NOTE: The information in this manual only applies to CIC Pro center software versions 5.0 or later. It does not
apply to earlier software versions. Due to continuing product innovation, specifications in this manual are
subject to change without notice.
NOTE: For technical documentation purposes, the abbreviation GE is used for the legal entity name, GE
Medical Systems Information Technologies.
Listed below are GE Medical Systems Information Technologies trademarks used in this document. All other trademarks
contained herein are the property of their respective owners.
APEX, Aware, CD TELEMETRY, Quantitative Sentinel, SOLAR, and MUSE are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
APEXPRO, CD TELEMETRY
®
-LAN, CENTRALSCOPE, CIC PRO, OCTACOMM, and UNITY NETWORK are
trademarks of GE Medical Systems Information Technologies.
T-2CIC ProTM Clinical Information Center2026419-002D
viiiCIC ProTM Clinical Information Center2026419-002D
1Introduction
2026419-002DCIC ProTM Clinical Information Center1-1
Equipment information
License agreement
It is important that you carefully read the terms and conditions of this license
agreement before commencing the use of the clinical information center workstation
(the “workstation”) and the clinical information center program recorded therein and
any accompanying user documentation (“program”). This license represents the
entire license agreement concerning the program between you and GE and
supersedes all other communications or advertising related to the program except
any terms and conditions of sale or warranties or warranty limitations relative to the
program and/or the workstation as may be embodied in any documentation supplied
with the workstation. By commencing the use of the workstation and the program
contained therein, you are accepting and agreeing to be bound by all the terms and
conditions of this license agreement. If you are not willing to be bound by the terms
and conditions of this license agreement, you should promptly return the
workstation to GE and you will receive a refund of the purchase price.
I.Grant
The Program is capable of coupling one to sixteen patient monitoring units to
the Workstation. GE hereby grants you a non-exclusive, non-transferable right
and license to use the Program for coupling the number of patient monitoring
units to the Workstation for which a per-unit royalty has been paid pursuant to
Article II hereof.
II.Royalty
You have paid GE a one-time, per-unit royalty equal to GE Medical Systems
Information Technologies’ current published price for the use of the Program.
The per-unit royalty is based on the actual number of patient monitors intended
to be coupled by the Program to the Workstation as stated in the Purchase Order
for the Workstation and the Program. If you use the Program to couple any
patient monitoring units to the Workstation in addition to the number for which
a per unit royalty was previously paid, you agree to pay GE an additional perunit royalty equal to GE Medical Systems Information Technologies ' then
current published royalty for the Program for each such additional patient
monitoring unit so coupled. The additional per-unit royalty shall be paid to GE
within 30 days of the use of the Program to couple any such additional patient
monitoring units to the Workstation.
1-2CIC ProTM Clinical Information Center2026419-002D
Introduction: Equipment information
III. Limitations
You hereby agree not to: (1) use the Program in any network or system other
than to couple patient monitoring units to the Workstation; (2) make any copy
of the Program for any reason, or allow or assist others to do so; (3) modify,
reverse engineer, de compile or disassemble the Program or merge any part of
the Program into any other program; (4) rent, sell, sublease, assign, transfer or
otherwise share the Program or any of your rights in the Program under thi s
Agreement with any third party; or, (5) remove or alter any copyright notice,
labels or trademarks from the Program or the Workstation.
IV. Title
This License is not a sale. Title and all copyrights to the Program and any copy
made by you remains the sole property of GE.
V.Term
This Agreement shall continue in force until terminated. This Agreement shall
terminate automatically when you cease using the Workstation and the Program
for their intended purpose. GE may terminate this Agreement on 30 days
written notice if you make any unauthorized copies of the Program or fail to
comply with any of the restrictions on use of the Program as set forth herein.
VI. Limited warranty, disclaimer and limitation of liability
A. Licensor warrants that on the acceptance date the Program shall be free from
significant programming errors and shall operate and conform to the
published functional specifications applicable thereto, and that the Program
shall conform to the standards generally observed in the industry for similar
software.
B. This warranty shall be invalidated by your modification of the Program if
such modification or the interaction between such modification and the
Program as supplied by GE is the cause of the defect, error or nonconformity.
C. Except as stated above, the warranty covering the Program and the
W orkstation shall be either GE S tandard Warranty or Limited Extended Parts
Warranty as published by GE and hereby made a part hereof.
D. Except for the express warranties stated herein, GE disclaims all warranties
with regard to the program including implied warranties of merchantability
or fitness for a particular purpose.
