Fluke 8808A SU Data Sheet

Fluke Bench Multimeters
8845A/8846A 6.5 digit multimeters: Precision
multimeter for bench or automated test system applications
8808A 5.5 digit multimeter: Versatile
multimeter for manufacturing, development
and service applications
With 14 measurement functions, math and analysis modes, the 8845A/8846A will meet the needs of any test, research and development, or service application.
Exceed your
expectations for
performance and value
The Fluke 8845A and 8846A 6.5 digit precision
CAT II 600 V multimeters have the precision and versatility to handle your most demanding measurements, on the bench or in a system. These meters are both high performance and feature rich, yet also remarkably easy to use.
These digital multimeters perform the functions you would expect to see in a multifunction DMM, including measuring volts, ohms, and amps. Basic V dc accuracy of up to 0.0024 %, 100 uA to 10 A current ranges, and a wide ohms range from 10 Ohms to 1 GOhm give you an unbeatable combination of measurement capability.
You can also use the 8845A and 8846A to measure tem­perature, capacitance, period, and frequency—the functions of a counter, capacitance meter, and thermometer are built in for unparalleled versatility.
Extend the meters’ utility even more with their graph­ical display modes, including Trendplot mode, statistics and histo­grams—Log readings to a USB flash drive and easily transfer them over to a PC with the 8846A USB device host port —features you won’t find on other multimeters.
Of course, these meters are also durable and depend­able, features you expect from any Fluke meter. This unique combination of features and performance makes the 8845A and 8846A an unbeat­able value for a wide variety of applications, including manufacturing test, research and development, and service.
measurements with
paperless recorder
Handle even the most demanding
high accuracy
and 6.5 digit
resolution.
8845A/8846A features at a glance
6.5 digit resolution
Basic V dc accuracy of
up to 0.0024 %
Dual display
100 μA to 10 A current
range, with up to 100 pA resolution
Wide ohms ranges from
10 Ω to 1 GΩ with up to 10 μΩ resolution
2 x 4 ohms 4-wire
measurement technique
Measures frequency and
period
Measures capacitance and
RTD temperature (8846A)
USB memory drive port
(8846A)
Fluke 45 and Agilent
34401A remote command emulation
Graphical display
Trendplot paperless
recorder mode, statistics, histogram
CAT I 1000 V, CAT II 600 V
Use the built-in
TrendPlot paperless
chart recorder to
graphically identify
the extent of drift
and intermittent
events in analog
circuits.
View results in
Histogram mode to
reveal stability or
noise problems in
analog circuits.
33
A perfect fit for many
applications, in a
system or on the bench
Dual display and versatile graphical capabilities
The 8845A/8846A feature a unique dual display that allows you to measure two different parameters of the same signal from one test connection. Looking at dual interrelated parameters like voltage and current simulta­neously can reveal conditions that might go unnoticed otherwise, greatly simplifying test and troubleshooting.
Versatile solutions for advanced test applications
These meters let you take test and troubleshooting to a new level. Set up the 8845A or 8846A to take measure­ments over a period of time and display them graphically on the DMM’s screen, for real time analysis. To reveal signal quality issues like drift, inter­mittants and stability, view data as a real time trend plot or histogram, with the unique analyze mode. You won’t find graphical display capabilities like these on any other instru­ment in this class.
Graphical trending analysis can highlight drift and signal fluctuations, as well as inter­mittent errors that can’t be seen on a numerical display.
Using the graphical display, the 8845A and 8846A can simultaneously display a his­togram and multiple statistical values such as mean, min, max and standard deviation calculated using real time measurements. Analyzing a measured values distribu­tion over time can highlight potential reoccurring out­of-tolerance problems. The ability to make dual simul­taneous measurements and display them in graphic or numeric format adds new tools to the test system and design engineer’s toolbox of tricks for solving or validating analog circuits.
Save measurement results to USB memory and transfer the data to a PC for detailed analysis. Or use the Ethernet connection to transport data over a network.
Extended current and ohms ranges and additional capabilities such as tempera­ture and capacitance increase the variety of measurements you can make and tests you can perform with a single instrument.
