Test Leads: 91.4 cm (36 in.) safety set with finger
guards on probes and shrouded banana plugs
* Specifications subject to change without notice.
Fuse: 5 x 20 mm type F2AH250V fuse
Environmental:
Operating Temperature: 5°C to 35°C
Storage Temperature: -20°C to 50°C
Humidity: 0 to 80%
Dimensions and Weight: 15 cm wide x 5.0 cm deep
x 8.75 cm high (6 in x 2 in x 3-1/2 in);
0.35 kg (0.78 lb)
1-1
Page 6
35035
Operators Manual
1.2 Introduction
The following safety precautions should be observed before and while operating the Model 35035. They
must be observed in conjunction with those outlined in the manual for the x-ray generator to which the
instrument is connected.
Circuit protection provisions are built into the Model 35035. However, they primarily protect the meter
circuit itself. The x-ray generator must therefore contain voltage and current limiting components within
its circuitry for its own protection as well as for the safety of the technician or operator making the
connections to and performing the tests on the generator.
1.3 Safety Precautions
This instrument is intended for use by qualified personnel who recognize shock hazards and are familiar
with the safety precautions required to avoid possible injury. Read this manual carefully before operating
the Model 35035 mA/mAs Meter.
Exercise extreme caution when a shock hazard is present at the instrument’s input. The American
National Standards institute (ANSI) states that a shock hazard exists when voltage levels greater than 30
V RMS or 42.4 V Peak are present. A good safety practice is to expect that a hazardous voltage is
present in any unknown circuit before measuring.
Do not exceed 500 V peak between input low and earth ground.
Inspect the test leads for possible wear, cracks, or breaks before each use. If any defects are found,
replace the test leads with ones that have a reputable measure of safety. For optimum safety, do not
touch the test leads or the instrument while power is applied to the circuit under test. Turn the x-ray
generator power off before connecting or disconnecting the instrument.
Do not exceed the instrument’s maximum allowable input as defined in the Sections 1 and 3.
The symbol on the instrument indicates that a potential shock hazard may exist during use.
Always expect a shock hazard at the input terminals when the unit is connected to the x-ray generator
and the generator is powered.
Do not make any contact with the unit during exposures. The voltage potential at the input could exceed
the insulating capacity of the case and test leads under fault conditions.
1-2
Page 7
Theory of Operation
G
Introduction
Section 2
Theory Of Operation
2.1 Introduction
The purpose of this manual is to outline the operation and maintenance of the instrument. IT IS THE
RESPONSIBILITY OF THE USER TO DETERMINE THE SUITABILITY OF THIS PRODUCT FOR HIS
PARTICULAR MEASUREMENT CONDITIONS. The entire manual should be read before using the
meter.
2
Failure to observe all of the safety precautions
presented in Sections 1.2 and 1.3 of this manual as
well as those provided by the x-ray generator
manufacturer may result in serious injury.
The Inovision Model 35035 mA/mAs Meter has two functions: an mA function with three (3) ranges that
measures current in mA, and an mAs function with two (2) ranges that connects an integrator to read the
product of current and time in mA-seconds. Both range and function are selected using rotary switches.
Either DC or AC input mode is selected using a two-position rocker switch. The mAs function is reset by
means of a momentary push-button switch.
Operation of the meter requires a standard 9 V akaline battery. A low-battery warning appears in the
display when the battery is nearly depleted (7.2 V). Polarity is displayed as well. Blanking of display
digits indicates overrange.
A schematic of the input circuit of the Model 35035 is shown on the following page. R1, R2, and R3 are
current shunt resistors for range scaling and changing the input current flow into a voltage. F1, R15, D2,
and D3 are input protection devices that protect the shunt resistors and input circuit electronics from
excessive current and voltage. F1 is a 2 A, 250 V fuse that limits the input current. R15 is a 140 V
varistor that clamps high voltage pulses. D2 and D3 are 24 V zener diodes that clamp lower voltages.
