C.2 Bill of Materials ............................................................................................C-3
ii
Introduction
General Description
1
Section 1
Introduction
1.1 General Description
Containment Monitor 875
The 875 Containment Monitor has been qualified
for use in Nuclear Safety related applications.
Qualification Test Reports, 950.301, 950.308A, and
950.310A define the parameters that have been
verified by test and are available for purchase.
Containment Monitor 875 functions as an accident monitor for reactor containments, refer to Figure 1-1. It
consists of the 877-1 Ion Chamber Detector that is located within the reactor containment, and the 876A-1
Readout Module that is contained in a rack in the control room. The readout module contains a power
supply that provides the necessary power both for itself and the detector. The readout module is
connected to the detector by two cables, a coaxial high voltage cable and a coaxial signal cable.
Specifications for the detector and readout module are contained below.
NOTE
1.2 Specifications
Detector 877-1
Radiation Detected Photons above 60 keV
Range 1 R/h to 1E7 R/h
Energy Response ± 20% from 80 keV to 2 MeV
Chamber Construction Stacked parallel plate, 316 stainless steel housing, 3
terminal, guarded ion chamber
Saturation >80% up to full scale on instrument
Collection Voltage 450 to 550 VDC
Current Output 7E-11 to 7E-4 Amp/R/Hr (nominal)
Chamber Capacitance 435 picofarads (nominal)
Voltage Withstand and Leakage 5E10-5 amperes @ 500 VDC between either ground and
connector pins
Capacitance Approximately 430 pf
1-1
Victoreen 875
Operators Manual
Hermetic Integrity 1E10-5 cm
3
of helium per second @ one standard
atmosphere
Chamber Fill Gas Nitrogen/helium (2%) at atmospheric pressure
Dimensions (W x H x D) with Mounting Bracket 9.25 x 12.5 x 10.06 in (234.95 x 317.5 x 255.52 mm)
Weight Approximately 18 lb (8.17 kg)
Temperature, Storage 40° to 160°F (4° to 71°C)
Temperature, Operating 40° to 160°F (4° to 71°C)
375°F (180°C) peak, accident, 3 hours
Relative Humidity 100% (waterproof)
60
Total Integrated Dose 2E8 Rads + 10% margin,
Chemical Spray 0.45 Gallons/minute/ft
Co
2
(0.28 molar H3BO3, 3000 ppm
Boron, and NaOH; ph of 11, 24 hrs
LOCA Test Duration 28 days
(Consult Fluke Biomedical, Radiation Management Services, for a complete description of Qualification
Test Results)
Readout Module 876A-1
Range 1 to 10
Input Current Minimum: 6.5 to 7.5 x 10
Maximum: 6.5 to 7.5 x 10
Recorder Output0 to +1 VDC, logarithmic (0 - 10 mV and 0 – 5 V
optional)
Computer Output0 to +5 VDC, logarithmic (0
optional)
(Other Outputs Available Upon Request)
System Accuracy (during "all" conditions)Accumulative @ Meter +36% of input radiation Analog Outputs +28% of input radiation
Power Requirements:a) AC Voltage: 108 to 132 VAC, RMS @ 60 ± 3 Hz b) Battery Power: 22 to 32 VDC @ 600 mA DC
maximum
Temperature Coefficient
0.40%/°C, + 0.25 R/h/°C
Dimensions (W x H x D) 8.46 x 5.25 x 15.21 in
(214.88 x 133.35 x 386.33 mm)
7
R/h
-11
A
-4
A
- 100 mV and 0 - 50 mV
Weight 20 lbs (9.07 kg)
Environmental Parameters Temperature (Storage): 40° to 140°F (4° to 60°C)
Temperature (Operating): 40° to 120°F (4° to 49°C)
1-2
Introduction
Specifications
Relative Humidity (Storage): 0 to 95% (noncondensation)
Relative Humidity (Operating): 0 to 90% (noncondensation)
Irradiation: 3.5 to 1 x 10
40 years life.
Mounting 876-1-55, Rack Chassis
Available Options (consult factory) 876A-100: Readout, for use with digital systems
Outputs: 0
- 100 mV
3
Rads @ 60°C integrated over
1
0
0
0
Figure 1-2 illustrates the typical energy response for the detector, Figure 1-3 illustrates typical linearity for
the detector and Figure 1-4 illustrates radiation input versus meter reading.
