High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases
can be released unexpectedly and with extreme force. High pressure systems should be assembled and
operated only by personnel who have been instructed in proper safety practices.
This instrument is not to be operated in any other manner than that specified by the manufacturer.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any
form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, a
Fluke Company 4765 East Beautiful Lane, Phoenix AZ 85044-5318, USA.
DH Instruments makes sincere efforts to ensure accuracy and quality of its’ published materials; however, no warranty, expressed
or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the
use of the information in this manual or products described in it. Mention of any product or brand does not constitute an
endorsement by DH Instruments of that product or brand. This manual was originally composed in English and was subsequently
translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version
and other language versions, the English version predominates.
Products described in this manual are manufactured under international patents and one or more of the following U.S. patents:
5,142,483; 5,257,640; 5,331,838; 5,445,035. Other U.S. and international patents pending.
AutoRange, AutoZ, DH Instruments, DH, DHI, COMPASS, PPC, PPCH, PPCK+, QDUT, Q-RPT, RPM, RPM4 and SDS are
trademarks, registered and otherwise, of DH Instruments, a Fluke Company.
Loctite is a trademark of Loctite Corporation
Teflon is a registered trademark of the Dupon de Nemours Company.
2.3.1 PREPARING FOR OPERATION ...................................................................................................................7
2.3.2 FRONT AND REAR PANELS ........................................................................................................................ 7
2.3.2.1 FRONT PANEL ..........................................................................................................................................7
2.3.3 POWER CONNECTION................................................................................................................................. 8
2.3.4 REMOTE [ENTER/SET P] CONNECTION (FOOTSWITCH OR OTHER SWITCH) .....................................8
2.3.5 CONNECTING PNEUMATIC POWER (DRIVE PORT) .................................................................................9
2.3.6 SETTING UP AND FILLING THE FLUID RESERVOIR ................................................................................9
2.3.11 TURN OFF ABSOLUTE MEASUREMENT MODE ......................................................................................18
2.4 POWER-UP AND VERIFICATION.........................................................................................................18
2.4.1 SWITCH POWER ON ..................................................................................................................................18
2.4.2 APPLY PNEUMATIC POWER (DRIVE AIR) ...............................................................................................18
2.4.6 CHECK PRESSURE CONTROL OPERATION...........................................................................................19
2.5 STORAGE AND SHIPPING ...................................................................................................................20
2.5.1 SHORT TERM STORAGE ...........................................................................................................................20
2.5.2 LONG TERM STORAGE .............................................................................................................................20
3.1 USER INTERFACE ................................................................................................................................21
3.1.1 MAIN RUN SCREEN....................................................................................................................................21
3.1.2 FUNCTION / DATA KEYPAD LAYOUT AND PROTOCOL.........................................................................23
3.1.3 DIRECT PRESSURE CONTROL KEYS......................................................................................................24
3.2 GENERAL OPERATING PRINCIPLES .................................................................................................25
3.2.1 PRESSURE CONTROL PRINCIPLES ........................................................................................................25
3.2.2 AUTOMATED PRESSURE CONTROL .......................................................................................................28
3.2.2.1 IDLE, NO AUTOMATED CONTROL (MEASURE MODE).......................................................................28
3.2.2.2 DYNAMIC CONTROL ..............................................................................................................................29
3.5.1.2 RUN AUTOZ.............................................................................................................................................78
5.10.1 PRIMING THE PUMP.................................................................................................................................152
5.10.1.1 PRIMING THE PUMP CHECK VALVE ..................................................................................................153
5.14.1.1 POWER ENTRY AND FUSE..................................................................................................................167
5.14.1.2 POWER SUPPLIES ...............................................................................................................................167
5.14.1.4 MINI MICRO BOARD.............................................................................................................................168
5.14.2.1 PUMP DRIVE AIR PRESSURE TRANSDUCER ...................................................................................169
5.14.2.2 HI Q-RPT OR UTILITY SENSOR...........................................................................................................169
5.14.2.3 TEST VENT VALVE...............................................................................................................................170
5.14.2.4 LO Q-RPT SHUTOFF VALVE................................................................................................................170
5.14.2.5 LO Q-RPT...............................................................................................................................................170
5.14.2.6 LO Q-RPT VENT VALVE .......................................................................................................................170
5.14.2.8 FRONT PANEL ......................................................................................................................................170
5.14.2.11 INLET CONTROL VALVE ......................................................................................................................170
5.14.2.12 OUTLET CONTROL VALVE ..................................................................................................................171
Figure 1. Front panel.................................................................................................................................... 7
This manual is intended to provide the user with the basic information necessary to operate a PPCH
pressure controller/calibrator. It also includes a great deal of additional information provided to allow you
to optimize PPCH use and take full advantage of its many features and functions.