2026419-002DCIC ProTM Clinical Information Center1-3
Introduction: Equipment information
E. GE Medical Systems Information Technologies’ entire liability to you
arising out of or in connection with this Agreement shall not exceed the perunit royalty paid to GE for use of the Program. You acknowledge that the
amount paid to GE for use of the Program is insufficient for GE to undertake
any greater risk. In no event shall GE be liable for any indirect, incidental,
consequential, special or exemplary damages (including without limitation,
lost profits, business interruption, loss of business information, personal
injury or any other pecuniary loss) arising from the use of the program, even
if GE has been advised of the possibility of such damages.
VII. Governing law
This Agreement shall be governed by the laws of the State of Wisconsin.
VIII.Partial invalidity
If any provision of this Agreement is held invalid or unenforceable, the
remaining portions of the Agreement shall continue in full force and effect.
1-4CIC ProTM Clinical Information Center2026419-002D
Introduction: Safety information
Safety information
Responsibility of the manufacturer
GE is responsible for the effects of safety, reliability, and performance only if:
Assembly operations, extensions, readjustments, modifications, or repairs are
carried out by persons authorized by GE;
The electrical installation of the relevant room complies with the requirements
of the appropriate regulations.
The equipment is used in accordance with the instructions for use.
General
This device is intended for use under the direct supervision of a licensed health care
practitioner.
This device is not intended for home use.
Federal law restricts this device to be sold by or on the order of a physician.
Hazard definitions
Contact GE for information before connecting any devices to the equipment that are
not recommended in this manual.
Parts and accessories used must meet the requirements of the applicable IEC 60601
series safety standards, and or the system configuration must meet the requirements
of the IEC 60601-1-1 medical electrical systems standard.
Parts and accessories used must meet all local building and safety requirements.
Periodically, and whenever the integrity of the device is in doubt, test all functions.
The use of accessory equipment not complying with the equivalent safety
requirements of the device may lead to a reduced level of safety of the resulting
system. Consideration relating to the choice shall include:
use of the accessory in the patient vicinity; and
evidence that the safety certification of the accessory has been performed in
accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized
national standards.
If the installation of the equipment, in the USA, will use 240V rather than 120V, the
source must be a center-tapped, 240V, single-phase circuit.
The terms danger, warning, and caution are used throughout this manual to point out
hazards and to designate a degree or level of seriousness.
A hazard is defined as a source of potential injury to a person.
DANGER
Indicates an imminent hazard which, if not avoided, will result in
2026419-002DCIC ProTM Clinical Information Center1-5
Introduction: Safety information
death or serious injury.
WARNING
Indicates a potential hazard or unsafe practice which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice which, if not
avoided, could result in minor personal injury or product/property
damage.
NOTE
Provides application tips or other useful information.
CIC Pro center hazards
WARNING
BEFORE USE — Before putting the system into operation visually
inspect all connecting cables for signs of damage. Damaged cables
and connectors must be replaced immediately.
Before using the system, the operator must verify that it is in
correct working order and operating condition.
Periodically, and whenever the integrity of the product is in doubt,
test all functions.
WARNING
POWER SUPPLY — The device must be connected to a properly
installed power outlet with protective earth contacts only. If the
installation does not provide for a protective earth conductor,
disconnect the monitor from the power line and operate it on
battery power, if possible.
GE recommends the use of an Uninterrupted Power Supply (UPS)
with the CIC Pro center. If a UPS is not used, improper shutdowns
of the system could result in the event of a power outage and
cause a lengthy disk scan procedure when the unit reboots. You
could also lose data in the event of a power outage if you do not
use a UPS.
All devices of a system must be connected to the same power
supply circuit. Devices which are not connected to the same
circuit must be electrically isolated when operated.
1-6CIC ProTM Clinical Information Center2026419-002D
Introduction: Safety information
WARNING
DISCONNECTION FROM MAINS — When disconnecting the
system from the power line, remove the plug from the wall outlet
first. Then you may disconnect the power cord from the device. If
you do not observe this sequence, there is a risk of coming into
contact with line voltage by inserting metal objects, such as the
pins of leadwires, into the sockets of the power cord by mistake.
WARNING
NETWORK INTEGRITY — The CIC Pro center resides on the
hospital’s computer network, and it is possible that inadvertent or
malicious network activity could adversely affect patient
monitoring. The integrity of the computer network is the
responsibility of the hospital.
