windows 2000, XP or Vista
Chapter 11.3 update: added resource update
Chapter 8 updated and renamed as ‘Checkout Procedure’
Chapter 12 EasyWare PC Software Installation - new
This page with additional addresses
August 31, 2009
Octobre 27, 2009
Header info modified to “EasyOne-CS Field Service Manual”
Chapter 7.2 Accessories – CD pictures adapted to V2.9.5
Chapter 11 modified to document update FW v4.1 to v4.3
Document History
Europe:
ndd Medizintechnik AG
Technoparkstrasse 1, CH-8005 Zurich, Switzerland
Tel.: +41 44 445 29 70
Fax: +41 44 445 25 31
Web: www.ndd.ch
USA:
ndd Medical Technologies
Two Dundee Park
Andover, MA 01810
Tel: +1 877 904 0090
Fax: +1 978 470 0924
Web: www.nddmed.com
Asia - Please contact GE Healthcare:
Asia Headquarters
GE China Co., Ltd.
No1 Huatuo Road,
Zhangjiang Hi-Tech Park Pudong,
Shanghai, P.R.China 201203
Tel: +86 21 38777888
Fax: +86 21 38777402
Distributed by: GE Medical Systems Information Technologies, Inc.
8200 West Tower Avenue, Milwaukee, WI 53223 USA
Tel.: +1 414 355 5000
+800 558 5120 (USA only)
Fax: +1 414 355 3790
1 Manual Purpose
This manual supplies technical information for service representative and technical personnel so they
can maintain the equipment to the assembly level. Use it as a guide for maintenance and electrical
repairs considered field repairable. Where necessary the manual identifies additional sources of
relevant information and or technical assistance.
See the operator manual for the instructions necessary to operate the equipment safely in accordance
with its function and intended use.

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 3
2 Intended Audience
This manual is intended for the person who uses, maintains, or troubleshoots this equipment.
3 Introduction
The EasyOne-CS Spirometer uses digital ultrasonic flow measurement technology for fast, accurate,
reliable operation. The EasyOne-CS was designed for full range testing in primary care, specialty
physician, industrial and hospital settings. It offers automated comparison to predicted values, preand post Broncho Dilator comparisons, interpretation of results and test quality control that
automatically assesses patient effort offering helpful suggestions to obtain optimum results. Optional
easy to read and interpret color printout with graphic display for instant test review is available. The
single-patient-use spirette minimizes risk of cross-infection.
4 Operation Principle
This EasyOne-CS flow sensor is equipped with two ultrasonic transducers, mounted at different sides of
the flow channel. In order to determine gas flow (and molar mass), ultrasonic pulses are transmitted in
up- and downstream direction. Precisely measured transit-times are the basis for determining flow and
molar mass. The following figure shows a diagram of the flow sensor including its internal sound
transmission path.
Figure 1: Schematic of EasyOne-CS flow sensor

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 4
5 Technical Details
5.1 EasyOne-CS Spirometer Handheld Unit
The following pictures show the top, rear and bottom views of the EasyOne-CS.
Figure 2: Top view of EasyOne-CS
Figure 3: Rear view with empty battery compartment
The two screws on the back side are used to open EasyOne-CS. With the exception of the lithium
battery there are no serviceable parts inside the unit. Therefore do not open the unit!
Figure 4: Bottom view of EasyOne-CS
The bottom view shows the name plate, the serial number and the EasyOne-CS-to-Screen-Connector.

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 5
5.2 Screen-Connector
The following pictures show the top and bottom views of the EasyOne-CS Screen-Connector.
Figure 5: Screen-Connector top view
Figure 6: Screen-Connector bottom view
The Screen-Connector uses a standard USB connection to the PC.

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 6
Microcontroller
M16C62
Real Time
Clock (RTC)
Serial Data
Flash
Graphic LC
Display
Flow Sensor
ASIC
Power
Management
Buzzer
Keypad
ON/OFF
EZ1-BlockDiagram.sdr
Screen
Connector
PC (USB)
5.3 Block Diagram
The following picture shows the functional block diagram of the EasyOne-CS Spirometer.
5.4 Printed Circuit board
All electrical components of EasyOne-CS are mounted on a single PCB.
With the exception of the lithium battery no components can be serviced.
The following two pictures show a top and bottom view of the PCB.
Figure 7: Bottom view of EasyOne-CS PCB
The lithium battery located on the top right of the PCB has an approximate life time of 5 to 10 years. In
early revisions of the hardware the lithium battery is soldered on the board.

