The Fluke 8845A and 8846A 6.5 digit precision multimeters have the precision and versatility to handle your most
demanding measurements, on the bench or in a system.
These meters are both high performance and feature rich,
yet also remarkably easy to use.
Features at a glance
6.5 digit resolution
•
Basic V dc accuracy of up to 0.0024 % (1 yr.)
•
Dual display
•
100 μA to 10 A current range, with up to 100 pA resolution
•
Wide ohms ranges from 10 W to 1 GW with up to 10 μW resolution
These digital multimeters
perform the functions you would
expect to see in a multifunction
DMM, including measuring
volts, ohms, and amps, with
performance that exceeds expectations. Basic V dc accuracy of
up to 0.0024 %, 10 A current
range, and a wide ohms range
from 10 W to 1 GW with up
to 10 μW resolution give you
an unbeatable combination of
measurement capability.
You can also use the 8845A
and 8846A to measure temperature, capacitance, period, and
frequency—the functions of a
counter, capacitance meter, and
thermometer are built in for
unparalleled versatility. Extend
the meters’ utility even more
with their graphical display
modes, including Trendplot™
paperless recorder mode, statistics and histograms—features you
won’t find on other multimeters.
Of course, these meters are
also durable and dependable,
features you expect from any
Fluke meter. This unique combination of features and performance makes the 8845A and
8846A an unbeatable value for
a wide variety of applications,
including manufacturing test,
research and development, and
service.
Page 2
General Specifications
Power
Voltage
100 V Setting............................................................. 90 V to 110 V
120 V Setting............................................................. 108 V to 132 V
220 V Setting............................................................. 198 V to 242 V
240 V Setting............................................................. 216 V to 264 V
Frequency........................................................................ 47 Hz to 440 Hz. Automatically sensed at power-on.
Power Consumption....................................................... 28 VA peak (12 Watt average)
Dimensions
Height .............................................................................. 88 mm (3.46 in.)
Width ............................................................................... 217 mm (8.56 in.)
Depth................................................................................ 297 mm (11.7 in.)
Weight ............................................................................. 3.6 kg (8.0 Ibs.)
Shipping Weight ............................................................5.0 kg (11.0 lbs)
Display
Vacuum Fluorescent Display, dot matrix
Environment
Temperature
Operating.................................................................... 0 °C to 55 °C
Storage........................................................................ -40 °C to 70 °C
Warm Up..................................................................... 1 hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating.................................................................... 0 °C to 28 °C <90 %
28 °C to 40 °C <80 %
40 °C to 55 °C <50 %
Storage........................................................................ -40 °C to 70 °C <95 %
Vibration and Shock......................................................Complies with Mil-T-28800F Type III, Class 5 (Sine Only)
Safety
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT II 600V
EMC
Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This Meter has shown
susceptibility to radiated frequencies greater than 1 V/m from 250 to 450 MHz in the 100 μA range.
Triggering
Samples per Trigger..................................................1 to 50,000
Trigger Delay..............................................................0 S to 3600 S; in 10 µS increments
External Trigger Delay..............................................<1 mS
Trigger Output............................................................5 V maximum. (Open collector)
Memory
8845A.........................................................................5,000 measurements, Internal only
8846A.........................................................................5,000 measurements, Internal and up to 2 Gigabyte capacity with USB memory module
(available separately. see “Accessories”) through front-panel USB port
Math Functions
Zero, dBm, dB, MX+B, Trend-plot, Histogram, Statistics (min/max/average/standard deviation), and Limit Test
8845A and 8846A Fluke Corporation 1
Page 3
Electrical
Input Protection............................................................1000 V all ranges
Overrange.......................................................................20 % on all ranges except 1000 V dc, 1000 V ac (8846A),
750 V ac (8845A), Diode, and 10 A ranges
Remote Interfaces
RS-232 (RS-232 to USB cable available to connect the Meter to a PC USB port. See accessories)
IEEE 488.2
LAN
Warranty
One year
Electrical Specifications
Accuracy specifications are valid for 6½ digit mode after at least a 1-hour warm-up with Auto Zero enabled.