Perfect for bench­top research and development
Research and development applications demand mea­surement performance and flexibility. The 8845A and 8846A meet these needs, with excellent performance in all measurement func­tions. It’s easy to use and adapts easily to almost any benchtop application.
Fluke 45 and Agilent 34401A emulation
The 8845A/8846A emulate programming commands of the Fluke 45 and Agilent 34401A. Emula­tion shortens the learning curve and makes it easy to fit the 8845A or 8846A into existing test systems.
4
Easy to integrate into your automated test system
Front and rear inputs let you easily make connections where it’s most convenient, whether the meter is rack­mounted or used on a bench. Multiple interfaces provide compatibility to existing and new standards.
Reading rates are up to 1000 readings per second, giving you the throughput you
need for systems applications.
Multiple connectors give you maximum flexibility
Choose from several interfaces to connect the 8845A/8846A to a personal computer: serial, IEEE-488, and Ethernet come standard on both models. A USB device port is included on the 8846A meter for conve­nient data transfer to and from a PC via a portable USB storage device.
With 6.5 digit performance, legacy DMM emulation and the latest interfaces, the 8845A/8846A will easily integrate into any legacy or new test system.
5
Perform 4-wire measurements with only two leads
Patented split terminal jacks for the 2x4-wire ohms func­tion allow you to perform 4-wire measurements using only two leads instead of four. Today’s surface mount components make it difficult to make connections. The task becomes even more difficult when you need to use a 4-wire technique for accurate low ohms mea­surements. The Fluke test lead accessory combines the four wires into two test lead pairs, making it easy for you to establish connections. You get excellent resolution and accuracy, plus the con­venience and ease of using a single pair of leads.
Patented split terminal jacks enable 4-wire measurements with two leads.
Input terminals are located on the front and back of the meter to simplify connections within a system.
Fluke support completes the equation of value
Like all Fluke products, the 8845A/8846A and 8808A are built to provide years of dependable service. However, if you do need cal­ibration or repairs, you can turn to our global network of direct and representative service centers strategically located around the world. Each offers a full range of support services.
6
Patented split terminal jacks enable 4-wire measurements with just two leads.
Clear and bright dual display lets you measure two different parameters of the same signal from one test connection.
Limit compare mode compares the current reading to predefined High and Low limits. If an out of tolerance reading occurs the display clearly shows the operator whether a test passes or fails. Used in combination with the front panel setup keys makes manual test processes simple, repeatable and efficient.
Clear and bright dual display shows data in graphic or numeric format.
Context sensitive soft function keys make it easy to navigate through measurements and instrument settings.
USB port for flash memory devices (8846A) provides convenient data storage and transport.
Single button per function for ease of use.
FlukeView® Forms Basic
The Fluke 8845A/8846A and 8808A ship with a free copy of FlukeView Basic. FlukeView Forms increases the power of your Fluke tool by enabling you to document, store and analyze individual readings or series of measurements, then convert them into professional­looking documents. Transfer data points from your meter to your PC and use the two standard, non-customizable forms to display your readings in table or graphical form. For a larger array of forms or to use FlukeView® Forms Designer to customize your forms, upgrade to FlukeView® Forms version
3.0 with FVF-UG.
i-Lkg function adds two additional hi-impedance ranges for sensitive low current measurements
Setup keys make performing your most common measurements as simple as pressing a button. Increase quality and test times by improving test repeatability by simply saving known measurement setups to front panel accessible setup keys. All the operator has to do is select the right setup key.
7
The Fluke 8808A has six dedicated setup buttons on the front panel that function as ‘pre-sets’ similar to those on a car radio, making it fast and easy for production operators to quickly follow a test procedure without having to push multiple buttons. Setup buttons with pass/fail test mode provide measurement consistency that will improve quality and efficiency in manufacturing test applications. With measurement functions including volts, ohms, amps and frequency, the Fluke 8808A also provides the performance and flexibility required from a bench meter in R&D, development and service applications.
8
Making measurements
is as simple as
pushing a button
The Fluke 8808A 5.5 digit multimeter has a broad
range of functions, measuring volts, ohms and amps with a basic V dc accuracy of 0.01 %. It is remarkably easy to use, even by unskilled operators, because it makes the mea­surements you perform most often extremely easy and fast to do.
