Each instrument manufactured by GE Panametrics is warranted to
be free from defects in material and workmanship. Liability under
this warranty is limited to restoring the instrument to normal
operation or replacing the instrument, at the sole discretion of GE
Panametrics. Fuses and batteries are specifically excluded from
any liability. This warranty is effective from the date of delivery
to the original purchaser. If GE Panametrics determines that the
equipment was defective, the warranty period is:
• one year from delivery for electronic or mechanical failures
• one year from delivery for sensor shelf life
If GE Panametrics determines that the equipment was damaged
by misuse, improper installation, the use of unauthorized
replacement parts, or operating conditions outside the guidelines
specified by GE Panametrics, the repairs are not covered under
this warranty.
The warranties set forth herein are exclusive and are in
lieu of all other warranties whether statutory, express
or implied (including warranties or merchantability and
fitness for a particular purpose, and warranties arising
from course of dealing or usage or trade).
iii
April 2004
Return Policy
If a GE Panametrics instrument malfunctions within the warranty
period, the following procedure must be completed:
1. Notify GE Panametrics, giving full details of the problem, and
provide the model number and serial number of the
instrument. If the nature of the problem indicates the need for
factory service, GE Panametrics will issue a RETURN
AUTHORIZATION number (RA), and shipping instructions
for the return of the instrument to a service center will be
provided.
2. If GE Panametrics instructs you to send your instrument to a
service center, it must be shipped prepaid to the authorized
repair station indicated in the shipping instructions.
3. Upon receipt, GE Panametrics will evaluate the instrument to
determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
• If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and
returned.
• If GE Panametrics determines that the damage is not covered
under the terms of the warranty, or if the warranty has expired,
an estimate for the cost of the repairs at standard rates will be
provided. Upon receipt of the owner’s app roval to proceed, the
instrument will be repaired and returned.
Appendix D: Material Safety Data Sheets for Couplants
xiii
April 2004
Chapter 1
Features and Capabilities
The TransPort® Model PT878GC is a transit-time clamp-on gas
flowmeter which combines all the features of a full-size gas
flowmeter with the advantages of a portable instrument.
This section describes the TransPort features and general system,
and explains the theory of operation.
Overview
Depending on the selected application, the TransPort PT878GC
measures the flow of clean, dry gases in pipes from 4 to 24 in. in
diameter. The flowmeter provides one linear 0/4-20 mA analog
output of flow velocities or volumetric flow rate of these gases,
measuring velocities from ±1 to ±120 ft/sec (±0.3 to ±40 m/sec),
along with one selectable frequency or pulsed totalizer output.
Figure 1-1 illustrates the PT878GC in use on a typical pipe, along
with cases for the PT878GC and accessories.
Figure 1-1: PT878GC in Use and in Transport
Features and Capabilities1-1
April 2004
Overview (cont.)
The TransPort has the ability to store site data in files which can
be accessed at a later time. Within the Main Menu, a set of forms
(windows) asks you all the necessary setup information for a
particular site. Once you answer the necessary questions, you
simply save the information to a file.
The TransPort stores these files and other data in non-volatile
memory, which retains the information even if power is turned
off. The flowmeter itself runs on rechargeable or alkaline
batteries for up to 8 hours.
This small, lightweight flowmeter displays measurements in both
numeric and graphical form on a 240 x 200 pixel, EL-backlit
LCD graphic screen. The TransPort also has the capability of
logging over 100,000 flow data points internally.
Using an infrared communications port, the TransPort can
transmit or print logged data, as well as site files, real time data
and other stored data.
To as sist you, the TransPort is fully equipped with contextsensitive on-line help which is accessible at any time by simply
pressing the “?” (Help) key. Internal diagnostic and
troubleshooting features help isolate and remedy common
flowcell and transducer problems.
1-2Features and Capabilities
April 2004
System Description
The TransPort is only one part of the flowmeter system. The
other part of the system is the flowcell, which consists of the pipe
and the transducers.
The Flowcell
The flowcell is that part of the system that uses ultrasonic pulses
to interrogate the flow. It consists of the flowcell pipe and the
transducers.
A. FLOWCELL PIPE - The flowcell can either be created in the
existing piping (for example, by clamping non-wetted
transducers onto the pipe), or inserted as a substitute pipe
section (spoolpiece). The flowcell must provide mechanical
support for the transducers and assure stable conditions for
accurate flow measurement.
