RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
High pressure liquids and ga ses a re po tentially haza rdous. Energy stored in these liquids and gas es can
be released unexpectedly and with extreme force. High pressure systems should be assembled and
operated only by p ers o nnel who have b een instructed in prop er s a fety pra ctices.
transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express written
permission of DH Instruments, I nc. 4765 East Beaut iful Lane Phoenix AZ 85044-5318 USA.
DH Instruments makes sincere efforts to ensure accuracy and quality of its published materials; however, no
warranty, expressed or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or
indirect damages resulting from t he use of the information in this manual or product s descr ibed in it. M ent ion of any
product does not constitute an endorsement by DH Instruments of that product. This manual was originally
composed in English and was subsequently translated into other languages. The fidelity of the t ranslation cannot
be guaranteed. In case of conflict between the English version and other language versions, t he English version
predominates.
DH Instruments, DH, DHI, AutoZ, CalTool, HGC, HPMS, OPG1, PPC, PPC2+, RPM3, SDS (Self Defense System)
are trademarks, r egister ed and ot herwise.
LabVIEW is a registered trademar k of Nat ional Inst r um ent s Corpor at ion.
2.1UNPACKING AND INSPECTION .........................................................................................................5
2.1.1REM OVING FROM PACKAGING.........................................................................................................................5
2.3.1PREPARING F OR OPERATION...........................................................................................................................6
2.3.2RPM3/HPMS FRONT AND REAR PANEL S.........................................................................................................7
2.3.6TEST PORT CONNE CTION................................................................................................................................11
2.3.7SETTING UP FILE SEQUENCES .......................................................................................................................12
2.4POWER UP AND VERIFICATION .....................................................................................................12
2.4.1APPLY POWER ...................................................................................................................................................12
3.1.1KEYPAD LAY OUT AND PROTOCOL................................................................................................................15
3.1.2MAIN RUN SCREE N............................................................................................................................................16
3.2.11 [ENTER] RUN TEST SEQUENCE......................................................................................................................49
3.3[SETUP] MENU KEY.........................................................................................................................54
3.4.5.5RESET - ALL .......................................................................................................................................75
4.3REMOTE COMMAND SYNTAX AND STYLE.....................................................................................83
4.3.1LOCAL AND REM OTE SETTING.......................................................................................................................83
4.3.2COMM AND S YNTAX........................................................................................................................................... 84
4.3.3QUERIES AND REPL IES....................................................................................................................................84
4.3.5COMMAND P ARAMETERS................................................................................................................................85
4.4.2COMMAND DE S CRIPTIONS.............................................................................................................................. 94
5.2.1AUTOZER O OF RPTS.......................................................................................................................................137
5.3.3SET-UP AND PREPARAT ION.......................................................................................................................... 142
5.3.4RPT CALI BRATION USING RPM3 CALTOOL SOFTWARE..........................................................................142
5.3.5EDITING AND V IEWING RPT CALIBRATION INFORMATION...................................................................... 143
5.5.1HPMS OV E RHAUL............................................................................................................................................ 150
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
A
A
This manual provides the user with the basic information necessary to operate an RPM3/HPMS
A30000/A6000-AF Reference Pressure Monitor with High Pressure Mounting System. It also
includes a great deal of addit ional information provided to help you opt imize use of t he instrument
and take full advantage of its many features and functions.
Before using this manual, take a moment to f amiliarize yours elf with the T able of Content s struc ture.
All first time users should read Chapter 2. Chapter 3 provides a comprehensive description of
general RPM3/HPMS operating principles. Section 3.1.7 describes the automated Sequence
function used to run calibrations and verifications of typical ins truments under test . This is the way
the instrument is typically used. Chapter 4 is for remote operation from an external computer.
Chapter 5 provides maintenance and calibration information. Chapter 6 is a quick troubleshooting
guide. Use it to troubleshoot unexpected RPM3/HPMS behavior based on the symptoms of that
behavior. Certain words and expr essions hav e specif ic meaning as t hey pert ain to t his produc t. The
Glossary is usef ul as a quick ref erence f or exact definit ion of specif ic words and expr essions as t hey
are used in this manual.
RPM3/HPMS A30000/A6000-AF is usually delivered as part of an HGC-30000-AF system which
includes an OPG1 hydraulic pressure generator/controller. The OPG1 has its own Operation and
Maintenance M a nual.
