1.6 General Safety Information......................................................................... 1-4
1-1
1.1 Introduction
Read these pages carefully before beginning to install and use the instrument.
The following paragraphs contain information, cautions and warnings which must be
followed to ensure safe operation and to keep the instrument in a safe condition.
Servicing described in this manual is to be done only by
qualified service personnel. To avoid electrical shock, do not
service the instrument unless you are qualified to do so.
1.2 Safety Precautions
For the correct and safe use of this instrument it is essential that both operating and
service personnel follow generally accepted safety procedures in addition to the safety
precautions specified in this manual. Specific warning and caution statements, where
they apply, will be found throughout the manual. Where necessary, the warning and
caution statements and/or symbols are marked on the instrument.
Warning
Safety Instructions
1.1 Introduction
1
1.3 Caution and Warning Statements
Caution
Used to indicate correct operating or maintenance procedures
to prevent damage to or destruction of the equipment or other
property.
Warning
Calls attention to a potential danger that requires correct
procedures or practices to prevent personal injury.
1.4 Symbols
Read the safety information in the Users
Manual
Equal potential inputs, connected
internally
Live voltageRecycling information
EarthDisposal information
DOUBLE INSULATION (Protection Class)
Static sensitive components
(black/yellow).
Conformité Européenne
1-3
123
Service Manual
1.5 Impaired Safety
1.6 General Safety Information
Whenever it is likely that safety has been impaired, the instrument must be turned off
and disconnected from line power. The matter should then be referred to qualified
technicians. Safety is likely to be impaired if, for example, the instrument fails to
perform the intended measurements or shows visible damage.
Warning
Removing the instrument covers or removing parts, except
those to which access can be gained by hand, is likely to
expose live parts and accessible terminals which can be
dangerous to life.
The instrument shall be disconnected from all voltage sources before it is opened.
Capacitors inside the instrument can hold their charge even if the instrument has been
separated from all voltage sources.
Components which are important for the safety of the instrument may only be replaced
by components obtained through your local FLUKE organization. These parts are
indicated with an asterisk (*) in the List of Replaceable Parts, Chapter 8.
FLUKE guarantees the properties expressed in numerical values with the stated
tolerance. Specified non-tolerance numerical values indicate those that could be
nominally expected from the mean of a range of identical ScopeMeter test tools.
Environmental Data
The environmental data mentioned in this manual are based on the results of the
manufacturer’s verification procedures.
Safety Characteristics
The test tool has been designed and tested in accordance with Standards ANSI/ISA
S82.01-1994, EN 61010-1 (1993) (IEC 1010-1), CAN/CSA-C22.2 No.1010.1-92
(including approval), UL3111-1 (including approval) Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use. Use of this equipment in a
manner not specified by the manufacturer may impair protection provided by the
equipment.
Characteristics
2.1 Introduction
2
2.2 Dual Input Oscilloscope
2.2.1 Vertical
Frequency Response
DC Coupled:
excluding probes and test leads:DC to 20 MHz (-3 dB)
with STL120 1:1 shielded test leads:DC to 12.5 MHz (-3 dB)
with PM8918 10:1 probe:DC to 20 MHz (-3 dB)
(optional accessory)
AC Coupled (LF roll off):
excluding probes and test leads<10 Hz (-3 dB)
with STL120 <10 Hz (-3dB)
with PM8918<1 Hz (-3 dB)
Rise Time
excluding probes and test leads<17.5 ns
Input Impedance
excluding probes and test leads1 MΩ//12 pF
with BB1201 MΩ//20 pF
with STL1201 MΩ//225 pF
with PM891810 MΩ//15 pF
DC to 20 MHz (-6 dB)
Sensitivity5 mV to 500 V/div
Display ModesA, -A, B, -B
2-3
123
Service Manual
2.2.2 Horizontal
Max. Input Voltage A and B
direct or with test leads600 Vrms
with BB120300 Vrms
(For detailed specifications see “2.7 Safety”)
Max. Floating Voltage
from any terminal to ground600 Vrms, up to 400Hz
Resolution8 bit
Vertical Accuracy±(1% + 0.05 range/div)
Max. Vertical Move±4 divisions
Max. Base Line JumpAfter changing time base or sensitivity
Normal & Single mode±0.04 divisions (= ±1 pixel)
Scope ModesNormal, Single, Roll
Ranges
Normal:
equivalent sampling20 ns to 500 ns/div
real time sampling1 µs to 5 s/div
Single (real time)1 µs to 5 s/div
Roll (real time)1s to 60 s/div
Sampling Rate (for both channels simultaneously)
Equivalent sampling (repetitive signals)up to 1.25 GS/s
Real time sampling:
1 µs to 5 ms/div25 MS/s
10 ms to 5 s/div5 MS/s
Time Base Accuracy
Equivalent sampling±(0.4% +0.04 time/div)
Real time sampling±(0.1% +0.04 time/div)
Glitch Detection≥40 ns @ 20 ns to 5 ms/div
≥200 ns @ 10 ms to 60 s/div
Glitch detection is always active.