Before using the manual, take a moment to familiarize yourself with the Table of Contents structure:
Sections 1, 2 and 3 should be read by all first time PPCH users. Section 3 is most important for those
using the local front panel interface but should be read over
general PPCH operating principles. Section 4 is for remote operation from an external computer. Section 5
provides maintenance and calibration information. Section 6 is a quick troubleshooting guide. Use it to
troubleshoot unexpected PPCH behavior based on the symptom of that behavior.
expressions have specific meaning as they pertain to PPCH. The Glossary, Section 9, is useful as a
quick reference for exact definition of specific words and expressions as they are used in the manual.
For those of you who “don’t read manuals”, go directly to Section 2.3 to set up your PPCH and then go
to Section 2.4 for power-up and verification. This will get you up and running quickly with a minimal risk of
causing damage to yourself or your new PPCH. THEN… when you have questions or start to wonder about all
the great features you might be missing, get into the manual!
(CAUTION) is used in throughout the manual to identify user warnings and cautions.
T
T
HHIISS
M
AANNUUAAL
M
L
by all users to familiarize themselves with
Certain words and
(NOTE) is used throughout the manual to identify operating and applications advice and
additional explanations.
[ ] indicates direct function keys (e.g., [RANGE]).
PPCH is a stand-alone, pressure controller intended to precisely generate, adjust and control hydraulic
pressure into a closed volume as is commonly needed for the calibration and testing of pressure
measuring instruments. It has been designed to provide very high performance combined with maximum
versatility and ease of use.
PPCH can be equipped with a low cost utility sensor for pressure monitoring or one or two Quartz
Reference Pressure Transducers (Q-RPTs) to allow it to set and measure pressure with higher precision
and very low measurement uncertainty. Up to four external Q-RPTs in one or two external Reference
Pressure Monitors (RPM4s), can also be integrated into a PPCH system. A barometer is also included
for atmospheric pressure compensation of gauge mode measurements and as a reference for AutoZero
in absolute measurement mode.
PPCH generates high liquid pressure internally using a hydropneumatic pump driven by a drive air supply.
Pressure control is achieved by servo-motor controlled, high pressure needle valves and a thermal
pressure control system.
PPCH is controlled locally by the operator using a front panel display, keypad and function keys or
remotely by a computer using ASCII character command strings over an RS232 or IEEE-488.2 interface.
PPCH models are available to measure and control pressure in ranges from as low as 20 MPa (3 000 psi)
to 200 MPa (30 000 psi) in absolute and gauge measurement modes.
85 to 264 VAC, 50/60 Hz, 75 W max consumption
15 to 35 °C
- 20 to 70 °C
Meets MIL-T-28800D
50 kg (110 lb) approx
Without optional enclosure: 26 cm H x 44 cm W x 55.5 cm D (10.2 in. x 17.3
in. x 21.9 in.) Front panel is 19” wide X 6U high with mounting holes for
standard
19 in. rack mount systems
With optional enclosure: 30 cm H x 52 cm W x 55.5 cm D
(11.8 in. x 20.5 in. x 21.9 in.)
To prevent product overheating, provide proper ventilation. Allow 10 cm
(4 in.) clearance from rear panel cooling fan.
RS232 (COM1), RS232 (COM2), IEEE-488.2
1 A, 250 VAC fuse, 5 x 20 mm, time lag type fuse
Internal power supply fuse not replaceable by operator: 2A, 250 V (UV 440-2
power supply), 3.15A, 250 V (NFS40-7612 power supply)
Four controller models from 70 MPa (10 000 psi max to 200 MPa (30 000 psi)
max.
Low uncertainty measurement provided by selection of Quartz Reference
Pressure transducer(s) (Q-RPTs)
Di-2-Ethyl Hexyl Sebacate (other fluids optional)
300 cc
DRIVE: 1/8 in. NPT F
TEST: DH500 F
NOTE: DH500 is a gland and collar type
fitting for ¼ in. (6.35 mm) coned and left
hand threaded tube. DH500 is equivalent to
EXTERNAL: 1/8 in. NPT F
AE F250C, HIP HF4, etc.
OVERFLOW: 1/8 in. NPT F
FILL/DRAIN: 1/4 in. swage M
ATM: 10-32 UNF
Pressure Limits
Maximum Working:Test Pressure: Controller or Hi Q-RPT maximum
Maximum Test Pressure:without Damage: 115 % controller or Hi Q-RPT maximum
Quartz reference pressure transducers (Q-RPTs) can be installed in PPCH to obtain
high precision and low uncertainty pressure measurement. One or two Q-RPTs can
be included in the PPCH and/or additional Q-RPTs can be used externally mounted
in DHI RPM4 Reference Pressure Monitors (see Section 3.2.5).
ll PPCH Q-RPTs are of the absolute pressure type (Axxx) us ing an evacuated,
A
permanently sealed reference. Axxx Q-RPTs used in PPCH can measure
absolute and gauge pressure. Gauge pressure with an Axxx (absolute) Q-RPT is
obtained by offsetting atmospheric pressure and applying dynamic compensation for
atmospheric changes using the on-board barometer (see Section 3.2.4).
See Section 3.3.3,
gauge measurement modes.
PPCHs configured with two Q-RPT modules have only one TEST port. PPCH internal
valves and logic handle switching between the two Q-RPTs as needed.
