Fluke 8808A Specifications

Page 1
Fluke 8808A
Digital Multimeter
Extended Specifications
Making measurements is as simple as pushing a button
The Fluke 8808A 5.5 digit multimeter has a broad range of functions, measuring volts, ohms and amps with a basic V dc accuracy of 0.01 %. It is remarkably easy to use, even by unskilled operators, because it makes the measurements you perform most often extremely easy and fast to do.
Six setup buttons on the 8808A front panel operate like a car radio’s station presets. Simply set up the meter for a common measurement, then press shift followed by a setup button (S1 to S6) to save the setup. Now each time you perform that measurement, you simply press the appropriate setup key. It’s that easy!
The setup buttons eliminate the need to follow complex work instruction sheets. Operators no longer need to press multiple buttons to set up a measurement function and range, test limits, or enter other parameters to make a measurement.
Eliminate production mistakes
The Fluke 8808A 5.5 digit multimeter dependably performs the most common measurements required by today’s applications.
Whether you are performing functional tests or making critical measurements on test points, using the limit compare mode with pass/fail indicators eliminates production mistakes, especially those where results are “on the edge.”
Features at a glance
5.5 digit resolution
Basic V dc accuracy of 0.01 %
Dual display
Dedicated dc leakage current measurement
2x4 ohms 4-wire measurement technique
Six dedicated buttons for fast access to instrument setups
Hi.Lo limit compare for Pass/Fail testing
Fluke 45 remote command emulation
The 8808A display has built-in enunciators that clearly show the operator whether a test passes or fails. The pass/fail indicators take the guesswork out of testing: the result is either within limits or it’s out.
Page 2
General Specifications
Voltage
100V Setting ..............................................................90 V to 110 V
1
20V Setting .............................................................108 V to 132 V
2
20V Setting .............................................................198 V to 242 V
2
40V Setting .............................................................216 V to 264 V
F
requency ...................................................................47 Hz to 440 Hz
P
ower Consumption ..................................................15 VA peak (10 W average)
Dimensions
Height..........................................................................88 mm (3.46 in)
W
idth...........................................................................217 mm (8.56 in)
D
epth .......................................................................... 297 mm (11.7 in)
W
eight.........................................................................2.1 kg (4.6 Ib)
Display
Vacuum Fluorescent Display, segment
Environment
Temperature
Operating................................................................0 rC to 50 rC
Storage....................................................................-40 rC to 70 rC
Warm Up.................................................................½ hour to full uncertainty specifications
R
elative Humidity (non-condensing)
Operating................................................................Uncontrolled (< 10qC)
<90 % (10 qC to 28 qC) <75 % (28 qC to 40 qC) <45 % (40 qC to 50 qC)
Storage ...................................................................-40 rC to 70 rC <95 %
A
ltitude
Operating ...............................................................2,000 Meters
S
torage ........................................................................12,000 Meters
V
ibration .....................................................................Complies with MIL-PRF-28800F Class 3
Safety
Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V
EMC
Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
Triggering
Trigger Delay ............................................................400 ms
E
xternal Trigger Delay .............................................<2 ms
E
xternal Trigger Jitter ..............................................<1 ms
T
rigger Input .............................................................TTL Levels
T
rigger Output............................................................5 V max
Math Functions
Min/max, relative, hold, compare and dB functions
Electrical
Input Protection ........................................................1000 V all ranges
O
verrange ...................................................................10 % on the largest ranges of all functions except continuity and diode
test
Remote Interfaces
RS-232C
Warranty
One year
2 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
Page 3
Electrical Specifications
Specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
Maximum Input ..........................................................1000 V on any range
C
ommon Mode Rejection..........................................120 dB at 50 or 60 Hz r0.1 (1 k: unbalance)
N
ormal Mode Rejection.............................................80 dB at Slow Rate
A
/D Nonlinearity ........................................................15 ppm of range
I
nput Bias Current .....................................................<30 pA at 25 rC
S
ettling Considerations.............................................Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(5-1/2 Digits)
S
low Medium Fast
Input Impedance
200 mV 199.999 mV 1 PV 10 PV 10 PV
>10 G:
[1]
2 V 1.99999 V 10 PV 100 PV 100 PV
>10 G:
[1]
20 V 19.9999 V 100 PV 1000 PV 1000 PV
10 M:r1 %
200 V 199.999 V 1 mV 10 mV 10 mV
10 M:r1 %
1000 V 1000.00 V 10 mV 100 mV 100 mV
10 M:r1 %
Notes: [1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M:.
U
ncertainty
[1]
90 days 1 year
Range
2
3 °C r 5 °C 23 °C r 5°C
T
emperature Coefficient/°C
Outside 18 – 28 °C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005 2 V 0.01 + 0.002 0.015 + 0.003
0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004
0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003
0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003
0.0015 + 0.0005
Notes: [1] Uncertainty given as r (% of reading + % of range)
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
Page 4
AC Voltage Specifications
AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz, add an additional error of 0.1 % of range, and for 50 kHz to 100 kHz, add 0.13 % of range.