Horizontal Move10 divisions
Trigger point can be positioned anywhere
across the screen.
2-4
2.2.3 Trigger
Screen UpdateFree Run, On Trigger
SourceA, B, EXT
EXTernal via optically isolated trigger
probe ITP120 (optional accessory)
Sensitivity A and B
@ DC to 5 MHz0.5 divisions or 5 mV
@ 25 MHz1.5 divisions
@ 40 MHz4 divisions
NormalCaptures up to 40 ns glitches and displays analog-like persistence
waveform.
SmoothSuppresses noise from a waveform.
EnvelopeRecords and displays the minimum and maximum of waveforms
over time.
Characteristics
2.3 Dual Input Meter
2
Auto Set
Continuous fully automatic adjustment of amplitude, time base, trigger levels, trigger
gap, and hold-off. Manual override by user adjustment of amplitude, time base, or
trigger level.
2.3 Dual Input Meter
The accuracy of all measurements is within ± (% of reading + number of counts) from
18 °C to 28 °C.
Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage
measurements with 10:1 probe, add probe uncertainty +1%.
More than one waveform period must be visible on the screen.
2.3.1 Input A and Input B
DC Voltage (VDC)
Ranges500 mV, 5V, 50V, 500V, 1250V
Accuracy±(0.5% +5 counts)
Turnover±12 counts
Normal Mode Rejection (SMR)>60 dB @ 50 or 60 Hz ±1%
Common Mode Rejection (CMRR)>100 dB @ DC
>60 dB @ 50, 60, or 400 Hz
Full Scale Reading5000 counts
Move influence±6 counts max.
2-5
123
Service Manual
True RMS Voltages (VAC and VAC+DC)
Ranges500 mV, 5V, 50V, 500V, 1250V
Accuracy for 5 to 100% of range
DC coupled:
DC to 60 Hz (VAC+DC)±(1% +10 counts)
1 Hz to 60 Hz (VAC)±(1% +10 counts)
AC or DC coupled:
60 Hz to 20 kHz± (2.5% +15 counts)
20 kHz to 1 MHz±(5% +20 counts)
1 MHz to 5 MHz±(10% +25 counts)
5 MHz to 12.5 MHz±(30% +25 counts)
5 MHz to 20 MHz±(30% +25 counts), excluding test leads or
probes
AC coupled with 1:1 (shielded) test leads:
60 Hz (6 Hz with 10:1 probe)-1.5%
50 Hz (5 Hz with 10:1 probe)-2%
33 Hz (3.3 Hz with 10:1 probe)-5%
10 Hz (1 Hz with 10:1 probe)-30%
DC Rejection (only VAC)>50 dB
Common Mode Rejection (CMRR)>100 dB @ DC
>60 dB @ 50, 60, or 400 Hz
Full Scale Reading5000 counts
The reading is independent of any signal
crest factor.
Move influence±6 counts max.