O
BBO
UTT
U
T
HIISS
H
T
M
M
N
AAN
UAALL
U
FOR THOSE OF YOU W HO “DO N’T RE AD M ANUAL S”, GO DIRE CTL Y T O SEC TION 2 .3 TO SE T UP YOUR
RPM3/HPMS AND THEN 2.4 FOR POWER UP AND VERIFICATION. THIS WILL GET YOU RUNNING
QUICKLY WITH MINIMAL RISK OF CAUSING D AMAGE TO YOURSE LF OR YOUR NE W RPM3. THE N…
WHEN YOU HAVE QUESTIONS OR START TO WONDER ABOUT ALL THE GREAT FEATURES YOU
MIGHT BE MISSING, GET IN TO THE M ANUAL!
Manual Conventions
(CAUTION) is us ed throughout the m a nua l to identify user w a rnings a nd cautions.
(NOTE) is used throughout the manual to identify operating and applications advice and
additional explanations.
[ ] indicates direct function keys (for example [RANGE]).
< > indicates RPM3 scr een dis play s ( for example <1yes>).
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
1. INTRODUCTION
1.1 PRODUCT OVERVIEW
RPM3/HPMS A30000/A6000-AF is the combination of an RPM3 A5000/A6000-AF reference
pressure monitor and an HPMS High Pressure Mounting System. The combination of an
RPM3/HPMS A30000/A6000-AF and an OPG1-AF oil pressure generator make up the
HGC-30000-AF hydraulic gauge calibrator. The OPG1-AF has it s own Operation and Maintenance
manual.
RPM3 A30000/A6000-AF is a stand-alone, microprocessor driven, reference pressure monitor
intended to accurat ely measure oil pressure in a variety of pressure calibr ation, measurement and
testing applications . It has been designed t o provide very high perfor mance and extensiv e features
combined with maximum v er s atility and ease of use. RPM3 A30000/A6000-AF is a special version of
the standard DHI RPM3 product configured specifically for the USAF HGC-30000-AF hydraulic
gauge calibrator.
The HPMS mounts the RPM3 at a conv enient viewing angle and c ontains t he hardware t o isolat e the
RPM3’s low pressure tr ans ducer (A6000) when the high press ure transducer (A30000) is in use.
RPM3 A30000/A6000-AF uses tw o high accurac y ref erence pres sure t ransduc ers (RP Ts) and an onboard barometer t o m eas ur e pr es s ur e.
RPM3 A30000/A6000-AF is controlled locally by the operator using its front panel display, keypad
and foot pedal or remot ely by a computer using ASCI I character command strings over its RS-232
and IEEE-488 interfaces.
RPM3 A30000/A6000-AF has s ix pressure m easurement r anges from 0 t o 2 000 psi (7 MPa) t o 0 to
30 000 psi (200 MPa) in gauge and absolut e m eas ur em ent modes.
RPM3/HPMS A30000/A6000-AF is usually delivered as part of an HGC-30000-AF system which
includes an OPG1 hydraulic pressure generator/controller. The OPG1 has its own Operation and
Maintenance M a nual.
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
1.2 SPECIFICATIONS
1.2.1 GENERAL SPECIFICATIONS
Power Requirements:85 to 264 VAC, 47 to 440 Hz , 18 VA max consumption
Operating Temperature Range:15 to 45 °C
Storage Temper ature Range:-20 to 70 °C
Weight:RPM3: 3.5 kg ( 7.7 lb)
Dimensions:21.5 cm H x 29 cm W x 38 cm D
Microprocessor:Motorola 68302, 16 MHz
Communication Port s :RS-232 (COM1), RS-232 (COM2), I EEE-488
Pressure Ranges:2 000 psi (14 MPa), 4 000 ps i ( 28 M P a) , 6 000 psi
Operating Medium:Oil (Di-ethyl hexy l s ebac ate)
Pressure Connections:
Test port (RPT > 10 000 psi):DH500 (gland and c ollar type for coned and left hand
Pressure Limits:
Maximum working pressure:
Maximum pressur e w/o damage:
HPMS: 5.8 kg (12.8 lb)
RPM3/HPMS: 9.3 kg (20.5 lb)
(8.5 in. x 11.5 in. x 15.0 in.) approx .
(42 MPa), 10 000 psi ( 70 M P a) , 18 000 psi (126 MPa),
30 000 psi (200 MPa) gauge and absolut e.
threaded tube, equivalent to AE F250C, HIP HF4, etc.)
30 000 psi (214 MPa)
37 000 psi (255 MPa)
1.2.2 PRESSURE MEASUREMENT SPECI FICATIONS
RPM3 A30000/A6000-AF is configured with two reference pressure transducers (RPT).