B. TRANSDUCERS - The transducers convert electrical energy
into ultrasonic pulses when in a transmit cycle, and convert
the ultrasonic pulses back to electrical energy when in a
receive cycle. In other words, they act like loudspeakers when
transmitting the signal and microphones when receiving it. In
the TransPort system, each transducer acts as both a receiver
and transmitter, since a series of ultrasonic pulses are
alternately sent upstream and then downstream through the
flowcell.
Features and Capabilities1-3
April 2004
Electronics Package
The TransPort consists of circuits that generate, receive, and
measure the travel time of the ultrasonic pulses. It also contains a
microcomputer that controls operation and calculates flow
measurement parameters. Specific circuits function as follows:
A. TRANSMIT SIGNAL GENERATOR - The transmit signal
generator, under control of the microcomputer and timing
circuit, synthesizes the signal that drives the transmitter.
B. TRANSMITTER - The transmitter amplifies the signals from
the transmit signal generator to a signal that drives the
transmit transducer.
C. RECEIVER - The receiver amplifies the received signals to a
level suitable for the data acquisition circuitry.
D. DATA ACQUISITION - The data acquisition circuitry
digitizes the received signal and stores it in a buffer for
processing by the microcomputer.
E. TIMING CIRCUIT - The timing circuit generates the
transmitter frequency, receive window, controls the data
acquisition circuit and the direction of the transmission.
F. MICROCOMPUTER - The microcomputer controls the
TransPort flowmeter’s operation and calculates flow
measurements derived from the transmitted and digitized
received signals. Also, the microcomputer continually checks
for faults and allows the use of built-in diagnostics for
troubleshooting.
G. INPUT/OUTPUT - The input/output circuitry allows the
flowmeter to indicate the measured flow with the
0/4 to 20-mA current loop, and to output to a printer or other
remote device.
1-4Features and Capabilities
April 2004
Theory of Operation
The TransPort is a transit-time ultrasonic flowmeter. When
ultrasonic pulses are transmitted through a moving liquid or gas,
the pulses that travel in the same direction as the fluid flow
(downstream) travel slightly faster than the pulses that travel
against the flow (upstream). The TransPort uses various digital
signal processing techniques, including cross-correlation, to
determine transit times and then uses the difference in transit
times to calculate flow velocity.
During operation, two transducers serve as both ultrasonic signal
generators and receivers. When mounted on a pipe, they are in
acoustic communication with each other, so that each transducer
can receive ultrasonic signals transmitted by the other transducer.
Each transducer thus functions as a transmitter generating a
certain number of acoustic pulses, and as a receiver for an
identical number of pulses.
The flowmeter measures the time interval between transmission
and reception of the ultrasonic signals in both directions. When
the gas in the pipe is not flowing, the transit-time downstream
equals the transit-time upstream. When the gas is flowing, the
transit-time downstream is less than the transit-time upstream.
The difference between the downstream and upstream transittimes is proportional to the velocity of the flowing gas, and its
sign indicates the direction of flow.
Features and Capabilities1-5
April 2004
Chapter 2
Initial Setup
Before making measurements, you must prepare the TransPort
for operation. This includes the following procedures:
• Making Electrical Connections
• Charging and/or Replacing Batteries
• Powering On and Off
• Using the Screen and Keypad
• Obtaining On-Line Help
Figure 2-1 below shows the PT878GC in its specially designed
case. The interior is molded for optimal protection of the meter
and its accessories.
Figure 2-1: The PT878GC and Accessories
Initial Setup2-1
April 2004
Making Electrical Connections
Before making measurements with the TransPort, you must make
all the necessary connections to the unit. This section describes
how to connect the following:
• Power
• Transducer
• Analog Input/Output
• Infrared Interface
Make all connections to the top of the TransPort unit as shown in
Figure 2-2 below. Please note that you only need to make the
proper transducer connections. The other connections are
required for particular functions, but are not necessary for basic
operation.
Input/Output
(See Table 2-1
on page 2-4)
Infrared
Transceiver
Power
Upstream
XDCR
Downstream
Figure 2-2: Connection Locations
Power Connections
The PT878GC is powered by either a 100-120/200-260 VAC
wall mount plug-in module, or by 5 internal C
energy rechargeable batteries or by a pack of 3.0 Ahr NiMH
batteries. (An optional power supplement, part #703-1283, uses 6
AA alkaline batteries.) When you receive the PT878GC, the
batteries are not charged; therefore, to make remote
measurements using the batteries, follow the instructions on
page 2-5 to charge the batteries. In either case, you must connect
the power cord to the appropriate terminal as shown in Figure 2-2
above.
-size NiCad high-
s
2-2Initial Setup
April 2004
!WARNING!
To ensure the safe operation of the TransPort,
you must install and operate it as described in
this manual. In addition, be sure to follow all
applicable safety codes and regulations for
installing electrical equipment in your area.
Transducer Connections
The transducer cables connect to the TransPort with LEMO®
coaxial type connectors. Each color-coded cable has a collar
labeled UPSTREAM or DOWNSTREAM. Make transducer
cable connections to the top of the flowmeter as shown in Figure
2-2 on the previous page. Transducer installation is discussed
separately in Chapter 3.
Analog Input/Output Connections
The TransPort provides one 0/4-20 mA current output and two
4 to 20-mA analog inputs with switchable 16-V supply for
loop- powered temperature transmitters. Connect the inputs/
outputs using a LEMO
2-2 on the previous page. The pin numbers for the connector and
the color code for the standard input/output cable are shown in
Table 2-1 on the next page.
®
multi-pin connector as shown in Figure
Initial Setup2-3
April 2004
Analog Input/Output Connections (cont.)
Table 2-1: Cable Assembly for Analog Inputs/Outputs
Wire
Pin Number
ColorDescription
1 Black Analog Out 1
2 Red16 V (switched)
3White Supply Temperature
4YellowReturn Temperature
5GreenAnalog Ground
6OrangeDigital Output
7BlueDigital Ground
8VioletReceive Monitor
The Infrared Wireless Interface
The PT878GC comes equipped with an internal infrared
transceiver (see Figure 2-2 on page 2-2) that enables
communication between the meter and other IR devices,
particularly the IR ports or dongles (IR to RS232 adapters) of
Windows
data. The PT878GC was designed for use with products that
comply to the IrDA protocol. For more information on
establishing IR communications between the PT878GC and your
PC, refer to Appendix B.
®
-based PCs. Users can send and receive site and log
2-4Initial Setup
April 2004
Charging and/or Replacing Batteries
The PT878GC comes with self-contained, built-in rechargeable
batteries to support portable operation. For optimum
performance, these batteries require a minimum of maintenance.
Charging the Batteries
When you receive the P T878GC, you will need to i nitially charge
the batteries. The batteries must be charged up to 8 hours to
receive the maximum charge. When fully charged, the batteries
provide 8 hours of continuous operation. An internal battery
gauge indicates the remaining power in the batteries.
To charge the batteries, simply plug the AC power module cord
into the power jack (shown in Figure 2-2 on page 2-2) and be sure
the battery pack is installed. When the PT878GC is plugged into
line voltage, the internal battery charger automatically charges
the batteries, whether the PT878GC is on or off. If the PT878GC
is on, the Battery icon in the upper right corner of the screen
indicates battery status (as shown in Table 2-2 below).
IMPORTANT: For CE compliance, the PT878GC is classified
as a battery-powered device, not to be used with
the AC adaptor.
Table 2-2: Battery Status Icons
IconBattery Status
Full battery
Partially full battery
Empty battery
Fully charged battery,
connected to AC power
Charging battery
Discharging battery
Failure/missing battery
Notification to check battery
form (see page 7-5)
Initial Setup2-5
April 2004
Replacing the Batteries
Caution!
Replace batteries only with the specified
rechargeable batteries. The battery charges when
the unit is off. Do not attempt to recharge non-
rechargeable batteries.
If you need to replace the rechargeable batteries, use 5 C
-size
s
NiCad high energy rechargeable batteries (GE Panametrics Part
Number 200-058) or 3.0 Ahr NiMH batteries (part number 200-
081). To replace the batteries, remove the rubber boot, open the
panel located on the back of the PT878GC unit, disconnect the
batteries, and replace with new ones (see Figure 2-3 below).
Battery Location
(Behind Panel)
Figure 2-3: Rear View of PT878GC
To further extend the battery power on the PT878GC, the GE
Panametrics Part #705-1283 power pack option uses 6 AA
alkaline batteries.
2-6Initial Setup
April 2004
Powering On and Off
To operate the PT878GC, the power cord must be plugged into
line voltage or the battery must be charged as described
previously.
IMPORTANT: For CE compliance, the PT878GC is classified
as a battery-powered device, and cannot be used
with the AC adaptor.
To turn the meter on, press the red button in the upper-right-hand
corner of the keypad. Immediately upon power up the PT878GC
emits a short beep and displays a “PCI Loader” message. It then
validates the instrument programming, and then displays the GE
Panametrics logo and the software version and emits a long beep.
If the meter fails any of these tests, contact GE Panametrics.
Caution!