Six setup buttons on the 8808A front panel operate like a car radio’s station pre­sets. Simply set up the meter for a common measurement, then press shift followed by a setup button (S1 to S6) to save the setup. Now each time you perform that mea­surement, you simply press the appropriate setup key. It’s that easy!
The setup buttons eliminate the need to follow complex work instruction sheets. Operators no longer need to press multiple buttons to set up a measurement function and range, test limits, or enter other paramters to make a measurement.
Eliminate production mistakes
The Fluke 8808A 5.5 digit multimeter dependably performs the most common measurements required by today’s applications. Whether you are performing func­tional tests or making critical measurements on test points, using the limit compare mode with pass/fail indicators eliminates production mis­takes, especially those where results are “on the edge.”
The 8808A display has built-in enunciators that clearly show the operator whether a test passes or fails. The pass/fail indicators take the guesswork out of testing: the result is either within limits or it’s out!
The 8808A can
measure small
leakage currents
with a resolution
of up to 100 nA,
without loading the
circuit under test.
8808A features at a glance
5.5 digit resolution
Basic V dc accuracy
of 0.01 %
Dual display
Dedicated dc leakage
current measurement
2x4 ohms 4-wire
measurement technique
Six dedicated buttons for
fast access to instrument setups
Hi/Lo limit compare for
Pass/Fail testing
Set up six common
measurements via
the front panel
buttons, then
simply press the
appropriate key to
perform each one.
The limit compare
mode with pass/
fail indicators can
help you eliminate
production mistakes.
9
Improve quality
and efficiency in
manufacturing test, R&D
or service applications
Manufacturing test, R&D, development and service applications demand perfor­mance and flexibility from a bench meter. This Fluke 8808A makes measurements that stand up to any scrutiny. It delivers a wide variety of measurement functions, including volts, ohms, and amps, plus frequency—all at superior accuracy and
resolution.
Measure two parameters at once
The 8808A features a unique dual display that allows you to measure two different parameters of the same signal from one test connection. Looking at dual interrelated parameters like voltage and current simultaneously can reveal conditions that might go unnoticed otherwise, greatly simplifying test and troubleshooting.
Measure sensitive leakage current
Using a typical multimeter to perform a sensitive low cur­rent measurement of less than 100 mA can load the circuit under test while the measure­ment is made. This makes it difficult, if not impossible, to perform tasks such as deter­mining the leakage current on a battery-powered device while it is powered down. The Fluke 8808A is the only multi­meter in its class to use a high impedance input technique to perform this type of critical leakage current measure­ment. In this special mode, the 8808A can measure small currents with a resolution of up to 100 nA, without loading the circuit under test.
Fluke quality is built in
Fluke is known around the world for its rugged, reliable, and accurate multimeters. With our experience, design standards and quality assur­ance processes, we have designed a multimeter that will stand up to any scrutiny of its specifications. In fact, a Fluke meter typically per­forms better than specified.
All of our efforts are designed to ensure that you can make an accurate mea­surement—repeatedly, and with confidence.
10
Perform 4-wire measurements with only two leads
Patented split terminal jacks for the 2x4-wire ohms func­tion allow you to perform 4-wire measurements using only two leads instead of four. Today’s surface mount components make it difficult to make connections. The task becomes even more difficult when you need to use a 4-wire technique for accurate low ohms measure­ments. The Fluke test lead accessory combines the four wires into two test lead pairs, making it easy for you to establish connections. You get excellent resolution and accuracy, plus the con­venience and ease of using a single pair of leads.
You can depend on 8808A measurements for critical applications in manufacturing test. But it is also a versatile tool for R&D and design engineers.