WARNING
INTERFACING OTHER EQUIPMENT — Devices may only be
interconnected with each other or to parts of the system when it
has been determined by qualified biomedical engineering
personnel that there is no danger to the patient, the operator, or the
environment as a result. In those instances where there is any
element of doubt concerning the safety of connected devices, the
user must contact the manufacturers concerned (or other informed
experts) for proper use. In all cases, safe and proper operation
should be verified with the applicable manufacturer's instructions
for use, and system standards IEC 60601-1-1/EN 60601-1-1 must
be complied with.
WARNING
EXPLOSION HAZARD — Do NOT use this equipment in the
presence of flammable anesthetics, vapors or liquids.
WARNING
DISCONNECTION FROM MAINS — When disconnecting the
system from the power line, remove the plug from the wall outlet
first. Then you may disconnect the power cord from the device. If
you do not observe this sequence, there is a risk of coming into
contact with line voltage by inserting metal objects, such as the
pins of leadwires, into the sockets of the power cord by mistake.
WARNING
ACCIDENTAL SPILLS — To avoid electric shock or device
malfunction, liquids must not be allowed to enter the device. If
liquids have entered a device, take it out of service and have it
2026419-002DCIC ProTM Clinical Information Center1-7
Introduction: Safety information
checked by a service technician before it is used again.
WARNING
ACCURACY — If the accuracy of any value displayed on the
screen or printed on a graph strip is questionable, first determine
the patient's vital signs by alternative means. Then, verify the CIC
Pro center and printer are working correctly.
WARNING
ALARMS — Do not rely exclusively on the audible alarm system
for patient monitoring. Adjustment of alarm volum e to a low level
or off during patient monitoring may result in a hazard to the
patient. Remember that the most reliable method of patient
monitoring combines close personal surveillance with correct
operation of monitoring equipment.
After connecting the monitor to the central station and/or nursealert system, verify the function of the alarm system. Repeat this
verification periodically, including a check of all connected
speakers.
CIC Pro center audible alarms will not sound for patients with
bedside monitoring devices configured to “Operating Room”
mode.
WARNING
DISPOSAL — Dispose of the packaging material, observing the
applicable waste control regulations and keeping it out of
children’s reach.
WARNING
LOSS OF MONITORING — If the monitoring at the CIC Pro
center is temporarily interrupted, alternate monitoring devices or
close observation of the patients must be used until the monitoring
function at the CIC Pro center is restored.
Indications of a loss of the monitoring function at the CIC Pro
center are as follows.
RED SCREEN indicates the CIC Pro center application is
restarting itself and patient monitoring at the CIC Pro center
is NOT occurring. The monitoring function at the CIC Pro
center will automatically resume in less than 30 seconds. No
user action is required.
BLUE SCREEN indicates the Windows
®
operating system
has a functional error and patient monitoring at the CIC Pro
1-8CIC ProTM Clinical Information Center2026419-002D
Introduction: Safety information
center is not occurring. If the CIC Pro center does not
automatically restart after 90 seconds, the monitoring
function at the CIC Pro center will not resume until you turn
off the power to the CIC Pro center and then turn the power
back on. The monitoring function should resume in
approximately 90 seconds
Once the monitoring function at the CIC Pro center has been
restored, you should verify the correct monitoring state and alarm
function.
WARNINGS
CIC Pro center V5.0.x is in-unit compatible with CIC Pro center
V4.0.x and V4.1.x. The CIC Pro center is not in-unit compatible
with CIC Pro center V3.1 or earlier and is not in-unit compatible
with any versions of Centralscope. Sharing of the same care unit
name across CIC Pro centers having incompatible software
versions can result in lost or corrupted telemetry alarm defaults
data and loss of audible alarms.
Both hardwire and telemetry beds are limited in the number of
remote view connections that can be supported.
Do not exceed a maximum of 15 CIC Pro centers in a single
logical care unit.
Attempting simultaneous displays of a patient monitor
(bedside or telemetry) at too many CIC Pro centers may cause
lost or intermittent communication between CIC Pro centers
and the patient monitor. This is evidenced by NO COMM or
intermittent communication conditions for the beds.
The maximum CIC Pro centers viewing a patient monitor can vary
depending on patient monitor capabilities and network design.
Do not load any software other than that specified by GE onto the
CIC Pro. Installation of software not specified by GE may cause
damage to the CIC Pro center or loss or corruption of data.
CAUTION
NEGLIGENCE — GE does not assume r esponsibil ity for damage
to the equipment caused by improperly vented cabinets, improper
or faulty power, or insufficient wall strength to support equipment
mounted on such walls.
CAUTION
MPSO — Do not use a multiple portable socket outlet (MPSO) for
a system because it could result in unacceptable enclosure leakage
currents.