WARNIN
2-1
Page 8
35035
Operators Manual
HI
INPUT
LO
R15
D2
D3
V130
LA
10A
IN
5359A
IN
5359A
F1
2A
FUNCTION
AND MODE
SWITCHING
AC/DC MODE
SWITCHING AND
RECTIFIER
TO METER ELECTRONICS
R1R2
FA2526-
1.003
FA2526-
9.026
R3
FA2526-
90.262
Figure 2-1. Input Circuit Schematic
2.2 Instrument Features
Figure 2-2 represents a front panel of the Model 35035 mA/mAs Meter and shows the location of its
display, input jacks, fuse, switches, and trimmer adjustment controls. The functions of these features are
described below.
DISPLAY
RESET
DC
RESET
SWITCH
AC
MODE SWITCH
FUNCTION SWITCH
RANGE SWITCH
mA/mAs METER Model 35035
2000
200
20 ( mA)
OFF
mA
mAs
WARNING! High voltage may be present during use.
DISPLAY ZERO ADJUST
mA GAIN ADJUST
mAs GAIN ADJUSTMENT
HI
LO
500V
MAX
F 2A H
250V
INPUT
JACKS
FUSE
ASSEMBLY
Figure 2-2. Model 35035 mA/mAs Meter Front Panel
2-2
Page 9
Theory of Operation
Instrument Features
2
INPUT JACKS
Touch-protected 4mm laboratory banana jacks for the use of safety laboratory leads and all spring-loaded
4mm banana plugs.
FUSE ASSEMBLY
Allows field replacement of the input current protection fuse with a small flat-blade screwdriver.
DISPLAY
3-1/2 digit liquid crystal display indicates the measured value of the signal applied to the input in mA or
mAs.
RESET SWITCH
Resets the display to zero after a mAs measurement has been made and the display reading has been
recorded.
MODE SWITCH
Selects either AC or DC measurement mode. The position of this switch depends upon the type of
current signal from the x-ray generator to be measured.
FUNCTION SWITCH
Turns the Model 35035 mA/mAs Meter power ON and selects the desired measurement function (mA or
mAs).
RANGE SWITCH
Selects the full-scale measurement range of the Model 35035 (20, 200, or 2000). All three ranges
operate in the mA function. Only the 200 and 2000 ranges operate in the mAs function.
DISPLAY ZERO ADJUST
Trimmer potentiometer (R33) provides adjustment for a zero display reading with no signal applied to the
input jacks.
MA GAIN ADJUST
Trimmer potentiometer (R10) provides calibration adjustment of the mA function.
MAS GAIN ADJUST
Trimmer potentiometer (R7) provides calibration adjustment of the mAs function.
2.3 Manual Addenda
Any improvements or changes concerning the instrument or manual will be explained in an addenda
included with the manual. Be sure to note these changes and incorporate them into the manual.
2-3
Page 10
35035
Operators Manual
Page 11
Operation
Battery Installation
Section 3
Operation
3.1 Battery Installation
The Model 35035 mA/mAs Meter is shipped from the factory with the battery separate from the unit.
Remove the battery compartment cover on the bottom of the unit to install the battery. Make certain that
the battery’s clip end is inserted at the end of the compartment where the clip wires come out from the
case and that the wires get looped around the top of the battery. This method will aid the user in
replacing the battery when its life has been depleted.
3.2 Operating Instructions
1. Before making any connections to the x-ray generator, MAKE CERTAIN THAT THE
GENERATOR'S POWER IS TURNED OFF. Connect the Model 35035 to the x-ray generator
under test. Ensure that the test leads have no defects and the connections are secure.
WARNING
3
Failure to observe all of the safety precautions
presented in Sections 1.2 and 1.3 of this manual as
well as hose provided by the x-ray generator
manufacturer may result in serious injury.
2. Select either the AC or DC measurement mode on the Model 35035 using the MODE rocker switch
depending upon the type of input signal expected from the x-ray generator. Again, knowledge of
the signal from a specific generator is the responsibility of the technician.
3. Select the desired measurement function (mA or mAs) on the Model 35035 with the FUNCTION
rotary switch. This switch turns the unit’s power ON as well.