11
Summary Specifications
8808A 8845A 8846A
Display VFD multi segment
Resolution 5.5 digits 6.5 digits
Measurement function Accuracy, ± (% of reading + % of range)
V dc ranges 200 mV to 1000 V 100 mV to 1000 V Max. resolution 1 uV 100 nV Accuracy 0.015 + 0.003 0.0035 + 0.0005 0.0024 + 0.0005
V ac
Ranges 200 mV to 750 V 100 mV to 750 V 100 mV to 1000 V Max. resolution 1 uV 100 nV Accuracy 0.2 + 0.05 0.06 + 0.03 0.06 + 0.03 Frequency 20 Hz to 100 KHz 3 Hz to 300 KH z
Resistance
2x4 Wire Yes Yes Yes Ranges 200 Ω to 100 MΩ 10 0 Ω to 100 MΩ 10 Ω to 1 GΩ Max. resolution 1 mΩ 100 μΩ 10 μΩ Accuracy 0.02 + 0.003 0.010 + 0.001 0.010 + 0.001
A dc
Ranges 200 μA to 10 A 100 μA to 10 A 100 μA to 10 A Max. resolution 1 nA 100 pA 100 pA Accuracy 0.02 + 0.005 0.050 + 0.005 0.050 + 0.005
A ac
Ranges 20 mA to 10 A 10 mA to 10 A 100 μA to 10 A Max. resolution 100 uA 10 μA 100 pA Accuracy 0.3 + 0.06 0.10 + 0.04 0.10 + 0.04 Frequency 20 Hz to 2 kHz 3 Hz to 10 kHz 3 Hz to 10 kHz
Freq/Period
Ranges 20 Hz to 1 MHz
Max. Resolution 0.1 mHz 1 μHz 1 μHz Accuracy 0.01 % 0.01 % 0.01 %
Continuity/Diode Test Yes Capacitance
Ranges 1 nF to 0.1 F Max. Resolution 1 pf Accuracy 1 %
Temperature
Type Platinum RTD Range -200 °C to +600 °C Max. resolution 0.01 ° Accuracy 0.06 °
Math functions
Types Zero, Min/Max/Average, Std Dev; mx+b dB/dBm Yes Yes
Advanced functions
Statistics/Histogram No Yes Trendplot No Yes Limit Test Yes Yes
Input output
USB memor y USB memor y drive port Real time clock No No Yes Interfaces RS-232, USB with
Programming Languages/Modes
General
Weight 2.1 kg (4.6 lbs) 3.6 kg (8.0 lbs) Size ( HxWxD) 88 mm x 217 mm x 297 mm (3.46 in x 8.56 in x 11.7 in) Safety Designed to comply
Warranty One year
display
(freq only)
optional adapter Simplified ASCII,
Fluke 45
with EN 61010-1:2001,
ANS I/ISA 61010-1
(S82.02.01):2004, U L
61010-1:2004, CAN/CSA
C22.2 No. 61010.1:2004
3 Hz to 300 kHz 3 Hz to 1 MHz
SCPI (IEEE-488.2), Ag ilent 34401A, Fluke 45
Designed to comply with IEC 61010-1 2000-1,
ANSI/ISA-S82.01-1994, CAN/CSA-C22.2 No.
1010.1-92 CAT I 1000 V, CAT II 600 V
VFD dot matrix
RS 232, IEE-488.2, Ethernet,
USB with optional adaptor
Ordering Information
Models
8845A 6.5 digit precision multimeter, 35 ppm
8846A 6.5 digit precision multimeter, 24 ppm USB mem
8808A 5.5 digit multimeter, .01 %
Options and accessories 8808A/8845A/8846A
T L910 Precision Electronic Probe Set
884X-SHORT 4-Wire Short
TL2X4W-PT II 2x4 Wire Ohms Test
Lead 2 mm Probe Tip
884X-USB USB to RS232 cable adapter
FVF-UG FlukeView Forms Software
Upgrade— No Cable
Y8846S Rack Mount Kit, Single
Y8846D Rack Mount Kit, Dual
8845A/8846A
884X-RTD 100 Ohm RTD Temperature Probe
884X-512M USB Memory 512 M
884X-1G USB Memory 1 GB
Y8022 IEEE488 cable (2 m)
Fluke supports you by supplying information about Fluke tools and how to use them. Visit the Fluke web site at www.fluke.com for product descrip­tions, application notes and white papers. You will also find newsletters, online communities, downloadable manuals, and much more.
Fluke. Keeping your world up and running.
Fluke Corporation PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands
For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
©2007 Fluke Corporation. All rights reserved. Specifications subject to change without notice. Printed in U.S.A. 7/2007 2814536 B-EN-N Rev A Pub_ID: 11262-eng Rev 01
®
Loading...