Table 1. Reference pressure transducer (Q-RPT) designations and ranges
PRINCIPLE, for additional information on absolute and
2. Predicted Q-RPT measurement stability limit (k=2) over one year assuming regular use of AutoZero function. AutoZero
occurs automatically in gauge mode whenever vented, by comparison with barometric reference in absolute mode.
Absolute mode predicted one year stability without AutoZ is ± (0.005 % Q-RPT span + 0.005 % of reading).
3. Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted one year
stability limit, temperature effect and calibration uncertainty, combined and expanded (k=2) following the ISO “Guide to the
Expression of Uncertainty in Measurement.”
4. Maximum deviation of the PPCH controlled pressure from the true value including measurement uncertainty and dynamic
control hold limit.
5. % of reading applies to 30 to 100 % of Q-RPT span. Under 30 % of Q-RPT span, the value is a constant which is the % of
reading value times 30 %.
A20M 20 3 000
Warm Up Time
Resolution
Acceleration Affect
Precision
Predicted Stability2
SI VERSION US VERSION
MAXIMUM PRESSURE
Absolute and Gauge [MPa]
30 minute temperature stabilization recommended for
best performance from cold power up.
To 1 ppm, user adjustable
5 to 35 °C
± 0.008 % /g maximum, worst axis
Allows operation at ± 20° from reference plane without
significant effect
1
<A200M A200M
± 0.012 % of reading5 ± 0.015 % of reading5
± 0.005 % of reading5 ± 0.005 % of reading5
± 0.013 % of reading5 ± 0.018 % of reading5
± 0.015 % of reading5 ± 0.020 % of reading5
Base PPCH pressure controllers include a utility sensor. If a Hi Q-RPT
(see Section 3.2.5) is installed, there is no utility sensor.
Utility sensors are used
for pressure control, system monitoring and safety
functions. They are intended for indication only, not to provide reference
pressure measurement. Q-RPTs (see Section 1.2.2.1) are used for reference
measurement.
PPCH-100M: 100 MPa (15 000 psi) absolute and gauge
PPCH-140M: 140 MPa (15 000 psi) absolute and gauge
PPCH-200M: 200 MPa (30 000 psi) absolute and gauge
0.001 % of span
0.1 % of span
1.2.2.3 ON-BOARD BAROMETER
The on-board barometer is used only to measure changes in atmospheric
pressure to provide dynamic compensation of the Q-RPT’s atmospheric pressure
offset when operating in gauge measurement mode. If calibrated,
it can also be used as an AutoZ reference for AutoZeroing in absolute
measurement mode.
PPCH is delivered in a wooden crate with high density polyethylene inserts to hold it in place.
Remove the PPCH and its accessories from the shipping container and remove each
element from its protective plastic bag.
PPCH weighs more than 50 kg (110 lb). Take appropriate precautions to lift and place it safely.
2.1.2 INSPECTING CONTENTS
Check that all items are present and have no visible damage. A standard PPCH includes all
items indicated in Table 2.
PPCH weighs more than 50 kg (110 lb). Be sure that the location in which it’s installed is adequate to
support the load.
The PPCH can be installed on any flat, stable surface at a convenient height. The PPCH can also be
mounted in a standard 19 in. rack (see Section 2.2.1).
Minimizing the distance between
the PPCH and the device or system under test will enhance control
performance and reduce pressure setting times.
Ready access to the PPCH rear panel should be considered to facilitate making and breaking
pressure connections.
Also consider:
• Access to the internal reservoir FILL/DRAIN port if an external reservoir is not being used (see
Section 2.3.6).
•
Location and connections to an external reservoir, if used (see Section 2.3.6.2).
Hydraulic and RS-232 connections to an external RPM4 should be considered if RPM4 is to be
•
used as an external reference pressure measurement device (see Section 2.3.9.1).
Location and connections to the device or system into which pressure is to be controlled (see
•
Section 2.3.9.4).
Support facilities required include:
•An electrical pow
er source of 85 to 264 VAC, 50/60 Hz.
•A pneumatic power supply, of “drive air” to be connected to the PPCH DRIVE port. Required
pressure varies with PPCH model (see Section 2.4.2).
ALWAYS use external tubing and fittings rated for pressures equal to or greater than the maximum
pressure PPCH will generate.
2.2.1 INSTALLATION IN A 19 IN. RACK MOUNT
PPCH is dimensioned to be ready for direct mounting in a standard 19 in. rack mount system.
If PPCH was delivered in the optional enclosure, it must be removed from the enclosure prior
to rack mounting.
The front panel is 6U (10.5 in.) high.
Install PPCH into the rack mount system using heavy duty front panel bolts provided by the rack
vendor. Due to PPCH’s weight, it is imperative that a rail system also be used to support it along
the sides and/or rear. These are also supplied by the rack mount system vendor.
PPCH weighs more than 50 kg (110 lb). Take appropriate precautions to lift and place it safely.
It is the user’s responsibility to support PPCH adequately on the sides and/or rear when installing it in
a rack mount system.
