Fluke PV350 Operating Manual

PV350
Pressure/ Vacuum Transducer Module
Instruction Sheet
ETRIC
M
PV350
ENGLISH ZE
cmHg
kPa
OFF
RE / VACUUM MOD
inHg
psi
ULE
®
Introduction
The PV350 Pressure/Vacuum Transducer Module (the module) measures gauge pressure; the difference between atmospheric (barometric) pressure and the pressure (or vacuum) applied to the pressure port on the transducer. It then converts that measurement to 1 mV dc per unit. A module and a high-performance digital multimeter (DMM) with min/max function become a datalogger capable of measuring peak high and peak low.
Box Contents
Pressure/Vacuum Transducer Module
1/4 in. male flare to 1/4 in. female pipe thread connector
9 V alkaline battery
Instruction Sheet
Quick Reference Guide
Safety Information
Warning
To avoid eye injury, always wear approved safety glasses when working with this instrument.
To avoid injury from pressure explosion:
Be sure that pressure is removed before
disconnecting any hoses or tubing from the PV350 transducer.
Follow accepted safety procedures for systems on
which pressure or vacuum measurements are made.
Make sure the PV350 is serviced by a qualified
technician.
PN 926873 May 1992 Rev. 6, 5/15
© 1999-2015 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies.
To ensure that the module is used safely, follow all
safety and operating instructions in this Instruction Sheet. If the module is not used as described in this Instruction Sheet, the safety features of the module might be impaired.
Do not use the module if the module or fittings look
damaged, or if you suspect that the module is not operating properly.
 This symbol on the module indicates that an
operator should refer to an explanation in the instruction sheet.
Follow accepted safety procedures for the systems on
which pressure or vacuum measurements are being made.
Materials Compatibility
The PV350 is compatible with industrial gases and liquids that are compatible with 316 stainless steel. To determine the compatibility of a liquid or gas, refer to the manufacturer's specifications. When making measurements on more than one type of pressure system, make sure the fluids or gases in those systems are compatible. If they are not, contamination or damage to the system being measured may result.
Cleaning
Clean the module using a soft cloth and mild soap and water. Do not immerse the PV350 in liquids.
Clean the transducer and connectors after each use with a cleaner appropriate for the liquid or gas connected during measurement. Refer to the manufacturer’s instructions for the proper cleaning material.
Measuring Pressure and Vacuum
Caution
The sensor diaphragm can be damaged by solid or sharp objects. Never insert any object into the inlet port.
Clean the sensor immediately after use with appropriate solvents (refer to the solvent manufacturer's instructions).
Note
Fluke recommends that Teflon sealing tape (or its equivalent) be used in fitting threads to ensure good seals between the PV350 and any hoses or fittings.
1. Plug the PV350 into the DMM (red polarity dot to volts input) and set the multimeter to mV dc.
2. Check the battery voltage (see Testing the Battery).
3. Press the English/Metric button to select the desired units.
4. Slide the switch to the desired setting.
5. Before applying pressure, zero the module by turning the ZERO potentiometer until the DMM displays zero. (Zero on the range you will be using and rezero the DMM whenever the scale is changed.)
To
pressure/vacuum
input
TRUE RMS MULTIMETER
87
MIN MAX
mV
H
D L
O
H
RANGE
z
H
L
E
R
X
A
M
IN
M
K
A
E
P
A
m
A
V
m
A
μ
V
V
F
F
O
A
Module
E
L U D
O
M N M
U
C A V
E
R U S
V
S
E I
C
R R T
P
E M
I
C
g
R T
E
H
M
M
n
O
C
A A
m
400mA FUS
X
A MA
10
FUSED
E
i
i
.
s
K
g
p
.
H
F
U
a
F
L
m
P
c
PV350
O
F
k
ED
!
Sensor
Figure 1. Operation
Note
The Relative or Zero function on your DMM (if available) can be used instead of the ZERO potentiometer to zero the meter. Refer to the multimeter's user manual for instructions on using this function. Zeroing the meter does not affect the module's specified accuracy.
6. Apply pressure/vacuum to the PV350 and read the DMM display.
When measuring pressure, if the DMM displays OL,
switch the multimeter range from millivolts to dc volts. Increasing the range to dc volts moves the decimal position three places to the left (500 psi displays as
0.500 V).
When measuring vacuum the DMM displays a minus sign.
Testing the Battery
1. Turn the DMM and the module OFF and connect the PV350 to the DMM.
2. Turn the DMM on and set the DMM to the mV dc range.
3. If the DMM displays less than 100 mV, replace the battery.
How to Replace the Battery
Warning
To avoid electrical shock, disconnect the module from all devices before opening the case. Be sure to completely reassemble the module before attempting to use it.
Battery
Figure 2. Battery Replacement
1. Turn the module OFF and unplug it from the DMM.
2. Remove the screw from the back of the module, separate the case, and replace the battery.
3. To reassemble, line up the holes in the case with the METRIC/ENGLISH switch and zero knob. Then snap the two halves together. Reinstall the screw.
