Fluke 8808A Specifications

Fluke 8808A
Digital Multimeter
Extended Specifications
Making measurements is as simple as pushing a button
The Fluke 8808A 5.5 digit multimeter has a broad range of functions, measuring volts, ohms and amps with a basic V dc accuracy of 0.01 %. It is remarkably easy to use, even by unskilled operators, because it makes the measurements you perform most often extremely easy and fast to do.
Six setup buttons on the 8808A front panel operate like a car radio’s station presets. Simply set up the meter for a common measurement, then press shift followed by a setup button (S1 to S6) to save the setup. Now each time you perform that measurement, you simply press the appropriate setup key. It’s that easy!
The setup buttons eliminate the need to follow complex work instruction sheets. Operators no longer need to press multiple buttons to set up a measurement function and range, test limits, or enter other parameters to make a measurement.
Eliminate production mistakes
The Fluke 8808A 5.5 digit multimeter dependably performs the most common measurements required by today’s applications.
Whether you are performing functional tests or making critical measurements on test points, using the limit compare mode with pass/fail indicators eliminates production mistakes, especially those where results are “on the edge.”
Features at a glance
5.5 digit resolution
Basic V dc accuracy of 0.01 %
Dual display
Dedicated dc leakage current measurement
2x4 ohms 4-wire measurement technique
Six dedicated buttons for fast access to instrument setups
Hi.Lo limit compare for Pass/Fail testing
Fluke 45 remote command emulation
The 8808A display has built-in enunciators that clearly show the operator whether a test passes or fails. The pass/fail indicators take the guesswork out of testing: the result is either within limits or it’s out.
General Specifications
Voltage
100V Setting ..............................................................90 V to 110 V
1
20V Setting .............................................................108 V to 132 V
2
20V Setting .............................................................198 V to 242 V
2
40V Setting .............................................................216 V to 264 V
F
requency ...................................................................47 Hz to 440 Hz
P
ower Consumption ..................................................15 VA peak (10 W average)
Dimensions
Height..........................................................................88 mm (3.46 in)
W
idth...........................................................................217 mm (8.56 in)
D
epth .......................................................................... 297 mm (11.7 in)
W
eight.........................................................................2.1 kg (4.6 Ib)
Display
Vacuum Fluorescent Display, segment
Environment
Temperature
Operating................................................................0 rC to 50 rC
Storage....................................................................-40 rC to 70 rC
Warm Up.................................................................½ hour to full uncertainty specifications
R
elative Humidity (non-condensing)
Operating................................................................Uncontrolled (< 10qC)
<90 % (10 qC to 28 qC) <75 % (28 qC to 40 qC) <45 % (40 qC to 50 qC)
Storage ...................................................................-40 rC to 70 rC <95 %
A
ltitude
Operating ...............................................................2,000 Meters
S
torage ........................................................................12,000 Meters
V
ibration .....................................................................Complies with MIL-PRF-28800F Class 3
Safety
Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V
EMC
Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
Triggering
Trigger Delay ............................................................400 ms
E
xternal Trigger Delay .............................................<2 ms
E
xternal Trigger Jitter ..............................................<1 ms
T
rigger Input .............................................................TTL Levels
T
rigger Output............................................................5 V max
Math Functions
Min/max, relative, hold, compare and dB functions
Electrical
Input Protection ........................................................1000 V all ranges
O
verrange ...................................................................10 % on the largest ranges of all functions except continuity and diode
test
Remote Interfaces
RS-232C
Warranty
One year
2 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
Electrical Specifications
Specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
Maximum Input ..........................................................1000 V on any range
C
ommon Mode Rejection..........................................120 dB at 50 or 60 Hz r0.1 (1 k: unbalance)
N
ormal Mode Rejection.............................................80 dB at Slow Rate
A
/D Nonlinearity ........................................................15 ppm of range
I
nput Bias Current .....................................................<30 pA at 25 rC
S
ettling Considerations.............................................Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(5-1/2 Digits)
S
low Medium Fast
Input Impedance
200 mV 199.999 mV 1 PV 10 PV 10 PV
>10 G:
[1]
2 V 1.99999 V 10 PV 100 PV 100 PV
>10 G:
[1]
20 V 19.9999 V 100 PV 1000 PV 1000 PV
10 M:r1 %
200 V 199.999 V 1 mV 10 mV 10 mV
10 M:r1 %
1000 V 1000.00 V 10 mV 100 mV 100 mV
10 M:r1 %
Notes: [1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M:.
U
ncertainty
[1]
90 days 1 year
Range
2
3 °C r 5 °C 23 °C r 5°C
T
emperature Coefficient/°C
Outside 18 – 28 °C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005 2 V 0.01 + 0.002 0.015 + 0.003
0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004
0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003
0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003
0.0015 + 0.0005
Notes: [1] Uncertainty given as r (% of reading + % of range)
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3