Each transducer has three ranges. The RPTs are of the absolute pressure type with an
evacuated, permanently sealed reference. Absolute RPTs measure both absolute and
gauge pressure. Gauge pressures are defined by offsetting atmospheric pressure
dynamically with compensation for atmos pher ic c hanges us ing an on board barometer.
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
1.2.2.1 RPT MEASUREMENT SPECIFICATIONS (% FS OF ACTIVE RANGE)
Transducer Type:Oscillating quart z resonator
Warm Up Time:30 minutes for operation within s pec ifications
Resolution:To 1 ppm, user settable by individual range
Overpressur e Lim its:
Without effect on calibration:
Without perm anent damage:
Temperature E ffect:± 0.008 % max imum temper ature eff ect in normal
1Precision: Combined linearity, hysteresis, r epeatability of measurement s made by the
reference pressure transducer. When using an absolute RPT for gauge mode
measurement add
account the resolution and short term stability of the on-board barometer used for
dynamic atmospheric pressure compensation.
2Stability: Maximum change in zero and span over 180 days for t ypical tr ansducer us ed
under typical conditions. As stability can only be predicted and varies from transducer to
transducer, stability for a specif ic RPT should be established from exper ience.
3Measurement Uncertainty: Combined precision, st ability, temperature effect.
± 2.5 Pa (0.00035 psi) to the precision specification to take into
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
1.2.2.2 ON-BOARD BAROMETER
Sensor Technology:Micro-machined silicon
Warm Up Time:None required
Resolution:1.25 Pa (0.00018 psi)
The on-board barom eter is NOT used as a so urce of absolute accuracy. It is used
only to measure changes in atmospheric pressure for dynamic compensation of the
atmospheric pressure offset when using an absolute reference pressure transducer
to make gauge pressure measurements (see Section 3.4.1, Gauge Mode with an
Absolute RPT, Dynamic Compensation for Atmospheric Pressure).
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
2. INSTALLATION
2.1 UNPACKING AND INSPECTION
RPM3/HPMS A30000/A6000-AF is usually delivered as part of an HGC-30000-AF system which
includes an OPG1 hydraulic pressure generator/controller. The OPG1-3000-AF is shipped in a
separate corrugated conta ine r a nd ha s its O p er a tio n a nd Maintenance Manual.
2.1.1 REMOVING FROM PACKAGING
RPM3/HPMS A30000/A6000-AF is delivered, along with its accessories, in a reusable
molded plastic, shipping c ontainer with polyurethane ins er ts to hold it in place.
Remove the RPM3/H P M S A 30000/A6000-AF and its ac c es s ories from the shipping container
and remove each element from its protec tive plastic bag.
Retain the shipping conta iner for repacking the RPM3/HPMS A3 0000/A6000-AF when it is
shipped for recalib ration of repair.
2.1.2 INSPECTING CONTENTS
Check that all item s are present and have NO vis ible dam age.
A standard RPM3/ HP M S A 30000/A6000-AF includes all items listed in Table 1.
Table 1. RPM3/HPMS A30000/A6000-AF Supplied Items
DESCRIPTIONPART NO.
RPM3/HPMS A30000/A6000-AF Reference Pressure Monitor/ High Pressure
Mounting System
Transport case (with inserts) ( used f or or iginal shipment)123122
Foot Switch Assembly401613
Operation and Maintenance Manual, RPM3/HPMS A30000/A6000550115
Calibration Certificate550100
Test Report, HGC-30000-AF ( if RPM3/ HPMS delivered as par t of HG C-30000- AF
system)
Accessories:401605
Power Cord (7.5 ft.)100770
General Accessories Disk (Important: I ncludes syst em suppor t sof t war e and
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
2.2 SITE REQUIREMENTS
The RPM3/HPMS A30000/A6000-AF is typically delivered as part of an HGC-30000-AF hydraulic
gauge calibrator that includes an OPG1-30000-AF that has its own Operation and
Maintenance Manual. See the OPG1-30000- AF Operation and Maint enance Manual for infor mation
on site requirement s for the HGC-30000-AF.
Install RPM3/HPM S A30000/A6000-AF on any stable surface at a convenient height. Consider t he
placement of the FOOT SWITCH which may need to be accessed frequently while
running calibrations.
Support facilities required f or RPM3/HPMS A30000/A6000- AF include an electrical power sour ce of
85 to 264 VAC, 47 to 440 Hz .
2.3 INITIAL SETUP
2.3.1 PREPARING FOR OPERATION
To prepare RPM3/ HP M S A 30000/A6000-AF for c hec k out and operation:
• Remove the plastic plug from the RPM3/HPMS A30000/A6000-AF rear panel
TEST connection.