24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326-1:2000-11
DC Voltage Specifications
Maximum Input.............................................................1000 V on any range
Common Mode Rejection............................................140 dB at 50 or 60 Hz ± 0.1 % (1 kΩ unbalance)
Normal Mode Rejection ..............................................60 dB for NPLC of 1 or greater with dc filter off and power line frequency ±0.1 %
100 dB for NPLC of 1 or greater with dc filter on and power line frequency ±0.1 %
6½ 100 0 % of range
6½ 10 0 % of range
5½ 1 0.001 % of range
5½ .2 0.001 % of range + 20 μV
4½ 0.02 0.01 % of range + 20 μV
AC Voltage Specifications
AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz, add an additional
error of 0.1 % of range, and for 50 to 100 kHz, add 0.13 % of range.
Maximum Input.............................................................750 V rms or 1000 V peak (8845A), 1000 V rms or 1414 V peak (8846A) or 8 x 10
Volts-Hertz product (whichever is less) for any range.
Measurement Method..................................................AC-coupled true-rms. Measures the ac component of input with up to 1000 V dc bias on
Specifications are for 4-wire resistance function, 2 x 4-wire resistance, or 2-wire resistance with zero. If zero is not used, add 0.2 Ω for 2wire resistance plus lead resistance, and add 20 mΩ for 2 x 4-wire resistance function.
Measurement Method..................................................Current source referenced to LO input.
Max. Lead Resistance (4-wire ohms)......................10 % of range per lead for 100 Ω, 1 kΩ ranges. 1 kΩ per lead on all other ranges.
Input Protection............................................................1000 V on all ranges.
6½ 100 0 % of range
6½ 10 0 % of range
5½ 1 0.001 % of range
5½ 0.2 0.001 % of range ± 20 mΩ
4½ 0.02 0.01 % of range ± 20 mΩ
Resolution
1 Year
(23 ± 5 °C)
1 Year
(23 ± 5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
Temperature
Coefficient/ °C
Outside 18 to 28 °C
Source Current
8845A and 8846A Fluke Corporation 5
Page 7
DC Current
Input Protection............................................................Tool-accessible 11 A/1000 V and 440 mA/1000 V fuses.
Input Characteristics
Range
Full Scale
(6½ Digits)
4½ Digits 5½ Digits 6½ Digits
100 µA 100.0000 µA 10 nA 1 nA 100 pA 100Ω <0.015
1 mA 1.000000 mA 100 nA 10 nA 1 nA 100Ω <0.15
10 mA 10.00000 mA 1 μA 100 nA 10 nA 1 Ω <0.025 V
100 mA 100.0000 mA 10 µA 1 μA 100 nA 1 Ω <0.25 V
1 A 1.000000 A 100 µA 10 µA 1 μA 0.01 Ω <0.05 V
[1]
3 A
3.00000A 1 mA 100 µA 10 µA 0.01 Ω <0.15 V
10 A 10.00000 A 1 mA 100 µA 10 µA 0.01 Ω<0.5 V
[1] Part of 10 A range.
Accuracy (8846A)
Accuracy is given as ± (% measurement + % of range)
[1] Part of 10 A range
[2] Available at front panel connectors only
Additional Current Errors
Digits NPLC Additional Noise Error
6½ 100 0 % of range
6½ 10 0 % of range
5½ 1 0.001 % of range
5½ 0.2 0.001 % of range ± 4 μA
4½ 0.02 0.01 % of range ± 4 μA
24 Hour
(23 ± 1 °C)
24 Hour
(23 ± 1 °C)
Resolution
90 Days
(23 ± 5 °C)
90 Days
(23 ± 5 °C)
1 Year
(23 ± 5 °C)
1 Year
(23 ± 5 °C)
Shunt Resistance
(Ohms)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
Temperature
Coefficient/ °C
Outside 18 to 28 °C
Burden Voltage
6 Fluke Corporation 8845A and 8846A
Page 8
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs from 1% to 5 % of
range, add an additional error of 0.1 % of range.