2. INSTALLATION
2.3 INITIAL SETUP
2.3.1 PREPARING FOR OPERATION
To prepare PPCH for check out and operation:
Remove the plastic caps from the PPCH rear panel pressure connections.
Remove the protective plastic sheet from the front panel display. Familiarize yourself briefly with the front and rear panel (see Section 2.3.2).
1. Electrical power connector (IEC-320C13) and fuse
2. Cooling fan
3. Remote [ENTER] connector
4. DRIVERS connector
5. RS-232 port, COM2
6. RS-232 port, COM1
7. Product label
8. IEEE-488 connector
9. ATM (barometer) port
10. Internal reservoir fill funnel
11. Internal reservoir level sight tube
12. Reservoir FILL/DRAIN port
13. PUMP DRIVE OUT gas exhaust port
14. Reservoir FILL/DRAIN valve
15. EXTERNAL reservoir return port
16. Reservoir OVERFLOW port
17. RESERVOIR RETURN valve
18. DRIVE IN air supply port
19. Mechanical module pull out handles
20. TEST port
Figure 2. Rear panel
2.3.3 POWER CONNECTION
Check that the PPCH power switch is OFF.
Connect the supplied power cable to the rear panel power input module.
Connect the other end of the power cable to an electrical supply of 85 to 264 VAC, 50/60 Hz.
2.3.4 REMOTE [ENTER/SET P] CONNECTION (FOOTSWITCH
OR OTHER SWITCH)
Connect the optional remote ENTER footswitch, if available or a user supplied switch fitted to
the optional cable (see Section 7.3). Connect the cable to the PPCH rear panel connection
labeled REMOTE ENTER. Activating the switch is equivalent to
key on the front panel (see Section 3.1.4).
Using a pressure connecting hose or tube of appropriate pressure rating, connect drive air supply
to the DRIVE port on the rear panel of PPCH. It is important to proper PPCH pressure control
that the supply connection not excessively restrict flow. The tube should have an ID of at least 3
mm (1/8 in.) and not be more than 2 m (6 ft) long. Generally, a 3 mm ID tube has OD of at least
6.35 mm (1/4 in.). The DRIVE port connection is 1/8 in. NPT female.
The drive air supply is regulated internally by PPCH. Drive air requirements for the different
PPCH models are listed in Table 3.
The PPCH fluid reservoir holds the fluid that supplies PPCH’s hydropneumatic pump and
returns from its exhaust valve.
PPCH has an internal reservoir with a capacity of 300 cc that is filled from and whose level can
be viewed on the rear panel of the instrument. If greater capacity and/or remote reservoir filling
and level viewing are needed, external reservoir (part number 402102) may be installed.
See Section 2.3.6.1 to set up and fill the internal reservoir.
See Section 2.3.6.2 to set up and fill an external reservoir.
Take care not to let the PPCH reservoir become empty. Operating PPCH with an empty
reservoir will cause the hydropneumatic pump to draw air and lose its prime. A pump that has
lost its prime must be reprimed using a special procedure (see Section 5.10.1). PPCH has a low
level warning displayed on the front panel that can be also checked by remote command (see
fluid
Section 3.5.7.3).
The system connected to the TEST port is connected to the reservoir when the PPCH is
vented ([VENT] LED ON). If there is an open point in the system below the fluid level in the
reservoir, liquid will run from the reservoir and out of the open point.
2.3.6.1 INTERNAL RESERVOIR
The PPCH is shipped with its internal reservoir filled with PPCH’s working fluid.
To operate without an external reservoir, the internal reservoir attachments must
be installed on the PPCH rear panel. These are included with the PPCH
accessories (see Table 2).
- Reservoir fill tube bracket, p/n 123832 (grey painted bracket)
- (2) M3 x 8 bolts, p/n 101010-Z
- Reservoir fill tube assembly, p/n 402039 (1/2 in. PFA tube with fittings
attached)
To prevent draining the reservoir accidentally, leave the RESERVOIR FILL/DRAIN
valve in the CLOSED position whenever the internal reservoir or external reservoir
attachments are not installed. Opening the FILL/DRAIN valve opens the FILL/DRAIN
port to the reservoir.
To set up to use the internal reservoir without an external reservoir, proceed as
follows (numeric references refer to Figure 3):
Install the reservoir fill tube bracket(1) to the PPCH rear panel using 2 ea. M3
x 8 bolts(3).
- Funnel p/n 123959
Install the reservoir fill tube assembly(4) by connecting the 1/4 in. swage
female on the assembly to the 1/4 in. swage male FILL/DRAIN connection.
Remove and retain the stainless steel cap that is installed on the connection
as delivered. Orient the PFA tube vertically. Bend the tube slightly to extend
through the reservoir fill bracket.
There is a red plug with a center pull tab in the end of the PFA tube. This
should be left in place when the funnel is not being used to prevent
contamination and/or evaporation of the fluid. The funnel(2) can be stored in
the holder just behind the tube.
Open the FILL/DRAIN valve(5). Fluid may enter the fill drain tube(4).