• Remove the protective plastic sheet from the RPM3 f r ont panel display.
• Familiarize yourself briefly with the RPM3 and HPMS f ront and rear panels ( see Sect ions
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
2.3.3.2 REAR PANEL
1
6
5
1.Accessories Connection Cable (connects to HPMS 4.ATM Post (On-Board Barometer)
Rear panel FOOT SWITCH and HPMS LEDs)(10-32 UNF)
2.Electrical Power Connector (IEC320-C13)5.Label, Product (on bottom of case)
3.LO TEST Port (A5000 RPT) (1/8 in. NPT F)6.HI TEST Port (A30000 RPT) (DH500 F)
4
3
2
Figure 5. RPM3 Rear Panel
2.3.4 POWER CONNECTION
• Connect the power cable t o the RPM3 rear panel electr ic al power connector.
• Do NOT connect the other end of the power c able to a power source yet.
RPM3 is always powered and active when power is supplied through the rear panel
power co nne ctor. The front p a nel O N/OFF key co ntrols a soft O N /O F F (s e e Se ctio n 3.1.4).
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
2.3.5 FOOT SWITCH CONNECTION
Connect the foot switch supplied in the RPM3/HPMS ac cessories to the HPMS rear panel
electrical connection labeled FOOT SWITCH. Place the foot switch on the floor at a
convenient location. The FOOT SWITCH function is equivalent to the RPM3 front panel
[ENTER] key and may be used fr equently while running tests and c alibr ations.
2.3.6 TEST PORT CONNECTION
A single high pressure TEST port is prov ided on the rear panel of the HPM S .
The test port connection is a DH500 F. DH500 is a gland and collar ty pe fitting f or 1/4 in.
(6 mm) coned and left hand threaded tube. DH500 is equivalent to AE F250C, HIP HF4, etc.
Connecti on to OP G1 if the RPM3/HPMS is Part of an HGC-30000-AF S yst em
If the RPM3/HPMS was delivered as part of an HG C-30000-AF Hydraulic G auge Calibrator,
an OPG1 Oil Pressure Generator/Controller was delivered with it. The OPG1 includes a
fittings kit with the necessary fittings to interconnect the RPM3/HPMS and OPG1 and
instructions on making the connection. S ee the OPG1-30000-A F manual.
USE THE CORRECT PRESSURE CONNECTORS: The RPM3/HPMS TEST port fitting is a
DH500 F (see Section 1.2.1). It is NOT a 1/8 in. NPT F . Never use fittings other than the
correspo nding male fittings in these connec tors. D amage to the connecto rs and d angerous
failure under pressure co uld res ult from using incorrect fittings.
DO NOT AP PLY PRESSURE UN TIL YOU ARE FAM ILIAR WITH OPERATION : The RPM3/HPMS
rear panel test port connects internally to both the RPM3’s 6 000 psi (40 MPa) and
30 000 psi (200 M Pa)RPTs. T he valve on the front o f the HPMS isolate s the low press ure
RPT when the high pressure RPT is in use. Do not app ly pressure to the RPM3/HPMS until
you are familiar with its operation and know how to protect the low pressure RPT from
overpressure (see Section 3.1.8). FAILURE TO PROTECT THE LOW PRESSURE RPT FROM
OVERPRESSURE MAY DESTROY IT. DAMAGE D UE T O RP T OV ERPRESSURE IS NO T COVERE D
BY THE PRODUCT WARRANTY.
2.3.6.1 THE RPM3 AT M PORT
The ATM pass through on the RPM3 rear panel is connected to the RPM3’s onboard barometer. This connection assures that the on-board barometer measures
ambient atmospher ic pressure rather than the press ure inside the RPM3 case that
may vary slightly from ambient pressure. The ATM port should be left open
and unobstructed.
NEVE R plug or obs truct the ATM pas s through as this ma y adverse ly affect gauge
mode operation and autozeroing on an absolute transducer.
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
2.3.7 SETTING UP FILE SEQUENCES
RPM3 supports automated test/calibration sequences. Sequence parameters for testing
specific DUTs can be stored in Sequence Files and recalled when a Sequence is run.
Consider setting up Sequence Files for frequently t ested DUTs as part of t he RPM3 set up
process (see Section 3.3.6).
2.4 POWER UP AND VERIFICATION
2.4.1 APPLY POWER
Connect the RPM3 power cable to an electric supply of 85 to 264 VAC (47 to 440 Hz).
Observe the front panel display as RPM3 initializes, error checks and goes to the main run
screen (see Sect ion 3.1. 2). Check t hat one of the t wo green Valve St atus LEDs on t he front
panel of the HPMS is lit (the red LED should NO T be ON).