Input Protection............................................................Tool accessible 11 A/1000 V and 440 mA/1000 V fuses.
Measurement Method..................................................ac-coupled true-rms, dc-coupled to the fuse and shunt (no blocking capacitor)
AC Filter Bandwidth
Slow.............................................................................3 Hz to 10 kHz
Medium.......................................................................20 Hz to 10 kHz
Fast..............................................................................200 Hz to 10 kHz
Maximum Crest Factor................................................5:1 at full scale
Additional Crest Factor Errors (<100 Hz) ..............Crest Factor 1-2, 0.05 % of full scale
Input Characteristics
Range
100 μA
1 mA
10 mA 10.00000 mA 1 μA 100 nA 10 nA 1 Ω <0.025 V
100 mA 100.0000 mA 10 µA 1 μA 100 nA 1 Ω <0.25 V
1 A 1.000000 A 100 µA 10 µA 1 μA 0.01 Ω <0.05 V
3 A[2] 3.00000 A 1 mA 100 µA 10 μA 0.01 Ω <0.05 V
10 A 10.00000 A 1 mA 100 µA 10 µA 0.01 Ω <0.5 V
[1] 8846A Only
[2] Part of 10 A range
Full Scale
(6½ Digits)
[1]
100.0000 μA 10 nA 1 nA 100 pA 100Ω <0.015 V
[1]
1.000000 mA 100 nA 10 nA 1 nA 100Ω <0.15 V
4½ Digits 5½ Digits 6½ Digits
8846A Accuracy
Accuracy is given as ± (% measurement + % of range)
Gate Times.....................................................................Programmable to 1 s, 100 ms, and 10 ms.
Measurement Method..................................................Flexible counting technique. AC-coupled input using the ac voltage measurement
Settling Considerations..............................................When measuring frequency or period after a dc offset voltage change, errors may occur.
Measurement Considerations....................................To minimize measurement errors, shield inputs from external noise when measuring low-
[1] Stated accuracy is attained when Zero function is used.
[1]
Temperature Coefficient/ °C
Outside 18 to 28 °C
Temperature (8846A only)
Test Current...................................................................1 mA
Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 ohms lead resistance.
Specifications do not include probe accuracy, which must be added.
Range Resolution
90 Days
(23 ± 5 °C)
1 Year
(23 ± 5 °C)
-200 °C 0.001 °C 0.06 0.09 0.0025
-100 °C 0.001 °C 0.05 0.08 0.002
0 °C 0.001 °C 0.04 0.06 0.002
100 °C 0.001 °C 0.05 0.08 0.002
300 °C 0.001 °C 0.1 0.12 0.002
600 °C 0.001 °C 0.12 0.14 0.002
Accuracy
Temperature
Coefficient/ °C
Outside 18 to 28 °C
8845A and 8846A Fluke Corporation 9
Page 11
Continuity
Continuity Threshold ..................................................Selectable between 1 Ω and 1000 Ω
Test Current...................................................................1 mA
Response Time..............................................................300 samples/sec with audible tone
Accuracy is given as ± (% measurements + % of range)
[1] Typical measurement rates with autozero off.
[2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required.
[3] For remote operation or external trigger using default settling delay
[4] Settling delay = 0
[2]
6½ 100 NPLC1.67 (2) s 0.6 (0.5) 0.6 (0.5)
6½ 10 NPLC 167 (200) ms 6 (5) 6 (5)
5½ 10 NPLC 16.7 (20) ms 60 (50) 60 (50)
5½ 0.2 NPLC3 ms 300 300
6½ 3 Hz 0.14 0.14
6½ 20 Hz 1 1
6½ 200 Hz
6½ 200 Hz
6½ 1 s 1 1
5½ 100 ms 9.8 9.8
4½ 10 ms 80 80
Integration Time
60 Hz (50 Hz)
[3]
1.6 1.6
[4]
6 6
Measurements/Second
8845A 8846A
[1]
10 Fluke Corporation 8845A and 8846A
Page 12
Fluke. Keeping your world
up and running.™
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com