Peak
ModesMax peak, Min peak, or pk-to-pk
Ranges500 mV, 5V, 50V, 500V, 1250V
Accuracy:
Max peak or Min peak5% of full scale
Peak-to-Peak10% of full scale
Full Scale Reading500 counts
Frequency (Hz)
Ranges1Hz, 10Hz, 100Hz, 1 kHz, 10 kHz,
100 kHz,1 MHz, 10 MHz, 40 MHz
Frequency Range for Continuous Autoset15Hz (1Hz) to 30 MHz
2-6
Accuracy:
@1Hz to 1 MHz±(0.5% +2 counts)
@1 MHz to 10 MHz±(1.0% +2 counts)
@10 MHz to 40 MHz±(2.5% +2 counts)
Full Scale Reading10 000 counts
Duty Cycle (DUTY)
Range2% to 98%
Frequency Range for Continuous Autoset15Hz (1Hz) to 30 MHz
Accuracy:
@1Hz to 1 MHz±(0.5% +2 counts)
@1 MHz to 10 MHz±(1.0% +2 counts)
@10 MHz to 40 MHz±(2.5% +2 counts)
Resolution0.1%
Pulse Width (PULSE)
Frequency Range for Continuous Autoset15Hz (1Hz) to 30 MHz
Accuracy:
@1Hz to 1 MHz±(0.5% +2 counts)
@1 MHz to 10 MHz±(1.0% +2 counts)
@10 MHz to 40 MHz±(2.5% +2 counts)
Full Scale reading1000 counts
Characteristics
2.3 Dual Input Meter
2
Amperes (AMP)with optional current probe
Rangessame as VDC, VAC, VAC+DC, or PEAK
Scale Factor1 mV/A, 10 mV/A, 100 mV/A, and 1 V/A
Accuracysame as VDC, VAC, VAC+DC, or PEAK
(add current probe uncertainty)
Temperature (TEMP)with optional temperature probe
Range200 °C/div (200 °F/div)
Scale Factor1 mV/°C and 1 mV/°F
Accuracyas VDC (add temperature probe
uncertainty)
Decibel (dB)
0 dBV1V
0 dBm (600Ω /50Ω)1 mW, referenced to 600Ω or 50Ω
dB onVDC, VAC, or VAC+DC
Full Scale Reading1000 counts
Crest Factor (CREST)
Range1 to 10
Accuracy±(5% +1 count)
Full Scale Reading100 counts
Phase
ModesA to B, B to A
Range0 to 359 degrees
Accuracy±(1 degree +1 count)
Resolution1 degree
2-7
123
Service Manual
2.3.2 Input A
Ohm (ΩΩΩΩ)
Ranges500Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ,
30 MΩ
Accuracy±(0.6% +5 counts)
Full Scale Reading
500Ω to 5 MΩ5000 counts
30 MΩ3000 counts
Measurement Current0.5 mA to 50 nA
decreases with increasing ranges
Open Circuit Voltage<4V
Continuity (CONT)
Beep30Ω ± 5Ω in 50Ω range
Measurement Current0.5 mA
Detection of shorts of ≥1 ms
Diode
Maximum Voltage:
@0.5 mA>2.8V
@open circuit<4V
Accuracy±(2% +5 counts)
Measurement Current0.5 mA
Polarity+ on input A, - on COM
Capacitance (CAP)
Ranges50 nF, 500 nF, 5 µF, 50 µF, 500 µF
Accuracy±(2% +10 counts)
Full Scale Reading5000 counts
Measurement Current5 µA to 0.5 mA, increases with increasing
Measurement principleDual slope integrating measurement with
2.3.3 Advanced Meter Functions
ranges
parasitic serial and parallel resistance
cancellation.
2-8
Zero SetSet actual value to reference
Fast/Normal/Smooth
Meter settling time Fast1s @ 1µs to 10 ms/div
Meter settling time Normal2s @ 1µs to 10 ms/div
Meter settling time Smooth10s @ 1µs to 10 ms/div
Touch Hold (on A)Captures and freezes a stable measurement
TrendPlotGraphs meter readings of the Min and
Fixed Decimal PointPossible by using attenuation keys.
2.4 Miscellaneous
Display
Size72 x 72 mm (2.83 x 2.83 in)
Characteristics
2.4 Miscellaneous
result. Beeps when stable. Touch Hold
works on the main meter reading , with
threshholds of 1 Vpp for AC signals and
100mV for DC signals.
Max values from 15 s/div (120 seconds) to
2 days/div (16 days) with time and date
stamp. Automatic vertical scaling and time
compression.
Displays the actual and Minimum,
Maximum, or average (AVG) reading.
2
Resolution240 x 240 pixels
Waveform display:
Vertical 8 divisions of 20 pixels
Horizontal9.6 divisions of 25 pixels
BacklightCold Cathode Fluorescent (CCFL)
Power
External:via Power Adapter PM8907
Input Voltage10 to 21V DC
Power5W typical
Input Connector5 mm jack
Internal:
Battery PowerRechargeable Ni-Cd 4.8V
Operating Time4 hours with bright backlight
5 hours with dimmed backlight
Charging Time4 hours with test tool off
12 hours with test tool on
12 hours with refresh cycle
Allowable ambient temperature
during charging0 to 45 °C (32 to 113 °F)
Memory
Number of Screens2
Number of User Setups10
Mechanical
Size232 x 115 x 50 mm (9.1 x 4.5 x 2 in)
Weight1.1 kg (2.5 lbs), including battery pack.
2-9
123
Service Manual
2.5 Environmental
InterfaceRS-232, optically isolated
To Printersupports Epson FX, LQ, and HP Deskjet
Laserjet
, and Postscript
Serial via PM9080 (optically isolated
RS232 adapter/cable, optional).
Parallel via PAC91 (optically isolated
print adapter cable, optional).
To PCDump and load settings and data.