RPM3 is always powered and active when power is supplied through the rear panel
power connector. The front panel ON/OFF key controls a soft ON/OFF function (see
Section 3.1. 4).
If the RPM3 fails to reac h the main run sc reen, servic e is required. Recor d the sequence of
operation and displays observed and contact a DHI Authorized Service Center (see
Section 7.2, Table 18).
If neither of the green LEDs on the HPMS front panel lights, check that the 9-pin D-Sub
accessory cable at the rear of the RPM3 itself is properly connected to the ACC. J1
connector.
The active range o n power up is the sam e as the range that was active at the last power
down (see Section 3.1.6).
2.4.2 CHECK PROPER PRESSURE MEASUREMENT OPERATION
2.4.2.1 CHECKING ABSO L UT E MODE PRESSURE MEASUREMENT
Check that the RP M 3 RP Ts operate properly in abs olute mode.
Open the RPM3/HP MS TES T port t o atmosphere. Put the valv e knob in the HPMS
front panel into the Lo RPT Active position. This opens the Lo RPT t o the HPMS
TEST port.
Press [MODE] on the RPM3 and select absolute. Change the pressure unit of
measure using [UNIT] if desired (see Sec tion 3.2.3).
Observe the RPM3 indicated pressure. Check that t he value is equal to at mospheric
pressure
[RANGE] and [ENTER] to change ranges. Check that the values of atmospheric
pressure measured by the different ranges agree with each other within RPM3
measurement tolerances (see Section 1.2.2.1). If they do NOT agree within
± 5 psi. Repeat this process for all the ranges on both RPTs using
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
tolerances, RPM3 may need calibration or repair.
2.4.2.2 CHECKING GAUGE MODE PRESSURE MEASUREM ENT
Open the RPM3/HP MS TES T port t o atmosphere. Put the valv e knob in the HPMS
front panel into t he Lo RPT Active position. This opens the Lo RPT t o the TEST
port.
Press [MODE] on the RPM3 and select gauge. Change the pressure unit of
measure using [UNIT] if desired (see Sec tion 3.2.3).
The value indicated should be near z ero (
then press [AutoZ]. This runs AutoZ to zero the range (see S ection 3.2.10). Upon
return to the m ain run screen, observe that the indication of m easured pr essure has
zeroed.
Use [RANGE] and [ENTER] to change ranges (see Section 3.2.2) and repeat the
zeroing process for each range (the two minute wait does not need to be repeated
each time).
If a range fails to zero properly, RP M 3 m ay need r epair .
It is normal for RP M3 to indicate a value o ther than zero when vented when gauge
mode is first entered or ra nges are changed, especially if Auto Z is OFF, RPM3 ha s
been OFF for so me time, its location ha s changed or the fluid head to which it is
connected has changed.
± 5 psi, 35 kPa). W ait two minutes and
2.5 SHORT TERM STORAGE
The following is recomm ended for short ter m s torage of RPM3/HPM S :
• Vent the RPM3/ H P M S TEST port.
• Disconnect the power s upply .
When RPM3 will NOT be us ed for some t ime, it m ay be left powered, but use the soft ON/OFF key
to turn OFF the display.
2.6 LONG TERM STORAGE AND SHIPPING
The following is recommended for long term storage and or shipping of RPM3/HPMS:
• Plug the HPMS TE S T port.
• Place the RPM3/HPMS in a plastic bag.
Place the RPM3/HP M S in the custom shipping/ s torage container in which it was deliv er ed.
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
3. OPERATION
3.1 GENERAL/MANUAL OPERATION
RPM3/HPMS is designed to offer the optimum balance bet ween simple, str aight forward operation
and the availability of a wide v ar iety of funct ions with a high level of operator disc r etion if desired.
The RPM3 provides the local operator interface is through a front panel 2 x 20 character alphanumeric display, a 4 x 4 multi-function k ey pad and an [ENTER] foot switch.
The HPMS mounts the RPM3 at a convenient viewing angle and includes a valve and visual
indicators to isolat e and protect the RP M3’s 6 000 psi RPT (Ranges L1, L2, L3) when using t he 30
000 psi RPT (Ranges H1, H2, H3) (see Section 3.1.6).
3.1.1 KEYPAD LAYOUT AND PROTOCOL
The RPM3 has a 4 x 4 keypad and an [ENTER] foot switc h for loc al operator access to dir ect
functions, function menus and for data entry.
3
1. The Editing and Execution keys are for
execution, suspending execution, backing up
in menus and editing entries.
The Menu/Data keys provide access to
2.
function menus from the main run screen.