Check that the EXTERNAL port(6) is plugged. It is plugged when delivered.
If it is not, use a 1/8 in. NPT M plug with Teflon® tape on the threads to plug
it. Not plugging the EXTERNAL port when using the internal reservoir only
will cause the fluid exhausted from the PPCH to leak out rather than
returning to the reservoir.
Check that the OVERFLOW port(7) is open to atmosphere and
unobstructed. This assures the reservoir is open to atmosphere so proper
system venting and zeroing can occur.
Be sure the RESERVOIR RETURN valve(8) is on the INTERNAL position. Check the level of the fluid in the fill/drain tube. If the level is below the MAX
FILL position, top off the tank. See Section 5.6.1 for information on filling
and draining the PPCH internal reservoir.
The PFA reservoir fill tube is delivered with a red plug with a center pull tab
installed. This plug can be used after the tube is installed to cover the opening
and reduce evaporation of the fluid. The plug is vented.
The PPCH is shipped with its internal reservoir filled with PPCH’s working fluid.
An external reservoir, if desired, is connected in series to supply the internal
reservoir so that the internal reservoir’s features are still available.
When using an external reservoir the internal reservoir attachments are NOT
installed. If they have been installed, remove them prior to setting up the
external reservoir (see Section 2.3.6.1). Be sure to put the FILL/DRAIN valve in
the CLOSED position before removing the fill/drain tube assembly.
To prevent draining the reservoir accidentally, leave the RESERVOIR FILL/DRAIN
valve in the CLOSED position if the internal or external reservoir attachments are not
installed. Opening the FILL/DRAIN valve opens the FILL/DRAIN port to the internal
reservoir.
When installing an external reservoir, the bottom of the reservoir should be
higher than the PPCH instrument. Avoid low points in the tubing between the
PPCH and the external reservoir.
To set up to use an external reservoir, proceed as follows (numerical references
refer to Figure 4):
With the reservoir FILL/DRAIN valve in the CLOSED position, connect a
tube coming from the bottom of the external reservoir to the FILL/DRAIN
port(1). The FILL/DRAIN port is1/4 in. M swage.
Connect an external reservoir return tube from the EXTERNAL port(3) to the
inside of the reservoir. Remove and retain the 1/8 in. NPT plug installed in
EXTERNAL port as shipped. The reservoir return tube returns fluid
exhausted by PPCH to the external reservoir. The top of the tube should
extend vertically to a level slightly higher than the top surface of the fluid in
the reservoir. The top of the tube defines the fluid head applied to the PPCH
Q-RPT when the PPCH is vented (see Section 5.6.2, Figure 20).
Fill the external reservoir with fluid. This should be the same fluid that is in
the PPCH. PPCH will not operate properly with a fluid other than the liquid
with which it was filled when delivered. The MAX FILL label on the rear of
the PPCH does not apply when using an external reservoir.
Purge air from the internal reservoir and overflow circuit:
a. Prepare an 1/8 in. NPT plug (one was delivered installed in the
EXTERNAL port) by wrapping its threads with Teflon tape.
b. While watching the OVERFLOW port(4), slowly turn the
RESERVOIR FILL/DRAIN valve(2) to the OPEN position.
c. When air free fluid begins to flow from the OVERFLOW port, plug
the port with the 1/8 in. NPT plug.
Be sure the RESERVOIR RETURN valve(5) is in the EXTERNAL position.
Be sure the external reservoir is open to atmosphere so proper system
venting and zeroing can occur.
Fill the reservoir return tube connected between the reservoir and the
EXTERNAL port with fluid. See Section 5.6.2 for additional information in
filling and draining an external reservoir.
1. Reservoir FILL/DRAIN port
2. PUMP DRIVE OUT gas exhaust port
3. Reservoir FILL/DRAIN valve
Figure 4. External reservoir setup
When the PPCH is vented ([VENT] LED ON), the system connected to the TEST
port is connected to the reservoir. If there is an open point in the system below the
fluid level in the reservoir, liquid will run out of the reservoir through the open point.
2.3.7 RESERVOIR RETURN SETTINGS
The PPCH rear panel has a section labeled RESERVOIR RETURN. The settings and
connections in this panel vary depending on whether the internal reservoir is being used
alone or an external reservoir is being used.
Refer to Table 4 for correct reservoir return settings and other conditions when using the
internal reservoir alone or an external reservoir. Also see Section 2.3.6 for internal and
external reservoir setups.
Table 4. Reservoir return settings for internal or external reservoir
RESERVOIR
RESERVOIR
Internal only INTERNAL Open to
External EXTERNAL Plugged Connected to external
RETURN VALVE
POSITION
OVERFLOW
PORT EXTERNAL PORT
Plugged
atmosphere
reservoir return tube
See Section 5.6, Figure 19 and Figure 20 for schematics of the PPCH reservoir system.
Incorrect settings or port conditions in the RESERVOIR RETURN panel may lead to fluid
leaks and faulty measurements.
