The 80K-15 Electronic Air Cleaner Probe (the probe or Product) is a low
energy probe that is able to extend the voltage measuring capability of an
ac/dc voltmeter or multimeter.
Properly connected, the probe is rated to 15 kV dc or 10 kV ac rms. The
use of the probe is restricted to low-energy applications that are energylimited circuits such as electrostatic air cleaners.
The probe is a precision 1000:1 voltage divider formed by two matched
resistors. The high input impedance offered by these resistors minimizes
circuit loading and optimizes measurement accuracy.
A Warnin g identifies conditions and procedures that are dangerous to the
user. A Caution identifies conditions and procedures that can cause
damage to the Product or the equipment under test.
XW Warning
To prevent possible electrical shock, fire, or personal injury:
• Read all safety information before you use the Product.
• Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
• Examine the case before you use the Product. Look for
cracks or missing plastic.
• Do not use test leads if they are damaged. Examine the test
leads for damaged insulation and measure a known voltage.
• Do not alter the Product and use only as specified, or the
protection supplied by the Product can be compromised.
• When you do a measurement, never make body contact with
the probe tip or red portion of the Product. Always hold the
Product by its black handle.
• Do not use this probe to measure high voltages on power
distribution systems.
• Connect the protective earth alligator clip lead to a known
good earth ground.
• Do not apply more than the rated voltage between the
terminals or between each terminal and earth ground.
• Have an approved technician repair the Product.
• Do not use the Product if it is altered or damaged.
• Disable the Product if it is damaged.
Symbols
SymbolDefinition
X
W
T
WARNING. HAZARDOUS VOLTAGE. Risk of electric
shock.
WARNING - RISK OF DANGER. Consult user
documentation.
Double Insulated
Earth
PConforms to European Union directives.
)
~
Certified by CSA Group to North American safety
standards.
This product complies with the WEEE Directive marking
requirements. The affixed label indicates that you must
not discard this electrical/electronic product in domestic
household waste. Product Category: With reference to
the equipment types in the WEEE Directive Annex I,
this product is classed as category 9 "Monitoring and
Control Instrumentation" product. Do not dispose of this
product as unsorted municipal waste.
The probe is a passive attenuator. See Table 1. High input impedance,
accuracy, and stability characteristics are achieved with thick-film
resistors. When the probe is connected to a voltmeter with a 10 MΩ input
resistance the probe becomes an accurate 1000:1 divider. This divider
depends on a ground lead to complete the low side of the circuit path. The
connection must always be secure before you do the voltage
measurement. Otherwise, instrument damage or a shock hazard can
result.
Table 1. Schematic Diagram
Probe Tip
RA
RB
LO
HI
10 MΩ
(Voltmeter)
Protective
Earth Lead
Probe
RARB
80K-15999 MΩ 1.11 MΩ
(Dual Banana Plug)
Resistor
Voltmeter Compatibility
Accuracy of the meter is not included in the accuracy of the probe and
must be added to the probe accuracy to determine system accuracy.
The probe is mechanically compatible with any ac or dc voltmeter or
multimeter that accepts a standard double banana connector (spaced
0.75" [19 mm]) with standard plugs (0.160" [4 mm]).
The probe is electrically compatible with any ac or dc voltmeter or
multimeter that has an input impedance of 10 MΩ ±1 %.
Voltmeters or multimeters with other input impedances require the use of
an external shunt or a correction factor to obtain an accurate
measurement. Higher impedance voltmeters or multimeters should be
equipped with a shunt, and lower impedance voltmeters or multimeters
should be assigned correction factors.
Use this formula to determine the value of an external shunt resistor
Rs
Rm x 10
Rm 10=−
Rs
20 10×
20 10–
-------------------
200
10
----------20 MΩ===
Cf
1.11 + Rm
1.11 x Rm
=
Cf
1.11 + 1
1.11 x 1
=
2.11
1.11
= 1.901=
(meter impedance >10 MΩ):
Where:Rs = Shunt resistance in MΩ
Example: If Rm = 20 MΩ,
Use this formula to calculate a correction factor (meter impedance
<10 MΩ):
Where:Cf = Correction factor (multiplier for meter reading)
Example: If Rm = 1 MΩ,
A meter reading of 0.526 volts represents an input of:
0.526 x 1.901 = 1 or 1 kV
Rm = Voltmeter input impedance in ΜΩ (>10 MΩ)
Rm = Voltmeter input impedance in MΩ
Circuit Loading
The probe represents a load to the circuit of 1000 MΩ , or 1 µA/1 kV.
Maintenance
The probe has no user-serviceable parts.
Use a soft cloth dampened with distilled water to clean the probe. Never
use solvents of abrasive cleaners. Make sure the probes are dry before
use.
Performance Test
Verify the accuracy of the probe with a 15 kV dc (±0.25 %) voltage source.
When used with a compatible dc voltmeter, the probe should measure the
source with ±1 % accuracy. No calibration adjustments are available on
the probe.
Voltage Range:0 kV to ±15 kV dc or peak ac, 10 kV rms ac
Input Resistance:1000 MΩ
Division Ratio:1000:1 (1000 x attenuation)
Accuracy dc:
Accuracy ac:±5 % @ 60 Hz, in 20 °C to 30 °C
Safety:IEC 61010-031, Type B, Pollution Degree 2,
Temperature
Operating:0 °C to 50 °C
Storage:-20 °C to +60 °C
Relative Humidity:80 % @ 30 °C
Altitude:2000 m
Dimensions (L x W)36 cm x 7.5 cm (14 in x 3 in)
Weight204 g (7.2 oz)
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship
for one year from the date of purchase. This warranty does not cover
fuses, disposable batteries, or damage from accident, neglect, misuse,
alteration, contamination, or abnormal conditions of operation or handling.
Resellers are not authorized to extend any other warranty on Fluke’s
behalf. To obtain service during the warranty period, contact your nearest
Fluke authorized service center to obtain return authorization information,
then send the product to that Service Center with a description of the
problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE,
ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES
OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some
states or countries do not allow the exclusion or limitation of an implied
warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.