The menu name is on the bottom half of
the key. The SETUP menu is for more
frequently used functions (see Section 3.3).
2
1
The SPECIAL menu is for functions that are
NOT generally used as a part of day to day
operation (see Section 3.4). These keys enter
numerical values when editing.
The Function/Data keys allow very
3.
commonly used functions to be accessed
directly from the main run screen by a single
keystroke. The name of the function is on the
bottom half of the key (see Section 3.2.1).
These keys enter numerical values when
editing.
[ENTER] Foot Switch (not shown) acts in exactly the same manner as t he keypad [ENTER]
4.
key. It serves as a remot e [ENTER] key to be used when the operat or’s hands are occupied or
touching the RPM3 front panel is not convenient.
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
Key press confirmation is provided by both tactile and audible feedback. A single tone
confirms a valid entry, a descending t wo note t one signals an invalid ent ry. The audible v alid
entry feedback can be suppressed or modified by pressing [SPECIAL] and selecting
<7Intern>, <2sound> (see S ection 3.4.7.2) .
Pressing the [ENTER] key or foot swit ch generally c auses exec ution or f orwar d movement in
the menu tree.
Pressing the [ESCAPE] key moves back in the menu tree and/or causes execution to c ease
or suspend without changes being implemented. Pressing [ESCAPE] repeatedly event ually
returns to the main run screen. From the main run screen, pressing [ESCAPE] allows
momentary viewing of the RPM3 introduction screen.
Pressing the [+/ -] key changes a numerical sign when editing. It also toggles through
multiple screens when available.
Pressing the [
←←←←] and [→→→→] keys allow reverse and forward cursor movement when editing
data entry. These keys are also used to scroll through choic es .
Some screens go beyond the two lines provided by the display. This is ind icated b y a flashing
arrow in the second line of the displa y. Use [←←←←] and [→→→→] to mo ve the cursor to access the
lines that are NOT visible or directly enter the number of the hidden menu choice if yo u know
it.
3.1.2 MAIN RUN SCREEN
The RPM3 main run scr een is its home display that is reac hed on power up and from which
other functions and menus are accessed. It is the top level of all menu struct ur es .
The main run screen is where the RPM3 is left in normal operation. It display s the current
measured pressure as well as a v ar iety of additional information if desired.
2
1
D DISPLAY FUNCTION
9
4.<M>: Pressure measurement mode: <g> for gauge, <a> for absolute (see Section 3.2.4).
5.<h>: Indicates whether a head correction is applied. <h> if applied, blank if NOT (see Section 3.2.8).
6.<z>: Indicates whether the autozero function is ON or OFF. <z> if ON; blank if OFF (see Section 3.4.1).
7.<RR>: Indicates active RPT (<H> = high, <L> = low) and range (<1> = low, <2> = mid, <3> = hi) (see
Section 3.2.2).
8.<DISPLAY FUNCTION>: Information displayed depends on current display function.
9.<D>: Indication of w hat is being displayed on the bottom line of the display as set by the DISP LAY func tion (see
Section 3.2.7). Choices include:
• <σσσσ>: Current DISPLAY mode is average (see Section 3.2.7.1).
• <R>: Current DI SPLA Y mode is rate (s ee Sec tion 3. 2.7.2); or if <n avg > is in the bottom r ight hand c or ner of
the display, c urrent DISPLAY mode is average and this is the instantaneous reading average screen (s ee
Section 3.2.7.1).
•<H>: Current DISPLAY mode is hi/lo (see Section 3.2.7.5).
4
3
8
5
7
6
1.<•> Ready/Not Ready indication, <•> when Ready,
↑↑↑↑> or <↓↓↓↓> indicating direction of measur ed pressure
<
evolution when Not Ready (see Section 3.1.5).
2.<PRESSURE1>: Numerical value and sign of
pressure meas ur ed by ac tiv e R PT and r ange. S hows
result of last average in average display
mode (see Section 3.2.7.1).
3.<UNIT>: Current unit of measure (see Section 3.2.3).
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
• <D>: Current DISPLAY mode is deviation (see Section 3.2.7.3).
• <✳✳✳✳>, <↑↑↑↑> or <↓↓↓↓>: Current DISPLAY mode is RPT (see Section 3.2.7.4).
• <F>: Current DISPLAY mode is freeze (see Section 3.2.7.6).
• Blank, NO character: Current DISPLAY mode is clean (see Section 3.2.7.7).
RPM3 has a screen saver function that causes the display to dim if NO key is pressed for
10 minutes. Pressing a key restores full power to the display. The screen saver activation
time can be changed or screen sa ving can be co m pletely suppressed (see Sectio n 3.4.7.1).
