Manual Title: Ti20 Users Supplement Issue: 3
Print Date: January 2006 Issue Date: 9/06
Revision/Date: Page Count: 1
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
Manual Supplement
CD Title: Ti20 Thermal Imager
CD Rev. & Date: 1, 2/2006
CD PN: 2492146
© 2006 Fluke Corporation. All rights reserved. Printed in U.S.A. x
Ti20 Users Manual Supplement
Change #1
On page 1-11, change Figure 1-7 as shown below:
1
Note that this change also applies to the Quick Reference Guide.
Change #2
On page 2-9, under Adjusting the Span, replace the first sentence with the following:
The full temperature range of the Imager is -10 °C to 350 °C (14 °F to 662 °F).
Under step 4, replace the last sentence with the following:
The minimum temperature span is 10 °C.
On page D-1, change the entries for Detector Type, Instantaneous Field of View, Reflected
Background Temperature and Relative Humidity to:
Detector Type………………………….128 x 96 thermal element focal plane array (FPA)
Instantaneous Field of View……………2.7 mrad
Reflected Background Temperature……-50 to 460 °C (-58 to 860 °F)
Relative Humidity………………………10 to 95% noncondensing
On page D-2, under Vibration change the following:
From: MIL-PRF-28800, paragraph 4.5.5.3.1, Class 2
To: MIL-PRF-28800, Class 2
9/06 1