2026419-002DCIC ProTM Clinical Information Center1-9
Introduction: Safety information
CAUTION
SECURITY — The web browser which runs in conjunction with
the CIC Pro center is intended for hospital intranet use only. If
confidential patient information is made available from the
hospital intranet, the security of the data is the responsibility of the
hospital.
CAUTION
EMC — Magnetic and electrical fields are capable of interfering
with the proper performance of the device. For this reason make
sure that all external devices operated in the vicinity of the
monitor comply with the relevant EMC requirements. X-ray
equipment or MRI devices are a possible source of interference as
they may emit higher levels of electromagnetic radiation.
1-10CIC ProTM Clinical Information Center2026419-002D
Equipment symbols
The following symbols appear on the equipment, but some symbols may not appear
on all equipment.
SymbolDescription
Introduction: Equipment symbols
Equipment Symbols
ATTENTION: Consult accompanying documents before using the
equipment.
Provide electrostatic discharge damage protection.
USB connector port
Ethernet connector port used to connect to the Unity Network MC network.
Ethernet connector port used to connect to the Unity Network IX network.
External speaker connector port
Primary video output connector port
Secondary video output connector port
Primary serial communication connector port
Secondary serial communication connector port
Power switch indicator. Indicates the power switch is in the on position
when filled
Power Off: Indicates the power switch is in the OFF position.
2026419-002DCIC ProTM Clinical Information Center1-11
Introduction: Equipment symbols
Equipment Symbols
SymbolDescription
Equipotential stud. A ground wire from another device can be tied here to
ensure the devices share a common reference point.
Type B applied part: Non-isolated applied part suitable for intentional
external and internal application to the patient excluding direct cardiac
application.
[Medical Standard Definition:] Applied part complying with the specified
requirements of IEC 60601-1/UL 60601-1/CSA 601.1 Medical Standards to
provide protection against electric shock, particularly regarding allowable
leakage current.
Type BF applied part: Isolated (floating) applied part suitable for intentional
external and internal application to the patient excluding direct cardiac
application. “Paddles” outside the box indicate the applied part is
defibrillator proof.
[Medical Standard Definition:] F-type applied part (floating/isolated)
complying with the specified requirements of IEC 60601-1/UL 60601-1/
CSA 601.1 Medical Standards to provide a higher degree of protection
against electric shock than that provided by type B applied parts.
NOTE
The rating of protection against electric shock (indicated by symbol for
CF or BF) is achieved only when used with patient applied parts
recommended by GE Medical Systems Information Technologies.
Type CF applied part: Isolated (floating) applied part suitable for intentio nal
external and internal application to the patient including direct cardiac
application. “Paddles” outside the box indicate the applied part is
defibrillator proof.
[Medical Standard Definition:] F-type applied part (floating/isolated)
complying with the specified requirements of IEC 60601-1/UL 60601-1/
CSA 601.1 Medical Standards to provide a higher degree of protection
against electric shock than that provided by type BF applied parts.
Fuse. Replace the fuse with a fuse of the same type and rating.
Power
Writer door button
Silence Alarm keyboard key
1-12CIC ProTM Clinical Information Center2026419-002D
Introduction: Service requirements
Equipment Symbols
SymbolDescription
Medical Equipment. With respect to electric shock, fire and mechanical
hazards only in accordance with UL 60601-1 and CAN/CSA C22.2
NO.601.1.
4P41
This symbol indicates that the waste of electrical and electronic equipment
must not be disposed as unsorted municipal waste and must be collected
separately. Please contact the manufacturer or other authorized disposal
company to decommission your equipment.
This symbol indicates the date of manufacture of this device. The first four
digits identify the year and the last two digits identify the month.
2005-08
Service requirements
Follow the service requirements listed below, and in the “Maintenance” chapter of
this manual.
Refer equipment servicing to GE authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids that
warranty.
It is the user’s responsibility to report the need for service to GE or to one of
their authorized agents.
Failure on the part of the responsible individual, hospi tal, or inst itution using
this equipment to implement a satisfactory maintenance schedule may cause
undue equipment failure and possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure that the
equipment is always functional when required.
2026419-002DCIC ProTM Clinical Information Center1-13
Introduction: Equipment identification
Equipment identification
Every GE device has a unique serial number for identification. A sample of the
information found on a serial number label is shown below.