3.1.3 SOUNDS
RPM3 is equipped with a variable frequency tone device to provide audible feedback
and alarms. The beeper is us ed for the following indications :
•Valid key press: Brief beep. Choice between three frequencies or NO sound is
available (see Section 3. 4.7.2).
• Invalid key press or out of tolerance reading when running an automated
test sequence: Descending two tone blurp (see Section 3.2.11.2).
•In tolerance reading when running an automated test sequence: Ascending three
tone fanfare (see Section 3.2.11.2).
• Leak check rout ine compl et ed: Three two sec ond beeps ( s ee S ec tion 3.3.5).
• UL (upper limit) exceeded: Intermittent one s ec ond beeps (see Section 3.2.5).
• Pmax! (overpressure limit) exceeded: Eight second high frequency beep (see
Section 3.2.5.1).
•Lo RPT is active but a Hi RPT range is selected: Intermittent two second high
frequency beep. T he audible alarm is combined wit h flashing of t he red, Lo RPT Active
LED on the HPMS (see Section 3.2.5. 2) .
3.1.4 SOFT [ON/OFF] KEY
RPM3 is equipped with a soft [ON/OFF] key and indicator LED on the bottom left hand
corner of the front panel. The purpose of the soft ON/OFF key is to put RPM3 into a
dormant mode in which the dis play is turned OFF but power is still s upplied and overpres s ur e
functions are still active. When RPM3 is ON, the ON/OFF indicator is ON continuously.
When RPM3 is soft OFF, the ON/OFF indicator blinks ever y five seconds.
The soft [ON/OFF] key can also be used to reset from an overpressure condition (see
Section 3.2.5.1).
When RPM3 is soft OFF, r ec eiv ing a r em ote command turns it ON.
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
3.1.5 PRESSURE READY <*>/NOT READY (<↑↑↑↑> OR <↓↓↓↓>) INDICATION
The character to the left of the measured pressure on the main run screen provides a
pressure Ready <*>/Not Ready (<
provide the user with a clear and objective indication of when a stable pressure has
been achieved. Ready <*> is indicated when the current stability (rate of change) of
pressure is less t han the stability limit. The user can set the stability limit (see Section 3.3.4) .
The Ready indicat ion is often used when c omparing the RPM3 and a t est dev ice to indicate
when a valid reading can be made.
↓↓↓↓> or <↑↑↑↑>) indication. This indication is intended to
Ready <*>/Not Ready (<
• <*>: Pressure Ready (stable).
• <↓↓↓↓>: Pressure Not Ready (unstable) and dec r eas ing.
• <↑↑↑↑>: Pressure Not Ready (unstable) and inc r eas ing.
↓↓↓↓> or <↑↑↑↑>) character indications are:
3.1.6 MULTIPLE PRESSURE RANGES
RPM3 A30000/A6000-AF has t wo ref erence pres sure t ransducer s (RP Ts) eac h of which has
three ranges for a total of six pressure ranges. This multi-ranging feature allows
measurement uncertaint y to be optimized for the range of pressur e in which you ar e working.
Generally, the bes t range to select ( see Section 3.2.2) is that whose full scale is closest t o,
but greater than, the maximum pressure of the device or system under test.
RPM3 handles all of the dat a operat ions needed to mak e r ange c hanges occur t ranspar ently
to the user when the RA NGE f unct ion is used f or range s election. For a range change t o be
executed, the current pressur e applied to the RPT on which the r ange is being select ed must
be lower than the current upper limit (UL) of that range (see Section 3.2.5). The RPM3
Sequence feature automatically selects the most appropriate range based on the range of
the device under test (see Section 3. 2.11).
When the range is changed, the upper limit automatically changes to the default for that
range or to the last upper limit set for that range. In addition, most other functions and
settings are range specific (see Section 3.2.2).
RPM3 A30000/A6 000- AF ha s s ix ranges. In general, s ettings and operationa l adjustments
are specific to the range currently in use, as if you had six instruments rather than one.
The DISP LAY function, HEAD functions and AUTO RE ADRT function are NOT range specific.
In remote mode, m ost settings are RPT specific rather than range specific (see Section 4.3 ).
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
3.1.6.1 RANGES AND IDENTIFICATION
The active RPT and range is cont inuously indic ated in t he upper right hand cor ner of
the main run screen and m os t other screens.
Hi RPT: The A30000 RPT whose maximum range is 30 000 psi (200 MPa) is
referred to as the Hi RPT.