- 50 mV
- 10 mV
- 5 V
1.3 Equipment Overview
High-Range Containment Monitor Detector (877-1)
The high-range containment monitor detector is an ion chamber detector that has the appearance of a
six-inch diameter domed cylinder about seven inches long. Inside the cylinder are two flange-mounted
electrodes consisting of 31 flat, disk-shaped plates, each about four inches in diameter, stacked, and
mounted on disk rods. The assembly has the appearance of a large air capacitor. The mounting posts are
attached to the mounting flange through insulating spacers, so the flange and housing will be neutral with
respect to the charges applied to the electrodes.
The whole assembly is covered by the six-inch diameter housing which contacts only the neutral
mounting flange. The mounting flange is pierced by three holes. One hole supports the exhaust tube used
for exhausting and back filling the chamber. The other two support two 2-pin connectors, one for each
electrode. One pin in each pair is connected to the neutral mounting flange. When the coaxial signal
cable is connected to this connector, the cable shield is connected to the neutral pin.
The entire chamber is filled with a mixture of helium and nitrogen at atmospheric pressure and sealed.
For further information refer to Section 4, Functional Description.
Readout Module (876A-1)
The 876A-1 Readout Module is designed to give an indication, on an analog meter, of radiation levels
within the containment area. The readout is composed of an analog meter, indicator lights and operating
switches. The meter has a range of 1 to 10
readout is mounted in the 876-1-55 Rack Chassis located in the control room.
For further information refer to Section 4, Functional Description.
7
R/h and is controlled by an eight position rotary switch. The
Rack Chassis (876-1-55)
The 876-1-55 Rack Chassis permits mounting of two readout modules or one readout module and one
optical isolator in a 19-inch wide RETMA equipment rack. The panel height is 5.21 inches.
For further information refer to applicable drawing located in Appendix C.
1-3
Victoreen 875
Operators Manual
Pull Box Assembly (878-12-5)
The 878-12-5 Pull Box Assembly is designed to allow for thermal expansion of the detector cables and to
provide a service loop. Various optional pull-box configurations are available per Table 1-1. For more
information refer to Section 2, Appendix B, and Appendix C.
Table 1-1. Pull-box Variations
Model Inlet, Size and Location
878-12 1" 270° 2, ¾", 90° 180°
878-12-M 1 1" MNPT 270° 2, ¾", 90° 180°
878-12-M2 1" MNPT 270° 2, 1", 90° 180°
878-12-M3 1" 270° 1, 1", 90° 180°
878-12-M4 1" 270° 1, 1", 90° 90°
878-12-M5 ¾" 270° 2, ¾", 90 180°
878-12-M6 2, ¾" 0°/360° 2, ¾", 90 90°
878-12-M7 1" MNPT 270° 2, ¾", 90 90°
878-12-M8 2, ¾", 0° 1,1", MNPT, 90° 90°
878-12-M9 1,1" MNPT 90° 1, ¾" MNPT, 120° 150°
1, ½" MNPT, 210° 210°
878-12-M10 1,1" MNPT 2, ¾ 90° 0°
878-12-M11 2, ¾", 90° 1,1" MNPT 90° 0°
878-12-M12 2, ¾", 270° 1" MNPT, 0° 90°
Outlet, Size and
Location
Inlet/Outlet Orientation
Flexible Tubing (878-12-30TAB)
1.0 inch diameter flexible stainless steel tubing is available in various lengths to interconnect the 877-1
Detector to the first pull-box.
Cables/Connectors/Panel
In-Containment Cable: (878-1-9-TAB)
Ex-Containment Cable: (50-103-TAB)
Replacement Detector Connector Kit: (878-7-5)
Blank Panel: (844-8-5)
Optional Equipment
The following optional equipment is available:
137
878-10 Field Calibrator - 250 mCi
1-4
Cs
Introduction
Equipment Overview
Figure 1-1. High-Range Containment Monitor 875
1
1-5
Victoreen 875
Operators Manual
Figure 1-2. Typical Energy Response Curve – Detector 877-1
1-6
Introduction
Equipment Overview
Figure 1-3. Typical Linearity Detector 877-1
Figure 1-4. Radiation Input vs. Meter Reading (Calculated Curve)
1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and
notify Fluke Biomedical, Radiation Management Services at 440.248.9300.
2. Remove the contents from the packing material.
3. Verify that all items listed on the packing list have been received and are in good condition.
If any of the listed items are missing or damaged,
notify Fluke Biomedical.
1-8
NOTE
Introduction
Storage
1
1.5 Storage
Storage of Victoreen instruments must comply with level B storage requirements as outlined in ANSI
N45.2.2 (1972) Section 6.1.2(.2). The storage area shall comply with ANSI N45.2.2 (1972) Section 6.2
Storage Area, paragraphs 6.2.1 through 6.2.5. Housekeeping shall conform to ANSI N45.2.3 (1972).
Level B components shall be stored within a fire resistant, tear resistant, weather tight enclosure in a wellventilated building.
Storage of Victoreen instruments must comply with the following:
1. Inspection and examination of items in storage must be in accordance with ANSI N45.2.2 (1972)
Section 6.4.1.
2. Requirements for proper storage must be documented and written procedures or instructions must
be established.
3. In the event of fire, post-fire evaluation must be in accordance with ANSI N45.2.2 (1972), Section
6.4.3.
4. Removal of items from storage must be in accordance with ANSI N45.2.2 (1972), Sections 6.5 and
6.6.
1.6 Procedures, Warnings, and Cautions
The equipment described in this manual is intended to be used for the detection and measurement of
ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of
its readings and the appropriate safety procedures to be followed in the presence of radiation.
Although the equipment described in this manual is designed and manufactured in compliance with all
applicable safety standards, certain hazards are inherent in the use of electronic and radiometric
equipment.
WARNINGS and CAUTIONS are presented throughout this document to alert the user to potentially
hazardous situations. A WARNING is a precautionary message preceding an operation that has the
potential to cause personal injury or death. A CAUTION is a precautionary message preceding an
operation that has the potential to cause permanent damage to the equipment and/or loss of data.
Failure to comply with WARNINGS and CAUTIONS is at the user's own risk and is sufficient cause to
terminate the warranty agreement between Fluke Biomedical and the customer.
Adequate warnings are included in this manual and on the product itself to cover hazards that may be
encountered in normal use and servicing of this equipment. No other procedures are warranted by Fluke
Biomedical. It shall be the owner’s or user's responsibility to see to it that the procedures described here
are meticulously followed, and especially that WARNINGS and CAUTIONS are heeded. Failure on the
part of the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical
and its agents from any resulting liability.
Indicated battery and other operational tests must be performed prior to each use to assure that the
instrument is functioning properly. If applicable, failure to conduct periodic performance tests in
accordance with ANSI N323-1978 (R1983) Radiation Protection Instrumentation Test and Calibration,
paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the same
standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by this
reference, a part of this operating procedure.
1-9
Victoreen 875
Operators Manual
(Blank page)
Installation
Installation
Section 2
Installation
2.1 Installation
Installation of the monitoring system consists of selecting suitable mounting sites for each component of
the system, mounting each of the components, and connecting the components into the system
configuration.
Installation of this system is as follows:
2
Refer to the applicable drawings in Appendix C for
further installation Instructions.
Rack Chassis
The 876-1-55 Rack Chassis is a standard 19-inch chassis with a flame barrier. When seismic qualification
is required for the readout, seismic support brackets (P/N 876-1-114) are needed to support the rear of
the rack chassis. The brackets are designed to mount on the 19-inch qualified equipment rack.
Recommended mounting is shown on drawings GEL876-1-55, and 876-1-114.
Readout Module
Readout Module 876A-1 is designed to fit into one-half of an 876-1-55 Rack Chassis. Insert the readout
module in the rack chassis (see drawing GEL876A-1), then insert and tighten the two holding screws in
the rear flanges of the rack chassis. The pawl fastener on the front panel of the readout must be
tightened.
Optical Isolator (No longer manufactured, consult factory)
Detector
The detector case MUST BE PHYSICALLY
GROUNDED TO EARTH GROUND. The readout
instrument circuit common SHOULD NOT be
grounded to earth ground.
The 877-1 Detector is designed to mount on the containment wall. A mounting bracket attached to the
detector has four holes that are used for mounting. Studs must be placed in the containment wall before
mounting the detector. Dimensions for the studs are the same as the dimensions of the holes in the
mounting bracket (refer to drawing GEL877-1). Recommended studs are 5/16 inch Grade 5. No lock
washers are to be used and recommended torque for the nuts is 18 ft. lbs. Orient the detector so that the
cable connectors are on the underside. Attach the detector to the mounting bracket with the four clamps
provided, securing the bolts with a torque of 132 in. lbs. (Figure 1-5 and GEL877-1).
NOTE
NOTE
Pull Box
A cable pull box is required to allow for thermal expansion of the detector cables and to provide a service
loop. The pull box (drawing 878-12-5) is a typical type that mounts to the containment wall. It should be
mounted directly below the detector as shown in Figure 1-1. Depending on actual detector location, more
than one pull box may be necessary.
2-1
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.