If the meter fails the backup battery test, you must
send the unit back to the factory for a battery
replacement. Make sure you keep the NiCad
batteries charged until you are ready to ship the unit
back to the factory. Before shipping, print out all the
log and site data, or transfer it to your PC, as this
data will be lost during the service procedure.
Initial Setup2-7
April 2004
Powering On and Off (cont.)
After the meter conducts all the self checks, the screen then
appears similar to the one shown in Figure 2-4 below.
ABC.SIT
Velocity, ft/s
2004/11/30 09:53 AM
Signal, dB
32
0.00
Delta-T, ns
Volume, l/s
0.10
E0: No Errors
Figure 2-4: Screen After Powering On
0.0
2-8Initial Setup
April 2004
Powering On and Off (cont.)
To turn the PT878GC off, press the red key for 3 seconds. The
screen now appears similar to Figure 2-5 below.
Velocity, ft/s
SHUTDOWN: Meter OFF
Signal, dB
SLEEP: Meter Idle
CANCEL: Resume Operations
Delta-T, ns
Shutdown
Figure 2-5: The Shutdown Menu
Three options are available:
Sleep
Volume, l/s
Resume
• Press [F1] to shut down the PT878GC, turning it completely
off.
• Press [F2] to send the PT878GC into sleep mode. In this mode,
some of the power supplies shut down, but the PT878GC
remains in a standby mode. Users can resume taking
measurements immediately by pressing the power button.
• Press [F3] to cancel the command and return the PT878GC to
normal operation.
If the PT878GC locks up, you can reset it by holding the power
key (the red key in the upper right corner) for 15 seconds.
Initial Setup2-9
April 2004
Using the Screen and Keypad
The essential features for operating the TransPort are the screen
and keypad. Although these features are common on portable
instruments, the PT878GC design offers unique features to
simplify and speed operation.
Screen
The primary function of the screen is to display information in
order for you to accurately and easily take measurements. The
TransPort screen consists of seven parts (see Figure 2-6 below).
Current
Site
DEFAULT
Velocity, ft/s
Status Bar
(alternates with
Menu Bar)
2000/11/30 09:53 AM
Signal, dB
Battery
Status
32
0.00
(Work
Area)
Delta-T, ns
0.10
E0: No Errors
Error Messages
Figure 2-6: The PT878GC Screen in Operate Mode
The top line of the screen is the status bar, which normally
displays the time, date and current site. However, when you press
[MENU] (the menu key), the Menu Bar replaces the status bar.
System
Tray
Volume, l/s
0.0
Function Keys
2-10Initial Setup
April 2004
Screen (cont).
The middle of the screen is the work area, which displays the
measured parameters, numeric measurements, and both bar and
line graphs. (When you enter a selection on the Main Menu
discussed in Chapter 4, Programming Site Data, this area
displays menu prompts.) A line at the bottom of the area also
displays error code messages, which are described in more detail
in Chapter 10, Diagnostics and Trou bleshooting.
The system tray, shown in Figure 2-6 on page 2-10, displays
icons that indicate meter operations not otherwise shown. Table
2-3 below lists the icons and their meanings.
Table 2-3: Icons in the System Tray
IconFunctionMeaning
IR TransferIR data transfer in progress.
AlertIndicates the meter encoun-
tered an error in operation.
Log RunningIndicates a log is running
(marks).
Log PendingIndicates a log is pending (no
marks).
StopwatchCalibration Gate Operation:
Watch is stopped when the
gate is closed, or runs when it
is open. (See page 3-46.)
Snapshot
(To file)
(To Printer)
Indicates that the Snapshot
function has been activated,
so users can take screen captures (see page 6-23).
The bottom of the screen displays the three feature key options:
F1, F2 and F3. The feature keys have different functions,
depending on the task you are performing.
Initial Setup2-11
April 2004
Keypad
The PT878GC keypad has 25 keys. The functions for each key
are as follows (see Figure 2-7 on the next page):
• 3 function keys ([F1], [F2], [F3]) — enable you to select the
special functions which appear at the bottom of the screen.
• 12 numeric keys (including - and .) — enable you to enter
numeric data.
• 4 arrow keys ([W ], [X], [S], [T]) — enable you to move
through the menu options.
• [?] — Help key enables you to access on-line help.
• [MENU] — Menu key enables you to access the Menu Bar.
• [ENTER] — enables you to enter a particular menu, and
enters selected values into the TransPort memory.
• [SEL] — enables you to move between data measurements on
the screen.
• [ESC] — enables you to exit menus or menu options at any
time; cancels numeric entry.
• Red key [ ] turns the power on or off, and toggles the
backlight on or off.
2-12Initial Setup
Loading...
+ 278 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.