4. Select the desired range of the Model 35035 with the RANGE rotary switch. For greatest accuracy
in the mA function, select the range that provides the most resolution on the unit’s display. (Make
certain that the signal input limits specified for that particular range are not exceeded.)
For mAs readings, select either the 200 range or the 2000 range (the 20 range is for mA only) and
when necessary, reset the Model 35035 integrator by depressing the RESET button. Since noise
comparators are built into the Model 35035 mAs circuitry, a minimum current of approximately 3.2
mA on the 200 mAs range or 32 mA on the 2000 mAs range is required before the signal registers
on the display.
3-1
Page 12
35035
G
G
Operators Manual
For greatest accuracy in the mAs function, use as high a current as possible within the proper
signal limits for the selected range. The duration of the current pulse should be correspondingly
short. This is especially important when measuring mAs in AC mode for single-phase x-ray
generators.
5. Turn the x-ray generator power ON and proceed to take measurements with the Model 35035. It is
mandatory to OBSERVE ALL SAFETY PRECAUTIONS contained in Section 1 of this manual.
6. Turn the x-ray generator power OFF after the measurements are complete.
7. Turn the Model 35035 power OFF and disconnect the test leads.
3.3 Signal Input Limits
WARNIN
Exceeding the maximum input overload limits may
result in fuse or other component failure. Such
failure can result in the generation of dangerously
high voltage at the input.
Function Ranges Maximum Input Limit
OFF Input shorted 2 A max (fuse protected)
mA 20 250 mA for 30 sec*
200 1 A for 30 sec*
2000 2 A max (fuse protected)
mAs 20 Input shorted 2 A max (fuse protected)
200 1 A for 30 sec*
2000 2 A max (fuse protected)
*Limit set by power dissipation rating of current range shunt resistors.
WARNIN
A possible electrical shock hazard may occur if either
input terminal is connected to a source of more than
500V DC or RMS AC above earth ground.
3-2
Page 13
Maintenance
General
4
Section 4
Maintenance
4.1 General
The Inovision Model 35035 mA/mAs Meter is designed to perform with an accuracy of 1%. The
performance of the meter should be checked periodically to ensure accurate, reliable operation. Under
normal operating conditions, this instrument will perform to specifications for a period of at least one (1)
year after calibration. The calibration cycle is, therefore, one (1) year. Any major, abrupt changes in
performance indicate that the meter might not be operating properly. The unit should then be returned to
the factory for calibration and repair. There are no user-serviceable parts within the meter. See Section
4.8 for calibration and repair information.
The calibration procedure for a normal operating unit does not require any disassembly of the instrument.
Adjustments are all accessible from the enclosure exterior. They are set using a small flat-blade
screwdriver with an insulated handle or an adjustment tool specifically designed for trimmer potentiometer
adjustment purposes. Refer to Figure 2-2 for the exact location of the trimmer potentiometer access
holes.
4.2 Cleaning the Instrument
To clean the exterior of the instrument, follow these steps:
1. Turn off the instrument and disconnect all input connectors.
2. Moisten a soft cleaning cloth with a solution of three parts water and one part liquid dishwashing
detergent.
Do not soak the cloth in the solution; you must not let the solution drip inside the instrument.
3. Use the moistened cloth to wipe the instrument's exterior cover.
4. Let the instrument dry completely before using.
4.3 Test Equipment
A calibrated DC current source is required to properly calibrate the Model 35035 mA/mAs Meter. Its dwell
time should be adjustable from 100 mS to 10 s and accurate to within 0.5%. Its current output should be
adjustable from 1 mA to 2 A and accurate to within 0.5%. Finally, its compliance (burden) voltage should
be adjustable to 25 volts DC.
In addition, a precision digital multimeter (0.25% accuracy in DC Amps) can be in the circuit to monitor the
output of the current source.
