2 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
Fluke 199C Fluke 196C
Fluke 199B Fluke 196B Fluke 192B
Bandwidth 200 MHz 100 MHz 60 MHz
Rise time 1.7 ns 3.5 ns 5.8 ns
Bandwidth limiter: User selectable: 10 kHz, 20 MHz
or off
Number of inputs: 2 plus external trigger. All inputs
isolated from each other and ground.
Input coupling: AC or DC, with ground level indicator
Input sensitivity: 2 mV/div to 100 V/div (Fluke 190C
Series); 5 mV/div to 100 V/div (Fluke 190B Series)
Variable attenuator: Variable gain on input channel A
Input voltage: 1000 V CAT II, 600 V CAT III rated -
See “general specifications” for further details.
Vertical resolution: 8 bit
Accuracy: ±(1.5% of reading + 0.04 x range/div)
Input impedance: 1 MΩ ± 1% // 15 pF ± 2 pF
Connect-and-View
™
: Advanced automatic triggering
that recognizes signal patterns, automatically sets up
and continuously adjusts triggering, time base and
amplitude. Automatically displays stable waveforms of
complex and dynamic signals like motor drive and
control signals.
Video triggering: NTSC, PAL, PAL+, SECAM. Includes
field 1, field 2 and line select.
Pulse width triggering: Pulse width qualified by time.
Allows for triggering <t, >t, =t, ≠t, where t is selectable in minimal steps of 0.01 div. or 50 nsec.
Time delay: One full screen of pre-trigger view or up
to 100 screens (1200 divisions) of post-trigger delay
Automatic Capture of 100 Screens
The instrument ALWAYS memorizes last 100 screens
(no user interaction or setup required). When an
anomaly occurs on screen, the HOLD button can be
pressed to review the full screen sequence over
and over.
Instrument can be set up for triggering on glitches
or intermittent anomalies and will operate in
“baby-sit” mode and will capture 100 events.
Alternatively, the 199C and 196C can be set up in
waveform compare mode to store only matching
(“Pass”) or only non-matching (“Fail”) acquired waveforms in the replay memory bank for further analysis.
Replay: Manual or continuous replay. Displays the
captured 100 screens as a “live” animation. Each
screen is labelled with date and timestamp. The
contents can also be viewed by manually scrolling
backwards and forwards “screen by screen.”
Replay storage: Up to 2 sets of 100 screens each
can be saved for later recall and analysis
Waveform Compare and Pass/Fail Testing
Waveform compare: Provides storage and display of a
reference waveform for visual comparison with newly
acquired waveforms. Reference is derived from an
acquired waveform and can be modified in the
ScopeMeter or externally using FlukeView®Software.
Pass/Fail Testing (199C, 196C): In waveform compare
mode, the Color Scopemeter can be set up to store
only matching (“Pass”) or only non-matching (“Fail”)
acquired waveforms in the replay memory bank for
further analysis.
Automatic Scope Measurements
Vdc, Vac rms, Vac+dc, Vpeak max, Vpeak min, Vpeak to
peak, Aac, Adc, Aac+dc, frequency (Hz), risetime, falltime, power factor, Watts, VA, VA reactive, phase, pulse
width (pos/neg), dutycycle (pos/neg), temperature °C,
temperature °F, dBV, dBm into 50 Ω and 600 Ω
Vpwm ac, Vpwm ac+dc for measurement on pulse
width modulated motordrives and frequency inverters
Cursor Measurements
Source: Input A, input B or the Mathematical Result
trace
Dual horizontal lines: Voltage at cursor 1 and 2,
voltage between cursors
Dual vertical lines: Time between cursors, 1/T
between cursors (in Hz), voltage between markers,
risetime with markers, falltime with markers
Single vertical line: Min-Max and Average voltage
at cursor position
Zoom: Up to 8x horizontal zoom
Technical Specifications 190C and 190B Series
Oscilloscope Mode
Vertical Deflection
Display and Acquisition
Fluke 199C Fluke 196C
Fluke 199B Fluke 196B Fluke 192B
Maximum 2.5 GS/s 1 GS/s 500 MS/s
real-time
sample rate
Number of 2 2 2
digitizers
Time base 5 ns/div 10 ns/div
range to 5 s/div to 5 s/div
Horizontal
Fluke 199C Fluke 199B
Fluke 196C
Fluke 196B Fluke 192B
Display 144 mm (5.67˝) full 144 mm (5.67˝)
color LCD monochrome LCD
Display modes Input A, input B, dual, Input A, input B, dual,
average, invert, replay, average, invert,
Digital Persistence mode replay, persistence
(short/medium/ (on/off)
long/infinite)
Maximum record length: 1,200 points per input in
Scope mode; 27,500 points per input in
ScopeRecord™roll mode (5 ms/div to 2 min/div)
Accuracy: ±(0.01 % of reading + 1 pixel)
Glitch capture: 50 nsec (at 5 µsec/div to 1 min/div);
250 nsec (at 2 min/div)
Visible screen width: 12 divisions in scope mode
Waveform mathematics: A + B, A - B, A * B, all
with user selectable scaling of resultant; A versus B
(X - Y-mode)
Acquisition modes: Normal, auto, single shot,
ScopeRecord™roll, glitch capture, waveform
compare, waveform compare with automatic
“Pass / Fail testing” (in 199C and 196C only)
Trigger and Delay
Source: Input A, input B, external trigger input.
All input references isolated from each other and from
ground
Modes: Automatic Connect-and-View,™free run,
single shot, edge, delay, video, video line, selectable
pulsewidth