FCR PRIMA T2 (CR-IR 392)
Quick Guide
X-ray Exposure
Registration of patient information
Select Exposure menu
X-ray exposure and image conrmation
Leave 10 cm (4 in.) or more space at both sides
of the Image Reader to allow the cooling fans
inside to intake and exhaust air.
10 cm (4 in.)
or more
10 cm (4 in.)
or more
1
1
Power ON
The monitor lamp changes from
orange to blue.
2
Power ON
After the Console (PC) has
booted, an application starts.
1
Click
When is clicked, the
conrmation window is shown.
2
Click
Conrm that “Shutdown the PC” is se-
lected and click [OK].
3
Power ON
Equipment initialization starts.
3
Check
When the Console (PC) shuts
down, the power supply is
switched off automatically.
To switch off the power
independently, press the
(Stand-by Switch) for
approximately 4 seconds.
4
Power OFF
Turn off the Main Power Switch.
1
Select
Select the Exposure
menu.
The Read Image
window is displayed
automatically.
1
Register
Register Patient ID,
Patient Name, Sex,
and Date of Birth.
Click
when
the input of patient
information has been
completed.
1
X-ray Exposure
2
Image Reading
• Slowly insert the cassette straight along
the guide at the right.
• Before inserting the
cassette, remove
the exposure marker
from the cassette.
3
Check
Start the Image Viewer
window and conrm
the image.
4
Click
When this is clicked,
the display returns to
the Main window.
When a printer is
connected, the image
is put out to lm.

897N102548
Frequent messages and countermeasures
is displayed. / Images are not transferred.
Not transferred image data remain in the
Image Reader. Do not switch off the power.
►Cause
- The read image data could not be transferred to the
Console (PC).
- The power supply of the Console (PC) is switched off.
- No Exposure menu has been registered.
The cassette has not been inserted correctly.
►Cause
The cassette has not been inserted correctly into the
Image Reader.
Do not pull out the cassette forcibly.
A part of the device may become broken.
If the IP protrudes out of the cassette pulled
out, return the IP into the cassette and close
the lid of the cassette.
Image reading is not completed.
►Cause
A cassette has been ejected without reading the images
exposed to IP.
Images remain on the IP.
Repeat image reading.
2
1
Stop the buzzer.
Press the
(Stand-by Switch) to stop the buzzer.
2
Conrm the Console.
- Conrm correct operation of the Console (PC).
- When the main power of the Console (PC) is off, turn
on the main power to start up the Console (PC).
- Conrm that an Exposure menu has been registered.
3
Conrm the cable connection.
- Conrm that the network cable connecting the Image
Reader and the HUB has not been disconnected.
- Conrm that the network cable connecting the Console
(PC) and the HUB has not been disconnected.
- Sometimes the error may be corrected by reconnecting
the network cable.
- Conrm that the power cord for the HUB has not been
disconnected.
1
Stop the buzzer.
Press
(Stand-by Switch) to stop the buzzer. Press
(Stand-by Switch) again to clear the error indication.
2
Remove the cassette slowly.
Conrm that the
(Cassette Removal Lamp) and the
(Cassette Ready Lamp) are blinking alternately,
press the
(Stand-by Switch) again and then remove
the cassette.
3
Align the cassette to the right and insert it
slowly.
Conrm that the cassette is not upside down or switched
front/back.
Insert the cassette following the operation instruction
attached to the Image Reader.
1
Stop the buzzer.
Press
(Stand-by Switch) to stop the buzzer. Press
(Stand-by Switch) again to clear the error indication.
2
Remove the cassette.
Conrm that the
(Cassette Removal Lamp) and the
(Cassette Ready Lamp) are blinking alternately,
press the
(Stand-by Switch) again and then remove
the cassette.
3
Perform image reading.
Insert the removed cassette into the unit and perform
image reading again.