4.4 Performance Tests
Connect the Model 35035 to the current source with appropriate test leads. Place its mode switch in the
DC position. For the mA performance tests, apply the current listed and verify that the Model 35035
display reads within the limits specified. For the mAs performance tests, apply the current for the dwell
4-1
Page 14
35035
Operators Manual
time specified and verify that the Model 35035 display reads within the limits listed. Reversing the polarity
of the current source signal should reverse the polarity indicated on the Model 35035 display without
significantly altering the numeric value for either mA or mAs. On mAs, the reading generated by a 50 mA,
5 s pulse should be effectively nulled out by an equal pulse of reverse polarity. (Note that except when
performing the mAs tracking test, the RESET button must be depressed after each measurement
reading.)
4.4.1 MA Performance Tests
35035 Range Setting Current Source Output 35035 Display Reading
20 5 mA 4.95 - 5.05
20 15 mA 14.85 - 15.15
200 50 mA 49.5 - 50.5
200 150 mA 148.5 - 151.5
2000 500 mA 495 - 505
2000 1500 mA 1485 - 1515
4.4.2 MAS Performance Tests
35035 Range Setting Current Source Output
20 15 mA 1 s 0
200 50 mA 1 s 49.5 - 50.5
200 500 mA .1 s 49.5 - 50.5
200 1500 mA .1 s 148.5 - 151.5
2000 50 mA 5 s 248 - 253
2000 500 mA 1 s 495 - 505
2000 1500 mA 1 s 1485 - 1515
NOTE
Current Source
Dwell Time
The following test will verify the mAs tracking
performance. Do not depress the RESET button
between applied current pulses.
2000 50 mA 5 s DO NOT RESET!
35035 Display
Reading
2000 -50 mA 5 s
4-2
LESS THAN ± 005
Page 15
Maintenance
Instrument Calibration
4
4.5 Instrument Calibration*
A thorough checkout and adjustment of the Inovision Model 35035 mA/mAs Meter should be performed
on a yearly basis*. Three (3) trimmer potentiometers (accessible from the front side of the enclosure) are
the only adjustments necessary to calibrate a properly functioning unit. The following 28 steps outline the
calibration procedure:
NOTE
* It is recommended that the unit be returned to
Fluke Biomedical, Radiation Management Services
for calibration. Removing the tamper-proof seal will
void the Fluke Biomedical, Radiation Management
Services warranty pertaining to the calibration.
1. With a fresh battery installed, remove the tamper-proof seal covering the three (3) trimmer
potentiometers. Remove any connections to the input terminals. Turn the Model 35035
FUNCTION switch to the mA position, the RANGE switch to the 200 position, and the MODE switch
to the DC position.
2. Adjust the display zero trimmer potentiometer (R33) for a stable 00.0 display reading. The “-”
polarity indicator on the display should flicker on and off.
3. Turn the Model 35035 FUNCTION switch to the mAs position, depress the mAs RESET button,
and, if necessary, re-adjust the display zero trimmer potentiometer (R33) for a stable 00.0 display
reading.
4. While alternately turning the FUNCTION switch between the mAs and mA positions, re-adjust the
display zero trimmer potentiometer (R33) once again so that equal flicker of the “-” polarity indicator
on the display is maintained between the two functions.
5. Connect the current source output to the Model 35035 input terminals.
6. Turn the Model 35035 FUNCTION switch to the mA position and ensure that the RANGE switch is
set to the 200 position.
7. Alternately apply +190.0 mA and -190.0 mA to the Model 35035 input and adjust the mA gain
trimmer potentiometer (R10) for equal readings on either side of 190.0 on the display.
8. Turn the Model 35035 RANGE switch to the 20 position.
9. Alternately apply +19.00 mA and -19.00 mA to the Model 35035 input and verify that its display
reads between 18.95 and 19.05 and that the correct polarity is indicated.
10. Turn the Model 35035 RANGE switch to the 2000 position.
11. Alternately apply +1900 mA and -1900 mA to the Model 35035 input and verify that its display
reads between 1895 and 1905 and that the correct polarity is indicated.
12. Turn the Model 35035 FUNCTION switch to the mAs position and the RANGE switch to the 200
position.