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 7
83 x 158 x 43 mm (3.3 x 6.2 x 1.7 inches)
242 grams (8,6 ounces), with batteries
Volume: ±2% or 0.050 l
Flow: ±2% or 0.020 l/s, (except PEF)
PEF: ±5% or 0.200 l/s
MVV: ±5% or 5 l/min.
Volume: ±12 l
Flow: ±16 l/s
For up to 700 tests, stored in EasyOne CS
FVC, FVL, Slow VC, MVV, Pre/Post
FVC, MVV, FEV6, FEV1, FEV1/FVC, FEV1/VC , FEV1/FEV6, MEF25 (FEF75),
MEF50 (FEF50), MEF75 (FEF25), MEF25%-75%, PEF, FET, FIVC, PIF, IVC, ERV,
IRV, pre-post % variation, Lung Age
Disposable spirette respiratory tube
Ultrasound transit-time measurement
NHANES-III, Knudson_83, Knudson_76, Crapo, Morris, ERS (ECCS/EGKS),
Forche (Austria), Sapaldia (Switzerland), Roca (Spain), Berglund, Gulsvik,
Hedenström, Gore, Cherniak (only for MVV)
Dockery (Harvard), Hsu, Zapletal, Polgar, Hibbert
Power supply
Power consumption:
2 alkaline batteries, Type AA, 1.5V
Typical 0.6 W
Temperature: -20 to 50 °C
Relative humidity: 5% to95%
Ambient pressure: 500 to 1060 hPa
Figure 8: Top view of EasyOne-CS PCB (without EMI shielding)
6 Specifications
6.1 EasyOne-CS Model 2001 Spirometer

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 8
Temperature: 0 to 40 °C
Relative humidity: 0% to 95%
Ambient pressure: 500 to 1060 hPa
Certifications and standards:
CE Declaration of Conformity
C CSA US approval, CAN/CSA-C22.2 No. 601.1-M90, S1-94, CSA 601.1
Amendment 2:1998, UL Std No. 2601.1
FDA 510 (k) approval, K993921
EasyOne-CS meets or exceeds the published targets of the European
Respiratory Society (ERS), the American Thoracic Society (ATS) and the
National Lung Health Education Program (NLHEP).
Instrument classification:
Type BF applied part (when used with a PC, protection class 2
according to IEC 60601-1)
Instrument not suitable for use in flammable anesthetic gases in mixtures
with O2 or NO.
64 x 44 x 25 mm (2.5 x 1.7 x 1.0 inches)
Connects the EasyOne-CS spirometer to a PC
Standard USB type A connector for connection to PC
Pins: 1 = VBus, 2 = D–, 3 = D+, 4 = GND
Temperature: -20 to 50 °C
Relative humidity: 5% to 95%
Ambient pressure: 500 to 1060 hPa
Temperature: 0 to 40 °C
Relative humidity: 0% to 95%
Ambient pressure: 500 to 1060 hPa
The PC must comply with corresponding IEC standard (ex. IEC 60950-1. The
user is responsible that requirements of IEC 60601-1-1 for safety of medical
electrical systems are met.
Part Number
Short Description
2030434-006 MNL SVCE EASYONE-CS
Field Service Manual (this document)
Operator’s Manual English
Operator’s Manual Italian
Operator’s Manual Spanish
Operator’s Manual Portuguese
Operator’s Manual Chinese
Operator’s Manual English-US
6.2 EasyOne-CS Model 2010 Screen-Connector
7 Replacement Parts List
The following parts are available for replacement:
7.1 Manuals

2030434-006 Rev. D EasyOne-CS Field Service Manual Page 9
Part Number
Short Description
2030431-003
CABLE EASYONE-CS - PC
Screen-Connector, connects EasyOne-CS with
the PC via USB connection
2030431-005
CD EASYWARE V2.9.5
CD with EasyWare V2.9.5 software
2030431-006
SPLY SPIRETTES #50
2030431-007
SPLY SPIRETTES #200
2030431-008
SPLY NOSECLIP #25
7.2 Accessories