Serial via PM9080 (optically isolated
RS232 adapter/cable, optional), using
SW90W (FlukeView software for
Windows).
EnvironmentalMIL 28800E, Type 3, Class III, Style B
Temperature
Operating0 to 50 °C (32 to 122 °F)
Storage-20 to 60 °C (-4 to 140 °F)
,
Humidity
Operating:
@0 to 10 °C (32 to 50 °F) noncondensing
@10 to 30 °C (50 to 86 °F)95%
@30 to 40 °C (86 to 104 °F)75%
@40 to 50 °C (104 to 122 °F)45%
Storage:
@-20 to 60 °C (-4 to 140 °F)noncondensing
Altitude
Operating4.5 km (15 000 feet)
Max. Input and Floating Voltage 600
Vrms up to 2 km, linearly derating to 400
Vrms @ 4.5 km
Storage12 km (40 000 feet)
Vibrationmax. 3g
Shockmax. 30g
Fungus ResistanceMIL28800E, Class 3, 3.7.7 & 4.5.6.1
Salt ExposureMIL28800E, Class 3, 3.7.8.2 & 4.5.6.2.2.
Structural parts meet 48 hours 5% salt
solution test.
Electromagnetic Compatibility (EMC)
2-10
EmissionEN 50081-1 (1992): EN55022 and
EN60555-2
ImmunityEN 50082-2(1992): IEC1000-4-2, -3, -4, -5
(see also Section 2.8, Tables 2-1 to 2-3)
Enclosure ProtectionIP51, ref: IEC529
2.6 Service and Maintenance
Calibration Interval1 Year
2.7 Safety
Designed for measurements on 600 Vrms Category III Installations, Pollution Degree 2,
per:
• ANSI/ISA S82.01-1994
• EN61010-1 (1993) (IEC1010-1)
• CAN/CSA-C22.2 No.1010.1-92 (including approval)
• UL3111-1 (including approval)
Max. Input Voltage Input A and B
Direct on input or with leads600 Vrms. For derating see Figure 2-1.
With Banana-to-BNC Adapter BB120300V rms. For derating see Figure 2-1.
Max. Floating Voltage
from any terminal to ground600 Vrms up to 400Hz
Characteristics
2.6 Service and Maintenance
2
Figure 2-1. Maximum Input Voltage vs Frequency
ST8112.CGM
2-11
123
Service Manual
2.8 EMC Immunity
No visible disturbanceE= 3 V/mE= 10 V/m
The Fluke 123, including standard accessories, conforms with the EEC directive 89/336
for EMC immunity, as defined by IEC1000-4-3, with the addition of tables 2-1 to 2-3.
Trace Disturbance with STL120See Table 2-1 and Table 2-2.
Table 2-1. No Visible Trace Disturbance
Frequency range 10 kHz to 27 MHz
Frequency range 27 MHz to 1 GHz
Table 2-2. Trace Disturbance < 10%
Disturbance less than 10% of full scaleE= 3 V/mE= 10 V/m
Frequency range 10 kHz to 27 MHz
Frequency range 2 MHz to 1 GHz
(-): no visible disturbance
Test tool ranges not specified in Table 2-1 and Table 2-2 may have a disturbance of more than 10% of full
scale.
50 mV/div to 500 V/div
50 mV/div to 500 V/div
10 mV/div to 20 mV/div
5 mV/div to 20 mV/div
500 mV/div to 500 V/div
50 mV/div to 500 V/div
50 mV/div to 200 mV/div
-
Multimeter disturbance See Table 2-3.
• VDC, VAC, and VAC+DC with STL 120 and short ground lead
• OHM, CONT, DIODE, and CAP with STL120 and black test lead to COM
Table 2-3. Multimeter Disturbance < 1%
Disturbance less than 1% of full scaleE= 3 V/mE= 10 V/m
2-12
Frequency range 10 kHz to 27 MHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
Frequency range 27 MHz to 1 GHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
Test tool ranges not specified in Table 2-3 may have a disturbance of more than 10% of full scale.
3.3.4 Digital Circuit ...................................................................................... 3-25
3-1
123
Service Manual
3-2
Figure 3-1. Fluke 123 Block Diagram
ST7965.EPS
3.1 Introduction
Section 3.2 describes the functional block diagram shown in Figure 3-1. It provides a
quick way to get familiar with the test tool basic build-up.
Section 3.3 describes the principle of operation of the test tool functions in detail, on the
basis of the circuit diagrams shown in Figures 9-1 to 9-5.
For all measurements, input signals are applied to the shielded input banana jackets.
Traces and readings are derived from the same input signal samples. So readings are
related to the displayed readings.