2.3.8 INSTALLING PUMP DRIVE OUT GAS EXHAUST
FILTER/MUFFLER
A stainless steel filter (P/N 103897) is supplied in the PPCH accessory kit. The filter dampens
the sound of supply gas being exhuasted from the hydro-pneumatic pump. To install the filter,
first wrap 1 to 2 turns of teflon tape around the threads then insert the threaded end of the muffler
into the Pump Drive Out port using a 9/16" wrench as shown in Figure 5. Do not overtighten.
It is important to use a wrench to install the filter instead of a screwdriver. The button head
screw on the end of the filter adjusts the conductance of the muffler. If the muffler is too restrictive
then the hydropneumatic pump will stall and the PPCH may not be able to increase pressure.
The filter/muffler is adjusted for conductance at the factory. If the filter adjustment is
altered it can be returned to the factory setting by using a screwdriver to turn the button head
screw clockwise until it stops then turning it one and half turns counterclockwise.
ALWAYS use external tubing and fittings rated for pressure equal to or greater than the
maximum pressure PPCH will be used to generate.
The PPCH TEST port is a DH500 female fitting.
DH500 is a gland and collar type fitting for 1/4 in. (6.35 mm) coned and left hand threaded
tube. DH500 is equivalent to AE F250C, HIP HF4, etc.
DH500 fittings are rated for working pressure up to 500 MPa (72 500 psi). Be sure to use
stainless steel tubing rated for your maximum operating pressure with the DH500 fittings.
2.3.9.1 INSTALLING A LINE FILTER TO PROTECT THE PPCH
PPCH uses pressure shutoff valves that are needle valves that close by the
metal to metal contact of the needle in its seat. These valves can be damaged
by hard particles. If the test system to which PPCH will be connected may
contain hard particles, such as metal shavings from a production process, a line
filter should be installed on the PPCH test port to prevent the particles from
entering the PPCH when reducing pressure. The filter should not be smaller
then 40 micron. DHI recommends high pressure filter kit p/n 402376.
2.3.9.2 CONNECTING TO AN RPM4 TO BE USED AS AN EXTERNAL
REFERENCE DEVICE
ALWAYS use external tubing and fittings rated for pressure equal to or
greater than the maximum pressure PPCH will be used to generate.
PPCH can be connected hydraulically and by RS-232 to one or two RPM4
reference pressure monitors to use up to four external Q-RPTs (two in each RPM4)
as external reference pressure measurement devices (see Section 3.2.5).
The PPCH then manages communications and other
integrate the RPM4 measurement capabilities into the PPCH system. The user
is responsible for setting up hardware and following procedures to isolate
lower pressure Q-RPTs from possible overpressure if pressures higher
than the Q-RPT maximum are generated.. PPCH and RPM4 do not have hardware provisions for isolating external Q-RPTs from the TEST circuit.
As a general rule, making the hydraulic connection between the PPCH TEST
port and the remote Q-RPT as direct as possible favors good pressure control. As
distance, volumes and restrictions between the PPCH TEST port and the remote QRPT increase, the possibility of difficulty with pressure control when using the
external Q-RPT increases.
To connect a PPCH to an RPM4 to be used as part of the PPCH system proceed
as follows:
Set up the RPM4 for use as a PPCH external device following the instructions
in the RRPM4 Operation and Maintenance Manual, Using RPM4 With a PPC3
Controller/Calibrator Section.
Using tubing of appropriate pressure rating and a tee fitting, connect the
RPM4 Q-RPT TEST(+) port to the PPCH TEST port (see Figure 6 for
instructions
on making this connection using the standard interconnection
accessories supplied with the PPCH). The third leg of the tee is for the
connection to the device or system under test.
When external Q-RPTs are used with PPCH, the maximum set pressure is
the maximum pressure of the PPCH controller/calibrator. The maximum pressure
of the PPCH is determined by the PPCH controller model or the Hi Q-RPT if one is
installed (see Section 1.2.1).
Using a standard pin-to-pin DB-9M to DB-9F RS232 cable, connect COM2 of
the PPCH to COM1 of the RPM4 (see Section 4.2.1.3).
Turn ON the RPM4 and the PPCH.
Press [RPT] on the PPCH key pad. This causes PPCH to execute the
Q-RPT search function (see Section 3.3.5). Following the search, Q-RPTs
with whom communication is established
are shown on the PPCH display,
identified by their position (see Table 6).
For PPCH to identify external RPM4s, the RPM4 COM port settings
must be:
Baud rate: between 1200 and 19200
Parity: Even
Data bits: 7
Stop bits: 1
If PPCH is not able to establish communications with an RPM4 and its
Q-RPTs, check that the RPM4 COM1 port settings conform to the requirements
above. If the COM1 port settings are correct, check that the correct communications
cable is being used (standard pin-to-pin DB-9M to DB-9F RS232) and is connected
to the correct communications ports.