### ## ## #### # #
123456
Description
1
Product code
2Year manufactured
3Fiscal week manufactured
4Production sequence number
5Manufacturing site
6Miscellaneous characteristic
1. The product code for the CIC Pro center is SCH.
1
646A
1-14CIC ProTM Clinical Information Center2026419-002D
Manual information
Manual purpose
This manual supplies technical information for service representatives and technical
personnel so they can maintain the equipment to the assembly level. Use it as a
guide for maintenance and electrical repairs considered field repairable. Where
necessary, the manual identifies additional sources of relevant information and
technical assistance.
See the operator’s manual for the instructions necessary to operate the equipment
safely in accordance with its function and intended use.
Intended audience
This manual is intended for use by service representatives and technical personnel
who maintain, troubleshoot, or repair the equipment.
Conventions used
Introduction: Manual information
Revision history
Bold textIndicates keys on the keyboard, text to be entered, or hardware items
such as buttons or switches on the equipment.
Bold Italicized
text
Ctrl+EscIndicates a keyboard operation. A plus (+) sign between the names of
<Space>Indicates you must press the spacebar. When instructions are given
EnterIndicates you must press the Enter or Return key on the keyboard.
The document part number and revision letter are listed at the bottom of each page
in this manual. The revision letter identifies the document’s update level. The
revision history of this document is summarized below.
Indicates software terms that identify menu items, buttons, or options
in various windows.
two keys indicates that you must press and hold the first key while
pressing the second key once.
For example, “Press Ctrl+Esc” means to press and hold down the
Ctrl key while pressing the Esc key.
for typing a precise text string with one or more spaces, the point
where the spacebar must be pressed is indicated as <Space>.
Do not type “enter”.
Revision History
RevisionComment
AInitial release of the document item number to develop the Bills of Material.
BInitial release of document content for limited customer use.
2026419-002DCIC ProTM Clinical Information Center1-15
Introduction: Manual information
Revision History
RevisionComment
CInitial release of document for all customers world wide.
DRelease of document with revisions to meet additional service requirements.
1-16CIC ProTM Clinical Information Center2026419-002D
2Equipment overview
2026419-002DCIC ProTM Clinical Information Center2-1
1Processor boxRun the CIC Pro center application.
2Primary displayDisplay real-time and stored patient data,
control windows, and various system-level
operations. Up to two displays may be
connected to the CIC Pro center
simultaneously.
3Secondary display (optional) Display stored patient data and browser
information.
4Standard mouse and keyboardEnter data, navigate menus, and choose
options.
3External speakers Sound audible patient status and system
status alarm tones.
2-2CIC ProTM Clinical Information Center2026419-002D
Equipment overview: Standard components
Primary display
The CIC Pro center supports either a standard 19-inch or 20-inch color display or a
standard 19-inch or 20-inch color touchscreen display. See “Optional components”
on page 8-9.
Processor box
System interconnection diagram
Laser printer
(Optional)
MC Unity
Network
IX Unity
Network
Speaker
Speaker
Mouse
Keyboard
Touchscreen
USB cable
USB to parallel
printer adapter
Network
Network
cable
cable
USB
PORT
USB
PORT
USB
PORT
USB
PORT
PORT
ETHERNET
PORT
ETHERNET
PORT
SPEAKER/AUDIO
CIC Pro
Computer
COM 1
SERIAL PORT
COM 2
DDW cable
SERIAL PORT
DVI
PORT
Monitor cable
DVI
PORT
Monitor cable
Touchscreen serial cable
(Optional)
PRN-50, digital
writer (Optional)
Primary display
Secondary display
(Optional)
(Optional)
Uninterruptable
power supply
(Recommended)
695A
2026419-002DCIC ProTM Clinical Information Center2-3
Back panel
Equipment overview: Standard components
The back panel of the processor box has the following connectors, ports, receptacle,
and switch.
1
129
2
10
3
4115
8
7
6
Back panel of the processor box
ItemFunction
1Equipotential studConnect a ground wire from another device
to ensure the devices share a common
reference point.
2Ventilation portsVent internal processor heat to the outside of
the processor box.
3Unity Network MC Ethernet port
Interface with other networked GE patient
monitoring and telemetry system devices.
Display waveform, parameter, and alarm
condition data from other networked
devices.
031A
4Unity Network IX Ethernet port
Connect to an optional network laser
printer.
Share licenses with other CIC Pro centers
within the same care unit.
Display full disclosure data.
Access remote serviceability.
Provide access to data and servers
outside of your facility.
5COM 1 portConnect to the touchscreen display.
6External speaker portConnect to external speakers to hear patient
and system status alarm notification.
7COM 2 portConnect to the PRN 50-M digital writer.
2-4CIC ProTM Clinical Information Center2026419-002D
Loading...
+ 260 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.