3.2 Block Diagram
In the overall block diagram Figure 3-1, the test tool is divided in five main blocks. Each
block represents a functional part, build up around an Application Specific Integrated
Circuit (ASIC). A detailed circuit diagram of each block is shown in Section 9.
See Table 3-1. for an overview of the blocks in which the test tool is broken down, the
main block function, the ASIC name, and the applicable circuit diagram.
Circuit Descriptions
3.1 Introduction
3
Table 3-1. Fluke 123 Main Blocks
BlockMain FunctionsASICCircuit
Diagram
CHANNEL AInput A signal (V-Ω-F) conditioningC(hannel)-ASIC OQ0258Figure 9-1
CHANNEL BInput B signal (V) conditioningC(hannel)-ASIC OQ0258Figure 9-2
TRIGGERTrigger selection and conditioning
Current source for resistance, capacitance,
continuity, and diode measurements
AC/DC input coupling and Ω/F relay control
Voltage reference source
DIGITALAnalog to Digital Conversion
Acquisition of ADC samples
Micro controller (µP-ROM-RAM)
Keyboard- and LCD control
POWERPower supply, battery charger
LCD back light voltage converter
Optical interface input
T(rigger)-ASIC OQ0257Figure 9-3
D(igital)-ASIC MOT0002Figure 9-4
P(ower)-ASIC OQ0256Figure 9-5
All circuits, except the LCD unit and the KEYBOARD, are located on one Printed
Circuit Board (PCB), called the MAIN PCB.
The ASIC’s are referred to as C-ASIC (Channel ASIC), T-ASIC (Trigger ASIC), P-ASIC
(Power ASIC), and D-ASIC (Digital ASIC).
3-3
123
Service Manual
3.2.1 Channel A, Channel B Measurement Circuits
The Channel A and Channel B circuit are similar. The only difference is that Channel A
can do all measurements, whereas Channel B does not provide resistance, diode, and
capacitance measurements.
Volts, and derived measurements (e.g. current with optional probe)
The input voltage is supplied to the C-ASIC, via the LF and HF path. The C-ASIC
converts (attenuates, amplifies) the input signal to a normalized output voltage ADCA/ADC-B, which is supplied to the Analog to Digital Converters (ADC-A and ADC-B)
on the DIGITAL part. The D-ASIC acquires the digital samples to build the trace, and to
calculate readings. For the HF and LF attenuation section of the C-ASIC some external
components are required: the HF DECade ATTenuator and LF DECade ATTenuator
section.
Resistance, continuity, and diode measurements (Input A only)
The T-ASIC supplies a current via the Ω/F relays to the unknown resistance Rx,
connected to the Input A and the COM input jacket. The voltage drop across Rx is
measured as for voltage measurements.
Capacitance measurements (Input A only)
The T-ASIC supplies a current via the Ω/F relays to the unknown capacitance Cx,
connected to the Input A and the COM input jacket. Cx is charged and discharged by
this current. The C-ASIC converts the charging time and the discharging time into a
pulse width signal. This signal is supplied to the T-ASIC via the C-ASIC trigger output
TRIG-A. The T-ASIC shapes and levels the signal, and supplies the resulting pulse
width signal ALLTRIG to the D-ASIC. The D-ASIC counts the pulse width and
calculates the capacitance reading.
When the capacitance function is selected no other measurement or wave form display is
possible. There is only a numeric readout of the capacitance value.
Frequency, pulse width, and duty cycle measurements
The input voltage is measured as described above. From the ADC samples to built the
trace, also the frequency, pulse width, and duty cycle of the input signal are calculated.
Miscellaneous
Control of the C-ASIC, e.g. selecting the attenuation factor, is done by the D-ASIC via
the SDAT and SCLK serial communication lines.
An offset compensation voltage and a trace position control voltage are provided by the
D-ASIC via the APWM bus.
The C-ASIC’s also provide conditioned input voltages on the TRIG-A/TRIG-B line.
These voltages can be selected as trigger source by the T-ASIC.
3.2.2 Trigger Circuit
The T ASIC selects one of the possible trigger sources TRIG-A (Input A) or TRIG-B
(Input B). For TV triggering the selected trigger source signal is processed via the
Sync(hronization) Pulse Separator circuit (TVOUT-TVSYNC lines). Two adjustable
trigger levels are supplied by the D-ASIC via the PWM FILTERS (TRIGLEV1 and
TRIGLEV2 line). Depending on the selected trigger conditions (- source, - level, - edge,
- mode), the T-ASIC generates the final trigger signal TRIGDT, which is supplied to the
D-ASIC.
3-4
Loading...
+ 140 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.