Lo RPT: The A6000 RPT whose maximum r ange is 6 000 psi (40 MPa) is referred
to as the Lo RPT.
Range 1, 2 or 3: The three ranges of each RPT are r eferred to as 1 = lo range,
2 = mid range, 3 = hi range. The Lo RPT ranges, low to high, are L1, L2, L3.
The high PRT ranges, low to high, are H1, H2, H3.
Table 2. RPM3 Range Identification Summar y
REFERENCE PRESSURE
TRANSDUCER AND RANGEDESIGNATIONDISPLAY SYMBOL *
Lo RPT, Lo rangeLo, 1L12 000 ( 14)
Lo RPT, Mid rangeLo, 2L24 000 (28)
Lo RPT, Hi rangeLo, 3L36 000 ( 42)
Hi RPT, Lo rangeHi, 1H110 000 (70)
Hi RPT, Mid rangeHi, 2H218 000 (125)
Hi RPT, Hi rangeHi, 3H330 000 (200)
* The display symbol is included in the upper, right hand corner of most RPM3 menu displays as
a convenient indicator of active range.
RANGE
PSI (MPa)
3.1.7 AUTOMATED TEST/CALIBRATION SEQUENCES
The RPM3 SEQUE NCE function supports “quick” and “file” automated calibrati on sequences.
These automatically set the appropriate RPM3 range, resolution and stability limit based on
the characteristics of the device under test. They also prompt the user through the
increments of the calibration sequence and log calibrat ion data. The SEQUENCE funct ion
should be used for most common calibration tasks, especially calibration of analog pressure
gauges (see Section 3. 2.11).
RPM3/HPMS A30000/ A6000- AF Operation and Mai ntenance Man ual
3.1.8 HPMS (HIGH PRESSURE MOUNTING SYSTEM)
The HPMS (high pressure mount ing system) holds the RPM3 at a convenient viewing angle
and includes a valve and visual indicat ors to isolate and prot ect the RPM3’s 6 000 psi RPT
(Ranges L1, L2, L3) when us ing the 30 000 psi RPT (Ranges H1, H2, H3).
m
m OPERATION
mm
Numerical references in this Section ref er to Figure 1.
In normal use, t he operator’s only interaction wit h the HPMS is t o operate the isolation
valve knob to connect and disconnect the Lo RPT from the TEST port. The valve’s
operation is prompt ed by the Valve Posit ion LEDs ( 1, 3).
The Lo RPT isolation valve (2) isolates and protects t he Lo RPT from the TES T port
when it is closed (knob f ully CCW) and opens the Lo RPT to the T EST port when it is
open (knob fully CW).
The Valve Position LEDs (1, 3) indicate the position in which the valve knob s hould be
set based on the current range selection on t he RP M 3. If a range on the Hi RPT (H1, H2
or H3) is selected, the LO RPT SHUT OFF LED (1) is lit indicating that t he valve knob
should be turned fully CW to close t he valve, prot ecting the Lo RPT from high pressur e.
If a range on the Lo RPT (L1, L2 or L3) is selected, t he LO RPT ACTIVE LE D (3) is lit
indicating that the valve knob should be turned fully CCW to open t he valve, connect ing
the Lo RPT to t he test pressure.
The HPMS isolatio n va lve (2) must always b e in the clo s ed position (knob fully CCW) when
operating a t p re ss ure grea te r tha n 6 000 psi (40 MPa ).
The CAUTIO N LO RPT ACTIVE LED (4) is used to indicate t hat the Lo RPT is active
and provide a warning when the Lo RPT is active but a Hi RPT range is selected on
the RPM3.
The CAUTION LO RPT ACTIVE LED (4) is driven by the value of pressure currently
measured by the Lo RPT, it is not driven by the valve posit ion as there is no sensing of
the valve position.
The CAUTION LO RPT ACT IVE LED (4) has three possible conditions:
•LED is OFF: The Lo RPT is NOT act ive. The Lo RP T is considered ac tive when it
measures a pressur e gr eater than 30 psig (200 kPa) .
•LED ON continuously: The Lo RPT IS active and a Lo RPT range is selected on
the RPM3. The LED is lit as a reminder that the Lo RPT is act ive and press ure over
its maximum wor k ing pr es s ur e of 6 600 (46 MPa) should not be applied.
• LED is flashing on accompanied by high frequency beeps and <!!!LO RPT
ACTIVE!!!> disp layed by RPM3: Lo RPT IS ac tive and a Hi RP T range is s elected
on RPM3. The Lo RPT should NOT be active when a Hi RPT range is in use.
Hi RPT ranges will overpres s ur e and dam age the Lo RPT.