IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
Important Precautions
• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN
ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING
AVERTISSEMENT
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.
• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE PRESTATAIRE DE SERVICES DU CLIENT A BESOIN DE CE MANUEL DANS
UNE AUTRE LANGUE QUE L’ANGLAIS, IL INCOMBE AU CLIENT DE LE FAIRE
TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL DE MAINTENANCE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
• DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST
ES AUFGABE DES KUNDEN, FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU
SORGEN.
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN
ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND
NACHVOLLZIEHEN KÖNNEN.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
i
Page 4
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
• ESTE MANUAL DE SERVICIO SÓLO ESTÁ DISPONIBLE EN INGLÉS.
• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL
CLIENTE.
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO SIN HABER
AVISO
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL USUARIO O EL PACIENTE SUFRAN LESIONES
PROVOCADAS POR DESCARGAS ELÉCTRICAS, PROBLEMAS MECÁNICOS O
PELIGROS DE OTRA NATURALEZA.
• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM
INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
ATENÇÃO
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE PÔR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.
AVVERTENZA
HOIATUS
• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN
INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
• KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA
TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE
VÕI MUU OHU TAGAJÄRJEL.
ii -
Page 5
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
• TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.
• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN
ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
VAROITUS
JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN
VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.
• ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ ΔΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ.
• ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ
ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ΔΕ ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝΔΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.
FIGYELMEZTETÉS
VIÐVÖRUN
• EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ
EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
• ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ
ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
iii
Page 6
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
• TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.
• V PŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V
JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA
ÚKOLEM ZÁKAZNÍKA.
• NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI
VÝSTRAHA
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
• V PŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB,
OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.
• DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
ADVARSEL
DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE
PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN,
OPERATØREN ELLER PATIENTEN.
WAARSCHUWING
BRĪDINĀJUMS
• DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS
VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.
• ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
• ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪ
T ELEKTRISKĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
iv -
Page 7
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
• ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.
• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
• NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
ĮSPĖJIMAS
VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MECHANINIŲ AR
KITŲ PAVOJŲ.
• DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
ADVARSEL
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
OSTRZEŻENIE
ATE NŢIE
• NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU
ANGIELSKIM.
• JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
• NIE PRÓBOWAĆ SERWISOWAĆ
NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO
ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOżE GROZIĆ
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEŃ.
• ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.
v
Page 8
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
• ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯТОЛЬКОНААНГЛИЙСКОМЯЗЫКЕ.
• ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО
НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО
ОБЕСПЕЧИТЬ ПЕРЕВОД.
ОСТОРОЖНО!
• ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ
К ДАННОМУ РУКОВОДСТВУ И ПОЙМИТЕ ИЗЛОЖЕННЫЕ В НЕМ СВЕДЕНИЯ.
• НЕСОБЛЮДЕНИЕ УКАЗАННЫХ ТРЕБОВАНИЙ МОЖЕТ ПРИВЕСТИ К ТОМУ,
ЧТО СПЕЦИАЛИСТ ПО ТЕХОБСЛУЖИВАНИЮ
, ОПЕРАТОР ИЛИ ПАЦИЕНТ
ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ
ДРУГОЕ ПОВРЕЖДЕНИЕ.
• TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE.
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
UPOZORNENIE
NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA
ELEKTRICKÝM PRÚDOM, PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO
NEBEZPEČENSTVA.
VAR NING
DİKKAT
• DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
• BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.
• EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ
BİR DİLDE
OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN
SORUMLULUĞUNDADIR.
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
vi -
Page 9
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
vii
Page 10
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to position ing of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Othe r conn ections between pieces o f
electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel.
In performing all electrical work on these products, GE will use its own specially trained field engineers.
All of GE’s electrical work on these products will comply with the requirements of the applicable
electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Healthcare Global Documentation Group with specific information listing the system type, manual
title, part number, revision number, page number and suggestion details. Mail the information to:
Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI 53226, USA.
GE Healthcare employees should use the iTrak System to report all documentation errors or omissions.
viii -
Page 11
GE HEALTHCARE
D
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE Healthcare.
GE Healthcare may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
IRECTION 5184024-100, REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL
This page was intentionally left blank.
x -
Page 13
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
Revision History
RevisionDateReason for change
12006-07-25Initial Version for LOGIQworks 3.0
22006-08-07Released Version for LOGIQworks 3.0
32007-01-09Added Dell Precision 390 and description about TripleMonitor
42007-03-21Released Version for LOGIQworks 3.0.4
52007-09-04Version for LOGIQworks 3.0.5
List of Effected Pages
PagesRevisionPagesRevisionPagesRevision
TitleN/A
Important Precautions
i - vi
Revision History
vii - viii
Table of Contents
ix to xx
Chapter 1 - Introduction
on page 1-1 to 1-20
Chapter 2 - Pre-
Installation on page 2-1
to 2-20
Chapter 3 - Installation on page 3-1
to 3-207
5
5
5
5
5
Chapter 4 - Functional Checks on
page 4-1 to 4-32
Chapter 5 - Components and
Functions (Theory) on page 5-1 to
5-42
Chapter 6 - Service Adjustments on
page 6-1 to 6-6
Chapter 7 - Diagnostics/
Troubleshooting on page 7-1 to 7-7
Chapter 8 - Replacement
Procedures on page 8-1 to 8-28
5
5
5
5Back CoverN/A
5
5
Chapter 9 - Renewal Parts on page 9-1
to 9-10
Chapter 10 - Quality Assurance on page
10-1 to 10-16
Index
I to II
5
5
5
xi
Page 14
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
This page was intentionally left blank.
xii -
Page 15
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5 LOGIQWORKS® BASIC SERVICE MANUAL
Technical Specifications of System based on Dell Precision 340 . . 5 - 3
Theoretical Description of System based on Dell Precision 340 . . . 5 - 3
Technical Specifications of System based on Dell Precision 360 . . 5 - 3
Technical Specifications of System based on Dell Precision 370 . . 5 - 3
Technical Specifications of System based on Dell Precision 380 . . 5 - 4
Technical Specifications of System based on Dell Precision 390 . . 5 - 4
Theoretical Description of System based on Dell Precision 360, Dell Precision
Theoretical Description of System based on Dell Precision 340 . . . 5 - 8
Theoretical Description of System based on Dell Precision 360, 370, 380, 390
5 - 9
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Chapter 1
Introduction
Section 1-1
Overview
1-1-1Purpose of Chapter 1
This chapter describes important issues related to safely servicing the LOGIQworks.