Typical Applications
Refer to Figures 3 and 4 for typical applications.
AUTOMOTIVE TESTER
88
DC
mV
20 2
46
3
82468
4
4000
mV
H
MIN MAX RANGE HOLD
ZERO
RPM
% DUTY
ALERT
_
+
Hz
SMOOTH
ms-PULSE
TRIGGER
mV
mA
V
A
RPM
mA
V
A
OFF
V
mA A
COM
RPM
A
400mA MAX
FUSED
10A MAX
1000V MAX
FUSED
!
IC
R
T
E
M
Z H
IS
L
G
N
E
P
V
5
0
0
P
R
E
S
S
U
R
E
/
V
A
C
U
U
M
M
O
D
U
L
E
inHg
cmHg
psi
kPa
OFF
Figure 3. Pressure for HVAC/R Super Heat Measurements
AUTOMOTIVE TESTER
88
RANGE
DC
mV
20 2
46
3
82468
4
4000
mV
H
MIN MAX RANGE HOLD
ZERO
RPM
% DUTY
ALERT
_
+
Hz
SMOOTH
ms-PULSE
TRIGGER
mV
mA
V
A
RPM
mA
V
A
OFF
V
mA A
COM
RPM
A
400mA MAX
FUSED
10A MAX
1000V MAX
FUSED
!
IC
R
T
E
M
H
IS
L
G
N
E
P
V
5
0
0
P
R
E
S
S
U
inHg
cmHg
psi
kPa
OFF
Figure 4. Measure Pneumatic Pressures
Z
R
E
/
V
A
C
U
U
M
M
O
D
U
L
E
Conversion Factors
psi = inches of H2O x (3.6127 x 10-2) inches of H2O = psi x 27.68 psi = mm of H2O x (1.4223 x 10-3) mm of H2O = psi x 703.1 psi = cm of H2O x (14.223 x 10-3) cm of H2O = psi x 70.3 psi = bar x (14.503) bar = psi x 0.0689 psi = lbs/ft2 x (6.9444 x 10-3) lbs/ft2 = psi x 144 psi = mbar x (1.4503 x 10-2) mbar = psi x (68.9513)
psi = Pascals x (1.4503 x 10-4) Pascals = psi x (6.895 x 103) All H2O conversion factors are @ 4°C
Specifications
The following specifications apply at 23 ±5 °C for 1 year after calibration when the module is zeroed prior to each measurement.
Pressure
Range Accuracy
0 psi to 500 psi: (0 kPa to 3447 kPa)
Range Accuracy
0 in-Hg to -29 in-Hg (0 cm-Hg to -76 cm-Hg)
Temperature coefficient: Add 0.05 % of reading per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C.
Output: 1 mV per unit (psi, kPa, cm-Hg or in-Hg) Resolution: Determined by the resolution of the measuring device. RFI Specification: Typically <1 % error from 80 to 200 MHz at 3 V/m. Electromagnetic Compatibility (EMC)
International IEC 61326-1: Portable; IEC 61326-2-2; CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes.
Korea (KCC) Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Meter Input Impedance: 1 MΩ Maximum working pressure: 500 psi Burst Pressure: 1000 psi
±0.25 % of reading ±0.3 psi ±0.25 % of reading ±2.1 kPa)
Vacuum
±0.25 % of reading ±0.5 in-Hg (±0.25 % of reading ±1.3 cm-Hg)
Storage Temperature: -40 °C to 71 °C (Battery removed)
Operating Environment
Temperature Humidity
-10 °C to 10 °C 10 °C to 30 °C 30 °C to 40 °C 40 °C to 50 °C 50 °C to 55 °C
Non-condensing 0 °C to 95 % Relative Humidity 0 °C to 75 % Relative Humidity 0 °C to 45 % Relative Humidity 0 °C to 35 % Relative Humidity
General
Weight: 12 oz, (336 grams) Battery: Standard 9 V Alkaline battery (NEDA #1604, IEC 6LR61) Battery Life: 300 hours (typical) Electrical Cable: Black PVC, will withstand up to 105 °C Cable Length: 8 ft Pressure Port Connection: 1/4 in. NPT Connector: Brass 1/4 in. male flare to 1/4 in. female pipe thread
List of Replaceable Parts
Part Number Description
614487 2813197 2816172 913207 926873 926881
To contact Fluke or for service, call one of the following telephone numbers:
USA: 1-888-44-FLUKE (1-888-443-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-6714-311481 Singapore: +65-6799-5566 Anywhere in the world: +1-425-446-5500 Or, visit Fluke's Web site at www.fluke.com.
Fluke Corporation Fluke Europe B.V. P.O. Box 9090 P.O. Box 1186 Everett, WA 5602 B.D. 98206-9090 Eindhoven The Netherlands
Battery, Primary, 9 V, 0-15 mA Case Bottom, PV350 Case Top, PV350 Fitting, BR, SAE, 45, 1/4 FP PV350 Instruction Sheet PV350 Quick Reference Guide
Loading...