M
aximum Input ..........................................................750 V rms or 1000 V peak or 8 x 10
7
Volts-Hertz product
M
easurement Method ...............................................AC-coupled true-rms. Measures the ac component of input with up to
1000 V dc bias on any range.
A
C Filter Bandwidth..................................................20 Hz – 100 kHz
C
ommon Mode Rejection..........................................60 dB at 50 Hz or 60 Hz (1 k: unbalance)
M
aximum Crest Factor .............................................3:1 at Full Scale
A
dditional Crest Factor Errors (<100 Hz) ..............Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
O
nly applies for non-sinusoid signals
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(5-1/2 Digits)
S
low Medium Fast
I
nput Impedance
200 mV 199.999 mV 1 PV 10 PV 10 PV 2 V 1.99999 V 10 PV 100 PV 100 PV 20 V 19.9999 V 100 PV 1000 PV 1000 PV
200 V 199.999 V 1 mV 10 mV 10 mV
750 V 750.00 V 10 mV 100 mV 100 mV
1 M: r2 % shunted by
<100 pf
U
ncertainty
[1]
90 days 1 year
Range Frequency
2
3 °C r 5 °C 23 °C r 5 °C
T
emperature
Coefficient/°C
O
utside 18 – 28 °C
200 mV 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005 45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005 20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005 50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01 2 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005 45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005 20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005 50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01 20 V 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005 45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005 20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005 50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01 200 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005 45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005 20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005 50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01 750 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005 45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005 20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005 50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
Notes: [1] Uncertainty given as r (% of reading + % of range)
4 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
Page 5
Resistance
Specifications are for 4-wire resistance function, or 2-wire resistance with REL. If REL is not used, add 0.2 : for 2-wire resistance plus lead resistance.
M
easurement Method ...............................................Current source referenced to LO input
M
ax Lead Resistance (4-wire ohms) ......................10 % of range per lead for 200 :, 2 k: ranges. 1 k: per lead on all
other ranges.
I
nput Protection .........................................................1000 V on all ranges
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(
5-1/2 Digits)
S
low Medium Fast
C
urrent Source
200 : 199.999 : 0.001 : 0.01 : 0.01 :
0.8 mA
2 k: 1.99999 k: 0.01 : 0.1 : 0.1 :
0.8 mA
20 k: 19.9999 k: 0.1 : 1 : 1 :
0.08 mA
200 k: 199.999 k: 1 : 10 : 10 :
0.008 mA
2 M: 1.99999 M: 10 : 100 : 100 :
0.9 PA
20 M: 19.9999 M: 100 : 1 k: 1 k:
0.16 PA
100 M: 100.000 M: 1 k: 10 k: 10 k:
0.16 PA || 10 M:
U
ncertainty
[1]
90 days 1 year
Range
2
3 °C r 5 °C 23 °C r 5 °C
T
emperature Coefficient/°C
Outside 18 – 28 °C
200 : 0.02 + 0.004 0.03 + 0.004
0.003 + 0.0006
2 k: 0.015 + 0.002 0.02 + 0.003
0.003 + 0.0005
20 k: 0.015 + 0.002 0.02 + 0.003
0.003 + 0.0005
200 k: 0.015 + 0.002 0.02 + 0.003
0.003 + 0.0005
2 M: 0.03 + 0.003 0.04 + 0.004
0.004 + 0.0005
20 M: 0.2 + 0.003 0.25 + 0.003
0.01 + 0.0005
100 M: 1.5 + 0.004 1.75 + 0.004
0.2 + 0.0005
Notes: [1] Uncertainty given as r (% of reading + % of range)
DC Current
Input Protection ........................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses.
S
hunt Resistance .......................................................0.01 : for 2 A and 10 A ranges
1 : for 20 mA and 200 mA Burden voltage < 5 mV for 200 PA and 2 mA range.
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(5-1/2 Digits)
S
low Medium Fast
B
urden Voltage
200 PA 199.999 PA 0.001 PA 0.01 PA 0.01 PA
<5 mV
2 mA 1999.99 PA 0.01 PA 0.1 PA 0.1 PA
<5 mV
20 mA 19.9999 mA 0.1 PA 1 PA 1 PA
<0.05 V
200 mA 199.999 mA 1 PA 10 PA 10 PA
<0.5 V
2 A 1.99999 A 10 PA 100 PA 100 PA
<0.1 V
10 A 10.0000 A 100 PA 1 mA 1 mA
<0.5 V
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 5
Page 6
U
ncertainty
[1]
90 days 1 year
Range
2
3 °C r 5 °C 23 °C r 5 °C
T
emperature Coefficient/°C
O
utside 18 – 28 °C
200 PA 0.02 + 0.005 0.03 + 0.005 0.003 + 0.001 2 mA 0.015 + 0.005 0.02 + 0.005 0.002 + 0.001 20 mA 0.03 + 0.02 0.04 + 0.02 0.005 + 0.001 200 mA 0.02 + 0.005 0.03 + 0.008 0.005 + 0.001 2 A 0.05 + 0.02 0.08 + 0.02 0.008 + 0.001 10 A 0.18 + 0.01 0.2 + 0.01 0.008 + 0.001
Notes: [1] Uncertainty given as r (% of reading + % of range)
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs from 1 % to 5 % of range, add an additional error of 0.1 % of range.