The service provider must read and understand all the information presented in this manual before
installing or servicing a unit.
Table 1-1 Contents in Chapter 1
SectionDescriptionPage Number
1-1
1-2
1-3
1-4
Overview
Important Conventions
Safety Considerations
Customer Assistance
1-1-2Purpose of Service Manual
This Service Manual provides installation and service information for the LOGIQworks and contains the
following chapters:
1.) Chapter 1 - Introduction: Contains a content summary and warnings.
2.) Chapter 2 - Pre-Installation: Contains any pre-installation requ ir em ents for the LOGIQworks.
3.) Chapter 3 - Installation: Contains the LOGIQworks installation procedure with installation
checklist.
4.) Chapter 4 - Functional Checks: Contains functional checks that must be performe d as part of the
installation, or as required during servicing and periodic maintenance.
5.) Chapter 5 - Components and Functions (Theory): Contains block diagrams and functional
explanations of the LOGIQworks electr on ics .
6.) Chapter 6 - Service Adjustments: Contains instructions on how to make available adjustments to
the LOGIQworks.
7.) Chapter 7 - Diagnostics/Troubleshooting: Prov ides procedures for running diagnostic or related
routines for the LOGIQworks.
8.) Chapter 8 - Replacement Procedures: Provides disassembly procedures and reassembly
procedures for all changeable Field Replaceable Units (FRU).
9.) Chapter 9 - Renewal Parts: Contains a complete list of replacement parts for the LOGIQworks.
10.)Chapter 10 - Quality Assurance: Provides periodic maintenance procedures for the LOGIQworks.
1-1
1-4
1-7
1-19
Chapter 1 Introduction1-1
Page 28
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Table 1-5 LOGIQworks Standard Hardware of System based on Dell Precision 380 (delivered
from November 2005) and on Dell Precision 390 (delivered from December 2006)
PARTPCLCD1 LCD2Modem
BTU/h1280195195164
Power (W)375575748Total BTU/h
H49021AKxxx1639
H49021ALxxxx1834
H49021AMxxx1639
H49021ANxxx1639
LOGIQworks Models
1-1-4-1How to Turn the LOGIQworks workstation on/off
See chapter 3, Section 3-4-4 on page 3-60.
1-1-4-2How to Check for Hardware/Software version, Options
See chapter 3, Section 3-5-13 on page 3-203.
1-1-5Purpose of Operator Manual(s)
The Operator Manual(s), Part Numbers available in chapter 9, should be fully read an d understood
before operating the LOGIQworks workstation and also kept near the unit for quick reference.
Chapter 1 Introduction1-3
Page 30
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Section 1-2
Important Conventions
1-2-1Conventions Used in Book
Model Designations
This manual covers the LOGIQworks PC.
(See Chapter 1 - LOGIQworks Models covered by this Manual on page 1-2.)
Icons
Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
Safety Precaution Messages
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identifi ed by a flag word that precedes the precautionary
message. Known or potential hazards are labeled in one of following ways:
WARNING
WARNING
Warnings that must take heed of to protect the patients / users against the danger of injury.
CAUTION
NOTE
Warnings that must be heeded to avoid causing damage to the equipment.
General or additional information relating to specific topics.
1-4Section 1-2 - Important Conventions
Page 31
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-2-2Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within a triangle, as
seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be
used to make you aware of specific types of hazards that could cause harm.
Table 1-6 Standard Hazard Icons
ELECTRICALMECHANICALRADIATION
LASERHEATPINCH
LASER
LIGHT
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-7 Standard Icons Indicating a Special Procedure be Used
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
TAG
&
LOCKOUT
Date
Signed
EYE
PROTECTION
Chapter 1 Introduction1-5
Page 32
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-2-3Product Icons
The following table describes the purpose and locatio n of safety labels and other important in formation
provided on the equipment.
Table 1-8 Product Icons
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
Manufacturer's name and address
Identification and Rating Plate
Device Listing/Certification Labels
Type/Class LabelUsed to indicate the degree of safety or protection.
“CAUTION
This unit weighs...
Special care must be used to avoid..."
Model and serial numbers
Electrical ratings
Laboratory logo or labels denoting
conformance with industry safety standards
such as UL or IEC.
This precaution is intended to prevent injury
that may result if one person attempt to move
the unit considerable distances or on an incline
due to the weight of the unit.
"CAUTION" The equilateral triangle is usually
used in combination with other symbols to
advise or warn the user.
On top of PC
Rear side of the unit
Rear side of the monitor
Service Manual
Various
ATTENTION - Consult accompanying
documents " is intended to alert the user to
refer to the operator manual or other
instructions when complete information cannot
be provided on the label.
"CAUTION - Dangerous voltage" (the lightning
flash with arrowhead in equilateral triangle) is
used to indicate electric shock hazards.
"Mains OFF" Indicates the power off position of
the mains power switch.
Various
Various
ON/OFF switches
1-6Section 1-2 - Important Conventions
Page 33
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Table 1-8 Product Icons (Continued)
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
"OFF/Standby" Indicates the power off/
standby position of the power switch.
CAUTION
This Power Switch DOES NOT ISOLATE
Mains Supply
"Mains ON" Indicates the power on position of
the mains power switch.
"Protective Earth" Indicates the protective
earth (grounding) terminal.
ON/OFF switches
Section 1-3
Safety Considerations
WARNING
WARNING
1-3-1Safety Hints
For hardware related risks please pay attention to the safety instructions of the DELL hardware
manuals (e.g. System Information Guide).