Figure 6. Connecting PPCH to an RPM4 or PG7302 using the standard
PPCH interconnection accessories
2.3.9.3 CONNECTING TO A PG7302 PISTON GAUGE
7. DH500 union (RPM4) or tee (PG7000)
8. 5 in. (127 mm) DH500 x 1/8 in. tube
weld nipple (slightly flexible)
9a. RPM4 or RPM3
9b. PG7000
ALWAYS use external tubing and fittings rated for pressure equal to or
greater than the maximum pressure PPCH will be used to generate.
PPCH is frequently used as the automated pressure generation and control
component in a PG7302 oil operated piston gauge system. To do so, it must be
connected hydraulically and by RS-232 to the PG7302.
The PG7302 puts the PPCH into remote mode and manages its operation.
To connect a PPCH to a PG7302 to be used as an automated pressure
generation and control component in the PG7302 system, proceed as follows:
Using tubing of appropriate pressure rating and a tee fitting, connect the
PG7302 TEST port to the PPCH TEST port (see Figure 6 for instructions on
making this connection using the standard interconnections
accessories
supplied with the PPCH). The third leg of the tee is for the connection to the
device or system under test.
Using a standard pin-to-pin DB-9M to DB-9F RS232 cable, connect COM1 of
the PPCH to COM3 of the PG7302.
See the PG7000 Operation and Maintenance Manual for information on
operating the PG7302 with an automated pressure controlling device.
For PPCH to be identified by a PG7302 platform, the PPCH-COM1 port settings
must be the same as the PG7302 COM3 settings (baud rate, parity, data bits, stop
bits). If the PG7000 is not able to establish communications with PPCH, check that
the PPCH COM1 port settings are the same as the PG7302 COM3 settings. If the
COM port settings are correct, check that the correct communications cable is being
used (standard pin-to-pin DB-9M to DB-9F RS232) and is connected to the correct
communications ports.
See the PG7302 Operation and Maintenance Manual for additional information on
PG7302 RS232 communications and COM port settings.
2.3.9.4CONNECTING TO A DEVICE UNDER TEST (TEST PORT)
ALWAYS use external tubing and fittings rated for pressure equal to or
greater than the maximum pressure PPCH will be used to generate.
Using a pressure connecting hose or tube of appropriate pressure rating,
connect the device or system to be tested to the PPCH TEST port. The PPCH TEST
connection is DH500 female.
DH500 is a gland and collar type fitting for 1/4 in. (6.35 mm) coned and left
hand threaded tube. DH500 is equivalent to AE F250C, HIP HF4, etc.
Minimizing the length of the test connection tubing will enhance control
performance and reduce pressure setting time. For normal operation, the optimum
total volume of the device or system under test including connecting tubing is 50 cc
3
) and should be less than 100 cc (6 in3).
(3 in
PPCH pressure control will not operate properly if there are excessive leaks in
the system connected to its TEST port. In general, the maximum acceptable leak
rate for optimal PPCH automated pressure control operation and to assure in
tolerance measurements with default pressure control parameters is 0.005 % of
reading/second. In DYNAMIC CONTROL mode, to handle higher test system leak
rates, increase the hold limit using CUSTOM CONTROL (see Section 3.4.6.1) or use HI
VOLUME DYNAMIC CONTROL (see Section 3.2.2.2).
When the PPCH is vented ([VENT] LED ON), the system connected to the TEST
port is connected to the reservoir. If there is an open point in the system below the
fluid level in the reservoir, liquid will run out of the reservoir through the open point.
2.3.10 CHECK/SET SECURITY LEVEL
PPCH has a security system based on user levels. By default, the security system is set to
“low”, which includes certain access restrictions, and there is no password required to
change the security level. See Section 3.5.5.5 for information on the security level system.
As part of the PPCH startup, determine the security level that is appropriate for the PPCH
and set a password if desired.
PPCH is delivered with the security level set to “low” to avoid inadvertent altering of critical
internal settings but with access to changing security levels unrestricted. It is recommended that
the low security level be maintained at all times and password protection be implemented if
control over setting of security levels is desired.
PPCH Q-RPTs can operate in both absolute and gauge measurement modes (see Section
3.3.3). If the PPCH will only be used in gauge mode, absolute measurement mode may be
turned off for its Q-RPT(s) so it is no
accidental use of the wrong measurement mode. See Section 5.3.5 for complete information
on turning off absolute measurement mode.
longer accessible. This can avoid confusion and/or
2.4 POWER-UP AND VERIFICATION
2.4.1 SWITCH POWER ON
Actuate the power switch on the PPCH rear panel. Observe the front panel display as PPCH
initializes, error checks and goes to the MAIN RUN screen (see Section 3.1.1).
PPCH
pressure measured by the Hi Q-RPT is more than 100 kPa (15 psi) away from standard
atmospheric pressure.
If the PPCH fails to reach the MAIN RUN screen, service is required. Record the sequence
of operations and displays observed. Contact a DHI Authorized Service Provider (see Table 31).
power-up condition is Internal, Hi Q-RPT or utility sensor active, VENT ON unless the
2.4.2 APPLY PNEUMATIC POWER (DRIVE AIR)
Apply pneumatic power of the appropriate pressure and flow rate to the PPCH DRIVE port
(see Section 2.3.5). It is normal for the hydropneumatic
drive air is applied.