I
nput Protection .........................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses
M
easurement Method ...............................................AC-coupled True RMS
S
hunt Resistance .......................................................0.01 : for 2 A and 10 A ranges
1 : for 20 mA and 200 mA
A
C Filter Bandwidth..................................................20 Hz – 100 kHz
M
aximum Crest Factor ..............................................3:1 at Full Scale
A
dditional Crest Factor Errors (<100 Hz) ..............Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale Only applies to non-sinusoid signals
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(5-1/2 Digits)
S
low Medium Fast
B
urden Voltage
20 mA 19.9999 mA 0.1 PA 1 PA 1 PA
<0.05 V
200 mA 199.999 mA 1 PA 10 PA 10 PA
<0.5 V
2 A 1.99999 A 10 PA 100 PA 100 PA
<0.1 V
10 A 10.0000 A 100 PA 1 mA 1 mA
<0.5 V
U
ncertainty
[1]
90 days 1 year
Range Frequency
2
3 °C r 5 °C 23 °C r 5 °C
T
emperature
Coefficient/°C
Outside 18 – 28 °C
20 mA 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005 45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005 200 mA 20 Hz - 45Hz 0.8 + 0.05 1 + 0. 06 0.015 + 0.005 45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005 2 A 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005 45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005 10 A 20 Hz - 45Hz 1 + 0.1 1.25 + 0.12 0.015 + 0.005 45 Hz - 2 kHz 0.35 + 0.1 0.5 + 0.12 0.015 + 0.005
Notes: [1] Uncertainty given as r (% of reading + % of range)
6 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
Page 7
Frequency
Gate Time ...................................................................131 ms
M
easurement Method ...............................................AC-coupled input using the ac voltage measurement function.
S
ettling Considerations.............................................When measuring frequency after a dc offset voltage change, errors
may occur. For the most accurate measurement, wait up to 1 second to allow input blocking RC time constant to settle.
M
easurement Considerations ..................................To minimize measurement errors, shield inputs from external noise
when measuring low voltage, low frequency signals.
U
ncertainty
90 days 1 year
Range Frequency
2
3 °C r 5 °C 23 °C r 5 °C
T
emperature
C
oefficient/°C
O
utside 18 – 28 °C
20 Hz – 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz – 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz – 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
100 mV to 750 V
[1,2]
200 kHz – 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
Notes: [1] Input > 100 mV [2] Limited to 8* 10
7
V Hz
Continuity
Continuity Threshold ................................................20 :
T
est Currents ..............................................................1 mA
R
esponse Time .........................................................100 samples/sec with audible tone
R
ate..............................................................................Fast
M
aximum Reading.....................................................199.99 :
R
esolution...................................................................0.01 :
Diode Test
Response Time...........................................................100 samples/sec with audible tone
R
ate..............................................................................Fast
M
aximum Reading.....................................................1.9999 V
R
esolution...................................................................0.1 mV
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 7
Page 8
Ordering information
Models Description
8808A 120V 5.5 Digit Multimeter 8808A 220V 5.5 Digit Multimeter 8808A 100V 5.5 Digit Multimeter 8808A 240V 5.5 Digit Multimeter
8808A/SU includes
8808A package plus, FlukeView Forms basic software, USB to RS-232 interface adapter cable.
8808A/SU 120V 5.5 Digit Multimeter, SW USB Cable Kit 8808A/SU 220V 5.5 Digit Multimeter, SW USB Cable Kit 8808A/SU 100V 5.5 Digit Multimeter, SW USB Cable Kit 8808A/SU 240V 5.5 Digit Multimeter, SW USB Cable Kit
8808A includes
Meter, TL71 test leads, line cord, spare line fuse, statement of cal prac­tices, WEEE information sheet, Warranty statement, Getting Started guide (English, French, German, Spanish, Italian, Simplified Chinese, Japanese), CD Rom with user manual (English).
8 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
Fluke. Keeping your world up and running.
Fluke Corporation PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands
For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
©2007-2008 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 5/2008 3053965 D-EN-N Rev C Pub_ID: 11288-eng Rev 02
®