If hardware from different manufacturers as ViewPoint is used Safety Notes are in their responsibility.
So also see in the relevant manuals.
"Equipotentiality" Indicates the terminal to be
used for connecting equipotential conductors
when interconnecting (grounding) with other
equipment.
“Nonionizing Radiation” indicates that the
system applies RF energy.
Rear of printer near power inlet
Chapter 1 Introduction1-7
Page 34
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-1Safety Hints (cont’d)
WARNING
WARNING
Danger of Electrical Shock!
Danger of Electrical Shock!
A connection between the patient and a personal computer can result in risks for the patient from
electric current. Consider, therefore, these safety references:
– Use only CE marked electrical equipment!
– Install a galvanic isolator to insulate the connection between the personal computer and all
connected medical equipment (e.g. fibre optic cable or certified video isolator) !
– Set up the personal computer outside the patient's vicinity (1.5 m/4.9 ft.)!
Use only CE marked galvanic isolators or medical devices appropriate under
EN/IEC 60601-1 and EN/IEC 60601-1-1!
Figure 1-1 Sketch of the patient’s environment
WARNING
WARNING
Danger of Electromagnetic Incompatibility!
Devices that are not CE marked can create an electromagnetic field which could impair the
operation of other devices.
To prevent this, check if the device is CE marked or FCC-marked.
Use only CE marked devices and medical devices appropriate under
EN/IEC 60601-1-2.
1-8Section 1-3 - Safety Considerations
Page 35
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-1Safety Hints (cont’d)
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
Danger of False Diagnosis!
Danger of Data Loss!
Personal computers can be affected by electromagnetic fields from other devices.
Do not place devices which could cause electromagnetic interference near the personal computer!
Danger of False Diagnosis!
An incorrect monitor or incorrect monitor configuration can lead to false diagnosis.
Adjust the monitor to display the images in adequate quality according to the manufacturer's
instructions for use. If in doubt, please contact your service representative.
Danger of Data Loss!
System connected to the inter-/intranet could b e infected by a computer virus. This infection can resu lt
in a loss of data or falsified data or an unoperable system.
WARNING
WARNING
CAUTION
CAUTION
The user has to ensure that the LOGIQworks system is protected from viruses, e.g. by using antivirus
programs, firewalls and regular updates for antivirus programs, firewalls and the operation system.
Danger of Data Loss!
Activation of the function for automatic deletion of patient studies might cause data loss.
To avoid this back up your data frequently.
Danger of Power Failure!
We recommend to install an UPS (Uninterruptible Power Supply) if frequent power failure occurs.
Danger of Inoperable System!
LOGIQworks should be installed by a GE service technician to avoid unoperable system.
Chapter 1 Introduction1-9
Page 36
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-2Label Locations of System based on Dell Precision 340
Sony Labels only. Guarantee agreement
3
14
2
5
Figure 1-2 Labeling
1.) CE-Label
2.) VDE-Label
3.) UL-Label
4.) Identification plate
5.) Homologation label (for Japan only)
1-10Section 1-3 - Safety Considerations
Page 37
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-2Label Locations of System based on Dell Precision 340 (cont’d)
Power Input
Information
Tower Rear
Tower Top Side
(Seen from front)
Front
Serial Number
Figure 1-3 Computer Tower Labelling
Chapter 1 Introduction1-11
Page 38
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-2Label Locations of System based on Dell Precision 340 (cont’d)
CRT Monitor, seen
from the rear
Serial Number
Figure 1-4 CRT Monitor Labels
LCD Monitor seen
from the rear
Figure 1-5 LCD NEC 1880 Monitor Labels
1-12Section 1-3 - Safety Considerations
Page 39
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-2Label Locations of System based on Dell Precision 340 (cont’d)
Modem seen from below
Figure 1-6 Recovery CD Label with Rating Plate Label
Figure 1-7 Modem Labels
Chapter 1 Introduction1-13
Page 40
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-3Label Locations of System based on Dell Precision 360 and Dell Precision 370
Power Input
Information
Tower Rear
Serial Number
Tower Top Side
(Seen from front)
Front
Figure 1-8 Computer Tower Labeling
1-14Section 1-3 - Safety Considerations
Page 41
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-3Label Locations of System based on Dell Precision 360 and Dell Precision 370
(cont’d)
Tower Front
Figure 1-9 Additional Computer Labeling for Dell Precision 370 only
Chapter 1 Introduction1-15
Page 42
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-3Label Locations of System based on Dell Precision 360 and Dell Precision 370
(cont’d)
LCD Monitor seen
from the rear
Figure 1-10 LCD DELL 2001 Monitor Labels
LCD Monitor seen
from the rear
Figure 1-11 LCD DELL 1901 Monitor Labels
1-16Section 1-3 - Safety Considerations
Page 43
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-4Label Locations of System based on Dell Precision 380 and Dell Precision 390
Tower Rear
Tower Bottom Side
Tower Top Side
Figure 1-12 Computer Tower Labeling
Figure 1-13
Chapter 1 Introduction1-17
Page 44
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
1-3-5Label Locations
Modem seen from below
Figure 1-14 Recovery CD Label with Rating Plate Label
Figure 1-15 Modem Label
1-18Section 1-3 - Safety Considerations
Page 45
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Section 1-4
Customer Assistance
1-4-1Contact Information
If this equipment does not work as indicated in this service manual or in the Basic User Manual, or if
you require additional assistance, please contact the local distributor or appropriate support resource,
as listed below.
Prepare the following information before you call:
•System ID serial number.
•Software version.