13. Depress the Model 35035 RESET button.
14. Apply +100.0 mA for 1.9 seconds to the Model 35035 input and immediately adjust the mAs gain
trimmer potentiometer (R7) for a reading of 190.0 on the display.
15. Depress the RESET button.
16. Apply -100.0 mA for 1.9 seconds to the Model 35035 input and immediately re-adjust the mAs gain
potentiometer (R7) for a reading of -190.0 on the display.
17. Depress the RESET button.
4-3
Page 16
35035
G
Operators Manual
18. Repeat steps 13 through 17 until equal readings on either side of 190.0 are attained on the Model
35035 display.
19. Turn the Model 35035 RANGE switch to the 2000 position.
20. Apply +1000 mA for 1.9 seconds to the Model 35035 input and verify that the display reads
between 1895 and 1905.
21. Depress the RESET button.
22. Apply -1000 mA for 1.9 seconds to the Model 35035 input and verify that the display reads between
-1895 and -1905. Before depressing the RESET button, verify that the display drifts no more than 5
display counts over a period of one minute.
23. Depress the RESET button.
24. Place the MODE switch in the AC position.
25. Again apply -1000 mA for 1.9 seconds to the Model 35035 input and verify that the display now
reads between 1895 and 1905. This verifies correct operation of the front-end rectifier circuit in the
Model 35035.
26. Depress the RESET button, place the MODE switch back in the DC position, and turn the RANGE
switch to the 20 position.
27. With the Model 35035 FUNCTION switch still in the mAs position, apply +100.0 mA to for 1.9
seconds to the input and verify that the display reads 0.00. This verifies lockout of the 20 range in
the mAs function.
28. Turn the RANGE switch of the Model 35035 to the 2000 position and the FUNCTION switch to the
OFF position. With the current source output turned off, remove the test leads from the Model
35035 input terminals. This concludes the adjustment and operational checkout of the meter.
4.6 Battery Replacement
The Model 35035 meter operates on a standard 9 V alkaline battery such as the Eveready 522. A
appears in the display when the battery voltage reaches 7.2 V, indicating the battery needs to be
replaced. The use of a battery type other than alkaline is not recommended.
To avoid electrical shock, the instrument must be
disconnected from the x-ray generator.
1. Turn the meter OFF and remove the test leads.
2. Turn the instrument over and remove the battery compartment cover.
3. Grasp the battery clip leads near the clip, gently lift the battery out of the compartment, and remove
the old battery from the clip.
4. Install the new battery. Be certain that the clip end of the battery is inserted at the end of the
compartment where the clip wires come out from the case and that the wires get looped around the
top of the battery. This will aid the user the next time the battery needs to be replaced.
5. Re-install the battery compartment cover.
WARNIN
+
4-4
Page 17
Fuse Replacement
G
4.7 Fuse Replacement
The fuse is a 5 x 20 mm type F2AH250V. Do not use fuses of a different type or rating.
Maintenance
4
If the fuse has blown, the meter operation should
be checked to ensure other damage has not
occurred.
1. Turn the meter OFF and remove the test leads.
2. Place the end of a flat-blade screwdriver into the slot in the fuse holder. Press inward slightly and
rotate the fuse carrier one-quarter turn counterclockwise. Release pressure and extract the fuse
carrier and the fuse.
3. Remove the defective fuse and replace it with the new fuse.
4. Re-install the fuse carrier with the new fuse turning the carrier one-quarter turn clockwise into its
holder.
WARNIN
4.8 Recalibration & Repair Information / Technical Assistance
If your instrument needs recalibration or repair, we request that you consult Fluke Biomedical, Radiation
Management Services at 800.850.4606, 440.498.2564, or by Fax at 440.542.3682.
More information concerning the operation, application, or service of your instrument may be obtained
from the applications engineer at the above office.
4.9 Warranty Information
Should your instrument require warranty service, consult Fluke Biomedical, Radiation Management
Services at 800.850.4606, 440.498.2564, or by Fax 440.542.3682.
4-5
Page 18
Page 19
Fluke Biomedical
Radiation Management Services
6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms