Fluke 80T-150UA User Manual

80T-150UA
Universal Temperature Probe
Instruction Sheet
80T-150U
C
TEMPERATURE PROBE
OFF
Introduction
The Model 80T-150UA Universal Temperature Probe is a self-contained temperature-to-voltage converter. The probe is designed to provide a direct temperature reading when it is connected to any high impedance DMM that is capable of 1 mV resolution and at least a 300-count full scale readout capability. Output is 1 mV per degree (Celsius or Fahrenheit). Two switch-selected temperature output scalings are provided: -50 to +150 °C or -58 to +302 °F. The probe will stand off 60 V dc or 30 V ac rms (42.4 V peak).
The unit is housed in two separate but attached assemblies: a temperature probe and a temperature-to-voltage converter. The probe contains the temperature-sensing element and is electrically connected to the temperature-to-voltage converter through a 46-inch cable. A three­position switch on the converter acts as a power switch and is used for selecting Celsius or Fahrenheit scaling for the output. Two banana plugs with standard 0.75-inch spacing are provided for connecting the 80T­150UA to the DMM.
Operating power for the 80T-150UA is derived from a standard 9 V battery. Typically, an alkaline battery provides more than 600 hours of continuous operation before replacement is necessary. An OFF switch is provided on the temperature-to-voltage converter to allow battery conservation when the unit is not in use. In addition, the OFF position of the power switch allows the battery condition to be determined via the external DMM.
Temperature is measured by exposing the probe tip directly to the material to be measured (non-corrosive liquid, gas, or solid). A direct temperature reading is displayed on the DMM.
®
Symbols
XWarning: Risk of electric shock. Refer to the manual for
W Attention: Refer to the manual for information about this
PN 3208738 February 2008
© 2008 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.
information.
feature.
Specifications
The 80T-150UA will achieve rated accuracy when it is used with any
0.25 % DMM that has an input impedance of 1 MΩ.
Electrical
Measurement Range: (See Probe Limitations)
-50 to +150 °C
-58 to +302 °F Accuracy: (Applies for one year after purchase or calibration)
AMBIENT °C ACCURACY +15 to +35 °C ±1 °C from 0 to +100 °C, decreasing
0 to +15 °C and ±2 °C from 0 to +100 °C, +35 to +50 °C decreasing linearly to ±4 °C at -50 and +150 °C
AMBIENT °F ACCURACY +59 to +95 °F ±1.8 °F from +32 to +212 °F,
+32 to +59 °F and ±3.6 °F from +32 to +212 °F, +95 to +122 °F decreasing linearly to ±7.2 °F at -58
Sensitivity (80T-150UA output): 1 mV dc / °C or °F Voltage Standoff: 60 V dc or 30 V ac rms (42.4 V peak) Settling Time: 9.5 seconds to settle within 2 ° for a 50 ° change RF Field: The 80T-150UA product is unspecified in environments with
EMC fields greater than 100 mV/meter. ESD performance is specified as Criteria C per EN61326.
linearly to ±3 °C at -50 and +150 °C
decreasing linearly to ±5.4 °F at -58 and +302 °F
and +302 °F
Environmental
Ambient Operating Range for Temperature Module: 0 to +50 °C (+32 to +122 °F)
Maximum Temperature Probe Body and Cable: +70 °C (+160 °F) See Probe Limitations
Minimum Temperature Probe Body and Cable: Minimum Operating Temperature for Cable: -12 °C (+10 °F)
Minimum Operating Temperature for Probe Body: -40 °C (-40 °F) See Probe Limitations
Storage Temperature for Unit: -40 to +70 °C (-40 to +160 °F) Humidity: 0 % to 90 % (0 °C to +35 °C)
0 % to 70 % (+35 °C to +50 °C) (noncondensing) Altitude: Operating: ≤3048 m ( 10,000 ft)
Storage: ≤15240 m (≤50,000 ft)
Application Force: 15 pounds maximum (probe tip to measured surface)