Table 1-9 Phone Numbers for Customer Assistance
LocationPhone Number
USA/Canada
GE Medical Systems
Ultrasound Service Engineering
4855 W. Electric Avenue
Milwaukee, WI 53219
Customer Answer Center
Europe
GE Ultraschall Deutschland GmbH & Co. KG
Beethovenstraße 239
Postfach 11 05 50, D-42665 Solingen
Germany
Asia Singapore/Japan)
GE Ultrasound Asia
Service Department - Ultrasound
298 Tiong Bahru Road #15-01/06
Central Plaza
Singapore 169730
Service
1-800-437-1171
1-800-321-7937 (Latin America)
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Chapter 2
Pre-Installation
Section 2-1
Overview
2-1-1Purpose of Chapter 2
This chapter provides the information required to plan and prepa re for the installation of a LOGIQworks.
Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit. A
checklist is also provided at the end of this section to help determine whether the proper planning and
preparation is accomplished before the actual equipment installation is scheduled.
Table 2-1 Contents in Chapter 2
SectionDescriptionPage Number
2-1
2-2
2-3
Overview
General Console Requirements
Facility Needs
2-1
2-2
2-11
Chapter 2 Pre-Installation2-1
Page 48
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
Section 2-2
General Console Requirements
2-2-1Environmental Requirements of System based on Dell Precision 340
Table 2-2 Dell Precision 340 Workstation
OperationNon-OperationStorageTransport
Temperature
Humidity
Altitude
10 to 35 oC
(50 to 95
20% to 80%
non-condensing
-52 - 10000ft
-16 - 3048 m
o
F)
Table 2-3 NEC LCD 1880 Monitor
OperationNon-OperationStorageTransport
Temperature
Humidity
Altitude
5 to 35 oC
o
(41to 95
30% to 80%
non-condensing
0 - 15912ft
0 - 4850m
F)
Table 2-4 Global Modem
o
-40 to 65
(- 40 to 149
20% to 80%
non-condensing
-52 - 34777ft
-16 - 10600 m
-10 to 60
(- 14 to 140
10% to 85%
non-condensing
0 - 45000ft
0 - 13716m
C
o
F)
o
C
o
F)
o
-40 to 65
(- 40 to 149
20% to 80%
non-condensing
52 - 34777ft
-16 - 10600 m
-10 to 60
(-14 to 140
10% to 85%
non-condensing
0 - 45000ft
0 - 13716m
C
o
F)
o
C
o
F)
o
-40 to 65
(-40 to 149
20% to 80%
non-condensing
52 - 34777ft
-16 - 10600 m
-10 to 60
(-14 to 140
10% to 85
non-condensing
0 - 45000ft
0 - 13716m
C
o
o
C
o
F)
F)
OperationNon-OperationStorageTransport
Temperature
Humidity
Altitude
0 to 50 oC
(32 to 120
20% to 90%
non-condensing
o
F)
N/AN/AN/AN/A
(32 to 120
20% to 90%
non-condensing
0 to 50
o
C
o
F)
-20 to 70
(-4 to 158
20% to 90%
non-condensing
2-2-1-1Cooling
The cooling requirement for the LOGIQworks is within the ran ge 1220-2100 BTU/hr . This is depe nding
on whether a CRT Monitor or LCD Monitor is used, a nd on the har dware options insta lled. For detailed
BTU/hr. see table 2-7 and table 2-8. The BTU/hr. value relates to the maximum operational Power
consumption of the LOGIQworks Works ta tion . Th is figure does not include cooling needed for lights,
people, or other equipment in the room. Each person in the room places an additional 300 BTU/hr.
Demand on the cooling system.
2-2Section 2-2 - General Console Requirements
o
C
o
F)
-20 to 70
(-4 to 158
20% to 90%
non-condensing
o
C
o
F)
Page 49
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-2-1-1Cooling (cont’d)
Table 2-5 LOGIQworks Standard Hardware of System based on Dell Precision 340
PARTPCCRT1 CRT2 LCD1LCD2Modem
BTU/HR860495495195195164
Power (W)250135135575748Total BTU/HR
H49021LAxxx1520
H49021LBxxxx2015
H49021LCxxx1220
H49021LDxxxx1415
LOGIQworks Models
2-2-1-2Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a sou rce
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
Chapter 2 Pre-Installation2-3
Page 50
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
2-2-2Environmental Requirements of System based on Dell Precision 360, on Dell
Precision 370, on Dell Precision 380 or on Dell Precision 390
The cooling requirement for the LOGIQworks is within the r ange 122 0-210 0 BTU/hr. T his is depend ed
on which hardware options that have been installed. For detailed BTU/hr. see table 2-7 and table 2-8.
The BTU/hr. value relates to the maximum operational Power consumption of the LOGIQworks
Workstation. This figure does not include cooling needed for lights, people, or other equipment in the
room. Each person in the room places an additional 300 BTU/hr. Demand on th e cooling system.
2-4Section 2-2 - General Console Requirements
o
C
o
F)
Page 51
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-2-2-1Cooling (cont’d)
Table 2-9 LOGIQworks Standard Hardware of DELL Precision 360 and DELL Precision 370
PARTPCLCD1LCD2Modem
BTU/h860195195164
Power (W)250575748Total BTU/h
Table 2-10 LOGIQworks Standard Hardware of DELL Precision 380 and of Dell 390
PARTPCLCD1LCD2Modem
BTU/h1280195195164
Power (W)357575748Total BTU/h
H49021AKxxx1220
H49021ALxxxx1415
H49021AMxxx1220
H49021ANxxx1415
LOGIQworks Models
Chapter 2 Pre-Installation2-5
Page 52
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
2-2-2-1Cooling (cont’d)
Table 2-11 LOGIQworks Optional Hardware
PARTModem
BTU/h164
Power (W)48Total BTU/h
Allx164
2-2-2-2Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2-2-3Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
Table 2-12 Average Installation Time
DescriptionAverage Installation
Time
Unpacking 15 minutes1 person
Checking delivery15 minutes1 person
Installation of
LOGIQworks
Check connectivity
and setup
1.0 hour1 person
15 minutes1 person
Comments
2-6Section 2-2 - General Console Requirements
Page 53
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-2-4Electrical Requirements
NOTE
LOGIQworks requires a dedicated power and ground for the proper operation of its workstation. This
dedicated power shall originate at the last distribution pane l before the system. The dedicated lin e shall
consist of one phase, a neutral (not shared wi th any o ther circuit) , and a full size gr ound wir e fro m the
distribution panel to the workstation. Please note that image artifacts can occur, if at any time within the
facility, the ground from the main facility's incoming power source to the workstation is only a conduit.
2-2-4-1LOGIQworks Power Requirements
Electrical Specifications for LOGIQworks
Table 2-13 Electrical Specifications of LOGIQworks System based on Dell Precision 340
It is recommended that the branch circuit breaker for the machine be readily accessible.
2-2-4-4Site Power Outlets
Three to five AC power outlet must be less than 1.2 meters (3.9feet) from the LOGIQworks Wor kstation.
Electrical installation must meet all current local, state, and national electrical codes.
Chapter 2 Pre-Installation2-7
Page 54
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
2-2-4-4Site Power Outlets (cont’d)
WARNING
WARNING
POWER OUTAGE MAY OCCUR.
The LOGIQworks requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.
2-2-4-5Unit Power Plug
If the unit arrives without a power plug, or with the wrong plug, you must contact GE Service; they must
supply what is locally required.
2-8Section 2-2 - General Console Requirements
Page 55
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-2-5EMI Limitations
WARNING
WARNING
WARNING
WARNING
Danger of Electromagnetic Incompatibility!
Devices that are not CE marked can create an ele ctromagnetic field which could impair the operation
of other devices.
To prevent this, check if the device is CE marked or FCC-marked.
Use only CE marked devices and medical devices appropriate under
EN/IEC 60601-1-2.
Danger of False Diagnosis!
Danger of Data Loss!
Personal computers can be affected by electromagnetic fields from other devices.
Do not place devices which could cause electromagnetic interference near the personal computer!
Computers are susceptible to Electromagnetic Interference (EMI) from radio freque ncies, magnetic
fields, and transients in the air or wiring. They also ge nerate EMI. The LOGIQworks complies with limits
as stated on the EMC label. However there is no guarantee that interference will not occur in a particular
installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect.
These sources include:
•medical lasers,
•scanners,
•cauterizing guns,
•computers,
•monitors,
•fans,
•gel warmers,
•microwave ovens,
•light dimmers
•portable phones.
The presence of a broadcast station or broadcast van may also cause interference.
see Table 2-16 for EMI Prevention tips
Table 2-16 EMI Prevention/Abatement
EMI RuleDetails
Keep the unit at least 5 meters or 15 feet away from other EMI sources. Special shielding may
Be aware of RF sources
be required to eliminate interference problems caused by high frequency, high powered radio
or video broadcast signals.
Ground the unit
Poor grounding is the most likely reason a unit will have noisy images. Check grounding of the
power cord and power outlet.
Chapter 2 Pre-Installation2-9
Page 56
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
Table 2-16 EMI Prevention/Abatement
EMI RuleDetails
Take care with cellular
phones
Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
2-10Section 2-2 - General Console Requirements
Page 57
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Section 2-3
Facility Needs
2-3-1Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay,
confusion, and waste of manpower can be avoided by comple ting pre installation work before d elivery.
Use the Pre installation checklist to verify that all needed steps have been taken. Purchaser
responsibility includes:
•Procuring the materials required.
•Completing the preparations before de liver y of th e LO GI Qw or ks.
•Paying the costs for any alterations and modification s not specifically provided in the sales contract.
WARNING
WARNING
All electrical installations that are preliminary to the positionin g of the equipment at the site prepared for
the equipment must be performed by licensed electrical contractors. Other connections between pieces
of electrical equipment, calibrations, and testing must also be performed by qualified personnel. The
products involved (and the accompanying electrical installations) are highly sophisticated and special
engineering competence is required. All electrical work on these products must comply with the
requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified
personnel to perform electrical servicing on the equipment.
The desire to use a non–listed or customer provided product or to place an approved prod uct further
from the workstation than the interface kit allows presents challenge s to the installation team. To avoid
delays during installation, such variances should be made known to the in dividuals or group performing
the installation at the earliest possible date (preferably prior to the purchase).
The workstation suite must be clean prior to delivery of the machine. Carpet is not recommended
because it collects dust and creates static. Potential sources of EMI (electromagnetic interference)
should also be investigated before delivery. Dirt, static, and EMI can negatively impact system reliability.
2-3-2Required Features
•Dedicated single branch power outlets of adequate amperage (see Table 2-14) meeting all local
and national codes which is located less than 1.5 m (5 ft.) from the workstation’s proposed location.
•Proposed location for the workstation is at least 0.3 m (1 ft.) from the wall for cooling.
•Power outlet and place for any external peripheral are within 2 m (6.6ft) of each other with peripheral
within 1 m of the unit to connect cables.
•Power outlets for test equipment within 1 m (3.3 ft.) of PC.
•Either an Analog Phone Line, or a GE Approved Broadband/VPN connection, is required for iLinq.
A functional LAN is required for Network connectivity.
•An Ethernet network wall jack is required.
2-3-3Desirable Features
•Circuit breaker for dedicated power outlet is easily accessible.
•Dual level lighting (bright and dim).
•Lockable cabinet ordered by GE for its software and proprietary manuals.
Chapter 2 Pre-Installation2-11
Page 58
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
2-3-3Desirable Features (cont’d)
Work Station parts
A dedicated Ethernet network wall outlet
243,84 cm
243,84 cm
Figure 2-1 Minimum room needs for LOGIQworks
LOGIQworksparts
1m14cm
304.8 cm
Vivid 3
1.5m
3 to 4 dedicated Power outlets
A dedicated Analog Telephone jack
A dedicated Ethernet network wall outlet
3 to 4 dedicated Power outlets
A dedicated Analog Telephone jack
Hot & Cold water
A dedicated Ethernet
network wall outlet
A dedicated Power
outlet
105cm
609.6 cm
Figure 2-2 Suggested Maximum Room with LOGIQworks and Ultrasound Scanner
2-12Section 2-3 - Facility Needs
A dedicated Analog
Telephone jack
Page 59
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-3-3-1Patient environment IEC 60601-1-1
An area in which patients are normally cared for, the patient vicinity is the space with surfaces likely to
be contacted by the patient or attendant who can touch the patient.
In practice a distance of 2,5 m (8.2 ft.) above the floor on which the medical personnel stand and a
horizontal distance of 1,5 m (4.9 ft.) have justified themselves as indicative of the dimensions of the
Patient Environment.
The patient environment/vicinity will be depicted as a dashed line in this procedure. See example below.
WARNING
WARNING
Danger of Electrical Shock!
Danger of Electrical Shock!
A connection between the patient and a personal computer can result in risks for the patient from
electric current. Consider, therefore, these safety references:
– Use only CE marked electrical equipment!
– Install a galvanic isolator to insulate the connection between the personal computer and all
connected medical equipment (e.g. fibre optic cable or certified video isolator)!
– Set up the personal computer outside the patient's vicinity (1.5 m/4.9 ft.)!
Use only CE marked galvanic isolators or medical devices appropriate under
EN/IEC 60601-1 and EN/IEC 60601-1-1!
Figure 2-3 Patient safety information
2-3-3-2Patient environment IEC 601-1
Sub clause 2.204
Such an area is an environment in which medical diagnosis, monitoring or treatment is carried out. It is
very difficult to attach unique dimensions to the PATIENT ENVIRONMENT.
Chapter 2 Pre-Installation2-13
Page 60
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
2-3-4Networking Pre-installation Requirements
2-3-4-1LOGIQworks Connected to a Network
Supported networks:
•10 or 100 Megabit Network can be used.
2-3-4-2Stand Alone Network or Integrated into Hospital’s Network
If LOGIQworks is to be used in a separate network together with a LOGIQ 9 or LOGIQ 7 and a network
printer with no connection to the Hospital’s network or LOGIQworks is to be on the Hospital network
either the predefined factory settings should be used.
To configure the LOGIQworks to work with other network connections, the site’s network administrator
must provide information to complete the Data Co llection Documen t on p age 14. Ensure th at there are
no spaces in any field of the form except for those fields under PHYSICAL LOCATION. Entries must
include:
•Verify that there is a Conformancy Statement that documents the DICOM server you are about to
use supports Ultrasound images.
•A host name, local port number, AE Title, IP address and Net Mask for the LOGIQworks.
•The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
•The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQworks for DICOM APPLICATION INFORMATION. Fields for Description, Institution Name,
Institution Department Name and Station Name are also available to describe the sending devices
in more detail. This information may be useful for error solv ing .
2-3-4-4Insite 2
For Insite 2 a connection to the internet is required.
Recommendation: A permanent broadband to the internet. The customer is responsible for the
connectivity, configuration and operation of this internet access funktionality.
2-14Section 2-3 - Facility Needs
Page 61
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-3-4-4Insite 2 (cont’d)
Figure 2-4 Data Collection Document Page 1
Chapter 2 Pre-Installation2-15
Page 62
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
2-3-4-4Insite 2 (cont’d)
Figure 2-5 Data Collection Document Page 2
2-16Section 2-3 - Facility Needs
Page 63
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-3-4-4Insite 2 (cont’d)
Figure 2-6 Data Collection Document Page 3
Chapter 2 Pre-Installation2-17
Page 64
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
2-3-4-4Insite 2 (cont’d)
Figure 2-7 Data Collection Document Page 4
2-18Section 2-3 - Facility Needs
Page 65
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
2-3-4-4Insite 2 (cont’d)
Figure 2-8 Data Collection Document Page 5
Chapter 2 Pre-Installation2-19
Page 66
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
Your Notes:
2-20Section 2-3 - Facility Needs
Page 67
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Chapter 3
Installation
Section 3-1
Overview
3-1-1The Purpose of Chapter 3
This chapter contains information NEEDED to install the UNIT, a procedure that desc rib es how to
receive and unpack the equipment and how to file a damage or loss claim.
How to prepare the facility and unit of the actual installation, and how to check and test the unit and
external peripherals for electrical safety are included in this procedure.
Table 3-1 Contents in Chapter 3
SectionDescriptionPage Number
3-1
3-2
3-3
3-4
3-5
3-6
Overview
Receiving and Unpacking the Equipment
Preparing for Installation
Completing the Installation
Configuration
Software Upgrades
3-1-2Average Installation Time
For time and personnel estimates during the installation see: Section 2-2-3 "Time and Manpower
Requirements" on page 2-6.
3-1
3-3
3-22
3-25
3-68
3-205
Chapter 3 Installation3-1
Page 68
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-1-3Safety Reminders
WARNING
WARNING
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
WARNING
WARNING
Two people should unpack the unit because of its weight. Two people are required whenever a part
weighing 22 kg (50 lbs.) or more must be lifted.
CAUTION
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to its
operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a
three to two prong adapter. This defeats safety grounding.
WARNING
WARNING
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQworks and
kept near the unit for quick reference.
3-2Section 3-1 - Overview
Page 69
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Section 3-2
Receiving and Unpacking the Equipment
3-2-1Shipment Description
From the source factory all LOGIQworks parts are packed in the original packaging a nd put into a cardboard box.
3-2-1-1Check of Delivered Box
1.) Check that cardboard box is undamaged as it was shipped.
CAUTION
Do not accept if damaged!
2.) Check that they match the contents on the included Delivery Note.
3.) Check that each box is undamaged and unopened.
4.) Open each of these cardboard boxes.
5.) Check that contents are there an d un d am aged.
6.) Move the shipment to the place where it is to be installed.
7.) Unpack the Parts, run the check list shown below.
8.) Handling Incomplete or Damaged Shipment.
Please read that procedure before packing/unpacking the LOGIQworks.
User Manual(s):
Check that the correct User Manual(s) for the system and software revision is included with the
installation. Specific language versions of the User Manual may also be available. Check with your GE
Sales Representative for availability.
We strongly advise you to store the LOGIQworks packing material in undamaged condition in case of
future transportation.
Chapter 3 Installation3-3
Page 70
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-2-1-1Check of Delivered Box (cont’d)
WARNING
WARNING
Transport only with forklift or stracker truck.
During transport pay attention to the point of gravity!
Have two people available to unpack the LOGIQworks.
Attempts to move the unit considerable distances or on an incline by one person could result
in injury or damage or both.
The envelope with delivery address, packing list and
invoice is located on the front panel of the crate.
Check whether delivery is complete (according to
packing list) and check visual damage!
CAUTION
Shipping packaging, cut the black ribbon
with a knife and unpack the system components
Figure 3-1 Shipping packaging and envelope at front panel of the crate
The device must only be transported in the original packaging!
Unpack the devices such a way that packaging can be reused.
3-4Section 3-2 - Receiving and Unpacking the Equipment
Page 71
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-2-1-1Check of Delivered Box (cont’d)
3
2
Figure 3-2 The LOGIQworks H49021LC delivery
Table 3-3 The LOGIQworks H49021LC parts
NameQty
1Dell Precision 3401
2Monitor NEC LCD 1880 Sx1
3Modem1
1
Chapter 3 Installation3-5
Page 72
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-2-1-1Check of Delivered Box (cont’d)
3
2
Figure 3-3 The LOGIQworks H49021LD delivery
Table 3-4 The LOGIQworks H49021LD parts
NameQty
1Dell Precision 3401
2Monitor NEC LCD 1880 Sx2
3Modem1
1
3-6Section 3-2 - Receiving and Unpacking the Equipment
Page 73
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-2-1-1Check of Delivered Box (cont’d)
3
2
Figure 3-4 The LOGIQworks H49021LA delivery
Table 3-5 The LOGIQworks H49021LA parts
ItemNameQty
1Dell Precision 3401
2Monitor Sony GMD-F5201
3Modem1
1
Chapter 3 Installation3-7
Page 74
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-2-1-1Check of Delivered Box (cont’d)
3
1
2
Figure 3-5 The LOGIQworks H49021LB delivery
Table 3-6 The LOGIQworks H49021LB parts
NameQty
1Dell Precision 3401
2Monitor Sony GMD-F5202
3Modem1
3-8Section 3-2 - Receiving and Unpacking the Equipment
Page 75
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-2-1-1Check of Delivered Box (cont’d)
2
3
1
Figure 3-6 The LOGIQworks H49021AC/H49021AA or H49021AK/H49021AM delivery.
Table 3-7 The LOGIQworks 1FP H49021AA/H49021AC parts
3-18Section 3-2 - Receiving and Unpacking the Equipment
Page 85
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-2-1-5Sony CRT Monitor
1
4
3
2
Figure 3-15 Parts in the CRT delivery
Table 3-18 Contents of CRT Monitor Box
ItemNameQty
1Monitor CRT1
2Signal Cable analog1
3
4Documentation1
(*If not shipped to Germany, the EU
power cord is not used; see Table 3-22
Power Cable EU
and Table 3-23 for power cord
information)
*
1
Chapter 3 Installation3-19
Page 86
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-2-1-6Modem Kit Parts Delivery Inspection
3-2-1-6-1Modem Kit Parts Delivery Inspection of System based on Dell Precision 340
ITE
M
10
6
4
2
Figure 3-16 Modem Kit 066E0796 Parts overview.
Table 3-19 Modem Kit 066E0796 Parts Table
NamePart NumberQty.
1
5
7
8
9
3
1
2
3
4
5
6
7
8
9
10
Read Me First DocumentPart of GE Part Number 2245794 1
ModemGE Part Number 2245794 1
Telephone cablePart of GE Part Number 2245794 1
Telephone WirePart of GE PartNumber 22457941
AC/DC Power convertorPart of GE Part Number 22457941
Power CablePart of GE Part Number 2245794 1
Plastic bag with 4 Stand-OffsPart of GE Part Number 22457941
Velcro-fastenersPart of GE Part Number 2245794 1
Installation manualPart of GE Part Number 22457941
Telephone jack adaptersPart of GE Part Number 22457941
3-20Section 3-2 - Receiving and Unpacking the Equipment
Page 87
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-2-1-6-2Modem Parts Delivery Inspection of System based on Dell Precision 360, Dell Precision 370, Dell
Precision 380 and Dell Precision 390
1
2
3
5
ITE
M
1
2
3
4
5
6
4
6
3
Figure 3-17 Modem Parts overview.
Table 3-20 Modem Parts Table
NamePart NumberQty.
Installation CDPart of GE Part Number 24133381
ModemGE Part Number 24133381
Telephone cablePart of GE Part Number 24133381
Quick Start GuidePart of GE Part Number 2413338 1
Telephone WirePart of GE PartNumber 24133381
Plastic bag with 4 Stand-OffsPart of GE Part Number 24133381
Chapter 3 Installation3-21
Page 88
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
Section 3-3
Preparing for Installation
3-3-1Physical Inspection
3-3-1-1LOGIQworks Voltage Settings
•Verify that the LOGIQworks has the correct voltage delivery.
Switch 110/220V
Figure 3-18 Correct Voltage Settings
Table 3-21 International Voltage and Frequencies settings
CountryVHz
Argentina22050
Australia23050
Austria23050
Bahamas12060
Belgium23050
Brazil110/22060
Canada12060
China, People’s Republic of22050
Croatia23050
Czech Republic23050
Denmark23050
Egypt22050
Finland23050
3-22Section 3-3 - Preparing for Installation
Page 89
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Table 3-21 International Voltage and Frequencies settings
CountryVHz
France23050
Germany23050
Great Britain23050
Greece22050
Hong Kong22050
Hungary23050
Iceland22050
India24050
Israel23050
Italy23050
Japan10050/60
Korea, South22060
Luxembourg22050
Mexico12760
Monaco127/22050
Netherlands23050
New Zealand23050
Norway23050
Poland23050
Portugal22050
Russian Federation22050
Singapore23050
Slovak Republic23050
South Africa220/23050
Spain23050
Sweden23050
Switzerland23050
Taiwan11060
Turkey23050
United States of America12060
Chapter 3 Installation3-23
Page 90
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-3-1-2Power Cables
Table 3-22 Power Cable for DELL PC and DELL Monitor FP1901, Multitech Modem
NamePart Number
Power Cord US2369732
Power Cord French2369733
Power Cord German2369734
Power Cord Italian2369735
Power Cord Spanish2369736
WARNING
WARNING
Connecting The LOGIQworks to a wrong voltage level destroys the LOGIQworks.
Table 3-23 Power Cable for DELL Monitor 2001
NamePart Number
Power Cord US2409670
3-24Section 3-3 - Preparing for Installation
Page 91
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
Section 3-4
Completing the Installation
3-4-1System Specifications
3-4-1-1Physical Dimensions of LOGIQworks
The physical dimensions of the LOGIQworks main parts are summarized in Table 3-25.
Table 3-27 lists the weight of LOGIQworks, with monitor and without on-board peripherals.
Table 3-24 Physical Dimension Table of System based on Dell Precision 340
Unit DescriptionHeightWidthDepthUnit
DellPrecision 340
Sony GDM F520
NEC LCD 1880
Modem
42.518.144.7cm
17.07.117.5.Inches
49.949.748.7cm
19.019.419.0Inches
39.438.718.0cm
15.015.17.0Inches
10.92.514.5cm
4.31.05.7Inches
Table 3-25 Physical Dimensions table of System based on Dell Precision 360, Dell Precision
370, Dell Precision 380 or on Dell Precision 390
Unit DescriptionHeightWidthDepthUnit
Dell Precision 360
42.518.144.7cm
Dell Precision 370
17.07.117.5Inches
Dell Precision 380
44.516.545.7cm
Dell Precision 390
17.56.518Inches
DELL LCD 1901
Modem
DELL LCD 2001
39.438.718.0cm
15.015.17.0Inches
10.92.514.5cm
4.31.05.7Inches
39.438.718.0cm
15.015.17.0Inches
Chapter 3 Installation3-25
Page 92
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-4-1-2Required Space for LOGIQworks Installation
min. 100
70 cm
min. 80cm
recommended 100 cm
Figure 3-19 Required Space for System with 1 CRT Monitor
CRT 1
Keyboard
Mouse
min. 140cm
120 cm
min. 80cm
recommended 100 cm
CRT 1
Keyboard
CRT 2
Mouse
Figure 3-20 Required Space for System with 2 CRT Monitors
3-26Section 3-4 - Completing the Installation
Page 93
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-4-1-2Required Space for LOGIQworks Installation (cont’d)
min. 80
60 cm
min. 80cm
recommended 100 cm
Figure 3-21 Required Space for System with 1 LCD Monitor
LCD 1
Keyboard
Mouse
min. 120 cm
100 cm
min. 80cm
recommended 100 cm
Figure 3-22 Required Space for System with 2 LCD Monitors
LCD 1
Keyboard
LCD 2
Mouse
Chapter 3 Installation3-27
Page 94
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-4-1-3Weight
3-4-1-3-1Weight of System based on Dell Precision 340
Table 3-26 Weight table of System based on Dell Precision 340
ModelWeight [kg]Weight [lbs]
Dell Precision 340
LOGIQworksw/CRT Monitor
LOGIQworksw/LCD Monitor
Modem
Table 3-27 Weight table of System based on Dell Precision 360, Dell Precision 370, on Dell
Precision 380 or on Dell Precision 390
ModelWeight [kg]Weight [lbs]
Dell Precision 360
Dell Precision 370
Dell Precision 380
Dell Precision 390
LOGIQworks/LCD Monitor
Modem
3-4-1-3-2Transport of PC and Monitor
The weight of PC and Monitor exceeds 25lbs. To prevent injuries form carrying PC and Monitor, cut
the delivered box (See “3-2-1-1 Check of Delivered Box (cont’d )” on page 4.) and move PC and
Monitor to the installation location.
9.921.83
30.567.24
8.218.08
0.20.5
9.921.83
17.739
8.218.08
0.20.5
3-4-1-4Acoustic Noise Output
max. 57dB(A)
3-4-2Electrical Specifications
To see the Electrical Specifications for the LOGIQworks see Table 2-14 on page 2-7.
Power Consumption nominal 550 VA including all options.
.
WARNING
WARNING
3-28Section 3-4 - Completing the Installation
Modification of voltage setting only by an authorized service person! Output power: 550VA per mains
outlet, max. power of all connected accessorie s mu st no t exc eed 550 VA.
Page 95
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-4-3Installing LOGIQworks at the Site
3-4-3-1Place the Parts ergonomically correct.
mounted on the floor
Tower PC
Figure 3-23 Work Station w/ Monitor, Keyboard, PC, Mouse
1.) Place the Monitor, Keyboard, Mouse, Peripherals, Desk and stool ergonomically cor rect and place
the Tower Computer either on the floor below the desk or on the desk top. Some prefer the
computer topside while others prefer it down below.
Chapter 3 Installation3-29
Page 96
GE MEDICAL SYSTEMS
DIRECTION 5184024-100, REVISION 5LOGIQWORKS BASIC SERVICE MANUAL
3-4-3-2Single LCD Monitor Installation of System based on Dell Precision 340
Items needed:
4
1
5
3
2
10
6
7
Figure 3-24 Items needed for Single LCD Monitor Installation of System based on Dell Precision 340
9
8
3-30Section 3-4 - Completing the Installation
Page 97
GE MEDICAL SYSTEMS
D
IRECTION 5184024-100, REVISION 5LOGIQWORKS® BASIC SERVICE MANUAL
3-4-3-2Single LCD Monitor Installation of System based on Dell Precision 340 (cont’d)
Table 3-28 Items needed for Single LCD Monitor Installation