Check that the PPCH operates properly in absolute mode.
If the PPCH is not vented (VENT LED ON), press the [VENT] direct pressure
control key (see Section 3.1.3).
Use the [RANGE]
Q-RPT’s DF range (see Section 3.3.1). Press the [MODE] function key and
select
<absolute> (see Section 3.3.3). Use [UNIT] to change the pressure unit if
desired (see Section 3.3.2).
function key to change ranges if necessary and select the
Observe the current value of atmospheric pressure.
with the local value of atmospheric pressure. Repeat this process for all the QRPTs and or the utility sensor in the PPCH system. Check that the values of
atmospheric pressure measured by the different devices agree with each other
within PPCH or RPM4 measurement tolerances as applicable (see Section
1.2.2.1, 1.2.2.2). Remember to consider that head pressures can be
with liquids (see Section 3.3.7). If the Q-RPTs do not agree within tolerances,
the PPCH or RPM4 Q-RPT may need calibration or repair.
If the PPCH is not vented (VENT LED ON), press the [VENT] direct pressure
control key to vent it (see Section 3.1.3).
Use the [RANGE]
function key to change ranges if necessary and select the
Q-RPT’s DF range (see Section 3.3.1). Press the [MODE] function key and
select
<gauge> mode (see Section 3.3.3). Use [UNIT]
to change the pressure
unit if desired (see Section 3.3.2).
t is normal for PPCH to indicate a value other than zero for up to thirty seconds
I
when first entering gauge mode. After ten to thirty seconds, the VENT LED
should flash and zero should be indicated. If this does not occur, check that the
AUTOZERO function is ON (see Section 3.5.1). If AUTOZERO is ON and the
displayed
pressure will not zero when vented in gauge or bi-directional gauge
measurement mode, PPCH may need repair.
If a non-zero head correction value is active, PPCH will not indicate zero when
vented in gauge mode. Use [HEAD] to set the head height to zero (see Section 3.3.7).
Using the [RANGE] function key to change ranges, observe that zero is
indicated for each Q-RPT within 10 to 30 seconds.
2.4.4 PRIME AND PURGE TEST SYSTEM
For PPCH pressure control to operate properly, it is important to minimize the amount of gas
present in the system connected to the TEST port. The prime and purge functions are
intended to assist in removing gas from the test system before starting a test.
Prior to controlling pressure with PPCH, it is recommended that a priming and/or purging
procedure be completed (see Section 3.3.8).
2.4.5 LEAK CHECK
PPCH cannot control pressure properly if the system connected to the TEST port has excessive
leaks. See Section 3.3.9 or information on the PPCH leak
test function.
For PPCH to operate properly, the leak rate in the test system after stabilization should not
exceed 0.005% of reading/second.
Verify the maximum pressure rating of the system connected to the PPCH TEST port before
entering a purge pressure value. Do not enter a purge pressure greater than the pressure rating
of the system or instruments connected to the PPCH TEST port.
2.4.6 CHECK PRESSURE CONTROL OPERATION
Select a pressure range using [RANGE] (see Section 3.3.1).
Press [SETUP], <6control>, <2mode>, <1dynamic>, <1standard>) (see Section 3.4.6). If
PCH was not already set to dynamic control mode, wait 5 minutes for the thermal pressure
controller to stabilize.
P
ress [ENT]. Key in a target pressure within the active range and press [ENT] again
Verify the maximum pressure rating of the system connected to the PPCH TEST port before
entering a target pressure. Do not enter a target pressure greater than the pressure rating of
the system or instruments connected to the PPCH TEST port.
PPCH should set the target pressure and indicate Ready (see Section 3.2.3) continuously in
60 to 120 seconds. If it does not, see Section 6 to troubleshoot.
2.5 STORAGE AND SHIPPING
2.5.1 SHORT TERM STORAGE
The following procedure is recommended for short term storage of PPCH:
Vent the PPCH test pressure.
If the PPCH TEST port is open to a point lower than the reservoir, when PPCH is vented
([VENT] LED ON), the liquid contained in the reservoir may leak out through the opening.
Turn the power OFF using the rear panel power switch. Shut OFF or disconnect the drive air pressure supply.
2.5.2LONG TERM STORAGE
The following procedure is recommended for long term storage of PPCH:
Complete the short term storage procedure (see Section 2.5.1).
Plug the TEST port with a DH500 steel plug. Set the RESERVOIR FILL/DRAIN valve to CLOSED and the RESERVOIR RETURN
valve to INTERNAL.
2.5.3PREPARATION FOR SHIPPING
The following procedure is recommended to prepare PPCH for shipment:
Complete the long term storage procedure (see Section 2.5.2).
Remove the internal reservoir accessories if they have been installed (see Section 2.3.6.1). Pack PPCH carefully being sure to use packing materials appropriate for shipment of a
delicate instrument weighing approximately 50 kg (110 lb.). If available, use the original
packing materials.